Pneumatic Seal II - Fail Safe Access Control Options
Standard
If power fails, RF
Lock-seal will release
and door can be
opened. Release
time is less than 1
second.
Door System with
Battery Back-up
If power fails, Battery will
keep RF Lock-seal
functional to complete
scan and operate for 24-
48 hours. Access controls
will remain operational as
long as air remains in
compressor.
Door system on
Emergency Power
If power fails, RF Lock-
seal will remain
functional and access
controls will remain
operational as long as air
remains in compressor.
Compressor and
Door system on
Emergency Power
RF Lock-seal will remain
functional and all access
controls will remain
operational.
Keypad Access
Enter code to release the RF Lock-seal enable
door to be opened. Several programming
techniques available, including multiple user
codes and lockout codes.
Push Button
Entrance
Push button will release the RF Lock-seal and
unlock the door.
Push Button
Entrance with
Keyed Lock-out
Switch
Key will enable or disable push button. Once
switch is in on position the push button will
release the RF Lock-seal and unlock the door.
3rd Party Card
Reader Control
3rd Party Card Reader will provide 24V power
to seal and release the RF Lock-seal and
unlock the door.
1730 E. Pleasant Street Noblesville, IN 46060
Bus: 317.773.8500 Fax:317.773.8508 Web: www.imedco.net
Pneumatic Seal II - Emergency Release, Request to Exit and Features
Emergency Open
Exterior
The Exterior Emergency Open button will
remove power from the door and release RF
Lock-seal until reset. Audible alarm will
sound when cover is lifted to alert staff that
the lock system is being bypassed.
Emergency Exit
Interior
The Interior Emergency Exit button will
remove power from the door and release RF
Lock-seal until reset. Audible alarm is
disabled by default.
Push to Exit Button
Interior
Located on the interior side of the door, this
Push to Exit will release RF Lock-seal from
the inside if door is closed.
Door Position Switch
Inflation time less
than 2 seconds
When door is closed and makes contact with
the door frame, sensor will activate and
engage the RF Lock-seal, placing the door in
a secured condition.
Compressor
Included with system
Airlines are run by IMEDCO to Door System
controls.
Compatible with most
hospital air systems
Acceptable range of supply air pressure is 50
to 120 PSI.
1730 E. Pleasant Street Noblesville, IN 46060
Bus: 317.773.8500 Fax:317.773.8508 Web: www.imedco.net
Pneumatic Seal II – Optional Features
Compatible with FerrAlert Halo II Plus Ferrous Metal Detection
FerrAlert
Integration in
Pneumatic
Seal II –
Inswing
FerrAlert
Integration in
Pneumatic Seal II –
Outswing
Compatible with IV Pass-Thru Waveguide integrated into RF Door System
IV Pass-through allows patients to remain
connected to a device that is left outside
the MRI room while scanning. A
secondary seal at the IV pass through
retracts each time the RF Lock-seal is
released to allow for easy insertion of
lines. Up to four lines can be connected at
one time.
Compatible with Automatic Opener/Closer
Fully automated door allows for hands
free operation.