physics mi diploma mi 10708-3
TRANSCRIPT
-
8/14/2019 Physics MI Diploma MI 10708-3
1/47
Physics of
Medical ImagingDiploma of Medical Imaging
1/2007/2008
Siti Zurina Mat Noor
-
8/14/2019 Physics MI Diploma MI 10708-3
2/47
Prime Exposure Factors
X-ray Panel (control board)
KVp
mA
Exposure time
Distance
Generator
X-ray circuit
-
8/14/2019 Physics MI Diploma MI 10708-3
3/47
Objectives
Be able to list the four prime exposure factors
Be able to discuss mAs and kVp in relation to
x-ray beam quantity and quality
Be able to describe characteristics of the
imaging system that affect x-ray beam qualityand quantity
-
8/14/2019 Physics MI Diploma MI 10708-3
4/47
Exposure Factors
Proper exposure of a patient to x-radiation is
necessary to produce a diagnostic radiograph
-
8/14/2019 Physics MI Diploma MI 10708-3
5/47
Exposure Factors
(control)
The emission of x-ray photons from an x-ray tube iscontrolled by a number of factors
These factors are
Milliamperage (mA) Exposure time (s) Kilovoltage (kVp)
SID (distance (d))
These factors and filtration control the quantityand/or quality of x-ray emission
mAs
-
8/14/2019 Physics MI Diploma MI 10708-3
6/47
Exposure Factors
The x-ray beam can be described in terms of
both its quality and its quantity
mAs and kVp are most important factorsresponsible for x-ray quantity and quality
Focal spot size, distance and filtration are
secondary factors
-
8/14/2019 Physics MI Diploma MI 10708-3
7/47
X-ray quality & quantity
X-ray quantity is a measure of the number of x-rayphotons in the useful beam (x-ray output/intensity)
Measured in mR or mR/mAs
X-ray quality is a measurement of the penetrating
ability of the x-ray beam (x-ray energy/penetrability)
Represented/measured by the half-value layer (HVL)
-
8/14/2019 Physics MI Diploma MI 10708-3
8/47
X-ray Image Formation
-
8/14/2019 Physics MI Diploma MI 10708-3
9/47
kVp
kVp is the primary control of beam quality (penetrability) The ability of x-ray beam to penetrate tissue
The higher-quality x-ray beam is one with higher energy andthus is more likely to penetrate the anatomic part of interest
kVp Controls radiographic contrast
Increasing kVp, more x-rays are emitted that have higherenergy and greater penetrability
Produce more scatter radiation resulting Increase image noise
Reduce image contrast
-
8/14/2019 Physics MI Diploma MI 10708-3
10/47
Body Penetration
-
8/14/2019 Physics MI Diploma MI 10708-3
11/47
X-ray Beam Quality
-
8/14/2019 Physics MI Diploma MI 10708-3
12/47
X-ray Image Contrast
-
8/14/2019 Physics MI Diploma MI 10708-3
13/47
mA
mA determines the number of x-raysproduced and therefore the radiation quantity
Unit Ampere = 1 C/s = 6.3 x 1018 electrons/s
More electron flow through the x-ray tube,
more x-ray are produced
An increase in mA will increase the radiation doseto the patient
-
8/14/2019 Physics MI Diploma MI 10708-3
14/47
X-ray Image Contrast
-
8/14/2019 Physics MI Diploma MI 10708-3
15/47
X-ray Beam Penetration
-
8/14/2019 Physics MI Diploma MI 10708-3
16/47
Exposure Time
Exposure times are usually kept as short as possible Not to minimize patient radiation dose but to minimize
motion blur
When exposure time is reduced, the mA must beincreased
Three-phase and high-frequency generators cannormally provide exposure times as short as 1 ms
-
8/14/2019 Physics MI Diploma MI 10708-3
17/47
Exposure Time
Image blur
-
8/14/2019 Physics MI Diploma MI 10708-3
18/47
Blur
-
8/14/2019 Physics MI Diploma MI 10708-3
19/47
mAs
mA and exposure time are usually combined andused as one factor (mAs)
mAs controls
Radiation quantity Doubling of the mAs doubles the x-ray quantity Optical density
OD is constant for any combination of mA and exposure time thatresults same mAs
Patient dose
mAs is not influence radiation quality
-
8/14/2019 Physics MI Diploma MI 10708-3
20/47
Distance
SID determines the intensity of the x-ray beam at theimage receptor and has no effect on radiation quality
Standard SID for tabletop radiography is 100cm
whereas chest examination is usually conducted at
180cm
The use of longer SID results in
less magnification,
less focal-spot blur, and
improve spatial resolution but more mAs
-
8/14/2019 Physics MI Diploma MI 10708-3
21/47
Distance
-
8/14/2019 Physics MI Diploma MI 10708-3
22/47
Distance and mAs
mAs (Second exposure) / mAs (first exposure)= SID2 (Second exposure) / SID2 (first
exposure)
-
8/14/2019 Physics MI Diploma MI 10708-3
23/47
Exposure Factors
(Fixed)
All factors are under the control of theradiologic technologist except those fixed by
the design of the x-ray imaging system
Focal spot size
Added x-ray beam filtration is fixed
High voltage generator provides characteristic
voltage ripple that cannot be changed
Fil i d
-
8/14/2019 Physics MI Diploma MI 10708-3
24/47
Filtration and
voltage ripple
-
8/14/2019 Physics MI Diploma MI 10708-3
25/47
Focal spot
-
8/14/2019 Physics MI Diploma MI 10708-3
26/47
Focal-spot size
Most x-ray tubes have small and large focal-spot 0.6mm/1.2mm, 0.5mm/1.0mm or 1.0mm/2.0mm
0.3mm/1.0mm for magnification radiography
Large focal spot is used for normal radiography
Produce many x-ray Provide shorter exposure time (minimize blur)
Use for thick or dense body part
Small focal spot is used in most mammography tube to checkmicrocalcification at short distance (0.1mm/0.3mm)
Fine-detail radiography Magnification radiography
Use for extremities radiography
-
8/14/2019 Physics MI Diploma MI 10708-3
27/47
Focal spot
-
8/14/2019 Physics MI Diploma MI 10708-3
28/47
Filtration
Inherent filtration(total 2.5mm aluminum - no control)
Properties of glass or metal envelope
Approximately 0.5 mm aluminum equivalent
The variable-aperture light-localizing collimator provides
additional 1.0mm aluminum equivalent
Mirror of collimator
Additional filter of 1.0mm of aluminum between x-ray tubehousing and collimator
-
8/14/2019 Physics MI Diploma MI 10708-3
29/47
Filtration
-
8/14/2019 Physics MI Diploma MI 10708-3
30/47
Filtration
Added filtration Higher filtration is used during examination of tissue with
high subject contrast such as extremities, joints and chest
Lower patient dose
As added filtration increases, x-ray beam quality and penetrability is increased
More scattered radiation
Image noise increased
Reduced image contrast
-
8/14/2019 Physics MI Diploma MI 10708-3
31/47
High-voltage generation
The radiation quantity and quality produced inx-ray tube are influenced by the type of high-
voltage generator
The three type of high-voltage generators are
Single phase
Three phase
High frequency
-
8/14/2019 Physics MI Diploma MI 10708-3
32/47
Tube kVp
-
8/14/2019 Physics MI Diploma MI 10708-3
33/47
Transformer
-
8/14/2019 Physics MI Diploma MI 10708-3
34/47
Waveform
-
8/14/2019 Physics MI Diploma MI 10708-3
35/47
Voltage ripple
-
8/14/2019 Physics MI Diploma MI 10708-3
36/47
Single phase
Half-wave rectification
It has 100% voltage ripple
During exposure, x-ray are produced and emitted only half
the time Radiation quantity is halved compare to full-wave
rectification
During each negative half-cycle, no x-ray are emitted
Mobile and dental x-ray are half-wave rectified
-
8/14/2019 Physics MI Diploma MI 10708-3
37/47
Single phase
Full-wave rectification
The voltage waveform is identical to that half-waverectification except there is no dead time
During exposure, x-ray are continually emitted inpulses
The radiation quantity is doubled compare to half-wave rectification
The required exposure time is half that for half-wave rectification
-
8/14/2019 Physics MI Diploma MI 10708-3
38/47
Three phase
6 or 12 pulse
The difference between two forms cause a detectiblechange in x-ray quantity and quality
It is more efficient than single-phase power More x-ray produced per mAs
The average energy of x-ray is higher
Higher x-ray quantity and quality The radiation emitted is nearly constant rather than pulsed
-
8/14/2019 Physics MI Diploma MI 10708-3
39/47
High frequency
Is used with low-power x-ray systems
The voltage waveform is nearly constant withless than 1% ripple
Higher x-ray quantity and quality
Is used with mammography and mobile x-ray
-
8/14/2019 Physics MI Diploma MI 10708-3
40/47
High-voltage generation
70/24112/3300
-
8/14/2019 Physics MI Diploma MI 10708-3
41/47
Summery
Radiographic exposure factors mAs, kVp and distance
The exposure factors influence
Quantity (number of x-rays) Quality (penetrability of x-rays)
Proper selection of these factors optimize
Spatial resolution Contrast resolution
S
-
8/14/2019 Physics MI Diploma MI 10708-3
42/47
Summery
IncreaseDecreaseFiltrationDecreaseDecreaseVoltage ripple
No changeDecreaseDistance
No changeIncreasemAsNo changeIncreaseExposure time
No changeIncreasemA
IncreaseIncreasekVpQualityQuantityFactor increased
Ch kV
-
8/14/2019 Physics MI Diploma MI 10708-3
43/47
Change kVp
Ch kV
-
8/14/2019 Physics MI Diploma MI 10708-3
44/47
Change kVp
Test
-
8/14/2019 Physics MI Diploma MI 10708-3
45/47
est
your understanding
What is the electron flow from cathode toanode when the 500-mA station is selected?
At 200 mA the entrance skin exposure (ESE)is 752 mR (7.5 mGy). What will be the ESE at
500 mA if no change in exposure time?
A radiographic technique calls for 600 mA at200 ms. What is the mAs?
Test
-
8/14/2019 Physics MI Diploma MI 10708-3
46/47
your understanding
A radiograph of the abdomen requires 300 mA and500 ms. The patient is unable to hold his breath,
which results in motion blur. A second exposure is
made with an exposure time of 200 ms. Calculate
the new mA that is required?
An examination requires 100 mAs at 180cm SID. If
the distance is changed to 90cm SID, what should
be the new mAs?
-
8/14/2019 Physics MI Diploma MI 10708-3
47/47
Thank You