p~~~~ fcs~ y,
TRANSCRIPT
![Page 1: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/1.jpg)
_).
-~ } '
l
' ' I
EXTERNAL HEART ASSIST DEVICE
by
., ,. PET AR FRAN 0 BASIC
submitted in fulfilment of the
requirements for the degree
·Of
MASTER OF SCIENCE IN ENGINEERING
UNIVERSITY OF CAPE TOWN
Deeem-BeP 1973
Th e copvric'·4 r ... ·-
U . , <:>"~. o.- r-:}:~ t!. . n•versit-., .. - r- ...... ••-:?Sis is held b
' Rc ' c.. ·-.::;::," T::. ··m Y the / prod t:cq.- ,1 • • • I may be m .. ·., cr V;e wf:o/e or
l not for p~~~~ fcs~ study purposes o~~y p~rt •cat•on. . y, and
![Page 2: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/2.jpg)
I -{ii)-
ABSTRACT
The work described herein is an independent and original work
carried out in the Department of Chemical Engineering on a
heart assist device which was designed, built an'd tested in the
laboratory.
The project can be divided into a mechanical and an electrical
task, both of which were undertaken by the author. The electri
cal task involved design and construction of a portable console
adapted to receive a triggering signal and to activate a series
of controls either in a sequential or singular fashion. The
system also comprises a pulse frequency counter and a safety
shut- off control circuits.
A portable pressure-vacuum supply system and a pressure pul
sating mechanism comprise the work done in completing the
mechanical task. Of the two pressure pulsating mechanisms
built and tested in the laboratory it is shown that a minimum
flow, rigidly enclosed system is superior to a high flow, variable
volume one.
The design of the electronic console, pressure supply system
and pressure pulsating mechanism is described and a descrip
tion of the laboratory tests carried out so far is given.
![Page 3: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/3.jpg)
- ( iii) -
ACKNOWLEDGEMENTS
The author wishes to express his sincere appreciation and
gratitude to Prof. A. D. Carr, Dr. B. M. Kennelly and
especially to Ass. Prof. H. 0. Buhr for their valuable
assistance and encouragement throughout this study.
The author is also grateful to the Workshop Staff of the
Chemical Engineering Department for their assistance
rendered in the fabrication of various parts of the system.
Finally, the generous financial assistance given to the
author by the Chris Barnard Fund, the Medical Research
Council, the Council for Scientific and Industrial Research
and the Harry Crossley Foundation is gratefully acknow-
ledged.
![Page 4: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/4.jpg)
- ( iv) -
TABLE OF CONTENTS
ABSTRACT ii ACKNOWLEDGEMENTS iii TABLE OF FIGURES vi NOMENCLATURE viii
CHAPTER 1
1.1 INTRODUCTION 1 1.2 AR TERIO-ARTERIAL PUMPING 1 1.3 INTRA-A OR TIC BALLOON PUMPING 4 1.4 EXTERNAL COUNTERPULSATION 6 1.4.1 Theory of External Counterpulsation 6 1.5 OBJECTIVES 8
CHAPTER 2
2.1 MEDICAL BACKGROUND 10 2.2 THE HEART 10 2.3 HEART BEAT 13 2.4 CARDIAC CYCLE AND EGG 14 2.5 HEART AS A PUMP 16 2.6 CORONARY BLOOD FLOW 19 2.7 HEART WORK AND EFFICIENCY 19
CHAPTER 3
3.1 EXTERNAL HEART ASSIST DEVICE 23 3.2 DESIGN REQUIREMENTS 24 3.3 ELECTRONIC CONSOLE 26 3.3.1 Test Signal Circuit 29 3.3.2 Shaper and Delay Circuits 29 3.3.3 Pulse Frequency Counter 32 3.3.4 Safety Shut- of£ Circuit 37 3.3.5 Timing Circuit 39 3.3.6 Actuating Circuit 43 3.3.7 Power Supplies 45 3.3.8 Control Circuits 47 3.4 EXTERNAL COUNTERPU LSATOR 52 3.4.1 Pressure Pulsating Medium 53 3.4.2 Sequenced Pulsator 54 3.4.3 Minimum Flow, Rigidly Walled Pulsator 58
1 3.5 HYDRAULIC PRESSURE - SUCTION SUPPLY 60 I
)
![Page 5: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/5.jpg)
-.(v)-
CHAPTER 4
4.1 RESULTS 4.1.1 Sequenced Pulsator 4.1.2 Rigidly Walled Pulsator 4.1.3 Undesirable Effects 4.1.3.1 High Flow 4.1.3.2 Entrapped air 4.1.4 Test on a Volunteer
CHAPTER 5
5.1 CONCLUSIONS AND RECOMMENDATIONS
REFERENCES
APPENDIX A
HEART EFFICIENCY
APPENDIX B
ELECTRONIC COMPONENTS
SN 74121 Monostable Multivibrator High Speed Flip- Flops Phototransistor Coupled Pair MCT2 The Pressure Transducer
APPENDIX C
PROPAGATION OF A PRESSURE WAVE
APPENDIX D
65 65 68 70 70 70 73
75
78
A1
B1
B1 B3 B3 B5
C1
VOLUMETRIC RATIOS D1
Ratio of Air- Water Volumes in the Pressure Tank D1 Ratio of Air- Water Volumes in the Vacuum Tank D3 Volume of Displaced Water due to the Movement of the Leg in and out of the Pulsator Cylinder D4
![Page 6: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/6.jpg)
- (vi) -
TABLE OF FIGURES
Fig. 1. Schematic block diagram of the SIMAS system as used in clinical trials Page 3
Fig. 2. Intra- aortic balloon in position Page 5
Fig. 3. Schematic representation of changes in intra-vascular volume and transmural pressure Page 7
Fig. 4. Anatomic components of the heart Page 11
Fig. 5. Aortic volume pressure relations in different age groups (isolated aortas) Page 12
Fig. 6. Cardiac events cycle Page 15
Fig. 7. Ventricular pressure changes Page 18
Fig. 8. Schematic block diagram for an external heart assist device Page 23
Fig. 9. Pulsation System Page 25
Fig. 10. Block diagram for the external counterpulsator Page 27
Fig. 11. Test Signal Circuit Page 30
Fig. 12. Shaper and Delay Circuits Page 31
Fig. 13. Pulse Frequency Counter Circuit Page 33
Fig. 14. Graph of Pulse Frequency Counter Output Voltage vs Time Page 34
Fig. 15. Graph of Pulse Frequency Counter Output Voltage vs Cardiac Cycle Page 35
Fig. 16. Graph of Tp Period vs Cardiac Cycle Page 36
Fig. 17. Safety Shut- Off Circuit Page 38
Fig. 18. Logic Diagram for the Safety Shut- Off Circuit Page 38
Fig. 19. Articulated Pulsating Mechanism Page 39
Fig. 20. Timing circuit logic Page 40
Fig. 21. Timing Circuit Page 42
![Page 7: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/7.jpg)
Fig. 22.
Fig. 23.
Fig. 24.
Fig. 25.
(vii)
Actuating Circuit for each of the eight solenoid valves
5V DC Power Supply
Front end elevation of the Console
Side view of the Console
Fig. 26. a) Plan view of the leg sheath
Fig. 26. b) Cross- sectional view A-A. of the leg sheath with rubber cuffs placed in pockets
Fig. 27.
Fig. 28.
Fig. 29
Fig. 30
Fig. 31
Fig. 32
Fig. 33
Fig. 34
Fig. 35
Fig. 36
Fig. 37
Fig. 38
Fig. 39
Plan view of the sequenced pulsator
Cross- sectional elevation of the minimum flow, rigidly walled pulsator with leg fitted therein
Front view of the Hydraulic Pressure-Suction Supply
Plan view of the Hydraulic Pressure-Suction Supply
Schematic diagram for the Hydraulic PressureVacuum Supply
Pressure Response Curve for the Sequenced Pulsator
Pressure Response Curve for a Rubber Cuff
Pressure Response Curve for the Rigidly Walled Pulsator
Improved Timing Circuit Output
Graph of Pressure Rise Time vs Entrapped Air Volume
Pressure Response Curve for the Rigidly Walled Pulsator when tested on a Volunteer
SN 74121 logic
MCT 2 to TTL interface
Page 44
Page 46
Page 48
Page 49
Page 55
Page 55
Page 57
Page 59
Page 62
Page 63
Page 64
Page, 66
Page 67
Page 69
Page 71
Page 72
Page 74
Page Bl
Page BS
![Page 8: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/8.jpg)
I
- (viii) -
NOMENCLATURE
A Ameter
'a' Atrial pressure wave
C Capacitor
Cx Pressure propagation wave velocity
'c' Delay due to the electrical and mechanical response of the external counterpulsator
'D' ~ Total delay period
DpF Constant width pulse generator output
'd' Delay period effected by the Delay circuit
ECG The electrocardiagram·
Ek Kinetic energy imparted to the expelled blood from the ventricles
F/F Flip-flop
G The Pulse Frequency Counter circuit input
GND Ground potential
IC Isometric contraction
IR Isometric relaxation
m The mass of blood expelled from a ventricle per beat
ME Maximum ejection
MCT2 Phototransistor-diode coupled pair
M Monostable multivibrator
P,Q,R,S,T ECG waves
rch Chamber ~ressure
~i Intravascular pressure
Ptr Transmural pressure
![Page 9: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/9.jpg)
'·
I
t.Pv
Q
.·
Q
Qc
QRS
R
RE
Tp
TR
Tv
ts
v
Vee
Vs
v
We
ZD
- (ix) -
Mean ventricular pressure rise
The monostable multivibrator and flip-flop output
The monostable multivibrator and flip-flop inverse output
Cardiac output
The ECG complex signal
Resistance
Felaxed ejection
External pressure application period
Transistor
External vacuum application period
Physiological time delay
Voltmeter
Supply Voltage
The Pulse Frequency Counter circuit output voltage
The mean velocity with which the blood is expelled
Cardiac work
Zener diode
![Page 10: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/10.jpg)
I
CHAPTER 1
1.1. rnTRODUCTION
For many years Cardiologists have felt the need for a cardiac
equivalent of the artificial kidney, a device which could support the
circulation temporarily, enabling the patient to survive an episode
of circulatory failure and giving the heart time to 'recover from
some temporary insult such as myocardial infarction (coronary
thrombosis). The concept of circulatory assist is then to provide
means to augment the cardiac output with the objective of increasing
the perfusion of blood to all organs, the heart included, without
increasing the energy requirements of the heart.
1.2. AR TERIO - ARTERIAL PUMPrnG
Since 1957 a circulatory assistance pump, designed to reduce pres
sure within the aorta and left ventricle during systole and to elevate
aortic pressure during diastole, has been in development and experi
mental use /1,2/. In principle, this method relies upon the concept
that reduction of the amount of blood and pressure in the aorta during
cardiac systole results in a reduction in cardiac work~ Catheters
are placed into the central arterial system by way of the femoral
arteries and blood is removed in association with ventricular systole.
This lowers the aortic pressure with the result that the left ventricle
develops less pressure to eject its volume of blood. The volume of
blood removed from the system during systole is then replaced into
the aorta in a retrograde fashion during diastole, augmenting aortic
diastolic pres sure and coronary blood flow. Such a device also
works as an in- series pump, in that it often provides an additional
force for propulsion of blood through other noncoronary vascular
beds, and will, therefore, aid in the force which determines net
venous return.
- 1 -
![Page 11: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/11.jpg)
J
I
- 2-
To be effective, this method requires precise timing. Consideration
must be given to allow for delay from inertia of the blood, particu
larly if the location of the catheters is distal in the arterial tree.
Also, the catheters should be positioned in such a way to enable the
withdrawal of a significant amount of blood at the right time. There
is limitation of the size of catheters which can be introduced into the
aorta through a peripheral blood vessel. Such phased mechanism is
usually triggered by the electrical activity of the heart. However,
the heart in failure may present either a chaotic rhythm or rhythm
which is too rapiQ. for efficient activity of the pump. Past studies
have indicated that counterpulsation is capable of reducing mean
systolic pressure and the time-tension index. Soroff et al I 3/
reported a reduction in left ventricular work and myocardial oxygen
consumption with no significant change in cardiac output.
In later studies 14,51 of Soroff et alit was shown that with synchro
nized counterpulsation in dogs, mean systolic pressure could be
lowered by approximately 40 mm Hg, the time-tension index was
reduced by 39 percent and myocardial oxygen consumption reduced
by 22 percent. Mean diastolic pressure remained at the original
levels but coronary blood flow increased by 50 percent. These
findings suggest that counterpulsation might be a useful technique
in the management of selected patients with ischemic heart disease
and cardiogenic shock.
Rosensw~ig et al /61 have also shown that counterpulsation substan
tially increases coronary blood flow. They attributed this increase
to the increase in collateral coronary flow which resulted from the
increased diastolic pressure provided by the counterpulsation.
One of the counterpulsation devices now available commercially is
SIMAS, the Standard Isochronous Myocardial Augmentation System
I 7 I, which has received limited clinical trial. The operation of the
![Page 12: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/12.jpg)
I
'
- 3 -
PH::IS/OLOG/CRL
MONITOR
ELECTRONIC CONTROl-S
ACTUATOR. AND
H~DRf\ULI C 60PPL~
Fig.l. Schematic block diagram of the SIMAS system as used in clinical trials. Femoral arteries are exposed and cannulated at A and B
![Page 13: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/13.jpg)
I
I
- 4-
system is illustrated1n Fig. 1.
A modification of this principle has resulted in what is known as the
auxiliary ventricle technique I Bl. In this situation, the pumping
chamber is attached directly to the aorta (through a thoractomy)
and the blood is removed from the arterial circuit immediately
adjacent to the aorta and returned directly to the aorta at the end of
a relatively large- sized conduit 19,101.
1.3 INTRA-AORTIC BALLOON PUMPING
This technique may be considered a derivative of arteria-arterial
pumping but it differs in its mechanics in an important way. Whereas
with the arteria-arterial pumping there is the withdrawal and reintro
duction of blood to the cardiovascular tree, no blood is extracted from
the arterial system with the balloon technique I 11 I.
A catheter- balloon system is introduced into the aorta from the
femoral region and placed in the intrathoracic aorta as shown in
Fig.2. The balloon is filled during ventricular diastole, thereby
increasing the aortic root pressure and improving coronary flow,
and deflated during ventricular systole. Such system has been built
and tested at the University of Cape Town /12,13 I.
The major disadvantages of both the arterio- arterial and intra- aortic
balloon pumping are that such counterpulsation methods involve expo
sure and cannulation of femoral arteries, the use of an extracorporeal
blood handling device, the requirement of sterile technique and the
necessity for administration of ~nticoagulants. A potential hazard
is rupture of the balloon and rupture of the vascular wall by over
distention of the balloon.
The ability to augment pressure is somewhat limited by the size of
![Page 14: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/14.jpg)
I
\,
{ I
\ J
I \
- 5 -
Fig. 2. Intra-aortic balloon in position
·-------- _____ .. ____ _
![Page 15: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/15.jpg)
I I
-1
RA.
l.\1.
v.c. Ao.
- 7 -
Ptr = Pi- Pch
t
L.V.
?ressure Chamber
Pch = 1 bar Pch = 0,9 bar Pi = 1,25 bar Ptr = 0,35 bar
Pch = 1,1 bar Pi = 1,25 bar Ptr = 0,15 bar
Pi = 1,25 bar Ptr = 0,25 bar
Fig.3. Schematic representation of changes in intravascular volume and transmural pressure ( Ptr) with changes in chamber pressure ( Pch) while the intravascular pressure (Pi) remains constant.
![Page 16: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/16.jpg)
- 8 -
l>etween the internal pressure and the pressure which surrounds the
tube.
In the vascular bed the transmural pressure is the difference between
the intravascular pressure, which changes continuously, and the extra
vascular pressure, which is normally essentially constant and very
nearly equal to atmospheric pressure. Certain portions of the vascular
bed which are affected by muscular activity respond ,to variations in the
extravascular pressure so created. For example, venous return from
the limbs is increased when the limbs are exercised and the coronary
vascular system ~s affected by myocardial contractions.· If means are
provided to vary the pressure external to portions of the vascular bed,
the capacity of the affected portion of the vascular bed will respond to
the changes in transmural pressure which are produced.
Fig. 3 schematically illustrates that as the external pressure is
lowered in relation to the ambient pressure, the capacity of the disten
sible vascular bed increases and vice versa. The resulting change in
capacity is determined by the pressure-volume characteristic of the
cardiovascular system.
In the circulatory system, the venous bed acts as a large reservoir
which readily compensates for changes in blood volume distribution
when cardiovascular equilibrium is disturbed. This is due to the
fact that in the typical distribution of the blood volume there is approxi
mately 15 percent of the blood in the arterial system, 75 to 80 percent
in the venous bed and the rest is accommodated in the capillary bed
/19/. Because of such high capacity the venous bed will be respon
sive to relatively small changes in external pressure whereas larger
excursions in external pressure will be required to affect the arterial
system.
1.5. OBJECTIVES
A project was undertaken to design, construct and test a device which
![Page 17: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/17.jpg)
- 9 -
would apply positive pressure to the legs of a patient during diastole
and switch to vacuum for the rest of the cardiac cycle. Such appli
cation would be synchronous with the cardiac activity so that:
( i) the heart work is reduced,
( ii) the cardiac output is increased,
(iii) the perfusion of the myocardium is increased.
The obvious advantage of using an external mechanism is that it does
away with surgery and many logistic problems that accompany it.
However, the clinical environment of the typical patient in cardia
genic shock places many constraints on the design of the system.
The ideal unit would have to be a portable one, which is relatively
quiet in operation and requires no elaborate adjunct equipment, does
not interfere with normal patient care requirements and in the case
of an emergency is readily applied. Furthermore the application of
the system should not hinder standard forms of medical treatment
and preferably cause minimal disturbance to the patient.
![Page 18: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/18.jpg)
CHAPTER 2
2.1. MEDICAL BACKGROUND
The human body is composed of a vast number of cells all of which
are grouped and organised according to their specific functions.
Being contained in a confined space, the body cells can function and
survive only when the immediate environment is adequately supplied
with the essential nutrients, i.e. when the pH, temperature and con
centration of various substances is maintained within certain limits.
The conveying system for such substances is the cardiovascular
system, which comprises the heart, arteries, veins and capillaries
on one hand and the blood on the other. . The transportation of the
nutrients and wastes to and from the cells is effected by blood, which
in turn is circulated along the cardiovascular bed and distributed to
the various tissues in a manner according to their functions. The
source of the kinetic and pressure energies, required by the blood
to be circulated through the body, is the heart.
2.2. THE HEART
The fibrous 'skeleton' of the heart, as illustrated in Fig. 4.c.,
consists of four valve rings joined together to which the two major
arterial trunks and all four cardiac chambers are fastened.
The two atria, which resemble a thin-walled, shallow cup of myo
cardium, divided by a partition down the centre, are fastened to
the superior surface of the mitral { M) and tricuspid ( T) valve
rings as shown in Fig. 4.a. The remaining two chambers, the
right and left ventricles, are fastened to the entire circumference
of the fibrous skeleton and form the lower part of the heart. This
arrangement is illustrated in Fig. 4.e. The ventricles serve as
- 10 -
![Page 19: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/19.jpg)
)
right--atrium
A. V. valves~.
- 11 -
---aorta
{a) {b)
fibrous skeleton
{c)
---left ventricle
{d) {e)
Fig. 4. Anatomic components of the heart
![Page 20: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/20.jpg)
200
a' . :c 140
E E
uJ ~ ~ 80 (/) (/) LJ o! tl-
2.0
- 12 -
AGE
2.6 42.
20 40 60 80 100 12.0 140 160
VOLUME. cc/rn'" BOD~ SURFACE
Fig.S. Aortic volume pressure relations in different age groups (isolated aortas)
![Page 21: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/21.jpg)
- 13 -
the major source of energy for the circulation of blooq ~dare
composed of sheets of myocardial fibres encircling the ventricula,r
chambers I 20-22 I. From a functional point of view, this muscu
lature can be divided into two groups of myocardial bundles, the
spiral muscles and the deep constrictor muscles 1231.
The two semilunar valves, pulmonary ( P) and aor,tic (A) as shown
in Fig. 4.d., are very similar, each consisting of three sy~metrical I
valve cusps. The atrioventricular (A-V) valves are fasten~d to the
inferior surface of the mitral ( M) and tricuspid ( T) valve rings with
chordae tendineae extending from the inferior margins of each valve
leaflet and fastened directly to the internal surface of the ventr;icul~r
walls. All four valves open and close rapidly and seal completely
against high pressures with minimal obstruction to flow.
Like the atria, the two major arterial trunks, the aorta anc;:l the p\ll
monary artery originate at the superior surface of the semilunar
valve rings as shown in Fig. 4.b. In the external counterJ?ulsation
the location and the pressure-volume characteristic of these two
vessels influence the design of the assist device, since the prima:~."y
objective of such an external mechanism is to raise and lower
pressure inside the arterial tree. The pressure-volume relations
vary widely in different individuals in the same age group ~d the
curve also tends to shift towards larger volume and less distensi
bility as subjects grow older I 24 I. Such a characteristic is
illustrated in Fig. 5.
2.3. HEART BEAT
To produce efficient pumping, the complex mass of myoc~rdial
bundles must contract more or less sim1:11taneously. Co-ordin~te<;l
contraction of the complex pattern of myocardial bundles stems
from the syncytial arrangement of the myocardial fibre~; excitatio:p.
![Page 22: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/22.jpg)
- 14-
beginning at one site spreads to all other contiguous areas. Exci
tation of the heart is normally initiated by an impulse which is
generated by the sino-atrial (S-A) node (a small mass of specialised
myocardial tissues embedded in the atrial wall near the entrance of
the superior vena cava) and which spreads rapidly in all directions
through the atrial musculature (atrial polarisation) . As it approaches
the interatrial septum, the wave of excitation reaches another mass of
spec;i.alised conducting tissue, the atrioventricular'( A- V) node, which
is located near the posterior margin of the interatrial septum close to
the entrance of the coronary sinus. After a slight delay ( 0,08 to 0,12
second) I 25/ at the A- V node, impluses are conducted by the Purkinje
system into the ventricles where a wave of excitation spreads from
the endocardial surfaces through the ventricular musculature. This
normal sequence of excitation initiates myocardial contraction and
establishes the mechanical events of the cardiac cycle ..
2.4. CARDIAC CYCLE AND ECG
A recording of electrical potentials, associated with waves of exci
tation which spread through the heart, is known as the electro
cardiogram ( ECG) . The normal ECG trace consists of five waves
(:I? ,Q,R,S and T) which correlate with the major events of the cardiac
cycle as shown in Fig. 6. The P- wave, the QRS complex and the T
wave appear during atrial depolarisation, ventricular depolarisation
and ventricular repolarisation respectively. The P-wave is normally
100 milliseconds in duration, followed by the P-Q interval during
which the atria pumps blood into the ventricular chambers ('a'
wave of the atrial pressure as illustrated in Fig. 6.) Following
this brief isoelectric period there is a series of deflections, the
QRS complex.
With the onset of the Q-wave the isometric contraction ( IC) of the
ventricles begins, accompanied by a rise in intraventricular pressure.
![Page 23: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/23.jpg)
> f
p
- 15 -
R
ECG T
Q s ,Q 5
IC ME RE tR DIASTOLIC DRAIN
AORTIC PRESSURE
VENTRICULAR VOLUME
400 800 12.00
TIME IN MSECS
Fig.6. Cardiac events cycle
T
inlra ve.,tricufctr p..essu.re
![Page 24: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/24.jpg)
- 16 -
When the intraventricular pressure reaches that within the aorta,
the aortic valve opens and the maximum ejection (ME) phase takes
place. In man the duration of QRS is about 80 milliseconds with
a normal upper limit of 100 - 120 milliseconds. The final T-wave
is a broad deflection. Its rise, peak and fall coincide approximately
with the relaxed ejection ( RE) phase, the closing of the aortic valve
and the isometric relaxation ( IR) phase respectively, as shown in
Fig.6.
With the average duration of the S- T interval equal to 270 milli
seconds, the duration of QRST phase is between 350 and 400 milli
seconds for a normal heart rate of 70 beats per minute. In the
design of the external counterpulsation mechanism this criterion
will be used as well as the phenomenon that a QRS complex cannot
be followed by another before a T -wave.
The heart performs its greatest work during the isometric contrac
tion and maximum ejection phases. Therefore if myocardial work
is to be reduced, the external counterpulsation must be timed to
reduce pressure in the ascending aorta before the aortic valve opens.
To maintain system flow and pressure, the pressure is to be increased
during the diastolic period.
2.5. HEART AS A PUMP
The average adult human heart weighs about 290 grams and has a
stroke volume of approximately 70 ml. of blood. For a normal
heart rate of 70 beats per minute the cardiac output is nearly 5 litres
and since this is about equal to the total amount of blood in the vascular
system, the blood is therefore circulated once every minute. Heart
output can, however, increase six-fold I 26/ in a severe exercise.
The atrial pressure curve as depicted in Fig.6. shows three major
![Page 25: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/25.jpg)
- 17 -
pressure elevations, known as the 'a', 'c' and 'v' waves. Under
normal conditions there is a continuous flow of blood from the
great veins into the atria and approximately 70 percent of this flows
directly into the ventricles before the atria contract, causing addi
tional 30 percent filling ('a' wave). The left atrial pressure then
rises to 7-8 mm Hg while that of the right atrium reaches a some
what lower peak: 4-6 mm Hg. The 'c' wave is caused during
ventricular contraction by reflux of blood out of the ventricles into
the atria on one hand and bulging of the closed A- V valves towards
the atria on the other. Because of the large amounts of blood
having accumulated in the atria by the end of systole, the atrial
pressure rises ( 'v' wave) but falls off steeply as the ventricular
pressure returns to its low diastolic value when the pressure in
the atrial chambers push open the A- V valves and allow the blood
to flow rapidly into the ventricles.
The function of the ventricles as a pump is entirely different from
that of the atria. The rapid filling of the ventricles occurs during
the first third of diastole, followed by diastasis - the middle third
of diastole when blood flows from veins directly into the ventricles
- to be completed by the last third phase during which atria contract
giving an additional thrust to the in-flow of blood. Both the filling
and the emptying of the ventricles is depicted in Fig.6. by the ven
tricular volume curve.
With the onset of systole, the ventricular contraction is accompa
nied by a rise in ventricular pressure which causes the A- V valves
to close immediately and only after the pressure level has risen
to 80 mm Hg in the left and 8 mm Hg in the right ventricle, as
illustrated in Fig. 7., do the aortic and pulmonary artery valves
open. Blood then begins to pour out of the ventricles, about half
of the outflow occurring during the first quarter of systole and
most of the remaining half during the next two quarters. In the
![Page 26: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/26.jpg)
- 18 -
LEFT VENTRICLE AND AORTA PRESSURE CURVES
I J.. r I I aonk vaive I 1/ cfoseci I
HOR.rtc. Pi?!£ . l I Ss vt~e
L.V. ----t-1
SYSTOLE UlRSTOLE
RIGHT VENTRICLE AND
-- - - - -- '70 wem 1-t~
PULMONAR~ ARTER~ PRESSURE CURVES
- -- - - 25 rnm H~
_ _ _ _ _ _ _ 8 m m 1-1~
~~::::=--1----~~~~~~f---- -- -·- 0 mm H~
S~STOLE DIASTOLE:
Fig. 7. Ventricular pressure changes
![Page 27: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/27.jpg)
- 19 -
last quarter of the systole ventricular relaxation begins suddenly,
allowing the intraventricular pressure to fall rapidly, whereupon
the elevated pressures in the large arteries immediately snap the
aortic and pulmonary valves closed. When the ventricular pressure
reaches the low diastolic levels below atrial pressures, the A- V
valves open and a new cycle begins.
2.6. CORONARY BLOOD FLOW
The coronary blood flow of a normal heart has been found to be 7 to
10 percent I 27/ of the total cardiac output. Since the coronary
circulation depends essentially on the difference in the pressure
between the right atrium and the aorta, an increase in diastolic
pressure will increase the coronary flow and at the same time the
oxygen supply. Under normal conditions, myocardial oxygen con
sumption is approximately 8 to 10 ml of 02 per 100 grams of heart,
but it can increase severalfold with exercise and decrease moder
ately under conditions such as hypotension and hypothermia. The
myocardial oxygen consumption correlates reasonably with the
area under systolic portion of the left ventricular pres sure curve
I 281 so that the effect and usefulness of a cardiac assist device can
be evaluated using this index.
When the blood, i.e. the oxygen supply is cut off tc;> a part of the
myocardium, myocardial infarction takes place; The muscle cells
die and the area becomes necrotic. One of the main causes of
myocardial infarction is coronary thrombosis which occurs when a
coronary artery is occluded by a thrombus that develops within the
lumen of a sclerotic vessel.
2.7. HEART WORK AND EFFICIENCY
Since William Harvey' s disclosure of the nature of the circulatory
system in 1632 the pumping action of the heart remains a difficult
![Page 28: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/28.jpg)
- 20 -
parameter to measure and a great need for a simple, non-invasive
method to obtain information concerning the mechanical activity of
the heart, exists. In recent years a wide variety of techniques I 29 I
has been proposed to determine cardiac output by indirect methods,
most of which have serious limitations and all of which are subject
to various sources of error. For the purpose of this study the
work done by the heart will be approximated as that due to two
factors only:
If
( i) Cardiac work done in expelling the blood out of the
ventricles against the external pressures, a·nd'
( ii) Kinetic energy imparted to the expelled blood.
We = Cardiac work
Qc = Cardiac output
APv = Mean ventricular pressure rise
then We = Qc x APv ....................... ( 1.1)
Considering the complexity of the situation, the cardiac output is
not an easy parameter to measure. The stroke volume (output
per beat) is not directly proportional to arterial pressure since
the pressure-volume relationship of the arterial system is not
constant nor uniform among individuals, but also non-linear from
high pressure to low. Hence various approximation techniques
have been developed. By the dye dilution technique advocated by
Hamilton and Remington I 30 I the pulse pressure correlates
roughly with stroke volume, namely a pressure rise of 1 mm Hg
is equivalent to about 1 tc of stroke volume/ sq m over the normal
pressure range. Even if the cardiac output were to be computed
according to the Fick principle /31/, such method is at best± 10
percent accurate.
![Page 29: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/29.jpg)
If Ek
m
v
=
=
=
- 21 -
Kinetic energy imparted to the expelled
blood
The mass of blood expelled
The mean velocity with which the blood
is expelled
then Ek = .................. ( 1. 2)
In this case too, both parameters ( m) and ( v) are difficult to mea
sure. However, Erlanger and Hooker I 32 I have shown that it is
justifiable to use the product of the pressure pulse by the pulse rate
as an index to the relative velocity of the ejected blood but only upon
the assumption that the following factors remain constant:
a) the rate of flow from the arteries into the veins;
b) the rate of systolic output;
c) the distensibility of human arteries;
d) the amount of blood in the arterial system.
From the Example on Heart Efficiency in the Appendix it follows that
the work done by the left ventricle per beat is 0,8935 + 0,00875 Nm or
0,90225 Nm and for the right ventricle 0,1401 + 0,00875 Nm or 0,14885
Nm. It follows then that the work expenditure of the left ventricle is
nearly 7 times that of the right ventricle, hence any reduction in the
work load on the left ventricle will greatly assist the heart.
Also, the equation ( 1.1) above emphasises the important fact that
the work of the heart is increased by factors raising either its output
or pressure against which it has to operate, therefore all effort in
the design of an external heart assist ought to be directed primarily
toward the latter, i.e. the reduction of systolic pressure and also
toward increasing diastolic pressure which would simultaneously
![Page 30: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/30.jpg)
I
- 22-
increase the perfusion of the myocardium.
If, during the time the aortic valve is open, the mean pressure in
the aorta were to be reduced from 108 mm Hg to typically 88 mm Hg,
17 percent reduction in work would be achieved. It is toward this
end that the project is directed.
![Page 31: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/31.jpg)
CHAPTER 3
3.1. EXTERNAL HEART ASSIST DEVICE
The design and construction of the external pressure pulsating
device to produce the effects of counterpulsation, was undertaken
in three stages, the first an electrical one and the remaining two
mechanical.
In its desired form the external heart assist, based upon the prin
ciples discussed in sub- section 1.4.1., comprises an electronic
console, an external counterpulsator and a pressure/vacuum
supply. The three modules are integrated as shown in Fig.8. to
provide a mechanical support of the active vascular system and,
at the same time, to reduce the myocardial work demand placed
on the heart under various conditions.
PRESSURE/
/VACUUM'
SUPPLY
.ELECTRONIC CONSOLE
~ 00
d. ... z 0 \J
c::::> PRESS.
... 0 t1 .. z 0 u
VAG.
EGG SIGNAL
PRESSURE SIGNAL
{1, /\
I. ,! .. ~;
SYSTEM
~· 't} .. ,.• ·-, . ' ·~·
EXTERNAL C OUNTERPU LSAT OR
Fig.8. Schematic block diagram for an external heart assist device
- 23 -
I
![Page 32: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/32.jpg)
- 24-
3.2. DESIGN REQUIREMENTS
With the onset of systole, i.e. the receipt of the QRS triggering
signal, the system is to apply pressure to the vascular bed for
approximately half the cardiac cycle and vacuum for the remainder
of the cycle. However, the pres sure in the aorta is· to be affected
by the externally applied pressure only after the aortic valve has
closed, i.e. during diastole, hence the application of pressure
period Tp is to be delayed for a period 1 d 1 as shown in Fig. 9.
Ideally the delay 1 d' would be equal to the systolic period if it
were not for two more delaying factors, namely the physiological
·time delay 1 ts' and delay 1 c' due to the electrical and mechanical
response of the external counterpulsator. These three delays
comprise the total delay period 1 D' as illustrated in Fig.9.
The type of external counterpulsator used in this project is a·
pressure pulsating leg unit into which the patient' s legs fit. The
reason for using such a unit and its design are discussed in detail
in sub- section 3.4. After inserting the legs, the container is filled
with working fluid. This fluid is connected to both the pressure
and the suction supply tank via a solenoid valve. On receipt of the
control signal the pressure solenoid valve opens, connecting the
working fluid with the pressure supply thus exerting pressure on
the legs. When the pressure valve closes, the suction valve
opens causing vacuum to be effected on the legs. The two solenoid
valves operate alternatively.
Since heart rate changes, the system should accelerate or decele
rate its pulsing action accordingly. Also, should there be a
spurious beat during diastole, i.e. while the pressure is being
applied to the vascular system, such pressure is to be removed in
![Page 33: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/33.jpg)
1\ I \
' \ I I \ ' l 1 \ I I I l I
~ : I ~ I I ' ' I ' I l \ I \,. • .,
- 25 -
ECO PHVS.
MONITO'R.
ELECTRONIC
CONSOLE
,-------- --1
t====+:i~<:::J~:::;t=l PR.E SS. TANK
SUCTION
TANK I I 1 I
L - - - - ·- --- - - - - -- - - - ~' HYORRUL.IC SUPPLY
'PRESSURE 1>Ut.SATING UNIT
ECG
AORTIC PRESSURE I
I ,
. AORTIC VALVE
CLOSED
' '
SPURIOUS BEAT
~ II
'
\ I I \ I \ --- ,, ....... _________ ,.. •!
SYSTOLE I DIASTOLE
I
SYSTOLE
.. D~-t -d J-c~ I ,......--........ --:--~ ~-d -~!
Tv
I ' j DELAYED EFFECT OF EXTERNALLY APPLIED
I r-tsj PRESSURE ON THE CARDIOVASCULAR SYSTEM phy~:ioloqico.l ~im
aldav
Fig. 9. Pulsation System
![Page 34: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/34.jpg)
I
- 26 -
the shortest possible time, as shown by the dotted lines in Fig.9.
Both the time delay 'd' and the pressure/vacuum applied should
be adjustable.
To meet the above objectives the electrical task involved building
and testing an electronic console to monitor, control and synchro
nise the action of the external counterpulsator.
3.3. ELECTRONIC CONSOLE
In the construction of the electronic console extensive use has
been made of solid state integrated circuit chips. Truth tables
together with other relevant data is given in the Appendix B.
The ECG signal, which is picked up by dermal sensors placed
on the patient's skin, as shown in Fig.9., is processed and ampli
fied by a physiological monitor. The greatest portion of this work
was expended on the design and construction of an electronic con
sole, the block diagram of which is given in Fig.lO. The console
was designed to accept a preconditioned triggering signal from an
existing physiological monitor. The monitor is fitted with a dual
channel cathode- ray oscilloscope specifically designed to display
ECG and aortic pressure wave signals. In this manner the monitor
provides a continuous ECG triggering signal to the console and also
provides means for visually setting the triggering level at which
the system will be activated.
For experimental purposes the console can provide a test signal
at a rate approximately that of the two heart- beat limits, namely
60 and 180 beats per minute. Both the output of this Test Signal
circuit and the input for physiological monitor processed trigger
output are fed to a selector switch. The selected triggering
·signal is then fed into a Shaper, which is a highly noise- immune
![Page 35: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/35.jpg)
I t
I
L&J _. 0
~ 0 (.)
~ z 0 (/. 1-u w ...J au
r---------, PH~SIOLOGICAL
MONITOR
, PULSE FRE~:
COUNTER
POWER SUPPUI
sv
I
~
/
...
ECG / SIGNAL /
TEST SIGNAL
/ /
/ /
,, t SWITCH
SHAPER
D£LA~
CCT
,, ltMING CCT
~ , ~ , , ACTUATING
CCT
... -
TEST SIGNAL CCT
, SAFET~
SHUT-OFF
I I L _______ .... _____________ ..... __ -- ____ _j I
PRESSURE PULSATING LEG UNlT
~-------------------, PRESS. SUPP~
t----r PUMP,__ __ ...,
Fig. 10. Block diagram for the external counterpulsator
- 27 -
I
I SUCTION
SIJPPLV
![Page 36: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/36.jpg)
I
- 28-
mono stable multi vibrator. The Shaper will accept any type of
trigger with a positive slope and shape it into positive- step
trigger suitable for a Delay circuit. The Delay circuit introduces
a time delay between receipt of the ECG signal and initiation of the
pressure pulse. The time delay 'd' produced can be fully adjusted
from 0 to 120 milliseconds. At the end of the time delay 'd' the
pressure phase Tp is applied as shown in Fig.9. When properly
set to allow for system and physiological time lags,:the period of
the effect of pressure application Tp corresponds with cardiac
diastole. The use of the ECG signal to trigger the delay 'd' , i.e.
the pressure removal Tv, is advantageous for reasons of patient
safety since with such a mode the heart will never be pumping
against a pressure wave produced by the assist device. This is
especially important in the case of a premature beat.
The triggering signal from the Shaper is fed into a Pulse Frequency
Counter circuit and a Safety Shut- off circuit. The latter monitors
the ECG signal and detects any spurious beats, whereupon it
immediately removes the externally applied pressure as illustrated
by the dotted lines in Fig.9. The output of the Pulse Frequency
Counter circuit is the pulse size controlling voltage of the Timing
circuit which determines the length of Tp periods of the cardiac
cycle. Each triggering signal represents a heart beat and since
heart rate changes, the rate of application of external pressure will
have to change accordingly and the period Tp of the pressure appli
cation as shown in Fig.9. will have to be increased or decreased
proportionately. Hence the input to the Pulse Frequency Counter
circuit is made proportional to its output, i.e. the number of
cardiac cycles per unit time which the Counter receives determines
the pulse size voltage, thus compensating automatically for varying
pulse rates. In this project one beat per second (i.e. 1000 milli
seconds cardiac cycle) was chosen as the lower and 3 beats per
second (i.e. 333 milliseconds cardiac cycle) as the upper heart'
rate limit.
![Page 37: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/37.jpg)
I
- 29 -
From the Delay circuit the signal is channelled to a Timing circuit
which triggers an Actuating circuit which, in turn, energises two
electrical solenoid valves as illustrated in Fig.10. On opening, one
solenoid valve connects the pulsating mechanism with a positive
pressure supply and the other with a negative pressure supply. The
valves operate alternatively and are timed to remain open through
out the period of application of positive (or negative) pressure.
Being of an industrial type, they operate on 220 volt~, hence the
Actuating circuit acts as a shunting system between the high 220
volts solenoid valve side and the low 5 volts electronic circuitry.
The entire console is mounted in a portable cabinet on wheels and
uses only a single source of 220V, 50Hz power supply. Thus need
for special power line is obviated.
3.3.1. Test Signal Circuit
The Test Signal circuit is illustrated in Fig.11 and comprises two
monostable multivibrators I 33 I TTL units ( SN 74121) connected
in series to provide 1 beat or 3 beats per second. The monostable
multivibrator No. 1. output gives the OFF pulse of constant width
A ( 230 milliseconds), while the multivibrator No.2. determines
the ON pulse of the two widths B and Bl. · The two set values B
and Bl ( 770 and 100 milliseconds respectively) are obtained by
shunting two parallel resistors, 10 kilo-ohms and 4,7 kilo-ohms
across the 10 kilo-ohms trimpot variable resistor.
3.3.2. Shaper and Delay Circuits
Both the Shaper and Delay circuits comprise a single SN 74121
monostable multivibrator unit, as shown in Fig.12. The Shaper
receives the triggering signal through a selector switch, either
from the physiological monitor or the Test Signal circuit and
![Page 38: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/38.jpg)
I
OUTPUT
I
A
A
- 30 -
.------+--~+ H~ 011pF'
ISO A
tOk~
f/
B'
Monostab1e Mu1tivibrator No. 1. SN 74121
A
Monostab1e Mu1tivibrator No. 2. SN 74121
Vee =5V
B
B'
A
second
Fig. 11. Test Signal Circuit and Output Logic
![Page 39: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/39.jpg)
I
t
- 31 -
PHYSIOLOGICAL MONITOR TEST SIGNAL CIRCUIT OUTPUT OUTPUT n--n
Vcc----4
D E L A Y Monostab1e Mu1tivibrator
SN 74121
I t
I I t I
I ' l - - - --- - - - -- _,
~----------------~l~-------------------.•Vcc
S H A P E R ( Monostab1e
Multi vibrator SN 74121
L---------~ TO P. F. C. CCT
TO TO SAFETY SHUT-OFF CIRCUIT
Fig. 12. Shaper and Delay Circuits
![Page 40: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/40.jpg)
I
I
- 32-
shapes it to an acceptable step trigger for the Delay circuit. In
the case of the Shaper multivibrator, the internal timing resistor
( 2 kilo- ohms nominal) and the internal tim~ng capacitor are used
to obtain minimal delay of 30 nanoseconds.
The .Pelay circuit employs an external timing resistor ( 10 kilo- ohms
variable) and a capacitor ( 15 microfarads) to obtain delay 'd' from
0 to 120 milliseconds, the value of which is set manu.ally. The
Timing circuit requires a positive- step trigger, so Q is used as the
Delay circuit output.
3.3.3. Pulse Frequency Counter
The Pulse Frequency Counter circuit, as shown in Fig. 13., is of a
discriminator type which, depending on the number of triggering
signals it receives at input c.:;., supplies a proportional voltage at the
output terminals A and B. The circuit comprises a constant width
pulse generator ( monostable multivibrator type SN 74121 with
external timing components R = 20 kilo- ohms variable and C = 22
microfarads giving the output pulse of constant width DpF = 230
milliseconds) and two BC 169 n-p-n transistors TR1 and TR2 con
nected in cascade with a 220 microfarads capacitor between the out
pl,lts A ( 0 volts) and B ( Vs volts}. The resistors R2, R3, R5 and
R6 are adjusted so that the output voltage Vs is approximately 4 volts
at 3 beats per second and 0,6 volts (average} at 1 beat per second.
Fig.14 illustrates the Pulse Frequency Counter circuit output for
1 and 3 beats per second inputs as well as its response to abrupt
changes in heart rate from 3 to 1 beat and from 1 to 3 beats per
second. The relationship between the Pulse Frequency Counter
output voltage and the cardiac cycle time length, which is given in
Fig. 15. is given for values of cardiac cycles between the lower
( 1000 milliseconds) and the upper limit ( 333 milliseconds} and is
not linear.
![Page 41: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/41.jpg)
s""
Mo,-to.s.to..bte I SN '7412.1 Q 1 ""./Wt, l K Mu.tti. vibrnhr
INPUT G
(FROM SHAPER) \ CIRCUIT
o"
•w ~~
TRI
"' -s"" 'POWER SUPPLY
__ Cz 2.2.0~lF
Fig. 13. Pulse Frequency Counter Circuit
.;
..... --
s"
oB L, ----,·r-vvs
TR2. BCt6'
--'-- c3 220pF
.~ A
l.V l.V
![Page 42: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/42.jpg)
~ !j 4
~ I
l.aJ (!) ~
0: !::; g2 1-:l 0..
~ 0
n n+l
• .l
Fig.l4. Graph o£ Pulse Frequency Counter Output Voltage vs Time
th .. ee beats /second
one beQ~ / second
TIME IN SECONDS
Vma.x
- - - - 97 % V mox
---- 80 '/o Vmo...x
40 Jo Vmax
w ~
![Page 43: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/43.jpg)
~ LIJ t- II)
-/
Fig.l5. Graph of Pulse Frequency Counter Output Voltage vs Cardiac Cycle
5 ~ 0 0 4 U"> '·""·
3~----------~.~ •
, LLI Ue> Za: w!:i :lo ~·> 2
a:t-L&. :J
' • -'t
Cl ~ 1- I (J) :J ..J 0 :::> p..
·~·~. I ------- '"--- __:__ -- -- --- - . I ........_• ............._
_ _ _ - - - - - I I ----·---- I I
I I I
200 :500 400 500 600 .,00 800 qoo 1000
CRRDIRC C~CLE IN MSECS
J
w U1
![Page 44: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/44.jpg)
$00
U) 400
0 w Ul ~ 300
0 0 -~ L&J
200
P-
f-L I
/./·""
100
300 400
I • ·c' ~-
Fig.l6. Graph of Tp Period vs Cardiac Cycle
/./·
.~·/
././
500 600 700 800 900
CARDIAC C~CLE T MSECS
w ~
.<.;0.,.
![Page 45: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/45.jpg)
I
I
I
- 37 -
For experimental use Tp and Tv, the two cardiac cycle pha~es when
either pressure or vacuum is applied externally to a section of the
vascular bed, have been made nearly equal but their ratio can
readily be adjusted. When tested, the Timing circuit, in conjunction
with the Pulse Frequency Counter, gave 480 milliseconds and 160
millisec;:onds long Tp periods for cardiac cycles of 1000 and 333
milll.seconds respectively. Fig.l6. illustrates the relationspip
between Tp period and the cardiac cycle time len~th . .' This relation
ship is linear with a maximum deviation of 3 percent.
3.3.4. Safety Shut-Off Circuit
The purpose of the Safety Shut-off circuit is to monitor the ECG
triggering signal and, should there be a spurious signal or a sudden
acceleration in heart rate during the Tp phase of the previous beat,
to remove the external pressure as quickly as possible and reset
the system to a new cycle, i.e. to effect the vacuum phase Tv
. immediately.
The Safety circuit makes use of a quadruple 2-input NOR gate TTL
unit (SN 7402 N) as illustrated in Fig.17. One gate monitors Q
output from the Delay circuit. This Q output suppresses the pres
sure signal and its suppressive effect is achieved by enabling th~
controlli:n.g NOR gate output (pin No.4.) to be at the logical 1 during
the Tp period only as shown in Fig. 18. The controlling output
must be at logical 1 for pressure signal to be on. It is normally
at such enabled state when the Timing circuit requires Tp, but
can only maintain 1 while Q indicates diastole. Should Q change,
i.e. when there is a spurious beat, Tp will immediately be switched
off as indicated by the dotted lines in Fig.18. The output signal
from the controlling gate will cear the appropriate flip- flops in
the Timing circuit, resetting the counterpulsator to vacuum phase
Tv of a new cycle.
![Page 46: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/46.jpg)
' I
• ·'
r
I
I
OUTPUT TO TIMING CCT
+Qo FROM DELAY CIRCUIT
t Q, I PRESSURE ON I SIGNAL FROM TIMING CIRCUIT
Fig. 17. Safety Shut-Off Circuit
ECG ! ..
SYSTOLE ..,
J
OUTPUT (1 o) .. I
#I., J
OUTPUT (1 J - I
~ {INPUT (.6
:i"' '5 ._ \N'PUT ( S IX([
t-" % · OUTPUT e
) .. I
) I -('-If I
AORTIC VAL \If' CLOSED
I
I I
I I I I
I
I j I I I
I I I I
i I I
Tp
SPURIOU$ .BERT
I
I .1 j
..: ; ;t .. I
DIASTOlfE i 1-- ---- - ---! r--------i i i I
i L i
I J
---------I i J
~J Tv
Fig. 18. Logic Diagram for the Safety
Shut-Off Circuit
- 38 -
IY
'
I'"---
I
'- ---
r----I
I
L.--
I t----
... ____
r--.1.
![Page 47: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/47.jpg)
I
--t ' '
r
I
- 39 -
3.3.5. Timing Circuit
The timing circuit we de signed for use in conjunction with an
. compartmented pressure pulsating mechanism where pressure
and vacuum are applied in a sequential fashion, from one end of
the pulsator to the other. If the leg pressure pulsating unit were
divided into four adjacent but pressure independent sections as
shown in Fig.19., then with the onset of the Tp phase of the
cardiac cycle all four sections would be pressurized one after
the other, starting with section No. 1.
solenoid
valve
I I I I
I
common vacuum
I I I
I I
I , ' ... ~
manifold
' ' ,. ~
' ' I I '4" I l \31 I I \ I
:0
CD I I ",
'
I I I I
' I \ I I
• I
I
' ' \ ' ,~·
solenoid
valve
I I I
common pressure
manifold
Fig. 1?. Articulated Pulsating MecP.anism
![Page 48: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/48.jpg)
r
\
ECG
SHAPING CCT Q
"""
SRFm SHUT Ol'F CCT ~
DELAY Q
M 1 Q
-M 1 Q
M 2 Q
M 3 Q
M 4 Q
F/F 1
F/F 2
F/F 3
F/F 4
--. . •
.
.
~
4
. ..
- 40 -
CARDIAC CYCLE __. EARLY )( BEAT '-.t
I I
~ SYSTOLE DIASTOLE
1\ SYSTOLE
y '!" '• n I I
,.__d. ~I~ Tp ~ t I I
I
:
n ll l
I
OFF -ON I OFF I J
ON I OFF I
ON I OFF
.J ON l I I
. Tv4 f. Tpf+ I
'X
Tv3 ·I--Tp3 ~
Tv2--,-Tp2
Tv1----!--- Tp1----
TIME t __.
Fig.20. Timing circuit logic
![Page 49: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/49.jpg)
I
I \
I
- 41 -
Obviously pressure would be applied for the longest period Tpl in
the first section and for the shortest period Tp4 in the fourth section
as illustrated in Fig.20. Such action would produce the desired move
ment of the blood from the leg towards the heart. The ratio of Tpl,
Tp2, Tp3 and Tp4 may be readily adjusted to suit any condition. On
the basis of one beat per second an interval of 100 milliseconds is
used between opening of two consecutive pressure valves. At the end
of the pressure application period Tp the pressure in. all four sections
is switched simultaneously to vacuum as shown in Fig.20.
Timing circuits have been built to control the period of vacuum/
pressure application of each of the four sections. However, in the
case where the pressure pulsating mechanism is not divided into
individual sections but is merely used as a single unit, only Tpl
timing circuit is then used.
Fig.21. illustrates the Timing circuit which consists of four sections,
each section comprising a monostable multivibrator (TTL chip SN
74121) driving a flip-flop (TTL chip SN 7473 N). The SN 7473 N
is a binary unit so only two are used for the whole system. The mono
stable multivibrator determines the length of Tp period. A variable
potentiometer is used as one of the external timing components, hence
the Tp period of each section is readily adjustable. The vacuum
application period Tv is complementary to Tp.
The pulse size control voltage from the Pulse Frequency Counter
is fed to each monostable multivibrator as the power supply to
external timing resistor thus controlling the change in time length
of each output Q (pin No.6.).
The inverse output Q of the Delay circuit is used to provide a common
trigger for the Timing circuit as shown in Fig.21. The monostable
multi vibrator ou~put will remain enabled for a period of time, the
![Page 50: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/50.jpg)
-\
•
' '-_L l>OL5E St%.£ CONTROL VOLT~CSE f {FROM "PUt.SE FRc~. COLINTE~) 'lJ·D (FROM DELAY CIRCUIT}
TO THE SAFETY _..... SHUT OFF CCT
t SAFETY SHUT-OFF
Fig. 21. Timing Circuit
- 42 -
MCT~
![Page 51: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/51.jpg)
I
' \ ___.
=4 I )
\ (' .
\.
t.
- 43 -
length of which is determined by the sizes of the external timing
components R and C. Each multivibrator drives a flip-flop TTL
unit ( SN 74 73) which, through an actuating circuit, keeps either
the pressure or vacuum valve open. Since a negative- step pulse
triggers the flip- flop, the inverse output Q of the first mono stable
is used. Thus at the end of delay period 'd' the inverse output -Q will immediately clock the first flip-flop; causing the first pres-
sure valve to be opened. The output Q of the second monostable
multivibrator is enabled for about quarter of the time of the first
one as shown in Fig.20. At the end of this time period the negative
step will clock the second flip-flop, i.e. open the second pressure
valve for a period Tp2. In this manner the opening of the second
valve is delayed by approximately 100 milliseconds. Using the
same technique, the third pressure valve is opened about 100 milli
seconds later and the fourth pressure valve another 100 milliseconds . later. The pressure application times Tp1 - Tp4, are illustrated
in Fig.20. At the end of the period Tp1 flip-flops are cleared and
reset by a negative- step pulse from output Q of the first mono stable
multivibrator. The common 'clear' signal ensures that the pres
sures in all four sections are switched to vacuum at the same time.
Should the ECG trigger occur anywhere in the Tp phase of the pre
vious cycle, as illustrated by the dotted line X - X, the output sig-
nal of the Safety Shut- Off circuit will clear all the flip- flops. These
units can be clocked into pressure again only in ~he new cardiac cycle
initiated by the early signal (X - X).
Each flip-flop has two outputs Q and Q (pin No.12. and 13. in the
case of the first flip-flop and No.9. and 8. of the second) which
are connected to light emitting diodes ( MCT2' s) of the suction and
pressure solenoid valve controls respectively in the Actuating circuit.
3. 3. 6. Actuating Circuit
The Actuating circuit acts as a shunting circuit between the low
![Page 52: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/52.jpg)
•
I ,_
- 45 -
voltage electronic control system on one hand and the high voltage
solenoid valve system on the other. Each solenoid valve has the
same actuating circuit as illustrated in Fig. 22. Altogether there
are eight such circuits, four for pres sure and four for suction lines.
The solenoid valve is connected across the 220 V supply through a
gate controlled, high peak current, full wave, silicon AC Triac
( 2N 5756). A shunting arm, consisting of a 47 ohms resistor and
0,1 microfarads, 630V capacitor joined in series, is connected across
the triac as a safety measure to prevent backlashing as the 8W solenoid
valve coil is being switched off and on.
The gate of the triac is connected to the flip-flop output through a
phototransistor- diode coupled pair MGT 2 unit. This diffused planar
GaAs diode, which is optically coupled to the n-p-n silicon planar
phototransistor, will, on receipt of the control signal from the flip
flop, emit light causing the phototransistor to conduct while the flip
flop output is at logical 1. At the same time the gate will trigger
the triac open and will remain open as long as the flip-flop output
is at logical 1. No danger of leakage from 220V to SV side exists
since the isolation resistance is in the order of 1011
ohms and the
voltage isolation from emitter to detector is 1500 volts.
The necessary lOV DC supply is provided for the L.E.D. unit and the
presence of 100 ohms resistance load on the gate is to ensure that
the gate voltage is within the prescribed limits of 0,2 to 2,2 volts.
3.3. 7. Power Supplies
The three power supplies required by the circuitry of the console,
i.e. 220V, lOV and SV, are tapped from a common point thus the
console requires only a single 220V AC, 50 Hz power source.
The lOV DC power supply makes use of an auto-transformer, as·
![Page 53: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/53.jpg)
~ 0'
O,S)lF
220 V AC --r c3, SOH~
12."
o"
..,
t;.tF C;s2.
Fig.23.
SV ZENER DIODE
tooo,...uF c~
SV DC Power Supply
Z..D.
GND
1 Amp
1 v sv
![Page 54: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/54.jpg)
}
J
- 47 -
illustrated in Fig.22., to keep the negative terminal tied to the ground.
Such arrangement is essential to enable the triac gate voltage to be
with respect to ground potential, otherwise the floating voltage will
keep the gate always open. On the low voltage side of the auto-trans.,.
former a 500 microfarads capacitor smooths the output to an accep
table ripple effect.
The 5V DC supply provides power to all console circuits at a stabilised
5V with the current sonsumption varying from 0,2 to b, 7 amps. It
utilises a full-wave silicon rectifier bridge ( SILEC 110B8 029N)
as illustrated in Fig.23. and a zener diode for a stable 5V output. To
protect the ammeter A, which has a full- scale deflection of 1 amp,
the circuit is provided with an 1 amp fuse. Both the 5V and lOV
supplies have a 0,5 microfarads capacitor connected across the high
voltage windings to reduce mains borne interference.
3.3.8. Controls
The entire console circuitry is housed in a cabinet, mounted on
wheels for ease of movement, as shown in Fig.24 and 25. A single
power cable ( 1), on the side of the Console cabinet is to be plugged
into a 220V AC 50 Hz power source. On the same side of the Console
there is a row of 5-pin sockets ( 2) each socket controlling a pair of
valves - one a pressure and the other a suction valve - connected to
a single stage. The console was designed to control a compartment~d
pressure pulsating mechanism having four stages only but provisions
have been made to enable the console to monitor a sequential pulsator
having up to 6 stages. Immediately above the row of solenoid valve
sockets, a pressure transducer socket ( 3) is provided, thus the power
cable, pressure transducer cable and the leads from 5-pin sockets to
solenoid valves are all on the same side of the cabinet, away from the
front panel.
All the controls and monitoring instruments are mounted on the front
![Page 55: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/55.jpg)
- 48 -
Fig. 24. Front end elevation/6£ the Console
" ..
![Page 56: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/56.jpg)
( ,.
\ \
I )
'
- 49 -
Fig. 25. Side view of the Console
![Page 57: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/57.jpg)
,, \
- 50 -
panel of the Console as shown in Fig.25. A double-trace, storage
· oscilloscope ( 4) in the top left section of the cabinet face acts as a
monitor and a visual aid. It can monitor the pressure of the pressure
pulsating mechanism, delay time 1 d', pulse size control voltage,
Timing circuit pressure application periods Tp' s and Test Signal
circuit output. Of these, delay period 1 d 1 and the pressure effected
inside the external counterpulsator are most important.
Next to the oscilloscope there is the ammeter ( 5) ana the voltmeter
( 6) dials of the SV power supply, with a circuit power switch ( 7)
positioned immediately below. With the system running, any change
in the SV power supply output voltage can be easily detected and the
change in the current consumption used to spot any malfunctioning
of the Timing circuit stages.
Below the oscilloscope and running across the width of the Front
panel, there is a row of solenoid valve switches. Altogether twelve
switches are provided, i.e. six pairs, each pair controlling a pressure (8)
. {q) 1 "d . 1 H b f 1 and a suctlon.-.so enol va ve. ence any num er o va ves or any
particular valve can be utilised at will. A selector type of switch is
used, to render the operation of a solenoid valve manual or automatic.
At the end of this row, an operational switch (10) is provided which
controls the power supply to all solenoid valves to enable simultan
eous initiation of the pulsating system. Each solenoid valve switch
is associated with an indicator light, hence there are altogether six
yellow ( 11) lights representing vacuum and six red ( 12) representing
pressure.
In the middle of the panel, there is situated a main power switch ( 13)
flanked by a 1 amp fuse ( 14) and an indicator light ( 15) . On the
other side of the power switch a removable cover ( 16) protects six
potentiometers and a Delay circuit external timing resistor. They
are all manually adjustable to provide the desired length of pressure
application periods Tp' s and the delay time 1 d' .
![Page 58: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/58.jpg)
' I J '
- 51 -
Should, at any stage of clinical trials, application of the external
pressure be required to be removed, an emergency switch ( 17) has
been provided immediately below the main power switch. It will
cut the power supply to all pressure valves but leave the vacuum valve
functional. Lastly, there are provided on the front panel of the Con
sole a number of test points so that the pres sure in the external
counterpulsator mechanism ( 18) , delay time ' d' ( 19) , pulse size
control voltage ( 20), Timing circuit pressure application periods Tp 1 s
( 21) and Test Signal circuit output ( 22) can be monitored. The
Timing circuit pressure application periods test point is also provided
with a selector switch thus any one of the stages can be selected.
All the electronic circuits are mounted on one side of board which
slides in and out of the cabinet; on the other power supply circuits
are mounted. The two sides of the board are shielded from each
other by a metallic plate which is grounded. Also, wherever the
high voltage installations are proximal to low voltage electronic
circuitry, maximum use of shielded cables has been made.
The electronic circuitry has been built on two printed circuit cards
for ease of removal, replacement, adjusting and checking. These
cards slide onto the board between guides. There is enough room
available on the lower shelf of the Console cabinet to carry spare
parts for the electronic circuitry, necessary tools and an instruction
manual for the oscilloscope. The cabinet can be wheeled about the
room with ease.
![Page 59: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/59.jpg)
\ \
- 52-
3.4. THE EXTERNAL COUNTERPULSATOR
Since the objective of the external heart assist device is to provide
pulsating pressure changes, the mechanical task involved designing,
building and testing of an external counterpulsator. The pressure
pulsation, applied externally to the body of a patient, is to effect the
vascular bed so that it responds to the resulting change in transmural
pressure. A negative pressure is to be applied during systole to
achieve the desired reduction in the systolic pressure and a positive
pressure during diastolic interval of the cardiac cycle to increase
myocardial perfusion.
At first a suit, similar to that of an aviator in a "G11 suit or that of
a 11hard hat 11 diver, was considered as means for enclosing the
vascular bed, i.e. the entire human body. However, due to poor
accessibility of the patient enclosed in such a suit, increased
sweating, obstruction of normal patient care requirements, negative
psychological effort on the patient of being enclosed in a suit and
ability of the patient tohear or be heard being substantially impaired
by the suit and head piece, such idea was abandoned for a more rea
listic lower body enclosure. Because of its easy accessibility and
the fact that below the abdomen line there is about one third of the
entire blood volume in the circulatory system, the lower body pro
vides the most convenient locus of the vascular system to effect the
pressure changes. Also utilisation of the legs leaves the upper body
free for medical examination.
For initial clinical trials only one leg is to be utilised and, once the
desired result is achieved, the total area of the affected vascular bed
is to be increased. Hence a design of a leg unit was embarked upon
with provisions that later both legs and a large portion of the lower
posterior would be included. •
![Page 60: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/60.jpg)
- 53 -
3.4.1. Pressure Pulsating Medium
The initial idea was to have a double-walled trouser-leg unit fitted
over the patient's leg and pressurised with air. Since the shape
and length of a leg varies among individuals to a great extent, the
system was designed for a particular leg with means provided to
accept smaller and larger legs within a limit.
The circumference of the upper end of the typical leg is 57cm and
22cm just above the ankle. The length between the two circum
ferences is 7lcm hence the total affected leg area is approximately
2800 sq em. If a cushion of air, contained in the trouser leg unit
was 0, 7 em thick, the total volume of the pres sur is e d air would be
about 2 litres. With two litre s of air needed for each leg unit of the
eventual system, at fast heart rates of 3 beats per second, the air
compressor required would have to supply 12litres of pressurised
air per second, i.e. 720 litres a minute. Likewise a suction pump
would have to remove about the same amount of air.
Because of such an enormous volume of air which is to be moved in
a relatively short time, the need would arise for a large compressor
and an equally large vacuum pump. Also, air tends to dampen abrupt
pressure changes (from 200 mm Hg positive pressure to 200 mm Hg
suction pressure and vice versa) towards which this project is
directed. To overcome these difficulties water is to be used, pri
marily because of its incompressibility. As shown in the Appendix
C, water, compared with air, propagates a pressure wave nearly
five times faster: Cx (air} = 335,28 m/ sec; Cx (water) = 1432,56
m/ sec. The possibility of using hydraulic brake fluid was also con
sidered due to its relatively low compressibility, but was discarded
because of its high costs, inflammability, corrosive effect on trouser
material and noxious effect on the human skin.
![Page 61: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/61.jpg)
J
t
) !
)
- 54-
Using water as the working medium and a pulsing mechanism
based on the concept of a double walled trouser leg unit, two
mechanisms were built and tested in the laboratory, namely a
sequenced pulsator comprising a number of inflatable cuffs and a
:rp.inimum flow, rigidly walled system.
3.4.2. Sequenced Pulsator
An air-tight container, made of clear perspex and able to receive
a patient' s leg, comprises the outer shell and a trouser-type leg
unit, which is wrapped around the leg, the inner shell of ·the sequenced
pulsator. The latter is a double walled sheath, tailored to encase a
conically shaped leg. Its outer wall is made of very thin, flexible
but non- stretchable nylon- reinforced PVC sheet, whereas the inner
wall is made of a soft wool fabric which does not cause irritation to
the skin, but provides a warm cover for the leg. The vacuum or
suction is effected by the external shell and the positive pressure by
the inner shell so that the mean applied pressure is nearly equal the
atmospheric.
As illustrated in Fig. 26.a., the leg sheath, which fits the typical
leg from the ankle to the thigh, is provided with laterally positioned
pockets. Each pocket accommodates a rubber cuff. To avoid
undesirable discontinuity in the area of pressure application, the
pockets have been made in such a way as to have the cuffs, positioned
therein, overlapping, as shown in Fig.26.b. Made of 3mm black
rubber, each cuff has a centrally located spout, acting as both the
inlet and the outlet.
By using a sealing type of material, known as 'Vulcro' , a simple
method.of fastening the sheath about the leg is. employed. When the
sheath is wound about the leg, with two opposite edges overlapping,
the male strapping, provided on one edge, is laid over female strips,
provided on the other edge, causing a strong interlocking. The
fastening, which prevents the edges from sliding away from each
![Page 62: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/62.jpg)
. J
nylon reinforced PVC sheet
- 55 -
~A 1 spout
aperture
pockets
.:
female vulcro
straps
Fig. 26. a.) Plan view of the Leg Sheath
PVC sheath
male vulcro straps
wool
spout
male vulcro straps
,;pig. 26. b.l Cross"-"'secttonal vtew A""A of the Leg
Sheath with Rubber Cuffs placed in
Pockets
![Page 63: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/63.jpg)
- 56 -
·other when the rubber cuffs are pressurised, is undone by merely
peeling off the male straps.
The leg, with the compartmented sheath fastened about it, is positioned
into the rectangular per spex box, as depicted in Fig.27. The box,
which is air-tight, is shown with its lid open, The short, upright
wall, through which the leg fits, is removable for ease of placing the
leg inside the box. This wall is provided with a rubber seal in the
form of a bell stretching from the leg diameter, where it is sealed,
to the circular opening in the wall, to which it is secured.
The rubber cuff spouts are easily connected to the pressure/vacuum
line by plugging them into quick- couple connector sockets mounted
on the inside of the long side wall. Each socket is connected through
the wall to a free arm of a T-piece, mounted on the outside as illu
strated in Fig.27. On the remaining two arms of the T-piece there
are two solenoid valves, one controlling the passage to the common
pressure supply manifold and the other to vacuum. From a hydraulic
pressure- suction supply liquid under pressure and vacuum is supplied
to the two manifolds mounted on the outside of the per spex box.
The console, as described in previous sub- section, activates the
pressure solenoid valves in a sequential manner, switches them off
together and, at the same time, switches on all the vacuum valves.
During the pressurisation phase of the pulsating cycle, the working
fluid distends the leg sheath, causing the inner wall to press against
the leg. The pressure valve controlling the passage to the widest
cuff, i.e. the one which fits around the leg immediately above the ·
ankle, is open the longest- typically 480 milliseconds - and the one
controlling the narrowest cuff, the shortest- 160 milliseconds. In
this way a propagating squeezing effect is achieved from the ankle
upwards towards the thigh. On depressurisation, when the vacuum
valves are opened, a certain amount of liquid is withdrawn from the
pressurised cuffs releasing the tightening grip about the leg. The
leg is then exposed to the surrounding pressure i.e. the negative
![Page 64: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/64.jpg)
- 57 -
rubber bell seal
lid in open position
compartmented sheath
removable end wall
pair of solenoid valves per each
\ rubber cuff
\ ....... ---.n l==;:;:::::t>l¢::::1 \ \ _______ l -----1 I \ I
pressure manifold
I I .._ ____ ...., l=ri=:f><l==l -·(------;--
/ I ( ' \ \
-\---+-\ ' \ I \ I
0
to vacuum pump
quick couple connec;:or
•
vacuum manifold
drain from pressure
supply
to vacuum supply
~ig. 27. Plan vtew ot the Sequenced Pulsator
![Page 65: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/65.jpg)
( \. \
- 58 -
'pressure which is continuously maintained in the perspex box.
Thus the two shells, working together in phase, effect the counter
pulsation.
3.4.3. Minimum Flow, Rigidly Walled Pulsator
To minimize the amount of flow of the working fluid into and out of
the pulsator a system with rigid walls was designed1 Instead of
having two separate shells - an outer and an inner one, the minimum
flow pulsator, as built and tested in the laboratory, has a combination
of the two. It also retains water as the working medium.
As illustrated in Fig. 28., the minimum flow pulsator comprises a
rigidly walled unit of cylindrical shape, closed at one end and with
an aperture on the other to receive a leg. The aperture is provided
with a specially constructed seal which consists of a stocking, made
of imperm·eable neoprene material 0,2 mm thick and adapted to sheathe
the leg, a rubber cover and a vacuum seal wrap. The space between
the stockinged leg and cylinder walls is filled with working fluid. The
purpose of the stocking is two fold: to keep the leg dry and to pre
vent spillage of the working fluid through the small circumferential'
gap between the leg and the flange. In the case of the typical leg the
aperture is of such a diameter that it leaves a gap of approximately
4 mm at the apex.
To prevent the stocking from bulging out through the small gap in the
aperture, during the pressurisation phase, a thick rubber cover,
which fits between the leg and the stocking is provided. Likewise,
to prevent the thick rubber cover and the stocking from being pulled
into the cylinder and away from the leg when suction within the cylin
der is effected, the same negative pressure is created in the small.
circumferential gap in the aperture and just outside it. As shown
in Fig. 28. this is achieved by employing a vacuum seal wrap around
![Page 66: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/66.jpg)
N 0 T
pressure transducer testing nipple
\_____._
suction outlet
~ ·- ~·· , . ..., ..
T 0 S C A L E
leg stocking air outlet
outlet valve
C> 0 C
. . ·- . ~ ·- ---.---;-; '"
rubber cover
c:r-o -,,
J I
flanges
Fig. 28. Cross-sectional elevation of the minimum flow, rigidly
walled pulsator with leg fitted therein
vacuum seal wrap
gap U'l. \0
![Page 67: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/67.jpg)
. 'J
- 60 -
the protruding thigh and by connecting to the vacuum supply. When
the leg is fitted into the stocking and the cylinder is filled up with
water, the leg stocking collapses tightly about the leg. To withdraw
any air which may be trapped, a perforated tygon tube is positioned
alongside the leg as shown. To facilitate air removal the cylinder is
tilted so that the unit is filled from the closed end up towards the
aperture.
The closed end of the cylinder is made of a double walled PVC disc
which is provided with an inlet and an outlet nozzle for the working fluid,
both of which are 19 mm in diameter and each of which is connected
through a solenoid valve to the hydraulic pressure and suction supply
tanks. Just above the nozzles there is a testing nipple onto which the
pressure transducer nozzle connects. Furthermore, two air outlet
valves have been fitted on the top ridge of the cylinder to allow entry,
or escape of air during filling or draining. The narrow air outlet
near the seal is connected to the vacuum system to allow bleeding out air.
From the above it is clear that the system can function only in a non
sequential i.e. single-pulse fashion. However, pressurisation and·
depressurisation inside the vessel will be damped to a degree by the
movement of the leg in and out ofthe cylinder. The force which causes
the outward movement is equivalent to the pressure effected on the leg
multiplied by the leg skin area inside the cylinder. Likewise, the in
ward movement is due to the force equivalent to the product of suction
pressure and the affected leg area. There is, however, a resistance
to these movements due to a number of factors, the largest one resul.;.
ting from the fact that the patient' s seat is fixed in relation to the
cylinder.
3.5. HYDRAULIC PRESSURE - SUCTION SUPPLY
The pressure - vacuum generating system, as shown in Fig. 29 and
Fig. 30., comprises two upright tanks, a pressure and a vacuum
![Page 68: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/68.jpg)
i
t ~
)
i. I l
j. I
1
:.
- 61 -
tank, partially filled with water and interconnected through a
pump. Pressure and vacuum are caused by pumping the water
from one and into the other tank. A return from the pressure tank to
the pump inlet is channelled through a control valve of much smaller
orifice than the pipe diameter used in the system. The re suiting
pressure drop across this valve (due to the constricted flow) is
used to control a pressure differential of desired magnitude between
the pressure and vacuum tanks. Both tanks are provided with a sight
glass, pressure or vacuum gauge and an air stop- cock valve at the
top end for manual control of the air pressure (positive or negative)
within the tank. The pipe, leading from the vacuum tank to the pump
inlet is provided with two valves, one to isolate vacuum tank from
the pump and the other, fitted on a side inlet tube, to fill the system .
with working medium. Likewise the U- bend on the pump discharge
side is fitted with two valves, one on a side discharge line to drain
the system and the other to isolate the pressure tank from the pump.
A schematic diagram for the system is shown in Fig.31.
A pressure relief valve is also provided between the pressure and
the vacuum tank. For this project a pressure ceiling of 500 mm Hg
has been chosen so the relief valve has been set at that value.
The pressure control valve is operated manually and is of such
type and size that the pressure in the pressure supply tank can be
set at any value between atmospheric ( 0 mm Hg gauge) and the
ceiling pressure ( 500 mm tlg gauge) . When the system is running
the two tanks are one third full, the air above the working fluid being
at 200 mm Hg pressure in the pressure tank and 200 mm Hg suction
in the vacuum tank. The ratios of air /water are calculated as shown
in Appendix D and are based on the desired result that the volume
of air in the tanks would be such that a withdrawal or addition of
one litre of working medium to a tank would alter the pressure of
air above it by a negligible amount. Calculations for the volume
of water flow, i.e. one litre per pulse, are also given in Appendix D.
![Page 69: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/69.jpg)
- 62 -
:>t ..-1 P-t P-t ::s (/)
s:: 0
•r-l +l u ::s (/)
I (!) !-1 ::s Ul Ul (!)
H ~
u ·r-l .--{
g ~ >-t . n:: (!)
/C. +l
!H 0
~ (!)
"r-l :> .jJ
s:: 0 H ~
. o-. N
![Page 70: p~~~~ fcs~ y,](https://reader030.vdocument.in/reader030/viewer/2022011901/61d6291cca64475d2145a0fc/html5/thumbnails/70.jpg)
"-
Fig. 30. -Plan view -of the Hydr a ulic Pressure-Suction Supply
CT\ w