Phase Technology PSA-70Xi Pour, Cloud & Freeze Point Analyzer. About Phase Technology. Expertise: Design & manufacture of analyzers to test cloud, pour and freeze point Cold Flow is All We Do! Scientific Background: 29 year history of research; 30 patents Market Leader: - PowerPoint PPT Presentation

TRANSCRIPT

  • Phase Technology PSA-70Xi Pour, Cloud & Freeze Point Analyzer

  • About Phase TechnologyExpertise: Design & manufacture of analyzers to test cloud, pour and freeze pointCold Flow is All We Do! Scientific Background: 29 year history of research; 30 patents

    Market Leader: 20 years of inter-laboratory round robin research Phase ranks highest for repeatability, reproducibility, sensitivity and equivalency to reference manual test methods

  • Phase Technology customers

  • Why do more companies in the world choose Phase Technology analyzers?

  • Companies choose Phase Technology because:Worlds best precision Worlds fastest test speed Worlds most complete set of features and easy-to-use analyzer

  • Phase Technology 70Xi

  • 70Xi analyzer benefitsFastest speed of any automatic or manual method: Average test duration 5 10 min Average operator time 0.5 2 minEasy and fast to use saves time in the labBest overall precision of any automatic or manual ASTM methodPowerful cooling system; minimum sample temperature to below -88C [-126F]

  • 70Xi productivity featuresFast, one-touch testing press one button to run a test instead of entering multiple screens of informationFull-color, touch-sensitive, 15 high resolution screenEasier to read and view more informationDisplays several views of information at the same time. Touch sensitive design is faster than keyboard & mouse.Easy-to-use interfaceNew colorful graphic icon buttons for faster navigation.Intuitive and easy to understand on-screen prompts.

  • 70Xi productivity featuresUSB and Ethernet connectivityQuickly export analyzer data to a portable USB flash drive. Ethernet port available for connecting to LIMS or internal computer networkImport and store any user documents (.doc, .pdf, .ppt and .pps) for customized operating procedures (SOP) or training Reliability improvements Modular component design boards can be easily replaced in the field; no need to return to Canada for serviceMore stable coolerFaster multiprocessor

  • 70Xi intellinosticsBuilt-in answers and operator information Quick, responsive solutions built-in alert messages give troubleshooting options to quickly solve problemsSelf-guided tutorials easy, on-screen instructions show how to use analyzer In-depth system diagnostics system checks verify performance and detect problems with critical components

  • Phase Technology 70Xi Autosampler System

  • Automatic sampling & cleaningSet-and-forget convenience set 1 to 48 test runs, then walk away and let the analyzer do the rest! Automatic sample injection no need to individually pipette samples or change pipette tips. Always uses the exact amount of sample required by the ASTM method.Self-cleaning Saves time and ensures perfect cleaning every time.No more need for cotton swabs or risk of damage to sample cup.

  • 70Xi home screen

  • 70Xi Favorites featureOne-touch preset FavoritesFrequently-used test settings can be stored in the analyzer for quick access. Favorites can be created, organized, sorted or deleted as you desire. With the single push of a button, lab operators can efficiently perform tests in significantly less time

  • Built-in tools to validate test resultsOverlay and compare multiple phase plotsControl and trend charts Automatic QC runOption to retest sample Optional alert or lockdown if out of acceptable limits

  • 70Xi control chart consistent runs

  • 70Xi control chart bad run

  • 70Xi plot overlay consistent runs

  • 70Xi plot overlay bad run

  • On-screen pour point phase plot

  • Understanding phase plots

  • Phase plot can tell you when there is a problem:example of possible moisture in sample

  • On-screen cloud point phase plot

  • Understanding phase plots

  • Phase plot can tell you when there is a problem:example of possible air bubble in sample

  • On-screen freeze point phase plot

  • Understanding phase plots

  • Phase plot can tell you when there is a problem: example of contaminated sample with tail

  • ASTM Methods Comparison

  • POUR POINT ASTM Method Comparison

  • POUR POINT ASTM Method ComparisonISL/PAC CPP5GsThe CPP5GS is extremely slow and has poor precision uses a mechanical tilt method which takes 80 140 minutes to complete a pour point testCompare to Phase Technology pour point test takes 8 20 minutes

    ISL/PAC MPP5GsThe ISL MPP5Gs uses ASTM D7346 to determine the no flow point of petroleum products. ASTM stated that there is no correlation between no flow and pour point.ASTM D7346 no flow point cannot be used to certify lube oil products for release because it is not a pour point method. The only accepted standards for product release are pour point methods, such as ASTM D5949 (as used by Phase Technology 70Xi analyzers).

  • POUR POINT ASTM Method ComparisonTanaka MPC-102Tanaka is the slowest automatic method reports of 90 120 minutes for low temperature lubesPhase Technology 70Xi can perform at least eight test runs in the time that it takes Tanaka to perform one single test!The MPC-102 requires 4 mL of sample (contrast to Phase Technology use of 0.15 mL). Larger sample volume introduces non-uniform temperature.Even with external liquid cooler, MPC-102 is not able to reach the low temperature level that Phase Technology does. Tanaka is only capable of going down to -30C with internal coolerPhase Technology reaches -88C with internal cooler

  • CLOUD POINT ASTM Method Comparison

  • CLOUD POINT ASTM Method ComparisonISL CPP-5GsWorst precision of any automatic methodRequires significant amount of time for sample preparation; water bath or oven must be used to bring sample to at least 14C above the expected cloud point.

    ISL MPP-5GsInterlaboratory Round Robin results showed there is a statistically significant bias of 1.23C relative to ASTM D2500.Slow test speed Phase Technology 70Xi can perform at least five test runs in the time that it takes MPP-5Gs to perform one single test!

  • CLOUD POINT ASTM Method ComparisonTanaka MPC-102Interlaboratory Round Robin results showed there is a statistically significant bias of 1.68C relative to ASTM D2500.Tanaka is the slowest automatic method takes almost as long as the manual method.The MPC-102 requires 4 mL of sample (complare to Phase Technology use of 0.15 mL). Larger sample volume introduces non-uniform temperature.Even with external liquid cooler, MPC-102 is not able to reach the low temperature level that Phase Technology does. Tanaka is only capable of going down to -30C with internal coolerPhase Technology reaches -88C with internal cooler

  • FREEZE POINT ASTM Method Comparison

  • FREEZE POINT ASTM Method ComparisonISL/PAC takes 2 to 3 times longer to run a freeze point test compared to Phase TechnologyISL/PAC FZP52Gs freeze point analyzer has serious false warming issuesNumerous studies have shown high incidence of false warming as high as 60% in a BP studyShell Bulletin GLN 33a (September 14, 2006) warned that ISL can occasionally falsely show a failure if samples contain small amounts of high boiling point material

  • The Phase Technology ADVANTAGE:Phase Technologys 70Xi analyzer is 4 to 10 times faster than the competition!

    Phase Technologys 70Xi analyzer has the worlds best precision

    The 70Xi requires much less sample: 4 to 14 ml less needed per test!

  • The Phase Technology ADVANTAGE:Phase Technology has the oldest, most established, continuous-in-production automatic cold flow ASTM Methods 17 years of proven, in-the-field use! The PSA-70Xi tests pour, cloud and freeze point in one single unit. Youd have to buy two or three separate analyzers from our competitor!

  • Thank you!

    For more information:phase-technology.com

    At the end of a freeze point test, the display should resemble the following plot. When a sample is cooled, wax crystals form. The freeze point is the temperature at which the last crystals disappear; on the plot, it appears as the point where the curve turns flat during the warm-up, or the verify stage of the test.

    Green line shows where we are warming sample

    Analyzer continues warming a little just to make sure

    : Contaminated

    When you have off spec or contaminated sample the plot can tell you why.

    Quick melting is Jet, but real freeze point is -47C

    Slow diesel melting causing a tail could also be water or other contaminants

    Red Colour is cooling / as is blue