operating instructions precision and analytical balances ms-ts...1 introduction thank you for...

84
MS MS MS MS MS-TS Operating Instructions Precision and Analytical Balances

Upload: others

Post on 25-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

MS

MS

MS

MS

MS-TSO

pera

ting

Inst

ruct

ions Precision and Analytical Balances

Page 2: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 3: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Table of Contents

Introduction1 7

Conventions and symbols used in these operating instructions1.1 7

Safety Information2 8

Definition of signal warnings and symbols2.1 8Product safety information2.2 8

Design and Function3 10

Components3.1 10Keys on the terminal3.2 11User Interface3.3 11Application home screen3.3.1 12Input Dialogs3.3.2 13Entering Text and Numbers3.3.2.1 13Changing time and date3.3.2.2 14Lists and tables3.3.3 15

Installation and Putting into Operation4 16

Scope of delivery4.1 16Installing the components4.2 17Selecting the location4.3 18Connecting instrument4.4 19Leveling the balance4.5 20Levelling the balance with the level indicator4.5.1 20Leveling the balance with the leveling assistant4.5.2 21Transporting the balance4.6 22Weighing below the balance4.7 22General requirements4.8 23Warming up the balance4.8.1 23Adjusting the balance4.8.2 23

Weighing made simple5 24

Switching on the balance5.1 24Switching off the balance5.2 24Main settings and activities at a glance5.3 25Navigation on the touch screen5.4 26Opening an application5.4.1 26Scrolling5.4.2 26Using shortcuts5.4.3 27Performing a simple Weighing5.5 27Switching the weighing unit5.6 28Changing the resolution5.7 30Zero setting5.8 31Taring5.9 31Closing an application5.10 31

Table of Contents 3

Page 4: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Working with IDs5.11 32Autoincrement5.11.1 32

Activities6 34

Activities - Weighing and other applications6.1 34Check Weighing6.1.1 35Totaling6.1.2 39Activities - Adjustments and tests6.2 41FACT Fully automatic adjustment6.2.1 42Internal Adjustment6.2.2 43External Adjustment6.2.3 44

General settings7 45

Quick setting/Preferences7.1 45Date and time7.1.1 45Brightness7.1.2 45Sound7.1.3 45Language7.1.4 46Glove mode7.1.5 46Leveling assistant7.1.6 46General configuration and data7.2 47Weighing options7.2.1 47Publishing7.2.2 48Devices and connectivity7.2.3 49System settings7.2.4 49ISO-log7.2.5 49Access protection7.2.6 49

Communication with Peripheral Devices8 51

Function PC-Direct8.1 51USB Device Interface8.2 52

Firmware (Software) Updates9 54

Operating principle9.1 54Update procedure9.2 54

Maintenance10 56

Cleaning the glass draft shield (0.1 mg and 1 mg models)10.1 56Disposal10.2 57

Interface Specification11 58

RS232C interface11.1 58USB host11.2 58USB device11.3 59MT-SICS interface commands and functions11.4 59

Technical Data12 60

General data12.1 60

Table of Contents4

Page 5: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Model-specific data12.2 61Balances with readability of 0.1 mg, S platform with draft shield12.2.1 61Balances with readability of 1 mg, S platform with draft shield12.2.2 63Balances with readability of 0.01 g, S platform12.2.3 65Balances with readability of 0.1 g to 1 g, S platform12.2.4 68Dimensions12.3 69Balances with readability of 0.1 mg and draft shield12.3.1 69Balances with readability of 1 mg and draft shield12.3.2 70Balances with readability of 0.01 g12.3.3 71Balances with readability of 0.1 g to 1 g12.3.4 72

Accessories and Spare Parts13 73

Accessories13.1 73Spare parts13.2 77

Index 78

Table of Contents 5

Page 6: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 7: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

1 IntroductionThank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of theMS-TS line combine a large number of weighing possibilities with easy operation.

These operating instructions apply to all balance models of the MS-TS line and are based on the ini­tially installed firmware (software) version V 3.20.

u www.mt.com/balances

1.1 Conventions and symbols used in these operating instructions

Key and button designations are indicated by a picture or text in square brackets (e.g.[Weighing]).

These symbols indicate an instruction:prerequisites

1

2 …

steps

results

7Introduction

Page 8: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

2 Safety Information

2.1 Definition of signal warnings and symbolsSafety notes are marked with signal words and warning symbols. These show safety issues and warn­ings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctionsand false results.

WARNING for a hazardous situation with medium risk, possibly resulting insevere injuries or death if not avoided.

CAUTION for a hazardous situation with low risk, resulting in damage to thedevice or the property or in loss of data, or minor or medium injuries ifnot avoided.

Attention (no symbol)for important information about the product.

Note (no symbol)for useful information about the product.

General hazard Electrical shock

2.2 Product safety information

Intended useYour balance is used for weighing. Use the balance exclusively for this purpose. Any other type of useand operation beyond the limits of technical specifications without written consent from Mettler-ToledoAG, is considered as not intended.

It is not permitted to use the instrument in explosive atmosphere of gases, steam, fog,dust and flammable dust (hazardous environments).

General safety informationThis balance complies with current industry standards and the recognized safety regulations; however,it can constitute a hazard in use. Do not open the balance housing: The balance contains no user-ser­viceable parts. In the event of problems, please contact a METTLER TOLEDO representative.

Always operate and use your instrument only in accordance with the instructions contained in thismanual. The instructions for setting up your new instrument must be strictly observed.

If the instrument is not used according to these Operating Instructions, protection of the instru­ment may be impaired and METTLER TOLEDO assumes no liability.

Staff safetyThese operating instructions must be read and understood before using the balance. These operatinginstructions must be retained for future reference.

The balance must not be altered or modified in any way. Only use METTLER TOLEDO original spareparts and accessories.

8 Safety Information

Page 9: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Safety notes

CAUTIONDamage to the balancea) Only use indoors in dry locations.

b) Do not use pointed objects to operate the touch screen!The balance is of a very sturdy design, but is still a precision instrument. Itmust be handled with care.

c) Do not open the balance:The balance contains no user-serviceable parts. In the event of problems,please contact a METTLER TOLEDO representative.

d) Only use METTLER TOLEDO original accessories and peripheral devices forthe balance.These are specifically designed for the balance.

WARNINGRisk of electric shockUse only the original universal AC/DC adapter delivered with your balance, andcheck that the voltage printed on it is the same as your local power supplyvoltage. Only plug the adapter into a socket which is grounded.

9Safety Information

Page 10: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3 Design and Function

3.1 Components

C

0.1 g

10

9

7

1 2

5

0.01 g

9

5

101 2

7

6

MS MS

3

101 2

5

7

4

3

MSMS

0.1 mg

101 2

5

7

46

1 Capacitive color TFT touch screen2 Operation keys3 Glass draft shield4 Handle for operation of the draft shield door5 Weighing pan6 Draft ring7 Level indicator8 Kensington slot for anti-theft purposes9 Safety feet (with 0.01 g and 0.1 g models)10 Leveling foot11 USB Device connection12 RS232C serial interface13 Socket for AC/DC adapter14 USB Host connection

11 12 813

15

14

15 Product label

10 Design and Function

Page 11: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3.2 Keys on the terminal

53 41 2

No. Key Name Explanation1 ON/OFF To switch the instrument on or off.

2 Capacitive color TFT touch screen General navigation3 Tare To tare the scale.

4 Zero To zero the scale.

5 Home To return from any menu level, or other win­dow to the application home screen.

3.3 User InterfaceThe screen is a capacitive color TFT touch screen. The screen not only displays information, it alsoallows you to enter commands by tapping on certain areas on its surface. You can choose the infor­mation displayed on the screen, change terminal settings or perform certain operations on the instru­ment.

Only those elements which are available for the current dialog appear on the display.

CAUTIONDo not touch the touch screen with pointed or sharp objects!This may damage the touch screen.

11Design and Function

Page 12: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3.3.1 Application home screenThe application home screen appears after the startup of the instrument. It is always the last applica­tion that was used before the balance was switched off. The application home screen is the mainscreen of the balance, from where every function can be accessed. You can always return to the appli­cation home screen by tapping on the home button [ ] in the right bottom corner.

Information and work bars

Name Explanation1 Weighing information bar Shows the weighing-in aid and general balance information.2 Work title bar Shows information about the current activity.3 Value bar Shows information about the current weighing process.4 Main navigation Work-related functions.

Information fields

Name Explanation5 Weighing-in aid A dynamic graphic indicator showing the used amount of the

total weighing range.6 Short balance information Readability and maximum capacity of the balance.*7 Weighing value field Shows the value of the current weighing process.8 Coach text field Shows instructions to the current weighing process.

* For approved balances: Min (minimum capacity) and e (Verification scale interval) are shown in theleft upper corner.

Action buttons

Name Explanation9 Application main configura­

tionTo configure the current application (e.g. Weighing).

10 Detailed balance Information Shows detailed technical data of the balance.11 Weighing unit Shows the unit of the current weighing process.12 Activities To open the activities selection.13 Print To produce a print.14 Settings/Preferences To configure the quick settings/preferences.15 Status information field Shows information about the system status.

12 Design and Function

Page 13: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3.3.2 Input Dialogs

3.3.2.1 Entering Text and NumbersThe keyboard dialog allows you to enter characters including letters, numbers and a variety of specialcharacters.

Name Explanation1 Input field Shows the characters that have been entered.2 Switch If a switch appears located on the right side in the content title bar, it must

be switched on for entering a value.3 Discard To discard the entered data and to exit the dialog.4 Delete all To delete every entered character.5 Scroll To navigate when the dialog consists of several steps.6 Delete To delete the last entered character.7 Confirm To confirm the entered data.8 Shift To enter lower or upper case letters.9 Specialized tabs To switch the keyboard mode for entering letters, numbers or special

characters.10 Explanation field Extra information about the value to enter (e.g. the maximum number of

characters).

13Design and Function

Page 14: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3.3.2.2 Changing time and dateThe dialog (Picker view) allows you to set the time/date.

Name Explanation1 Change time/date format Different time/date formats can be chosen.2 Pick button Increment3 Picker field Shows the chosen time/date.4 Pick button Decrement

14 Design and Function

Page 15: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

3.3.3 Lists and tablesBasic elements of a simple list includes a content title and a list of the sub-elements. Tapping on anelement opens a list of sub-elements or an input dialog.

Name Explanation1 List title Title of the current list2 Back button Goes one step back.3 List element title Title of the list element4 List element description Shows the sub-elements.5 Scroll position The list can be scrolled.6 Selection bars Shows bars of the selectable sub-categories.

15Design and Function

Page 16: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4 Installation and Putting into OperationIn this section you will find information how to put your new instrument into operation.

4.1 Scope of deliveryComponents 0.1 mg 1 mg 0.01 g 0.1 g

236 mm – – –Draft shield

168 mm – – –

Ø 90 mm – – –

127 x 127 mm – – –

170 x 200 mm – – –

Weighing pan

190 x 226 mm – – –

Draft shield element – –

Pan support

Bottom plate – –

ESD plate – –

Protective cover

Universal AC/DC adapter (country specific)

AC/DC adapter (without power cable) – –

Operating instructions printed or on CD-ROMdepending on the countryQuick guide

EC declaration of conformity

16 Installation and Putting into Operation

Page 17: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.2 Installing the components

Balances with readability of 0.1 mg and draftshield (236 mm)

Place the following components on the balance inthe specified order:

NotePush the side glass back as far as it will go andgrasp the draft shield (1) with both hands on thebars at the top. 1 Turn draft shield lock (2) to position " "

(unlock).2 Place draft shield on the balance.

3 Turn draft shield lock to " " (lock) and placebottom plate (3).

4 Place draft shield element (4) and weighingpan (6) with pan support (5).

NoteFor cleaning the draft shield see section "Mainte­nance and cleaning".

3

4

5

6

1 22

MS

Balances with readability of 1 mg and draftshield (168 mm)

Place the following components on the balance inthe specified order:

Note: Push the side glass back as far as it will goand grasp the draft shield (1) with both hands onthe bars at the top.1 Turn draft shield lock (2) to position " "

(unlock).2 Place draft shield on the balance.

3 Turn draft shield lock to " " (lock) and placebottom plate (3).

4 Place weighing pan (5) with pan support(4).

NoteFor cleaning the draft shield see section "Mainte­nance and cleaning".

3

4

5

1 2

MS

17Installation and Putting into Operation

Page 18: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Balances with readability of 0.01 g– Place the following components on the bal­

ance in the specified order:

ESD plate (1) Pan support (2) Weighing pan (3) Draft shield element (4)

MS

2

3

4

1

Balances with readability of 0.1 g– Place the following components on the bal­

ance in the specified order:

ESD plate (1) Pan support (2) Weighing pan (3)

MS

3

2

1

4.3 Selecting the locationYour balance is a precision instrument and will thank you for an optimum location with high accuracyand dependability. Select a stable position that is as horizontal as possible. The surface must be ableto safety carry the weight of a fully loaded balance.

Observe ambient conditions (see Technical Data).

Avoid the following:

Direct sunlight Powerful drafts (e.g. from fans or air conditioners) Excessive temperature fluctuations Vibrations

18 Installation and Putting into Operation

Page 19: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.4 Connecting instrument

WARNINGRisk of electric shocka) To connect the balance, only use the supplied three-core power cable with

equipment grounding conductor.

b) Only connect the balance to a three-pin power socket with earthing con­tact.

c) Only standardized extension cable with equipment grounding conductormust be used for operation of the balance.

d) Intentional disconnection of the equipment grounding conductor is forbid­den.

Your balance is supplied with an country-specific AC/DC adapter or with a country-specific powercable. The power supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz. Fordetailed specifications, see Technical Data.

Attention:Allow your balance to warm up for 30 minutes (0.1 mg models 60 minutes) to enable it to adaptitself to the ambient conditions.

Connect the AC/DC adapter to the connection socket on the back of yourbalance (see figure) and to the power line.

Attention Check if your local power supply falls within this range. If this is not the case, under no circum­

stances connect the AC/DC adapter to the power supply, but contact a METTLER TOLEDO repre­sentative.

The power plug must be accessible at all times. Prior to use, check the power cable for damage. Route the cable in such a way that it cannot be damaged or cause a hindrance when working. Ensure that no liquid comes into contact with the AC/DC adapter.

19Installation and Putting into Operation

Page 20: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.5 Leveling the balanceExact horizontal positioning and stable installation are prerequisites for repeatable and accurateresults.

NoteTo compensate for small irregularities or inclinations (±2 %) at the location, the instrument must beleveled and adjusted each time it is moved to a new location.

The balance can be levelled with the leveling assistant and/or with the level indicator on top of termi­nal. There are two adjustable leveling feet to compensate for slight irregularities in the surface of thewieghing bench.

4.5.1 Levelling the balance with the level indicatorEvery balance has a level indicator with an air bubble to level the balance manually.

Leveling balances with readability of 0.1 mg and 1 mg

– Adjust the two leveling feet appropriately until the airbubble comes to rest exactly in the middle of the glass:

Air bubble at "12 o'clock" turn both feet clockwise.Air bubble at "3 o'clock" turn left foot clockwise, right

foot counterclockwise.Air bubble at "6 o'clock" turn both feet counterclock­

wise.Air bubble at "9 o'clock" turn left foot counterclock­

wise, right foot clockwise.

Leveling balances with readability of 0.01 g and 0.1 g

1 Remove the clamps (A) for the safety feet by turning themoutwards.

2 Turn the clamps (A) outwards as far as they will go (~ 90°),so that the safety feet can move freely.

3 Now level the balance by turning both leveling screws (B)until the air bubble is in the inner circle of the level indicator(see procedure above).

4 Secure the safety feet by turning the clamps (A) inwards asfar as they will go. BB

A A

20 Installation and Putting into Operation

Page 21: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.5.2 Leveling the balance with the leveling assistantWhen the balance was switched on at its new location, the symbol The instrument is out of levelappears in the status information field at the left side of the screen.

1 Tap on [ ].

The window Notifications opens.

2 Select The instrument is out of level.

The function Leveling assistant opens.The function Leveling assistant is a step-by-step guidance that helps you leveling your balance.

Navigation: [ ] Quick settings/Preferences > [ ] Leveling assistant

When the instruction is done and accepted, the leveling assistant will proceed with the next steps. Fol­low the steps until the balance is levelled again.

NoteAlways use the air bubble of the level indicator as reference indicator, see section Levelling the bal­ance with the level indicator (page 20)).

21Installation and Putting into Operation

Page 22: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.6 Transporting the balanceSwitch off the balance and remove the power cable and any interface cable from the balance. Refer tothe notes in Section "Selecting the location" regarding the choice of an optimal location.

Transporting over short distances

CAUTIONFor balances with a draft shield:Never lift the balance by the glass draft shield. The draft shield is not sufficient­ly fastened to the balance.

Transporting over long distances

CAUTIONIf you would like to transport or send your balance over long distances, usethe complete original packaging.

4.7 Weighing below the balanceThe balances are equipped with a hanger for carrying out weighings below the work surface.

CAUTION

Do not place the balance on the pan support location bolt (0.1 mg and 1 mg models). Models with a glass draft shield: Carefully lift the draft shield from the weighing platform and put it

aside.

1 Switch off the balance and remove the power cable and anyinterface cable from the balance.

2 Remove the draft shield element (10 mg models).

3 Remove the weighing pan and pan support.

4 Remove the bottom plate and unlock the draft shield (modelswith draft shield). Carefully lift the draft shield from theweighing platform and put it aside.

5 Turn the balance carefully on its side.

6 Remove and retain the 2 screws (A) and the cover plate (B).The hanger is now accessible.

7 Then turn the balance to its normal position and simply rein­stall all components in the reverse order.

B

A

22 Installation and Putting into Operation

Page 23: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

4.8 General requirements

4.8.1 Warming up the balanceBefore working with the balance, it must be warmed up in order to obtain accurate weighing results.To reach operating temperature, the balance must be connected to the power supply for at least:

30 minutes on balances with a readability of 1 mg (0.01 ct) to 5 g. 60 minutes on balances with a readability of 0.1 mg (0.001 ct) and better.

4.8.2 Adjusting the balanceTo obtain accurate weighing results, the balance must be adjusted to match the gravitational accelera­tion at its location and depending on the ambient conditions. After reaching the operation temperature,adjusting is necessary.

before the balance is used for the first time. when the balance (readability of 0.1 mg) was disconnected from the power or in case of power

failure. after a change of the location and after the warming-up phase. at regular intervals during weighing service.

23Installation and Putting into Operation

Page 24: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5 Weighing made simpleThis section describes how to make a simple weighing. In addition to that, the basic concept of thenavigation and the basic functions of the balance will be explained.

5.1 Switching on the balance1 Remove any load from the weighing pan.

2 Connect the balance via AC/DC adapter to the mains.

After the start screen disappeared, the balance starts with the application home screen.

NoteWhen the balance is switched on for the first time, the application home screen of Weighing opens. Ifthe balance is switched on again, it always starts with the home screen of the application that wasused before switching off.

5.2 Switching off the balance1 Tap on [ ] until the switch off dialogue appears on the screen.

2 Tap on [ ].

the balance switches off and goes into standby mode.Note

After switching on from standby mode, your balance needs no warm-up time and is immediatelyready for weighing.

If the balance has been switched off after a preselected time, the display is dimly lit and showsthe time and the date.

If the balance has been switched off manually, the display is off.

To completely switch off the balance, it must be disconnected from the power supply.

24 Weighing made simple

Page 25: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5.3 Main settings and activities at a glanceThis picture gives you an overview of the main settings within an application (e.g. Weighing).Depending on the application, the content of the selectable options can differ. Every application isbased on this concept.

25Weighing made simple

Page 26: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5.4 Navigation on the touch screenTo interact with the balance, use the screen and the operating keys at the bottom of the screen. Thenavigation on the screen is similar to the navigation of a smart phone or a tablet.

5.4.1 Opening an applicationTo open settings or applications, tap with your finger on the symbol of the application (e.g. [ ]Weighing).

5.4.2 ScrollingWhen the options are too numerous to be shown on one screen, a blue bar on the right side appears.That means you can scroll up and down. For scrolling, put your finger somewhere in the list and dragit up or down.

26 Weighing made simple

Page 27: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5.4.3 Using shortcutsTo simplify the navigation on the capacitive color TFT touch screen, there are a few shortcuts thatallow quick access to the most important areas of the balance. The weighing value field on the appli­cation home screen for example works as shortcut (see picture below), so does the weighing unit nextto the weighing value field. Depending on the activity, there are other shortcuts, that can be used.

NoteEvery setting that can be changed directly via shortcut, can also be changed in the contextual settingsof the current activity.

5.5 Performing a simple WeighingWhen you switch on the balance for the first time, the application home screen of Weighing opensautomatically. When the balance was switched on already, the last used application before the bal­ance was switched off opens. If another activity is running, end this activity and restart Weighing.

Navigation: [ ] > Activities – Weighing and other applications > [ ] Weighing

1 Tap on [ ] to zero the balance or [ ] to tare the balance.

The initial screen looks like follows:

2 Place a weighing sample on the weighing pan.

The unstability detector symbol appears and the value in the weighing value field becomeslight blue.

27Weighing made simple

Page 28: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

After a short period of time, the weighing pan is stable. The unstability detector symbol dis­appears and the value in the weighing value field becomes dark blue again.

The weighing process is finished. The result appears on the screen.

5.6 Switching the weighing unitThere are several weighing units available. The default value is country-specific.

The weighing unit can be chosen via the contextual settings of the current application or via shortcut.This example describes how to change the weighing unit via shortcut.

1 Tap on the weighing process unit (shortcut) gram 'g'.

28 Weighing made simple

Page 29: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

the screen Unit 1 opens.

2 Put your finger somewhere in the list and drag up to scroll down.

3 Choose another weighing unit (e.g. ounce 'oz') by tapping on it.

4 Tap on [ ] to confirm the chosen weighing unit or tap on [ ] to dismiss the chosen weighingunit.

The weighing unit gram 'g' has been changed to ounce 'oz'.

NoteWith approved balances, this menu topic has a fixed setting and cannot be changed.

29Weighing made simple

Page 30: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5.7 Changing the resolutionThere are several resolutions available. The default resolution is device-specific.

The resolution can be changed as follows:

1 Tap on the Weighing value field

2 Tap on 10d - 0.001 g

3 Confirm the chosen resolution with .

30 Weighing made simple

Page 31: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

The resolution has been changed.

5.8 Zero setting1 Unload the balance.

2 Tap on [ ] to set the balance to zero.All weight values are measured in relation to the zero point.

NoteTap on [ ] before you start with the weighing process.

5.9 Taring1 Tap on [ ] to set the balance to zero.

2 Place the empty container on the balance.

The weight is displayed.

3 Tap on [ ] to tare the balance.

The display shows Net and 0.000 g.

5.10 Closing an applicationA current application can simply be closed by choosing another application or activity. Therefore, tapon [ ] Activities.

31Weighing made simple

Page 32: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

5.11 Working with IDsIdentifications (IDs) must be defined in the contextual settings of an application. How the ID is usedand which parameters of the ID can be defined differs depending on the application in which the ID isused.

Navigation: > Open an application (e. g. Weighing) > Open the contextual settings of the applica­tion > [ ID ] Identifications

Define an ID1 Activate the function with the switch in the title bar.

2 Tap on [ID 1]

The input dialog ID 1 opens.

3 Activate the dialog with the switch in the title bar.

The title bar shows ID - Label.Define ID - Label.

4 Confirm the definition by tapping on [ ].

The input dialog ID 1 - Value opens.

5 Define ID 1 - Value.

6 Confirm the definition by tapping on [ ].

The Identifications window opens showing the defined ID.

7 Tap on [ ] to accept the defined IDs.

NoteThe maximal length of an ID is 16 characters.

5.11.1 AutoincrementAutoincrement specifies some part of the ID that is being incremented with each use of that ID. Thereare two basic functions depending on how the ID is defined:

If there is no counter in the ID, the system automatically adds a counter to the ID starting by -1(e.g. the ID Process will be Process-1 in the next use).

If a counter is part of the ID, the system automatically increments the ID starting at that counter(e.g. the ID Process 1 will be Process 2 in the next use).

Note

32 Weighing made simple

Page 33: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

The counter must be set at the end of the ID, otherwise the system doesn't recognize the numberas a counter (e.g. in 567Apple the system doesn't recognize 567 as a counter).

If the ID has no counter and a maximal length of 16 characters, the last few characters will beoverwritten by the counter.

33Weighing made simple

Page 34: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6 ActivitiesThe Activities can be opened by tapping on [ ] Activities. The icon is placed between the weighingprocess unit and the home button in the right corner of the current home application screen.

The Activities have the following two sub-sections:

Activities – Weighing and other applications Activities – Adjustments and tests

6.1 Activities - Weighing and other applications

Activities – Weighing and other applications implies the following applications:

Weighing see at section Weighing made simple (page 24) Check Weighing see at section Check Weighing (page 35) Totaling see at section Totaling (page 39)

34 Activities

Page 35: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.1.1 Check WeighingThe application Check Weighing allows to check the deviation of a sample weight within a tolerancelimit to a reference target weight. The target weight can be determined manually or by weighing, thetolerance limit must be defined manually.

Navigation: [ ] > Activities – Weighing and other applications > [ ] Check Weighing

Before performing a Check WeighingWhen the application is starting for the first time, the weighing value field is all red. That is becausethe relevant parameters haven't been defined yet. Before performing a check weighing, the followingparameters must be defined first:

Target weight ± Tolerance limits

Defining the target weight manually by entering the nominal weight1 Tap on [ ] Check weighing - Main configuration.

2 Tap on [Target, tolerance].

3 Tap on [ ] Target weight.

4 Delete the default value by tapping on [ ].

5 Enter the nominal weight.

Tap on [ ] to confirm the defined nominal weight.

Defining the target weight by weighing1 Tap on [ ] Check weighing - Main configuration.

2 Tap on [Target, tolerance].

3 Tap on [ ] Target weight.

4 Tap on [ ] to open the weighing dialog.

5 Place the reference weight on the weighing pan and wait until the unstability detector symbol disappears.

6 Tap on to accept the defined target weight.

35Activities

Page 36: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Defining the lower and upper limits manually by entering a percentage value or weight1 Tap on [ ] Check weighing - Main configuration.

2 Tap on [Target, tolerance].

3 Tap on [ ] ± Tolerance limits

4 Delete the default value by taping on [ ].

5 Enter the tolerance limit.

6 Tap on [ ] to accept the new tolerance limit.

36 Activities

Page 37: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Performing Check WeighingAfter defining the target weight and the tolerance limits, Check Weighing can be performed. Theweighing-in aid at the top bar visualizes, if the weight sample is within the defined tolerances.

In this example, the defined target weight is 100.000 g, the tolerance limit is ±2.5 %. The sampleweight is 97.0000 g.

– Put the target weight on the weighing pan.

The weight is stable and the unstability detector symbol disappears.

The value is out of tolerance, the weighing-in aid bar and the weighing value field are red.

In this example, the defined target weight is still 100.000 g and the tolerance limit is ±2.5 %. Thesample weight is 99.0000 g.

– Place the sample weight on the weighing pan.

The weight is stable and the unstability detector symbol disappears.

The value is within tolerances, the weighing-in aid bar and the weighing value field are green.

37Activities

Page 38: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Check weighing - Main configurationYou can define the following parameters:

Parameter Explanation Values

Target, tolerance Defines the target weight in thedefined unit and its tolerancelimits in %.

Target weight: 100 g*± Tolerance limits: 0.10...100(2.5 %*)

Within tolerance beeps Defines the acoustic signalwhen the result is within the tol­erance.

OFF* | ON

Unit 1 Defines the unit of the weighingprocess.

g* | mg | µg | ct | N | oz | ozt |GN | dwt | mom | msg | tlh | tls |tlt | tcl | tola | baht | lb:oz

Resolution Defines the resolution of theweighing process.

1d - 0.0001 g* | 2d -0.0002 g | 5d - 0.0005 g |10d - 0.001 g | 100d - 0.01 g| 1000d - 0.1 g

* Factory setting

NoteUnit 1: The selectable weighing units can differ depending on the balance model.

Only those weight units allowed by the appropriate national legislation are selectable. With approved balances, this menu topic has fixed settings and cannot be changed.

38 Activities

Page 39: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.1.2 TotalingTotaling allows to weigh different samples to add their weight values and to totalize them.

Navigation: [ ] Activities > [Activities – Weighing and other applications] > [ ] Totaling

Performing Totaling1 Tap on [ ] to zero the balance.

2 If using a container: Place empty container on the weighing pan and tap on [ ] to tare thebalance.

3 Place the first weight on the weighing pan.

4 Wait until the unstability detector symbol disappears.

When the balance is stable, the weighing value becomes dark-blue.

5 Tap on [ ] to accept the weight and to start the procedure.

6 Place next sample weight on the weighing pan.

7 Tap on to accept the second sample weight.

The work title bar shows the number of samples (2 samples) and the total weight of thesamples ( = 30.0000 g).

8 To open the result view, tap on [ ] in the work title bar or tap on [ ] and then tap on [ ] Viewresult

The list shows every sample that has been weighted within the weighing process. The list canbe published by tapping on the Printer symbol in the right bottom corner.

39Activities

Page 40: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Totaling – Main configurationThe following parameter can be defined:

Parameter Explanation ValuesUnit 1 Defines the unit of the weighing

process.g* | mg | µg | ct | N | oz | ozt | GN | dwt| mom | msg | tlh | tls | tlt | tcl | tola |baht | lb:oz

Resolution Defines the resolution of the weighingprocess.

1d - 0.0001 g* | 2d - 0.0002 g | 5d -0.0005 g | 10d - 0.001 g | 100d -0.01 g | 1000d - 0.1 g

* Factory settings

NoteUnit 1: The selectable weighing units can differ depending on the balance model.

Only those weight units allowed by the appropriate national legislation are selectable. With approved balances, this menu topic has fixed settings and cannot be changed.

Terminate the application1 Tap on [ ].

2 Tap on [ ] Discard data or tap on [ ] Finish and discard data (Autoprint).

The home screen of the application Totaling appears.

40 Activities

Page 41: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.2 Activities - Adjustments and tests

Activities – Adjustments and tests contains the following elements:

FACT see at section FACT Fully automatic adjustment (page 42) Internal Adjustment see at section Internal Adjustment (page 43) External Adjustment see at section External Adjustment (page 44)

41Activities

Page 42: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.2.1 FACT Fully automatic adjustmentFACT stands for Fully Automatic Calibration Technology. It is activated as default value.

FACT means, that the balance adjusts itself:

in case the conditions change (temperature difference > 2°C), which could lead to a noticeabledeviation in the measurement.

on a predefined day time set by the user.

Setting FACTYou can define the day time of FACT as follows:

1 Tap on FACT in the Activities – Adjustments and tests overview.

2 Activate Fully automatic adjustment with the switch at the right top corner.

The Fully automatic adjustment dialog opens.

3 Select the time (hours : minutes) with the pick buttons.

4 Tap on [ ] to confirm the defined time.

The time underneath FACT has been updated and shows the time of the daily adjustment.

42 Activities

Page 43: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.2.2 Internal AdjustmentNoteThe Internal Adjustment is available for models with internal weight only (see technical data).

Proceed an internal adjustment1 Tap on [Adjust internal].

The internal adjustment procedure starts. The coach text field shows Adjustment ongoing,please wait...

When the internal adjustment procedure is successfully completed, the results of the internaladjustment appear.

2 Tap on [ ] to confirm the results.NoteIf the balance is configured and connected to a printer, the results of the adjustment process are print­ed automatically.

See also

Publishing (page 48)

43Activities

Page 44: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

6.2.3 External AdjustmentNoteBecause of certification legislation, the approved balances can not be adjusted with an external weight(depending on the certification legislation of the selected country).

Performing an external adjustment1 Tap on Adjust external in the Activities – Adjustments and tests overview.

The External Adjustment screen opens.

2 Tap on [ ] to define the adjustment weight according to the weight certificate.

3 Confirm the adjustment weight by tapping on [ ].

4 Prepare the adjustment weight and tap on [ ] to start the adjustment process.

5 Load the adjustment weight in the center of the weighing pan.

6 Unload the adjustment weight from the weighing pan.

After the adjustment process, The result screen is showing date and time and other informa­tion about the external adjustment.

44 Activities

Page 45: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

7 General settingsThis section describes, how to define the general settings.The General settings are devided into two sub-sections:

Quick settings/Preferences

General configuration and data

7.1 Quick setting/Preferences

7.1.1 Date and timeIn this menu topic, date and time can be defined. Switch between Time and Date by tapping on thebars on the left side. Tap on [ ] for Time and tap on [25] for Date.

The following parameters can be defined:

Parameter Explanation ValuesTime Defines the time. Hours | MinutesTime format Defines the time format. 24:MM* | 12:MM | 24.MM | 12.MMDate Defines the date. Day | Month | YearDate format Defines the date format. MMM DD YYYY* | D.MMM YYYY |

MM/DD/YYYY | MMM DD YYYY

* Factory setting

7.1.2 BrightnessIn this menu topic the brightness can be defined.

Parameter Explanation ValuesBrightness Defines the level of brightness. 10...100 % (50 %*)

* Factory setting

7.1.3 SoundIn this menu topic, the sound can be defined.

Parameter Explanation ValuesStability beep Notifies when an unstable weight becomes stable. OFF | Low | Medium |

HighWorkflow feed­back beep

Notifies when zero/tare bar actions are not available,provides additional feedback in case of input errors,messages and status notifications.

OFF | Low | Medium |High

Touch beep Notifies every touch of interactive elements on touchdisplay and zero/tare bar.

OFF | Low | Medium |High

45General settings

Page 46: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

7.1.4 LanguageThe following languages can be chosen:

English Русский Português BRA.Deutsch Polski 中文Français Česky 日本語Español Magyar 한국어/조선말Italiano Nederlands

7.1.5 Glove modeWith the Glove Mode activated, the touch-screen is more sensitive and better to navigate when wear­ing gloves.

Parameter Explanation ValuesGlove mode Activate or deactivate the Glove mode. ON | OFF*

* Factory setting

7.1.6 Leveling assistantThe Leveling assistant helps to adjust the balance (e.g. when the balance has changed its location).

For more information, see section Leveling the balance with the leveling assistant (page 21).

46 General settings

Page 47: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

7.2 General configuration and dataThe following options are available:

Weighing options Publishing Devices and connectivity System settings ISO-log Access protection

7.2.1 Weighing optionsThe following parameters can be defined:

Parameter Explanation ValuesWeighingmode

Defines the weighing mode. Universal* = for standard weighing applicationsDosing = for dosing liquid or powdery samples

Environment Adapts the balance to the envi­ronmental circumstances.

Standard* = for standard environmentsUnstable = for unstable environmentsStable = for stable environments

Autozero This menu topic allows to switchthe automatic zero on or off.

ON* = Autozero is switched on. The automaticzero setting continuously corrects possible varia­tions in the zero point that might be causedthrough small amounts of contamination on theweighing pan.

OFF = Autozero is switched off. The zero point isnot automatically corrected. This setting isadvantageous for special applications (e.g.evaporation measurements).

* Factory setting

NoteAutozero is generally not available for approved balances (except some selected countries).

MinWeighMinWeigh notifies when a measurement value is not fulfilling the required weighing accuracy. Theusage of MinWeigh is application-specific (contextual settings).

Defining the MinWeigh method

1 Tap on MinWeigh.

The dialog MinWeigh configuration opens.

2 Tap on MinWeigh method.

The dialog MinWeigh method opens.

3 Select method and confirm by tapping on [ ].Depending on the chosen method, the selectable options in the dialog MinWeigh configuration vary.

The following methods are available:

OIML Customized Certificate

47General settings

Page 48: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

7.2.2 PublishingIn this section, the print and output options can be defined.

PrintThis option can be activated or deactivated with the switch in the work title bar.

The following parameters can be defined:

Parameter ValuesSingle values Manual, stable* = Print next stable weight at will

Manual, all values = Print any stable or unstable weight at willAutomatic, stable = Print next stable weight automatically

Workflow reports Automatic* = Print on finishingManual = Print at will and finish

FACT report ON* | OFF

* Factory setting

Printouts data (Weighing applications only)The following parameter can be defined:

Parameter Explanation ValuesDate, time To add date and time. ON | OFF*Balance information To add information

about the balance.ON | OFF*

Gross/Tare To add gross/tare tosingle values.

ON | OFF*

Signature line To add the signatureline.

ON | OFF*

Empty lines To add empty lines. ON | OFF*

* Factory setting

Send weight valueThe option can de activated or deactivated with the switch in the work title bar.

The following parameters can be chosen:

Parameter DescriptionManual, stable Send next stable weight at willManual, all values Send any stable or unstable weight at willAutomatic, stable Send next stable weight automatically

'Autmatic, continuous' 'Send any stable or unstable weight automati­cally'

* Factory setting

AutopublishThe option Autopublish can be activated or deactivated with the switch in the work title bar.

A numerical value for the Autopublish can be defined.

48 General settings

Page 49: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

7.2.3 Devices and connectivity

RS232The following parameters can be defined:

Parameter ValuesAllocated device RS-P2X | P5X | Printer | Host* | PC-Direct |

Second displayBaudrate 600 | 1200 | 2400 | 4800 | 9600* | 19200 |

38400 | 57600 | 115200Bit/Parity 8/No* | 7/No | 7/Mark | 7/Space | 7/Even |

7/OddStop bits 1 bit* | 2 bitsHandshake Xon/Xoff* | RTS/CTS | NoneCharacter set IBM/DOS* | ANSI/WIN | UTF-8

* Factory setting

USB DeviceThe following parameters can be defined:

Parameter ValuesAllocated device Host* | PC-DirectCommand set MT-SICS* | MT-PM | Sartorius 16 | Sartorius 22

* Factory setting

7.2.4 System settings

7.2.5 ISO-logIn this section, detailed history information about performed adjustments, intensity of use and settingscan be displayed.

Information type ExplanationAdjustments Shows detailed information about the performed adjustments. Balance Shows detailed information about the balance history.Settings Shows detailed information about changes of the settings.

7.2.6 Access protectionWith the function Access protection, certain functionalities of the balance can be protected by anumerical passcode.

The following parameters can be defined:

Parameter Explanation ValuesApplications To protect the application. ON | OFF*Adjustments andtests

To protect adjustments and tests. ON | OFF*

Settings To protect all settings. ON | OFF*Passcode To define the passcode by the user. 1....9 (1-16 digits)

* Factory setting

49General settings

Page 50: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

User passcodeThe user passcode can be defined by the user. The default value is 12345678 and will automaticallybe reset after a factory reset. The length of the freely changeable user passcode is restricted to 16 dig­its.

Service passcodeThe service passcode is a fixed passcode that can be obtained from the Mettler Toledo Balance sup­port site or from a Mettler Toledo representative. It is used when the user passcode has been lost.

50 General settings

Page 51: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

8 Communication with Peripheral Devices

8.1 Function PC-DirectThe numerical value displayed at the balance can be transferred to the cursor position in WindowsApplications (e.g. Excel, Word) as by typing with the keyboard.

Note: The units will not be transferred.

Requirements PC with one of the Microsoft Windows® operating system 32bit/64bit: XP (SP3), Vista (SP2),

Win 7 (SP1) or Win 8 Serial interface RS232, USB or BT option Administrator rights for installing software (for USB not required) Windows Application (e.g. Excel) Balance to PC connection with cable RS232 or USB

Settings at the balance:CAUTIONDisconnect the USB connection from the balance prior to change the settings.

Balance Interface Settings (see Interface Menu):

Topic RS232 or USB Device: set PC-Direct and select the most appropriate option for the desiredweighing result.

Topic End of line (transmitted end of line character):– set <TAB> to write into the same row (e.g. in Excel).– set <CR><LF> to write into the same column (e.g. in Excel).

Save changes.

Settings at the PC:

Installing SerialPortToKeyboardOperation of PC-Direct via serial port RS232 requires the installation of SerialPortToKeyboard on yourhost computer.

Using CD-ROM1 Insert the product CD in the CD/DVD drive of the host computer.

2 Double click the folder SerialPortToKeyboard.

Using internet1 Go to the site http://www.mettler-toledo-support.com.

2 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of aMETTLER TOLEDO instrument required).

3 Click Customer Support

4 Click appropriate product folder and save the program file SerialPortToKeyboard.exe on yourspecified storage location.

51Communication with Peripheral Devices

Page 52: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Installation procedure1 Right-click on SerialPortToKeyboard.exe and select Run as Administrator from the menu.

2 Follow the installer's instructions.

Settings for SerialPortToKeyboard

1 Select the serial port (COM) to be used forconnection with the balance.

2 Set the baud rate to 9600.

3 Activate "Connect"

Note

The window can be minimized. Closing of the window terminates the session.

Checking operation1 Start SerialPortToKeyboard (RS232)

2 Start Excel (or another application) at the PC.

3 Activate a cell in Excel. According to your selected PC-Direct option, the displayed values will appear e.g. in the column oneafter the other one in the different rows.

8.2 USB Device InterfaceTo perform the functionality USB HOST with a PC equipped only with a USB Interface, you have toassign an appropriate USB Driver on the PC first.

Requirements Balance with USB Device Interface. PC with one of the Microsoft Windows® operating system 32bit/64bit: XP (SP3), Vista (SP2),

Win 7 (SP1) or Win 8. Administrator rights for installing software. PC to balance USB connection cable.

Installing USB driver on the PC:

Using CD-ROM1 Insert the product CD in the CD/DVD drive of the host computer.

2 Double click the folder USB Driver.

3 Click USBDriverInstaller.exe.

52 Communication with Peripheral Devices

Page 53: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Using internet1 Connect to the Internet

2 Go to the site http://www.mettler-toledo-support.com.

3 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of aMETTLER TOLEDO instrument required).

4 Click Customer Support.

5 Click appropriate product folder.

6 Click USB Driver.

7 Click USBDriverInstaller.exe.

Installing procedure

1 Click Save to download to your specified location.

2 Right-click on the downloaded install program: USBDriverIn­staller.exe and select Run as Administrator from the menu.

3 If a safety warning appears, allow Windows to install.

4 Click Next and follow the installer's instructions.

Installing instrument

1 Switch the balance off.

2 Connect the balance to the preferred USB Port on the PC.

3 Switch the balance on.

4 Follow the instructions of the Wizard and install the softwareautomatically (recommended)

Note: The wizard appears again for each USB port, either on yourPC or if another balance is connected.

Warning: Do not click Cancel as for the connected USB port, itmight not be possible anymore to perform the installationprocess.

53Communication with Peripheral Devices

Page 54: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

9 Firmware (Software) UpdatesMETTLER TOLEDO is continuously improving its balance firmware (software) for the benefit of cus­tomers, so that the customer can benefit quickly and easily from further developments, METTLERTOLEDO makes the latest firmware versions available on the Internet. The firmware made available onthe Internet has been developed and tested by Mettler-Toledo AG using processes that meet the guide­lines of ISO 9001. Mettler-Toledo AG does not, however, accept liability for consequences that mightarise from using the firmware.

9.1 Operating principleYou will find all the relevant information and updates for your balance on the METTLER TOLEDO web­site at the following address:

www.mettler-toledo-support.comA program known as the "e-Loader II" is loaded onto your computer together with the firmwareupdate. You can use this program to download the firmware to the balance. The "e-Loader II" can alsosave the settings in your balance before the new firmware is downloaded to it. You can reload thesaved settings into the balance manually or automatically after the software is downloaded.

If the selected update includes an application that is not described in these instructions (or that hasbeen updated in the meantime) you can download the corresponding instructions in Adobe Acrobat®

PDF format.

NoteNew applications might not be visible unless the type data are updated by a service technician.

RequirementsThe minimum requirements for obtaining applications from the Internet and downloading them intoyour balance are as follows:

PC with one of the following Microsoft Windows® operating system: • Microsoft® Windows® XP Home or Professional with Service Pack 3 (32 bit)• Microsoft® Windows Vista® Home Premium, Business, Ultimate, or Enterprise with Service

Pack 2 (32 bit and 64 bit)• Microsoft® Windows 7 with Service Pack 1 Home Premium, Professional, Ultimate, or Enter­

prise (32 bit and 64 bit)

Administrator rights for installing software. PC to balance connection cable (e.g. No. 11101051 see chapter accessories)

9.2 Update procedure

Installing the "e-Loader II" software from the Internet onto the PC.1 Connect to the Internet.

2 Go to the site http://www.mettler-toledo-support.com.

3 Log in to the METTLER TOLEDO Balance Support Site (registration with the serial number of aMETTLER TOLEDO instrument required).

4 Click Customer Support.

5 Click appropriate product folder.

6 Click the firmware version (e-Loader II) you need and save it on your specified storage location.

7 Right-click on the firmware SNxxx.exe and select Run as Administrator from the menu.

8 Follow the installer's instructions.

54 Firmware (Software) Updates

Page 55: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Loading the new firmware into the balance.1 Right-click on METTLER TOLEDO e-Loader II and select Run as Administrator from the menu.

2 Follow the instructions, which will take you step-by-step through the installation.

55Firmware (Software) Updates

Page 56: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

10 MaintenanceWARNING

Risk of electric shock– The instrument must be disconnected from the power supply, before cleaningor other maintenance work to be performed.– Use only the power cord from METTLER TOLEDO, if it needs replacing.

Please observe the following notes:NoteOn no account use cleaning agents which contain solvents or abrasive ingredients, as this can resultin damage to the touch screen.

10.1 Cleaning the glass draft shield (0.1 mg and 1 mg models)

Remove the following parts:1 Remove weighing pan, draft shield element and pan support.

2 Remove the bottom plate.

3 Unlock the draft shield, lift it off the balance and place it on a clean surface.

A

B

1 Push the glass doors (A) back.

2 Turn the two lock covers (B) on thefront as far as they will go.

C

1 Tilt the front glass (C) forward.

2 Remove the front glass.

– Pull the top glass door (D) out from thefront. E

F

– Lift the side glass doors (E) at (F) andpull them out from the front.

56 Maintenance

Page 57: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

G

– Push the lock button (G) to release therear glass.

H

– Remove the rear glass (H).

1 Turn the draft shield lock to the " "(Service) position.

2 Remove the draft shield lock.

After cleaning reinstall all components in the reverse order. For balance mounting see chapter "Set­ting up the Balance – Installing the Components".

10.2 Disposal

In conformance with the European Directive 2002/96/EC on Waste Electrical andElectronic Equipment (WEEE) this device may not be disposed of in domestic waste.This also applies to countries outside the EU, per their specific requirements.

Please dispose of this product in accordance with local regulations at the collectingpoint specified for electrical and electronic equipment. If you have any questions,please contact the responsible authority or the distributor from which you purchasedthis device. Should this device be passed on to other parties (for private or profession­al use), the content of this regulation must also be related.

Thank you for your contribution to environmental protection.

57Maintenance

Page 58: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

11 Interface Specification

11.1 RS232C interfaceEach balance is equipped with an RS232C Interface as standard for the attachment of a peripheraldevice (e.g. printer or computer).

Schematic Item SpecificationInterface type Voltage interface according to EIA

RS-232C/DIN66020 CCITT V24/V.28)Max. cable length 15 mSignal level Outputs:

+5 V … +15 V (RL = 3–7 kΩ)–5 V … –15 V (RL = 3–7 kΩ)Inputs:+3 V … +25 V–3 V … -25 V

Connector Sub-D, 9-pole, femaleOperating mode Full duplexTransmissionmode

Bit-serial, asynchronous

Transmissioncode

ASCII

Baud rates 600, 1200, 2400, 4800, 9600, 19200,38400, 57600, 115200 (softwareselectable)

Bits/parity 7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none(software selectable)

Stop bits 1 stop bitHandshake None, XON/XOFF, RTS/CTS (software

selectable)End-of-line <CR><LF>, <CR>, <LF> (software

selectable)

15

9 6

IN

GND

OUT

HAND

SHAKE

IN

DATA

OUT

RxD

TxD

CTS

RTS

OUT

POWER

SUPPLY

+12V

2nd display mode only

Power supply for2nd display

+ 12 V, max 40 mA (software selectable,2nd display mode only)

11.2 USB hostSchematic Item Specification

Standard In conformity with USB Specification Revi­sion 1.0/1.1

Speed Full speed 12 Mbps (requires shieldedcable)

Power usage Max. 500 mAConnector Type A

1 VBUS (+5 V DC)2 D- (Data -)3 D+ (Data +)4 GND (Ground)

Pin assignment

Shell Shield

58 Interface Specification

Page 59: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

11.3 USB deviceEach balance is equipped with an "USB Device" Interface as standard for the attachment of a peripher­al device (e.g. computer).Note: This interface is not suitable to communicate with a Printer.

Schematic Item SpecificationStandard In conformity with USB Specification Revi­

sion 1.1Speed Full speed 12 Mbps (requires shielded

cable)Function CDC (Communication Device Class) serial

port emulationPower usage Suspended device: Max 10 mA

1 VBUS (+5 VDC)

2 D– (Data –)

3 D+ (Data +)

4 GND (Ground)

Shield Shield

Connector Type B

11.4 MT-SICS interface commands and functionsMany of the instruments and balances used have to be capable of integration in a complex computeror data acquisition system.

To enable you to integrate balances in your system in a simple manner and utilize their capabilities tothe full, most balance functions are also available as appropriate commands via the data interface.

All new METTLER TOLEDO balances launched on the market support the standardized command set"METTLER TOLEDO Standard Interface Command Set" (MT-SICS). The commands available dependingon the functionality of the balance.

For further information please refer to the Reference Manual MT-SICS downloadable from the Internetunder

u www.mt.com/sics-newclassic

59Interface Specification

Page 60: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12 Technical Data

12.1 General data

Power supplyPrimary: 100 V–240 V, ±10%, 50/60 Hz, 0.3 ASecondary: 12 V DC, 0.84 A (with electronic overload protection)Power supply to the balance: 12 V DC, 10 W

AC/DC adapter

Use only with a tested AC/DC Adapter with SELV output cur­rent.Ensure correct polarity

Power supply to the balance 12 V DC, 10 W

Protection and standards Overvoltage category II. III

Degree of pollution 2

Degree of protection Protected against dust and water

Standards for safety and EMC See Declaration of Conformity

Range of application For use only in enclosed interior rooms

Environmental conditions Height above mean sea level up to 4000 m

Ambient temperature range 10 to 30 °C

Relative air humidity 10% up to 80% at 31 °C, linearly decreasing to 50% at 40 °C,noncondensing

• Warm-up time After connecting the balance to the power supply or switched onin battery operation at least

30 minutes on balances with a readability of 0.001 g(0.01 ct) to 5 g.

60 minutes on balances with a readability of 0.1 mg(0.001 ct) and better.

Materials Housing Die-cast aluminum, lacquered

Weighing pan Stainless steel X2CrNiMo 17-12-2 (1.4404)245 x 351 mm: Stainless steel X5CrNiMo 18-10 (1.4301)

Draft shield element with 0.1 mg models: Stainless steel X2CrNiMo 17-12-2(1.4404)with 10 mg models: Plastic (PBT)

Draft shield Plastic (PBT), glass

In-use-cover Plastic (PET)

60 Technical Data

Page 61: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.2 Model-specific dataNoteThe definition of the model-specific data in this section is still in progress and may change to the finalsoftware version.

12.2.1 Balances with readability of 0.1 mg, S platform with draft shield

MS104TS MS204TSLimit valuesMaximum capacity 120 g 220 gReadability 0.1 mg 0.1 mgRepeatability (at nominal load) 0.1 mg 0.1 mgLinearity deviation 0.2 mg 0.2 mg Sensitivity temperature drift 1.5 ppm/°C 1.5 ppm/°CTypical valuesRepeatability (at nominal load) 0.08 mg 0.08 mgLinearity deviation 0.06 mg 0.06 mgMinimum sample weight (acc. to USP) 0.16 g 0.16 gMinimum sample weight (U=1 %, k=2) 0.016 g 0.016 gMinimum sample weight OIML 0.01 g 0.01 gSettling time 2 s 2 sAdjustment Int.Cal / FACT Int.Cal / FACTBalance dimensions (W x D x H) 204x347x348 mm 204x347x348 mmWeighing pan dimensions ø 90 mm ø 90 mmUsable height of draft shield 236 mm 236 mmWeight of balance 5.9 kg 5.9 kgWeights for routine testingOIML CarePac

Weights

#11123002

100 g F2, 5 g E2

#11123001

200 g F2, 10 g E2

ASTM CarePac

Weights

#11123102

100 g 1, 5 g 1

#11123101

200 g 1, 10 g 1

61Technical Data

Page 62: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

MS304TSLimit valuesMaximum capacity 320 gReadability 0.1 mgRepeatability (at nominal load) 0.1 mgLinearity deviation 0.2 mg Sensitivity temperature drift 1.5 ppm/°CTypical valuesRepeatability (at nominal load) 0.08 mgLinearity deviation 0.06 mgMinimum sample weight (acc. to USP) 0.16 gMinimum sample weight (U=1 %, k=2) 0.016 gMinimum sample weight OIML 0.01 gSettling time 2 sAdjustment Int.Cal / FACTBalance dimensions (W x D x H) 204x347x348 mmWeighing pan dimensions ø 90 mmUsable height of draft shield 236 mmWeight of balance 5.9 kgWeights for routine testingOIML CarePac

Weights

#11123001

200 g F2, 10 g E2

ASTM CarePac

Weights

#11123101

200 g 1, 10 g 1

62 Technical Data

Page 63: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.2.2 Balances with readability of 1 mg, S platform with draft shield

MS303TS MS403TSLimit valuesMaximum capacity 320 g 420 gReadability 1 mg 1 mgRepeatability (at nominal load) 1 mg 1 mgLinearity deviation 2 mg 2 mg Sensitivity temperature drift 3 ppm/°C 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.7 mg 0.7 mgLinearity deviation 0.6 mg 0.6 mgMinimum sample weight (acc. to USP) 1.4 g 1.4 gMinimum sample weight (U=1 %, k=2) 0.14 g 0.14 gMinimum sample weight OIML 0.02 g 0.02 gSettling time 1.5 s 1.5 sAdjustment Int. Cal / FACT Int. Cal / FACTBalance dimensions (W x D x H) 204x347x283 mm 204x347x283 mmWeighing pan dimensions 127x127 mm 127x127 mmUsable height of draft shield 168 mm 168 mmWeight of balance 5.6 kg 5.6 kgWeights for routine testingOIML CarePac

Weights

#11123001

200 g F2, 10 g F1

#11123000

200 g F2, 20 g F1

ASTM CarePac

Weights

#11123103

200 g 1, 10 g 1

#11123100

200 g 1, 20 g 1

63Technical Data

Page 64: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

MS603TS MS1003TSLimit valuesMaximum capacity 620 g 1020 gReadability 1 mg 1 mgRepeatability (at nominal load) 1 mg 1 mgLinearity deviation 2 mg 2 mg Sensitivity temperature drift 3 ppm/°C 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.7 mg 0.7 mgLinearity deviation 0.6 mg 0.6 mgMinimum sample weight (acc. to USP) 1.4 g 1.4 gMinimum sample weight (U=1 %, k=2) 0.14 g 0.14 gMinimum sample weight OIML 0.02 g 0.02 gSettling time 1.5 s 1.5 sAdjustment Int. Cal / FACT Int. Cal / FACTBalance dimensions (W x D x H) 204x347x283 mm 204x347x283 mmWeighing pan dimensions 127x127 mm 127x127 mmUsable height of draft shield 168 mm 168 mmWeight of balance 5.6 kg 5.6 kgWeights for routine testingOIML CarePac

Weights

#11123007

500 g F2, 20 g F1

#11123008

1000 g F2, 50 g F1

ASTM CarePac

Weights

#11123107

500 g 1, 20 g 1

#11123108

1000 g 1, 50 g 1

64 Technical Data

Page 65: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.2.3 Balances with readability of 0.01 g, S platform

MS1602TS MS3002TS MS4002TSLimit valuesMaximum capacity 1620 g 3200 g 4200 gReadability 0.01 g 0.01 g 0.01 gRepeatability (at nominal load) 0.01 g 0.01 g 0.01 gLinearity deviation 0.02 g 0.02 g 0.02 g Sensitivity temperature drift 3 ppm/°C 3 ppm/°C 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.007 g 0.007 g 0.007 gLinearity deviation 0.006 g 0.006 g 0.006 gMinimum sample weight (acc. to USP) 14 g 14 g 14 gMinimum sample weight (U=1 %,k=2)

1.4 g 1.4 g 1.4 g

Minimum sample weight OIML 0.5 g 0.5 g 0.5 gSettling time 1.5 s 1.5 s 1.5 sAdjustment Int. Cal / FACT Int. Cal / FACT Int. Cal / FACTBalance dimensions (W x D x H) 194x347x99 mm 194x347x99 mm 194x347x99 mmWeighing pan dimensions 170x200 mm 170x200 mm 170x200 mmWeight of balance 4.6 kg 4.6 kg 4.6 kgWeights for routine testingOIML CarePac

Weights

#11123008

1000 g F2, 50 gF2

#11123009

2000 g F2, 100 gF2

#11123010

2000 g F2, 200 gF2

ASTM CarePac

Weights

#11123106

1000 g 1, 50 g 1

#11123109

2000 g 1, 100 g1

#11123110

2000 g 4, 200 g4

65Technical Data

Page 66: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

MS4002TSDR MS6002TS MS6002TSDRLimit valuesMaximum capacity 4200 g 6200 g 6200 gMaximum capacity, fine range 820 g – 1220 gReadability 0.1 g 0.01 g 0.1 gReadability, fine range 0.01 g – 0.01 gRepeatability (at nominal load) 0.1 g 0.01 g 0.01 gRepeatability, at fine range (at nominalload)

0.01 g – 0.01 g

Linearity deviation 0.2 g 0.02 g 0.2 g Linearity deviation, fine range 0.02 g – 0.02 g Sensitivity temperature drift 3 ppm/°C 3 ppm/°C 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.05 g 0.007 g 0.05 gRepeatability, at fine range (at nominalload)

0.007 g – 0.007 g

Linearity deviation 0.02 g 0.006 g 0.02 gLinearity deviation, at fine range 0.006 g – 0.006 gMinimum sample weight (acc. to USP) 14 g 14 g 14 gMinimum sample weight (U=1 %,k=2)

1.4 g 1.4 g 1.4 g

Minimum sample weight OIML 0.5 g 0.5 g 0.5 gSettling time 1.5 s 1.5 s 1.5 sAdjustment Int. Cal / FACT Int. Cal / FACT Int. Cal / FACTBalance dimensions (W x D x H) 194x347x99 mm 194x347x99 mm 194x347x99 mmWeighing pan dimensions 170x200 mm 170x200 mm 170x200 mmWeight of balance 4.6 kg 5.1 kg 5.1 kgWeights for routine testingOIML CarePac

Weights

#11123010

2000 g F2, 200 gF2

#11123011

5000 g F2, 200 gF2

#11123011

5000 g F2, 200 gF2

ASTM CarePac

Weights

#11123110

2000 g 4, 200 g4

#11123111

5000 g 4, 200 g4

#11123111

5000 g 4, 200 g4

66 Technical Data

Page 67: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

MS12002SLimit valuesMaximum capacity 12200 gReadability 0.01 gRepeatability (at nominal load) 0.01 gLinearity deviation 0.02 gSensitivity temperature drift 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.007 gLinearity deviation 0.006 gMinimum sample weight (acc. to USP) 14 gMinimum sample weight (U=1 %,k=2)

1.4 g

Minimum sample weight OIML 0.5 gSettling time 1.5 sAdjustment Int. Cal / FACTBalance dimensions (W x D x H) 194x347x99 mmWeighing pan dimensions 170x200 mmWeight of balance 5.2 kgWeights for routine testingOIML Weights 10000 g F2,

500 g F2ASTM Weights #11123110

10000 g 4,500 g 4

67Technical Data

Page 68: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.2.4 Balances with readability of 0.1 g to 1 g, S platform

MS8001TSLimit valuesMaximum capacity 8200 gReadability 0.1 gRepeatability (at nominal load) 0.1 gLinearity deviation 0.2 g Sensitivity temperature drift 3 ppm/°CTypical valuesRepeatability (at nominal load) 0.07 gLinearity deviation 0.06 gMinimum sample weight (acc. to USP) 140 gMinimum sample weight (U=1 %, k=2) 14 gMinimum sample weight OIML 5 gSettling time 1.5 sAdjustment Int. Cal / FACTBalance dimensions (W x D x H) 194x347x99 mmWeighing pan dimensions 170x200 mmWeight of balance 4.6 kgWeights for routine testingOIML CarePac

Weights

#11123011

5000 g F2, 200 g F2

ASTM CarePac

Weights

#11123111

5000 g 4, 200 g 4

68 Technical Data

Page 69: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.3 Dimensions

12.3.1 Balances with readability of 0.1 mg and draft shield

Models:MS104TSMS204TSMS304TS

23

6

77

86 245

21

173

17

4

14

0

347

19

4

20

4

34

8

74

98

218

85

ø 90

115

140 99

69Technical Data

Page 70: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.3.2 Balances with readability of 1 mg and draft shield

Models:MS303TSMS403TSMS603TSMS1003TS

127

127

168

85

104.5 (MS1003x)

86 245

21

173

174

140

347

194

204

283

74

100

218

77

140 99

115

70 Technical Data

Page 71: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.3.3 Balances with readability of 0.01 g

Models:MS1602TSMS3002TSMS4002TSMS4002TSDRMS6002TSMS6002TSDR

113 28

24586

18.5

74 9

4 99

14

0

17

0

19

4

26

8

200

17

0

235.5

347

104.5 (MS6002x)

71Technical Data

Page 72: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

12.3.4 Balances with readability of 0.1 g to 1 g

Models:MS8001TS

226

190

113 28

24586

14

74

99

140

170

194

268

240

347

72 Technical Data

Page 73: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

13 Accessories and Spare Parts

13.1 Accessories

Description Part No.

Density determination

Glass beaker, height 100 mm, Ø 60 mm 00238167

Sinker for density of liquids in conjunction with density kit 00210260Calibrated (sinker + certificate) 00210672Recalibrated (new certificate) 00210674

Calibrated thermometer with certificate 11132685

Weighing pans

Dynamic weighing pan MS-DWP-21 with 4 litre bowl (forbalances with readability of 0.01 g and 0.1 g )

30006471

Draft shields

C

Draft shield with sliding doors "mg" (usable heigh168 mm)

12122405

C

Draft shield with sliding doors "0.1 mg" (usable heigh236 mm)

12122404

Draft shield MS-DS-21 for models with readability of 0.1 gup to 0.01 g.

12121014

73Accessories and Spare Parts

Page 74: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Printers

RS-P25 printer with RS232C connection to instrument 11124300Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975

RS-P26 printer with RS232C connection to instrument(with date and time)

11124303

Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975

RS-P28 printer with RS232C connection to instrument(with date, time and applications

11124304

Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975

P-56RUE thermal printer with RS232C, USB and ethernetconnections, simple printouts, date and time, label print­ing (limited).

30094673

Paper roll, white, set of 10 pcs 30094723Paper roll, white, self-adhesive, set of 10 pcs 30094724Paper roll, white, self-adhesive labels, set of6 pcs

30094725

P-58RUE Thermal printer with RS232C, USB and ethernetconnections, simple printouts, date and time, label print­ing, balance applications: statistics, formulation, totaling,

30094674

Paper roll, white, set of 10 pcs 30094723Paper roll, white, self-adhesive, set of 10 pcs 30094724Paper roll, white, self-adhesive labels, set of6 pcs

30094725

Cables for RS232C interface

RS9 – RS9 (m/f): connection cable for PC, length = 1 m 11101051

RS9 – RS25 (m/f): connection cable for PC, length = 2 m 11101052

74 Accessories and Spare Parts

Page 75: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

RS232 - USB converter cable – Cable with converter toconnect a balance (RS232) to a USB port

64088427

Cables for USB interface

USB (A –B) connection cable for connection to PC,length = 1 m

12130716

Cable replacement (wireless)

Auxiliary displays

METTLER TOLEDO

000000

RS232 auxiliary display AD-RS-M7 12122381

Power supplies

AC/DC universal adapter (EU, USA, AU, UK) 100–240VAC, 50/60 Hz, 0.3 A, 12 VDC 0.84 A

11120270

AC/DC adapter (without power cable) 100–240 V AC,0.8 A, 50/60 Hz, 12 V DC 2.5 A

11107909

Protective covers

Protective cover for balaces with draft shield 12121850

Protective cover for balances without draft shield 12121851

75Accessories and Spare Parts

Page 76: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Anti-theft devices

Steel cable 11600361

Software

LabX direct balance (simple data transfer) 11120340

Weighing below the balance

Transport cases

Transport case 11124245

Adjustment weights

OIML / ASTM Weights (with calibration certificate) see www.mt.com/weights

Density kit MS-DNY-43 for MS-TS balances(d = 0.1 mg/1 mg)

11142143

Bluetooth RS232 serial adapter ADP-BT-S for wirelessconnection between printer and balance* or between bal­ance and PC*. Fits printers P-56 / P-58 and the followingbalance models (SW V3.20 or higher required): MS-TS.* Bluetooth interface required

1 Bluetooth RS232 serial adapter (slave) 1 MT-DB9 male to female connector 1 MT-DB9 male to male connector

30086494

76 Accessories and Spare Parts

Page 77: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Bluetooth RS232 serial adapter set ADP-BT-P for wirelessconnection between printer and balance. Fits printers P-56/ P-58 and the following balance models (SW V3.20 orhigher required): MS-TS.

2 Bluetooth RS232 serial adapter paired (slave/mas­ter)

1 MT-DB9 male to female connector 1 MT-DB9 male to male connector

30086495

13.2 Spare partsDraft shield

Drawing Pos Description Part No.5 Draft shield lock 121220136 Bottom plate 12122019

Draft shield "168 mm"1 Top glass with handle 121218842 Rear glass low 121220153 Side glass door left low with handle 121218814 Side glass door right low with handle 121218837 Front glass low 12122014

Draft shield "236 mm"1 Top glass with handle 121218842 Rear glass high 121220123 Side glass door left high with handle 121218804 Side glass door right high with handle 12121882

7

4

2

1

3

5

6

7 Front glass high 12122011

Weighing pans / draft shield elements / support

Drawing Pos Description Part no.

1 0.1 mg Weighing pan Ø 90 mm 121220102 0.1 mg Pan support Ø 90 mm 111242493 0.1 mg Draft shield element 121220084 1 mg Weighing pan 127x127 mm 121220095 1 mg Pan support 127x127 mm

for models up to 999 g12122017

5 1 mg Pan support 127x127 mmfor models from 1000 g

12122016

6 0.01 g Draft shield element170x200 mm

12122018

7 0.01 g Weighing pan 170x200 mm 111242478 0.01 g Pan support 170x200 mm 121210647 0.1 g Weighing pan 190x226 mm 111242488 0.1 g Pan support 190x226 mm 121210669 from

0.01 gPan support caps 11131029

10 Leveling foot 11106323

4

1

2

3 5

6

7

9

10

8

MS

MS

77Accessories and Spare Parts

Page 78: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

Index

A

Access protection 49Adjustments 41Applications 34Tests 41Adjusting 23Ambient conditions 18Application “Totaling” 39Application home screen 12Action buttons 12Information and work bars 12Information screen 12Check weighing 35Autopublish 48

C

Changing the resolution 30Cleaning 56

D

Dimensions 69Disposal 57Draft Shield 17, 77

F

FACT 42Firmware update 54Function PC-Direct 51

G

General configuration anddata

47

General safety information 8General settings 45General weighing options 47

I

Entering Text and numbers 13Changing time 14Installing the components 17Intended use 8Interface

MT-SICS 59Interface RS232C 58Interface USB device 59

L

Quick settings/Preferences 46Leveling assistant 21Leveling the balance 20Location 18

M

MinWeigh 47MT-SICS 59

N

Navigation 26

O

Operating temperature 23

P

PC-Direct 51Power supply 19Print 48

Q

Sound 45

R

RS232C interface 58

S

Safety InformationGeneral 8Intended use 8Signal words 8Staff safety 8Warning Symbols 8

Selecting the location 18Send weight value 48Setting FACT 42Signal words 8Simple weighing 27Software update 54Spare parts 77Staff safety 8Switching

On 23Switching off 24Switching on 24Switching the weighing unit 28

T

Taring 31Technical data dimensions 69Technical data general 60Totaling 39Transporting the balance 22

U

USB Device Interface 52, 59USB-Driver installing 52User

Interface 11

Index78

Page 79: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

W

Warm-up time 23, 60Warning Symbols 8Weighing below the bal­ance

22

Z

Zero setting 31

Index 79

Page 80: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 81: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 82: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 83: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS
Page 84: Operating Instructions Precision and Analytical Balances MS-TS...1 Introduction Thank you for choosing a METTLER TOLEDO balance. The precision and analytical balances of the MS-TS

GWP® is the global weighing standard, ensuring consistent accuracy

of weighing processes, applicable to all equipment from any manu-

facturer. It helps to:

• Choose the appropriate balance or scale

• Calibrate and operate your weighing equipment with security

• Comply with quality and compliance standards in laboratory and

manufacturing

www.mt.com/GWP

GWP®

Good Weighing Practice™

*30235435*

For more information

Mettler-Toledo AG, Laboratory WeighingCH-8606 Greifensee, SwitzerlandTel. +41 (0)44 944 22 11Fax +41 (0)44 944 30 60www.mt.com

Subject to technical changes.© Mettler-Toledo AG 11/201430235435A en