pitting corrosion due to deposits in cooling water systems
TRANSCRIPT
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
1/12
By Sang-Hea Shim, Donald A. Johnson and Barbara E. Moriarty, Nalco Chemical Company
SM
Characterization of Localized andUnderdeposit Corrosionin Cooling Water Systems
Presented at the National Association of Corrosion Engineers, Corrosion ‘88 meeting,St. Louis, Missouri, March 21–25, 1988.
1. Modification of the standard pitting scan technique
which allows for the development of inhibitor films
under oxygenated conditions
2. Galvanometric technique which measures the dif-
ferential current between a well-aerated cathode
and synthetic occluded cells of various dimensions
3. Alternating Current (AC) impedance method whichmeasures change in the surface impedance char-
acteristics caused by the presence of pits
The above methods have been used to characterize
the ability of some major classes of corrosion inhibi-
tor programs to prevent and passivate localized
corrosion. Data is presented on three significantly
different types of corrosion inhibitor packages:
• Neutral pH high-chromate program
• Neutral pH stabilized phosphate package
• High alkalinity phosphonate-based all-organic
inhibitor
Significant differences among these programs were
found. A mechanistic rationale to explain these dif-
ferences is proposed.
EXPERIMENTAL METHODS AND RESULTS
MODIFIED PITTING SCAN
There are many different electrochemical techniques
used to determine characteristic pitting potentials.
Generally, these methods can be divided into
potentiostatic or galvanostatic categories.1 Of thesemethods, potentiodynamic cyclic polarization (also
known as the “pitting scan”) is most widely used to
measure pitting nucleation and passivation potentials.
The standard pitting scan is typically1 conducted
by scanning anodically from the open circuit
potential until a specified current density
ABSTRACT
Localized attack is potentially the most damaging form
of corrosion encountered in cooling water systems. This
paper discusses the mechanism of localized corrosion,
presents some novel techniques for evaluating the
tendency toward and extent of localized corrosion on
carbon steel, and compares the ability of various
inhibitor approaches.
INTRODUCTION
The inhibition of general corrosion by cooling water
is a well-developed technology which has been dis-
cussed extensively in the literature. However, there
is a potentially more damaging form of corrosion
which has received much less attention, i.e., local-
ized corrosion. Pitting, underdeposit corrosion, and
crevice corrosion are related phenomena which can
be described by the generalized term of “localized
corrosion.”The microenvironment which gives rise to localized
corrosion is very different from that associated with
general corrosion. As a consequence, the requirements
for inhibition of these two forms of corrosion are very
different. The testing methods which are typically
used to evaluate inhibitors generally do a poor job of
simulating the conditions occurring on a real-world
“technical” surface on which significant deposits gen-
erally exist. Arguments for extensive cleaning proce-
dures or for invoking unrealistic expectations of
antifoulants or dispersants are really avoiding the
issue of underdeposit corrosion control. What is
needed are evaluation methods which address
underdeposit corrosion as a discreet problem, and
inhibitors which can effectively function in the micro-
environment which is the cause and the consequence
of localized corrosion. This paper discusses some
methods which can quantify localized corrosion. These
methods include:
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
2/12
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
3/12
3
2. If Eo is between Epp and Enp, no new pits will form,
but existing pits will not passivate.
3. If Eo is greater than both Epp and Enp, new pits will
form and will not passivate.
The following examples illustrate some parameters,
which influence Epp and Enp as measured by this
technique.
A phosphonate-based all-organic corrosion inhibitor
program which is designed to operate at elevated pH
and alkalinity was tested using the above procedure.
This inhibitor program has been described in previ-
ous publications2,3. This sequence of experiments
illustrates the effect of alkalinity, phosphonate and
polymer on Epp and Enp and on the anodic passive
current.
The first comparison (Figure 2) shows the effect of
the phosphonate combination on Epp, Enp and the
anodic passive current. While the anodic passive cur-
rent remains relatively unchanged, a 100 mV increase
in Enp is observed.
The second comparison (Figure 3) shows the effect of
adding an anionic acrylic copolymer to the phos-
phonates. This polymer is an antiscalant with activity
as a calcium phosphate and phosphonate inhibitor.4
The addition of the polymer further increases Enp and
Epp and also causes a decrease in the anodic passiva-
tion current.
The third comparison shows the effect of varying al-
kalinity on modified pitting scan results. Figure 4
shows the effect of alkalinity variation on Epp and
Enp. The pitting nucleation potential increases to a
greater extent than the pitting passivation potential.
The passive current also decreases with increasing
alkalinity (Figure 5). The carbon steel specimens were
examined under an optical microscope after each pit-
ting scan measurement. The number of pits was larg-
est on the specimen exposed to 100 ppm alkalinity,
and decreases with increasing alkalinity. These re-
sults show the correlation between the pitting nucle-ation potential and passive current results. When the
film is weaker, the passive current on the pitting scan
is higher and pits can easily nucleate on the surface,
causing the low Enp and large number of pitting sites.
Pitting scan techniques have been widely used as a
predictive tool for characterizing the tendency of a
material to experience pitting corrosion. Although
valuable information can be obtained from these tech-
niques, they have their limitations. Used in a cooling
water environment on carbon steel, they are prima-
rily a measure of the resistance of an inhibitor film
to disruption by anodic corrosion. In conducting anexperiment, a tremendous electrical perturbation is
applied to the specimen, essentially destroying it as
a relevant metal specimen. This limitation makes this
technique a “snapshot” method which cannot follow
the corrosion and inhibition processes over time. This
method, like other potentiodynamic methods, char-
acterizes the tendency of a material in a medium to
experience corrosion under conditions which are dras-
tically different than those typically encountered in
service. Therefore, other techniques are needed.
Figure 3 — Effect of polymer addition in the all-organic inhibitor solution
Figure 2 — Effect of phosphonate on pitting scan
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
4/12
4
AC IMPEDANCE MEASUREMENTS
The AC impedance technique has been extensively
used to study corrosion phenomena5,6. By applying a
low-amplitude oscillating potential to the working
electrode, the perturbation of the electrode is mini-
mized and useful information about the electrochemi-
cal characteristics of the inhibitive and corrosion pro-
cesses on the surface is obtained. This technique is a
powerful tool for characterizing inhibitive films on
metal surfaces. In addition, it has been recently sug-
gested7 that AC impedance is applicable to the studyof localized corrosion. Mansfeld proposed that char-
acteristic patterns are observed in the impedance
spectra in the presence of pitting corrosion.
Experiments were done in order to determine the
effect of pitting corrosion on the impedance charac-
teristics of metal surfaces and to determine differences
in the film characteristics of two radically different
inhibitor programs. The programs studied were a
chromate-only inhibitor which was dosed at a marginal
level, and the previously-described phosphonate-
based all-organic program.
The chromate was dosed at a marginal level (200 ppm)
without the presence of zinc or any other inhibitors.
The pH of the chromate experiment was controlled
at 7 using controlled addition of carbon dioxide gas
into the air sparger, while the all-organic program
was tested at high alkalinity levels, equilibrated withair. The resulting pH was typically greater than 9.
Both tests used air-saturated solutions. All measure-
ments were performed at 120°F.
The measurements were performed on mild steel
rotating cylinder electrodes using a Princeton Applied
Research model (PAR) 273 potentiostat and a PAR
model 5301 lock-in amplifier. The potentiostat and
lock-in were controlled by an Apple II™ computer
using the PAR AC impedance software.
Figure 6 shows the Bode plot obtained at 1.5 hours
after immersion of the specimen into the chromate
solution. The figure shows only one peak, which isrelated to the double layer capacitance. In Figure 6,
the Bode plot taken after 24 hours of immersion is
also shown. This plot shows an additional small peak
in the high frequency range. This phenomenon is also
clear when the Nyquist plot (Figure 7) is examined.
The Nyquist plot shows a small semi-circle in the high
frequency range. This phenomenon appears to be
related to the formation of a primarily anodic passive
film on the metal surface. At 72 hours into the ex-
periment, another distortion appears on the Bode plot.
Figure 8 illustrates that the low frequency phase shift
does not approach 0 degrees. Rather, it decreases and
approached –32 degrees at 0.001 Hz. The shape of the Bode plot did not change after this time for up to
108 hours. The rotating disk electrode was examined
under an optical microscope, and the presence of pits
Figure 4 — Effect of alkalinity on Enp and E pp in theall-organic inhibitor solution
Figure 5 — Effect of alkalinity on passive current in
pitting scan curves in all-organic inhibitor solution
Figure 6 — Bode plot at 1.5 and 24 hours after immersion of the mild steel electrode into thechromate solution
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
5/12
• Rct is used as a measure of the general corrosion
rate
• Cfilm and Rfilm are the capacitance and resistance of
the inhibitor film
• Rpit is related to the pitting corrosion rate
• Cpit is the double layer capacitance over the pit
• W is a Warburg impedance which is due to diffu-
sional processes of oxygen or corrosion products
through the pit
The measurements taken in the chromate solution
indicate that a detectable film was formed at about
24 hours after the immersion of the specimen. This
is demonstrated by the appearance of the high fre-
quency peak. Figure 10 shows the open circuit
5
was observed. The low frequency phase distortion of
the Bode plot appeared to coincide with the forma-
tion of pits on the metal surface.
In order to analyze these results, the development of an equivalent circuit to model the impedance behav-
ior is useful. The various processes at work in the
metal/liquid interface can be visualized as elements
in the equivalent circuit. A proposed equivalent cir-
cuit model for the chromate results is shown in Fig-
ure 9. The physical and chemical processes analogous
to these elements are:
• Rs is the resistance of the solution
• Cdl represents the double layer capacitance
Figure 7 — Nyquist plot at 24 hours after immersionof the mild steel electrode into the chromate solution
Figure 8 — Bode plot at 72 hours after immersion of the mild steel electrode into the chromate solution
Figure 9 — Equivalent circuit for the impedanceresponse in the chromate solution
Figure 10 — Change in open-circuit potential in thechromate solution
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
6/12
potential throughout the experiment. The increasing
tendency of Ecorr indicates the formation of an anodic
passive film. At 72 hours, visible pitting was observ-
able and the low frequency phase distortion began to
appear in the Bode plot.
Results of the measurements taken with the all-
organic inhibitor are quite different than those obtained
from the chromate solution. Up to 48 hours into the
experiment, the Bode plot showed a single peak, and
the Nyquist plot only one semicircle (Figures 11 and
12). However, after 72 hours, the shape of the Nyquist
plot changed significantly. The low frequency region
is linear with a slope of about 45 degrees. Figure 13
shows that the 45-degree region spans the frequency
range of 0.02 to 0.005 Hz. No visible pitting was
observed on the specimen throughout the test. The
test was continued for 96 hours with the results shown
in Figure 14. The Bode plot shows that the region
showing Warburg characteristics was extended to
lower frequencies (0.002 Hz), indicating a thickening
of the cathodic film. Figure 15 shows the open circuit
potential measured throughout this experiment. The
decrease with time is indicative of the formation of a
cathodic passive film.
The impedance response observed with the all-organic
program is typical of a diffusion layer of finite thick-
ness. As the all-organic program is predominantly a
cathodic inhibitor8, an electron insulator film is
formed on the metal surface. Therefore, it seems prob-
able that the impedance behavior is caused by oxygen
diffusion through this finite layer.
OCCLUDED CELL MEASUREMENT
There are many similarities between pitting, under-
deposit corrosion and crevice corrosion. Often, local-
Figure 11 — Bode plot at 48 hours after immersion of the mild steel electrode in the all-organic inhibitor solution
Figure 12 — Nyquist plot at 48 hours after immersionof the mild steel electrode in the all-organic inhibitor solution
Figure 13 — Bode plot at 72 hours after immersion of the mild steel electrode in the all-organic inhibitor solution
Figure 14 — Bode plot at 96 hours after immersion of the mild steel electrode in the all-organic inhibitor solution
6
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
7/12
ized corrosion contains elements of all three forms.
The test methods, which have been previously de-
scribed, all start with a clean surface and measure
some element of the resistance of passive films to
breakdown by anodic corrosion. In the case of
underdeposit corrosion, the ability of an inhibitor to
act may be limited by its ability to diffuse to a sur-
face through a porous deposit and by its ability to act
in the environment resulting from the localized cor-
rosion process. Although the previously discussed test-
ing methods simulate this process to some extent, thespecific geometry through which diffusion must oc-
cur is somewhat a matter of chance.
An occluded cell apparatus was discussed by Krisher8.
In order to specifically address the issue of inhibitor
mobility and effectiveness, a similar apparatus was
constructed to allow for variation of cell geometries,
simulating various pit depths. This apparatus is a
standard laboratory recirculator apparatus (Figure 16)
which contained an electrode assembly (Figure 17)
consisting of a free electrode and occluded electrodes
in various configurations.
The free electrode (cathode) was a section of carbonsteel tube. The occluded cells (anodes) were carbon steel
cylinders imbedded in a nonconductive plastic block.
The depth of the anodes was adjusted by using dif-
ferent lengths of plastic tubing inside the cells. Three
different cell depths were used with three replicates
of each cell depth. The depths of the shallow, medium
and deep cells were 4, 11 and 18mm, respectively.
The anodes were polished with #600 grit emery paper
and rinsed with acetone before they were inserted in
the occluded cells. The cell box was filled with test
solution and mounted in the recirculating loop. The
temperature of the recirculating solution and the pH
of the solutions were controlled by automatic control-lers. To maintain the inhibitor concentration, makeup
water was pumped continuously into the basin and
out an overflow. Inhibitor was added proportionately
to account for the overflow by a syringe pump. The
current between the anodes and cathode was moni-
tored by zero resistance ammeters and a computer
data acquisition system.
Three different cooling water inhibitor programs were
tested using this device. They were chosen to represent
a spectrum of inhibitors currently in wide commercial
use and a variety of inhibitor types. They included
the phosphonate all-organic, a chromate program, and
a neutral pH-stabilized phosphate program. The
conditions of the three experiments are shown in Table 1.
Figures 18, 19 and 20 show typical results obtained
from these experiments. The reproducibility of the
currents from the replicate anodes was quite good.
The passivation time of the occluded electrodes in the
all-organic and stabilized phosphate programs varied
depending on the depth of the cells. In the all-organic
experiment, the shallow cells took about 80 hours for
passivation to occur. The medium and deep cells took
Figure 15 — Change in the open circuit potential in
the all-organic inhibitor solution
Figure 16 — Recirculator apparatus for the occludedcell experiments
Figure 17 —Occluded cell assembly
7
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
8/12
8
140 and 200 hours respectively. The stabilized phos-
phate program took less time to achieve steady state
currents. The current from the shallow cells leveled
out at about 15 hours. The medium and deep cells
required 60 and 80 hours respectively. The steady-
state currents observed in the all-organic program
were far lower than those seen in the phosphate pro-gram. The average current from the deep occluded
cells after passivation was approximately 0.7 µ A for
the all-organic program and typically 9 µ A for the
stabilized phosphate program. The passivation times
observed with the chromate program were relatively
insensitive to the depth of the cell. All three depths
gave sudden and effective passivation after 75 hours.
After completion of the experiment, the anodes were
microscopically examined. The exposed area of the
anodes from the stabilized phosphate program
showed significant visible etch with depth. The
visible effects on the anodes from the other two pro-
grams were significantly less, with little visible etch.
DISCUSSION
The typical bulk cooling water corrosion environment
consists of air-saturated water at neutral to slightly
alkaline pH with varying amounts of hardness and
other solutes in the presence of corrosion inhibitors.
The metal surface environment is typically carbon
steel of varying cleanliness with numerous small
niches and crevices. These are obvious starting points
for localized corrosion. However, under these condi-
tions, even initially clean carbon steel equipment canexperience underdeposit corrosion. Defects in passive
films or variations in metal composition can cause
local dissolution of iron. The primary product of this
reaction is Fe+2. Compounds of Fe+2 are quire soluble
under these conditions. However, once the Fe+2
reaches the oxygenated bulk water, it is quickly
oxidized to Fe+3. There are numerous oxides and other
components of this ion which are very insoluble and
quickly precipitate. Thus, what starts out as a pit,
quickly acquires a layer of deposit. In the case of
Table 1 — Operating conditions for occluded cell experiments
Inhibitor program
All-organic stabilized phosphate Chromate
Product concentration (ppm) 200* 100** 300***
pH 9* 7** 7***
Temperature (°F) 120* 100** 120***
Ca (ppm as CaCO3) 360* 360** 360***
Mg (ppm as CaCO3) 200* 200** 200***
Alkalinity (ppm as CaCO3) 350* 50** 50*** Conductivity (microsiemens) 2200* 2200** 2200***
***corresponds to 12 ppm phosphonates as PO 4
* **corresponds to 21 ppm total phosphate as PO
4
***corresponds to 300 ppm chromate as chromate
carbon steel, pitting and underdeposit corrosion are
intimately related.
The local environment in an actively corroding
underdeposit site is significantly different than that
in the bulk water, in that:
• Oxygen is depleted• Fe+2 is the predominant iron form
• The pH is very low
• Ionic concentrations differ depending on electro-
phoretic mobilities
There are a number of issues which determine the
ability of an inhibitor to prevent and passivate
underdeposit corrosion. These are:
• The quality and integrity of the inhibitor film and
its ability to resist attack
• The presence of components in the inhibitor whichcan effectively passivate in the underdeposit
environment
• The ability of these components to diffuse to the
corroding site
The three experiments presented focus on various as-
pects of the problem. The pitting nucleation potential,
Enp, measured by the modified pitting scan, is an in-
dication of the resistance of preformed inhibitor films
to rupture by anodic dissolution of iron. In the two
examples presented, increasing the alkalinity and
including an appropriate polymer had beneficial ef-
fects on film strength. Increasing the alkalinity alsoincreased the pitting passivation potential Epp, indi-
cating an improved ability to passivate existing de-
fects in the film.
The AC impedance method provides a great deal of
information about the characteristics of the film. Low
frequency phase distortion in the Bode plot appears
to correlate well with the appearance of pits. The
appearance of a Warburg-like characteristic consist-
ing of a linear 45 degree region in the all-organic im-
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
9/12
9
pedance spectrum is indicative of the formation of a
diffusion layer of finite thickness. This provides evi-
dence of the formation of a well-defined cathodic film
on the mild steel electrode. The extension of the fre-
quency range of the Warburg-like region corresponds
to thickening of this layer. The formation of an an-
odic passive film by the chromate program was indi-
cated by the appearance of a small impedance peak
at high frequency, corresponding to a capacitance-re-sistance type of circuit.
The occluded cell experiment emphasizes the ability
of an inhibitor to diffuse to a remote surface and to
function in the oxygen-depleted, low pH, Fe+2-rich
environment encountered under deposits.
The ability of the chromate program to act as an anodic
inhibitor in the absence of oxygen and to promote the
conversion of Fe+2 to Fe+3 probably accounts for its ef-
fectiveness in this test. The insensitivity to pit depth
in this experiment indicates that this process is not
diffusion-controlled. The potential of the occluded
electrode is determined by the IR drop between the
free electrode and the occluded electrode, where I is
the current between the free electrode and R is mainly
solution resistance. The occluded electrode potential
can then be expressed as follows:
Ecell = Eo – IR
where Ecell is the potential of the occluded electrode
and Eo is the open-circuit potential of the free elec-
trode. Since Eo increases (Figure 10) and I decreases
(Figure 20) with time, Ecell increases as the occluded
cell measurement proceeds. This change in Ecell af-
fects the passivation of the occluded electrode sincethe solution contains chromate ions. When Ecellreaches the critical potential for passivation, the cur-
rent drops because of the passivation of the occluded
electrode. This decrease in current causes a further
increase in Ecell. Figures 21 and 22 show this process.
In Figure 21, Ecell is lower than the critical potential
for passivation; therefore, the corrosion rate is very
high. In Figure 22, Ecell increases to above the critical
potential for passivation and the resultant corrosion
rate is very low.
The passivation of the occluded cells in the two
nonchromate programs was different from that ob-
served with the chromate program. These programs
function primarily as cathodic inhibitors2 in contrast
to the anodic character of the chromate program. The
difference between the passivation currents observed
in the stabilized phosphate and all-organic programs
is probably due to the great difference in the alkalin-
ity at which these programs are operated. At a pH of
7, the alkalinity is typically lower than 50 ppm
(CaCO3). At a pH of 9, the alkalinity will be in the
400-500 ppm range.
Figure 18 — Results of occluded cell measurement inthe all-organic inhibitor solution
Figure 19 — Results of occluded cell measurement inthe stabilized phosphate inhibitor solution
Figure 20 — Results of occluded cell measurement inthe chromate solution
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
10/12
Alkalinity can influence the underdeposit environ-
ment in two ways. First, the increased buffer capacity
afforded by a high alkalinity will tend to minimize
the pH depression at an anodic occluded site, caused
by hydrolysis by ferrous and ferric ions. Secondly, thepresence of the CO3= ion provides the opportunity for
the precipitation of FeCO3(Siderite). While phosphate
could potentially form Fe3(PO
4)
2(Vivianite), this re-
action appears to be less effective in passivating
occluded cells. The Siderite model of corrosion inhi-
bition in drinking water circuits has been extensively
discussed by Sondheimer et al.9 They propose that it
is an important intermediate phase which undergoes
isomorphous transformation into oxidized iron phases
such as Geothite.
CONCLUSIONThe conditions under which underdeposit corrosion
occurs are different from those controlling generalized
corrosion. Issues of inhibitor diffusivity and ability to
control corrosion in deoxygenated environments be-
come important. The test methods used to evaluate
inhibitors should take these factors into account. The
experiments described in this study were intended
to provide information about various aspects of local-
ized corrosion such as:
Figure 21 — Occluded cell potential and currentbefore passivation
• The resistance of inhibitor films to pitting attack
• The ability of inhibitors to stop crevice and under-
deposit corrosion once it has begun
• The effect of water chemistry on localized corrosion• The kinetics of pit passivation
The described three localized corrosion measurement
techniques provided important interrelated informa-
tion. The major findings of the study are as follows:
1. In the all-organic program, phosphonates (which
comprise the major corrosion inhibitor in the pro-
gram) did not decrease the passive current in the
modified pitting scan curve confirming that they
are cathodic inhibitors. However, Enp increased
about 100mV, indicating a pitting inhibition prop-
erty of the chemical.
2. The polymer in the all-organic program improved
the protective property of the passive film and in-
creased pitting nucleation potential.
3. The cathodic passive film formed in the all-organic
program acted as a finite thickness diffusion layer
and was detected as a Warburg-type impedance in
the impedance measurements. The thickening of
the film with time was detected by the measurements.
Figure 22 — Occluded cell potential and current after passivation
10
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
11/12
4. Higher alkalinity and pH of a mainly cathodic in-
hibitor seemed to be beneficial for the passivation
of the occluded cells.
5. The anodic passive film formed in the chromate
solution was detected by impedance measurements
as a parallel capacitance-resistance type of circuit
at the high frequency range.
6. The pitting formation by the underdosed chromateprogram was detected by impedance measure-
ments at the low frequency range.
7. The passivation of the occluded electrodes in an
anodic inhibitor seemed to be related to the poten-
tial of the electrode determined by the open circuit
potential of the free electrode and the IR drop.
REFERENCES
1. Z. Szklarska-Smialowska, Pitting Corrosion of
Metals, National Association of Corrosion Engi-
neers, Houston, Texas, 1986.
2. G. Bohnsack, K.H. Lee, D.A. Johnson, E. Buss, “In-
vestigation of the Mechanisms of Organic Com-
pounds Used in Cooling Water Corrosion Control,”
Materials Performance, Vol. 25, No. 5, p.32, 1986.
3. K.E. Fulks, A.M. Yeoman, “Performance Evalua-
tion of Non-Metal Cooling Water Treatments,” pre-
sented at Corrosion/83, Anaheim, California, April,
1983.
4. J.E. Hoots, G.A. Crucil, “Role of Polymers in the
Mechanisms and Performance of Alkaline Cooling
Water Programs,” Materials Performance, Vol. 26,
No. 4, p. 17, 1987.
5. D.D. Macdonald, “Theoretical Analysis of Electro-
chemical Impedance,” Corrosion/87, Paper No. 479,
National Association of Corrosion Engineers, Hous-
ton, Texas, 1987.6. I. Epelboin, C. Gabrielli, M. Keddam, H. Takenouti,
“Alternating-Current Impedance Measurements
Applied to Corrosion Studies and Corrosion-Rate
Determination,” Electrochemical Corrosion Test-
ing, STP 727, ASTM, Philadelphia, Pennsylvania,
p. 150, 1981.
7. F. Mansfield, presented at the International Con-
ference on Localized Corrosion, Orlando, Florida,
1987.
8. A.S. Krisher, “A Synthetic Crevice Device to Moni-
tor Corrosiveness of Aqueous Systems,” Corrosion/
81, Paper No. 153, National Association of Corro-
sion Engineers, Houston, Texas, 1981.
9. H. Sontheimer, W. Kolle, A. Kuch, Internal Cor-
rosion of Water Distribution Systems, AWWA
Research Foundation, Denver, Colorado, 1985.
11
-
8/18/2019 Pitting Corrosion Due to Deposits in Cooling Water Systems
12/12
N A L C O C H E M I C A L C O M P A N YONE NALCO CENTER NAPERVILLE, ILLINOIS 60563-1198
Registered Trademarks of Nalco Chemical Company Printed in U.S.A. 12-88