powerpoint presentation › webassets › cms › library › docs › 2018asm… · xevo tq-s the...
TRANSCRIPT
Comparison between the tandem quadrupole and quadrupole time-of-flight quantification of exenatide in rat plasma Gordon J. Murray1, Jake Y. Hsu2 & David A. Johnson2
1Waters Corporation, 100 Cummings Center, Beverly, MA 01915, 2MicroConstants, 9050 Camino Santa Fe, San Diego, CA 92121
Sample extraction
Blank rat plasma (20 µL, K2EDTA) was spiked with
exenatide to produce concentrations ranging from 0.1 to 20
ng/mL. Deuterated exenatide (exenatide –d5) was added as
the internal standard. The samples were extracted using
solid phase extraction (SPE) on a Waters Oasis® MAX µ-
elution plate. The final SPE eluents were diluted 1:1 with DI
water and analyzed by reversed phase UPLC® coupled to
either a tandem quadrupole MS or to a bench-top QToF MS
system.
LC/MS and Data Processing
Data were collected on a Waters ACQUITY I-Class coupled
to either a Waters Xevo ® TQ-S or a Waters Xevo ® G2-XS
QToF (Figure 2)
LC Column: 3 µm, 2.1 x 50 mm HALO C4 at 65 oC
Mobile phase A and B: DI water and acetonitrile + 0.1%
formic acid
Table 1. LC gradient
Instruments: Xevo TQ-S or Xevo G2-XS operating in
positive electrospray mode
Acquisition software: Data was acquired using MassLynx
v4.1 software and processed using the TargetLynx
application manager.
Data acquisition: Xevo TQ-S data was acquired using
multiple reaction monitoring (MRM), 838.4 > 948.8
(exenatide) and 839.4 > 950 (exenatide–d5). Xevo G2-XS
data was acquired using ToF-MRM 838 > 948.4543
(exenatide) and 839 > 950.2135 (exenatide-d5)
Methods Introduction
The development of biotherapautic molecules continues to
be an attractive avenue of research with new biotherapautic
drug approvals outpacing traditional small molecule
therapeutics. One such biotherapeutic molecule, Exenatide,
is a synthetic version of Exendin-4, a 39-amino acid peptide
(MW 4186.6 Da) found in the saliva of the Gila monster and
is approved for the treatment of type 2 diabetes. (Figure 1).
Figure 1. Amino acid sequence of exenatide
Traditionally, exenatide plasma concentrations have been
determined be immunoassay (ELISA), with more recent
advances focusing on tandem quadrupole liquid
chromatography (LC-MS/MS) techniques. LC-MS/MS
methodologies offer significant advantages over
immunoassays such as greater dynamic range, specificity
and repeatability, lower cost and shorter assay
development times. However, the specificity gained gained
by using LC-MS/MS may not be sufficient to effectively
resolve the target analyte from endogenous interfering
species. In this regard, high resolution mass spectrometry
(HRMS) can offer advantages over the gold standard LC-
MS/MS quantitative approach, while maintaining significant
versatility to be effectively utilized in more qualitative
assays, such as, metabolite identification, peptide
catabolism, metabolomics etc.1,2 Here, we present a
comparison between a LC-MS/MS method and an accurate
mass quadrupole time-of-flight (QToF) LC/MS quantitative
assay for the analysis of exenatide in rat plasma.
Figure 2. Schematic of the Xevo G2-XS QToF
Time (mins) Flow rate (mL/min) A (%) B (%)
0.00 0.3 72 28
0.50 0.3 72 28
3.00 0.3 65 35
3.50 0.3 5 95
3.51 0.5 5 95
4.50 0.5 5 95
4.51 0.3 5 95
5.00 0.3 72 28
6.00 0.3 72 28
Xevo TQ-S
The LOQ obtained from the TQ-S tandem quadrupole MS
system was 100 pg/mL from a 25 µL injection (Figure 3 and
Table 2). The curve was linear from the LOQ to 20 ng/mL
(highest level tested) (Figure 4).
Table 2. Batch report for exenatide quantification on Xevo TQ-S
Figure 3. Blank, LOQ and ULOQ chromatograms from Xevo TQ-S
Figure 4. Calibration curve from Xevo TQ-S
Results – QQQ MS
Xevo G2-XS QToF
The LOQ obtained using ToF-MRM data was 100 pg/mL
from a 25 µL injection (Figure 5 and Table 3). The curve was
linear from the LOQ to the upper level tested (20 ng/mL)
(Figure 6).
Table 3. Batch report for exenatide quantification on Xevo G2-XS QToF
Figure 5. Blank, LOQ and ULOQ chromatograms from Xevo G2-XS QToF
Figure 6. Calibration curve from Xevo G2-XS QToF
Results – QToF MS
We have shown that QQQ and QToF platforms can exhibit
near equivalent sensitivity in the analysis of exenatide from
rat plasma. This assay could be expanded to utilize micro-
flow technology to further enhance sensitivity.
Conclusions
ng/mL-0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0
Response
-0.00
2.00
4.00
Sample ID Std. Conc.
(ng/mL) RT Peak Area ISTD Area
Response
Ratio
Calc. Conc.
(ng/mL) %Dev S/N
Matrix Blank_1
W1_1 0.1 2.33 15 348 0.04 0.09 -5.8 26
W1_2 0.1 2.34 18 418 0.04 0.10 -4.4 26
W2_1 0.2 2.33 26 346 0.07 0.20 1.9 27
W3_1 0.5 2.34 63 346 0.18 0.58 16.4 93
W4_1 1.0 2.34 74 260 0.28 0.94 -6.2 79
W5_1 2.0 2.34 275 474 0.58 1.98 -0.8 424
W6_1 5.0 2.34 622 437 1.42 4.95 -0.9 1022
W7_1 10.0 2.34 1229 433 2.84 9.93 -0.7 1496
W8_1 20.0 2.34 2280 386 5.91 20.76 3.8 3639
W8_2 20.0 2.34 2347 426 5.51 19.35 -3.2 1954
min1.00 2.00
%
0
100
Matrix Blank_1
0.58
0.902.77
2.50
min
%
0
100
Matrix Blank_1
0.84
2.3418.71
1.33 2.65
min1.00 2.00
%
0
100
W1_2 2.33418.36
min
%
0
100
W1_2 2.3418.13
1.280.51
2.11 2.62
min1.00 2.00
%
0
100
W8_1 2.33385.61
min
%
0
100
W8_1 2.342280.16
Blank 0.1 ng/mL 20 ng/mL
Exenatide
Exenatide
-d5
Sample ID Std. Conc.
(ng/mL) RT Peak Area ISTD Area
Response
Ratio
Calc. Conc.
(ng/mL) %Dev S/N
Matrix Blank_1
W1_1 0.1 2.39 436 32786 0.01 0.10 -3.4 95
W1_2 0.1 2.39 420 31455 0.01 0.10 -2.2 650
W2_1 0.2 2.39 718 29028 0.03 0.21 -5.7 2061
W3_1 0.5 2.39 1860 30416 0.07 0.56 11.4 805
W4_1 1.0 2.39 3315 29456 0.12 1.05 5.2 2616
W5_1 2.0 2.39 6723 31388 0.23 2.01 0.6 2804
W6_1 5.0 2.39 16466 30782 0.58 5.17 3.4 6575
W7_1 10.0 2.39 27843 27326 1.11 9.93 -0.7 2850
W8_1 20.0 2.40 45468 24803 2.01 17.95 -10.2 8431
W8_2 20.0 2.39 63796 34637 2.02 18.06 -9.7 9564
Blank 0.1 ng/mL 20 ng/mL
Exenatide
Exenatide
-d5
min2.00 4.00
%
0
100
AR01056
2.24
2.09
2.45
2.88
min
%
0
100
AR01056
2.38
2.99
min2.00 4.00
%
0
100
AR01054
2.3732785.64
min
%
0
100
AR01054
2.39435.66
2.86
min2.00 4.00
%
0
100
AR01001
2.3824803.29
min
%
0
100
AR01001
2.4045467.95
1. Dillen, L. et al. Bioanalysis 4 (5), 565-579
(https://www.future-science.com/doi/pdf/10.4155/bio.13.87)
2. Morin, L-P. et al. Bioanalysis 5 (10), 1181-1193 (https://www.future-science.com/doi/pdf/10.4155/bio.12.3
References
ng/mL-0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0
Response
-0.00
0.50
1.00
1.50
2.00