overview of ccd intensity and brightness distribution · brightness distribution on back-light is...

6
Measurement method for back-light/ coaxial light data Brightness distribution Luminosity characteristics Measuring luminosity change Luminosity graph Brightness distribution Uniformity of brightness is an important point when selecting a back- light and a coaxial light. To ensure that our products are of high quality, we perform luminosity measurements. As shown in the diagrams, luminosity is measured at 5-mm intervals along the X0-X1 and Y0-Y1 lines and recorded. Measurement points and luminosity changes are indicated. Differences in the characteristics of each light due to the type of LED used and the manufacturing method are clear. The lower diagram is a graph showing luminosity changes in the LDL-TP-63×60 and LFL-4012. The LDL-TP Series achieves high luminosity and high uniformity by mounting chip LEDs on a flat circuit board. The LFL Series features light emission using the light conduction method. A dim light is produced that is very even and does not have high luminosity. In this way, luminosity characteristics provide a guideline for selecting a light that matches the application. We perform measurements of brightness distribution of back- lights at an LWD of 0 mm. This information in combination with luminosity graphs helps provide a good understanding of the characteristics of each light. The diagram shows the brightness distributions of the LDL-TP-63×60 and LFL-4012. The LDL-TP shows a broad, even distribution around the center. The LFL clearly shows the characteristics of the light conduction method. Please use the brightness distributions and luminosity graphs as an aid in selecting the optimum light for your application. Devices used for testing Luminosity meter Power unit Two dimentional brightness distribution measuring system Minolta LS-110 PSD-3012-WW,PSD-3024-WW(CCS Co.) (CCS Co.) Test conditions Measurement start time Measurement distance Measurement interval Approximately 30 minutes after turning on power of LED light 180mm X0-X1, Y0-Y1 measured at 5-mm intervals Brightness distribution Devices used in testing Test conditions This curve shows the relationship between CCD intensity and LWD. The maximum value of CCD intensity is shown on the Y axis as 100%. In the case of the lower diagram, CCD intensity reaches a maximum when the LWD is approximately 40 mm. This maximum CCD intensity is shown on the graph and can be used as an indication of the brightness of each light. The two-dimensional brightness distribution measuring system selects three points of work distance (LWD) that are recommended as factory standard. The three points selected are those that are useful for understanding the characteristics of a light. Of these three points, the point with the highest brightness level is taken to be 100% and the other points assigned percentage values accordingly. Ten levels are set and expressed by contour lines and color. The diagram at right compares the brightness distributions of the LDR2-90RD. It can be seen that the center point at an LWD of 50 mm is the brightest of the three points. Moving outward from the center, the brightness decreases one level at a time and the different levels are indicated by color. At distances of LWD 100 to 150 mm, the number of levels decrease and the field becomes more uniform. In this way, it can be determined how the light strength and the illumination range change with varying LWD, and the result can be used as a guideline for selecting the most suitable LWD. CCD intensity vs. LWD CCD camera Power unit Two dimentional brightness distribution measuring system Reference plate NC300 (Takenaka System Device Co.) f 50 mm lens, 5-mm close-up ring, lens opening of 17, 1.1 focus PSD-3012-WW,PSD-3024-WW(CCS Co.) (CCS Co.) K3401 (Mitsubishi Plastic), white mat Measurement start time Measurement distance Measurement interval Camera position Approximately 30 minutes after turning on power of LED light 10 to 150 mm at 10 mm intervals Waited one minute after each move of 10 mm and then performed measurement 180mm 69 70 "CCD intensity" as used in our product literature refers the level of output from the CCD camera. The data provided in this literature is intended only as a guideline for light selection. It does not constitute a guarantee of the quality and performance of each light. Measuring brightness distribution Brightness distribution on back-light is measured using a two-dimensional brightness distribution measuring system that we independently developed. 400 light receptors are mounted at a pitch of 10 mm on a 200×200 mm receptor plate. These receptors detect the brightness of the light and the data is sent to a computer. For all points the computer converts the data into a brightness map, brightness pattern, and time-elapse chart, and stores this information. Two-dimensional brightness distribution measuring system As shown in the diagram, we use our own proprietary measurement unit to acquire and measure image signals from the CCD camera using a distance from the reference plate to the light (LWD) that is varied over a range of 10 to 150 mm. Measuring brightness distribution Comparing brightness distributions Two-dimensional brightness distribution measuring system Sketch of two dimensional brightness distribution measuring system 200mm 200mm 10mm LWD 5~150mm LED light Luminosity change in the x direction Luminosity change in the y direction 5 4 3 2 1 0 0 10 20 30 40 50 Distance (mm) 5 4 3 2 1 0 0 10 20 30 40 Distance (mm) Luminosity change in the x direction Luminosity change in the y direction 18 16 14 12 10 8 6 4 2 0 0 10 20 30 40 50 60 Distance (mm) 18 16 14 12 10 8 6 4 2 0 0 10 20 30 40 50 Distance (mm) LDL-TP-63x60 LFL-4012 X0 Y0 Y1 X1 5mm Emitting surface 180mm Luminosity meter Emitting surface LED light LDR2-90RD CCD intensity change characterstic VS distance to work piece. 100 80 60 40 20 0 0 25 50 75 LWD mm (Light Working Distance) 100 125 150 CCD intensity % (max: 100) Maximum output of CCD camera: 1.44V 180mm 10~150mm CCD camera LED light Standard white plate Luminosity (100 cd/m2) Luminosity (100 cd/m2) Luminosity (100 cd/m2) Luminosity (100 cd/m2) Overview of CCD Intensity and Brightness Distribution LDL-TP-63×60 LFL-4012 Brightness level 100 10 10mm 10mm Brightness level 100 10 LWD:50mm 10mm LWD:100mm 10mm LWD:150mm 10mm Brightness level 100 10 Sketch of two dimensional brightness distribution measuring system 200mm 200mm 10mm LWD LED light 0mm Brightness distribution is measured using a two- dimensional brightness distribution measuring system that we independently developed. 400 light receptors are mounted at a pitch of 10 mm on a 200×200 mm receptor plate. These receptors detect the brightness of the light and the data is sent to a computer. For all points the computer converts the data into a brightness map, brightness pattern, and time-elapse chart, and stores this information. Data measuring method of direct light /surface light emission ring illumination LWD characteristics of CCD luminance Measuring LWD characteristics of CCD luminance CCD luminance change characteristics in relation to the distance from the workpiece (LWD), distribution of brightness, and change of luminance are shown in a graph form as a guide for selecting illuminations.

Upload: dangphuc

Post on 20-Apr-2018

219 views

Category:

Documents


4 download

TRANSCRIPT

Measurement method for back-light/ coaxial light data

Brightness distributionLuminosity characteristics

Measuring luminosity change

Luminosity graph

Brightness distribution

Uniformity of brightness is an important point when selecting a back-light and a coaxial light. To ensure that our products are of high quality, we perform luminosity measurements. As shown in the diagrams, luminosity is measured at 5-mm intervals along the X0-X1 and Y0-Y1 lines and recorded.

Measurement points and luminosity changes are indicated. Differences in the characteristics of each light due to the type of LED used and the manufacturing method are clear.The lower diagram is a graph showing luminosity changes in the LDL-TP-63×60 and LFL-4012. The LDL-TP Series achieves high luminosity and high uniformity by mounting chip LEDs on a flat circuit board.The LFL Series features light emission using the light conduction method. A dim light is produced that is very even and does not have high luminosity.In this way, luminosity characteristics provide a guideline for selecting a light that matches the application.

We perform measurements of brightness distribution of back-lights at an LWD of 0 mm. This information in combination with luminosity graphs helps provide a good understanding of the characteristics of each light. The diagram shows the brightness distributions of the LDL-TP-63×60 and LFL-4012. The LDL-TP shows a broad, even distribution around the center. The LFL clearly shows the characteristics of the light conduction method. Please use the brightness distributions and luminosity graphs as an aid in selecting the optimum light for your application.

Devices used for testingLuminosity meterPower unit

Two dimentional brightnessdistribution measuring system

Minolta LS-110PSD-3012-WW,PSD-3024-WW(CCS Co.)

(CCS Co.)

Test conditionsMeasurement start timeMeasurement distanceMeasurement interval

Approximately 30 minutes after turning on power of LED light180mmX0-X1, Y0-Y1 measured at 5-mm intervals

Brightness distribution

Devices used in testing

Test conditions

This curve shows the relationship between CCD intensity and LWD. The maximum value of CCD intensity is shown on the Y axis as 100%.In the case of the lower diagram, CCD intensity reaches a maximum when the LWD is approximately 40 mm. This maximum CCD intensity is shown on the graph and can be used as an indication of the brightness of each light.

The two-dimensional brightness distribution measuring system selects three points of work distance (LWD) that are recommended as factory standard. The three points selected are those that are useful for understanding the characteristics of a light. Of these three points, the point with the highest brightness level is taken to be 100% and the other points assigned percentage values accordingly.Ten levels are set and expressed by contour lines and color. The diagram at right compares the brightness distributions of the LDR2-90RD.It can be seen that the center point at an LWD of 50 mm is the brightest of the three points. Moving outward from the center, the brightness decreases one level at a time and the different levels are indicated by color.At distances of LWD 100 to 150 mm, the number of levels decrease and the field becomes more uniform.In this way, it can be determined how the light strength and the illumination range change with varying LWD, and the result can be used as a guideline for selecting the most suitable LWD.

CCD intensity vs. LWD

CCD camera

Power unit

Two dimentional brightnessdistribution measuring system

Reference plate

NC300 (Takenaka System Device Co.)f 50 mm lens, 5-mm close-up ring, lens opening of 17, 1.1 focus

PSD-3012-WW,PSD-3024-WW(CCS Co.)

(CCS Co.)

K3401 (Mitsubishi Plastic), white mat

Measurement start timeMeasurement distanceMeasurement intervalCamera position

Approximately 30 minutes after turning on power of LED light10 to 150 mm at 10 mm intervalsWaited one minute after each move of 10 mm and then performed measurement

180mm

69 70

"CCD intensity" as used in our product literature refers the level of output from the CCD camera.

The data provided in this literature is intended only as a guideline for light selection. It does not constitute a guarantee of the quality and performance of each light.

Measuring brightness distributionBrightness distribution on back-light is measured using a two-dimensional brightness distribution measuring system that we independently developed.400 light receptors are mounted at a pitch of 10 mm on a 200×200 mm receptor plate. These receptors detect the brightness of the light and the data is sent to a computer. For all points the computer converts the data into a brightness map, brightness pattern, and time-elapse chart, and stores this information.

Two-dimensional brightness distribution measuring system

As shown in the diagram, we use our own proprietary measurement unit to acquire and measure image signals from the CCD camera using a distance from the reference plate to the light (LWD) that is varied over a range of 10 to 150 mm.

Measuring brightness distribution

Comparing brightness distributions

Two-dimensional brightness distribution measuring system

Sketch of two dimensional brightness distribution measuring system

200mm

200m

m

10mm

LWD5~150mm

LED light

Luminosity change in the x direction Luminosity change in the y direction5

4

3

2

1

00 10 20 30 40 50

Distance (mm)

5

4

3

2

1

00 10 20 30 40

Distance (mm)

Luminosity change in the x direction Luminosity change in the y direction1816141210

86420

0 10 20 30 40 50 60Distance (mm)

1816141210

86420

0 10 20 30 40 50Distance (mm)

LDL-TP-63x60

LFL-4012

X0

Y0

Y1

X15mm

Emittingsurface

180mm

Luminosity meter

Emittingsurface

LED light

LDR2-90RD CCD intensity change charactersticVS distance to work piece.

100

80

60

40

20

00 25 50 75

LWD mm (Light Working Distance)100 125 150

CC

D in

tens

ity %

(max

: 100

)

Maximum output of CCD camera: 1.44V

180mm

10~150mm

CCD camera

LED light

Standard white plate

Lum

inos

ity (1

00 c

d/m

2)

Lum

inos

ity (1

00 c

d/m

2)

Lum

inos

ity (1

00 c

d/m

2)

Lum

inos

ity (1

00 c

d/m

2)

Overview of CCD Intensity and Brightness Distribution

LDL-TP-63×60

LFL-4012

Brig

htne

ss le

vel

100

10

10mm

10mm

Brig

htne

ss le

vel

100

10

LWD:50mm

10mm

LWD:100mm

10mm

LWD:150mm

10mm

Brig

htne

ss le

vel

100

10

Sketch of two dimensional brightness distribution measuring system

200mm

200m

m

10mm

LWD

LED light

0mm

Brightness dis t r ibut ion is measured using a two-dimensional brightness distribution measuring system that we independently developed.400 light receptors are mounted at a pitch of 10 mm on a 200×200 mm receptor plate. These receptors detect the brightness of the light and the data is sent to a computer. For all points the computer converts the data into a brightness map, brightness pattern, and time-elapse chart, and stores this information.

Data measuring method of direct light/surface light emission ring illumination

LWD characteristics of CCD luminance

Measuring LWD characteristics of CCD luminance

CCD luminance change characteristics in relation to the distance from the workpiece (LWD), distribution of brightness, and change of luminance are shown in a graph form as a guide for selecting illuminations.

71 72

without with

With the conventional construction, the light was not able to dissipate heat with total effeciency due to the gap between the PCB and aluminum housing.By employing a special heat dissipating enclosure between the PCB and the housing in the new construction, there is substantial absorption of heat generation from LED, and efficient heat conductivity into the housing.This new construction of LDR2 suppresses the temperature rise of LED sharply, providing stable images for a long period of time.

Although the white LED ring light without heat dissipation has a range of fluctuation of about 14% over a period of 60 minutes at 100% light intensity, the new ring light, LDR2 maintains the highest intensity with a fluctuation range of only 2% over the same period.

Data Acquisition Environment:LED light is installed inside environmental chamger at 40 , and light intensity and temperature are measured (LED light is fixed to the aluminum plate of 5mm thickness).

While the light without heat dissipation material goes into a 'heat-balanced state'after 38 minutes with a temperature of 111.3 ,the new LDR2 ring light attains a 'heat-balanced state' after 42 minutes andtemperature is stabilized at only 57.2 .This shows the success of the LDR2 light in sharply suppressing temperature rise,compared with the light without heat dissipation.

Note: Less than 0.5 is a stable range.

With 120 white LED ring light, the temperature rise of the solder joint is measured with100% light intensity.Although the soldered part of the PCB of the light without heatdissipation material heated up to 106.6 after 60 minutes, the new LDR2 ring light wasmeasured at a maximum of 58.7 , representing a successful suppression of heat, adecrease of 45% less than the temperature of the earlier light in the same conditions.

LED Radiation Pattern

Light Spectrum

CCS has studied LED life cycles and intensity degeneration for many years. In particular, we have focused on the issue of "heat radiation and it's impact on the life of LED" and taken measures to extend the useful life of our standard LED arrays by incorporating life extending features. These features include manufacuturing the light cases out of aluminum, which has excellent heat conducting characteristics, as well as heat dissipating measures such as air connections and fans. Recently, we have taken this research to the next plateau by providing LED illumination that can be used for a longer period with ease by adding LDR2 LDM2 Series with our original heat dissipating architecture.

The following chart shows the relative directional characteristics. In general, the center shows the highest intensity with the intensity diminishingas the angle becomes lower.

Each LED type has a characteristic spectrum pattern which affects the sensitivity of CCD cameras. The optimal image can not be obtained if the light spectrum pattern and the CCD's sensitivity pattern are not well matched.

Directional pattern of Red LED N-type Directional pattern of Red LED W-type Directional pattern of White LED W-type Directional pattern of Blue and Green LED

AluminumHousing

PCB

LED

Radiation Fin

Heat DissipationMaterial

550 600 650 700 750 800 850 900 950 1000 1050500450Wavelength (nm)

4003503000

20

100

80

60

40

Rel

ativ

e em

issi

on in

tens

ity (%

)

UV light:365 nm

Green light:525 nm Infrared light: 850 nm

Red light: 660 nm

White light

Infrared light:940 nm

Blue light:470 nm

Ta=25

Technology OverviewTechnology Overview Technology OverviewTechnology OverviewAn introduction to the characteristics of LED lighting An introduction to the characteristics of LED lighting

Skillful use of LED lighting Skillful use of LED lighting

Using an LED light at high temperature shortens its life.LED lights generate heat that decreases luminosity and speeds degeneration. In particular, the luminosity of red LEDs decreases 1% for each degree (Celsius) the temperature rises. (The original luminosity is restored by cooling the LED.) Prolonged use at high temperature results in gradual degeneration and decreased luminosity. (The original luminosity is not restored by cooling the LED.)

To prevent decreased luminosity and degeneration due to heat Only turn light on during imaging

Strobe Power Supplies

Use the light with the light level turned as low as possible.

Install a fan Provide air flow Install on a bracket with good heat conductivity

PTU2-3012/PTU2-3024

As indicated in Figures 1 and 2, setting the light level to 50% keeps current flow low and therefore generates minimal heat, allowing you to prevent luminosity weakening. Reducing the current flowing to the light makes it possible to keep the amount of heat generated low. In general, we recommend a light level of 30% to 50%. If prolonged use has caused degeneration and reduced luminosity, you can raise the light level without concern.

Luminosity weakening can be reduced by installing a fan or providing air flow for heat dissipation. As indicating in Figure 3, luminosity decreases when the room temperature increases, and increases when the room temperature decreases.

LED lights are very durable with respect to switching. When a LED is used for strobe lighting or is only turned on as needed by means of an external signal, light intensity is stable and the life of the LED increases dramatically.

The standard CCS LED light can be used as a strobe light.The unit has two independent circuits so that the strobe Illumination time can be set from 10 sec to 990 sec.For more information, see P65

Measurement conditions and methodsMeasurement conditions

Measurement methods

Measuring instrument

Measurements performed with the light level set to maximum (100%) and 50%.Measurement time also included 24-hour continuous illumination.Digital power source PSD Series and analog power source PSB Series(EOL) are used as the power source.Ambient temperature is ordinary temprateure.

Light was reflected off of a standard white plate from a height of 50 mm and the luminosity(reflected light)at the center of the plate was measured.Once a week, the light was turned off for at least two hours to allow it to return to normal temperature, and then the light level was set to 100%, the light was turned on, and the luminosity was immediately measured.

Luminosity meter: Minolta LS-110Note: Although this measurement was held with LDR-50A.

We currently produce and sell LDR2-50 Series which has superior heat radiation characterstics.

Decrease deterioration of light intensity against heat generationStructure comparison LDR2-90SW vs. Light without heat dissipation

LDR2-120SW(white) vs. Light without heat dissipation LDR2-120SW (white) vs. Light without heat dissipation

Figure 1 Relation between light modulation and brightness(Heat radiation = n/a, Intensity Control = Max, LDR-50A)

Figure 3 Relation between intensity and room temperature.(Heat radiation = n/a, Intensity Control = Max, LDR-50A)

Figure 2 Relation between light modulation and brightness(Heat radiation = n/a, Intensity Control = 50% LDR-50A)

1.0

0.8

0.6

0.4

0.2

0400 500 600 700

Wavelength nm800 900 1000

Spe

ctra

l lum

inou

s ef

ficie

ncy

1.0

0.8

0.6

0.4

0.2

0400 500 600 700

Wavelength nm800 900 1000

Spe

ctru

m o

f CC

D s

ensi

tivity

Figure 2 Spectral sensitivity characteristics of a CCD camera

Figure 1 Standard spectral luminous efficiency curve

Figure 4 CCD sensitivity change

CCD sensitivity graph Test configuration

Figure 3 Brightness comparison graph

Spectral luminous efficiency and CCD sensitivity

Distance from the light to the work (the object of inspection)

Shape, condition, and color of the work

Size of the field of view

Obtaining an image with good contrast

908070605040302010

00 50 100 150 200 250

LWD mm (Light Working Distance)

Brig

htne

ss lx

(1000)

Green

Red

73 74

1.0

0.8

0.6

0.4

0.2

0400 500 600 700

Wavelength nm800 900 1000

Rel

ativ

e se

nsiti

vity

Spectral sensitivity characteristics (CCD only) Spectral sensitivity characteristics (with infrared filter)

Figure 1 shows a cross-section of a coaxial light (LFV Series). Light from the LED is reflected using a half-mirror, and thus the position of the emitting surface can be treated as if it is directly behind the mirror. In this case, the distance from the emitting surface to the work is called the "LWD'".The effective field of view of coaxial light is determined by 1) the LWD (distance from the light to the work) and 2) the WD (distance to the CCD camera). Figure 2 shows how to determine the field of view "V" when the WD is held constant and the LWD (distance to the light) is varied. The following is an explanation of what the effective field of view will be when the provisional emitting surface is at positions A and B.In the case of position A, if we assume that the work is a reflecting surface, we can say that there is an emitting surface at A opposite to the work (position A' of the LWD'a distance). Therefore, when the work piece is viewed through the camera, it appears as if the emitting surface is at A', and thus the effective field of view is Va.In the same way, in the case of B the emitting surface is at B' and the effective field of view is Vb. Comparing Va to Vb, we find that Va, which has the shorter LWD, has a greater effective field of view. In this way, the effective field of view grows as the LWD shortens.

<What is the effective field of view?>For example, when reading characters engraved on a shiny piece of metal, if we assume that the provisional emitting surface is at position B, Vb will be determined by the virtual provisional emitting surface position B'. For this reason, only the letters CDEFG will be visible as dark letters against a light background, and the letters AB and HI, which appear dark against a dark background, will not be discernible. In this way, the effective field of view Vb is smaller than the field of view V.

The human eye is most sensitive to a wavelength of 555 nm. As the wavelength grows longer or shorter, the human eye becomes less sensitive. (Although there is some variation depending on the person, most people can view a wavelength range of 380 nm to 760 nm.) This sensitivity of the eye to light is

called spectral luminous efficacy.Spectral luminous efficiency is a measure based on a value of 1 for the spectral luminous efficacy of 555 nm light. Figure 1 shows a standard spectral luminous efficiency curve for the human eye. Wavelengths longer than the visible range are called infrared light and wavelengths shorter than the visible range are called ultraviolet light.Figure 2 shows the spectral sensitivity characteristics of a typical CCD camera (NC300, Takenaka System Devices). It can be seen that sensitivity to infrared light is better than in Figure 1. In this way, there is a difference between the sensitivity of the human eye and that of a CCD camera.

Figures 3 and 4 show the relationship between brightness and distance for our LDR-50B (red) and GR (green) LED lights. Figure 3 shows the change of brightness, and the green light has a higher brightness than the red light. Figure 4 shows the change of CCD sensitivity, and in this case red has a higher brightness than green. (Brightness/luminosity meters are correlated to the spectral luminous efficiency of the human eye.)In this way, green appears brighter than red to the human eye, whereas red appears brighter than green to a CCD camera. A CCD camera is normally used for image processing, and thus we provide graphs showing CCD sensitivity change rather than brightness and luminosity values.

Determining the field of view of coaxial lightingTo help you better understand our LED light catalogue

Important points when using a red LED light and a CCD camera with a built-in infrared cutoff filter

Changes in CCD sensitivity due to illumination time for each color of LED

When using a red LED light, remove the infrared cutoff filter from the CCD camera. If the infrared cutoff filter is not removed, the light intensity will decrease by more than one half as the wavelength of red LED light is 660 nm.In addition, the light shifts toward infrared if it becomes hot, and thus heating will cause the image to darken.

120

100

80

60

40

20

00 10 20 30 40 50 60

Per

cent

age

of c

hang

e %

Elapsed time (minutes)

Measurement results Red light shows a sensitivity decrease of approximately 6%. Red light shows a sensitivity decrease of approximately 31% when an infrared filter is attached. White light shows an increase of approximately 2.6%. Green and blue show no change in sensitivity.

1.0

0.8

0.6

0.4

0.2

0400 500 600 700

Wavelength nm800 900 1000

Rel

ativ

e se

nsiti

vity

Color and model used Red LDR-50B Red (W/an infrared cut filter) LDR-50B White LDR-50B-SW Green LDR-50B-GR Blue LDR-50B-BL

Camera

Shutter speed

Camera Lens

F-Stop

Extension tube

Field of view

Power Supply

Volume Level

Temperature

Humidity

WD

LWD

Ingrared cut filter is used. The intensity is measured

on the light reflected from the standard white plate.

NC-300(Takenaka system)

1/60 Sec.

f25mm Lens

F-16

5mm

20mm

PSD-3012/3024

100%(Max)

25°C

50%

99mm

50mm

0.80.70.60.50.40.30.20.1

00 50 100 150

LWD mm (Light Working Distance)

CCD

sens

itivity

Red

Green

Object

Reflected light (1/2)

Diffusion Plate

LWD

'LW

D

Transmittedlight(1/2)

LED

Field of view V (field of view of camera WD, A to I)

Effective field of view Vb

Half mirror

CCD camera

VaVb

WD

LWD

'bLW

D'b

LWD

'aLW

D'a

CCD camera

Provisional emittingsurface positions A

Provisional emittingsurface positions B

Virtual emittingsurface positions B'

Virtual emittingsurface positions A'

A B C D E F G H I

Figure 2 Determining field of view by LWD

Figure 1 Cross-section of coaxial light

Figure 3 Field of view and effective field of view

Provisional emitting surface

Technology OverviewTechnology Overview An introduction to the characteristics of LED lighting

What to consider when selecting a lightWhat to consider when selecting a light

2

4

5

3

1

2

4

53

1

CCD sensitivity change and brightness distribution characteristics provide a good guideline for selecting a light based on the above conditions.

75 76

CCS Inc. will repair or replace the product free of charge if it should fail to function or if the radiant quantity of the product should drop to 50% or less of its initial radiant quantity within the specified warranty period.If either of these conditions occurs, please take the product to your CCS sales representative.

This warranty information provides the scope of CCS's product warranty within the specified period, and does not indicate or imply any further guarantee beyond the warranty terms.Contact CCS for inquiries or information on repairs to the product after the expiration of the warranty.

The radiant quantity refers to the wattage of physical energy radiated from a LED. It refers to the radiation luminosity of the LED measured under conditions specified by CCS or the radiation illumination of the LED under specified irradiation conditions. CCS specifies the radiant quantity for each LED light because the measurement and irradiation conditions vary from the form, the application and the irradiation wavelength.

Warranty Terms

Warranty Information

Warranty period:Two years (one year for radiant quantity), starting from CCS Inc. shipping date.

CCS Inc. will repair or replace the product free of charge if it should fail to function under normal use in accordance with the Instruction Guide and other written cautions during the indicated warranty period of two yearsCCS Inc. will repair or replace the product free of charge if its radiant quantity should drop to 50% or less of its initial radiant quantity under normal use in accordance with the Instruction Guide and other written cautions during the indicated warranty period of one year.CCS Inc. will charge a repair fee under the following conditions :

If the product has been subjected to misuse, unauthorized repairs, or modification from its original design.If the product has been damaged from impacts due to inappropriate handling.If damage to the product results from external causes including accidents, fire, pollution, riots, communication failures, earthquakes, thunderstorms, wind and flood damage, or any other act of providence, or from any extraordinary conditions such as electrical surges, water leakage, condensation, or the use of chemicals.If the damage results from connection to any power supply or to any equipment which CCS Inc. does not manufacture or does not specify for use.

1.

2.

3.

1) 2) 3)

4)

Note:

CCS's activities with regard to environmental issues

Environmentally friendlyEnvironmentally friendly

Please visit our website for detailed information on the RoHS Directive and product compliance with RoHS Directive.

http://www.ccs-grp.comPlease visit our website for detailed information on warranties.

http://www.ccs-grp.com

*1: Items listed in this brochure (Except for some products). *2,*3 : Visit our website for detailed information on the RoHS Directive and product compliance with the RoHS Directive.

CCS has achieved compliance with the RoHS

Directive for products including its standard

lineup of LED illuminators , power supplies ,

and optional accessories ,

and delivers RoHs compliant products

CCS has achieved compliance with the RoHS

Directive for products including its standard

lineup of LED illuminators , power supplies ,

and optional accessories ,

and delivers RoHs compliant products

*1 *1

*2 *3

I t is our desire to contribute to society through dissemination of LED illuminators.

*1

77 78

CCS Worldwide SupportWe will satisty your needs with variousproducts.

CCS World Wide NetworkFor your nearest sales representative, please visit our website.

CCS Worldwide SupportWe will satisty your needs with variousproducts.

CCS provides lighting solutions and technical consultation on locations worldwide so that you can choose the best illumination.Please feel free to contact us. All locations have testing rooms where you can do experiments of your workpieces with our lightings.(Please make an appointment in advance.)

Testing RoomProducts Display Shelf

CCS World Wide Network

CCS Europe NV/SA

Bergensesteenweg 423, Bus 13 1600 Sint-Pieters-Leeuw, Belgium

TEL : +32-(0)2-333-0080FAX : +32-(0)2-333-0081Email : [email protected]

RDV (S) PTE LTD.

63 Hillview Avenue #07-10, Lam Soon Industrial Building, Singapore 669569

TEL : +65-6769-1669FAX : +65-6769-3422Email : [email protected]

CCS Inc. Shanghai Office

Room 308B-309, CIMIC Tower No.1090 Century Avenue, Pu Dong New Area, Shanghai 200120, P.R. China

TEL : +86-21-5835-8728FAX : +86-21-5835-8928Email : [email protected]

CCS America Inc.5 Burlington Woods Suite 204Burlington, MA 01803 USA

TEL : +1-781-272-6900FAX : +1-781-272-6902Email : [email protected]

CCS Inc. (Headquarters)

Tokyo Sales Office/ Testing Room

Shimodachiuri-agaru, karasuma-dori, kamigyo-ku, Kyoto 602-8011 JAPAN International Sales Department

TEL : +81-75-415-8284FAX : +81-75-415-8278

Kyoto subway karasuma line Marutamachi station Exit2 (5 min,to North on foot)

PalacesideHotel

Goou shrine

Policeoffice

McDonald's

Karasum

a-dori

Kyotogyoen

KyotoPrefectualoffice

N

JR line Kyoto station

Kyoto subway karasuma line Marutamachi station Exit2

Shimochoja machi-dori

Shimodachiuri-dori

Testing RoomCCS's strong product lineup meets your diverse lighting needs.

Dedicated staff provide consultations for the selection of lighting and technical inquiries.

For your nearest sales representative, please visit our website.

CCS Inc. Shenzhen office17B,China Economic Trade Building,7Rd Zizhu,Zhuzilin,Futian District,Shenzhen 518040 P.R.China

TEL : +86-755-8279-0477FAX : +86-755-8279-0478 Email : [email protected]

NTT Kanto Hospital Ryoshinji Grand Prince Hotel

Shin-Takanawa

UFJ BankAM PM

Route 1

Seisen UniversitySeisen University

Red Baron

ShinagawaPrince Hotel

Toeiasakusa LineTakanawadai St.Toeiasakusa LineTakanawadai St.

Toeiasakusa LineGotanda St.Toeiasakusa LineGotanda St.

JR Shinagawa StationJR Gotanda Station

NNTakanawadai Green bldg. 8,9F, 4-6-6 Higashigotanda, Shinagawa-Ku, Tokyo 141-0022 JAPAN(9F:front counter, 8F:testing room)

TEL : +81-3-5475-3681 FAX : +81-3-5475-3682

2 minutes walk fromTakanawadai Station, Toeiasakusa Subway Line / 10 minutes walk from JR Gotanda Station, Yamanote Line / 17 minutes walk from JR Shinagawa Station

Kunball Ave

128

3A

95

95

Lahey ClinicMedical Center

Mildred R

d

Barbara Cir

Mich

ele Dr

Wayside Rd

Cam

bridge St

Burlington Ma

ll RdS Bedford St

UnitedCart Park

Burling

ton W

SD

E19/A7

Ruisbroeksesteenweg

ZUUN

Berg

ense

stee

nweg

Meerw

eg

RUISBROEKRUISBROEK

Fabrie

kstra

at

HUMEHILLVIEWGREEN

CHANTILLYRISE

HILLBROOKS

INTAN

BATU

CENTURYMANSIONS

HILLVIEWHOUSE

GUMILANG

JLN

REMAJA

KEWALRAMHILLVIEW

AVE

GLENSPRINGS

HILLVIEWAPARTMENTS

HILLVIEW CARSPLAYGROUND

LAM SOONIND BLDG

Pu D

ong

S.Rd

.

Dong

Fong

Rd.

Pu Dong Rj.

Zhang Yang Rd.

Huang Pu Jiang

Lao

Shan

E.R

d.

Lao

Shan

E.R

d.

Century Rj rd.

Century Rj rd.

Dong Chang Lu(Subwey)Dong Chang Lu(Subwey)

CIMIC TowerCIMIC Tower

Customer service support locations all over the worldProvides lighting solutions and technical support.

廣 深 高 速 公 路

紅 西路

農林路

交通局大廈

招商銀行大厦

光大銀行大廈

紅樹林路

敦煌大廈

財富廣場

本元大廈

深航酒店

地鐵1號綫竹子林站

地鐵1號綫竹子林站

園博園

人才公園( 擬建中)

濱   海   大   道

深   南   大   道

往皇崗口岸方向

往CB D中心區

往機場方向

中国経貿大廈中国経貿大廈China Economic

Trade BuildingChina Economic

Trade Building

Takanawadai Green Building9F:Sales Office(Reception)8F:Testing Room

Notes:

Examples of workpiece imaging in this catalog are a guide that may be informative for choosing illuminations. Please check the functions of the equipment and requirements when choosing.In addition, the sample workpieces that are used are processed by us and do not represent the original quality and performance.

Carefully read the product's instruction manual before use to ensure correct operation. Product specifications and design are subject to change without notice.

Many of our products are protected by industrial property rights (patents, industrial designs, and trademarks). Be warned against imitations of CCS brand.

*

HeadquartersShimodachiuri-agaru, Karasuma-dori, Kamigyo-ku, Kyoto 602-8011 JapanPhone: +81-75-415-8284 / Fax: +81-75-415-8278

URL: http://www.ccs-grp.com E-mail: [email protected]

Copyright(c) 2008 CCS Inc. All Rights Reserved.Description in this catalog is based on the information as of April 2008. 02001-10-0409