parameters for indirect viewing of visual signals used in
TRANSCRIPT
Parameters for Indirect Viewing of Visual
Signals Used in Emergency Notification
Final Report
Prepared by:
John D. Bullough Nicholas P. Skinner
Yiting Zhu
Lighting Research Center, Rensselaer Polytechnic Institute
© September 2013 Lighting Research Center, Rensselaer Polytechnic Institute
THE FIRE PROTECTION RESEARCH FOUNDATION ONE BATTERYMARCH PARK
QUINCY, MASSACHUSETTS, U.S.A. 02169-7471 E-MAIL: [email protected] WEB: www.nfpa.org/Foundation
iii
FOREWORD
LEDs and other innovative energy saving lighting technologies (e.g. fluorescents) are rapidly entering the marketplace and present themselves for application to emergency notification appliances. The existing requirements for the performance and application of visible notification appliances are based on relatively short duration, high peak intensity flashing lights – strobe lights. NFPA 72, National Fire Alarm and Signaling Code, and referenced listing standards define a method for calculating the equivalent or effective intensity of a flashing light source. The calculation method is subjective and does not produce an exact comparison and is intended only to approximate the perceived brightness for direct viewing of the light source. It has worked because all of the lights approved using the standard have all had relatively similar and short pulse durations. Thus, the peak intensities have been relatively similar. A review of research performed for the Fire Protection Research Foundation by the RPI Lighting Research Center suggested that effective intensity may not be predictive of visual detection of signal lights when these are viewed indirectly or in the far-peripheral field of view. In particular, observers see the change in illuminance on room surfaces rather than the flashing light itself when it is not in the central field of view. Based on previous literature, the previous Foundation study suggested that a flashing light should increase the illuminance on the opposite wall by at least 7% in order for this increase to be detected reliably. For an ambient horizontal illumination level of 100 footcandles (fc) on the work plane in a space such as an office, it was estimated that the vertical illuminance on the wall should increase by at least 2 fc to be reliably detected. This estimate has not been tested empirically. The objective of this project was to conduct a human factors laboratory study to identify whether the 7% increase in light level can be reliably detected by observers with normal vision when viewed indirectly. The results from this study provide technical basis to support the development of methods and criteria to evaluate performance of light sources used in emergency notification appliances for NFPA 72. The Research Foundation expresses gratitude to the report authors John D. Bullough, Nicholas P. Skinner, and Yiting Zhu with the Lighting Research Center at Rensselaer Polytechnic Institute located in Troy, New York. The Research Foundation appreciates the guidance provided by the Project Technical Panelists, the funding provided by the sponsors, and all others that contributed to this research effort. The content, opinions and conclusions contained in this report are solely those of the authors.
Keywords: fire protection building systems, visible notification appliances
iv
v
PROJECT TECHNICAL PANEL
Robert Elliot, FM Approvals
Bruce Fraser, Fraser Fire Protection Services
Warren Olsen, Fire Safety Consultants, Inc.
Lee Richardson, NFPA Staff Liaison
Robert Schifiliti, R.P. Schifiliti and Associates, Inc.
Andrew Trotta, Consumer Product Safety Commission
PROJECT SPONSOR REPRESENTATIVES
Dan Finnegan, Larry Grodsky, and Mark Pavlica, Siemens
Isaac Papier, Honeywell/System Sensor and National Electrical Manufacturers
Association
Rodger Reiswig, SimplexGrinnell
CONTRIBUTING PROJECT SPONSOR REPRESENTATIVES
Jack McNamara, Bosch
Dave Newhouse, Gentex Corporation
vi
vii
TableofContents1.Background..........................................................................................................................................................12.Methods..................................................................................................................................................................3 2.1.ExperimentalLaboratory3 2.2.TestLightSource4 2.3.ExperimentalConditions6 2.4.SubjectsandProcedure73.Results.....................................................................................................................................................................8 3.1.Experiment18 3.2.Experiment29 3.3.Experiment39 3.4.Experiment411 3.5.Experiment511 3.6.Experiment613 3.7.Experiment7134.Discussion...........................................................................................................................................................16 4.1.PartialTemporalSummation16 4.2.IntegrationTimes165.Conclusions........................................................................................................................................................216.Acknowledgments..........................................................................................................................................227.References..........................................................................................................................................................23Appendix:ExperimentalData......................................................................................................................24
1
1.BackgroundVisualsignalingappliancesusedforemergencynotificationhavecommonlyusedxenonstrobelightsthatproduceverybrief(<1ms),high‐intensityflashesoflight,andtheirspecifiedperformanceischaracterizedbytheireffectiveintensity,whichisanestimateoftheluminousintensity(incandelas[cd])ofasteady‐burninglightthathasequivalentvisualeffectivenessastheflashinglight.Evaluationsofvisualsignalsusedinawidevarietyofapplicationshavegenerallyconfirmedtheutilityoftheeffectiveintensityforvisualsignalswhentheyareviewedwithinthecentralportionofthefieldofview(IALA,2008;Bulloughetal.,2013;BulloughandSkinner,2013),regardlessofthespecifictemporalcharacteristicsoftheflashinglight.Theeffectiveintensity(Ie,incd)isdefined(IES,1964)intermsofEquation1: Ie=∫t1t2I(t)dt/(a+t2–t1) (Eq.1)wheret1andt2arethestartandendtimes(inseconds[s])oftheflashoflight,respectively;I(t)istheinstantaneousluminousintensity(incd)oftheflashattimet;andaisaconstantdeterminedempirically(BlondelandRey,1912)tohaveavalueofapproximately0.2s.Thisconstantisrelatedtothetemporalintegrationofthevisualsystemundervisualconditionssimilartothoseunderwhichnavigationalsignallightsarejustatthethresholdfordetectionatnight,whenthevisualsystemisdark‐adapted.Eventhoughmanyvisualsignals,includingthosedeployedinbuildingsforemergencynotificationapplications,arenotviewedunderdarkadaptation,andaredesignedtobeseenatwellabovethresholdlevels,theeffectiveintensityconcepthasheldupforon‐axisandnear‐on‐axisviewingconditions(referredinthisreportas"direct"viewing).InspectionofEquation1suggeststhatthesameeffectiveintensitycanbeachievedwithsignallightshavingverydifferenttemporalintensitycharacteristics.Thedurationoftheflashoflightanditsinstantaneousintensitycanbetradedoff,sothataverybrief,high‐intensityflashcouldhavethesameeffectiveintensityasalonger,lower‐intensityflashoflight.Thishasimplicationsforvisualsignalsusingsolid‐statelightingtechnologysuchaslight‐emittingdiodes(LEDs).LEDsareavailablewithincreasingefficiencyandbrightness,makingthesesourcespracticalforvisualsignalingdevices.Unlikethexenonsourcesinstrobelights,LEDscanbeflashedwithdifferentdurationsandtemporalwaveforms,sothattwoappliancescouldhavethesamecalculatedeffectiveintensitybuthaveverydifferenttemporalflashpatterns(Bulloughetal.,2012a).Areview(Bulloughetal.,2012b)ofpreviousliteraturedescribingtheresultsofstudiesconductedwithxenonstrobelightsources(UL1991)confirmedthatforindirectdetectionwhenthevisualsignalisinornearthefieldofview,aneffectiveintensityof15cdprovidedreliablelevelsofdetection.However,theresultsofthesamestudiesshowedthatflashingincandescentsourcesusedasvisualsignalsrequiredmuchhighereffectiveintensitiestobedetectedreliably.Incandescentflashingsourceshavelongerflashdurationsthatxenonstrobelightsbecauseoftheinherentpropertiesofthefilamentthatmustheatuptoproducelight,andcoolofftostopproducinglight.AmorerecenttestusingLEDsources
2
varyinginintensityanddurationtoachievethesame15cdeffectiveintensity(Savage,2011)reportedthatsourceswithlongerdurationstendedtohavelowerdetectability.Researchstudiesonlargefield,lowfrequencyflickerperception(Kelly,1961)suggestedthatwhenflashingofalargefield(asinindirectviewingofanemergencyvisualsignal)occurredatafrequencynear1‐2Hz,theabsolutemodulationlevelofabout7%producedreliabledetection.Basedonthesefindings,Bulloughetal.(2012b)postulatedthatperhapstheabsoluteorinstantaneousintensityfromasignalwhenviewedindirectlymightbemoremeaningfulthatitseffectiveintensitycharacterizedusingEquation1.ToassisttheFireProtectionResearchFoundation(FPRF)oftheNationalFireProtectionAssociation(NFPA)inidentifyingpredictivemetricsforcharacterizingtheperformanceofvisualsignalshavingtemporalpropertiesdifferentfromxenonstrobes,aseriesofhumanfactorsexperimentswasconductedbytheLightingResearchCenter(LRC);theseexperimentsaredescribedinthepresentreport.
3
2.Methods2.1.ExperimentalLaboratoryThetestlaboratoryusedforthehumanfactorsexperimentwastheSeminarRoomattheLRClaboratoryinTroy,NY.Thisisalargeclassroomspacewithwhitepaintonthreewallsandunpaintedbrickonthefourthwall.Figure1illustratesaschematicdiagramoftheoverallexperimentalset‐up,showingtherelativelocationsoftheexperimentalsubjects,thetestlightsourceandtheoppositewallinthefieldofview(nottoscale).Subjectswereseatedatatablefacingawayfromtheunpaintedbrickwall.Figure2showsthesubjects’viewofthewhitefallfacingthem.
Figure1.Schematicofexperimentallaboratory(positionofsubjectsnottoscale).
Figure2.Viewoffacingwallfromsubjects’position.
Thelightingsystemintheroomisabletobecontrolledthroughaseriesofdimmingswitched.Dependingupontheexperiment,theilluminanceintheseminarroomwasadjustedtoproducethefollowingconditions: Highambientcondition:Averagehorizontalilluminanceontabletopswas500lx,
averageverticalilluminanceonwallfacingsubjectswas200lx
4
Lowambientcondition:Averagehorizontalilluminanceontabletopswas250lx,averageverticalilluminanceonwallfacingsubjectswas100lx
2.2.TestLightSourceFigure3showsthetestlightsource,whichconsistsofanarrayofthreehighpowerwhiteLEDs(Manufacturer:Cree,Partnumber:XREWHT‐L1‐0000)overwhichlenses(Manufacturer:Khatod,Partnumber:KEPL1127)arefitted.Thelensesincludethreepossibletypesofopticaldistributions:40o,25oor6o.(ThebluefilmsvisibleoverthefrontlensaperturesinFigure3areprotectivecoversandareremovedduringtesting.)Thecorrelatedcolortemperature(CCT)rangeofthewhiteLEDswasapproximately3200KinordertomatchtheCCTofthefluorescent(3500K)andhalogen(2800K)lightinginthetestlaboratory.
Figure3.Testlightsource.
Twotypesofoptics(Manufacturer:Khatod)wereusedtogetherwiththeLEDs:onecreatesanarrowbeamangleof6(Partnumber:KEPL112706)whiletheothercreatesawidebeamangleof40(Partnumber:KEPL112740).AsshowninTable1,toincreasetheverticalilluminanceby15lx(anapproximate7%verticalilluminanceincrease)onthefacingwallatheightof6ft,itrequireddrivingcurrenttothewhiteLEDtobe61mAforthe6obeamangleopticand961mAforthe40obeamangleoptic.Figure4illustratesopticalray‐tracingsimulationresultsofdifferentbeampatternsonthewallwithdifferentbeamangleoptics:6and40.TheLEDandtheopticsweremountedatheightof6ft,andtheyareaimingperpendicularlytothewall,whichwas20ftawayfromthelightsource.AsillustratedinFigure3,6opticsilluminatea2ftby2ftarea,whilethe40opticsilluminatealmostthewhole12ftby12ftarea.Therefore,6,and40opticswerechoseninthisexperimenttorepresentnarrowandwidebeamtypes.Thelightsourcewasmountedintheceilingbehindthesubjects’seatingposition.Aplywoodbafflewasusedtoblockaportionofthelightfromthesourcetoavoidilluminatingashadowedportionofthewallfacingthesubjects,showninFigure5.
5
Figure4.Opticalray‐tracingsimulationofverticalilluminanceonthewallwhenusingdifferentbeamangleoptics(6,topand40,bottom)ona12ftby12ftarea
withthelightsourceaimedatthecenter.
Figure5.Photographofthebaffleandlightsourcemountedintheceiling.
6
2.3.ExperimentalConditionsThelightsourceswerecontrolledusingacustomLabview(NationalInstruments)programhatcouldproduceflashesoflightvaryinginintensity(intermsoftheilluminanceincrementonthewallfacingthesubjects’seatingposition)andduration,withflashratesfrom1Hz(oneflashpersecond)to2Hz(oneflashperhalf‐second).Insomeexperiments,subjectswererequestedtolookdirectlyaheadatthewallfacingthem,andinothers,subjectsperformedanumericalverificationtask(NVT)placedonthetableinfrontofthemandwouldhavedetectedtheflashingintheirperipheralvision.Thenumericalverificationtaskconsistedofprintedcolumnsofnearlymatching5‐digitnumbers,with3%ofthedigitsnotmatching.Subjectswereinstructedtoplaceacheckmarknearnon‐matching5‐digitnumbersandtoreportwhethertheysawflashingwhileperformingthistask.Inthefinalexperiment,subjectsunawareofthepurposeoftheexperimentperformedthetaskandthenansweredwhethertheynoticedtheflashingafteranexperimentalconditionwasdisplayed.Table1.Preliminarymeasurementresultsoftheilluminancelevel(“I”represents
drivingcurrent)Height(ft)
VerticalIlluminance(lx)Ambientlightonwall 6optics(I=61mA) 40optics(I=961mA)
8 117 7 206 6 219 15 155 208 4 198
Thesourcewascalibratedandadjustedtoproducethefollowingexperimentalconditionsineachofthesevenexperimentsthatwereconducted: Experiment1:Ambientilluminance500lx;Beamangle40o;Duration1,10,100ms;
Illuminanceincrement1%,2%,4%,8%;Frequency1Hz;Subjectslookingatwall Experiment2:Ambientilluminance500lx;Beamangle40o;Duration10,25,50,100
ms;Illuminanceincrement1%,2%,4%,8%;Frequency2Hz;Subjectslookingatwall Experiment3:Ambientilluminance250lx;Beamangle40o;Duration10,25,50,100
ms;Illuminanceincrement2%,4%,8%,16%;Frequency1Hz;Subjectslookingatwall Experiment4:Ambientilluminance500lx;Beamangle40o;Duration10,25,50,100
ms;Illuminanceincrement1%,2%,4%,8%;Frequency1Hz;SubjectsperformingNVT Experiment5:Ambientilluminance250lx;Beamangle40o;Duration10,25,50,100
ms;Illuminanceincrement2%,4%,8%;16%;Frequency1Hz;SubjectsperformingNVT
Experiment6:Ambientilluminance250lx;Beamangle6o;Duration10,25,50,100ms;Illuminanceincrement2%,4%,8%,16%;Frequency1Hz;SubjectsperformingNVT
Experiment7:Ambientilluminance250lx;Beamangle40o;Duration50ms;Illuminanceincrement4%,16%;Frequency1Hz;SubjectsperformingNVT
7
InExperiments4and5acommerciallyavailableemergencynotificationvisualsignalwasalsoincluded.Thesignalhadanadjustableeffectiveintensitysetting,whichwassetduringbothexperimentstothenominal15cdsetting.Basedonintensityprofiledataprovidedbythemanufacturer,anestimatedvalueof40cdfortheeffectiveintensity(whencalculatedusingEquation1)wasusedforsubsequentanalysis.Duringtheexperimentsinwhichthissignalwasused,itwasfittedwithabaffletoproduceadistributionsimilartothatoftheLEDtestsourcewhenequippedwiththe40olensoptics.2.4.SubjectsandProcedureTensubjectsparticipatedineachexperiment.InExperiments1through6,subjectswereexposedtoeachoftheexperimentalconditionsinarandomorderfor10safterwhichtheywereaskedwhethertheydetectedtheflashinglight.Threenullconditiontrialswithnoflashingpresentwerealsopresentedineachofthesesixexperimentstomeasureresponseswhennosignalwaspresent.InExperiment7,subjectswereshownonecondition(4%or16%illuminanceincrement)only;halfsaweachcondition.InExperiments1through7,subjectsalsoratedtheease/difficultyofdetectionusingthefollowingscale(iftheydidnotdetecttheflashing,avalueof‐3wasassignedastheresponsetothisquestion): +2 Veryeasy +1 Somewhateasy 0 Neithereasynordifficult ‐1 Somewhatdifficult ‐2 VerydifficultInExperiments2through7,subjectsalsoratedtheperceivedurgencyofdetectionusingthefollowingscale(iftheydidnotdetecttheflashing,avalueof‐1wasassignedastheresponsetothisquestion): 3 Veryurgent 2 Somewhaturgent 1 Slightlyurgent 0 Notatallurgent
8
3.ResultsTheresultsofeachexperimentareincludedinthissection,withthedetectionpercentages,ease/difficultyratings,andurgencyratingsforeachconditionplottedasafunctionoftheeffectiveintensityforeachconditioncalculatedusingEquation1.Forthenull‐conditiontrialsinalloftheexperiments,the“falsepositive”detectionratesinalloftheexperimentsrangedfrom0%to3%.Theserateswerelowenoughtofeelconfidentthatsubjectswereonlyveryrarelyrespondingthattheydetectedsomethingwhennoflashingconditionwaspresented,andthatthedetectionpercentagesreportedherearerepresentativeofthelikelihoodthatasignallightwouldbedetected.3.1.Experiment1Figure6ashowsthedetectionpercentagesandease/difficultyratingsforExperiment1.
a.
b. Figure6.a:DetectionpercentagesforExperiment1;b:EaseratingsforExperiment1.ItcanbeseeninFigure6thattheabilitytoseetheflashingsignalsindirectlywasnotwellpredictedbytheireffectiveintensitybasedonEquation1.Norwastheabsoluteilluminanceincrementpredictiveofperformance.Forexample,inFigure6a,the10msflashwitha2%illuminanceincrementyieldedonly20%detection,butthe100msflashwiththesameilluminanceincrementwasdetected90%ofthetime.The1mssignalswerehardlyeverdetected.
9
Basedontheseresults,the1msconditionswereeliminatedfromfutureexperiments,andinordertoprovidemoreresolutionbetween10and100ms,flashdurationsof25and50mswereincluded.Approximately,theflashdurationsforeachdurationincrementdoubledexceptforthestepbetween10and25increment,afactorof2.5increase.3.2.Experiment2Figure7ashowsthedetectionpercentages,Figure7btheease/difficultyratings,andFigure7ctheurgencyratingsforExperiment2.
a. b.
c.
Figure7.a:DetectionpercentagesforExperiment2;b:EaseratingsforExperiment2;c:UrgencyratingsforExperiment2.
AswithExperiment1,neithereffectiveintensitynortheabsoluteilluminanceincrementwaspredictiveofperformancefortheconditionsinExperiment2.Inaddition,althoughtheflashfrequencybetweenExperiments1and2differed(1HzinExperiment1and2HzinExperiment2),thedetectionperformanceandratedease/difficultyfortheconditionscommontobothexperiments(10and100ms,foreachilluminanceincrement)werehighlycorrelatedtoeachother(r2=0.90fordetection,andr2=0.96forease/difficulty)andverysimilarinmagnitude.Basedonthiscorrelation,allsubsequentexperimentsusedaflashrateof1Hz,sincetheredoesnotappeartobeaneffectoffrequencybetween1and2Hz.3.3.Experiment3Figure8ashowsthedetectionpercentages,Figure8btheease/difficultyratings,andFigure8ctheurgencyratingsforExperiment3.
10
a. b.
c.
Figure8.a:DetectionpercentagesforExperiment3;b:EaseratingsforExperiment3;c:UrgencyratingsforExperiment3.
a. b.
c. Figure9.Correlationofdetection(a),ease/difficulty(b),andurgency(c)forcorresponding
conditionsinExperiments2and3.Asexpected,sincetheambientlightlevelinExperiment3waslowerthaninExperiments1and2,theperformanceandsubjectiveratingswerehigher.Comparingtheconditionsfor
11
whichtheflashdurationandtherelativeilluminanceincrement(2%,4%and8%)werecommontobothexperiments,theresultsoftheseexperimentswereverysimilarandstronglycorrelatedwitheachother(Figure9).Thissuggeststhattherelativeilluminanceincrementratherthantheabsoluteincrementismoreimportantfortheperformanceofasignallightwhenviewedindirectly.Nonetheless,thepositivey‐interceptsofthebest‐fittinglinearfunctionsinFigure9suggestthatperformancewasslightlyimprovedunderthehigherambientlevel,forthesamedurationandrelativeilluminanceincrement.3.4.Experiment4Figure10ashowsthedetectionpercentages,Figure10btheease/difficultyratings,andFigure10ctheurgencyratingsforExperiment4.
a.
Experiment 4: 500 lx, 1 Hz (visual task)
0%
20%
40%
60%
80%
100%
1 10 100 1000
Effective Intensity (cd)
Detection (%)
Xenon
10 ms
25 ms
50 ms
100 ms
b.
Experiment 4: 500 lx, 1 Hz (visual task)
‐3
‐2
‐1
0
1
2
1 10 100 1000
Effective Intensity (cd)
Mean
Ease Rating
Xenon
10 ms
25 ms
50 ms
100 ms
c.
Experiment 4: 500 lx, 1 Hz (visual task)
‐1
0
1
2
3
1 10 100 1000
Effective Intensity (cd)
Mean Urgen
cy Rating
10 ms
25 ms
50 ms
100 ms
Xenon
Figure10.a:DetectionpercentagesforExperiment4;b:EaseratingsforExperiment
4;c:UrgencyratingsforExperiment4.Thedetectionperformanceislowerforthisexperiment,inwhichsubjectsperformedtheNVTduringthestudy,thanforExperiment2,inwhichsubjectswerepermittedtolookdirectlyatthewall.Itcanalsobeseenthatthedetectionandratingsforthexenonsignalwerehigherthanforanyoftheotherconditions.3.5.Experiment5Figure11ashowsthedetectionpercentages,Figure11btheease/difficultyratings,andFigure11ctheurgencyratingsforExperiment5.
12
a.
Experiment 5: 250 lx, 1 Hz (visual task)
0%
20%
40%
60%
80%
100%
1 10 100 1000
Effective Intensity (cd)
Detection (%)
10 ms
25 ms
50 ms
100 ms
Xenon
b.
Experiment 5: 250 lx, 1 Hz (visual task)
‐3
‐2
‐1
0
1
2
1 10 100 1000
Effective Intensity (cd)
Mean
Ease Rating
10 ms
25 ms
50 ms
100 ms
Xenon
c.
Experiment 5: 250 lx, 1 Hz (visual task)
‐1
0
1
2
3
1 10 100 1000
Effective Intensity (cd)
Mean Urgen
cy Rating
10 ms
25 ms
50 ms
100 ms
Xenon
Figure11.a:DetectionpercentagesforExperiment5;b:EaseratingsforExperiment
5;c:UrgencyratingsforExperiment5.
a.
y = 0.68x + 0.11R² = 0.60
0%
10%
20%
30%
40%
50%
60%
70%
80%
0% 20% 40% 60% 80% 100%
500 lx
250 lx
Detection
b.
y = 0.75x ‐ 0.54R² = 0.82
‐4
‐3
‐2
‐1
0
‐3 ‐2 ‐1 0 1 2
500 lx
250 lx
Difficulty
c.
y = 0.79x ‐ 0.01R² = 0.74
‐2
‐1
0
1
‐1 0 1 2 3
500 lx
250 lx
Urgency
Figure12.Correlationofdetection(a),ease/difficulty(b),andurgency(c)forcorresponding
conditionsinExperiments4and5.TheperformanceinExperiment5washigherthanforExperiment4,whichusedahigherambientlightlevel.However,whencomparedforthesamedurationandthesamerelative
13
illuminanceincrementscommontobothexperiments(2%,4%,8%),theresultsweresimilarandreasonablycorrelatedwitheachother(Figure12).Thepositivey‐interceptsinFigure12suggest,asinFigure10,thatperformancewasslightlyimprovedunderthehigherambientlevelforthesamerelativeilluminanceincrement.3.6.Experiment6Figure13ashowsthedetectionpercentages,Figure13btheease/difficultyratings,andFigure13ctheurgencyratingsforExperiment6.
a. b.
c. Figure13.a:DetectionpercentagesforExperiment6;b:EaseratingsforExperiment
6;c:UrgencyratingsforExperiment6.ItcanbeseenthatusingthenarrowbeamlightsourceinExperiment6resultedingenerallylowdetectionperformanceandlowratingsofease/difficultyandurgency.Onlythehighestilluminanceincrement(16%)foradurationof50mswasdetectedatleast50%ofthetime.3.7.Experiment7Figure14ashowsthedetectionpercentages,Figure14btheease/difficultyratings,andFigure14ctheurgencyratingsforExperiment7.
14
a. b.
c. Figure14.a:DetectionpercentagesforExperiment7;b:EaseratingsforExperiment
7;c:UrgencyratingsforExperiment7.
a. b.
c. Figure15.Correlationofdetection(a),ease/difficulty(b),andurgency(c)forcorresponding
conditionsinExperiments5and7.
15
FewerconditionswereusedinExperiment7thaninExperiment5,whichusedcorrespondingconditions.ThisisbecausesubjectsinExperiment7onlyviewedasingleconditioninordertoidentifytheresponsesofunalertedsubjectsunawareofthepurposeoftheexperiment;oncesubjectswereaskedwhethertheydetectedthefirstconditiontheysaw,theywouldbeawareofthenatureoftheexperimentforanysubsequentpresentations.Tocomparetheimplicationsofunalerted,unawaresubjects,thedatainExperiments5and7werecomparedforcorrespondingconditionsinFigure15.Similartothecomparisonsbetweenambientlightlevels,theperformanceandresponsesoftheunalertedsubjectswhowereunawareofthepurposeoftheexperiment,wasgenerallylowerthanforsubjectswhowereawareofthepurposeoftheexperiment,andthey‐interceptvaluesofthebest‐fittingfunctionsinFigure15arepositive.Theexceptiontothispatternwasforthe16%incrementcondition,whichwasdetected100%ofthetimeinbothexperiments,althoughthisconditionwasratedasmoredifficultandlessurgentbytheunalertedsubjects.
16
4.DiscussionAsdescribedintheresultssection,theeffectiveintensity(asdefinedinEquation1)wasnotausefulpredictivemetricfortheperformanceofvisualsignalshavingdifferentdurations.Norwastherelativeilluminanceincrementpredictiveofperformance,asconsideredbyBulloughetal.(2012b).Theeffectiveintensityformulationisbasedontheconceptthattheintensityanddurationofalightsourcecanbetradedoffinordertomaintaindetectionperformance.AsmentionedbytheIALA(2008),theoriginaleffectiveintensityformulationdevelopedbyBlondelandRey(1912)isprimarilyapplicabletothresholddetectionofsignallightsvieweddirectly(on‐axis)fromlongdistances(andseenaspointsourcesoflight)underdarkadaptationconditions.Incomparison,thevisualsignalsinthepresentstudywereviewedunderhighlightlevelscommontoofficeandotherinteriorapplications,andunderindirectviewing,couldfillarelativelylargeportionofthefieldofviewandofteninthevisualperiphery.Undersuchconditions,theuseofEquation1initscurrentformmaynotbewarrantedforspecifyingindirectdetection.Twoconceptswereexploredtodevelopalternativeformulationsfortheeffectivenessofflashinglightswhenviewedindirectlyunderinteriorlightlevels.4.1.PartialTemporalSummationEquation1assumesalinear,proportionaltradeoffbetweeninstantaneousintensityandflashduration.BaumgardtandHillman(1961)found,forrelativelylarge,7.5o‐diameterluminousstimulipresented20ooffaxisunderdarkconditions,thatforshortdurationstheintensityanddurationcouldbetradedoffproportionallysothatthethresholdoccurredwhentheproductoftheintensityanddurationwasaconstantvalue.Forlongerflashdurations,theproductofthesquareorcuberootoftheintensityanddurationwasconstantforthesamelevelofdetection.InEquation1,thetermI(t)dtwasraisedtoanexponenthavingdifferentvalueslowerorgreaterthan1wasusedtoestimateapossibleindirecteffectivenessquantity.Noneofthevaluesresultedinrelationshipsbetweentheindirecteffectivenessquantityanddetectionperformanceorratedease/difficultyorurgency,thatappearedmuchdifferentfromthegraphsfortheexperimentaldatashowninChapter3(Results)ofthisreport.Thissuggeststhattheconceptofpartialtemporalsummationwasnotapplicabletothedatapresentedhere.4.2.IntegrationTimesAnotheraspectofEquation1thatpertainstothespecificexperimentalconditionsusedbyBlondelandRey(1912)intheirstudiesunderlyingtheeffectiveintensityformulationisthevalueoftheconstantaintheequation.Avalueforaof0.2swasempiricallydeterminedby
17
BlondelandRey(1912).Theconstantaisthoughttoberelatedtothetemporalintegrationtimeofthevisualsystemduringwhichintensityanddurationcouldbetradedoffforequivalentperformanceatdetectionthresholdunderdarkconditionswhentheprimaryphotoreceptorsusedforvisionarerods,whichhaverelativelylongintegrationtimesof0.1‐0.2s.FortheconditionsusedbyBlondelandRey(1912),theconditionsincludedpoint‐sourcesizesignals,viewedunderverydarkconditionsusingon‐axisvision,whenthesignalswerejustbarelyabletobedetected.Othervaluesforahavebeenfoundinvariousstudies,buttherehasbeenrelativelylittlesystematicinvestigationbetweentheoptimalvalueforaandspecificviewingconditionsdifferingfromthoseusedbyBlondelandRey(1912).OneexemplarystudywasconductedbySchmidt‐Clausen(1971),whoinvestigatedtheroleofbackgroundluminance,sizeofthesignallight,andeccentricityinthefieldofviewontheoptimalvalueofainEquation1.Schmidt‐Clausen(1971)foundthatlowervaluesofawerefoundforhigherbackgroundluminances,largersignallightsizes,andgreatereccentricities.ThisisconsistentwithdatafromBattersbyandSchuckman(1970)whoreportedthattemporalintegrationtimesforconephotoreceptors,whicharetheprimaryreceptorsthevisualsystemusesunderdaytimelightlevels,wereontheorderof0.01s.ModificationofthevalueofainthedenominatorofEquation1found,consistentwiththeshorterintegrationtimesofthevisualsystemunderhighlightlevels(BattersbyandSchuckman,1970),andwiththeexperimentaldatafromSchmidt‐Clausen(1971).ThedatafromExperiments1through6werecomparedwithdifferentvaluesofaanditwasfoundthatavalueforaof0.01sresultedinmanyofthecurvesfordifferentsignaldurationsbeingsuperimposedovereachother.Figure16showsthedetectiondataforExperiments1through6(dataforExperiments1and2werecombinedsincetherewasnodifferencebetweenthe1Hzand2Hzdataintheseexperiments),Figure17showstheease/difficultydata,andFigure18showstheurgencydatawhenEquation1ismodifiedusingavalueof0.01sforainthedenominator.Inspectionofthesefiguresshowsthatthecurvesarelocatedclosertooneanotherthanthecurvesinthepreviouschapterofthisreport(Results)andinmanycasesyieldveryconsistentpredictionsfortheflashesoflightwithdifferentdurations.Inaddition,thecurvesseemtoalsobeconsistentwiththedataforthexenonsourceinExperiments4and5inFigures16through18.ThisalsosuggeststhatanindirecteffectivenessquantitybasedonEquation1,butusingavalueof0.01sfora,maybeausefulpredictivemetricforperformance.
18
a.
Detection: Experiments 1‐2
0%
20%
40%
60%
80%
100%
1 10 100 1000a=0.01 s
Detection
1 ms
10 ms
25 ms
50 ms
100 ms
b.
Detection: Experiment 3
0%
20%
40%
60%
80%
100%
1 10 100 1000a=0.01 s
Detection
10 ms
25 ms
50 ms
100 ms
c.
Detection: Experiment 4
0%
20%
40%
60%
80%
100%
1 10 100 1000
a=0.01 s
Detection
10 ms
25 ms
50 ms
100 ms
Xenonstrobe
d.
Detection: Experiment 5
0%
20%
40%
60%
80%
100%
1 10 100 1000
a=0.01 s
Detection
10 ms
25 ms
50 ms
100 ms
Xenonstrobe
e.
Detection: Experiment 6
0%
20%
40%
60%
80%
100%
1 10 100 1000
a=0.01 s
Detection
10 ms
25 ms
50 ms
100 ms
Figure16.ExperimentaldetectiondataforExperiments1‐6plottedasafunctionof
quantitiesusingEquation1withavalueforaof0.01s.
19
a.
Difficulty: Experiments 1‐2
‐3
‐2
‐1
0
1
2
1 10 100 1000
a=0.01 s
Difficulty
1 ms
10 ms
25 ms
50 ms
100 ms
b.
Difficulty: Experiment 3
‐3
‐2
‐1
0
1
2
1 10 100 1000
a=0.01 s
Difficulty
10 ms
25 ms
50 ms
100 ms
c.
Difficulty: Experiment 4
‐3
‐2
‐1
0
1
2
1 10 100 1000a=0.01 s
Difficulty
10 ms
25 ms
50 ms
100 ms
Xenonstrobe
d.
Difficulty: Experiment 5
‐3
‐2
‐1
0
1
2
1 10 100 1000
a=0.01 s
Difficulty
10 ms
25 ms
50 ms
100 ms
Xenonstrobe
e.
Difficulty: Experiment 6
‐3
‐2
‐1
0
1
2
1 10 100 1000a=0.01 s
Difficulty
10 ms
25 ms
50 ms
100 ms
Figure17.Experimentalease/difficultydataforExperiments1‐6plottedasa
functionofquantitiesusingEquation1withavalueforaof0.01s.
20
a.
Urgency: Experiments 1‐2
‐1
0
1
2
3
1 10 100 1000
a=0.01 s
Urgency
10 ms
25 ms
50 ms
100 ms
b.
Urgency: Experiment 3
‐1
0
1
2
3
1 10 100 1000
a=0.01 s
Urgency
10 ms
25 ms
50 ms
100 ms
c.
Urgency: Experiment 4
‐1
0
1
2
3
1 10 100 1000
a=0.01 s
Urgency
10 ms
25 ms
50 ms
100 ms
Xenonstrobe
d.
Urgency: Experiment 5
‐1
0
1
2
3
1 10 100 1000a=0.01 s
Urgency
10 ms
25 ms
50 ms
100msXenonstrobe
e.
Urgency: Experiment 6
‐1
0
1
2
3
1 10 100 1000
a=0.01 s
Urgency
10 ms
25 ms
50 ms
100 ms
Figure18.ExperimentalurgencydataforExperiments1‐6plottedasafunctionof
quantitiesusingEquation1withavalueforaof0.01s.
21
5.ConclusionsThepresentexperimentalresultsandanalysesconductedusingvariationsontheformulationforeffectiveintensitysuggestthattheconventionaleffectiveintensityformulationinEquation1,usingaconstantvalueof0.2sinthedenominator,isnotasuitablemetricforpredictingtheperformanceofasignallightviewedindirectly.Noristherelativeinstantaneousilluminanceincreaseausefulmetric.Itappearsthatausefulpredictivemetric(herecalledanindirecteffectivenessquantity)canbedevelopedfromEquation1,butusingasmallervaluefortheconstanta.Settinga=0.01sallowsthedataforflashesoflighthavingdifferentdurationstobesuperimposedovereachother.Asaroughapproximation,andusingtheperformancewhensubjectswereawareofthenatureoftheexperimentbutwereperformingtheNVTandnotlookingatthefacingwall,theindirecteffectivenessquantityneedstobeapproximately500cdwhentheambientilluminanceis500lx,and250cdwhentheambientilluminanceis250lx,inordertoachieveadetectionpercentageof75%‐80%.[Inordertoachieveadetectionpercentageof90%,acriterionusedbyUL(1991),theindirecteffectivenessquantityshouldbe750cdforanambientilluminanceof500lx,and375cdforanambientilluminanceof250lx.]Thesevaluesfortheindirecteffectivenessquantities(foradetectionlevelof~75%‐80%)alsoresultinameanease/difficultyratingnearzero(theborderlinebetweeneasyanddifficulttodetect)andameanurgencyratingofaboutone(slightlyurgent).Thusitseemsthatachievingthesevaluesorhigherwithasignallightthatisviewedindirectlywouldbegintoensurethatthesignalswereeasytodetectandinterpretedasurgent.Ofcourse,responsesfortheunalerted,naïvesubjectsinthefinalexperimentwerelowerthanfortheprevioussixexperimentsandthismaybeaconsiderationforsettingperformancespecifications.Inaddition,theresponsedataforthenarrowbeamconditionswerequitepoorintermsofdetectionpercentagesandrateddifficultyandurgency.Basedonthepresentdataanemergencynotificationsignalintendedtobeviewedindirectlyshouldilluminatearelativelylargerareaoftheroomsurfaces(e.g.,producingabeamangleofatleast40o)inordertobereliablydetected.ThepresentstudyusedanLEDlightsourcewithaCCTof3200K,matchingtheambientilluminationinthetestlaboratory.ThexenonsourcetestedhasaCCTcloserto5000K.AlthoughtheresultsfromthepresentstudydonotsuggestthedifferenceinCCTbetweenthesesourcesmadealargedifferenceindetection,muchlargerchromaticitydifferencessuchastheuseofcoloredillumination,mightbeeasiertodetectthanthenominally"white"lightsourcesemployedinthepresentexperiments.
22
6.AcknowledgmentsThisstudywassponsoredbytheFireProtectionResearchFoundationoftheNationalFireProtectionAssociation,andperformedunderthedirectionofprojectmanagerAmandaKimball.HelpfulreviewandfeedbackwasprovidedbyRobertElliot,DanFinnegan,BruceFraser,LarryGrodsky,JackMcNamara,DaveNewhouse,WarrenOlsen,IsaacPapier,MarkPavlica,RodgerReiswig,LeeRichardson,RobertSchifiliti,andAndrewTrotta.TheauthorsalsoacknowledgethetechnicalcontributionstothisreportofDr.N.NarendranfromtheLightingResearchCenteratRensselaerPolytechnicInstitute.
23
7.References
BattersbyWS,SchuckmanH.1970.Thetimecourseoftemporalsummation.VisionResearch10(3):263‐73.BaumgardtE,HillmanB.1961.Durationandsizeasdeterminantsofperipheralretinalresponse.JournaloftheOpticalSocietyofAmerica51:340‐344.BlondelA,ReyJ.1912.Theperceptionoflightsofshortdurationattheirrangelimits.TransactionsoftheIlluminatingEngineeringSociety7(8):625‐662.BulloughJD,ZhuY,NarendranN.2012a.Characteristicsoflight‐emittingdiodesources:Relevanceforvisualsignaldetection.Suppression,DetectionandSignalingResearchandApplications:ATechnicalWorkingConference,Phoenix,AZ,March5‐8.Quincy,MA:FireProtectionResearchFoundation,NationalFireProtectionAssociation.BulloughJD,ZhuY.2012b.PerformanceObjectivesofLightSourcesUsedinEmergencyNotificationAppliances[reporttotheFireProtectionResearchFoundation].Quincy,MA:FireProtectionResearchFoundation.BulloughJD,SkinnerNP,TarantaRT.2013.Characterisingtheeffectiveintensityofmultiple‐pulseflashingsignallights.LightingResearchandTechnology45(3):377‐390.BulloughJD,SkinnerNP.2013.Conspicuityofflashesoflight:Interactionsbetweenintensityandduration.JournalofModernOptics[inpress].InternationalAssociationofLighthouseAuthorities.2008.MarineSignalLights,Part4:DeterminationandCalculationofEffectiveIntensity,IALARecommendationE‐200‐4.St.Germain‐en‐Laye,France:InternationalAssociationofLighthouseAuthorities.IlluminatingEngineeringSociety.1964.IESGuideforCalculatingtheEffectiveIntensityofFlashingSignalLights.NewYork,NY:IlluminatingEngineeringSociety.KellyDH.1961.Visualresponsestotime‐dependentstimuli:I.Amplitudesensitivitymeasurements.JournaloftheOpticalSocietyofAmerica51(4):422‐429.SavageK.2011.Flashpulsewidtheffectivenessinnotificationappliances.SUPDET2011Conference.Quincy,MA:FireProtectionResearchFoundation.Schmidt‐ClausenHJ.1971.Theinfluenceoftheangularsize,adaptationluminance,pulseshapeandlightcolurontheBlondel‐Reyconstanta.Theperceptionandapplicationofflashinglights95‐111.AdamHilger,London.
UnderwritersLaboratories.1991.Emergencysignalingdevicesforusebythehearingimpaired,UL1971.Northbrook,IL:UnderwritersLaboratories.
24
Appendix:ExperimentalData
Datafromallsevensetsofexperimentsaresummarizedinthissppendix.Foreachexperiment,themeandetectionpercentagesandthemeansandstandarderrorsofthemean(S.E.M.)forthesubjectiveratings(easeandurgency)arelisted.Experiment1:Ambientilluminance500lx;Beamangle40o;Duration1,10,100ms;Illuminanceincrement1%,2%,4%,8%;Frequency1Hz;Subjectslookingatwall
Mean detection Duration (ms)
1 10 100
Illuminance
increm
ent 1% 10% 0% 30%
2% 0% 20% 90%
4% 0% 90% 100%
8% 0% 100% 100%
Mean Ease Rating Duration (ms)
1 10 100
Illuminance
increm
ent 1% ‐2.9 ‐3.0 ‐2.7
2% ‐3.0 ‐2.7 ‐1.0
4% ‐3.0 ‐0.7 0.4
8% ‐3.0 0.5 1.2
S.E.M. of Ease
Rating
Duration (ms)
1 10 100
Illuminance
increm
ent 1% 0.10 0.00 0.15
2% 0.00 0.21 0.39
4% 0.00 0.54 0.45
8% 0.00 0.34 0.39
Experiment2:Ambientilluminance500lx;Beamangle40o;Duration10,25,50,100ms;Illuminanceincrement1%,2%,4%,8%;Frequency2Hz;Subjectslookingatwall
Mean detection Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0% 10% 20% 10%
2% 40% 70% 90% 70%
4% 100% 100% 100% 100%
8% 100% 100% 100% 100%
25
Mean Ease Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% ‐3.0 ‐2.9 ‐2.5 ‐2.9
2% ‐2.4 ‐1.4 ‐0.4 ‐1.7
4% ‐0.4 0.6 0.7 0.7
8% 0.8 1.4 1.5 1.5
S.E.M. of Ease
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0.00 0.10 0.40 0.10
2% 0.27 0.45 0.52 0.33
4% 0.48 0.31 0.42 0.37
8% 0.47 0.31 0.31 0.31
Mean Urgency
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% ‐1.0 ‐0.9 ‐0.6 ‐0.9
2% ‐0.5 0.1 0.4 0.0
4% 0.9 1.2 1.2 0.9
8% 1.4 1.8 1.7 1.6
S.E.M. of
Urgency Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0.00 0.10 0.31 0.10
2% 0.22 0.28 0.27 0.26
4% 0.28 0.29 0.25 0.23
8% 0.37 0.33 0.33 0.27
Experiment3:Ambientilluminance250lx;Beamangle40o;Duration10,25,50,100ms;Illuminanceincrement2%,4%,8%,16%;Frequency1Hz;Subjectslookingatwall
Mean detection Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 20% 90% 70% 70%
4% 80% 100% 100% 90%
8% 100% 100% 100% 100%
16% 100% 100% 100% 100%
26
Mean Ease Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐2.8 ‐1.5 ‐1.6 ‐1.6
4% ‐0.9 ‐0.2 ‐0.1 ‐0.5
8% 0.0 0.8 0.7 0.8
16% 1.1 1.6 1.7 1.7
S.E.M. of Ease
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.13 0.37 0.48 0.48
4% 0.48 0.36 0.38 0.50
8% 0.45 0.42 0.33 0.36
16% 0.38 0.31 0.21 0.21
Mean Urgency
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐0.8 0.1 ‐0.1 0.0
4% 0.2 0.6 0.8 0.7
8% 0.9 1.6 1.3 1.1
16% 1.3 2.2 2.3 2.2
S.E.M. of
Urgency Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.13 0.18 0.23 0.30
4% 0.29 0.22 0.25 0.30
8% 0.31 0.31 0.26 0.28
16% 0.26 0.36 0.33 0.25
Experiment4:Ambientilluminance500lx;Beamangle40o;Duration10,25,50,100ms;Illuminanceincrement1%,2%,4%,8%;Frequency1Hz;SubjectsperformingNVT
Mean detection Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0% 0% 0% 0%
2% 0% 20% 0% 10%
4% 10% 20% 40% 40%
8% 50% 70% 60% 70%
27
Mean Ease Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% ‐3.0 ‐3.0 ‐3.0 ‐3.0
2% ‐3.0 ‐2.7 ‐3.0 ‐2.9
4% ‐2.5 ‐2.8 ‐1.9 ‐2.0
8% ‐1.7 ‐0.9 ‐0.9 ‐0.9
S.E.M. of Ease
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0.00 0.00 0.00 0.00
2% 0.00 0.21 0.00 0.10
4% 0.40 0.13 0.53 0.47
8% 0.47 0.60 0.64 0.55
Mean Urgency
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% ‐1.0 ‐1.0 ‐1.0 ‐1.0
2% ‐1.0 ‐0.8 ‐1.0 ‐0.9
4% ‐0.7 ‐0.7 ‐0.2 ‐0.4
8% 0.0 0.4 0.5 0.6
S.E.M. of
Urgency Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 1% 0.00 0.00 0.00 0.00
2% 0.00 0.13 0.00 0.10
4% 0.21 0.21 0.33 0.27
8% 0.37 0.37 0.45 0.43
Experiment5:Ambientilluminance250lx;Beamangle40o;Duration10,25,50,100ms;Illuminanceincrement2%,4%,8%;16%;Frequency1Hz;SubjectsperformingNVT
Mean detection Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0% 0% 20% 0%
4% 30% 10% 60% 30%
8% 20% 60% 90% 60%
16% 80% 100% 100% 90%
28
Mean Ease Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐3.0 ‐3.0 ‐2.6 ‐3.0
4% ‐2.6 ‐2.8 ‐1.5 ‐2.3
8% ‐2.3 ‐1.1 0.2 ‐1.0
16% ‐0.8 0.9 1.4 0.8
S.E.M. of Ease
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.00 0.00 0.27 0.00
4% 0.22 0.20 0.45 0.37
8% 0.47 0.62 0.47 0.60
16% 0.47 0.35 0.16 0.51
Mean Urgency
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐1.0 ‐1.0 ‐0.7 ‐1.0
4% ‐0.6 ‐0.8 ‐0.1 ‐0.7
8% ‐0.7 0.1 1.1 0.3
16% 0.4 1.6 1.7 1.5
S.E.M. of
Urgency Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.00 0.00 0.21 0.00
4% 0.22 0.20 0.28 0.15
8% 0.21 0.35 0.31 0.45
16% 0.27 0.22 0.15 0.37
Experiment6:Ambientilluminance250lx;Beamangle6o;Duration10,25,50,100ms;Illuminanceincrement2%,4%,8%,16%;Frequency1Hz;SubjectsperformingNVT
Mean detection Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0% 0% 20% 0%
4% 0% 10% 0% 10%
8% 20% 0% 20% 10%
16% 30% 20% 50% 20%
29
Mean Ease Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐3.0 ‐3.0 ‐2.8 ‐3.0
4% ‐3.0 ‐2.9 ‐3.0 ‐2.9
8% ‐2.6 ‐3.0 ‐2.8 ‐2.8
16% ‐2.6 ‐2.4 ‐1.4 ‐2.5
S.E.M. of Ease
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.00 0.00 0.13 0.00
4% 0.00 0.10 0.00 0.10
8% 0.27 0.00 0.13 0.20
16% 0.22 0.43 0.58 0.40
Mean Urgency
Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% ‐1.0 ‐1.0 ‐0.8 ‐1.0
4% ‐1.0 ‐0.9 ‐1.0 ‐0.8
8% ‐0.7 ‐1.0 ‐0.8 ‐0.9
16% ‐0.6 ‐0.6 0.2 ‐0.7
S.E.M. of
Urgency Rating
Duration (ms)
10 25 50 100
Illuminance
increm
ent 2% 0.00 0.00 0.13 0.00
4% 0.00 0.10 0.00 0.20
8% 0.21 0.00 0.13 0.10
16% 0.22 0.31 0.42 0.21
Experiment7:Ambientilluminance250lx;Beamangle40o;Duration50ms;Illuminanceincrement4%,16%;Frequency1Hz;SubjectsperformingNVT
Duration=50ms Mean
detection
Illuminance
increm
ent
4% 0%
16% 100%
30
Duration=50 ms Mean Ease Rating
S.E.M. of Ease Rating
Illuminance
increm
ent
4% ‐3.0 0.0
16% 1.0 0.5
Duration=50 ms Mean Urgency Rating
S.E.M. of Urgency Rating
Illuminance
increm
ent
4% ‐1.0 0.0
16% 1.0 0.4