performance of uv-vis spectrophotometers

Upload: huong-zam

Post on 04-Jun-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    1/6

    1

    Performance of UV-Vis spectrophotometers

    Introduction:

    UV-Visible (UV-Vis) spectrophotometry is commonly used in analytical laboratories forqualitative and quantitative analyses. The popularity of the UV-Vis technique stems from its ease

    of use and the speed of the analysis. The absorbance data from the UV-Vis measurement can berelated to the concentration of the sample by Beers Law. UV-Vis methods are so simple that the

    fundamental performance of the spectrophotometers required to generate reliable data issometimes neglected by the analysts. The various factors which affect the reliability of the UV-

    Vis measurement will be discussed in this article.

    The performance verification tests required by major Pharmacopoiaes for UV-Visspectrophotometers are listed in Table 1. The required performance tests include wavelength

    accuracy, stray light, resolution and photometric accuracy. There are other factors such as noise,baseline flatness and stability that will affect the performance of the spectrophotometers as well.

    Table 1: Performance Tests for UV-Vis Spectrophotometers

    Performance Attributes:

    1. Wavelength AccuracyWavelength accuracy is defined as the deviation of the wavelength reading at an absorption bandor emission band from the known wavelength of the band. The wavelength deviation can cause

    errors in the qualitative and quantitative results of the UV-Vis measurement. It is quite obviousthat if the spectrophotometer is not able to maintain an accurate wavelength scale, the UV profile

    of the sample measured by the instrument will be inaccurate. The true maxof the analyte cannot

    be characterized accurately.

    Pharmcopoeia

    Performance Test Standard USP 24 BP 1998 EP 3 Ed. JP (XII)

    Wavelength accuracy Deuterium lamp X X X X

    Mercury vapour lamp X X X X

    Holmium oxide glass filter X - - -

    Didymium glass filter X - - -

    Holmium oxide in HClO4 - X X -

    Stray Light Potassium chloride solution - X X -

    Resolution Toluene in hexane solution - X X -

    Photometric accuracy Neutral density glass filters X - - -

    Metal on quatz filter X - - -

    Potassium dichromate solution X X X X

    Noise --------- - - - -

    Baseline flatness --------- - - - -

    Stability --------- - - - -

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    2/6

    2

    In addition to the qualitative problem, wavelength deviation affects the quantition in two ways:

    the accuracy and sensitivity. Typically, most UV-Vis assays will specify that the measurements

    are to be taken at the max and the absorption profile of the analyte is fairly broad. A slight

    deviation in wavelength will have very little effect on the absorbance value (Figure 1). However,

    if the selected wavelength is not close to the max of the analyte of interest, then a small deviation

    in wavelength will cause a big change in the absorbance value of the measurement. Significantwavelength deviation will also steer the measurement away from the optimal wavelength at the

    max which effectively reduces the extinction coefficient and hence the sensitivity of the

    measurement.

    Figure 1: Effect of wavelength accuracy on UV-Vis measurements

    There are many standards that can be used to verify the wavelength accuracy of aspectrophotometer. The advantages and disadvantages of various wavelength verification

    standards are summarized in Table 2.

    Table 2: Wavelength standards comparison

    Wavelength Standard Advantage Disadvantage

    Emission lines from deuterium

    lamp

    Sharp spectral lines

    Present in the UV light source

    Limited to visible wavelengths

    Emission lines from mercuryvapour lamp

    Sharp spectral linesCover both UV and visible

    regions

    Not commonly built into theinstrument

    Holmium oxide and

    Didymium filters

    Easy to use

    NIST SRM available

    Independent of the resolutionpower of the instrument

    Limited to visible wavelengths

    Broad spectral lines

    Re-certification required onceevery two years

    4% Holmium oxide in 10%

    perchloric acid

    Easy to use

    NIST SRM availableCover both UV and visible

    regionsRe-certification required every

    eight years

    Dependency on the resolution

    power of the instrumentCorrosive solution

    Wavelength Accuracy

    0.0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    200 250 300 350 400Wavelength (nm)

    Absorban

    ce

    Error

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    3/6

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    4/6

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    5/6

    5

    In general, insufficient resolution leads to a decrease in extinction coefficient across thewavelength axis and therefore inaccurate quantitation. The sensitivity of the measurement is also

    compromised. From a qualitative point of view, the fine features in the spectrum may be lost. Asolution mixture of 0.02 % v/v toluene in hexane (UV grade) is used to test the resolution power

    of the spectrophotometer. The ratio of the absorbance at max (269 nm) and absorbance at min(266 nm) should be greater than 1.5.

    4. NoiseNoises in the UV-Vis measurement are mainly originating from the light source and electroniccomponents. Noise in the measurement affects the accuracy at both ends of the absorbance scale.

    Photon noise from the light source affects the accuracy of the measurements at low absorbance.Electronic noise from electronic components affects the accuracy of the measurements at high

    absorbance. High noise level will reduce the limit of detection and render the instrument less

    sensitive.

    5. Baseline FlatnessMost UV-Vis spectrophotometers have dual light sources. A deuterium lamp is used for the UV

    range and a tungsten lamp is used for the visible range. The intensity of the radiation comingfrom the light sources is not constant over the whole UV-Visible range. The response of the

    detector also varies over the spectral range. A flat baseline demonstrates the ability of theinstrument to normalize the output of the lamp and detector responses.

    6. StabilityVariations in lamp intensity and electronic outputs between the measurements of I oand I result

    in instrument drifts. These changes can lead to error in the value of the measurements especiallyover a long period of time. The resulting errors in the measurements may be positive or negative.

    The stability test checks the ability of the instrument to maintain a steady state over time so thatthe effect of the drift on the accuracy of the measurements is insignificant.

    7. Photometric Accuracy

    Photometric accuracy is determined by the difference between the measured absorbance and the

    established standard value. Most quantitation applications using UV-Vis would involve themeasurement of the standards and samples of comparable concentrations in rapid succession on

    the same instrument. As long as the photometric measurements are reproducible and the responseis linear over a defined range, the absolute photometric accuracy may not be critical. However,

    photometric accuracy is critical for measurement which comparing the extinction coefficientsbetween different instruments. Photometric inaccuracy will lead to errors in quantitation.

  • 8/13/2019 Performance of UV-Vis Spectrophotometers

    6/6