p~~~~ fcs~ y,

70
_). } ' l ' ' I EXTERNAL HEART ASSIST DEVICE by ., ,. PET AR FRAN 0 BASIC submitted in fulfilment of the requirements for the degree ·Of MASTER OF SCIENCE IN ENGINEERING UNIVERSITY OF CAPE TOWN Deeem-BeP 1973 Th e copvric'·4 r ... ·- U . , o.- t!. . n•versit-., .. - r- ...... ••-:?Sis is held b ' Rc ' c.. ·-.::;::," T::. ··m Y the / prod t:cq.- , 1 I may be m .. ·., cr V;e wf:o/e or l not for study purposes •cat•on. . y, and

Upload: others

Post on 06-Jan-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: p~~~~ fcs~ y,

_).

-~ } '

l

' ' I

EXTERNAL HEART ASSIST DEVICE

by

., ,. PET AR FRAN 0 BASIC

submitted in fulfilment of the

requirements for the degree

·Of

MASTER OF SCIENCE IN ENGINEERING

UNIVERSITY OF CAPE TOWN

Deeem-BeP 1973

Th e copvric'·4 r ... ·-

U . , <:>"~. o.- r-:}:~ t!. . n•versit-., .. - r- ...... ••-:?Sis is held b

' Rc ' c.. ·-.::;::," T::. ··m Y the / prod t:cq.- ,1 • • • I may be m .. ·., cr V;e wf:o/e or

l not for p~~~~ fcs~ study purposes o~~y p~rt •cat•on. . y, and

Page 2: p~~~~ fcs~ y,

I -{ii)-

ABSTRACT

The work described herein is an independent and original work

carried out in the Department of Chemical Engineering on a

heart assist device which was designed, built an'd tested in the

laboratory.

The project can be divided into a mechanical and an electrical

task, both of which were undertaken by the author. The electri­

cal task involved design and construction of a portable console

adapted to receive a triggering signal and to activate a series

of controls either in a sequential or singular fashion. The

system also comprises a pulse frequency counter and a safety

shut- off control circuits.

A portable pressure-vacuum supply system and a pressure pul­

sating mechanism comprise the work done in completing the

mechanical task. Of the two pressure pulsating mechanisms

built and tested in the laboratory it is shown that a minimum

flow, rigidly enclosed system is superior to a high flow, variable

volume one.

The design of the electronic console, pressure supply system

and pressure pulsating mechanism is described and a descrip­

tion of the laboratory tests carried out so far is given.

Page 3: p~~~~ fcs~ y,

- ( iii) -

ACKNOWLEDGEMENTS

The author wishes to express his sincere appreciation and

gratitude to Prof. A. D. Carr, Dr. B. M. Kennelly and

especially to Ass. Prof. H. 0. Buhr for their valuable

assistance and encouragement throughout this study.

The author is also grateful to the Workshop Staff of the

Chemical Engineering Department for their assistance

rendered in the fabrication of various parts of the system.

Finally, the generous financial assistance given to the

author by the Chris Barnard Fund, the Medical Research

Council, the Council for Scientific and Industrial Research

and the Harry Crossley Foundation is gratefully acknow-

ledged.

Page 4: p~~~~ fcs~ y,

- ( iv) -

TABLE OF CONTENTS

ABSTRACT ii ACKNOWLEDGEMENTS iii TABLE OF FIGURES vi NOMENCLATURE viii

CHAPTER 1

1.1 INTRODUCTION 1 1.2 AR TERIO-ARTERIAL PUMPING 1 1.3 INTRA-A OR TIC BALLOON PUMPING 4 1.4 EXTERNAL COUNTERPULSATION 6 1.4.1 Theory of External Counterpulsation 6 1.5 OBJECTIVES 8

CHAPTER 2

2.1 MEDICAL BACKGROUND 10 2.2 THE HEART 10 2.3 HEART BEAT 13 2.4 CARDIAC CYCLE AND EGG 14 2.5 HEART AS A PUMP 16 2.6 CORONARY BLOOD FLOW 19 2.7 HEART WORK AND EFFICIENCY 19

CHAPTER 3

3.1 EXTERNAL HEART ASSIST DEVICE 23 3.2 DESIGN REQUIREMENTS 24 3.3 ELECTRONIC CONSOLE 26 3.3.1 Test Signal Circuit 29 3.3.2 Shaper and Delay Circuits 29 3.3.3 Pulse Frequency Counter 32 3.3.4 Safety Shut- of£ Circuit 37 3.3.5 Timing Circuit 39 3.3.6 Actuating Circuit 43 3.3.7 Power Supplies 45 3.3.8 Control Circuits 47 3.4 EXTERNAL COUNTERPU LSATOR 52 3.4.1 Pressure Pulsating Medium 53 3.4.2 Sequenced Pulsator 54 3.4.3 Minimum Flow, Rigidly Walled Pulsator 58

1 3.5 HYDRAULIC PRESSURE - SUCTION SUPPLY 60 I

)

Page 5: p~~~~ fcs~ y,

-.(v)-

CHAPTER 4

4.1 RESULTS 4.1.1 Sequenced Pulsator 4.1.2 Rigidly Walled Pulsator 4.1.3 Undesirable Effects 4.1.3.1 High Flow 4.1.3.2 Entrapped air 4.1.4 Test on a Volunteer

CHAPTER 5

5.1 CONCLUSIONS AND RECOMMENDATIONS

REFERENCES

APPENDIX A

HEART EFFICIENCY

APPENDIX B

ELECTRONIC COMPONENTS

SN 74121 Monostable Multivibrator High Speed Flip- Flops Phototransistor Coupled Pair MCT2 The Pressure Transducer

APPENDIX C

PROPAGATION OF A PRESSURE WAVE

APPENDIX D

65 65 68 70 70 70 73

75

78

A1

B1

B1 B3 B3 B5

C1

VOLUMETRIC RATIOS D1

Ratio of Air- Water Volumes in the Pressure Tank D1 Ratio of Air- Water Volumes in the Vacuum Tank D3 Volume of Displaced Water due to the Movement of the Leg in and out of the Pulsator Cylinder D4

Page 6: p~~~~ fcs~ y,

- (vi) -

TABLE OF FIGURES

Fig. 1. Schematic block diagram of the SIMAS system as used in clinical trials Page 3

Fig. 2. Intra- aortic balloon in position Page 5

Fig. 3. Schematic representation of changes in intra-vascular volume and transmural pressure Page 7

Fig. 4. Anatomic components of the heart Page 11

Fig. 5. Aortic volume pressure relations in different age groups (isolated aortas) Page 12

Fig. 6. Cardiac events cycle Page 15

Fig. 7. Ventricular pressure changes Page 18

Fig. 8. Schematic block diagram for an external heart assist device Page 23

Fig. 9. Pulsation System Page 25

Fig. 10. Block diagram for the external counterpulsator Page 27

Fig. 11. Test Signal Circuit Page 30

Fig. 12. Shaper and Delay Circuits Page 31

Fig. 13. Pulse Frequency Counter Circuit Page 33

Fig. 14. Graph of Pulse Frequency Counter Output Voltage vs Time Page 34

Fig. 15. Graph of Pulse Frequency Counter Output Voltage vs Cardiac Cycle Page 35

Fig. 16. Graph of Tp Period vs Cardiac Cycle Page 36

Fig. 17. Safety Shut- Off Circuit Page 38

Fig. 18. Logic Diagram for the Safety Shut- Off Circuit Page 38

Fig. 19. Articulated Pulsating Mechanism Page 39

Fig. 20. Timing circuit logic Page 40

Fig. 21. Timing Circuit Page 42

Page 7: p~~~~ fcs~ y,

Fig. 22.

Fig. 23.

Fig. 24.

Fig. 25.

(vii)

Actuating Circuit for each of the eight solenoid valves

5V DC Power Supply

Front end elevation of the Console

Side view of the Console

Fig. 26. a) Plan view of the leg sheath

Fig. 26. b) Cross- sectional view A-A. of the leg sheath with rubber cuffs placed in pockets

Fig. 27.

Fig. 28.

Fig. 29

Fig. 30

Fig. 31

Fig. 32

Fig. 33

Fig. 34

Fig. 35

Fig. 36

Fig. 37

Fig. 38

Fig. 39

Plan view of the sequenced pulsator

Cross- sectional elevation of the minimum flow, rigidly walled pulsator with leg fitted therein

Front view of the Hydraulic Pressure-Suction Supply

Plan view of the Hydraulic Pressure-Suction Supply

Schematic diagram for the Hydraulic Pressure­Vacuum Supply

Pressure Response Curve for the Sequenced Pulsator

Pressure Response Curve for a Rubber Cuff

Pressure Response Curve for the Rigidly Walled Pulsator

Improved Timing Circuit Output

Graph of Pressure Rise Time vs Entrapped Air Volume

Pressure Response Curve for the Rigidly Walled Pulsator when tested on a Volunteer

SN 74121 logic

MCT 2 to TTL interface

Page 44

Page 46

Page 48

Page 49

Page 55

Page 55

Page 57

Page 59

Page 62

Page 63

Page 64

Page, 66

Page 67

Page 69

Page 71

Page 72

Page 74

Page Bl

Page BS

Page 8: p~~~~ fcs~ y,

I

- (viii) -

NOMENCLATURE

A Ameter

'a' Atrial pressure wave

C Capacitor

Cx Pressure propagation wave velocity

'c' Delay due to the electrical and mechanical response of the external counterpulsator

'D' ~ Total delay period

DpF Constant width pulse generator output

'd' Delay period effected by the Delay circuit

ECG The electrocardiagram·

Ek Kinetic energy imparted to the expelled blood from the ventricles

F/F Flip-flop

G The Pulse Frequency Counter circuit input

GND Ground potential

IC Isometric contraction

IR Isometric relaxation

m The mass of blood expelled from a ventricle per beat

ME Maximum ejection

MCT2 Phototransistor-diode coupled pair

M Monostable multivibrator

P,Q,R,S,T ECG waves

rch Chamber ~ressure

~i Intravascular pressure

Ptr Transmural pressure

Page 9: p~~~~ fcs~ y,

I

t.Pv

Q

Q

Qc

QRS

R

RE

Tp

TR

Tv

ts

v

Vee

Vs

v

We

ZD

- (ix) -

Mean ventricular pressure rise

The monostable multivibrator and flip-flop output

The monostable multivibrator and flip-flop inverse output

Cardiac output

The ECG complex signal

Resistance

Felaxed ejection

External pressure application period

Transistor

External vacuum application period

Physiological time delay

Voltmeter

Supply Voltage

The Pulse Frequency Counter circuit output voltage

The mean velocity with which the blood is expelled

Cardiac work

Zener diode

Page 10: p~~~~ fcs~ y,

I

CHAPTER 1

1.1. rnTRODUCTION

For many years Cardiologists have felt the need for a cardiac

equivalent of the artificial kidney, a device which could support the

circulation temporarily, enabling the patient to survive an episode

of circulatory failure and giving the heart time to 'recover from

some temporary insult such as myocardial infarction (coronary

thrombosis). The concept of circulatory assist is then to provide

means to augment the cardiac output with the objective of increasing

the perfusion of blood to all organs, the heart included, without

increasing the energy requirements of the heart.

1.2. AR TERIO - ARTERIAL PUMPrnG

Since 1957 a circulatory assistance pump, designed to reduce pres­

sure within the aorta and left ventricle during systole and to elevate

aortic pressure during diastole, has been in development and experi­

mental use /1,2/. In principle, this method relies upon the concept

that reduction of the amount of blood and pressure in the aorta during

cardiac systole results in a reduction in cardiac work~ Catheters

are placed into the central arterial system by way of the femoral

arteries and blood is removed in association with ventricular systole.

This lowers the aortic pressure with the result that the left ventricle

develops less pressure to eject its volume of blood. The volume of

blood removed from the system during systole is then replaced into

the aorta in a retrograde fashion during diastole, augmenting aortic

diastolic pres sure and coronary blood flow. Such a device also

works as an in- series pump, in that it often provides an additional

force for propulsion of blood through other noncoronary vascular

beds, and will, therefore, aid in the force which determines net

venous return.

- 1 -

Page 11: p~~~~ fcs~ y,

J

I

- 2-

To be effective, this method requires precise timing. Consideration

must be given to allow for delay from inertia of the blood, particu­

larly if the location of the catheters is distal in the arterial tree.

Also, the catheters should be positioned in such a way to enable the

withdrawal of a significant amount of blood at the right time. There

is limitation of the size of catheters which can be introduced into the

aorta through a peripheral blood vessel. Such phased mechanism is

usually triggered by the electrical activity of the heart. However,

the heart in failure may present either a chaotic rhythm or rhythm

which is too rapiQ. for efficient activity of the pump. Past studies

have indicated that counterpulsation is capable of reducing mean

systolic pressure and the time-tension index. Soroff et al I 3/

reported a reduction in left ventricular work and myocardial oxygen

consumption with no significant change in cardiac output.

In later studies 14,51 of Soroff et alit was shown that with synchro­

nized counterpulsation in dogs, mean systolic pressure could be

lowered by approximately 40 mm Hg, the time-tension index was

reduced by 39 percent and myocardial oxygen consumption reduced

by 22 percent. Mean diastolic pressure remained at the original

levels but coronary blood flow increased by 50 percent. These

findings suggest that counterpulsation might be a useful technique

in the management of selected patients with ischemic heart disease

and cardiogenic shock.

Rosensw~ig et al /61 have also shown that counterpulsation substan­

tially increases coronary blood flow. They attributed this increase

to the increase in collateral coronary flow which resulted from the

increased diastolic pressure provided by the counterpulsation.

One of the counterpulsation devices now available commercially is

SIMAS, the Standard Isochronous Myocardial Augmentation System

I 7 I, which has received limited clinical trial. The operation of the

Page 12: p~~~~ fcs~ y,

I

'

- 3 -

PH::IS/OLOG/CRL

MONITOR

ELECTRONIC CONTROl-S

ACTUATOR. AND

H~DRf\ULI C 60PPL~

Fig.l. Schematic block diagram of the SIMAS system as used in clinical trials. Femoral arteries are exposed and cannulated at A and B

Page 13: p~~~~ fcs~ y,

I

I

- 4-

system is illustrated1n Fig. 1.

A modification of this principle has resulted in what is known as the

auxiliary ventricle technique I Bl. In this situation, the pumping

chamber is attached directly to the aorta (through a thoractomy)

and the blood is removed from the arterial circuit immediately

adjacent to the aorta and returned directly to the aorta at the end of

a relatively large- sized conduit 19,101.

1.3 INTRA-AORTIC BALLOON PUMPING

This technique may be considered a derivative of arteria-arterial

pumping but it differs in its mechanics in an important way. Whereas

with the arteria-arterial pumping there is the withdrawal and reintro­

duction of blood to the cardiovascular tree, no blood is extracted from

the arterial system with the balloon technique I 11 I.

A catheter- balloon system is introduced into the aorta from the

femoral region and placed in the intrathoracic aorta as shown in

Fig.2. The balloon is filled during ventricular diastole, thereby

increasing the aortic root pressure and improving coronary flow,

and deflated during ventricular systole. Such system has been built

and tested at the University of Cape Town /12,13 I.

The major disadvantages of both the arterio- arterial and intra- aortic

balloon pumping are that such counterpulsation methods involve expo­

sure and cannulation of femoral arteries, the use of an extracorporeal

blood handling device, the requirement of sterile technique and the

necessity for administration of ~nticoagulants. A potential hazard

is rupture of the balloon and rupture of the vascular wall by over­

distention of the balloon.

The ability to augment pressure is somewhat limited by the size of

Page 14: p~~~~ fcs~ y,

I

\,

{ I

\ J

I \

- 5 -

Fig. 2. Intra-aortic balloon in position

·-------- _____ .. ____ _

Page 15: p~~~~ fcs~ y,

I I

-1

RA.

l.\1.

v.c. Ao.

- 7 -

Ptr = Pi- Pch

t

L.V.

?ressure Chamber

Pch = 1 bar Pch = 0,9 bar Pi = 1,25 bar Ptr = 0,35 bar

Pch = 1,1 bar Pi = 1,25 bar Ptr = 0,15 bar

Pi = 1,25 bar Ptr = 0,25 bar

Fig.3. Schematic representation of changes in intra­vascular volume and transmural pressure ( Ptr) with changes in chamber pressure ( Pch) while the intravascular pressure (Pi) remains con­stant.

Page 16: p~~~~ fcs~ y,

- 8 -

l>etween the internal pressure and the pressure which surrounds the

tube.

In the vascular bed the transmural pressure is the difference between

the intravascular pressure, which changes continuously, and the extra­

vascular pressure, which is normally essentially constant and very

nearly equal to atmospheric pressure. Certain portions of the vascular

bed which are affected by muscular activity respond ,to variations in the

extravascular pressure so created. For example, venous return from

the limbs is increased when the limbs are exercised and the coronary

vascular system ~s affected by myocardial contractions.· If means are

provided to vary the pressure external to portions of the vascular bed,

the capacity of the affected portion of the vascular bed will respond to

the changes in transmural pressure which are produced.

Fig. 3 schematically illustrates that as the external pressure is

lowered in relation to the ambient pressure, the capacity of the disten­

sible vascular bed increases and vice versa. The resulting change in

capacity is determined by the pressure-volume characteristic of the

cardiovascular system.

In the circulatory system, the venous bed acts as a large reservoir

which readily compensates for changes in blood volume distribution

when cardiovascular equilibrium is disturbed. This is due to the

fact that in the typical distribution of the blood volume there is approxi­

mately 15 percent of the blood in the arterial system, 75 to 80 percent

in the venous bed and the rest is accommodated in the capillary bed

/19/. Because of such high capacity the venous bed will be respon­

sive to relatively small changes in external pressure whereas larger

excursions in external pressure will be required to affect the arterial

system.

1.5. OBJECTIVES

A project was undertaken to design, construct and test a device which

Page 17: p~~~~ fcs~ y,

- 9 -

would apply positive pressure to the legs of a patient during diastole

and switch to vacuum for the rest of the cardiac cycle. Such appli­

cation would be synchronous with the cardiac activity so that:

( i) the heart work is reduced,

( ii) the cardiac output is increased,

(iii) the perfusion of the myocardium is increased.

The obvious advantage of using an external mechanism is that it does

away with surgery and many logistic problems that accompany it.

However, the clinical environment of the typical patient in cardia­

genic shock places many constraints on the design of the system.

The ideal unit would have to be a portable one, which is relatively

quiet in operation and requires no elaborate adjunct equipment, does

not interfere with normal patient care requirements and in the case

of an emergency is readily applied. Furthermore the application of

the system should not hinder standard forms of medical treatment

and preferably cause minimal disturbance to the patient.

Page 18: p~~~~ fcs~ y,

CHAPTER 2

2.1. MEDICAL BACKGROUND

The human body is composed of a vast number of cells all of which

are grouped and organised according to their specific functions.

Being contained in a confined space, the body cells can function and

survive only when the immediate environment is adequately supplied

with the essential nutrients, i.e. when the pH, temperature and con­

centration of various substances is maintained within certain limits.

The conveying system for such substances is the cardiovascular

system, which comprises the heart, arteries, veins and capillaries

on one hand and the blood on the other. . The transportation of the

nutrients and wastes to and from the cells is effected by blood, which

in turn is circulated along the cardiovascular bed and distributed to

the various tissues in a manner according to their functions. The

source of the kinetic and pressure energies, required by the blood

to be circulated through the body, is the heart.

2.2. THE HEART

The fibrous 'skeleton' of the heart, as illustrated in Fig. 4.c.,

consists of four valve rings joined together to which the two major

arterial trunks and all four cardiac chambers are fastened.

The two atria, which resemble a thin-walled, shallow cup of myo­

cardium, divided by a partition down the centre, are fastened to

the superior surface of the mitral { M) and tricuspid ( T) valve

rings as shown in Fig. 4.a. The remaining two chambers, the

right and left ventricles, are fastened to the entire circumference

of the fibrous skeleton and form the lower part of the heart. This

arrangement is illustrated in Fig. 4.e. The ventricles serve as

- 10 -

Page 19: p~~~~ fcs~ y,

)

right--­atrium

A. V. valves~.

- 11 -

---aorta

{a) {b)

fibrous skeleton

{c)

---left ventricle

{d) {e)

Fig. 4. Anatomic components of the heart

Page 20: p~~~~ fcs~ y,

200

a' . :c 140

E E

uJ ~ ~ 80 (/) (/) LJ o! tl-

2.0

- 12 -

AGE

2.6 42.

20 40 60 80 100 12.0 140 160

VOLUME. cc/rn'" BOD~ SURFACE

Fig.S. Aortic volume pressure relations in different age groups (isolated aortas)

Page 21: p~~~~ fcs~ y,

- 13 -

the major source of energy for the circulation of blooq ~dare

composed of sheets of myocardial fibres encircling the ventricula,r

chambers I 20-22 I. From a functional point of view, this muscu­

lature can be divided into two groups of myocardial bundles, the

spiral muscles and the deep constrictor muscles 1231.

The two semilunar valves, pulmonary ( P) and aor,tic (A) as shown

in Fig. 4.d., are very similar, each consisting of three sy~metrical I

valve cusps. The atrioventricular (A-V) valves are fasten~d to the

inferior surface of the mitral ( M) and tricuspid ( T) valve rings with

chordae tendineae extending from the inferior margins of each valve

leaflet and fastened directly to the internal surface of the ventr;icul~r

walls. All four valves open and close rapidly and seal completely

against high pressures with minimal obstruction to flow.

Like the atria, the two major arterial trunks, the aorta anc;:l the p\ll­

monary artery originate at the superior surface of the semilunar

valve rings as shown in Fig. 4.b. In the external counterJ?ulsation

the location and the pressure-volume characteristic of these two

vessels influence the design of the assist device, since the prima:~."y

objective of such an external mechanism is to raise and lower

pressure inside the arterial tree. The pressure-volume relations

vary widely in different individuals in the same age group ~d the

curve also tends to shift towards larger volume and less distensi­

bility as subjects grow older I 24 I. Such a characteristic is

illustrated in Fig. 5.

2.3. HEART BEAT

To produce efficient pumping, the complex mass of myoc~rdial

bundles must contract more or less sim1:11taneously. Co-ordin~te<;l

contraction of the complex pattern of myocardial bundles stems

from the syncytial arrangement of the myocardial fibre~; excitatio:p.

Page 22: p~~~~ fcs~ y,

- 14-

beginning at one site spreads to all other contiguous areas. Exci­

tation of the heart is normally initiated by an impulse which is

generated by the sino-atrial (S-A) node (a small mass of specialised

myocardial tissues embedded in the atrial wall near the entrance of

the superior vena cava) and which spreads rapidly in all directions

through the atrial musculature (atrial polarisation) . As it approaches

the interatrial septum, the wave of excitation reaches another mass of

spec;i.alised conducting tissue, the atrioventricular'( A- V) node, which

is located near the posterior margin of the interatrial septum close to

the entrance of the coronary sinus. After a slight delay ( 0,08 to 0,12

second) I 25/ at the A- V node, impluses are conducted by the Purkinje

system into the ventricles where a wave of excitation spreads from

the endocardial surfaces through the ventricular musculature. This

normal sequence of excitation initiates myocardial contraction and

establishes the mechanical events of the cardiac cycle ..

2.4. CARDIAC CYCLE AND ECG

A recording of electrical potentials, associated with waves of exci­

tation which spread through the heart, is known as the electro­

cardiogram ( ECG) . The normal ECG trace consists of five waves

(:I? ,Q,R,S and T) which correlate with the major events of the cardiac

cycle as shown in Fig. 6. The P- wave, the QRS complex and the T­

wave appear during atrial depolarisation, ventricular depolarisation

and ventricular repolarisation respectively. The P-wave is normally

100 milliseconds in duration, followed by the P-Q interval during

which the atria pumps blood into the ventricular chambers ('a'

wave of the atrial pressure as illustrated in Fig. 6.) Following

this brief isoelectric period there is a series of deflections, the

QRS complex.

With the onset of the Q-wave the isometric contraction ( IC) of the

ventricles begins, accompanied by a rise in intraventricular pressure.

Page 23: p~~~~ fcs~ y,

> f

p

- 15 -

R

ECG T

Q s ,Q 5

IC ME RE tR DIASTOLIC DRAIN

AORTIC PRESSURE

VENTRICULAR VOLUME

400 800 12.00

TIME IN MSECS

Fig.6. Cardiac events cycle

T

inlra ve.,tricufctr p..essu.re

Page 24: p~~~~ fcs~ y,

- 16 -

When the intraventricular pressure reaches that within the aorta,

the aortic valve opens and the maximum ejection (ME) phase takes

place. In man the duration of QRS is about 80 milliseconds with

a normal upper limit of 100 - 120 milliseconds. The final T-wave

is a broad deflection. Its rise, peak and fall coincide approximately

with the relaxed ejection ( RE) phase, the closing of the aortic valve

and the isometric relaxation ( IR) phase respectively, as shown in

Fig.6.

With the average duration of the S- T interval equal to 270 milli­

seconds, the duration of QRST phase is between 350 and 400 milli­

seconds for a normal heart rate of 70 beats per minute. In the

design of the external counterpulsation mechanism this criterion

will be used as well as the phenomenon that a QRS complex cannot

be followed by another before a T -wave.

The heart performs its greatest work during the isometric contrac­

tion and maximum ejection phases. Therefore if myocardial work

is to be reduced, the external counterpulsation must be timed to

reduce pressure in the ascending aorta before the aortic valve opens.

To maintain system flow and pressure, the pressure is to be increased

during the diastolic period.

2.5. HEART AS A PUMP

The average adult human heart weighs about 290 grams and has a

stroke volume of approximately 70 ml. of blood. For a normal

heart rate of 70 beats per minute the cardiac output is nearly 5 litres

and since this is about equal to the total amount of blood in the vascular

system, the blood is therefore circulated once every minute. Heart

output can, however, increase six-fold I 26/ in a severe exercise.

The atrial pressure curve as depicted in Fig.6. shows three major

Page 25: p~~~~ fcs~ y,

- 17 -

pressure elevations, known as the 'a', 'c' and 'v' waves. Under

normal conditions there is a continuous flow of blood from the

great veins into the atria and approximately 70 percent of this flows

directly into the ventricles before the atria contract, causing addi­

tional 30 percent filling ('a' wave). The left atrial pressure then

rises to 7-8 mm Hg while that of the right atrium reaches a some­

what lower peak: 4-6 mm Hg. The 'c' wave is caused during

ventricular contraction by reflux of blood out of the ventricles into

the atria on one hand and bulging of the closed A- V valves towards

the atria on the other. Because of the large amounts of blood

having accumulated in the atria by the end of systole, the atrial

pressure rises ( 'v' wave) but falls off steeply as the ventricular

pressure returns to its low diastolic value when the pressure in

the atrial chambers push open the A- V valves and allow the blood

to flow rapidly into the ventricles.

The function of the ventricles as a pump is entirely different from

that of the atria. The rapid filling of the ventricles occurs during

the first third of diastole, followed by diastasis - the middle third

of diastole when blood flows from veins directly into the ventricles

- to be completed by the last third phase during which atria contract

giving an additional thrust to the in-flow of blood. Both the filling

and the emptying of the ventricles is depicted in Fig.6. by the ven­

tricular volume curve.

With the onset of systole, the ventricular contraction is accompa­

nied by a rise in ventricular pressure which causes the A- V valves

to close immediately and only after the pressure level has risen

to 80 mm Hg in the left and 8 mm Hg in the right ventricle, as

illustrated in Fig. 7., do the aortic and pulmonary artery valves

open. Blood then begins to pour out of the ventricles, about half

of the outflow occurring during the first quarter of systole and

most of the remaining half during the next two quarters. In the

Page 26: p~~~~ fcs~ y,

- 18 -

LEFT VENTRICLE AND AORTA PRESSURE CURVES

I J.. r I I aonk vaive I 1/ cfoseci I

HOR.rtc. Pi?!£ . l I Ss vt~e

L.V. ----t-1

SYSTOLE UlRSTOLE

RIGHT VENTRICLE AND

-- - - - -- '70 wem 1-t~

PULMONAR~ ARTER~ PRESSURE CURVES

- -- - - 25 rnm H~

_ _ _ _ _ _ _ 8 m m 1-1~

~~::::=--1----~~~~~~f---- -- -·- 0 mm H~

S~STOLE DIASTOLE:

Fig. 7. Ventricular pressure changes

Page 27: p~~~~ fcs~ y,

- 19 -

last quarter of the systole ventricular relaxation begins suddenly,

allowing the intraventricular pressure to fall rapidly, whereupon

the elevated pressures in the large arteries immediately snap the

aortic and pulmonary valves closed. When the ventricular pressure

reaches the low diastolic levels below atrial pressures, the A- V

valves open and a new cycle begins.

2.6. CORONARY BLOOD FLOW

The coronary blood flow of a normal heart has been found to be 7 to

10 percent I 27/ of the total cardiac output. Since the coronary

circulation depends essentially on the difference in the pressure

between the right atrium and the aorta, an increase in diastolic

pressure will increase the coronary flow and at the same time the

oxygen supply. Under normal conditions, myocardial oxygen con­

sumption is approximately 8 to 10 ml of 02 per 100 grams of heart,

but it can increase severalfold with exercise and decrease moder­

ately under conditions such as hypotension and hypothermia. The

myocardial oxygen consumption correlates reasonably with the

area under systolic portion of the left ventricular pres sure curve

I 281 so that the effect and usefulness of a cardiac assist device can

be evaluated using this index.

When the blood, i.e. the oxygen supply is cut off tc;> a part of the

myocardium, myocardial infarction takes place; The muscle cells

die and the area becomes necrotic. One of the main causes of

myocardial infarction is coronary thrombosis which occurs when a

coronary artery is occluded by a thrombus that develops within the

lumen of a sclerotic vessel.

2.7. HEART WORK AND EFFICIENCY

Since William Harvey' s disclosure of the nature of the circulatory

system in 1632 the pumping action of the heart remains a difficult

Page 28: p~~~~ fcs~ y,

- 20 -

parameter to measure and a great need for a simple, non-invasive

method to obtain information concerning the mechanical activity of

the heart, exists. In recent years a wide variety of techniques I 29 I

has been proposed to determine cardiac output by indirect methods,

most of which have serious limitations and all of which are subject

to various sources of error. For the purpose of this study the

work done by the heart will be approximated as that due to two

factors only:

If

( i) Cardiac work done in expelling the blood out of the

ventricles against the external pressures, a·nd'

( ii) Kinetic energy imparted to the expelled blood.

We = Cardiac work

Qc = Cardiac output

APv = Mean ventricular pressure rise

then We = Qc x APv ....................... ( 1.1)

Considering the complexity of the situation, the cardiac output is

not an easy parameter to measure. The stroke volume (output

per beat) is not directly proportional to arterial pressure since

the pressure-volume relationship of the arterial system is not

constant nor uniform among individuals, but also non-linear from

high pressure to low. Hence various approximation techniques

have been developed. By the dye dilution technique advocated by

Hamilton and Remington I 30 I the pulse pressure correlates

roughly with stroke volume, namely a pressure rise of 1 mm Hg

is equivalent to about 1 tc of stroke volume/ sq m over the normal

pressure range. Even if the cardiac output were to be computed

according to the Fick principle /31/, such method is at best± 10

percent accurate.

Page 29: p~~~~ fcs~ y,

If Ek

m

v

=

=

=

- 21 -

Kinetic energy imparted to the expelled

blood

The mass of blood expelled

The mean velocity with which the blood

is expelled

then Ek = .................. ( 1. 2)

In this case too, both parameters ( m) and ( v) are difficult to mea­

sure. However, Erlanger and Hooker I 32 I have shown that it is

justifiable to use the product of the pressure pulse by the pulse rate

as an index to the relative velocity of the ejected blood but only upon

the assumption that the following factors remain constant:

a) the rate of flow from the arteries into the veins;

b) the rate of systolic output;

c) the distensibility of human arteries;

d) the amount of blood in the arterial system.

From the Example on Heart Efficiency in the Appendix it follows that

the work done by the left ventricle per beat is 0,8935 + 0,00875 Nm or

0,90225 Nm and for the right ventricle 0,1401 + 0,00875 Nm or 0,14885

Nm. It follows then that the work expenditure of the left ventricle is

nearly 7 times that of the right ventricle, hence any reduction in the

work load on the left ventricle will greatly assist the heart.

Also, the equation ( 1.1) above emphasises the important fact that

the work of the heart is increased by factors raising either its output

or pressure against which it has to operate, therefore all effort in

the design of an external heart assist ought to be directed primarily

toward the latter, i.e. the reduction of systolic pressure and also

toward increasing diastolic pressure which would simultaneously

Page 30: p~~~~ fcs~ y,

I

- 22-

increase the perfusion of the myocardium.

If, during the time the aortic valve is open, the mean pressure in

the aorta were to be reduced from 108 mm Hg to typically 88 mm Hg,

17 percent reduction in work would be achieved. It is toward this

end that the project is directed.

Page 31: p~~~~ fcs~ y,

CHAPTER 3

3.1. EXTERNAL HEART ASSIST DEVICE

The design and construction of the external pressure pulsating

device to produce the effects of counterpulsation, was undertaken

in three stages, the first an electrical one and the remaining two

mechanical.

In its desired form the external heart assist, based upon the prin­

ciples discussed in sub- section 1.4.1., comprises an electronic

console, an external counterpulsator and a pressure/vacuum

supply. The three modules are integrated as shown in Fig.8. to

provide a mechanical support of the active vascular system and,

at the same time, to reduce the myocardial work demand placed

on the heart under various conditions.

PRESSURE/

/VACUUM'

SUPPLY

.ELECTRONIC CONSOLE

~ 00

d. ... z 0 \J

c::::> PRESS.

... 0 t1 .. z 0 u

VAG.

EGG SIGNAL

PRESSURE SIGNAL

{1, /\

I. ,! .. ~;

SYSTEM

~· 't} .. ,.• ·-, . ' ·~·

EXTERNAL C OUNTERPU LSAT OR

Fig.8. Schematic block diagram for an external heart assist device

- 23 -

I

Page 32: p~~~~ fcs~ y,

- 24-

3.2. DESIGN REQUIREMENTS

With the onset of systole, i.e. the receipt of the QRS triggering

signal, the system is to apply pressure to the vascular bed for

approximately half the cardiac cycle and vacuum for the remainder

of the cycle. However, the pres sure in the aorta is· to be affected

by the externally applied pressure only after the aortic valve has

closed, i.e. during diastole, hence the application of pressure

period Tp is to be delayed for a period 1 d 1 as shown in Fig. 9.

Ideally the delay 1 d' would be equal to the systolic period if it

were not for two more delaying factors, namely the physiological

·time delay 1 ts' and delay 1 c' due to the electrical and mechanical

response of the external counterpulsator. These three delays

comprise the total delay period 1 D' as illustrated in Fig.9.

The type of external counterpulsator used in this project is a·

pressure pulsating leg unit into which the patient' s legs fit. The

reason for using such a unit and its design are discussed in detail

in sub- section 3.4. After inserting the legs, the container is filled

with working fluid. This fluid is connected to both the pressure

and the suction supply tank via a solenoid valve. On receipt of the

control signal the pressure solenoid valve opens, connecting the

working fluid with the pressure supply thus exerting pressure on

the legs. When the pressure valve closes, the suction valve

opens causing vacuum to be effected on the legs. The two solenoid

valves operate alternatively.

Since heart rate changes, the system should accelerate or decele­

rate its pulsing action accordingly. Also, should there be a

spurious beat during diastole, i.e. while the pressure is being

applied to the vascular system, such pressure is to be removed in

Page 33: p~~~~ fcs~ y,

1\ I \

' \ I I \ ' l 1 \ I I I l I

~ : I ~ I I ' ' I ' I l \ I \,. • .,

- 25 -

ECO PHVS.

MONITO'R.

ELECTRONIC

CONSOLE

,-------- --1

t====+:i~<:::J~:::;t=l PR.E SS. TANK

SUCTION

TANK I I 1 I

L - - - - ·- --- - - - - -- - - - ~' HYORRUL.IC SUPPLY

'PRESSURE 1>Ut.SATING UNIT

ECG

AORTIC PRESSURE I

I ,

. AORTIC VALVE

CLOSED

' '

SPURIOUS BEAT

~ II

'

\ I I \ I \ --- ,, ....... _________ ,.. •!

SYSTOLE I DIASTOLE

I

SYSTOLE

.. D~-t -d J-c~ I ,......--........ --:--~ ~-d -~!

Tv

I ' j DELAYED EFFECT OF EXTERNALLY APPLIED

I r-tsj PRESSURE ON THE CARDIOVASCULAR SYSTEM phy~:ioloqico.l ~im

aldav

Fig. 9. Pulsation System

Page 34: p~~~~ fcs~ y,

I

- 26 -

the shortest possible time, as shown by the dotted lines in Fig.9.

Both the time delay 'd' and the pressure/vacuum applied should

be adjustable.

To meet the above objectives the electrical task involved building

and testing an electronic console to monitor, control and synchro­

nise the action of the external counterpulsator.

3.3. ELECTRONIC CONSOLE

In the construction of the electronic console extensive use has

been made of solid state integrated circuit chips. Truth tables

together with other relevant data is given in the Appendix B.

The ECG signal, which is picked up by dermal sensors placed

on the patient's skin, as shown in Fig.9., is processed and ampli­

fied by a physiological monitor. The greatest portion of this work

was expended on the design and construction of an electronic con­

sole, the block diagram of which is given in Fig.lO. The console

was designed to accept a preconditioned triggering signal from an

existing physiological monitor. The monitor is fitted with a dual­

channel cathode- ray oscilloscope specifically designed to display

ECG and aortic pressure wave signals. In this manner the monitor

provides a continuous ECG triggering signal to the console and also

provides means for visually setting the triggering level at which

the system will be activated.

For experimental purposes the console can provide a test signal

at a rate approximately that of the two heart- beat limits, namely

60 and 180 beats per minute. Both the output of this Test Signal

circuit and the input for physiological monitor processed trigger

output are fed to a selector switch. The selected triggering

·signal is then fed into a Shaper, which is a highly noise- immune

Page 35: p~~~~ fcs~ y,

I t

I

L&J _. 0

~ 0 (.)

~ z 0 (/. 1-u w ...J au

r---------, PH~SIOLOGICAL

MONITOR

, PULSE FRE~:

COUNTER

POWER SUPPUI

sv

I

~

/

...

ECG / SIGNAL /

TEST SIGNAL

/ /

/ /

,, t SWITCH

SHAPER

D£LA~

CCT

,, ltMING CCT

~ , ~ , , ACTUATING

CCT

... -

TEST SIGNAL CCT

, SAFET~

SHUT-OFF

I I L _______ .... _____________ ..... __ -- ____ _j I

PRESSURE PULSATING LEG UNlT

~-------------------, PRESS. SUPP~

t----r PUMP,__ __ ...,

Fig. 10. Block diagram for the external counterpulsator

- 27 -

I

I SUCTION

SIJPPLV

Page 36: p~~~~ fcs~ y,

I

- 28-

mono stable multi vibrator. The Shaper will accept any type of

trigger with a positive slope and shape it into positive- step

trigger suitable for a Delay circuit. The Delay circuit introduces

a time delay between receipt of the ECG signal and initiation of the

pressure pulse. The time delay 'd' produced can be fully adjusted

from 0 to 120 milliseconds. At the end of the time delay 'd' the

pressure phase Tp is applied as shown in Fig.9. When properly

set to allow for system and physiological time lags,:the period of

the effect of pressure application Tp corresponds with cardiac

diastole. The use of the ECG signal to trigger the delay 'd' , i.e.

the pressure removal Tv, is advantageous for reasons of patient

safety since with such a mode the heart will never be pumping

against a pressure wave produced by the assist device. This is

especially important in the case of a premature beat.

The triggering signal from the Shaper is fed into a Pulse Frequency

Counter circuit and a Safety Shut- off circuit. The latter monitors

the ECG signal and detects any spurious beats, whereupon it

immediately removes the externally applied pressure as illustrated

by the dotted lines in Fig.9. The output of the Pulse Frequency

Counter circuit is the pulse size controlling voltage of the Timing

circuit which determines the length of Tp periods of the cardiac

cycle. Each triggering signal represents a heart beat and since

heart rate changes, the rate of application of external pressure will

have to change accordingly and the period Tp of the pressure appli­

cation as shown in Fig.9. will have to be increased or decreased

proportionately. Hence the input to the Pulse Frequency Counter

circuit is made proportional to its output, i.e. the number of

cardiac cycles per unit time which the Counter receives determines

the pulse size voltage, thus compensating automatically for varying

pulse rates. In this project one beat per second (i.e. 1000 milli­

seconds cardiac cycle) was chosen as the lower and 3 beats per

second (i.e. 333 milliseconds cardiac cycle) as the upper heart'

rate limit.

Page 37: p~~~~ fcs~ y,

I

- 29 -

From the Delay circuit the signal is channelled to a Timing circuit

which triggers an Actuating circuit which, in turn, energises two

electrical solenoid valves as illustrated in Fig.10. On opening, one

solenoid valve connects the pulsating mechanism with a positive

pressure supply and the other with a negative pressure supply. The

valves operate alternatively and are timed to remain open through­

out the period of application of positive (or negative) pressure.

Being of an industrial type, they operate on 220 volt~, hence the

Actuating circuit acts as a shunting system between the high 220

volts solenoid valve side and the low 5 volts electronic circuitry.

The entire console is mounted in a portable cabinet on wheels and

uses only a single source of 220V, 50Hz power supply. Thus need

for special power line is obviated.

3.3.1. Test Signal Circuit

The Test Signal circuit is illustrated in Fig.11 and comprises two

monostable multivibrators I 33 I TTL units ( SN 74121) connected

in series to provide 1 beat or 3 beats per second. The monostable

multivibrator No. 1. output gives the OFF pulse of constant width

A ( 230 milliseconds), while the multivibrator No.2. determines

the ON pulse of the two widths B and Bl. · The two set values B

and Bl ( 770 and 100 milliseconds respectively) are obtained by

shunting two parallel resistors, 10 kilo-ohms and 4,7 kilo-ohms

across the 10 kilo-ohms trimpot variable resistor.

3.3.2. Shaper and Delay Circuits

Both the Shaper and Delay circuits comprise a single SN 74121

monostable multivibrator unit, as shown in Fig.12. The Shaper

receives the triggering signal through a selector switch, either

from the physiological monitor or the Test Signal circuit and

Page 38: p~~~~ fcs~ y,

I

OUTPUT

I

A

A

- 30 -

.------+--~+ H~ 011pF'

ISO A

tOk~

f/

B'

Monostab1e Mu1tivibrator No. 1. SN 74121

A

Monostab1e Mu1tivibrator No. 2. SN 74121

Vee =5V

B

B'

A

second

Fig. 11. Test Signal Circuit and Output Logic

Page 39: p~~~~ fcs~ y,

I

t

- 31 -

PHYSIOLOGICAL MONITOR TEST SIGNAL CIRCUIT OUTPUT OUTPUT n--n

Vcc----4

D E L A Y Monostab1e Mu1tivibrator

SN 74121

I t

I I t I

I ' l - - - --- - - - -- _,

~----------------~l~-------------------.•Vcc

S H A P E R ( Monostab1e

Multi vibrator SN 74121

L---------~ TO P. F. C. CCT

TO TO SAFETY SHUT-OFF CIRCUIT

Fig. 12. Shaper and Delay Circuits

Page 40: p~~~~ fcs~ y,

I

I

- 32-

shapes it to an acceptable step trigger for the Delay circuit. In

the case of the Shaper multivibrator, the internal timing resistor

( 2 kilo- ohms nominal) and the internal tim~ng capacitor are used

to obtain minimal delay of 30 nanoseconds.

The .Pelay circuit employs an external timing resistor ( 10 kilo- ohms

variable) and a capacitor ( 15 microfarads) to obtain delay 'd' from

0 to 120 milliseconds, the value of which is set manu.ally. The

Timing circuit requires a positive- step trigger, so Q is used as the

Delay circuit output.

3.3.3. Pulse Frequency Counter

The Pulse Frequency Counter circuit, as shown in Fig. 13., is of a

discriminator type which, depending on the number of triggering

signals it receives at input c.:;., supplies a proportional voltage at the

output terminals A and B. The circuit comprises a constant width

pulse generator ( monostable multivibrator type SN 74121 with

external timing components R = 20 kilo- ohms variable and C = 22

microfarads giving the output pulse of constant width DpF = 230

milliseconds) and two BC 169 n-p-n transistors TR1 and TR2 con­

nected in cascade with a 220 microfarads capacitor between the out­

pl,lts A ( 0 volts) and B ( Vs volts}. The resistors R2, R3, R5 and

R6 are adjusted so that the output voltage Vs is approximately 4 volts

at 3 beats per second and 0,6 volts (average} at 1 beat per second.

Fig.14 illustrates the Pulse Frequency Counter circuit output for

1 and 3 beats per second inputs as well as its response to abrupt

changes in heart rate from 3 to 1 beat and from 1 to 3 beats per

second. The relationship between the Pulse Frequency Counter

output voltage and the cardiac cycle time length, which is given in

Fig. 15. is given for values of cardiac cycles between the lower

( 1000 milliseconds) and the upper limit ( 333 milliseconds} and is

not linear.

Page 41: p~~~~ fcs~ y,

s""

Mo,-to.s.to..bte I SN '7412.1 Q 1 ""./Wt, l K Mu.tti. vibrnhr

INPUT G

(FROM SHAPER) \ CIRCUIT

o"

•w ~~

TRI

"' -s"" 'POWER SUPPLY

__ Cz 2.2.0~lF

Fig. 13. Pulse Frequency Counter Circuit

.;

..... --

s"

oB L, ----,·r-vvs

TR2. BCt6'

--'-- c3 220pF

.~ A

l.V l.V

Page 42: p~~~~ fcs~ y,

~ !j 4

~ I

l.aJ (!) ~

0: !::; g2 1-:l 0..

~ 0

n n+l

• .l

Fig.l4. Graph o£ Pulse Frequency Counter Output Voltage vs Time

th .. ee beats /second

one beQ~ / second

TIME IN SECONDS

Vma.x

- - - - 97 % V mox

---- 80 '/o Vmo...x

40 Jo Vmax

w ~

Page 43: p~~~~ fcs~ y,

~ LIJ t- II)

-/

Fig.l5. Graph of Pulse Frequency Counter Output Voltage vs Cardiac Cycle

5 ~ 0 0 4 U"> '·""·

3~----------~.~ •

, LLI Ue> Za: w!:i :lo ~·> 2

a:t-L&. :J

' • -'t

Cl ~ 1- I (J) :J ..J 0 :::> p..

·~·~. I ------- '"--- __:__ -- -- --- - . I ........_• ............._

_ _ _ - - - - - I I ----·---- I I

I I I

200 :500 400 500 600 .,00 800 qoo 1000

CRRDIRC C~CLE IN MSECS

J

w U1

Page 44: p~~~~ fcs~ y,

$00

U) 400

0 w Ul ~ 300

0 0 -~ L&J

200

P-

f-L I

/./·""

100

300 400

I • ·c' ~-

Fig.l6. Graph of Tp Period vs Cardiac Cycle

/./·

.~·/

././

500 600 700 800 900

CARDIAC C~CLE T MSECS

w ~

.<.;0.,.

Page 45: p~~~~ fcs~ y,

I

I

I

- 37 -

For experimental use Tp and Tv, the two cardiac cycle pha~es when

either pressure or vacuum is applied externally to a section of the

vascular bed, have been made nearly equal but their ratio can

readily be adjusted. When tested, the Timing circuit, in conjunction

with the Pulse Frequency Counter, gave 480 milliseconds and 160

millisec;:onds long Tp periods for cardiac cycles of 1000 and 333

milll.seconds respectively. Fig.l6. illustrates the relationspip

between Tp period and the cardiac cycle time len~th . .' This relation­

ship is linear with a maximum deviation of 3 percent.

3.3.4. Safety Shut-Off Circuit

The purpose of the Safety Shut-off circuit is to monitor the ECG

triggering signal and, should there be a spurious signal or a sudden

acceleration in heart rate during the Tp phase of the previous beat,

to remove the external pressure as quickly as possible and reset

the system to a new cycle, i.e. to effect the vacuum phase Tv

. immediately.

The Safety circuit makes use of a quadruple 2-input NOR gate TTL

unit (SN 7402 N) as illustrated in Fig.17. One gate monitors Q

output from the Delay circuit. This Q output suppresses the pres­

sure signal and its suppressive effect is achieved by enabling th~

controlli:n.g NOR gate output (pin No.4.) to be at the logical 1 during

the Tp period only as shown in Fig. 18. The controlling output

must be at logical 1 for pressure signal to be on. It is normally

at such enabled state when the Timing circuit requires Tp, but

can only maintain 1 while Q indicates diastole. Should Q change,

i.e. when there is a spurious beat, Tp will immediately be switched

off as indicated by the dotted lines in Fig.18. The output signal

from the controlling gate will cear the appropriate flip- flops in

the Timing circuit, resetting the counterpulsator to vacuum phase

Tv of a new cycle.

Page 46: p~~~~ fcs~ y,

' I

• ·'

r

I

I

OUTPUT TO TIMING CCT

+Qo FROM DELAY CIRCUIT

t Q, I PRESSURE ON I SIGNAL FROM TIMING CIRCUIT

Fig. 17. Safety Shut-Off Circuit

ECG ! ..

SYSTOLE ..,

J

OUTPUT (1 o) .. I

#I., J

OUTPUT (1 J - I

~ {INPUT (.6

:i"' '5 ._ \N'PUT ( S IX([

t-" % · OUTPUT e

) .. I

) I -('-If I

AORTIC VAL \If' CLOSED

I

I I

I I I I

I

I j I I I

I I I I

i I I

Tp

SPURIOU$ .BERT

I

I .1 j

..: ; ;t .. I

DIASTOlfE i 1-- ---- - ---! r--------i i i I

i L i

I J

---------I i J

~J Tv

Fig. 18. Logic Diagram for the Safety

Shut-Off Circuit

- 38 -

IY

'

I'"---

I

'- ---

r----I

I

L.--

I t----

... ____

r--.1.

Page 47: p~~~~ fcs~ y,

I

--t ' '

r

I

- 39 -

3.3.5. Timing Circuit

The timing circuit we de signed for use in conjunction with an

. compartmented pressure pulsating mechanism where pressure

and vacuum are applied in a sequential fashion, from one end of

the pulsator to the other. If the leg pressure pulsating unit were

divided into four adjacent but pressure independent sections as

shown in Fig.19., then with the onset of the Tp phase of the

cardiac cycle all four sections would be pressurized one after

the other, starting with section No. 1.

solenoid

valve

I I I I

I

common vacuum

I I I

I I

I , ' ... ~

manifold

' ' ,. ~

' ' I I '4" I l \31 I I \ I

:0

CD I I ",

'

I I I I

' I \ I I

• I

I

' ' \ ' ,~·

solenoid

valve

I I I

common pressure

manifold

Fig. 1?. Articulated Pulsating MecP.anism

Page 48: p~~~~ fcs~ y,

r

\

ECG

SHAPING CCT Q

"""

SRFm SHUT Ol'F CCT ~

DELAY Q

M 1 Q

-M 1 Q

M 2 Q

M 3 Q

M 4 Q

F/F 1

F/F 2

F/F 3

F/F 4

--. . •

.

.

~

4

. ..

- 40 -

CARDIAC CYCLE __. EARLY )( BEAT '-.t

I I

~ SYSTOLE DIASTOLE

1\ SYSTOLE

y '!" '• n I I

,.__d. ~I~ Tp ~ t I I

I

:

n ll l

I

OFF -ON I OFF I J

ON I OFF I

ON I OFF

.J ON l I I

. Tv4 f. Tpf+ I

'X

Tv3 ·I--Tp3 ~

Tv2--,-Tp2

Tv1----!--- Tp1----

TIME t __.

Fig.20. Timing circuit logic

Page 49: p~~~~ fcs~ y,

I

I \

I

- 41 -

Obviously pressure would be applied for the longest period Tpl in

the first section and for the shortest period Tp4 in the fourth section

as illustrated in Fig.20. Such action would produce the desired move­

ment of the blood from the leg towards the heart. The ratio of Tpl,

Tp2, Tp3 and Tp4 may be readily adjusted to suit any condition. On

the basis of one beat per second an interval of 100 milliseconds is

used between opening of two consecutive pressure valves. At the end

of the pressure application period Tp the pressure in. all four sections

is switched simultaneously to vacuum as shown in Fig.20.

Timing circuits have been built to control the period of vacuum/

pressure application of each of the four sections. However, in the

case where the pressure pulsating mechanism is not divided into

individual sections but is merely used as a single unit, only Tpl

timing circuit is then used.

Fig.21. illustrates the Timing circuit which consists of four sections,

each section comprising a monostable multivibrator (TTL chip SN

74121) driving a flip-flop (TTL chip SN 7473 N). The SN 7473 N

is a binary unit so only two are used for the whole system. The mono­

stable multivibrator determines the length of Tp period. A variable

potentiometer is used as one of the external timing components, hence

the Tp period of each section is readily adjustable. The vacuum

application period Tv is complementary to Tp.

The pulse size control voltage from the Pulse Frequency Counter

is fed to each monostable multivibrator as the power supply to

external timing resistor thus controlling the change in time length

of each output Q (pin No.6.).

The inverse output Q of the Delay circuit is used to provide a common

trigger for the Timing circuit as shown in Fig.21. The monostable

multi vibrator ou~put will remain enabled for a period of time, the

Page 50: p~~~~ fcs~ y,

-\

' '-_L l>OL5E St%.£ CONTROL VOLT~CSE f {FROM "PUt.SE FRc~. COLINTE~) 'lJ·D (FROM DELAY CIRCUIT}

TO THE SAFETY _..... SHUT OFF CCT

t SAFETY SHUT-OFF

Fig. 21. Timing Circuit

- 42 -

MCT~

Page 51: p~~~~ fcs~ y,

I

' \ ___.

=4 I )

\ (' .

\.

t.

- 43 -

length of which is determined by the sizes of the external timing

components R and C. Each multivibrator drives a flip-flop TTL

unit ( SN 74 73) which, through an actuating circuit, keeps either

the pressure or vacuum valve open. Since a negative- step pulse

triggers the flip- flop, the inverse output Q of the first mono stable

is used. Thus at the end of delay period 'd' the inverse output -Q will immediately clock the first flip-flop; causing the first pres-

sure valve to be opened. The output Q of the second monostable

multivibrator is enabled for about quarter of the time of the first

one as shown in Fig.20. At the end of this time period the negative­

step will clock the second flip-flop, i.e. open the second pressure

valve for a period Tp2. In this manner the opening of the second

valve is delayed by approximately 100 milliseconds. Using the

same technique, the third pressure valve is opened about 100 milli­

seconds later and the fourth pressure valve another 100 milliseconds . later. The pressure application times Tp1 - Tp4, are illustrated

in Fig.20. At the end of the period Tp1 flip-flops are cleared and

reset by a negative- step pulse from output Q of the first mono stable

multivibrator. The common 'clear' signal ensures that the pres­

sures in all four sections are switched to vacuum at the same time.

Should the ECG trigger occur anywhere in the Tp phase of the pre­

vious cycle, as illustrated by the dotted line X - X, the output sig-

nal of the Safety Shut- Off circuit will clear all the flip- flops. These

units can be clocked into pressure again only in ~he new cardiac cycle

initiated by the early signal (X - X).

Each flip-flop has two outputs Q and Q (pin No.12. and 13. in the

case of the first flip-flop and No.9. and 8. of the second) which

are connected to light emitting diodes ( MCT2' s) of the suction and

pressure solenoid valve controls respectively in the Actuating circuit.

3. 3. 6. Actuating Circuit

The Actuating circuit acts as a shunting circuit between the low

Page 52: p~~~~ fcs~ y,

I ,_

- 45 -

voltage electronic control system on one hand and the high voltage

solenoid valve system on the other. Each solenoid valve has the

same actuating circuit as illustrated in Fig. 22. Altogether there

are eight such circuits, four for pres sure and four for suction lines.

The solenoid valve is connected across the 220 V supply through a

gate controlled, high peak current, full wave, silicon AC Triac

( 2N 5756). A shunting arm, consisting of a 47 ohms resistor and

0,1 microfarads, 630V capacitor joined in series, is connected across

the triac as a safety measure to prevent backlashing as the 8W solenoid

valve coil is being switched off and on.

The gate of the triac is connected to the flip-flop output through a

phototransistor- diode coupled pair MGT 2 unit. This diffused planar

GaAs diode, which is optically coupled to the n-p-n silicon planar

phototransistor, will, on receipt of the control signal from the flip­

flop, emit light causing the phototransistor to conduct while the flip­

flop output is at logical 1. At the same time the gate will trigger

the triac open and will remain open as long as the flip-flop output

is at logical 1. No danger of leakage from 220V to SV side exists

since the isolation resistance is in the order of 1011

ohms and the

voltage isolation from emitter to detector is 1500 volts.

The necessary lOV DC supply is provided for the L.E.D. unit and the

presence of 100 ohms resistance load on the gate is to ensure that

the gate voltage is within the prescribed limits of 0,2 to 2,2 volts.

3.3. 7. Power Supplies

The three power supplies required by the circuitry of the console,

i.e. 220V, lOV and SV, are tapped from a common point thus the

console requires only a single 220V AC, 50 Hz power source.

The lOV DC power supply makes use of an auto-transformer, as·

Page 53: p~~~~ fcs~ y,

~ 0'

O,S)lF

220 V AC --r c3, SOH~

12."

o"

..,

t;.tF C;s2.

Fig.23.

SV ZENER DIODE

tooo,...uF c~

SV DC Power Supply

Z..D.

GND

1 Amp

1 v sv

Page 54: p~~~~ fcs~ y,

}

J

- 47 -

illustrated in Fig.22., to keep the negative terminal tied to the ground.

Such arrangement is essential to enable the triac gate voltage to be

with respect to ground potential, otherwise the floating voltage will

keep the gate always open. On the low voltage side of the auto-trans.,.

former a 500 microfarads capacitor smooths the output to an accep­

table ripple effect.

The 5V DC supply provides power to all console circuits at a stabilised

5V with the current sonsumption varying from 0,2 to b, 7 amps. It

utilises a full-wave silicon rectifier bridge ( SILEC 110B8 029N)

as illustrated in Fig.23. and a zener diode for a stable 5V output. To

protect the ammeter A, which has a full- scale deflection of 1 amp,

the circuit is provided with an 1 amp fuse. Both the 5V and lOV

supplies have a 0,5 microfarads capacitor connected across the high

voltage windings to reduce mains borne interference.

3.3.8. Controls

The entire console circuitry is housed in a cabinet, mounted on

wheels for ease of movement, as shown in Fig.24 and 25. A single

power cable ( 1), on the side of the Console cabinet is to be plugged

into a 220V AC 50 Hz power source. On the same side of the Console

there is a row of 5-pin sockets ( 2) each socket controlling a pair of

valves - one a pressure and the other a suction valve - connected to

a single stage. The console was designed to control a compartment~d

pressure pulsating mechanism having four stages only but provisions

have been made to enable the console to monitor a sequential pulsator

having up to 6 stages. Immediately above the row of solenoid valve

sockets, a pressure transducer socket ( 3) is provided, thus the power

cable, pressure transducer cable and the leads from 5-pin sockets to

solenoid valves are all on the same side of the cabinet, away from the

front panel.

All the controls and monitoring instruments are mounted on the front

Page 55: p~~~~ fcs~ y,

- 48 -

Fig. 24. Front end elevation/6£ the Console

" ..

Page 56: p~~~~ fcs~ y,

( ,.

\ \

I )

'

- 49 -

Fig. 25. Side view of the Console

Page 57: p~~~~ fcs~ y,

,, \

- 50 -

panel of the Console as shown in Fig.25. A double-trace, storage

· oscilloscope ( 4) in the top left section of the cabinet face acts as a

monitor and a visual aid. It can monitor the pressure of the pressure

pulsating mechanism, delay time 1 d', pulse size control voltage,

Timing circuit pressure application periods Tp' s and Test Signal

circuit output. Of these, delay period 1 d 1 and the pressure effected

inside the external counterpulsator are most important.

Next to the oscilloscope there is the ammeter ( 5) ana the voltmeter

( 6) dials of the SV power supply, with a circuit power switch ( 7)

positioned immediately below. With the system running, any change

in the SV power supply output voltage can be easily detected and the

change in the current consumption used to spot any malfunctioning

of the Timing circuit stages.

Below the oscilloscope and running across the width of the Front

panel, there is a row of solenoid valve switches. Altogether twelve

switches are provided, i.e. six pairs, each pair controlling a pressure (8)

. {q) 1 "d . 1 H b f 1 and a suctlon.-.so enol va ve. ence any num er o va ves or any

particular valve can be utilised at will. A selector type of switch is

used, to render the operation of a solenoid valve manual or automatic.

At the end of this row, an operational switch (10) is provided which

controls the power supply to all solenoid valves to enable simultan­

eous initiation of the pulsating system. Each solenoid valve switch

is associated with an indicator light, hence there are altogether six

yellow ( 11) lights representing vacuum and six red ( 12) representing

pressure.

In the middle of the panel, there is situated a main power switch ( 13)

flanked by a 1 amp fuse ( 14) and an indicator light ( 15) . On the

other side of the power switch a removable cover ( 16) protects six

potentiometers and a Delay circuit external timing resistor. They

are all manually adjustable to provide the desired length of pressure

application periods Tp' s and the delay time 1 d' .

Page 58: p~~~~ fcs~ y,

' I J '

- 51 -

Should, at any stage of clinical trials, application of the external

pressure be required to be removed, an emergency switch ( 17) has

been provided immediately below the main power switch. It will

cut the power supply to all pressure valves but leave the vacuum valve

functional. Lastly, there are provided on the front panel of the Con­

sole a number of test points so that the pres sure in the external

counterpulsator mechanism ( 18) , delay time ' d' ( 19) , pulse size

control voltage ( 20), Timing circuit pressure application periods Tp 1 s

( 21) and Test Signal circuit output ( 22) can be monitored. The

Timing circuit pressure application periods test point is also provided

with a selector switch thus any one of the stages can be selected.

All the electronic circuits are mounted on one side of board which

slides in and out of the cabinet; on the other power supply circuits

are mounted. The two sides of the board are shielded from each

other by a metallic plate which is grounded. Also, wherever the

high voltage installations are proximal to low voltage electronic

circuitry, maximum use of shielded cables has been made.

The electronic circuitry has been built on two printed circuit cards

for ease of removal, replacement, adjusting and checking. These

cards slide onto the board between guides. There is enough room

available on the lower shelf of the Console cabinet to carry spare

parts for the electronic circuitry, necessary tools and an instruction

manual for the oscilloscope. The cabinet can be wheeled about the

room with ease.

Page 59: p~~~~ fcs~ y,

\ \

- 52-

3.4. THE EXTERNAL COUNTERPULSATOR

Since the objective of the external heart assist device is to provide

pulsating pressure changes, the mechanical task involved designing,

building and testing of an external counterpulsator. The pressure

pulsation, applied externally to the body of a patient, is to effect the

vascular bed so that it responds to the resulting change in transmural

pressure. A negative pressure is to be applied during systole to

achieve the desired reduction in the systolic pressure and a positive

pressure during diastolic interval of the cardiac cycle to increase

myocardial perfusion.

At first a suit, similar to that of an aviator in a "G11 suit or that of

a 11hard hat 11 diver, was considered as means for enclosing the

vascular bed, i.e. the entire human body. However, due to poor

accessibility of the patient enclosed in such a suit, increased

sweating, obstruction of normal patient care requirements, negative

psychological effort on the patient of being enclosed in a suit and

ability of the patient tohear or be heard being substantially impaired

by the suit and head piece, such idea was abandoned for a more rea­

listic lower body enclosure. Because of its easy accessibility and

the fact that below the abdomen line there is about one third of the

entire blood volume in the circulatory system, the lower body pro­

vides the most convenient locus of the vascular system to effect the

pressure changes. Also utilisation of the legs leaves the upper body

free for medical examination.

For initial clinical trials only one leg is to be utilised and, once the

desired result is achieved, the total area of the affected vascular bed

is to be increased. Hence a design of a leg unit was embarked upon

with provisions that later both legs and a large portion of the lower

posterior would be included. •

Page 60: p~~~~ fcs~ y,

- 53 -

3.4.1. Pressure Pulsating Medium

The initial idea was to have a double-walled trouser-leg unit fitted

over the patient's leg and pressurised with air. Since the shape

and length of a leg varies among individuals to a great extent, the

system was designed for a particular leg with means provided to

accept smaller and larger legs within a limit.

The circumference of the upper end of the typical leg is 57cm and

22cm just above the ankle. The length between the two circum­

ferences is 7lcm hence the total affected leg area is approximately

2800 sq em. If a cushion of air, contained in the trouser leg unit

was 0, 7 em thick, the total volume of the pres sur is e d air would be

about 2 litres. With two litre s of air needed for each leg unit of the

eventual system, at fast heart rates of 3 beats per second, the air

compressor required would have to supply 12litres of pressurised

air per second, i.e. 720 litres a minute. Likewise a suction pump

would have to remove about the same amount of air.

Because of such an enormous volume of air which is to be moved in

a relatively short time, the need would arise for a large compressor

and an equally large vacuum pump. Also, air tends to dampen abrupt

pressure changes (from 200 mm Hg positive pressure to 200 mm Hg

suction pressure and vice versa) towards which this project is

directed. To overcome these difficulties water is to be used, pri­

marily because of its incompressibility. As shown in the Appendix

C, water, compared with air, propagates a pressure wave nearly

five times faster: Cx (air} = 335,28 m/ sec; Cx (water) = 1432,56

m/ sec. The possibility of using hydraulic brake fluid was also con­

sidered due to its relatively low compressibility, but was discarded

because of its high costs, inflammability, corrosive effect on trouser

material and noxious effect on the human skin.

Page 61: p~~~~ fcs~ y,

J

t

) !

)

- 54-

Using water as the working medium and a pulsing mechanism

based on the concept of a double walled trouser leg unit, two

mechanisms were built and tested in the laboratory, namely a

sequenced pulsator comprising a number of inflatable cuffs and a

:rp.inimum flow, rigidly walled system.

3.4.2. Sequenced Pulsator

An air-tight container, made of clear perspex and able to receive

a patient' s leg, comprises the outer shell and a trouser-type leg

unit, which is wrapped around the leg, the inner shell of ·the sequenced

pulsator. The latter is a double walled sheath, tailored to encase a

conically shaped leg. Its outer wall is made of very thin, flexible

but non- stretchable nylon- reinforced PVC sheet, whereas the inner

wall is made of a soft wool fabric which does not cause irritation to

the skin, but provides a warm cover for the leg. The vacuum or

suction is effected by the external shell and the positive pressure by

the inner shell so that the mean applied pressure is nearly equal the

atmospheric.

As illustrated in Fig. 26.a., the leg sheath, which fits the typical

leg from the ankle to the thigh, is provided with laterally positioned

pockets. Each pocket accommodates a rubber cuff. To avoid

undesirable discontinuity in the area of pressure application, the

pockets have been made in such a way as to have the cuffs, positioned

therein, overlapping, as shown in Fig.26.b. Made of 3mm black

rubber, each cuff has a centrally located spout, acting as both the

inlet and the outlet.

By using a sealing type of material, known as 'Vulcro' , a simple

method.of fastening the sheath about the leg is. employed. When the

sheath is wound about the leg, with two opposite edges overlapping,

the male strapping, provided on one edge, is laid over female strips,

provided on the other edge, causing a strong interlocking. The

fastening, which prevents the edges from sliding away from each

Page 62: p~~~~ fcs~ y,

. J

nylon reinforced PVC sheet

- 55 -

~A 1 spout

aperture

pockets

.:

female vulcro

straps

Fig. 26. a.) Plan view of the Leg Sheath

PVC sheath

male vulcro straps

wool

spout

male vulcro straps

,;pig. 26. b.l Cross"-"'secttonal vtew A""A of the Leg

Sheath with Rubber Cuffs placed in

Pockets

Page 63: p~~~~ fcs~ y,

- 56 -

·other when the rubber cuffs are pressurised, is undone by merely

peeling off the male straps.

The leg, with the compartmented sheath fastened about it, is positioned

into the rectangular per spex box, as depicted in Fig.27. The box,

which is air-tight, is shown with its lid open, The short, upright

wall, through which the leg fits, is removable for ease of placing the

leg inside the box. This wall is provided with a rubber seal in the

form of a bell stretching from the leg diameter, where it is sealed,

to the circular opening in the wall, to which it is secured.

The rubber cuff spouts are easily connected to the pressure/vacuum

line by plugging them into quick- couple connector sockets mounted

on the inside of the long side wall. Each socket is connected through

the wall to a free arm of a T-piece, mounted on the outside as illu­

strated in Fig.27. On the remaining two arms of the T-piece there

are two solenoid valves, one controlling the passage to the common

pressure supply manifold and the other to vacuum. From a hydraulic

pressure- suction supply liquid under pressure and vacuum is supplied

to the two manifolds mounted on the outside of the per spex box.

The console, as described in previous sub- section, activates the

pressure solenoid valves in a sequential manner, switches them off

together and, at the same time, switches on all the vacuum valves.

During the pressurisation phase of the pulsating cycle, the working

fluid distends the leg sheath, causing the inner wall to press against

the leg. The pressure valve controlling the passage to the widest

cuff, i.e. the one which fits around the leg immediately above the ·

ankle, is open the longest- typically 480 milliseconds - and the one

controlling the narrowest cuff, the shortest- 160 milliseconds. In

this way a propagating squeezing effect is achieved from the ankle

upwards towards the thigh. On depressurisation, when the vacuum

valves are opened, a certain amount of liquid is withdrawn from the

pressurised cuffs releasing the tightening grip about the leg. The

leg is then exposed to the surrounding pressure i.e. the negative

Page 64: p~~~~ fcs~ y,

- 57 -

rubber bell seal

lid in open position

compartmented sheath

removable end wall

pair of solenoid valves per each

\ rubber cuff

\ ....... ---.n l==;:;:::::t>l¢::::1 \ \ _______ l -----1 I \ I

pressure manifold

I I .._ ____ ...., l=ri=:f><l==l -·(------;--

/ I ( ' \ \

-\---+-\ ' \ I \ I

0

to vacuum pump

quick couple connec­;:or

vacuum manifold

drain from pressure

supply

to vacuum supply

~ig. 27. Plan vtew ot the Sequenced Pulsator

Page 65: p~~~~ fcs~ y,

( \. \

- 58 -

'pressure which is continuously maintained in the perspex box.

Thus the two shells, working together in phase, effect the counter­

pulsation.

3.4.3. Minimum Flow, Rigidly Walled Pulsator

To minimize the amount of flow of the working fluid into and out of

the pulsator a system with rigid walls was designed1 Instead of

having two separate shells - an outer and an inner one, the minimum

flow pulsator, as built and tested in the laboratory, has a combination

of the two. It also retains water as the working medium.

As illustrated in Fig. 28., the minimum flow pulsator comprises a

rigidly walled unit of cylindrical shape, closed at one end and with

an aperture on the other to receive a leg. The aperture is provided

with a specially constructed seal which consists of a stocking, made

of imperm·eable neoprene material 0,2 mm thick and adapted to sheathe

the leg, a rubber cover and a vacuum seal wrap. The space between

the stockinged leg and cylinder walls is filled with working fluid. The

purpose of the stocking is two fold: to keep the leg dry and to pre­

vent spillage of the working fluid through the small circumferential'

gap between the leg and the flange. In the case of the typical leg the

aperture is of such a diameter that it leaves a gap of approximately

4 mm at the apex.

To prevent the stocking from bulging out through the small gap in the

aperture, during the pressurisation phase, a thick rubber cover,

which fits between the leg and the stocking is provided. Likewise,

to prevent the thick rubber cover and the stocking from being pulled

into the cylinder and away from the leg when suction within the cylin­

der is effected, the same negative pressure is created in the small.

circumferential gap in the aperture and just outside it. As shown

in Fig. 28. this is achieved by employing a vacuum seal wrap around

Page 66: p~~~~ fcs~ y,

N 0 T

pressure transducer testing nipple

\_____._

suction outlet

~ ·- ~·· , . ..., ..

T 0 S C A L E

leg stocking air outlet

outlet valve

C> 0 C

. . ·- . ~ ·- ---.---;-; '"

rubber cover

c:r-o -,,

J I

flanges

Fig. 28. Cross-sectional elevation of the minimum flow, rigidly

walled pulsator with leg fitted therein

vacuum seal wrap

gap U'l. \0

Page 67: p~~~~ fcs~ y,

. 'J

- 60 -

the protruding thigh and by connecting to the vacuum supply. When

the leg is fitted into the stocking and the cylinder is filled up with

water, the leg stocking collapses tightly about the leg. To withdraw

any air which may be trapped, a perforated tygon tube is positioned

alongside the leg as shown. To facilitate air removal the cylinder is

tilted so that the unit is filled from the closed end up towards the

aperture.

The closed end of the cylinder is made of a double walled PVC disc

which is provided with an inlet and an outlet nozzle for the working fluid,

both of which are 19 mm in diameter and each of which is connected

through a solenoid valve to the hydraulic pressure and suction supply

tanks. Just above the nozzles there is a testing nipple onto which the

pressure transducer nozzle connects. Furthermore, two air outlet

valves have been fitted on the top ridge of the cylinder to allow entry,

or escape of air during filling or draining. The narrow air outlet

near the seal is connected to the vacuum system to allow bleeding out air.

From the above it is clear that the system can function only in a non­

sequential i.e. single-pulse fashion. However, pressurisation and·

depressurisation inside the vessel will be damped to a degree by the

movement of the leg in and out ofthe cylinder. The force which causes

the outward movement is equivalent to the pressure effected on the leg

multiplied by the leg skin area inside the cylinder. Likewise, the in­

ward movement is due to the force equivalent to the product of suction

pressure and the affected leg area. There is, however, a resistance

to these movements due to a number of factors, the largest one resul.;.

ting from the fact that the patient' s seat is fixed in relation to the

cylinder.

3.5. HYDRAULIC PRESSURE - SUCTION SUPPLY

The pressure - vacuum generating system, as shown in Fig. 29 and

Fig. 30., comprises two upright tanks, a pressure and a vacuum

Page 68: p~~~~ fcs~ y,

i

t ~

)

i. I l

j. I

1

:.

- 61 -

tank, partially filled with water and interconnected through a

pump. Pressure and vacuum are caused by pumping the water

from one and into the other tank. A return from the pressure tank to

the pump inlet is channelled through a control valve of much smaller

orifice than the pipe diameter used in the system. The re suiting

pressure drop across this valve (due to the constricted flow) is

used to control a pressure differential of desired magnitude between

the pressure and vacuum tanks. Both tanks are provided with a sight

glass, pressure or vacuum gauge and an air stop- cock valve at the

top end for manual control of the air pressure (positive or negative)

within the tank. The pipe, leading from the vacuum tank to the pump

inlet is provided with two valves, one to isolate vacuum tank from

the pump and the other, fitted on a side inlet tube, to fill the system .

with working medium. Likewise the U- bend on the pump discharge

side is fitted with two valves, one on a side discharge line to drain

the system and the other to isolate the pressure tank from the pump.

A schematic diagram for the system is shown in Fig.31.

A pressure relief valve is also provided between the pressure and

the vacuum tank. For this project a pressure ceiling of 500 mm Hg

has been chosen so the relief valve has been set at that value.

The pressure control valve is operated manually and is of such

type and size that the pressure in the pressure supply tank can be

set at any value between atmospheric ( 0 mm Hg gauge) and the

ceiling pressure ( 500 mm tlg gauge) . When the system is running

the two tanks are one third full, the air above the working fluid being

at 200 mm Hg pressure in the pressure tank and 200 mm Hg suction

in the vacuum tank. The ratios of air /water are calculated as shown

in Appendix D and are based on the desired result that the volume

of air in the tanks would be such that a withdrawal or addition of

one litre of working medium to a tank would alter the pressure of

air above it by a negligible amount. Calculations for the volume

of water flow, i.e. one litre per pulse, are also given in Appendix D.

Page 69: p~~~~ fcs~ y,

- 62 -

:>t ..-1 P-t P-t ::s (/)

s:: 0

•r-l +l u ::s (/)

I (!) !-1 ::s Ul Ul (!)

H ~

u ·r-l .--{

g ~ >-t . n:: (!)

/C. +l

!H 0

~ (!)

"r-l :> .jJ

s:: 0 H ~

. o-. N

Page 70: p~~~~ fcs~ y,

"-

Fig. 30. -Plan view -of the Hydr a ulic Pressure-Suction Supply

CT\ w