P16081: SYSTEMIC CIRCULATION MODELSUBSYSTEM DESIGN REVIEW
John RayFabian PerezRobert Kelley
Mallory LennonJacob Zaremski
![Page 2: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/2.jpg)
Our goals for this review Updates from Phase II Review (10 minutes)
CAD Schematic Analysis Subsystem Analysis (25 minutes)
Risk Analysis (5 minutes) Bill of Materials (BoM) (5 minutes) Project Plans (5 minutes)
Agenda
![Page 3: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/3.jpg)
Goals
1. Introduce progress in engineering analysis
2. Address budget concerns
3. Demonstrate efforts to design for modification
4. Receive feedback
5. Create action plan
![Page 4: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/4.jpg)
Finalized Objective
Our goal is to deliver a functioning physical model of systemic circulation which, when
used in conjunction with P16080’s heart pump, will be used as a teaching tool, allowing
students to validate mathematical models of the circulatory system from Chapter 5 of
“Quantitative Human Physiology” An Introduction by Joseph Feher. The model will
ultimately enhance students' understanding of the circulatory system by enabling
them to analyze the circulatory system under normal, exercise, and pathological conditions
through the measurement of pressure and flow.
![Page 5: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/5.jpg)
Updated Use Case
![Page 6: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/6.jpg)
ER Metrics of Quality (1 of 3)
![Page 7: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/7.jpg)
ER Metrics of Quality (2 of 3)
![Page 8: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/8.jpg)
ER Metrics of Quality (3 of 3)
![Page 9: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/9.jpg)
ER mapping to F.D. (1 of 2)
![Page 10: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/10.jpg)
ER mapping to F.D. (2 of 2)
![Page 11: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/11.jpg)
Functional Decomposition
![Page 12: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/12.jpg)
Morph Chart (1 of 3)
![Page 13: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/13.jpg)
Morph Chart (2 of 3)
![Page 14: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/14.jpg)
Morph Chart (3 of 3)
![Page 15: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/15.jpg)
Undecided Concepts
Pressure• PASCO versus Honeywell
• LabVIEW versus DataStudio
Resistance• External Clamp vs. Valve
![Page 16: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/16.jpg)
Updated System Architecture
![Page 17: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/17.jpg)
CAD Schematic
![Page 18: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/18.jpg)
Main Components
Barb tube fitting: cheapest, easiest
connection
Pressure release value: easy release valve
Drill and screw into acrylic
Ball pump: cheapest,
easiest way to add air pressure
Pressure tap into tubing
![Page 19: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/19.jpg)
Pressure, PA
Pressure, PC Pressure, PV v
P16081 Pump
Arterial Compliance,
CA
VenousCompliance,
CV
Flow
Resistance
1
2
345
LabVIEW
6
7 8 9
![Page 20: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/20.jpg)
Subsystems Agenda
1 & 5 - Pressure
2 & 4 - Compliance
3 - Resistance
6 - Labview
7, 8 & 9 - Consult with P16080
![Page 21: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/21.jpg)
Pressure Sensor Analysis1. Flow Diagram
2. PASCO Sensors
3. Honeywell Sensors
4. Bill of Materials
5. Test Plan
6. Risks
![Page 22: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/22.jpg)
Pressure Flow Diagram
Analog Pressure Signal
Digital Pressure Signal
DAQ
Amplifier Board
ComputerLabView Program
AC Power
Energy
Information
Pressure vs. Time Graph
![Page 23: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/23.jpg)
PASCO Sensors
Will need more than one computer for real time system analysis
• LabView for heart, DataStudio for circulatory pressures
• LabView needed for automatic resistance control
Pros Cons
Already Owned No integration with LabView
Differential Pressure Capabilities
Special Pressure Taps Needed
User Friendly
http://www.pasco.com/file_downloads/product_manuals/PASPORT-Dual-Pressure-Sensor-Manual-PS-2181.pdf
Maximum Sampling Rate 1000 Hz
Absolute Pressure Range 0-1500 mmHg
Differential Pressure Range -750-750 mmHg
Resolution 0.075 mmHg at 10 Hz
Repeatability 7.5 mmHg
Tubing (Type) Polyurethane
Tubing Size (Diameter) 3.2 mm
Tubing Length 2.4 m
InterfaceData Studio (New one coming soon?)
Source:PasPort Instruction Sheet (012-09969A)
![Page 24: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/24.jpg)
Honeywell Sensors
Only one computer program needed for both heart and circulatory
• DAQ is already provided, but would need to build the circuit board/LabView Program
http://www.mouser.com/ProductDetail/Honeywell/HSCMRNT005PGAA5/?qs=%2fha2pyFaduhkciXVz6btFHLY3u79xkDhknp39AuPvmffYIGgrGx0aQ%3d%3d
Pros Cons
Integration With LabView
More Expensive than PASCO
Accuracy Will need a Sensor Board/DAQ
Liquid Friendly
TruStability Board Mount Pressure Sensors: HSC Series (HSCMRNT005PGAA5)
Operating Gage Pressure Range 0-258 mmHg
Output Type Analog
Pressure Type Gauge
Operating Supply Voltage 5 V
Operating Temperature -40-85 C
Operating Supply Current 20 mA
Accuracy 0.25%
Liquid Media Capable? Yes
Source:
HSCMRNT005PGAA5 Datasheet (Mouser.com)
![Page 25: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/25.jpg)
Honeywell SensorsDimensions
http://www.mouser.com/ProductDetail/Honeywell/HSCMRNT005PGAA5/?qs=%2fha2pyFaduhkciXVz6btFHLY3u79xkDhknp39AuPvmffYIGgrGx0aQ%3d%3d
![Page 26: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/26.jpg)
Pressure BOMComponent
IDComponen
tSupplier Supplier ID Quanitity/Dimensions Price/Unit Total
CostNotes
P1 Honeywell Board Mount Pressure Sensor
Mouser Electronic
HSCMRNT005PGAA5 2 $45.78 $91.56 May only need 1 (48.85)
P2 PASCO Sensor
PASCO PS-2181 2 Free - Already Owned
P3 Pressure Taps
PASCO ME-2224 6 $16.00 $96.00 Comes as a set of 6
Might be able to borrow for free (Rep mentioned it)
![Page 27: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/27.jpg)
Pressure Risks
5 Technical Not being able to generate required values
6 Technical Seal on the pressure tap leaks
7 Technical Not being able to calibrate within time constraints
9 Resource System components will be expensive
11 SafetyElectricity and water combination can cause
dangerous conditions
Owner: Jack
![Page 28: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/28.jpg)
PRESSURE TEST PLAN1.Prove Sensor Functionality
a. Flow through a pipe with decreased diameter in the center
b. Pressure taps at two points
c. Measure the pressure drop
2.Calibration of the Sensors
a. Obtain a full tank of known pressure
b. Measure the pressure and correct accordingly
![Page 29: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/29.jpg)
Subsystem - Compliance Tank1. Flow Diagram
a. Energy, Material, and Information I/O
b. Interfaces
2. Cylinder or rectangular prism?
a. Pros and cons
b. Feasibility
3. Bill of Materials (Draft)
4. Subsystem Risks
5. Preliminary Ideas for Testing Plans
![Page 30: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/30.jpg)
Pressure
Flow
AC Power
Air Height Compliance
Pressure TransducerLabVIEW
Pressure vs. Time
Flow
Pressure
Forced Air
Energy
Material
Information
![Page 31: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/31.jpg)
Cylindrical Tanks
ProsFewer pieces
Less interfaces to seal
ConsPrice limits diameter
Interfacing with tubing and pressure taps
![Page 32: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/32.jpg)
Rectangular Tanks
ProsFlat surfaces easier to machine and
interface
ConsMore pieces that need to be machined
and sealed
![Page 33: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/33.jpg)
Cylindrical Arterial Compliance Tank
2 3 4 5 6 7 8 9 10 11 120
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5 Polycarbonate 5.75ID [in]
Acrylic 5ID [in]
Height (inches)
Co
mp
lia
nc
e (
mL
/mm
Hg
)
![Page 34: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/34.jpg)
Cylindrical Venous Compliance Tank
0 5 10 15 20 250
5
10
15
20
25
30
35Acrylic 9.75ID [in]
Polycarbonate 5.75ID [in]
Height (inches)
Co
mp
lia
nc
e (
mL
/mm
Hg
)
![Page 35: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/35.jpg)
Rectangular Arterial Compliance Tank
2 4 6 8 10 12 14 160
1
2
3
4
5
6
7
8
9
4x4x12 [in]
4x8x16 [in] (Donovan)
Height (Inches)
Co
mp
lia
nc
e (
mL
/mm
Hg
)
![Page 36: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/36.jpg)
Rectangular Venous Compliance Tank
2 4 6 8 10 12 14 160
5
10
15
20
25
30
35
40
12x12x16 [in]12x8x16 [in] (Donovan)
Height (inches)
Co
mp
lian
ce
(m
L/m
mH
g)
![Page 37: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/37.jpg)
Technical Risks - Compliance Tank
![Page 38: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/38.jpg)
Resource & Safety Risks - Compliance Tank
![Page 39: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/39.jpg)
Compliance BOM - Cylindrical
![Page 40: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/40.jpg)
Compliance BOM - Rectangular
• 6-sided tank would be around $200
![Page 41: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/41.jpg)
Compliance - Preliminary Testing Plans
1.Air tight seal
a. pressurize, use soap, and look for bubbles
2.Generate chart to allow students to know which height
corresponds to a desired condition - scale on box
3.Excel spreadsheet for liquid height and corresponding
compliance values
![Page 42: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/42.jpg)
Resistance Agenda1. Flow Diagram
2. Valve and Resistance Analysis
3. Bill of Material and Alternative Bill of Materials
4. Risks
5. Test Plans
![Page 43: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/43.jpg)
Resistance Flow
![Page 44: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/44.jpg)
Valve and Resistance AnalysisResistance can be modeled after Ohm’s Law in that: R=ΔP/F where ΔP is
the height difference and F is the mean flow rate.
[http://circ.ahajournals.org/content/89/2/893.full.pdf]
![Page 45: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/45.jpg)
Valve and Resistance ContinuedFor Valves resistance and friction can be modeled as:
hf=(Kv^2)/2gWhere K is the resistance coefficient, f is the Darcy
friction factor, V is the velocity and hf is the Frictional Loss or Head Loss.
Considering using a Gate valve for purposes of the design.● Resistance will have to be tested manually
○ Test each resistor position and correlate them to the pressure output seen
■ Create a graph of resistor position vs pressure output
![Page 46: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/46.jpg)
Valve ConsiderationsOther potentials: Linear Actuator and Globe Valve
![Page 47: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/47.jpg)
Resistance BOM
Subsystem
Component ID Component Supplier
Supplier ID
Quanitity / Dimensions Price/Unit Total Cost Notes
Valve V1
Bronze Gate Valve-Class 125, 3/4" NPT Female, Non-rising Stem
McMaster Carr 4619K14 1 $27.96 $27.96
May be alternated for linear actuator
SubsystemComponent
ID Component SupplierSupplier
IDQuanitity /
Dimensions Price/Unit Total Cost Notes
Linear Actuator LA1
25mm Diamter Actuator Exteded strew with motor
Anaheim Automation
TSFCA25-150-21-023-LW4 11 $39.00 $39.00
Can only apply 10 Newtons of Force.
Globe Valve GV
Low-Pressure Bronze Globe Valve, 3/4" NPT Female, EPDM Disc
McMaster Carr 4695K65 1 $37.44 $37.44
Meant for low pressure flows, overall length of 2 5/16”
Current BOM
Alternative BOM
![Page 48: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/48.jpg)
Resistance Risks
![Page 49: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/49.jpg)
Resistance - Preliminary Testing Plans1.Calibration
a. Make sure when impedance is 0 R=ΔP/F is obeyed.
b. Make sure when impedance is at max there is no flow through the
system after this point.
2.Calculate theoretical head loss through circuit and perform
head loss experiment on the pipe and valve to confirm
compliance to the theoretical model (initial set up for
MSD)
![Page 50: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/50.jpg)
LabVIEW
![Page 51: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/51.jpg)
LabVIEW Considerations
• Waveform consistency
• Same parameters give
same waveforms each
time
• Interfacing with pressure
sensors
• One program for both
teams
F. M. Donovan (1975) Design of a Hydraulic Analog of the Circulatory System for Evaluating Artificial Hearts, Biomaterials, Medical Devices, and Artificial Organs, 3:4, 439-449
Arterial
Venous
![Page 52: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/52.jpg)
Preliminary Life Span Calculation
Considerations:
What will likely fail first?
Is that part expensive?
Is that part easy to replace?
![Page 53: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/53.jpg)
New risks
![Page 54: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/54.jpg)
Mitigated Risks
![Page 55: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/55.jpg)
Risk Chart
![Page 56: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/56.jpg)
Draft System Bill of Materials (1 of 2)
![Page 57: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/57.jpg)
Draft System Bill of Materials (2 of 2)
![Page 58: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/58.jpg)
Project Plan - what we achieved
![Page 59: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/59.jpg)
Project Plan - deliverables for next phase
![Page 60: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/60.jpg)
End of MSD I Deliverables1. “Working” theoretical model
2. Finalized, completed and accurately priced Bill of Materials
3. CAD drawing 100% done
4. Test plan for design 90% complete
5. Theoretical risk list complete with ideas as to how to minimize
potential effects
6. Understanding of deliverables for MSD II
7. Short list of contests this design could enter
![Page 61: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/61.jpg)
Ways to Improve Efficiency
1. More organized direction for research of parts and
materials
2. Better collaboration with P16080
3. More organized group meetings
![Page 62: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/62.jpg)
Goals for Phase IV
1. Choose the most efficient pressure sensor
2. Decide on dimensions for compliance tanks
3. Decide on internal versus external resistances
4. Interactions with P16080
• LabVIEW
• Flow meter
• Interfacing
![Page 63: P16081: SYSTEMIC CIRCULATION MODEL SUBSYSTEM DESIGN REVIEW John Ray Fabian Perez Robert Kelley Mallory Lennon Jacob Zaremski](https://reader036.vdocument.in/reader036/viewer/2022062309/5697bfd61a28abf838cadb90/html5/thumbnails/63.jpg)
ASEE 123rd Annual Conference & Exposition
• Abstract submitted on October 20th, 2015
• Abstract decision deadline: November 9, 2015