paper #: 7250-40 spie medical imaging...
TRANSCRIPT
Jong Chul Hana, Seungman Yuna, Chang Hwy Lima, Tae Woo Kimb, Ho Kyung Kima*
a School of Mechanical Engineering, Pusan National University, Busan 609-735, South Korea b Advanced Medical Engineering Laboratory, Vatech Co., Ltd., Giheung 449-904, South Korea* Correspondence: [email protected]
Feasibility study of CMOS detectors for mammographyPusan National University
Motivation•Many types of digital mammography detectors are currently used as clinical systems.•A photodiode array made by CMOS (complementary metal-oxide-semiconductor) process
in conjunction with a scintiallator has lately attracted considerable attention due to its unique advantages such as low read out noise, small pixel size, and high fill factor, etc.
•With the CMOS detector employing a thin phosphor screen as an x-ray converter the extra noise due to the direct absorption of x-ray photons unattenuated from the phosphor screen in the photodiode is white.- This additional white noise further raises the noise spectral density in the high spatial frequency
band, hence deteriorating the DQE performance.
Objective•To investigate the potential use of CMOS detectors employing a phosphor screen for
mammographic applications- The imaging performance of the detectors is characterized in the Fourier domain, such as
modulation-transfer function (MTF), noise-power spectrum (NPS) and detective quantum efficiency (DQE). - Componential analysis for various image noise sources due to the optical photons generated, the direct x-ray photons unattenuated from the phosphor screen, and the additive electronic noise charges is performed by experimental measurements.- The role of a fiber-optic plate (FOP) layer in the detector performance is also investigated.
Conclusions•The indirect-conversion detectors have an NPS, which
decreases with the spatial frequency, and the direct-conversion detectors have a nearly constant NPS with the spatial frequency.
•This explains that when a significant amount of x rays are not absorbed in the phosphor layer, then the additional absorption of x rays in the photodiodes with their white noise contributions degrades the total NPS performance.
•Although the introduction of the FOP layer reduces the sensitivity and MTF of the detector, it largely enhances the DQE performance because it plays as an efficient stopper for the direct x rays.
•The use of a combination of thicker columnar-structured CsI:Tl and FOP layers may provide a more efficient mammography detector with higher DQE performance but without the significant loss in the MTF performance.
AcknowledgementsThis work was supported by the Korea Research Foundation Grant (KRF-2008-313-D01339) and
the Korea Science and Engineering Foundation Grant (R01-2006-000-10233-0) funded by the Korea Goverment (MEST), and Vatech Research Grant (Year 2008).
CMOS detectors Cascaded model analysis
Results
•CMOS photodiode array - RadEyeTM, Rad-icon Imaging Corp.
CMOS photodiode array
MinR-2000TM phosphor screen
- Phosphor screen (MinR-2000TM, Carestream Healthcare, Inc.)• Mainly made up of terbium-doped gadolinium oxysulfide
(Gd2O2S:Tb) and a polyurethane elastomeric as a phosphor and binder
• 84 µm thickness & ~4 g/cm3 density
Property Description
Array format 512 × 1024 pixels
Pixel pitch 48 µm
Field of view 25 × 50 mm2
Dynamic range 85 dB (>14 bits)
ADC bit-depth 12 bits
Pixel fill factor 0.87
Saturation 2,800,000 electrons
Dark current < 104 electrons/sec
Read Noise (at 1 fps) < 200 electrons
•Modeling signal and noise characteristics using the cascaded linear-systems theory
•Theoretical DQE calculation
Parameter Description Parameter Description
q0 Incident x-ray fluence Tapert(ρ) MTF due to the aperture integration
αscn Quantum efficiency in the scintillator Tscn(ρ) MTF of scintillator
βscnQuantum amplification factor in the scintillator
Ipd Swank noise factor in photodiode
γpdLight quantum efficiency in the photodiode a Pixel aperature
αpd Quantum efficiency in the photodiode d Pixel pitch
βpdQuantum amplification factor in the scintillator Iscn Swank noise factor in the scintillator
σadd Additive electronic noise III(ρ) Sampling process
•Measured Characteristic curves of two different configurations
-The signal is expressed in the number of electrons (e–) considering the CMOS photodiode conversion gain.
- The introduction of FOP reduces the detector sensitivity (or the slope of the linear fit curve) by a factor of 2.- The measured detector sensitivities are extremely low compared to the reported data with an amorphous silicon-based detector.• Smaller pixel pitch of the detector used in this study maybe one
reason for this large differences.• The exact reason is unclear and will be further investigated.
•Measured MTFs of two different configurations of the detector with respect to various energies
- The MTF curves are quite similar to one another for a range of energy from 40 – 80 kVp. - Due to erroneous MTF results for 26 kVp, the MTF of 40 kVp was used for DQE calculations.- Comparing the graph (a) and (b), the FOP largely degrades the MTF performance over the entire spatial frequency region.
•Measured NPSs and DQEs of two different detector configurations with respect to various dose
- In the graph (a), • The use of FOP reduces the level of noise spectral densities over
the entire spatial frequency regions.• The configuration of the detector without the FOP layer shows a
more flat shaped spectrum, due to the contribution of direct x-ray absorption noise.- In the graph (b),
• DQE of the detector without the FOP is almost independent on the dose level.
• Although the FOP layer reduces the detector sensitivity and degrades the MTF performance, it greatly improves the DQE performance because it effectively prevents direct x-ray photons from the absorption within the photodiode.
•Componential analysis of noise spectral densities
- In the graph (a), in the case of detector without FOP• From the measurements, the scattered and direct x rays significantly
contribute to the noise spectral densities, especially in the high spatial frequency region because of the white noise characteristics. - In the graph (b), in the case of detector with FOP
• The contributions due to the scattered and direct x rays are negligible. - Although the cascade models developed in this study do not consider
the scattered x rays, the cascade models reasonably describe the measurements in both cases.
•DQE comparison between the measurement and theoretical calculation, and DQE(0) as a function of fluence for two detector configurations
- In the graph (a),• The discrepancy of the model for the detector with FOP is mainly
due to the incomplete model of the FOP layer. • The discrepancy of the model for the detector without FOP is mainly
due to incompleteness in the model.- In the graph (b),• The range of dose considered in this study is indicated as a hatched
box.• From the calculation results, maximally achievable DQE(0) is about
30% with the CMOS detectors investigated in this study.
Experimental
X-ray quality
IEC HVL Measured HVL
IEC Tube voltage
Adjusted voltage IEC q Measured
qX-ray quality
mmAlmmAl kVpkVp #/mm2/µGy#/mm2/µGy
W/Al 0.83 0.82 28 26 6575 6500
•Componential analysisDetector configurationsDetector configurations Noise sources
Detector without FOP
dark image σadd2
Detector without FOP
phosphor + photodiode σopt2 + σdirect2 + σscatter2 + σadd2Detector without FOP phosphor + black paper + photodiode σdirect2 + σscatter2 + σadd2
Detector without FOP
phosphor + air gap (+ carbon plate) + black paper +photodiode σdirect2 + σadd2
Detector with FOP
dark image σadd2
Detector with FOP
phosphor + FOP + photodiode σopt2 + σscatter2 + σadd2
Detector with FOP
phosphor + black paper + FOP + photodiode
σscatter2 + σadd2Detector with FOP
phosphor + air gap (+ carbon plate) + FOP + black paper + photodiode σadd2
The subscripts opt, direct, scatter, and add denote each corresponding component contributing to noise including optical photons, direct x rays, scattered x rays, and additive electronic noise, respectively.
•Source information - X-ray tube (UltraBright TM, Oxford Instrument)• Tungsten (W) target, 245 µm beryllium exit window, and DC type
•Standard radiation qualities and characteristics for the
mammographic application
•Blackout paper (Flock Paper #40, Emond Optic Inc.)- For the measurement of direct x-ray signals without lights- 100 µm thickness
•FOP (fiber optic plate: Income Corp)- 3 mm thickness & 4 g/cm3 density- To prevent the direct x-rays from the absorption within the
photodiode array
Cascaded model due to optical photon
Cascaded model due to direct x-ray absorption
•Measured DQE calculation
(a) (b)
DQE(ρ) = q0G2MTF2(ρ)NPS(ρ)
the optical quantum correlated noise term
NPS due to the direct x-ray absorption NPS due to the additive electron noise charges
WU =a2d 2q0 (1− α scn )α pdβ pd
2
I pdWadd = d
2σadd2
WQ (ρ) = a4q0α scnβscnγ pd 1+ γ pdβscn
Iscn−1
⎛⎝⎜
⎞⎠⎟Tscn2 (ρ ± k
d)
⎡
⎣⎢
⎤
⎦⎥sinc2 πa(ρ ± k
d)⎧
⎨⎩
⎫⎬⎭k=0
∞
∑
DQE(ρ) =q0 α scnβscnγ pd + (1− α scn )α pdβ pd⎡⎣ ⎤⎦
2Tscn2 (ρ)sinc2 (πaρ)
WQ (ρ) +WU +Wadd
SPIE Medical Imaging 2009Paper #: 7250-40
BiomedicalMechatronics Lab
BML