pasi - electron microscopy - chile 1 lyman - eds qual energy-dispersive x-ray spectrometry in the...

47
PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy- Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for Lehigh University semester courses and for the Lehigh Microscopy School

Upload: cornelius-sanders

Post on 27-Dec-2015

231 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

1Lyman - EDS Qual

Energy-Dispersive

X-ray Spectrometry

in the AEMCharles Lyman

Based on presentations developed for Lehigh University semester courses and for the Lehigh Microscopy School

Page 2: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

2Lyman - EDS Qual

Why Do EDS X-ray Analysis in TEM/STEM?

Spatial resolution» 2-20 nm (103 times better than SEM/EPMA)

Elemental detectability» 0.1 wt% - 1 wt%, depending on the specimen (~same as SEM/EDS)

Can use typical TEM specimens (t ~ 50-500 nm)» EELS requires specimens < 20-30 nm

Straightforward microanalysis» Qualitative analysis => Which element is present?» Quantitative analysis => How much of the element is present?» Easy x-ray mapping

Page 3: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

3Lyman - EDS Qual

Example of an X-ray Spectrum

2 Types of X-rays

» Characteristic x-rays– elemental identification– quantitative analysis

» Continuum x-rays– background radiation– must be subtracted for quantitative

analysis

Example of EDS x-ray spectrum

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 4: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

4Lyman - EDS Qual

Continuum X-rays

Interactions of beam electrons with nuclei of specimen atoms

Accelerating electric charge emits electromagnetic radiation

» Here the acceleration is a change in direction

The good» The shape of the continuum is a valuable

check on correct operation The not-so-good

» I bkg increases as ib increases» I bkg is proportional to Zmean of specimen» I max bkg rises as beam energy rises

Peak-to-background ratio » Ratio of Ichar / Ibkg sets limit on elemental

detectability

Continuum x-rays

Absorption of continuum

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 5: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

5Lyman - EDS Qual

Generation of Characteristic X-rays

Mechanism » Fast beam electron has enough energy

to excite all atoms in periodic table» Ionization of electron from the K-, L-, or

M-shell» X-ray is a product of de-excitation

Example» Vacancy in K-shell» Vacancy filled from L-shell» Emission of a K x-ray

(or a KLL Auger electron)

Important uses» Qualitative use x-ray energy to identify

elements» Quantitative use integrated peak

intensity to determine amounts of elements

Page 6: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

6Lyman - EDS Qual

Compute Energy of Sodium K Line

Energy levels EK and EL3 are in Bearden's "Tables of X-ray Wavelengths and X-ray Atomic Energy Levels" in older editions of the CRC Handbook of Chemistry and Physics

EK = 1072 eVEL3 = 31 eV

X-ray energy is the difference between two energy levels:

For sodium (Z=11):

If beam E > EK, then a K-electron may be ionized:

EKα1 = EK − EL III

EK a1

Na =1072 eV − 31 eV =1041 eV

K L M

For Na only see one peak since the K is only 26 eV from the K line

Beam electron

Beam electron loses EK

Page 7: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

7Lyman - EDS Qual

Families of Lines

Note: this is a simplified version of Goldstein Figure 6.9 showing only lines seen in EDS

If K-series excited, will also have L-series

Page 8: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

8Lyman - EDS Qual

Fluorescence Yield = fraction of ionization events producing

characteristic x-rays · the rest produce Auger e–

increases with Z» K typical values are:

– 0.03 for carbon (12) K-series @ 0.3 keV– 0.54 for germanium (32) K-series @ 9.9 keV– 0.96 for gold (79) K-series @ 67 keV

» X-ray production is inefficient for low Z lines (e.g., O, N, C) since mostly Augers produced

for each shell: K L M » X-ray production is inefficient for L-shell and M-

shell ionizations since » Land M always < 0.5:

L = 0.36 for Au (79)

M = 0.002 for Au (79)

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 9: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

9Lyman - EDS Qual

X-ray Absorption and Fluorescence

X-rays can be absorbed in the specimen and in parts of the detector

Certain x-rays fluoresce x-rays of other elements» X-rays of element A can excite x-rays from element B » Energy of A photon must be close to but above absorption edge

energy of element B» Example: Fe K (6.40 keV) can fluoresce the Cr K-series

(absorption edge at 5.99 keV)

I = Ioe−

μ

ρ

⎝ ⎜

⎠ ⎟ρt

Greater absorption when -- x-ray energy is just above absorber absorption edge-- path length t is large

Page 10: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

10Lyman - EDS Qual

EDS Dewar, FET, Crystal

LN dewar is most recognizable part

» To cool FET and crystal

Actual detector is at end of the tube

» Separated from microscope by x-ray window

Crystal and FET fitted as close to specimen as possible

» Limited by geometry inside specimen chamber

Schematic courtesy of Oxford Instruments

Page 11: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

11Lyman - EDS Qual

Electron-Hole Pair Creation

Absorption of x-ray energy excites electrons

» From filled valence band or states within energy gap

Energy to create an electron-hole pair

» = 3.86 eV @ 77K (value is temperature dependent)

Within the intrinsic region» Li compensates for impurity holes» Ideally # electrons = # holes» # electron-hole pairs is

proportional to energy of detected x-ray

Acceptor

Valence band

Conduction band

Donor

Energy gapE

nerg

y

Excited e-

Hole

For Cu K 8048 eV/3.8 eV = 2118 e-h pairs

(after drawing by J. H. Scott)

Page 12: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

12Lyman - EDS Qual

Details of Si(Li) Crystal

–1000 V bias

X-ray

Silicon inactive layer (p-type) ~100 nmGold

electrode20 nm

Active silicon(intrinsic)

3 mm

Ice?

WindowBe, BN,

diamond,polymer

0.1 m — 7 m

Anti-reflectiveAl coating 30 nm

(+)(–)

Holes Electrons

Gold electrode

(after drawing by J. H. Scott)

Page 13: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

13Lyman - EDS Qual

X-ray Pulses to Spectrum

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

slow amplifier

fast amplifier

charge staircase

analog-to-digital

converter

spectrum

energy binscollects e-h as charge

Page 14: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

14Lyman - EDS Qual

Slow EDS Pulse Processing

EDS can process only one photon at a time

» A second photon entering, while the first photon pulse is being processed, will be combined with the first photon

» Photons will be recorded as the sum of their energies

X-rays entering too close in time are thrown away to prevent recording photons at incorrect energies

Time used to measure photons that are thrown away is “dead time”

» Lower dead time -> fewer artifacts» Higher dead time -> more counts/sec into

spectrum

Processor extends the “live time” to compensate

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Counting is linear up to 3000 cps (20-30% dead time)

fast amplifier

slow amplifier

slower amplifier

Page 15: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

15Lyman - EDS Qual

Things for Operator to Check

Detector Performance» Energy resolution (stamped on detector)» Incomplete charge collection (low energy tails)» Detector window (thin window allows low-energy x-ray detection)» Detector contamination (ice and hydrocarbon)» Count rate linearity (counts vs. beam current)» Energy calibration (usually auto routine)» Maximum throughput (set pulse processor time constant to collect the

most x-rays in a given clock time with some decrease in energy resolution)

Topics in red explained in next few slides

Page 16: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

16Lyman - EDS Qual

Energy Resolution

Natural line width ~2.3 eV (Mn K » measured full width at half maximum (FWHM)

Peak width increases with statistical distribution of e-h pairs created and electronic noise:

Measured with 1000 cps at 5.9 keV» Mn K line

FWHM ≈ C2E + N 2( )

1/ 2

C = 2.35(Fε)1/ 2

E = x-ray energyN = electronic noiseF = Fano factor (~0.1 for Si)E = 3.8 eV/electron-hole pair

Mn K line

from Williams and Carter, Transmission Electron Microscopy, Springer, 1996.

Page 17: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

17Lyman - EDS Qual

X-ray Windows

Transmission curve for a “windowless” detector

» Note absorption in Si

Transmission curves for several commercially available windows

» Specific windows are better for certain elements

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 18: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

18Lyman - EDS Qual

Ice Build Up on Detector Surface

All detectors acquire an ice layer over time

» Windowless detector in UHV acquires ~ 3µm / year

Test specimens

» NiO thin film (Ni L / Ni K

» Cr thin film (Cr L / Cr K Check L-to-K intensity ratio

for Ni or Cr

» L/K will decrease with time as ice builds up

» Warm detector to restore (see manufacturer) Courtesy of J.R. Michael

After operating 1 year

Immediately after warmup

Windowless detector

Page 19: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

19Lyman - EDS Qual

Spectrometer Calibration

Calibrate spectrum using two known peaks, one high E and one low E

» NiO test specimen (commercial) – Ni K (high energy line) at 7.478 keV – Ni L (low energy line) at 0.852 keV

» Cu specimen– Cu K (high energy line) at 8.046 keV– Cu L (low energy line) at 0.930 keV

Calibration is OK if peaks are within 10 eV of the correct value

Calibration is important for all EDS software functions

0.930 keV 8.046 keV

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 20: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

20Lyman - EDS Qual

Artifacts in EDS Spectra

Si "escape peaks”» Si Ka escapes the detector» Carrying 1.74 keV» Small peak ~ 1% of parent» Independent of count rate

Sum peaks» Two photons of same energy enter detector simultaneously» Count of twice the energy» Only for high count rates

Si internal fluorescence peak» Photon generated in dead layer» Detected in active region

from Williams and Carter, Transmission Electron Microscopy, Springer, 1996

Page 21: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

21Lyman - EDS Qual

Expand Vertically to See EDS Artifacts

Escape peaks

Si internal fluor.

Sum peaks

System peaks

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.

Page 22: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

22Lyman - EDS Qual

EDS-TEM Interface

We want x-rays to come from just under the electron probe, BUT…

TEM stage area is a harsh environment» Spurious x-rays, generated from high energy x-rays originating from

the microscope illumination system bathe entire specimen» High-energy electrons scattered by specimen generate x-rays» Characteristic and continuum x-rays generated by the beam

electrons can reach all parts of stage area causing fluorescence

Detector can't tell if an x-ray came from analysis region or from elsewhere

Usually fixed by manufacturers of EDS and TEM

Page 23: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

23Lyman - EDS Qual

The Physical Setup

Want large collection angle, » Need to collect as many counts

as possible Want large take-off angle,

» But reduced as is increased Compromise by maxmizing

with ~ 20˚ at 0˚ tilt angle» can always increase by tilting

specimen toward detector -- but this increases specimen interaction with continuum from specimen

from Williams and Carter, Transmission Electron Microscopy, Springer, 1996

Page 24: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

24Lyman - EDS Qual

Orientation of Detector to Specimen

Detector should have clear view of incident beam hitting specimen

» specimen tilting eucentric» specimen at 0˚ tilt

Identify direction to detector within the image

Analyze side of hole "opposite the detector”

Keep detector shutter closed until ready to do analysis

EDSdetector

Rim

Thinned

Detector

Top view of disc

Edge of hole furthest from detector

from Williams and Carter, Transmission Electron Microscopy, Springer, 1996

Page 25: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

25Lyman - EDS Qual

Spurious X-rays in the Microscope

Pre-Specimen Effects» spurious x-rays => hole count due to column x-rays and stray

electrons » spurious x-rays => poor beam shape from large C2 aperture

Post-Specimen Scatter» system x-rays => elements in specimen stage, cold finger,

apertures, etc.» spurious x-rays => excited by electrons and x-rays generated in

specimen Coherent Bremsstrahlung

» extra peaks from specimen effects on beam-generated continuous radiation

The analyst must understand these effects to achieve acceptable qualitative and quantitative results

Page 26: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

26Lyman - EDS Qual

Test for Spurious X-rays Generated in TEM

Detector for x-rays from illumination system

» thick, high-Z metal acts as “hard x-ray sensor"

Uniform NiO thin film used to normalize the spurious "in hole" counts, thus

» NiO film on Mo grid*

% hole count = I on NiO film

MoK

I on NiO filmNiK

x100%

( Note: " on NiO film count" and

" in hole count" are similar)

Mo grid bar

NiO film on C Hole

Electron beam

Hard x-ray from illumination system

Spurious (bad) Mo K-series

Beam-generatedNi K x-ray (good)

* see Egerton and Cheng, Ultramicroscopy 55 (1994) 43-54

- “on film” results ~ “in hole” results- Usually take inverse “hole count” ratio

Page 27: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

27Lyman - EDS Qual

Spectrum from NiO/Mo

Spurious x-rays» Inverse hole count

(Ni K Mo K» Want high inverse

hole count Fiori P/B ratio

» Ni K/B(10 eV)» Increases with kV» Want high to

improve element detectability

Measure in center of grid square on NiO/Mo specimen

Minimize spurious Mo x-rays by using thick C2 aperture

Maximize P/B ratio for Ni K

Egerton and Cheng, Ultramicroscopy 55 (1994) 43-54

Page 28: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

28Lyman - EDS Qual

Figures of Merit for an AEM

Fiori PBR = full width of Ni K divided by 10 eV of background (Ni K / Mo K is inverse hole count

Better

Less good

Obviously, we want to use the highest kV

from Williams and Carter, Transmission Electron Microscopy, Springer, 1996

Page 29: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

29Lyman - EDS Qual

Beam Shape and X-ray Analysis

Calculated probes (from Mory, 1985)

Effect on x-ray maps (from Michael, 1990)

Properly limited Spherically aberrated

C2 aperture too largecorrect C2 aperture size

“witch’s hat” beam tail excites x-rays

Page 30: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

30Lyman - EDS Qual

Qualitative Analysis

Collect as many x-ray counts as possible » Use large beam current regardless of poor spatial resolution with

large beam» Analyze thicker foil region, except if light elements x-rays might be

absorbed

Scan over large area of single phase => avoid spot mode

Use more than one peak to confirm each element

Which elements are present?

Page 31: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

31Lyman - EDS Qual

Qualitative Analysis Setup 1

» Use thin foils, flakes, or films rather than self-supporting disks to reduce spurious x-rays (not always possible)

» Orient specimen so that EDS detector is on the side of the specimen hole opposite where you take your analysis

» Collect x-rays from a large area of a single phase

» Choose thicker area of specimen to collect more counts

» Tilt away from strong diffracting conditions– (no strong bend contours)

» Operate as close to 0˚ tilt as possible (say, 5˚ tilt toward det.)

Specimen

Page 32: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

32Lyman - EDS Qual

Qualitative Analysis Setup 2

Microscope Column» Use highest kV of microscope» Use clean, top-hat Pt aperture in C2 to minimize “hole count”

effect» Minimize beam tails

– (C2 aperture or VOA should properly limit beam angle)

» Use ~ 1 nA probe current to maximize count rate– This may enlarge the electron beam (analyze smaller regions

later)

» Remove the objective aperture

Page 33: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

33Lyman - EDS Qual

Qualitative Analysis Setup 3

X-ray Spectrometer» Ensure that detector is cranked into position» Keep detector shutter closed until you are ready to analyze

» Use widest energy range available (0-20 keV is normal)– 0–40 keV for Si(Li) detector– 0–80 keV for intrinsic Ge detector

» Choose short detector time constant (for maximum countrate)» Count for a long time – 100-500 live sec

Page 34: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

34Lyman - EDS Qual

Peak Identification

Start with a large, well-separated, high-energy peak» Try the K-family» Try the L-family» Try the M-family

– Remember -- these families are related

Check for EDS artifacts Repeat for the next largest peak

Important:» Use more than one peak for identification» If peak too small to "see", collect more counts or forget about

identifying that peak; peak should be greater than 3B1/2

Page 35: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

35Lyman - EDS Qual

Chart of X-ray Energies (0-20 keV)

M-series

L-series

K-series

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003

Page 36: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

36Lyman - EDS Qual

Chart of X-ray Energies (0-5 keV)

M-series

L-series

K-series

from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003

Page 37: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

37Lyman - EDS Qual

Know X-ray Family Fingerprints

Page 38: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

38Lyman - EDS Qual

Some Peaks will Look Similar

At low energies each series collapses to a single line

From 1 keV to 3 keV, the K, L, or M lines all look similar

At 2.0 keV: Z = 15 (P) K @ 2.013 keV Z = 40 (Zr) L @ 2.042 Z = 77 (Ir) M @ 1.977

M-series

L-series

K-seriesZ

after Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003

Page 39: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

39Lyman - EDS Qual

Unknown #1

Energy (keV)

Page 40: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

40Lyman - EDS Qual

Data Analysis for Unknown #1

26 6.4 Fe(26) K

7.0 Fe(26) K

4.5

4.9

Page 41: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

41Lyman - EDS Qual

Unknown #1

from xray.optics.rochester.edu/.../spr04/pavel/

Page 42: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

42Lyman - EDS Qual

Unknown #2

after www.pentrace.com/ nib030601003.html

Page 43: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

43Lyman - EDS Qual

Analysis of Unknown #2

Line Energy(keV) Int, Sh Candidate 1 Candidate 21 0.2 w C(6) K (0.25)

2 0.8 vw Ru(44) L escape (0.86) Ni(28) La (0.8)

3 1.4 w, sym W(74) M (1.4)

4 1.8 s, sym W(74) M (1.8) No possible K line

5 2.6 m, asym Ru(44) L (2.6)

6 6.4 w, sym Fe(26) K (6.4)

7 7.1 vw Fe(26) K (7.1)

8 7.4 m, sym W(74) Ll (7.4) Ni(28) Ka (7.5)

9 8.0 vw Cu(29) K (8.0) Ni(28) Kb (8.3)

10 8.3 m, sym W(74) L (8.4) No possible K line

11 9.6 m, overlap W(74) L1 (9.7)

12 10.0 m, overlap W(74) L2 (10.0)

13 11.3 w, sym W(74) L1 (11.3)

14 11.6 vw W(74) L3 (11.6)

Start

Page 44: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

44Lyman - EDS Qual

Qualitative Analysis

W M

W L

W L1

W L1

W LlW M Fe K

Cu

Ru L

C

Elements present major: W, Ru trace: Fe, Cu, C

Page 45: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

45Lyman - EDS Qual

Automatic Qualitative Analysis?

1. Are suggested elements reasonable? Tc and Pm are unusual, Cl and S are not

2. Do not use peak energy alone to identifyLines of other elements may have the same energy

3. Consider logic of x-ray excitationAll lines of an element are excited in TEM/STEM (100-300 kV)If L-series indicated, K-series must be presentIf M-series indicated, L-series must be present

Check the results of every automatic qualitative analysis

Page 46: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

46Lyman - EDS Qual

Automatic Qualitative Analysis Blunders

Identification by peak energy alone without considering x-ray families or peak shape

Identification without considering other lines of same element

from Newbury, Microsc. Microanal. 11 (2005) 545-561

Page 47: PASI - Electron Microscopy - Chile 1 Lyman - EDS Qual Energy-Dispersive X-ray Spectrometry in the AEM Charles Lyman Based on presentations developed for

PASI - Electron Microscopy - Chile

47Lyman - EDS Qual

Summary

EDS in the TEM has more pitfalls than in SEM» Use the highest kV available» Understand the effects of:

– detector-specimen geometry– spurious x-rays from the illumination system– post-specimen scatter– beam shape and spatial resolution => the “witch’s hat”

Identify every peak in the spectrum» Even artifact peaks» Forget peaks of intensity < 3 x (background)1/2

Collect as many counts as possible» Use large enough beam size to obtain about 1 nA current» Qualitative analysis use:

– use long counting times or – thicker electron-transparent regions with a short pulse processor time

constant, if appropriate» Assume data might be used for later quantitative analysis (determine

t if possible)