pitting corrosion due to deposits in cooling water systems

Upload: anonymous-1xhscfci

Post on 07-Jul-2018

223 views

Category:

Documents


1 download

TRANSCRIPT

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    1/12

    By Sang-Hea Shim, Donald A. Johnson and Barbara E. Moriarty, Nalco Chemical Company

    SM

    Characterization of Localized andUnderdeposit Corrosionin Cooling Water Systems

    Presented at the National Association of Corrosion Engineers, Corrosion ‘88 meeting,St. Louis, Missouri, March 21–25, 1988.

    1. Modification of the standard pitting scan technique

    which allows for the development of inhibitor films

    under oxygenated conditions

    2. Galvanometric technique which measures the dif-

    ferential current between a well-aerated cathode

    and synthetic occluded cells of various dimensions

    3. Alternating Current (AC) impedance method whichmeasures change in the surface impedance char-

    acteristics caused by the presence of pits

    The above methods have been used to characterize

    the ability of some major classes of corrosion inhibi-

    tor programs to prevent and passivate localized

    corrosion. Data is presented on three significantly

    different types of corrosion inhibitor packages:

    • Neutral pH high-chromate program

    • Neutral pH stabilized phosphate package

    • High alkalinity phosphonate-based all-organic

    inhibitor

    Significant differences among these programs were

    found. A mechanistic rationale to explain these dif-

    ferences is proposed.

    EXPERIMENTAL METHODS AND RESULTS

    MODIFIED PITTING SCAN

    There are many different electrochemical techniques

    used to determine characteristic pitting potentials.

    Generally, these methods can be divided into

    potentiostatic or galvanostatic categories.1 Of thesemethods, potentiodynamic cyclic polarization (also

    known as the “pitting scan”) is most widely used to

    measure pitting nucleation and passivation potentials.

    The standard pitting scan is typically1 conducted

    by scanning anodically from the open circuit

    potential until a specified current density

     ABSTRACT

     Localized attack is potentially the most damaging form

    of corrosion encountered in cooling water systems. This

     paper discusses the mechanism of localized corrosion,

     presents some novel techniques for evaluating the

    tendency toward and extent of localized corrosion on

    carbon steel, and compares the ability of various

    inhibitor approaches.

    INTRODUCTION

    The inhibition of general corrosion by cooling water

    is a well-developed technology which has been dis-

    cussed extensively in the literature. However, there

    is a potentially more damaging form of corrosion

    which has received much less attention, i.e., local-

    ized corrosion. Pitting, underdeposit corrosion, and

    crevice corrosion are related phenomena which can

    be described by the generalized term of “localized

    corrosion.”The microenvironment which gives rise to localized

    corrosion is very different from that associated with

    general corrosion. As a consequence, the requirements

    for inhibition of these two forms of corrosion are very

    different. The testing methods which are typically

    used to evaluate inhibitors generally do a poor job of 

    simulating the conditions occurring on a real-world

    “technical” surface on which significant deposits gen-

    erally exist. Arguments for extensive cleaning proce-

    dures or for invoking unrealistic expectations of 

    antifoulants or dispersants are really avoiding the

    issue of underdeposit corrosion control. What is

    needed are evaluation methods which address

    underdeposit corrosion as a discreet problem, and

    inhibitors which can effectively function in the micro-

    environment which is the cause and the consequence

    of localized corrosion. This paper discusses some

    methods which can quantify localized corrosion. These

    methods include:

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    2/12

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    3/12

    3

    2. If Eo is between Epp and Enp, no new pits will form,

    but existing pits will not passivate.

    3. If Eo is greater than both Epp and Enp, new pits will

    form and will not passivate.

    The following examples illustrate some parameters,

    which influence Epp  and Enp  as measured by this

    technique.

     A phosphonate-based all-organic corrosion inhibitor

    program which is designed to operate at elevated pH

    and alkalinity was tested using the above procedure.

    This inhibitor program has been described in previ-

    ous publications2,3. This sequence of experiments

    illustrates the effect of alkalinity, phosphonate and

    polymer on Epp and Enp and on the anodic passive

    current.

    The first comparison (Figure 2) shows the effect of 

    the phosphonate combination on Epp, Enp  and the

    anodic passive current. While the anodic passive cur-

    rent remains relatively unchanged, a 100 mV increase

    in Enp is observed.

    The second comparison (Figure 3) shows the effect of 

    adding an anionic acrylic copolymer to the phos-

    phonates. This polymer is an antiscalant with activity

    as a calcium phosphate and phosphonate inhibitor.4

    The addition of the polymer further increases Enp and

    Epp and also causes a decrease in the anodic passiva-

    tion current.

    The third comparison shows the effect of varying al-

    kalinity on modified pitting scan results. Figure 4

    shows the effect of alkalinity variation on Epp  and

    Enp. The pitting nucleation potential increases to a

    greater extent than the pitting passivation potential.

    The passive current also decreases with increasing 

    alkalinity (Figure 5). The carbon steel specimens were

    examined under an optical microscope after each pit-

    ting scan measurement. The number of pits was larg-

    est on the specimen exposed to 100 ppm alkalinity,

    and decreases with increasing alkalinity. These re-

    sults show the correlation between the pitting nucle-ation potential and passive current results. When the

    film is weaker, the passive current on the pitting scan

    is higher and pits can easily nucleate on the surface,

    causing the low Enp and large number of pitting sites.

    Pitting scan techniques have been widely used as a

    predictive tool for characterizing the tendency of a

    material to experience pitting corrosion. Although

    valuable information can be obtained from these tech-

    niques, they have their limitations. Used in a cooling 

    water environment on carbon steel, they are prima-

    rily a measure of the resistance of an inhibitor film

    to disruption by anodic corrosion. In conducting anexperiment, a tremendous electrical perturbation is

    applied to the specimen, essentially destroying it as

    a relevant metal specimen. This limitation makes this

    technique a “snapshot” method which cannot follow

    the corrosion and inhibition processes over time. This

    method, like other potentiodynamic methods, char-

    acterizes the tendency of a material in a medium to

    experience corrosion under conditions which are dras-

    tically different than those typically encountered in

    service. Therefore, other techniques are needed.

    Figure 3  — Effect of polymer addition in the all-organic inhibitor solution

    Figure 2 — Effect of phosphonate on pitting scan

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    4/12

    4

     AC IMPEDANCE MEASUREMENTS

    The AC impedance technique has been extensively

    used to study corrosion phenomena5,6. By applying a

    low-amplitude oscillating potential to the working 

    electrode, the perturbation of the electrode is mini-

    mized and useful information about the electrochemi-

    cal characteristics of the inhibitive and corrosion pro-

    cesses on the surface is obtained. This technique is a

    powerful tool for characterizing inhibitive films on

    metal surfaces. In addition, it has been recently sug-

    gested7 that AC impedance is applicable to the studyof localized corrosion. Mansfeld proposed that char-

    acteristic patterns are observed in the impedance

    spectra in the presence of pitting corrosion.

    Experiments were done in order to determine the

    effect of pitting corrosion on the impedance charac-

    teristics of metal surfaces and to determine differences

    in the film characteristics of two radically different

    inhibitor programs. The programs studied were a

    chromate-only inhibitor which was dosed at a marginal

    level, and the previously-described phosphonate-

    based all-organic program.

    The chromate was dosed at a marginal level (200 ppm)

    without the presence of zinc or any other inhibitors.

    The pH of the chromate experiment was controlled

    at 7 using controlled addition of carbon dioxide gas

    into the air sparger, while the all-organic program

    was tested at high alkalinity levels, equilibrated withair. The resulting pH was typically greater than 9.

    Both tests used air-saturated solutions. All measure-

    ments were performed at 120°F.

    The measurements were performed on mild steel

    rotating cylinder electrodes using a Princeton Applied

    Research model (PAR) 273 potentiostat and a PAR

    model 5301 lock-in amplifier. The potentiostat and

    lock-in were controlled by an Apple II™  computer

    using the PAR AC impedance software.

    Figure 6 shows the Bode plot obtained at 1.5 hours

    after immersion of the specimen into the chromate

    solution. The figure shows only one peak, which isrelated to the double layer capacitance. In Figure 6,

    the Bode plot taken after 24 hours of immersion is

    also shown. This plot shows an additional small peak

    in the high frequency range. This phenomenon is also

    clear when the Nyquist plot (Figure 7) is examined.

    The Nyquist plot shows a small semi-circle in the high

    frequency range. This phenomenon appears to be

    related to the formation of a primarily anodic passive

    film on the metal surface. At 72 hours into the ex-

    periment, another distortion appears on the Bode plot.

    Figure 8 illustrates that the low frequency phase shift

    does not approach 0 degrees. Rather, it decreases and

    approached –32 degrees at 0.001 Hz. The shape of the Bode plot did not change after this time for up to

    108 hours. The rotating disk electrode was examined

    under an optical microscope, and the presence of pits

    Figure 4 — Effect of alkalinity on Enp and E pp in theall-organic inhibitor solution

    Figure 5 — Effect of alkalinity on passive current in

     pitting scan curves in all-organic inhibitor solution

    Figure 6  — Bode plot at 1.5 and 24 hours after immersion of the mild steel electrode into thechromate solution

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    5/12

    • Rct is used as a measure of the general corrosion

    rate

    • Cfilm and Rfilm are the capacitance and resistance of 

    the inhibitor film

    • Rpit is related to the pitting corrosion rate

    • Cpit is the double layer capacitance over the pit

    • W is a Warburg impedance which is due to diffu-

    sional processes of oxygen or corrosion products

    through the pit

    The measurements taken in the chromate solution

    indicate that a detectable film was formed at about

    24 hours after the immersion of the specimen. This

    is demonstrated by the appearance of the high fre-

    quency peak. Figure 10 shows the open circuit

    5

    was observed. The low frequency phase distortion of 

    the Bode plot appeared to coincide with the forma-

    tion of pits on the metal surface.

    In order to analyze these results, the development of an equivalent circuit to model the impedance behav-

    ior is useful. The various processes at work in the

    metal/liquid interface can be visualized as elements

    in the equivalent circuit. A proposed equivalent cir-

    cuit model for the chromate results is shown in Fig-

    ure 9. The physical and chemical processes analogous

    to these elements are:

    • Rs is the resistance of the solution

    • Cdl represents the double layer capacitance

    Figure 7 — Nyquist plot at 24 hours after immersionof the mild steel electrode into the chromate solution

    Figure 8 — Bode plot at 72 hours after immersion of the mild steel electrode into the chromate solution

    Figure 9  — Equivalent circuit for the impedanceresponse in the chromate solution

    Figure 10 — Change in open-circuit potential in thechromate solution

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    6/12

    potential throughout the experiment. The increasing 

    tendency of Ecorr  indicates the formation of an anodic

    passive film. At 72 hours, visible pitting was observ-

    able and the low frequency phase distortion began to

    appear in the Bode plot.

    Results of the measurements taken with the all-

    organic inhibitor are quite different than those obtained

    from the chromate solution. Up to 48 hours into the

    experiment, the Bode plot showed a single peak, and

    the Nyquist plot only one semicircle (Figures 11 and

    12). However, after 72 hours, the shape of the Nyquist

    plot changed significantly. The low frequency region

    is linear with a slope of about 45 degrees. Figure 13

    shows that the 45-degree region spans the frequency

    range of 0.02 to 0.005 Hz. No visible pitting was

    observed on the specimen throughout the test. The

    test was continued for 96 hours with the results shown

    in Figure 14. The Bode plot shows that the region

    showing Warburg characteristics was extended to

    lower frequencies (0.002 Hz), indicating a thickening 

    of the cathodic film. Figure 15 shows the open circuit

    potential measured throughout this experiment. The

    decrease with time is indicative of the formation of a

    cathodic passive film.

    The impedance response observed with the all-organic

    program is typical of a diffusion layer of finite thick-

    ness. As the all-organic program is predominantly a

    cathodic inhibitor8, an electron insulator film is

    formed on the metal surface. Therefore, it seems prob-

    able that the impedance behavior is caused by oxygen

    diffusion through this finite layer.

    OCCLUDED CELL MEASUREMENT

    There are many similarities between pitting, under-

    deposit corrosion and crevice corrosion. Often, local-

    Figure 11 — Bode plot at 48 hours after immersion of the mild steel electrode in the all-organic inhibitor solution

    Figure 12 — Nyquist plot at 48 hours after immersionof the mild steel electrode in the all-organic inhibitor solution

    Figure 13 — Bode plot at 72 hours after immersion of the mild steel electrode in the all-organic inhibitor solution

    Figure 14 — Bode plot at 96 hours after immersion of the mild steel electrode in the all-organic inhibitor solution

    6

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    7/12

    ized corrosion contains elements of all three forms.

    The test methods, which have been previously de-

    scribed, all start with a clean surface and measure

    some element of the resistance of passive films to

    breakdown by anodic corrosion. In the case of 

    underdeposit corrosion, the ability of an inhibitor to

    act may be limited by its ability to diffuse to a sur-

    face through a porous deposit and by its ability to act

    in the environment resulting from the localized cor-

    rosion process. Although the previously discussed test-

    ing methods simulate this process to some extent, thespecific geometry through which diffusion must oc-

    cur is somewhat a matter of chance.

     An occluded cell apparatus was discussed by Krisher8.

    In order to specifically address the issue of inhibitor

    mobility and effectiveness, a similar apparatus was

    constructed to allow for variation of cell geometries,

    simulating various pit depths. This apparatus is a

    standard laboratory recirculator apparatus (Figure 16)

    which contained an electrode assembly (Figure 17)

    consisting of a free electrode and occluded electrodes

    in various configurations.

    The free electrode (cathode) was a section of carbonsteel tube. The occluded cells (anodes) were carbon steel

    cylinders imbedded in a nonconductive plastic block.

    The depth of the anodes was adjusted by using dif-

    ferent lengths of plastic tubing inside the cells. Three

    different cell depths were used with three replicates

    of each cell depth. The depths of the shallow, medium

    and deep cells were 4, 11 and 18mm, respectively.

    The anodes were polished with #600 grit emery paper

    and rinsed with acetone before they were inserted in

    the occluded cells. The cell box was filled with test

    solution and mounted in the recirculating loop. The

    temperature of the recirculating solution and the pH

    of the solutions were controlled by automatic control-lers. To maintain the inhibitor concentration, makeup

    water was pumped continuously into the basin and

    out an overflow. Inhibitor was added proportionately

    to account for the overflow by a syringe pump. The

    current between the anodes and cathode was moni-

    tored by zero resistance ammeters and a computer

    data acquisition system.

    Three different cooling water inhibitor programs were

    tested using this device. They were chosen to represent

    a spectrum of inhibitors currently in wide commercial

    use and a variety of inhibitor types. They included

    the phosphonate all-organic, a chromate program, and

    a neutral pH-stabilized phosphate program. The

    conditions of the three experiments are shown in Table 1.

    Figures 18, 19 and 20 show typical results obtained

    from these experiments. The reproducibility of the

    currents from the replicate anodes was quite good.

    The passivation time of the occluded electrodes in the

    all-organic and stabilized phosphate programs varied

    depending on the depth of the cells. In the all-organic

    experiment, the shallow cells took about 80 hours for

    passivation to occur. The medium and deep cells took

    Figure 15 — Change in the open circuit potential in

    the all-organic inhibitor solution

    Figure 16 — Recirculator apparatus for the occludedcell experiments

    Figure 17 —Occluded cell assembly

    7

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    8/12

    8

    140 and 200 hours respectively. The stabilized phos-

    phate program took less time to achieve steady state

    currents. The current from the shallow cells leveled

    out at about 15 hours. The medium and deep cells

    required 60 and 80 hours respectively. The steady-

    state currents observed in the all-organic program

    were far lower than those seen in the phosphate pro-gram. The average current from the deep occluded

    cells after passivation was approximately 0.7 µ A for

    the all-organic program and typically 9 µ A for the

    stabilized phosphate program. The passivation times

    observed with the chromate program were relatively

    insensitive to the depth of the cell. All three depths

    gave sudden and effective passivation after 75 hours.

     After completion of the experiment, the anodes were

    microscopically examined. The exposed area of the

    anodes from the stabilized phosphate program

    showed significant visible etch with depth. The

    visible effects on the anodes from the other two pro-

    grams were significantly less, with little visible etch.

    DISCUSSION

    The typical bulk cooling water corrosion environment

    consists of air-saturated water at neutral to slightly

    alkaline pH with varying amounts of hardness and

    other solutes in the presence of corrosion inhibitors.

    The metal surface environment is typically carbon

    steel of varying cleanliness with numerous small

    niches and crevices. These are obvious starting points

    for localized corrosion. However, under these condi-

    tions, even initially clean carbon steel equipment canexperience underdeposit corrosion. Defects in passive

    films or variations in metal composition can cause

    local dissolution of iron. The primary product of this

    reaction is Fe+2. Compounds of Fe+2 are quire soluble

    under these conditions. However, once the Fe+2

    reaches the oxygenated bulk water, it is quickly

    oxidized to Fe+3. There are numerous oxides and other

    components of this ion which are very insoluble and

    quickly precipitate. Thus, what starts out as a pit,

    quickly acquires a layer of deposit. In the case of 

    Table 1 — Operating conditions for occluded cell experiments

    Inhibitor program

    All-organic stabilized phosphate Chromate

      Product concentration (ppm) 200* 100** 300***

      pH 9* 7** 7***

      Temperature (°F) 120* 100** 120***

      Ca (ppm as CaCO3) 360* 360** 360***

      Mg (ppm as CaCO3) 200* 200** 200***

      Alkalinity (ppm as CaCO3) 350* 50** 50***  Conductivity (microsiemens) 2200* 2200** 2200***

      ***corresponds to 12 ppm phosphonates as PO 4 

     * **corresponds to 21 ppm total phosphate as PO 

     ***corresponds to 300 ppm chromate as chromate 

    carbon steel, pitting and underdeposit corrosion are

    intimately related.

    The local environment in an actively corroding 

    underdeposit site is significantly different than that

    in the bulk water, in that:

    • Oxygen is depleted• Fe+2 is the predominant iron form

    • The pH is very low

    • Ionic concentrations differ depending on electro-

    phoretic mobilities

    There are a number of issues which determine the

    ability of an inhibitor to prevent and passivate

    underdeposit corrosion. These are:

    • The quality and integrity of the inhibitor film and

    its ability to resist attack

    • The presence of components in the inhibitor whichcan effectively passivate in the underdeposit

    environment

    • The ability of these components to diffuse to the

    corroding site

    The three experiments presented focus on various as-

    pects of the problem. The pitting nucleation potential,

    Enp, measured by the modified pitting scan, is an in-

    dication of the resistance of preformed inhibitor films

    to rupture by anodic dissolution of iron. In the two

    examples presented, increasing the alkalinity and

    including an appropriate polymer had beneficial ef-

    fects on film strength. Increasing the alkalinity alsoincreased the pitting passivation potential Epp, indi-

    cating an improved ability to passivate existing de-

    fects in the film.

    The AC impedance method provides a great deal of 

    information about the characteristics of the film. Low

    frequency phase distortion in the Bode plot appears

    to correlate well with the appearance of pits. The

    appearance of a Warburg-like characteristic consist-

    ing of a linear 45 degree region in the all-organic im-

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    9/12

    9

    pedance spectrum is indicative of the formation of a

    diffusion layer of finite thickness. This provides evi-

    dence of the formation of a well-defined cathodic film

    on the mild steel electrode. The extension of the fre-

    quency range of the Warburg-like region corresponds

    to thickening of this layer. The formation of an an-

    odic passive film by the chromate program was indi-

    cated by the appearance of a small impedance peak

    at high frequency, corresponding to a capacitance-re-sistance type of circuit.

    The occluded cell experiment emphasizes the ability

    of an inhibitor to diffuse to a remote surface and to

    function in the oxygen-depleted, low pH, Fe+2-rich

    environment encountered under deposits.

    The ability of the chromate program to act as an anodic

    inhibitor in the absence of oxygen and to promote the

    conversion of Fe+2 to Fe+3 probably accounts for its ef-

    fectiveness in this test. The insensitivity to pit depth

    in this experiment indicates that this process is not

    diffusion-controlled. The potential of the occluded

    electrode is determined by the IR drop between the

    free electrode and the occluded electrode, where I is

    the current between the free electrode and R is mainly

    solution resistance. The occluded electrode potential

    can then be expressed as follows:

    Ecell = Eo – IR

    where Ecell is the potential of the occluded electrode

    and Eo is the open-circuit potential of the free elec-

    trode. Since Eo increases (Figure 10) and I decreases

    (Figure 20) with time, Ecell increases as the occluded

    cell measurement proceeds. This change in Ecell af-

    fects the passivation of the occluded electrode sincethe solution contains chromate ions. When Ecellreaches the critical potential for passivation, the cur-

    rent drops because of the passivation of the occluded

    electrode. This decrease in current causes a further

    increase in Ecell. Figures 21 and 22 show this process.

    In Figure 21, Ecell  is lower than the critical potential

    for passivation; therefore, the corrosion rate is very

    high. In Figure 22, Ecell increases to above the critical

    potential for passivation and the resultant corrosion

    rate is very low.

    The passivation of the occluded cells in the two

    nonchromate programs was different from that ob-

    served with the chromate program. These programs

    function primarily as cathodic inhibitors2 in contrast

    to the anodic character of the chromate program. The

    difference between the passivation currents observed

    in the stabilized phosphate and all-organic programs

    is probably due to the great difference in the alkalin-

    ity at which these programs are operated. At a pH of 

    7, the alkalinity is typically lower than 50 ppm

    (CaCO3). At a pH of 9, the alkalinity will be in the

    400-500 ppm range.

    Figure 18 — Results of occluded cell measurement inthe all-organic inhibitor solution

     Figure 19 — Results of occluded cell measurement inthe stabilized phosphate inhibitor solution

    Figure 20 — Results of occluded cell measurement inthe chromate solution

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    10/12

     Alkalinity can influence the underdeposit environ-

    ment in two ways. First, the increased buffer capacity

    afforded by a high alkalinity will tend to minimize

    the pH depression at an anodic occluded site, caused

    by hydrolysis by ferrous and ferric ions. Secondly, thepresence of the CO3= ion provides the opportunity for

    the precipitation of FeCO3(Siderite). While phosphate

    could potentially form Fe3(PO

    4)

    2(Vivianite), this re-

    action appears to be less effective in passivating 

    occluded cells. The Siderite model of corrosion inhi-

    bition in drinking water circuits has been extensively

    discussed by Sondheimer et al.9 They propose that it

    is an important intermediate phase which undergoes

    isomorphous transformation into oxidized iron phases

    such as Geothite.

    CONCLUSIONThe conditions under which underdeposit corrosion

    occurs are different from those controlling generalized

    corrosion. Issues of inhibitor diffusivity and ability to

    control corrosion in deoxygenated environments be-

    come important. The test methods used to evaluate

    inhibitors should take these factors into account. The

    experiments described in this study were intended

    to provide information about various aspects of local-

    ized corrosion such as:

    Figure 21  — Occluded cell potential and currentbefore passivation

    • The resistance of inhibitor films to pitting attack

    • The ability of inhibitors to stop crevice and under-

    deposit corrosion once it has begun

    • The effect of water chemistry on localized corrosion• The kinetics of pit passivation

    The described three localized corrosion measurement

    techniques provided important interrelated informa-

    tion. The major findings of the study are as follows:

    1. In the all-organic program, phosphonates (which

    comprise the major corrosion inhibitor in the pro-

    gram) did not decrease the passive current in the

    modified pitting scan curve confirming that they

    are cathodic inhibitors. However, Enp increased

    about 100mV, indicating a pitting inhibition prop-

    erty of the chemical.

    2. The polymer in the all-organic program improved

    the protective property of the passive film and in-

    creased pitting nucleation potential.

    3. The cathodic passive film formed in the all-organic

    program acted as a finite thickness diffusion layer

    and was detected as a Warburg-type impedance in

    the impedance measurements. The thickening of 

    the film with time was detected by the measurements.

    Figure 22 — Occluded cell potential and current after  passivation

    10

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    11/12

    4. Higher alkalinity and pH of a mainly cathodic in-

    hibitor seemed to be beneficial for the passivation

    of the occluded cells.

    5. The anodic passive film formed in the chromate

    solution was detected by impedance measurements

    as a parallel capacitance-resistance type of circuit

    at the high frequency range.

    6. The pitting formation by the underdosed chromateprogram was detected by impedance measure-

    ments at the low frequency range.

    7. The passivation of the occluded electrodes in an

    anodic inhibitor seemed to be related to the poten-

    tial of the electrode determined by the open circuit

    potential of the free electrode and the IR drop.

    REFERENCES

    1. Z. Szklarska-Smialowska, Pitting Corrosion of 

    Metals, National Association of Corrosion Engi-

    neers, Houston, Texas, 1986.

    2. G. Bohnsack, K.H. Lee, D.A. Johnson, E. Buss, “In-

    vestigation of the Mechanisms of Organic Com-

    pounds Used in Cooling Water Corrosion Control,”

     Materials Performance, Vol. 25, No. 5, p.32, 1986.

    3. K.E. Fulks, A.M. Yeoman, “Performance Evalua-

    tion of Non-Metal Cooling Water Treatments,” pre-

    sented at Corrosion/83, Anaheim, California, April,

    1983.

    4. J.E. Hoots, G.A. Crucil, “Role of Polymers in the

    Mechanisms and Performance of Alkaline Cooling 

    Water Programs,” Materials Performance, Vol. 26,

    No. 4, p. 17, 1987.

    5. D.D. Macdonald, “Theoretical Analysis of Electro-

    chemical Impedance,” Corrosion/87, Paper No. 479,

    National Association of Corrosion Engineers, Hous-

    ton, Texas, 1987.6. I. Epelboin, C. Gabrielli, M. Keddam, H. Takenouti,

    “Alternating-Current Impedance Measurements

     Applied to Corrosion Studies and Corrosion-Rate

    Determination,” Electrochemical Corrosion Test-

    ing, STP 727, ASTM, Philadelphia, Pennsylvania,

    p. 150, 1981.

    7. F. Mansfield, presented at the International Con-

    ference on Localized Corrosion, Orlando, Florida,

    1987.

    8. A.S. Krisher, “A Synthetic Crevice Device to Moni-

    tor Corrosiveness of Aqueous Systems,” Corrosion/ 

    81, Paper No. 153, National Association of Corro-

    sion Engineers, Houston, Texas, 1981.

    9. H. Sontheimer, W. Kolle, A. Kuch, Internal Cor-

    rosion of Water Distribution Systems, AWWA 

    Research Foundation, Denver, Colorado, 1985.

    11

  • 8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems

    12/12

    N A L C O C H E M I C A L C O M P A N YONE NALCO CENTER NAPERVILLE, ILLINOIS 60563-1198

    Registered Trademarks of Nalco Chemical Company Printed in U.S.A. 12-88