pet design: simulation studies using geant4 and gate - status report - martin göttlich desy

23
PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

Upload: serena-lownes

Post on 15-Dec-2015

220 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

PET Design: Simulation Studies using GEANT4 and GATE

- Status Report -

Martin Göttlich

DESY

Page 2: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

2

Outline

• Measurements

• Single crystal studies (GEANT4)

• PET scanner (GATE)

Page 3: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

3

Experimental Setup

• MPPC and crystal well aligned, optical photons well contained, high reproducibility• Reaching energy resolutions of 10-11 % FWHM (LSO and LFS)

3x3 mm2

3600 pixels

MPPC

LFSReflective Paint

Reflective Foil 3x3x15 mm3

Source Na-22

Page 4: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

4

Energy Resolution

Poisson stat. 8% for LSO/LFS negligible

• MPPC (3x3 mm, 3600 pixels) coupled to LFS (3x3x15 mm)

• Good separation between compton continuum and photo-electric peak.

• Mean at 2000 pixels.

• Saturation not an issue: signal width ~120 ns, peak amplitude ~400 p.e.

• Linearity checked with different sources

blue sensitivity

Page 5: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

5

Linearity

Na-22

Cs-137

Cs-137

(back scatter peak)

Linearity confirmed in the energy region of interest.

Page 6: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

6

Simulation Studies

Single Crystal + MPPC

• Detailed studies to optimize single crystal dimensions

• GEANT4 simulation• Good description of the

measurements in terms of energy resolution and photon yield.

PET System

• Design of a full PET system

• Spatial resolution (image reconstruction using FBP)

• Optimize crystal dimensions (DOI)

• GATE (GEANT4 Application for Tomographic Emission) toolkit based on GEANT4

• First results very promising.

Page 7: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

7

Single Crystal + MPPCGEANT4 Simulation

Important parameters:

• LSO/LFS: scintillation yield (23k), refraction index (1.82), emission spectrum, absorption length (50m)

• Optical grease BC-630 (10µm): refraction index (1.5)

• MPPC: quantum efficiency

• Teflon wrapping: reflectivity (0.90), refraction index (1.34)

Compare to measurements:

• number of pixels

• signal width

• signal shape

1.5 cm

3x3mm2

Compton-Effect

Photo-Effect

(LY: 100 photons/MeV)

30 reflections: P(absorbed)=45% (98% reflectivity)

crystal

MPPC

Page 8: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

8

Scintillation Yield and Photon Detection Efficiency

Incorporated published distributions.

PDE Hamamatsu:

Page 9: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

9

A Single EventNumber of photons produced : 10940

Absorbed in the crystal : 9

Absorbed at boundary : 9013

Detected : 1911

Leakage (optical grease) : 7

z of absorptions at boundary:

Crystal length: 1.5 cm

Reflectivity (Teflon): 90 %

MPPC

interaction

z of reflections at boundary:

optical grease

Page 10: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

10

Efficiency

~ 50 %

Interaction in crystal? Photo- or Compton-Effect?

5 cm > total length > 3 cm

Page 11: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

11

Energy Resolution

R = 95%

R = 90%

• Energy resolution is slightly decreasing for longer crystals due to absorption in the Teflon wrapping.

• The larger the reflectivity the smaller the effect.

Page 12: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

12

Time ResolutionHit-time distribution for a single event:

Triggering at low threshold, i.e. the very first photons which arrive:

Crystal length has only a small influence on the timing.

t [ns]

Hits t(0.1/e)

FWHM

Page 13: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

13

Measurements vs. Simulation

Reasonable agreement with data.

MC

R = 90%Tuning of the reflectivity

MC

Page 14: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

14

Crystal Length and DOI

Depth of interaction (DOI) information improves resolution considerably.

1 layer 2 layers

Gamma rayLOR

Simulation of a full PET scanner

Page 15: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

15

PET System - GATE• Simulation of a full PET system to

optimize camera design (high granularity, DOI, TOF).

• Image reconstruction using Filtered Back Projection (STIR)

• Determination of spatial resolution in compliance with NEMA NU 2-2007 standard*.

• Dead material, movement, noise, dead time, dose, phantom, scan time, efficiency

• Simulation of a simple PET system: 2x387 crystals (LSO, 3x3x15 mm3)

* National Electrical Manufacturers Association

Page 16: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

16

PET System – First Results

Tune energy and time resolution of single crystal+MPPC element according to measurements.

Crystal+MPPC

FWHM 11%

FWHM 600 pscoincident events

Sinogram

point source x=2cm

next steps: granularity, FBP (STIR)

Page 17: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

17

Conclusions and Outlook

• According to simulations no strong dependence of the energy and time resolution on the crystal length.

• DOI studies with simulation of full PET scanner.

• Two systems with two layers of crystals of different length. Total length 3.5 to 4.5 cm, First layer 1 to 1.5 cm.

Page 18: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

18

BACKUP FOILS

Page 19: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

19

Design Criteria

Camera design (matrix of crystals, each crystal read out individually by an

MPPC) guided by the following criteria:

• spatial resolution (DOI: two layers of crystals)

• time resolution

• efficiency

• exposure to radiation, scan-time

• costs

Prototype:

• Two matrices of 6x6 crystal double-layers

• Study uniformity of channels, multi-channel r/o, calibration

• Image reconstruction of a point source

LFS 3x3x15 mm3

Page 20: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

20

Sinogram

SinogramIntegration along

all LORs at fixed

Page 21: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

21

Sinogram - Example

Each point in the sinogram corresponds to a LOR.

Page 22: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

22

Filtered Back Projection(Analytic 2D image reconstruction)

Imaging system: p = Hf + n

p: observation, H: system model, f: image, n: error

Solution: unfolding, inverse problem ill-posed, regularization

2D central-section theorem:

(Radon transform closely related to Fourier transform. )

The Fourier transform of a 1D projection is equivalent to a section at the same angle through the center of the 2D Fourier transform of the object.

Ramp-filter:

Page 23: PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY

23

A Commercial Example

GEMINI TF PET/CT (Philips)

• diameter of 90 cm, axial field of view of 18 cm

• 4 x 4 x 22 mm3 LYSO (lutetium-yttrium oxyorthosilicate) crystals

• coincidence timing resolution: 585 ps

• resolution near the center: 4.8 mm (FWHM)

• single high-quality image: 1-2 min

• whole-body scan: 10-30 min (depending on patient size)

• available for $1.85M