polarimeters - meadowlark optics · 2014. 5. 5. · recalibration of your polarimeter. custom...
TRANSCRIPT
-
(303) 833-4333(303) 833-4335 Fax
5964 Iris Parkway, PO Box 1000Frederick, Colorado 80530 USA© 2008 Meadowlark Optics, Inc. All rights reserved.
Polarimeters
Meadowlark Optics has a longstanding reputation for precision polarization control and metrology. Our line of user-friendly Polarimeters provides this same high accuracy and reliability in an easy to use instrument, suitable for manufacturing and laboratory applications.
These Polarimeters are compact, state-of-the-art systems with convenient computer control that accurately measures Stokes Parameters 10 times per second that quantify the State of Polarization (SOP) and then graphically displays this state on the Poincaré Sphere, Polarization Ellipse, or running chart. It contains no spinning waveplates, motors, or other moving parts to wear or cause vibrations. Patented, proprietary algorithms provide high accuracy and calibration versatility. Two models offer calibrated polarization measurement over the wavelength ranges 450 nm to 1100 nm and 900 nm to 1700 nm. Meadowlark Optics Polarimeters are computer-controlled with USB interface for increased measurement rate and convenience.
Applications:• Ellipsometry • TestandMeasurement • MaterialsCharacterization • SpectroscopicApplications • TelecomDeviceManufacturing• PharmaceuticalDevelopmentandQC• StokesVectorAnalysis
PolarView3000UserInterfaceSoftwareFeatures:• DataAveraging • StokesParameters• PolarizationEllipse• DegreeofPolarization • ContinuousSaveToFile • ContinuousChartdisplayOption• PoincaréSpherewith2DProjections• ContinuousTrackingOnPoincaréSphere
PolarimeterPackageincludes:• TemperatureControlledLCPolarimeterOpticalHead• D3000DigitalElectronicController• PolarView3000softwareinstallationCD• UserManual • USBandopticsheadcables• PowersupplyandPowercable• Calibrationatuptothreewavelengthsincluded(Storageinmemoryofupto12wavelengthspossible)• LabView™VirtualInstrumentLibraryforPolarimeter• Oneyearofsoftwaresupportandupgrades• Oneyearwarrantyfordefectsinmaterials andworkmanship
-
(303) 833-4333(303) 833-4335 Fax
5964 Iris Parkway, PO Box 1000Frederick, Colorado 80530 USA© 2008 Meadowlark Optics, Inc. All rights reserved.
Polarimeters
KeyBenefits• Compact• Nomovingparts • StokesVectorAccuracybetterthan1%• Polarizationstatetracking • SoftwareInstrumentLibrary• Fiberandfreespaceversions • User-friendlygraphicalinterface• Visibleandnearinfraredversions
LabVIEW™ and National Instruments™ are trademarks of National Instruments Corporation. Microsoft® and Windows™ are trademarks of Microsoft Corporation.
SPECIFICATIONS
WavelengthRanges(nm)VisibleNearInfrared
450-1100 900-1700
MaxAbsoluteDegreeofPolarizationError
≤ 1%
WavelengthRangeforCalibrationAccuracy
± 3 nm
MeasurementFrequency 10 Hz
Resolution 0.001 of a Stokes Parameter
MaximumOperatingTemperature 40° C
OpticalHeadDimensions 2.83 x 1.75 x 1.75 in.
MinimumOpticalPowertomaintainaccuracy
10 mW sensitivity can be increased by special request
PowerRequirements 100 – 240 V ac47 – 63 Hz500 mA
InputAperture 2 mm
These specifications describe performance at 23 ± 3° C ambient temperature.
Minimum System Requirements: • PC with Pentium II class processor • Windows™ 98/ME/2000/XP/Vista • CD ROM drive • 20 MB hard drive space • 32 MB RAM • USB Port • Use of LabVIEW™ Instrument Library requires LabVIEW version 6.1 or newer full development system. system.
ORDERINgINFORMATION
Wavelength Range (nm)
Version Part Number
450-1100 Visible PMI - VIS
900-1700 Near Infrared PMI - NIR
OptionalAccessories:• EigenstategeneratorSet*• Fiberopticcableadapter -Visible -NearInfrared• Angle-polishedfibercollimatoradapter • Calibrationatadditionalwavelengths
*TheEigenstategeneratorSetenablesprecise recalibrationofyourPolarimeter.
Custom versions are available.
Please contact our Sales Department to obtain a price list of standard Meadowlark Optics components.
1.769
1.75
0.875
0.909
1.769
2.73
Front
0.83
0.8651.00
0.50
Tapped 8-32 MountingHole Locations
2mm
Mounting Holes onCenterline of Head