W HCAL simulations and test-beam plans

Peter Speckmayer, LCWS2010, Beijing

3/27/2010

Page 2: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 2

OutlineCalorimetry (HCAL) for higher collision

energies

Test-beamrequirementsconstraints

Test-beamstatusoutlook

3/27/2010

Page 3: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 3

Calorimetryfor a high energy future linear collider detector

3/27/2010

Page 4: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 4

Considerations for HCAL depth and materialshower leakage worsens energy resolution

to reduce leakage:deeper calorimeterdenser calorimeter (more interaction lengths)

depth limited by feasible coil size:larger coil with smaller B-fieldlarger B-field with smaller coil

depth limited by tracker size:larger tracker better p-resolution

IP beam

COIL

HCAL

Tracker

ECAL

Vertex

Barrel Endcaptail-catcher

3/27/2010

Page 5: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 5

HCAL absorber material Which material for the absorber?

steel, tungsten, ... ?

Tungsten expensive more contained showers (compared to Fe) with the same HCAL geometrical

depth less leakage smaller shower diameter better separation of showers (probably good for

particle flow) different ratios (Fe/W) of X0 and λ electromagnetic parts of showers get more

“compressed” than hadronic parts compared to Fe

final goal good energy resolution with whole detector (PFA)

3/27/2010

λ (cm)

X0 (cm)

RM (cm)

Fe 16.77 1.757 1.719

W 9.946 0.3504

0.9327

Fe/W 1.686 5.014 1.843

Page 6: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 6

Energy resolution: W/Fe

shorter HCAL more leakage worse resolution

numbers denote HCAL length inunits of interaction lengths

~40 λ

5 mm scintillator, 2.5 mm G10

- energy reconstruction with neural network(information from fine granularity of calorimeter not used traditional approach)-variables describe shower shape and size and energy-train NN with pion energy-QGSP_BERT_HP used

steel

tungsten

3/27/2010

approx. HCAL depth in CLIC detector

Page 7: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 7

Energy resolution: W/Fe

shorter HCAL more leakage worse resolution

numbers denote HCAL length inunits of interaction lengths

~40 λ

5 mm scintillator, 2.5 mm G10

- energy reconstruction with neural network(information from fine granularity of calorimeter not used traditional approach)-variables describe shower shape and size and energy-train NN with pion energy-QGSP_BERT_HP used

steel

tungsten

3/27/2010

approx. HCAL depth in CLIC detector

trade off intrinsic resolution against leakage

Page 8: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

Effect of physics list on predicted resolution

P. Speckmayer, LCWS2010, Beijing 8

5 mm scintillator, 2.5 mm G10

~10-15% improvement (at 40 GeV)

why: n are captured farther away from shower core (2γ) “halo” produced which reduces reconstruction performance. removing halo (with HP n tracking)

QGSP_BERT_HP less energy deposited by ionization than QGSP_BERT, but ... Improved resolution!

considerable effectbut: perfect readout assumed, no cut-off in time

3/27/2010

Page 9: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

Experimental validation of simulations with a test-beam Validate detailed picture of showers in W

in space ... ... and time validated simulations

trust more in detector simulations trust more in detector optimizations trust more in PFA

Time structure of signal broadened by n-content time stamping (important in CLIC environment)

used to separate signal/background on a time basis (slow) n-content smears out energy deposits in calorimeters know time-structure of n-content to set requirements for time stamping

dependent on active material (e.g. scintillator, gas) measurements necessary

validate physics lists (QGSP_BERT, QGSP_BERT_HP, etc.)

P. Speckmayer, LCWS2010, Beijing 93/27/2010

Page 10: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 10

W HCAL testbeamDesign

3/27/2010

Page 11: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 11

Simulations of test-beamSeveral set-ups simulated

(by C.Grefe, J.Blaha, A.Lucaci-Timoce)

energies: 3 to 200 GeV test-beam initially with E smaller than ~10 GeV ... but preserve option to go to higher energies later on

absorber materials: W, W alloy, Fe (for comparison)

active materials: Scintillator and gas

3/27/2010

Page 12: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 12

Longitudinal shower profiles10 mm WMix + µMegas (two 2 mm Fe covers)

10 mm W + 2 mm air + 5 mm scint, 38 layers AHCAL

Longitudinal containment of showers

95%

J. Blaha

J. BlahaA. Lucaci Timoce

C. Grefe

12 mm tungsten + 5 mm Scint + 2.5 G10

3/27/2010

Page 13: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 13

Longitudinal shower profiles10 mm WMix + µMegas (two 2 mm Fe covers)

10 mm W + 2 mm air + 5 mm scint, 38 layers AHCAL

Longitudinal containment of showers

95%

J. Blaha

J. BlahaA. Lucaci Timoce

C. Grefe

12 mm tungsten + 5 mm Scint + 2.5 G10

3/27/2010

• different simulations agree• active layers don’t change the general picture

Page 14: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 14

Longitudinal shower profiles10 mm WMix + µMegas (two 2 mm Fe covers)

10 mm W + 2 mm air + 5 mm scint, 38 layers AHCAL

Longitudinal containment of showers

95%

J. Blaha

J. BlahaA. Lucaci Timoce

C. Grefe

12 mm tungsten + 5 mm Scint + 2.5 G1095% of shower energy deposited in ~40 layers

3/27/2010

Page 15: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 15

Lateral shower size

10 mm WMix + µMegas (two 2 mm Fe covers)

10 mm W + 2 mm air + 5 mm scint, 38 layers AHCAL

95%

J. Blaha

A. Lucaci Timoce

C. GrefeC. Grefe

3/27/2010

95%95%

Page 16: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 16

Lateral shower size

10 mm WMix + µMegas (two 2 mm Fe covers)

10 mm W + 2 mm air + 5 mm scint, 38 layers AHCAL

95%

J. Blaha

A. Lucaci Timoce

C. GrefeC. Grefe

3/27/2010

95%95%

95% energy contained in ~40 cm

Page 17: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 17

Conclusions of study results

#layers: ~40~4.8 λcost constraints initial configuration with 20 layers

Radius: take into account:

limited length absorber material energy containment, fraction of fully contained events

(>95%) existing active modules

safe choice ~40 cm3/27/2010

Page 18: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 18

W HCAL testbeamPreparation

3/27/2010

Page 19: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 19

Tungsten platesStart with 20 layers

Market servey on tungsten supply 3 positive answers received

Invitation to tender (closes 12.04.2010)

Place order by 30.04.2010

Receive 4 batches of 25 squares and 20 triangles in June & July

Start to equip absorber plates from 15 June onwards

270

100 x

80 x

Tungsten plates

3/27/2010

Page 20: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 20

Design of Frames

270

Tungsten plates

3/27/2010

Page 21: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 21

Design of Frames

270

Tungsten plates

3/27/2010

•place order end of april (for 20 layers)

Page 22: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 22

Active layers: CALICE AHCAL modules

90 cm30 cm

active material: Scintillator read out: Silicon photo multipliers

3/27/2010

Page 23: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 23

Beam1.3 above floor

Beam Counter Delay wire chamber Area 10cm * 10cmexisting

Cherenkov (CO2)existingto separate e- from π-

1Gev 50% e- + 50% π-

5Gev 10% e- + 90% π-

ScintillatorsFingers 1cm *10 cmGeneral 10cm * 10cmVeto 25cm * 25 cm (with hole of 4cm)Veto behind < 50 cm * 50cm

Calorimeter Prototype

TelescopeGEM chamber σ ≈ 1mm

T9 Test Beam Setup

3/27/2010

Telescope fixed to Calo(have geomerical correlation)

Page 24: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 24

Test-beam@CERN

PSNovember 2010

20 absorber layers to begin with40 layers planned for a second stage

active layers: μMegas, Scintillators, RPC, GEM

3/27/2010

Page 25: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 25

ConclusionsTungsten is viable option for HCAL at CLIC

energiesValidation of simulations is necessaryW HCAL test-beam preparations are

proceedingW HCAL test-beam:

muons from september onwards2 weeks in November 2010 beam-time

Thank you!3/27/2010

Page 26: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 26

Backup

3/27/2010

Page 27: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 27

Time Development

• Most of energy deposited early

• Clear difference between steel and tungsten setups

C. Grefe

3/27/2010

Page 28: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 28

Energy deposit in first 20 layersmip-like events difference in time development between W and Fe

can still be seen

C. Grefe

3/27/2010

Page 29: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

P. Speckmayer, LCWS2010, Beijing 29

Proton synchrotron (PS) East-Hall/ T9 Layout

3/27/2010

Page 30: Peter Speckmayer, LCWS2010, Beijing 1P. Speckmayer, LCWS2010, Beijing3/27/2010

Particle flow results so farComparison of around 8 ½ interaction lengths of HCAL with Fe and WW delivers comparable resolution to Feno optimization of the PFA for W done

P. Speckmayer, LCWS2010, Beijing 30

Angela Lucaci-Timoce

3/27/2010