optimum uniformity
TRANSCRIPT
plus a wide temperature rangeOptimum uniformity
Thermo ScientificForma Environmental ChambersControlled Temperature Storage Solutions
* Lights (Cat. No. See page 11) do not meet ICH guidelines.
TypeThermo Scientific Cat. No. Temp Range Uniformity RH
Size (Cu. Ft.) CO2 Lights
Stability
3940 (3949) 0°C TO 60°C ±0.3°C @ 25°C to 37°C Above ambient to 95% @ 37°C 29.0 Optional Optional*
3920 0°C TO 60°C ±0.3°C @ 25°C to 37°C N/A 29.0 Optional Optional*
3911 (3913) 0°C TO 60°C ±0.3°C @ 25°C to 37°C Above ambient to 95% @ 37°C 11.0 Optional N/A
3960 (3961)5°C above ambient to 60°C
±0.4°C @ 30°C±0.3°C @ 37°C±0.5°C @ 45°C
N/A 29.0 N/A N/A
A Choice of Environmental Chambers to Meet Your Specific Needs
At-A-Glance Chamber Comparison
From stability testing to cell culturing, CO2, RH, and refrigeration control to temperature-only, and 11.0 to 29.0 cu. ft., Thermo Scientific™ Forma™ environmental chambers are built to last and will meet your needs now and in the future.
If you require a high precision cabinet to meet ICH, FDA, TAPPI, ASTM or national testing standards our 3920/3940/3911 Series is the ideal choice.
• Uniformity and a wide temperature range for easy compliance with regulatory requirements
• Rugged construction and durable shelving designed for long life
• Broad range of options to create the perfect fit for your applications
Consider our Thermo Scientific 3960 Reach-In Incubator/ Environmental Chamber if you are focused on elevated temperature-only segments of drug stability and culture.
• Optimum uniformity and an elevated temperature range for easy compliance with regulatory requirements
• Rugged construction and durable shelving for long life
• Valuable features designed for flexibility and cost-effectiveness
2
This series of environmental chambers meets International Conference on Harmonization (ICH) guidelines* for drug substance and product storage testing. The units are ideal for drug stability studies, shelf life testing for packaged products, and insect and other large-scale biological research.
Optimum uniformity and a wide temperature range for easy compliance with regulatory requirements
These chambers feature a 0°C to 60°C (32°F to 140°F) temperature range that can be set in 0.1°C increments. The broad temperature range, and optimum temperature uniformity and recovery contribute to an ideal test environment, even when the chamber contains large product loads and low output,
heat-generating equipment.
Advantages of our directed horizontal laminar airflow system
Our directed airflow system (shown below) promotes an ideal growth environment. The enhanced design includes a positive pressure feed plenum on the right
side of the chamber and a negative pressure return plenum on the left to distribute the airflow uniformly throughout the chamber.
This combination directs air across the surface of each solid shelf. Even when filled with samples or equipment, each shelf receives a consistent flow of conditioned air for optimum temperature uniformity and recovery.
As opposed to our horizontal airflow system, top-to-bottom (non-directed) airflow systems use a top-mounted fan to push air down through wire shelves.
Temperature uniformity and recovery can deteriorate quickly when shelves are filled because air movement is blocked. That variation in temperature, alone or when combined with frequent door openings, may compromise environmental conditions or make process validation difficult.
Standard 4 to 20 milliamp output connects to most alarm/monitoring systems, allowing you to meet internal and regulatory product documentation requirements. Chart recorders are also available.
The Thermo Scientific directed airflow design (left) versus a top-to-bottom, undirected airflow design (right).
The directed airflow minimizes the risk of product desiccation and loss, and wasted time and money due to poor temperature uniformity and recovery.
Thermo Scientific 3920/3940/3911 Series Stability Chambers for a Multitude of Applications
3
Dehumidify and Humidify Switches and Indicators (3940/3949, 3911/3913 only) Control humidification and refrigeration systems, show cycle status (dehumidify switch operates only with optional heatless dryer)
Humidity Controller (3940/3949, 3911/3913) only)
Displays actual RH (%) and setpoint (Ramp and Soak)
Defrost and Refrigeration Switches and Indicators Adjustable defrost timer is factory set for two 15 minute defrost cycles during a 24 hour period
Over-temperature Safety Thermostat and Indicator Set over-temp (alarm and light indicate over-temp condition)
Temperature Controller Displays actual chamber temperature and setpoint(Ramp and Soak)
Main Power Switch and Indicator
Under-temperature Safety Thermostat and Indicator Set under-temp control (alarm and light indicate under-temp condition)
Humidity Alarm and Indicator (3940/3949, 3911/3913 only)
Indicates deviation of cabinet humidity from set parameters
Rugged construction and durable shelving for long life
• Exterior cabinet is 18 gauge cold-rolled steel, powder coated for durability and a high quality appearance that resists scratches and chipping.
• 2.0" (5.1cm) thick, non-CFC, foamed-in-place polyurethane cabinet insulation provides maximum product thermal protection and increased rigidity for reliable daily use.
• Phenolic-coated evaporator protects against corrosive by-products or agents in the chamber atmosphere (e.g., secretions from insects such as flies), ensuring long life.
• Solid, stainless steel interior is more durable and corrosion-resistant than plastic or painted metal, ensuring long life and minimizing equipment costs.
• Heavy-duty, solid, stainless steel shelves are easy-to-clean, saving time and effort; more corrosion-resistant than coated wire shelves for long life; and adjustable for convenience.
• 29.0 cu. ft. models include 6 shelves that are adjustable on 3.0" (7.6cm) centers.
• 11.0 cu. ft. model includes 3 shelves that are adjustable on 2.0" (5.1cm) centers.
• Reinforced floor/shelf components for 29.0 cu. ft. models are available so you can use shakers, cell rollers, or other heavy equipment or product loads inside the chamber to meet a variety of applications.
• Swivel, locking casters ensure easy mobility during installation and cleaning. Leveling feet provide stability and added safety in the lab.
See page 11 for order numbers, model compatibility and additional information.4
Broad range of options create a perfect fit for your applications
Many options are available so you can tailor this series to your specific applications.
Factory-installed options (ordered separately) include:
• Choice of access ports for inserting data logging sensors or cables.
• Convenience receptacles to run electrical devices in chamber.
• Choice of chart recorders for data logging.
• Choice of data outputs for monitoring/alarm (3960 / 3961 models only).
• Infrared CO2 Control Package for control of a CO
2 environment.
• Heatless dryer for low humidity conditions – check temperature / humidity chart (see page 5).
• Door light package for plant growth, seed germination, algae, small animal, insect studies.
• Stainless steel door glass for light sensitive applications.
• Door options include door lock, left-hand door, and Lexan™ inner door.
• Choice of shaker support systems provides system safety.
• Temperature mapping and calibration certificate – customized.
Accessories/on-site options (ordered separately) include:
• Extra and special shelving.
• Reinforced floor options for heavy weights – up to 150 lb.
• Sensaphone™ telephone dialing systems – interfaces with standard touch-tone phone.
• Cell roll system for production of monolayer cell cultures (see page 7).
• On site IQ/OQ/PQ validation services.
Options for the humidity system:
• Condensate pump kit: recommended when the accessible drain on lab wall is well above the drain on back of environmental chamber.
• Condensate evaporator: external heating device which burns off water that drips out the back of the unit when there is no drain available.
Chamber Specifications
Temperature/Humidity Performance Chart for 3940 (3949) and 3911 (3913)
The following chart illustrates the maximum and minimum attainable humidity within the cabinet at a given set temperature under the following nominal conditions: refrigeration on, defrost off, no wet bulb recorder, no air exchange, set for 22.2°C (72°F) and 50% RH ambient. Optional heatless dryer is available for lower humidity levels.
*Six shelves in 3920 and 3940 (3949), three shelves in 3911 (3913); uniformity is ±0.5°C at all other temperature parameters.**All models require a drain unless fitted with the optional Condensate Evaporator (No. 1900031). 3940 (3949) and 3911 (3913) also require a constant water source.
Thermo Scientific Cat. No. 3920 ** 3940 ** 3949 ** 3911 ** 3913**
Description29 cu. ft. Incubated/Refrigerated Environmental Chamber
29 cu. ft. Incubated/Refrigerated Environmental Chamber with humidity control
11 cu. ft. Incubated/Refrigerated Environmental Chamber with humidity control
Voltage 120V, 60 Hz 208-220V, 60 Hz 220-240V, 50/60 Hz 208-220V, 60 Hz 220-240V, 50/60 Hz
Temperature
Control ±0.1°C @ 25°C to 37°C (77°F to 98.6°F) ±0.1°C @ 25°C to 37°C (77°F to 98.6°F) ±0.1°C @ 25°C to 37°C (77°F to 98.6°F)
Range 0°C to 60°C (32°F to 140°F) 0°C to 60°C (32°F to 140°F) 0°C to 60°C (32°F to 140°F)
Sensor RTD RTD RTD
Controller Microprocessor, PID Microprocessor, PID Microprocessor, PID
Programmability Ramp & Soak (to 40 steps) Ramp & Soak (to 40 steps) Ramp & Soak (to 40 steps)
Setpoint Digital Digital Digital
Display Digital dual LED Digital dual LED Digital dual LED
Readability & Setability 0.1°C 0.1°C 0.1°C
Uniformity* ±0.3°C @ 25°C to 37°C (77°F to 98.6°F) ±0.3°C @ 25°C to 37°C (77°F to 98.6°F) ±0.3°C @ 25°C to 37°C (77°F to 98.6°F)
Temperature Safety
Sensor & Controller Thermostat Thermostat Thermostat
Setpoint Analog reference dial Analog reference dial Analog reference dial
Alarm Audible/visual Audible/visual Audible/visual
Refrigeration
Compressor 1⁄4 HP, air-cooled 1⁄4 HP, air-cooled 1⁄4 HP, air-cooled
Refrigerant Non-CFC, R134A refrigerant Non-CFC, R134A refrigerant Non-CFC, R134A refrigerant
Humidity
Input Water Quality 50K-1Megohm resistance 50K-1Megohm resistance
Input Water Volume 1 gallon/hour 1 gallon/hour
Input Water Pressure Gravity to 40 psi Gravity to 40 psi
RH Above ambient to 95% @ 37°C (98.6°F) ±5.0% RH Above ambient to 95% @ 37°C (98.6°F) ±5.0% RH
Sensor Capacitive – thin film polymer Capacitive – thin film polymer
Controller Microprocessor, PID Microprocessor, PID
Programmability Ramp & Soak (to 40 steps) Ramp & Soak (to 40 steps)
Setpoint Digital (direct set in % RH) Digital (direct set in % RH)
Display Digital, dual LED (direct read in % RH) Digital, dual LED (direct read in % RH)
Readability & Setability 1.0% 1.0%
Steam Generator Initial fill approx. 1 qt. (1 liter) Initial fill approx. 1 qt. (1 liter)
Setable Alarm Audible/visual Audible/visual
Certification cULus cULus CE cULus CE
Effective temperature and humidity range
Extended range with optional dryer
Maximum humidity
Minimum humidity
Optional Dryer
TEMPERATURE °C
% R
ELAT
IVE
HUM
IDIT
Y
5
*All models require a drain unless fitted with the optional Condensate Evaporator (No. 1900031). 3940 (3949) and 3911 (3913) also require a constant water source.
Chamber Specifications continued
Thermo Scientific 3920/3940/3911 Series Stability Chambers
Thermo Scientific Cat. No. 3920 ** 3940 ** 3949 ** 3911 ** 3913**
Description29 cu. ft. Incubated/Refrigerated Environmental Chamber
29 cu. ft. Incubated/Refrigerated Environmental Chamber with humidity control
11 cu. ft. Incubated/Refrigerated Environmental Chamber with humidity control
Fittings**
Fill Port 1⁄8” female NPT 1⁄8” female NPT
Drain Port (bottom) 3⁄8” FPT & 3⁄8” O.D. P-trap right rear 3⁄8” FPT & 3⁄8” O.D. P-trap right rear 3⁄8” FPT & 3⁄8” O.D. P-trap right rear
Unit Heat Load
115V 5500 BTUH (1600 Watts) 5500 BTUH (1600 Watts) 5500 BTUH (1600 Watts)
220V 6000 BTUH (1750 Watts) 6000 BTUH (1750 Watts) 6000 BTUH (1750 Watts)
Dimensions
Exterior
Width 38.0” (96.5cm) 38.0” (96.5cm) 38.0” (96.5cm)
Height 88.5” (224.8cm) 88.5” (224.8cm) 51.5” (130.8cm)
F-B 32.0” (81.3cm) 32.0” (81.3cm) 32.0” (81.3cm)
Interior
Width 31.0” (78.7cm) 31.0” (78.7cm) 31.0” (78.7cm)
Height 60.0” (152.4cm) 60.0” (152.4cm) 24.0” (60.9cm)
F-B 27.0” (68.6cm) 27.0” (68.6cm) 27.0” (68.6cm)
Shelves
Dimensions 30.6”W x 25.8” F-B (77.8cm x 65.6cm) 30.6”W x 25.8” F-B (77.8cm x 65.6cm) 30.6”W x 25.8” F-B (77.8cm x 65.6cm)
Construction Solid stainless steel, reinforced Solid stainless steel, reinforced Solid stainless steel, reinforced
Surface Area per Shelf 5.4 sq. ft. (0.5 sq. m) 5.4 sq. ft. (0.5 sq. m) 5.4 sq. ft. (0.5 sq. m)
Max. Surface Area per Chamber
104.3 sq. ft. (9.7 sq. m) 104.3 sq. ft. (9.7 sq. m) 59.4 sq. ft. (5.5 sq. m)
Standard/Maximum Number of Shelves
6, 19 6, 19 3, 11
Load per Shelf35 lbs. (15.9 kg), slide in/out 50 lbs. (22.7 kg), stationary
35 lbs. (15.9 kg), slide in/out 50 lbs. (22.7 kg), stationary
35 lbs. (15.9 kg), slide in/out 50 lbs. (22.7 kg), stationary
Construction
Interior Volume 29.0 cu. ft. (821.2 liters) 29.0 cu. ft. (821.2 liters) 11.0 cu. ft. (311.5 liters)
Interior Material Type 304, 2B finish, stainless steel Type 304, 2B finish, stainless steel Type 304, 2B finish, stainless steel
Exterior Material Cold-rolled steel, powder coated Cold-rolled steel, powder coated Cold-rolled steel, powder coated
Outer Door Gasket Four-sided, vinyl compression Four-sided, vinyl compression Four-sided, vinyl compression
Electrical 120V, 60 Hz, 16.0 Amps 3 Wire (including ground)
208-220V, 60 Hz, 15.0 Amps 4 Wire (including ground) single phase
220-240V,50/60 Hz 14.0 Amps, 3 Wire (including ground) must be hardwired to a lockable fused disconnect switch
208-220V, 60 Hz 15.0 Amps 4 Wire (including ground) single phase
220-240V,50/60 Hz 14.0 Amps, 3 Wire (including ground) must be hardwired to a lockable fused disconnect switch
Dry Contacts Common, NO, NC Common, NO, NC Common, NO, NC
Data Output 4-20 milliamp, temperature 4-20 milliamp, temperature and RH 4-20 milliamp, temperature and RH
Power Switch 2 Pole 2 Pole 2 Pole
Plug NEMA 5-20P NEMA 14-20P None NEMA 14-20P None
Weight
Net 700 lbs. (317.5 kg) 765 lbs. (347.0 kg) 575 lbs. (260.8 kg)
Shipping (motor) 850 lbs. (385.6 kg) 1,040 lbs. (471.7 kg) 940 lbs. (426.4 kg) 715 lbs. (324.3 kg)
6
NEMA 5-20 NEMA 14-20PNEMA 5-20 NEMA 14-20P NEMA 5-20 NEMA 14-20P
Incubator/Environmental Chamber for Elevated Temperature Applications The 29.0 cu. ft. (821.2 liters) 3960 (3961) offers precise temperature control in a range of 5°C above ambient to 60°C (140°F), making this unit ideal for clinical applications, elevated temperature stability
3960 (3961) Models
• Large size for high volumes and a broad range of products
• Heated glass door for minimal condensation and a clear view
• Interior and exterior convenience receptacles, and five adjustable shelves
• Thermo Scientific Enviro-Scan™ controls with easy-to-read display of actual temp studies, and shelf life testing.
Cell Roll SystemThe optional cell roll system allows extensive production of monolayer cell cultures in standard roller culture vessels. Oxygenation and exposure of the cells to the media growth area are improved. Culture yields are increased by the uniform temperature control and cell roll system’s continuous, gentle rotation.
The environmental chamber accommodates a cell roller up to 7 decks high with 5 positions per deck for a maximum total of 35 positions, or bottles.
Achieving maximum capacity requires a 4862 (4868) three tier cell roller base (15 positions), four add-on tiers (20 positions), and a reinforced floor/ramp.
All position drive is standard. Adjustable speed control provides precise speeds of 0.125 to 6.25 RPM with ±1.0% accuracy, based on 110mm bottles.
Cell Roller System
Model 3961
7
Uniformity and an Elevated Temperature Range for Easy Compliance with Regulatory Requirements
Our directed airflow system promotes an ideal growth environment. The proven design includes a positive pressure feed plenum and a negative pressure return plenum. This combination directs air across the surface of each solid shelf. Even when filled with samples or equipment, each shelf receives a consistent flow of conditioned air for optimum temperature uniformity and recovery. By design, the feed plenum cannot be blocked by the chamber’s contents. (Refer to the diagram on page 3.)
Intuitive, powerful Enviro-Scan– a key Forma advantage > ensures precise temperature control without
complicated programming
> provides audible/visual under-temperature and over-temperature alarms for peace-of-mind
> includes an easy-to-read display for convenient, continuous monitoring.
• Enviro-Scan operating modes include Run, Setpoint, Calibration, and System Configuration.
• Standard remote alarm contacts and available data outputs allow connection to an in-house monitor/alarm system to track chamber conditions, helping you meet internal and regulatory documentation requirements. Chart recorders are available.
8
Durable construction and shelving for long life
• Exterior cabinet is 18 gauge cold-rolled steel, powder coated for durability and a high quality appearance that resists scratches and chipping.
• Maximum product thermal protection and added cabinet strength for reliable daily use are provided by the 2.0" (5.1cm) thick fiberglass cabinet insulation.
• Solid, stainless steel interior is more durable and corrosion-resistant than plastic or painted metal, ensuring long life and minimizing equipment costs.
• Five heavy-duty, solid, stainless steel shelves, adjustable on 2.0" (5.1cm) centers, are easy-to-clean, saving time and effort; more corrosion-resistant than coated wire shelves for long life; and adjustable for convenience.
• Reinforced floor/shelf components are available so you can use shakers, cell rollers, or other heavy equipment or product loads inside the chamber to meet a variety of applications.
Valuable features included for cost- effectiveness and flexibility
Casters, access ports, convenience receptacles, heated glass door, and remote alarm contacts – all are included up-front. Our 3960 Incubator/Environmental Chamber is prepared to meet your changing application needs without the added cost of expensive add-on options.
• Swivel, locking casters ensure easy mobility during installation and cleaning. Leveling feet provide stability and added safety in the lab.
• Thru-wall access ports, located on the right and left sides of the chamber, make it possible to add probes, sensors, etc. without altering the cabinet.
• Interior and exterior accessory receptacles provide a convenient power source.
– Interior GFCI duplex receptacle on the 115V 3960 unit (Single European 230V CEE 7 on 3961) located in the upper right corner of the rear wall, permits the use of shakers, cell rollers, and other equipment inside the chamber, eliminating the inconvenience of an extra external power strip.
– Exterior receptacle, located on the upper-right side of the control panel, is available for connecting an optional recorder or other Forma accessories.
9
Specifications are based on nominal voltages of 115V or 230V in ambients of 22ºC to 25ºC. Performance is plus or minus the least significant digit unless otherwise specified.
Thermo Scientific Cat. No. 3960 3961
Description 29 cu. ft. Incubated Environmental Chamber 29 cu. ft. Incubated Environmental Chamber
Voltage 115V, 50/60 Hz 200-230V, 50/60 Hz
Temperature
Control ±0.1°C ±0.1°C
Range 5°C above ambient to 60°C (140°F) 5°C above ambient to 60°C (140°F)
Uniformity ±0.4°C @ 30°C (86°F) ±0.3°C @ 37°C (98.6°F) ±0.5°C @ 45°C (113°F)
±0.4°C @ 30°C (86°F) ±0.3°C @ 37°C (98.6°F) ±0.5°C @ 45°C (113°F)
Low Alarm User programmable and indicator User programmable and indicator
Over-temperature
Sensor Precision thermistor Precision thermistor
Setability 0.1°C 0.1°C
Function User programmable, action (shuts off heat), and indicator User programmable, action (shuts off heat), and indicator
Temperature Safety (Cabinet)
Sensor Independent thermostat Independent thermostat
Type Independent analog electronic Independent analog electronic
Fittings
Access Port With Stopper 2.4” (6.1cm) I.D., one on each side 2.4” (6.1cm) I.D., one on each side
Unit Heat Load
115V/230V 510 BTUH (150 Watts) 129 kcal per hour 510 BTUH (150 Watts) 129 kcal per hour
Shelves
Dimensions 30.6”W x 25.8”F-B (77.7cm x 65.5cm) 30.6”W x 25.8”F-B (77.7cm x 65.5cm)
Construction 2B finish, solid stainless steel 2B finish, solid stainless steel
Surface Area 5.4 sq. ft. (0.5 sq. m) per shelf 5.4 sq. ft. (0.5 sq. m) per shelf
Max. per Chamber 140.4 sq. ft. (13.0 sq. m) 140.4 sq. ft. (13.0 sq. m)
Standard/Maximum No. of Shelves 5, 26 5, 26
Loading 30 lbs. (13.6 kg) per shelf, fully inserted and stationary; not to exceed 300 lbs. (136.0 kg) total per cabinet
30 lbs. (13.6 kg) per shelf, fully inserted and stationary; not to exceed 300 lbs. (136.0 kg) total per cabinet
Construction
Interior Volume 29.0 cu. ft. (821.2 liters) 29.0 cu. ft. (821.2 liters)
Interior Material Type 304, 2B finish, stainless steel Type 304, 2B finish, stainless steel
Exterior Material Cold-rolled steel, powder coated Cold-rolled steel, powder coated
Insulation 2.0” (5.1cm) fiberglass 2.0” (5.1cm) fiberglass
Exterior Door Heated, triple pane, tempered glass Heated, triple pane, tempered glass
Outer Door Gasket Molded vinyl Molded vinyl
Electrical
115V, 50/60 Hz, 9.0 Amps (Operating range 90-125V includes voltage fluctuations)
200/230V, 50/60 Hz, 5.0 Amps (Operating range 180-250V includes voltage fluctuations)
Circuit Breaker/Power Switch 15 Amps; 2 pole 8 Amps; 2 pole
Exterior Convenience Receptacle 75 Watts maximum (matches cabinet voltage) 75 Watts maximum (matches cabinet voltage)
Interior Convenience Receptacle 230 Watts maximum (matches cabinet voltage) 230 Watts maximum (matches cabinet voltage)
Plug 115V: NEMA 5-15P Plug 230V: CEE 7/7 Plug
Alarm Contacts Temperature deviation and power failure; Common, NO, and NC; customer connections through RJ11 jack on rear of unit
Temperature deviation and power failure; Common, NO, and NC; customer connections through RJ11 jack on rear of unit
Data Outputs (opt.) RS-485, 0-1V, 0-5V, 4-20 milliamp RS-485, 0-1V, 0-5V, 4-20 milliamp
Dimensions
Exterior 38.0”W x 80.0”H x 33.0”F-B (96.5cm x 203.2cm x 83.8cm) 38.0”W x 80.0”H x 33.0”F-B (96.5cm x 203.2cm x 83.8cm)
Interior 31.0”W x 60.0”H x 27.0”F-B (78.7cm x 152.4cm x 68.6cm) 31.0”W x 60.0”H x 27.0”F-B (78.7cm x 152.4cm x 68.6cm)
Weight
Net 500 lbs. (226.8 kg) 500 lbs. (226.8 kg)
Shipping (motor) 660 lbs. (299.4 kg) 660 lbs. (299.4 kg)
Certification cULus CE
Chamber Specifications
10
NEMA 5-20 NEMA 14-20P
NEMA 5-20 NEMA 14-20P
Thermo Scientific Cat. No DescriptionShelving
Stainless Steel Shelves for 3920, 3940 (3949), 3911 (3913), and 3960 (3961)
224139 Solid Stainless Steel Shelf and Channels for 3920, 3940 (3949), 3911 (3913), and 3960 (3961)
224155 Perforated Stainless Steel Shelf and Channels for 3920, 3940 (3949), 3911 (3913), and 3960 (3961)
224161 Reinforced Stainless Steel Shelf and Channels for 3920, 3940 (3949), 3911 (3913), and 3960 (3961); two per unit max., not for shakers; 150 lb. (68.0 kg) load maximum with shelf fully inserted and stationary
Shaker Support System, factory installed
1900005* Two shelf shaker support system without duplex outlet, each shelf will hold 200 lbs., for 3920, 3940, (3949) and 3960 (3961). Must be ordered with duplex outlet (505099 or 505094)
190761* Three shelf shaker support system includes duplex outlet, each shelf will hold 200 lbs., for 3920, 3940, (3949) and 3960 (3961)
Door Options, factory installed
190239 Lexan® Inner Door Kit (five doors) for 3920, 3940 (3949), and 3960 (3961); factory installed
190514 Door Lock Assembly for 3920, 3940 (3949), and 3960 (3961); factory installed
190597 Left-Hand Door Swing Kit for 3920 and 3940 (3949); factory installed
Stainless Steel Door Glass Covers
190591 For 3920, 3940 (3949), and 3960 (3961); factory installed
190892 For 3920, 3940 (3949), and 3960 (3961)**
Door Light Package, factory installed
190032 (220V, 60Hz)Door light package with 8 fluorescent light bulbs for plant growth, seed germination, algae, small animal and insect studies for 3920, 3940 and (3949); A 24-hour timer enables automated light on/off cycles. Intensity of light can be modified in 4 steps by enabling / disabling from 2 to 8 lights.
1900604 (120V, 60Hz)
1900605 (220V, 50Hz)
Reinforced Floor Options
500182*** Reinforced Floor with Removable Ramp for 3920, 3940 (3949), and 3960 (3961); accommodates a cell roller
190777*** Same as No. 500182 but factory installed
Access Ports, factory installed
505101 Port for 3920, 3940 (3949), and 3911 (3913); 1.5" (3.8 cm) I.D.; on center right wall
193004 2" nominal ID port with cover
190164 Port for 3960 (3961), 2.4" (6.1 cm) I.D., customer specified location
193005 4" nominal ID port with cover
193006 6" nominal ID port with cover
Convenience Receptacles, factory installed
505099 120V Duplex (includes separate line cord) for 3920, 3940, and 3911; located on center rear wall
505094 230V (includes separate line cord) for 3949, and 3913; located on center rear wall
Data Outputs (select one) for 3960 (3961), factory installed
190523 RS-485 interface (compatible with 1535 Monitor/Alarm only)
190512 4-20 milliamp
190543 0-5V
190544 0-1V
Sensaphone Telephone Dialing Systems, interface with standard touch-tone phone
400047 For up to four input channels
400134 For up to eight input channels
Accessories (All are customer installed unless indicated otherwise)
Warranty
We confidently back our environmental chambers with a two-year parts and labor warranty.
* Rubber isolators replace casters for maximum stability.** Must be installed by qualified personnel.*** Ramp extends approximately 23.0” (58.4cm) in front of the chamber.****Specification sheets for protocol and field validations are available for review upon request by calling Pre-Sales at 866-984-3766 or emailing: [email protected].
Thermo Scientific Cat. No DescriptionChart Recorders and Paper
12", 6 Hour to 31 Day Digital Recorders for 3920, 3940 (3949), and 3911 (3913); draw own chart as they record; single or dual pen; 120V, 50/60 Hz; factory installed
198036 Left Side
198039 Right Side
198042 Top, adds 20.3" (51.6cm) to cabinet height
197054 Replacement Chart Paper, 100 per box
6", 7 Day Chart Recorders for 3920, 3940 (3949), and 3911 (3913); 0°C to 60°C (32°F to 140°F); 115V, 50/60 Hz
201144 Single Pen, factory installed201145 Single Pen201146 Dual Pen, factory installed
201147 Dual Pen
6", 7 Day Chart Recorders for 3960 (3961), 0°C to 60°C (32°F to 140°F), single pen
201155 115V, 50/60 Hz
201156 230V, 50/60 Hz
6" Chart Paper for Recorders, 50 per box
180006 For Single Pen, 0°C to 60°C (32°F to 140°F)
197030 For Single Pen, -10°C to 70°C (14°F to 158°F)
197075 For Single or Dual Pen, 0°C to 100°C (32°F to 212°F)
RH System Options
184062 Condensate Pump Kit for 3920, 3940 (3949), and 3911 (3913) Elevated Drains; includes separate line cord, 115V
1900031 Condensate Evaporator for 3920, 3940, and 3911; includes separate line cord; 115V
1900391 Condensate Evaporator for 3949 and 3913; includes separate line cord; 240V
1900139
Heatless Dryer Kit for 3940 (3949) and 3911 (3913), for sub-ambient chamber RH, factory installed; The heatless dryer requires an air supply of 90 psi capable of 10 cfm, at minimum. It is recommended that the air supply be greater than the minimum required by the dryer.
Deionization Cartridge Assembly for chambers with humidity and no DI water source
019-168-00 DI cartridge assembly includes DI cartridge, wall bracket, pressure regulator and gauge and 1/4" flexible tubing
585-036-00 Replacement DI cartridge
Calibration Certificate and Temperature Mapping with Report
260045 Temperature Mapping Report (factory installed)
260049 Calibration Certificate (specify test parameters)
cGxP Protocols and on site Validation Services****
GLP Level Protocols (IQ OQ)
IOQDOCE89003499 For 3920, 3960
IOQDOCE89003710 For 3940 (3949) 3911 (3913)
GMP Level Protocols (IQ OQ PQ)
IOPQDOCE89003499 For 3920, 3960
IOPQDOCE89003710 For 3940 (3949) 3911 (3913)
GLP Level Field Validation (IQ OQ) Performed at the customer site (protocol included at no extra charge)
IOQPCKE89003499 For 3920, 3960
IOQPCKE89003710 For 3940 (3949) 3911 (3913)
GMP Level Field Validation (IQ OQ PQ) Performed at the customer site (protocol included at no extra charge)
IOPQPCKE89003499 For 3920, 3960
IOPQPCKE89003710 For 3940 (3949) 3911 (3913)
Miscellaneous Accessories Infrared CO2 Control Packages, factory installed
1900226 For 3920
1900227 For 3940 (3949) and 3911 (3913)
Cell Roll System
4862 Three Tier Cell Roller Base (15 positions), 120V, 50/60 Hz, 29.8"W x 27.8"H x 24.4"F-B (75.7cm x 70.6cm x 62.0cm)
4868 Same as 4862 but 230V, 50/60 Hz
190049 Add-On Tier (5 positions), 29.8"W x 7.1"H x 24.4"F-B (75.7cm x 18.0cm x 62.0cm), customer installed
190777 Reinforced Floor with Removable Ramp, ramp extends 23.0" (58.4cm), factory installed
500182 Same as No. 190777 but customer installed
228077 Rotation Alarm System, includes annunciator jack, factory installed
228078 Battery Back-Up, provides 24 hours of power if a power failure occurs, factory installed
475560 110mm x 285mm Glass Bottles (4 per case)
11
BRFORMACHB 0715
Australia +61 39757 4300Austria +43 1 801 40 0Belgium +32 53 73 42 41China +800 810 5118 or +400 650 5118 France +33 2 2803 2180Germany national toll free 0800 1 536 376 Germany international +49 6184 90 6000
India toll free 1800 22 8374 India +91 22 6716 2200Italy +39 02 95059 552Japan +81 3 5826 1616Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Nordic/Baltic/CIS countries +358 9 3291 0200
Russia +7 812 703 42 15Spain/Portugal +34 93 223 09 18Switzerland +41 44 454 12 22UK/Ireland +44 870 609 9203USA/Canada +1 866 984 3766
Other Asian countries +852 2885 4613 Countries not listed +49 6184 90 6000
thermoscientific.com/chambers
© 2015 Thermo Fisher Scientific Inc. All rights reserved. Sensaphone is a registered trademark of the Sensaphone Company. Lexan is a registered trademark of the General Electric Company. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.