Practical Faster

GC Applications

with Capillary

GC Columns

Simon Jones

Application Engineer

June 10, 2010

Page 2: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 2

Questions to Ask

What information do you need from your analysis?

Do you have more baseline than you need between your

peaks?

Do you need to resolve all of the components?

Page 3: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 3

•Stationary Phase

•Temperature Programming

•Carrier Gas: type and linear velocity

•Shorten Column Length

•Decrease Film Thickness

•Decrease Internal Diameter

Variables for Shortening Run Times

Page 4: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 4

Resolution

N = (gas, L, rc)

k = (T, df, rc)

a = (T, phase)

RN k

ks =

4 1

1

a

a

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 5: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 5

Resolution

RN k

ks =

4 1

1

a

a

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

N = (gas, L, rc)

k = (T, df, rc)

a = (T, phase)

Page 6: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 6

Stationary Phase - Common Types

Siloxane polymers

Poly(ethylene) glycols

Porous polymers

Page 7: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 7

Selectivity

Relative spacing of the chromatographic peaks

The result of all non-polar, polarizable and polar

interactions that cause a stationary phase to be

more or less retentive to one analyte than another

Page 8: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 8

Optimizing Selectivity

Match analyte polarity to stationary phase polarity

-like dissolves like (oil and water don’t mix)

Take advantage of unique interactions between analyte and

stationary phase functional groups

Page 9: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 9

Start with the Right Phase

Time (min.)

0 2 4 6 8 10 12 14

Time (min.)

DB-1

15m x 0.32mm, 0.25µm

Oven:

40°C for 2 min

40-120°C at 5°C/min

0 1 2 3 4 5

DB-Wax

15m, 0.32mm, 0.25µm

Oven:

80-190°C at 20°C/min

Page 10: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 10

Resolution

N = (gas, L, rc)

k = (T, df, rc)

a = (T, phase)

RN k

ks =

4 1

1

a

a

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 11: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 11

Column Length and Efficiency (Theoretical Plates)

0.25 mm ID

n/m = 4630 (for k = 5)

Length (m) n

15 69,450

30 138,900

60 277,800

Page 12: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 12

Column Length and Resolution

Length X 4 = Resolution X 2

R a n aL

t a L

Page 13: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 13

Column Length VS Resolution and Retention:

Isothermal

Double the plates, double the time

but not double the the resolution

15 m 60 m30 m

R=0.84

2.29 min

R=1.68

8.73 min

R=1.16

4.82 min

Page 14: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 14

BTEX

Carrier: Helium, 36 cm/sec at 40°c

Oven : 40°C for 3 min, 5°/min to 100°C

1. Benzene

2. Toluene

3. Ethylbenzene

4. m,p-Xylene

5. o-Xylene

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.00.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00

1

2 3

4

5

DB-5

30 m

0.53 mm I.D., 0.5 µm

DB-5

15 m

0.53 mm I.D., 0.5 µm

0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.801.0 2.0 3.0 4.0 5.0

1 2 3

4

5

6.0

DECREASE THE LENGTH

Page 15: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 15

Column Length and Cost

15m30m

60m

$ $ $ $ $ $ $

Page 16: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 16

Length Summary

If you Decrease Length:

Efficiency Decrease

Resolution Decrease

Analysis Time Decrease

Pressure Decrease

Cost Decrease

Page 17: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 17

Resolution

N = (gas, L, rc)

k = (T, df, rc)

a = (T, phase)

RN k

ks =

4 1

1

a

a

Efficiency

Retention

Selectivity

L = Length

rc = column radius

df = film thickness

T = temperature

Page 18: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 18

Column Diameter - Theoretical Efficiency

I.D. (mm)

0.05

0.10

0.20

0.25

0.32

0.45

0.53

0.18

n/m

23,160

11,580

5830

4630

3660

2840

2060

6,660

30 m

20 m

10 m

k = 5

5 m

N ~ 112,000

N ~ 112,000

N ~ 112,000

N ~ 112,000

Total Plates

Page 19: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 19

Different Column I. D. Equal Phase Ratios

Time (min)

0 5 10 15 20

Carrier:Oven: 65°CInjection: SplitDetector: FID

Column: DB-62430 m, 0.32 mm, 1.8 m

Helium, 40(cm/sec)

Column: DB-62430 m, 0.53 mm, 3 m

Page 20: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 20

PHASE RATIO ()

Film Thickness

Column Dimensions Phase Ratio β

30 m x .53 mm x 3.0 m 44

30 m x .32 mm x 1.8 m 44

KC = k β

r

2df

β =

Page 21: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 21

Column Diameter and Capacity

Like Polarity

Phase/Solute

0.25 µm film thickness

I.D. (mm) Capacity (ng)

0.05 1-2

0.18 25-55

0.20 35-70

0.25 80-160

0.32 110-220

0.53 1000-2000

0.45 600-800

0.10 6-13

Page 22: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 22

Column Diameter - Inlet Head Pressures (Helium)

30 meters

Hydrogen pressures x 1/2

I.D (mm) Pressure (psig)

0.10 90-130

0.20 25-40

0.25 15-25

0.32 10-20

0.45 3-7

0.53 2-4

0.18 30-45

0.05 275-400

Page 23: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 23

Column Diameter and Carrier Gas Flow

Lower flow rates: Smaller diameter columns

Higher flow rates: Larger diameter columns

Low flow rates : GC/MS

High flow rates: Headspace, purge & trap

Page 24: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 24

Diameter Summary

If you decrease the inside diameter:

Efficiency Increase

Resolution Increase

Pressure Increase

Capacity Decrease

Flow rate Decrease

Page 25: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 26

Carrier Gas Considerations

Best velocity?

Optimal range of velocities

Too low or high results in loss of resolution

Balance resolution and analysis time

BUT... there is no LAW against going FASTER

Page 26: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 27

60

Excessive Diffusion

Poor Mass Transfer

Carrier Gas Considerations

Van Deemter Curve

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

H (

min

)

He

N2

H2

uopt

Page 27: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 28

Carrier Gas ConsiderationsUtilizing Computer Controlled Flow Ramping

(EPC)

Lonely Late Eluters? = P

Decrease retention and overall run time by

increasing pressure

(Speed up the gas!)

Page 28: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Group/Presentation Title

Agilent RestrictedMonth ##,

200X

Fast Detector Electronics – 6890 FID

200 Hz

100 Hz

50 Hz

20 Hz

10 Hz

Proper peak quantitation typically requires a minimum of 10 data points per

peak

10 Hz data rate: (peaks less than one second wide will have less than 10

data points)

200 Hz data rate: (peaks as narrow as 50 msec can be handled reliably)

Page 29: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 30

Easy Options with Method Translation Software

Different Column Dimensions

Switch He to H2 Carrier Gas and Try Faster Velocities

Same Column & Gas Type but Faster Velocities

Combination of all of the above

Page 30: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Group/Presentation Title

Agilent RestrictedMonth ##,

200X

Temperature Programming

30m, 0.25mm ID

Temperature

10m, 0.1mm ID

Temperature

Temperature program must be modified to give same

temperature of elution (i.e. faster ramps, shorter hold

times)

Page 31: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 32

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

22

21

20

19

18

17

16

1514

13

12

11

10

9

8

7

5,6

4

3

2

1

Column: DB-XLB

30m x 0.32mm i.d., 0.25µm

Carrier: He, constant flow, 38 cm/s at 120°C

Injector: Pulsed Splittless, 220 °C

Pulse pressure & time: 35psi for 1.15min

2µL, 50ppb

Oven: 120°C for 1.17min

120°C to 160°C at 25°/min

160°C to 260°C at 10°/min

260°C to 300°C (4min) at 15°/min

Detector: µ-ECD, 320°C

Ar/CH4 (P5) makeup gas at 60mL/min

<16 minutes

CLP-Pesticides - Original “Improved” Method

0.32mm I.D., Helium Carrier Gas

Page 32: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 33

Input Original Method Parameters

Page 33: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 34

New Velocity

New Temp.

Program

Same Column and Gas Type

Fast Analysis (OPGV)

Transfer same

dimensions

Page 34: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 35

Only Time!

<14 minutes

No Real

Changes in

Resolution!

Same Column and Gas Type

Fast Analysis (OPGV)

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

Before

Page 35: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 38

Same Column, Hydrogen Carrier Gas

Translate Only

Page 36: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 39

Same Column, H2 Carrier Gas

Translate Only

33% faster!

11 minutes

No Real

Changes in

Resolution

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

Before

Page 37: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 42

New Column Dimensions, H2 Gas, Translate Only

Input NEW

dimensions

New Velocity

New Temp.

Program

Close enough

(got lucky)

Page 38: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 43

New Column Dimensions, H2 Gas, Translate Only

Page 39: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 44

10

11

5

6

New Column Dimensions, H2 Gas, Translate Only

Page 40: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 45

New Velocity

New Temp.

Program

New Column Dimensions, H2 Gas, Fast Analysis

Page 41: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 46

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0 7 . 5 0

2 0 0 0 0 0 0

3 0 0 0 0 0 0

4 0 0 0 0 0 0

5 0 0 0 0 0 0

6 0 0 0 0 0 0

7 0 0 0 0 0 0

8 0 0 0 0 0 0

9 0 0 0 0 0 0

1 e + 0 7

1 . 1 e + 0 7

1 . 2 e + 0 7

1 . 3 e + 0 7

1 . 4 e + 0 7

1 . 5 e + 0 7

1 . 6 e + 0 7

1 . 7 e + 0 7

1 . 8 e + 0 7

1 . 9 e + 0 7

2 e + 0 7

2 . 1 e + 0 7

2 . 2 e + 0 7

2 . 3 e + 0 7

2 . 4 e + 0 7

2 . 5 e + 0 7

2 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 4 5 . D \ E C D 2 BNew Column Dimensions, H2 Gas, Fast Analysis

Better Resolution

and Faster Analysis!

Page 42: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 47

New Column Dimensions, H2 Gas, Higher Velocities

Page 43: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 48

105 cm/sec

95 cm/sec

85 cm/sec

<6 minutes

5.5 minutes!

?

New Column Dimensions, H2 Gas, Higher Velocities

Page 44: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 49

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 2 B

22

21

20

19

18

17

16

1514

13

12

11

109

8

7

6

5

4

3

2

1

Column: DB-XLB

20m x 0.18mm i.d., 0.18µm

Carrier: H2, constant flow, 77.3cm/s at 120°C

Injector: Pulsed Splittless, 220 °C

Pulse pressure & time: 35psi for 0.5min

Flow ramp at 6.25min of 99mL/min2 to 3mL/min

2mm i.d. liner

0.5µL, 50ppb

Oven: 120°C for 0.49min

120°C to 160°C at 59.4°/min

160°C to 260°C at 23.7°/min

260°C to 300°C (1.69min) at 35.6°/min

Detector: µ-ECD, 320°C

Ar/CH4 (P5) makeup gas at 60mL/min

Final Method Used at EPA

Page 45: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 50

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 2 B

DB-XLB

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 1 A

V i e w M o d e : I n t e g r a t i o n

DB-17ms

22

21

20

19

18

17

161514

13

12

11

1098

7

6

5

4

3

2

1

22

21

20

19

1817

1615

14

1312

11

1098

7

6

5

4

32

1

Final Method Used at EPA

Page 46: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 51

Food/Fragrance – Method translation

Page 47: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 52

Food/Fragrance – Method translation

Page 48: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 53

Spearmint Oil

Δ -9.7 min (0.18 mm, He Carrier)

Δ -16.8 min (0.18 mm, H2

Carrier)

10.6 min

17.7 min

3 8 13 18 23 28

(0.25 mm, He

Carrier)

Time (min)

27.4 min

Page 49: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 54

10.6 min

17.7 min

3 8 13 18 23 28

27.4 min

(0.25 mm, He Carrier)

Δ -9.7 min (0.18 mm, He carrier)

Δ -16.8 min (0.18 mm, H2 Carrier)

Spearmint Oil – Resolution Check

Page 50: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 55

Resolution Maintained

Compound ResolutionCompound

s

0.25 mm

Helium

0.18 mm

Helium

0.18 mm

Hydrogen

Sabinene

β-Pinene1.52 1.59 1.56

α-Terpinene

p-Cymene1.61 1.73 1.86

Speed Gain N/A 35% 61%

Page 51: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 56

New Devices & Instrument Improvements

Provide Even Faster Analytical Cycle Times

Low Thermal Mass Technology Capillary Flow Technology

Page 56

Page 52: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 57

“LTM” (Low Thermal Mass) Technology (Patented)

Directly heat/cool fused silica GC columns

Page 57

Page 53: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 58

A Closer Look

LTM Column Module

LTM Column

Assembly

LTM Retrofit

Door

Page 58

Page 54: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Obtaining Faster GC, GC/MS Analytical Cycle Times

– A Variety of Approaches

Post-Run

Bake-Out

Chromatographic

Run

Post-Run

Cool-Down

Pre-Run

ALS Set-Up

Included “ALS Sample Overlap” 7890 GC

in in ChemStation SW Faster Cool-down

HW/SW

Faster GC Column & Gas Selection,

Analyses Method Translation

Page 59

Page 55: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Backflush

During GC Run

S/S Inlet

Purged CFT

DeviceColumn

25 psi

4 psi

MSD

Aux EPCSplit Vent Trap

Page 60

Page 56: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Backflush

S/S Inlet

Purged CFT

DeviceColumn

1 psi

45 psi

MSD

Aux EPCSplit Vent Trap

During GC Run

After GC Run

S/S Inlet

Purged CFT

DeviceColumn

25 psi

4 psi

MSD

Aux EPCSplit Vent Trap

Page 61

Page 57: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

5 10 15 20 25 30 35 40 45 50 55 60 65 70

Run stopped at 42 min and

backflushed at 280oC for 7 mins.

It took additional 33 mins

and column to 320oC to

remove these high boilers.

Blank run after backflushing

min

showing the column was clean.

Milk Extract

Page 62

Page 58: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Other Purged Capillary Flow Devices Also

Provide Dual Use (including Backflush Capability)

2-Way Splitter with

Makeup

3-Way Splitter with

Makeup

Deans Switch

Page 63

(Quick Swap Alternatives)

Page 59: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 64

CONCLUSIONS

Stationary Phases – Chosen for optimized selectivity

Diameter – Smaller allows shorter length but has less

capacity

Make Small Changes – Again -- think capacity

Carrier Gas – Hydrogen, high velocity, but can still go fast

with He

Temperature Program – Scale properly to preserve

elution pattern

Method Translation Software – FREE, reliable

Flow Ramp – Increase at end of run for late eluters (if necessary)

System Upgrades – LTM, Capillary Flow Devices (back flush)

Page 60: Practical Faster GC Applications with Capillary GC Columns · GC Applications with Capillary GC Columns Simon Jones Application Engineer June 10, 2010. Practical Fast GC Applications

Practical Fast GC Applications with Capillary

GC Columns

May 19, 2009Page 65

Thank you!