pacemaker and defibrillator troubleshooting - · pdf filepacemaker and defibrillator...
TRANSCRIPT
Pacemaker and Defibrillator
Troubleshooting
Jeffrey L. Williams, MD, MS, FACC, FHRS
Director, Invasive Cardiac Electrophysiology
Heart Rhythm Center
Lebanon Cardiology Associates
The Good Samaritan Hospital
Disclosures
• None
Objectives
• Be able to understand pacemaker and
defibrillator normal function and
troubleshoot common abnormal
conditions.
Pacemaker and ICD Topics for Boards
• Passing Score: ~65—70% correct is a pass, very few get
>90% correct. 80-85% first time pass rate
• Topics on EP Boards:
– Basic EP and Pharmacology: 5%
– Noninvasive Diagnosis and Treatment: 15%
– Invasive Diagnosis and Treatment: 42%
– Devices: 18%
– Clinical Syndromes: 7%
• Topics on General Cardiology Boards
– 13% Arrhythmias with up to 35% of these device related.
• Content not proportional to prevalence
Today’s Outline
• Pacing Functions and Timing Cycles
• Failure to Pace
• Loss of Capture
• Oversensing
• Undersensing
• Defibrillation Threshold Testing
• Arrhythmia Classification
• Sample Board Questions
Basics of Device Troubleshooting
• Use all the available information.
• Interrogate devices during your training.
• Look at the CXR’s.
Surface
EGM
Ventricular
EGM
Atrial
EGM
Markers
Pacemaker Function Codes ICHD Code
Position
Category
Letters Used
Manufac- turer’s Designation Only
I II III IV V
Chamber(s) Paced
Chamber(s) Sensed
Mode of Response(s)
Programmable Functions
Special Tachy- arrhythmia Functions
V- Ventricle A- Atrium D- Double O- None
V- Ventricle A- Atrium D- Double
T- Triggered I Inhibited D- Double O- None R- Reverse
P- Programmable rate and/or output
M- Multi- Programmable
C- Communicating O- None
B- Burst N- Normal Rate
Competition
S- Scanning E- External
S- Single Chamber
S- Single Chamber
, Comma Optional Here * Atrial Triggered & Ventricular Inhibited
Pacing Mode Selection
• AV block should have ventricular pacing support
• Sinus Node Dysfunction – atria should be supported with rate response.
• Neurocardiogenic syncope should have ―rate-drop‖ response activated.
• Atrial tachyarrhythmias need mode switching (atrial pacing paradigms likely not on boards).
Transitions of DDD Timing
Event Response
AV Interval Times Out Pace V, Begin VAI
VA Interval Times Out Pace A, Begin AVI
V sense during AVI Begin VA, No Pacing
V sense during VAI Restart VA, No Pacing
A sense during VAI Begin AVI, No Pacing
Timing Cycles
• Refractory Period: Sensed events are ignored for timing purposes.
• Blanking Period: Sensing amplifier is off and no sensing occurs.
• Maximum Tracking Rate in DDD
– TARP = AVI + PVARP
– Total Atrial Refractory Period (TARP), AV Interval (AVI), Post-Ventricular Atrial Refractory Period.
Pacemaker Timing Cycles • DDD, max track 100 ppm, AV delay 150
ms, PVARP 250 ms
Board Question #1
• Which of the following defines the
maximum tracking rate in DDD mode?
1. PVARP
2. VRP + AVI
3. TARP
4. AVI + Blanking Period
5. TARP - VRP
Troubleshooting
• Additional leads
• Magnet
• Manipulation of
pacemaker, pocket,
and leads.
• Isometric maneuvers
• CXR
• Exercise testing
– We had patient with
changing morphology
during exertion.
• Device reps and field
engineers.
Cheat Sheet
Voltage
Threshold
Current
Threshold
Lead Impedance
Wire Fracture High High/Nml/Low High
Insulation Break High/Nml/Low High Low
Lead
Dislodgement
High High Nml
Exit Block High High Nml
Pacemaker Syndrome:
Clinical Presentation
Pacemaker Syndrome
• Symptoms: malaise, weakness, cannon A
waves, CP, cough, confusion, syncope.
• Symptoms due to loss of AV synchrony.
• Most common with VVI or VVIR.
• May occur with any pacing mode if AV
synchrony lost.
Board Question 2
Patient complains of dizziness when running to his car the night prior to your
office visit. The rhythm strip is obtained during an treadmill exercise test.
The device parameters are: DDD, mode-switch detection rate 175bpm,
upper rate limit 150bpm, sensed AV interval 200msec, PVARP 400msec.
This EKG demonstrates:
A. Mode-switching
B. Rate-drop response
C. Rate-smoothing
D. Pacemaker-induced 2:1 block
E. None of the above
- Taken from Nacarelli et al, ACC EPSAPII.
Board Question 2 (Cont’d)
• The max tracking rate is 100bpm defined by TARP (AVI+PVARP=600ms).
• When spontaneous P wave rate exceeds max tracking rate of 100,
alternative P waves fall within PVARP, and are not responded to, creating a
2:1 AV block.
• Mode-switching not present since atrial rate does not exceed 175bpm; in
addition, PV intervals are the same during slower V pacing, suggesting VVI
pacing not present.
• Rate-drop response is not present, since this feature requires an abrupt
drop in spontaneous P wave rate for activation.
• Rate smoothing not present, since this strip shows an abrupt reduction in
paced ventricular rate rather than a smoothed one.. - Taken from Nacarelli et al, ACC EPSAPII.
Common Pacemaker Mediated Tachycardia
• Surface, atrial, and
ventricular EGMs
shown.
• The intracardiac
markers indicate that
the retrograde P
waves occur 280
milliseconds after the
ventricular-paced
beats.
• Occurs at
programmed upper
rate limit.
• Treatment is to
prolong PVARP.
Surface EGM
Ventricular EGM
Atrial EGM
Uncommon Endless Loop Tachycardia
• Repetitive Non-Reentrant Ventriculo-Atrial Synchrony
• Form of endless loop tachycardia (functional atrial non-capture with retrograde ―P‖ occurring in
PVARP
• Unlike the pacemaker mediated endless loop tachycardia, RNRVAS does not occur at the
programmed upper rate interval.
• The retrograde activation of the atria occurs within the PVARP so that it is not sensed by the
atrial circuit of the pacemaker (functional undersensing). Though it is not sensed by the atrial
circuit, it makes the atria refractory to the subsequent atrial paced output (functional loss of
capture). The ventricular output is delivered by the device at the end of the programmed AV
delay. {Johnson Francis, Indian Pacing Electrophysiol J. 2010; 10(5): 203–204.}
Automatic Mode Switching (AMS)
• AMS turns off atrial tracking in the presence of intrinsic atrial activity above a programmable atrial rate cutoff
• This stops rapid ventricular pacing in response to high-rate atrial activity
• AMS can cause a sudden rate decrease as atrial tracking in DDD at 140 ppm mode switches and becomes DDI at 70 ppm—in one cycle – Programmable AMS base rate allows an interim mode-switch rate to go
into effect—higher than the base rate
– Eliminates uncomfortable abrupt rate transitions
• Set atrial rate cutoff at which automatic mode switching (AMS) goes into effect – Program it to a value at least 20 bpm above the Maximum Tracking
Rate and/or Maximum Sensor Rate
• Set atrial rate cutoff high enough to accommodate: – Patient’s lifestyle and activity level
– Presence of sinus tachycardia
Automatic Mode Switching
- Taken from St. Jude Medical
Hysteresis
• Programming of hysteresis permits prolongation of the first pacemaker escape interval after a sensed event.
• A pacemaker programmed at a cycle length of 1000 milliseconds (60 bpm) and a hysteresis of 1200 milliseconds (50bpm) allows 200 milliseconds more for another sensed QRS complex.
• If another QRS complex is not recognized, then the pacemaker continuously stimulates the heart at the programmed rate of 60bpm, an escape interval of 1000 milliseconds, until a sensed event restarts the cycle.
• The advantage of hysteresis in a single chamber pacing mode is the ability to maintain spontaneous AV synchrony as long as possible
Hysteresis
Cardiac Pacing and ICDs, 5th Edition, Kenneth A. Ellenbogen, MD, Mark A. Wood, MD
p. 291.
Board Question 3
A 68 yo woman had a VVI pacing system placed during aortic valve replacement. The
tracing shows simultaneously recorded leads I and II. The tracing most clearly
demonstrates which one of the following?
1.There is failure to sense.
2.The mode of function is VVT
3.Ventricular lead dislodgement.
4.Normal VVI function
5.The patient has a right IVCD
- Taken from Nacarelli et al, ACC EPSAPII.
Board Question 3 (Cont’d)
• Rhythm is atrial flutter. Normal VVI function.
• Spontaneous QRS complexes are LBBB.
• 5th Beat: Fusion.
• 7th Beat: Pure paced from epicardially placed (during valve
surgery) LV lead with RBBB configuration.
• VVT with intermittent undersensing cannot be ruled out
without device interrogation but is unlikely mode.
- Taken from Nacarelli et al, ACC EPSAPII.
Failure To Pace (No Output)
• Battery Failure
• Circuit Failure
• Lead Fracture
• Internal Insulation Failure
• Oversensing
• Loose Set Screw
• Crosstalk
TENS Unit
- Testing revealed no inhibition.
Crosstalk • Lack of Anodal Circuit Contact
– All electricity travels in a circuit (circle) from a
positive pole (anode, ring) to a negative pole
(cathode, tip).
• Incompatible Lead / Header
• Pseudomalfunction
• Most Common: Atrial output sensed by the
ventricle and resetting the VAI.
Board Question 4 • DDI programmed to rate of 86 (VV=697ms), AVI=165ms,
blanking period=13ms.
• Referred for evaluation of abnormal pacemaker function.
• Most appropriate option: 1. Lengthen the PVARP
2. Lengthen the blanking period
3. Lengthen the AVI
4. Lower the rate -Taken from David Hayes, Cardiac Pacing Review for
EP Boards.
Crosstalk: Most Common Presentation
• DDI programmed to rate of 86 (VV=697ms), AVI=165ms, blanking period=13ms.
• Note: All intrinsic V activity, Rate~100?
• VV=AV+VA, VA=VV-AV, VA=697-165=532ms. – 532ms is longer than what we see on strip.
• Measured VV=575ms, 575-532=43ms .
• Blanking period of >43ms should avoid crosstalk. – Decrease atrial stimulus (if adequate safety margin)
– Decrease ventricular sensitivity.
-Taken from David Hayes, Cardiac Pacing Review for
EP Boards.
Crosstalk: Ventricular Safety Pacing • EGM confirming atrial pacing
coinciding with intrinsic ventricular
rhythm and retrograde P-wave in
alternate beats without pacing stimuli.
Note ventricular safety pacing in
alternate beats with atrial pacing (2
vertical lines in marker channel).
• Following atrial pacing, the pacer is
designed to trigger a ventricular pacing
output if ventricular sensing occurs
during the first portion–e.g., 110 ms of
the programmed A-V interval.
• It is difficult to determine whether these
ventricular paced events represent
ventricular safety pacing (VSP) or the
ventricular sense response (VSR)
algorithm. In cases in which both VSP
and VSR are enabled, the VSP
algorithm takes precedence during the
VSP interval within the paced AV delay
interval. Lim S, ―Ventricular safety pacing, ventricular sense response, and
ventricular tachycardia,‖ Heart Rhythm, Vol 7, No 4, April 2010.
Crosstalk: Ventricular Safety
Pacing
• Normal Atrial Capture
• Normal Ventricular Capture with two beats of functional
loss of capture
• Ventricular Sensing Unknown
• Atrial Undersensing with a Pseudo Pseudofusion beat
causing crosstalk with Ventricular Safety Standby seen
- Taken from St. Jude Medical
Failure to Pace:
Lead Fracture / Pseudofracture
Failure to Pace: Air in Header
Atrial EGM
Ventricular
EGM
This patient not pacer dependent.
Loss of Capture
• Dislodgement
• Elevated Thresholds
• Inappropriate Lead
Placement
• Fracture
• Insulation Failure
• Loose set screw
• Exit Block (>4 weeks)
• Perforation
• Battery/circuit Failure
• Air in Pocket
• Metabolic/Drugs
(Flecainide)
Loss of Atrial Capture?: 64 Year Old Female for
Device Check
Loss of Capture in Atrium
Friend of a Friend Calls Two
Weeks after Implant at OSH • Admitted with syncope and tachy-brady.
• DDD placed on right side. Pt is right handed and
hunts.
• Since implant patient sleeps sitting upright and
feels chest ―throb‖ when he lies flat.
• One week f/u with surgeon and device
functioning fine per report and told to ―live with
it.‖
• Pt. calls cardiologist and he echoes implanter’s
recs…‖you’ll have to get used to it.‖
Loss of Capture
• After discussion with friend, I told them to
mention the word ―perforation‖ to the
implanter.
• Implanter’s Response: ―If I perforated,
you’d be dead by now.‖ They did however
check a CXR and then scheduled patient
for ―heart surgery.‖
• Upon evaluation in our device clinic, R
waves sensed at 1.4mV and no V capture
at [email protected] (+diaph capture).
Loss of RV Capture
Other Examples of Perforation
Heart
Border
Signs of RV Perforation
• RBBB paced QRS morphology
– Can see RBBB configuration and
diaphragmatic stimulation in RV apical
position.
• Intercostal muscle or diaphragmatic
stimulation
• Friction rub after implant.
• Pericarditis, pericardial effusion, or cardiac
tamponade.
Failure to Pace after DDD?
Algorithm to Minimize Ventricular Pacing
Oversensing
• Lead fracture (increased impedance)
• Insulation Defect (decreased impedance)
• EMI
• Isoelectric ventricular event
• Sensing T/P waves
• Diaphragmatic potentials from integrated
bipole lead. (Bipole is tip to RV coil giving
larger antenna with which to receive signals)
Shock During Central PA Flooding
• Patient went into flooded basement with socks only.
• Intervals less than 140msec are nonphysiologic.
• Noise is on both atrial and ventricular channels.
13 Year Old Family Friend
Shocked at YMCA
T Wave Oversensing
Undersensing
• Change in intrinsic complex (BBB, VT/VF,
AF)
• MI
• Lead dislodgement / Poor positioning
– Tined leads
• Lead insulation failure
• Magnet application
• ERI
Undersensing
• Normal Atrial Capture with one beat
showing functional loss of atrial capture
• Atrial Undersensing
- Taken from St. Jude Medical
Management of High DFT’s • Reversible Causes: Amiodarone or PTX
• Reverse Shock Polarity
• Remove SVC Coil from Circuit (esp. if Z<40 Ohms)
• Reposition RV Lead
• Add Subcutaneous Array
• Additional Lead: Coronary Sinus or Azygous Vein.
• Optimize waveform tilt (SJM only)
• Consider using sotalol (or dofetilide)
- Mainigi S, Callans D, ―How to manage the patient with a high defibrillation
Threshold,‖ Heart Rhythm, V. 3, No. 4 (April 2006), pp. 492-495.
Failed DFT’s
- I know, there is no ICD.
Board Question 5
• Which of the following configurations is
likely to have the lowest DFT?
1. Distal coil to proximal coil
2. Proximal Coil and can to Distal Coil
3. Distal Coil to SC Array + Can
4. SC Array to Can
- Avoid using Distal Coil + Can to Proximal Coil
What Hardware is Present?
37yo Admitted with Failed DFT
RV lead with high septal position and there is old epicardial patch.
Failed DFT’s in Lab
• EGM1 = Rvtip-Rvring
• LEKG = Can-SVC
• EGM3 = Rvtip-Rvcoil
• EGM2 = Can-RVcoil
Initial Configuration
Corrected Configuration
Arrhythmia Classification
• Use all the available information: surface,
atrial, and ventricular EGMs.
• What ―drives‖ what.
• How does arrhythmia terminate?
• Response to ATP and shocks.
• Morphology during sinus versus
arrhythmia.
What are the Rhythms?
Atrial fibrillation Note morphology
Change to VT
Antitachycardia
Pacing
Diagnose this Arrhythmia
• VAV Response with AVNRT
• When atrium entrained during Vpacing, examine the EMG sequence
immediately after the last paced ventricular complex: – AVNRT = V-A-V response (may also indicate ORT)
– Atrial Tachycardia = V-A-A-V response
- EGM from Ellenbogen, ―ICD’s: Evaluation
and Troubleshooting, 2008 HRS Board Review.
Diagnose this Arrhythmia
• VAAV Response to Overdrive V pacing
with entrainment indicates atrial
tachycardia. - EGM from Ellenbogen, ―ICD’s: Evaluation
and Troubleshooting, 2008 HRS Board Review.
Board Question 6
• A 80 yo male with nonischemic cardiomyopathy (EF15%) and neurofibromatosis
underwent implantation of an ICD for rapid sustained hypotensive ventricular
tachycardia and syncope. He was treated with amiodarone and is undergoing ICD
implant from left subclavian position. He has an existing DDDR pacemaker in the
right subclavian position. The figure shows part of the testing procedure performed
at the time of device implantation. The pacemaker mode is DOO and the output is
programmed to maximal values. It demonstrates which of the following? 1. Testing for double counting.
2. Inhibition of atrial and ventricular pacing by VF.
3. Failure of the ICD to deliver a shock.
- Taken from Nacarelli et al, ACC EPSAPII.
4. Appropriate ICD function during dual
chamber pacing.
5. The requirements for multiple shocks
during pacing to terminate VF.
Board Question 6 (cont’d)
• Appropriate ICD function during dual-chamber pacing.
• Pacer is programmed to DOO at max outputs on atrial
and ventricular channels.
• VF is induced; DOO pacing continues.
• The device is able to sense VF and delivers a shock
to terminate it.
- Taken from Nacarelli et al, ACC EPSAPII.
Acknowledgements • Good Samaritan Hospital
• Lebanon Cardiology Associates
• PA Chapter of the American College of Cardiology
• Drs. Dave Schwartzman, Samir Saba, David Hayes,
Kenneth Ellenbogen.
• Our patients
- Please email me ([email protected]) with
any interesting device-related CXR’s.