Specifications Scan: Full band 950 to 2150 MHz scan or down to 160MHz at full zoom Range: RF input level range 40 dBuV to 90 dBuV Optical range +6 to -25 dBm. Accuracy: typically +/- 1 dB. MER: (SNR) Pass 8dB or more. Marginal 6-7 dB fail less than 5 dB BER: Pass must be greater than 1E-3 Input: 75 ohm BNC . BNC to F adaptor supplied . Optical type MC DiSEqC : Version 1.2 compatible, any DISEqC can be generated. On screen battery indicator: Battery life 5 hours when powering a typical LNB Charge time: 12 hours with auto regulation to ensure the unit does not overcharge Weight: 500 g Size: 243 mm by 125 mm by 55 mm deep Favourite channels: Storage of up to 99 favourites Accessories supplied: Mains charger BNC to F adaptor & Carrying cover. Global Invacom Ltd Winterdale Manor, Southminster Road, Althorne, Essex CM3 6BX. tel +44(0)1621 743440 fax +44(0)1621 743676 [email protected] 8 Optiscan Mk 2 Manual iss1

Upload: others

Post on 04-Jul-2020

14 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

Specifications

Scan: Full band 950 to 2150 MHz scan or down to 160MHz at full zoom

Range: RF input level range 40 dBuV to 90 dBuV Optical range +6 to -25 dBm.

Accuracy: typically +/- 1 dB.

MER: (SNR) Pass 8dB or more. Marginal 6-7 dB fail less than 5 dB

BER: Pass must be greater than 1E-3

Input: 75 ohm BNC . BNC to F adaptor supplied . Optical type MC

DiSEqC : Version 1.2 compatible, any DISEqC can be generated.

On screen battery indicator: Battery life 5 hours when powering a typical LNB

Charge time: 12 hours with auto regulation to ensure the unit does not overcharge

Weight: 500 g Size: 243 mm by 125 mm by 55 mm deep

Favourite channels: Storage of up to 99 favourites

Accessories supplied: Mains charger BNC to F adaptor& Carrying cover.

Global Invacom Ltd

Winterdale Manor,Southminster Road,Althorne, EssexCM3 6BX.

tel +44(0)1621 743440 fax +44(0)1621 743676 [email protected]

8

Optiscan Mk 2 Manual iss1

Page 2: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

The Optiscan Mk 2 Operators Manual

Thank you for buying the Optiscan Mk 2 Now you will be able to -

Measure optical power in dBm whilst viewing the spectrum

View the full satellite band in real-time ensuring that all transponders are available

Identify the channel and position, which is read from the satellite

Measure Level, SNR and BER at the touch of a button

Store all satellite settings as favourites in the memory

Use DiSEqC commands for the setting of motorised dishes and switches

Page 3: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

Operation

Using the stored favourite channels

The high eficiency backlight is on during operation

The ON button also acts as the UP key.

If the meter is left running with no button pushes detected it will automatically switch off after 10 minutes.The Digiscan when switched on recalls the favourite channel which was last used.

To select another favourite channel press the LEFT or RIGHT arrow key. When the favour-ite channel is selected press ENTER to go to the scan screen shown below.

The settings for gain frequency and bandwidth are all preset by the favourites memory, and the desired signal can now be searched for. Adjust the dish for maximum signal - the higher the peak the better the signal.When measuring optical signals the optical power is shown adjust for maximum reading,

Note The spectrum shown when in the optical mode is of the vertical low band only.

To change the sweep width use the UP and DOWN arrow keys to zoom in and out. Using these keys means that you can view the complete satellite band in one scanthese keys means that you can view the complete satellite band in one scanthese keys means that . This is ideal for identifying faulty cables and LNBs.

The centre frequency is set using the LEFT and RIGHT arrow keys. The figures at the lower left and right show the start and finish frequencies with the centre frequency in the middle below the dotted cursor.

2

Page 4: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

The custom menu allows any DiSEqC command to be generated in the Setup menu and subsequently stored in commands 1 to 5 in the custom menu.

Use the four arrow keys to navigate the Setup menu.

To enter text and numbers. Select the number in which the the code is to be stored(1-5) and press ENTER. Set the number of bytes with the RIGHT and LEFT arrow keys.

Press ENTER and the text line may now be entered. Select the letters using the UP/DOWN arrow keys. Press ENTER to input the code in the same way. Press ENTER once more and the code will be stored.

A four way DisSEqC switch control is loaded as shown in the screen above the outputs can be selected by setting the required switch position(1 to 4) and pressing ENTER.

Setting the custom menu (SETUP) requires a knowledge of the required DiSEqC com-mands and these can be obtained either from equipment manufacturers or the Eutelsat website:-http://www.eutelsat.com

7

Page 5: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

Finding a new channel and storing it as a favourite

To find and add your own channel to the favourites list, you will need know the re-quired centre frequency of the channel and the polarity of the signal (horizontal high or low or vertical high or low). All other settings are adjusted automatically by the Digiscan.

In the MENU screen select LNB volts. This brings up the screen shown below.

Select the required setting with the UP and DOWN keys Press ENTER to go to the scan screen and set the chosen LNB voltage.

Use the arrow keys on the scan to set the required centre frequency of the satellite.

Set up the dish to the expected angle and adjust it for the best signal level. Check that the desired signal is on the centre dotted line in the middle of the screen.

Adjust the GAIN controls on the Optiscan to bring the signal level on screen with the peak of the signal about three quarters of the way up the screen for comfortable viewing.

Press ENTER which brings up the MEASURE screen.

4

Page 6: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

Other features

In the setup MENU options selecting sound ON/ OFF emits a tone which increases in fre-quency as the signal level increases.

If any of the original factory preset favourites are accidentally overwritten, the setup menu enables them to be reset. These favourites occupy the first 20 channels. N.B. all added fa-vorites and new DiSEqC settings will be erased!

The MEASURE screen shows all the parameters of the signal. If LOCK is not achieved, the frequency may be changed by using the LEFT and RIGHT arrow keys.

When LOCK has been achieved, the identification will be read from the satellite; this can take up to 30 seconds as this data is not transmitted continuously.

PASS on screen indicates that SNR (MER) and BER are good.

If MARGINAL or FAIL is indicated then check both the SNR and BER figures. Note that the BER reading is Pre Viterbi which is the uncorrected data and shows the quality of the signal before any corrections are made. This gives the most critical indication of the quality of the installation.

These settings may now be stored as a favourite for future use. To store a new favour-ite go to MENU and select store favourite, the screen below will be shown. Select new favourite using the RIGHT or LEFT arrow keys (up to 99 can be stored) use an unused favorite position and press ENTER . The favourite is then stored for future use and can be selected from the start up favourites screen. Note favorites may be stored over exist-ing favorites , however when a factory reset is made the first 20 favorites will be over-written

5

Page 7: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

Select motor scan from the DiSEqC menu and press ENTER to go to the SCAN mode.

Use the LEFT and RIGHT arrow keys to move the dish to EAST or WEST. Press EN-TER to go to the MEASURE screen to check the signal quality. Pressing ENTER again returns to the scan screen. Pressing the UP key returns the dish to position A and pressing the down key sets the dish to position B.

The motor position may be saved and subsequently recalled by using SAVE MOTOR POSITION from the DiSEqC options menu shown above. Limits and centring can also be set in the DiSEqC menu shown below.

A simple A/B switch for multi-cable systems may also be set from the menu below.

Setting up a DiSEqC motorised dish

In the MENU screen select DiSEqC and the menu will appear as shown below.

6

Page 8: Optiscan MK 2 manual CS2 - Shop Online MK 2 manual.pdf · Optiscan Mk 2 Manual iss1. The Optiscan Mk 2 Operators Manual Thank you for buying the Optiscan Mk 2 Now you will be able

The MEASURE screen shows all the parameters of the signal. If LOCK is not achieved the frequency may be changed by using the LEFT and RIGHT arrow keys.If required, press ENTER to return to the scan screen.

The bar graph at the bottom of the screen shows the signal level. The level is also shown in dBuV at the end of the bar graph. When the signal has been found, LOCK will appear on the top right of the screen.

When LOCK has been achieved, the identification will be read back from the satellite. Read back can take up to 30 seconds as this data is not transmitted continuously.

PASS on the screen indicates that SNR (MER) and BER are good.

If MARGINAL or FAIL is indicated, check both the SNR and BER figures as one or both of these will be incorrect. More adjustment to the dish can be made to get the best possible readings

To select another favourite press MENU to go to the screen shown below.

The Measure Screen

3

The gain keys are used to vertically centre the display on the screen. When the highest signal level has been found by adjusting the dish, press ENTER to go to the MEASURE screen