optically stimulated luminescence dosimetry based on srs

13
DESY - August 7th 2007, Hamburg, Germany 1 Optically Stimulated Luminescence Dosimetry Based on SrS:Ce,Sm Phosphor for High-Energy Physics Applications 1 « RADIAtions et Composants » Research Group Institut d’Électronique du Sud (IES), University of Montpellier II, Montpellier, FRANCE 2 CERN, Geneva, SWITZERLAND D. Benoit 1 , F. Ravotti 1,2 , P. Garcia 1 , J-R. Vaillé 1 , and L. Dusseau 1

Upload: others

Post on 17-Oct-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 1

Optically Stimulated Luminescence Dosimetry

Based on SrS:Ce,SmPhosphor for High-Energy

Physics Applications

1 « RADIAtions et Composants » Research GroupInstitut d’Électronique du Sud (IES), University of Montpellier II, Montpellier, FRANCE

2 CERN, Geneva, SWITZERLAND

D. Benoit1, F. Ravotti1,2, P. Garcia1, J-R. Vaillé1, and L. Dusseau1

Page 2: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 2

Why OSL ? A bit of History…1950/1960: first suggestion of OSL as a dosimetry tool80’s years:

OSL technique became popular in archeological and geological dating communityDevelopment of OSL imaging technique

In Montpellier80’s years: first research program on OSL phenomenon90’s years: development of OSL materialsConception of dosimeters and developments of applications1998: first paper on OSL dose-mapping2001: first paper on integrated OSL sensor2007: first accepted paper on OSL fibered dosimetry system for RADECS Conference

INTRODUCTION

Page 3: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 3

INTRODUCTION

(1) e-/h+ pair generation and trapping

(2) Infrared stimulation (700-1300 nm)

(3) Visible emission (450-650 nm)

Readout “reset” the OSL!

OSL

IR stimulation

Photo-sensor

(Sec)

OSL Phenomenon

SrS:Ce,Sm

Conduction Band

Valence Band

Trap level

(1) Ionising Radiations

Recombinationlevel

(2) Optical Stimulation (IR) (3) Visible

Emission

Stimulation (IR)Emission (visible)300 500 700 900 1100 1300

λ (nm)

(V)

Page 4: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 4

FeaturesBroad dynamic range: 6 decadesDetection threshold: 10 µGyLinear response with dose up to 500 Gy

OSL Material: SrS:Ce,SmMATERIAL PROPRIETIES

Raw OSLDifferent packagingIntegrated OSL sensor(grain size ~ few 100 µm)OSL Films (grain size ~50 µm)OSL deported by optical fibre(cylindrical shape with a diameter ~900 µm)

SPACE APPLICATIONS

BEAM IMAGING

HEP and Medical Applications

(Not presented here)(on-line dosimetry)

OFF-LINE DOSIMETRY

ON-LINE DOSIMETRY

Page 5: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 5

Principle

Dose-Mapping SystemOFF-LINE DOSIMETRY

System FeaturesReproducibility: 5 %Scan Area: 20x20 cm²Spatial Resolution: ~250 µmSensitivity: < mGyLayer Homogeneity: 95%

Kapton Foil

OSL Layer

OSLFilm

IR Stimulation LED

Photomultiplier tube

Displa-cement stage

Glass tray

Aim: Record of spatial distribution of the dose

Film thickness from 80 to 250 µm

Page 6: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 6

Example of Beam Profiling

One method of beam profiling: measurement of activation of Al foils

OFF-LINE DOSIMETRY

23 GeV proton beam (IRRAD1 - CERN)

OSL filmsImprovement of spatial resolutionReduction of experiment time

Spatial resolution: 500 µm Spatial resolution: 5 mm

Page 7: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 7

ON-LINE DOSIMETRY

First Prototype of OSL-Fibered System

Polymicro fibre doesn't show significant attenuation at the OSL λ of interest

– Repeatability: 2 %

– Reproducibility: 6 %

– No evidence of radiation damage up

to Φeq = 5x1013 cm-2

[Courtesy M. Glaser, CERN]

Laser Light

Visible Light

Page 8: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 8

Collaboration between CERN, SCK-CEN and IES Montpellier

Second Prototype of Fibered System

Radiation AreaHead with OSL crystal

Stimulation Laser Diode

Optical Coupler

Opt

ical

Fib

reON-LINE DOSIMETRY

Main Advantages:- Integrated system- Single Optical Fibre (Polymicro fibres for fibres #02 and #03)- Design of a head for introduction of OSL material

Page 9: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 9

LASER(980 nm)

PMT

OPTICALCOUPLER

LAPTOP with Labview

OPTICAL FIBRE

ADC/DAC

Overview of the System

10 mm

ON-LINE DOSIMETRY

4 m long O.F.

45x51x13 cm3

Page 10: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 10

Calibration Curve

First Experimental Results with 60-CoRepeatability

D = 600 mGy

- Max. Error: 1.6 %- Mean Error: 0.6 %- 85% of values: ± 1%

- Good linearity up to 1 Gy- D > 1 Gy Modify the PMT

sensitivity Software control

4 m long optical fibre (Fibre #01)

ON-LINE DOSIMETRY

R² > 0.999S ~50mV.cGy-1

(for UPM = 0.55 V)

4 m long optical fibre (Fibre #01)

Page 11: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 11

Experimental Set-upMedical LINACEnergy X-rays: 6 MVDose rate: 2 Gy/minPMMA Blocks

Calibration with High Energy X-raysON-LINE DOSIMETRY

Fibre #02 (25 m) Fibre #03 (15 m)UPM (V)

Sensitivity(mV.cGy-1)

0.5 V 0.8 V 0.5 V 0.8 V

4.19 81.9 3.84 74.5

0 1 2 3 4 5 60,0

0,5

1,0

1,5

2,0

2,5

3,0

3,5

4,0

4,5

5,0

OS

L S

igna

l (V

)

Dose (Gy)

Fibre n°03. UPM = 0.5 V Fibre n°03. UPM = 0.8 V Fibre n°02. UPM = 0.5 V Fibre n°02. UPM = 0.8 V

Good Linearity up to 6 Gy for the two optical fibres

Page 12: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 12

PrincipleThermal and epithermal neutrons: OSL+B2O3

Fast neutrons: OSL+PERecoil protons following elastic scattering

First Results of two different neutrons fields

OSL Neutron-Sensitive Material

α+→<+ LieVEnB th710 )5.0(

MATERIAL DEVELOPMENT

First measurements match very well with the facility spectra

FACILITIES NEUTRON SPECTRA

OSL+B OSL+PE

CERN IRRAD2Ljubljana TRIGA Reactor

Page 13: Optically Stimulated Luminescence Dosimetry Based on SrS

DESY - August 7th 2007, Hamburg, Germany 13

Conclusions & OutlookOSL dosimetry

Interesting properties (sensitivity, dynamics, linear response with dose)OSL Films: spatial distribution of the doseOSL Fibered dosimetry system: real-time measurements in harsh environment

AdvantagesMaterial Advantages

Not needed to heat the sample (Optical readout)High Radiation Hardness

Optical Fibre AdvantagesInsensitive to electro-magnetic interferences

Possibility to build compact, low cost and flexible systems

OutlookComplete characterization of the system ongoing Comparison of results with others dosimetric tools