n Step 3: Prepare the SNAP* Device © 2017 IDEXX Laboratories, Inc. All rights reserved. • 06-18022-08 *Vacutainer is a registered trademark of Becton, Dickinson and Company. SNAPshot Dx, SNAP, SNAP cPL, SNAP fPL, 4Dx, and Catalyst are trademarks or registered trademarks of IDEXX Laboratories, Inc. or its affiliates in the United States and/or other countries. Preparing Samples for Use on the IDEXX SNAPshot Dx * Analyzer 2. Dispense the appropriate amount of conjugate into the same sample tube used in step 1. 1. Following the on-screen instructions on the SNAPshot Dx Analyzer, dispense sample into disposable sample tube. Ensure you use the correct pipette (as indicated on the screen). 5. Pour entire contents of sample tube into the sample well of a SNAP* device. 3. Gently invert the sample tube 4–5 times to mix. 4. Total T 4 and Cortisol ONLY: Incubate the sample tube for 5 minutes. Anticoagulated Whole Blood (e.g., EDTA, heparin) Test Type Anticoagulated Whole Blood Plasma Serum Pipette Used Components Requiring Room Temperature Warm-up Warm-up Time Required SNAP* Total T 4 Purple Device 15 minutes SNAP* Cortisol Gray or Blue (depending on protocol) Device 10 minutes SNAP* Bile Acids Gray Device and conjugate 30 minutes SNAP* FIV/FeLV Combo Clear (provided) Device and conjugate 30 minutes SNAP* Heartworm RT Clear (provided) Device and conjugate 30 minutes SNAP* 4Dx* Plus Clear (provided) Device and conjugate 30 minutes SNAP* cPL* Clear (provided) Device and conjugate 30 minutes SNAP* fPL* Clear (provided) Device and conjugate 30 minutes SNAP* Feline proBNP Clear (provided) Device and conjugate 30 minutes Lithium heparin-based plasma EDTA-based plasma 1. Use the appropriate sample collection device. 2. Draw the sample gently and transfer it to a treated sample tube. 1. When the sample drawer is ready, remove the Catalyst* whole blood separator from the sample drawer. 2. Aspirate the plasma from the whole blood separator. Serum Plasma OR n Step 2: Collect and Prepare the Sample activation circle 6. When color first appears in the activation circle, press the activator. 7. Immediately load the SNAP device into an available port on the analyzer. n Step 1: Verify Sample and SNAP* Device Requirements When using an evacuated tube, such as a BD Vacutainer* tube, allow the sample to draw naturally into the tube by vacuum. 1. Use the appropriate serum separator tube. 3. Draw the sample gently and transfer it to a serum tube. 4. Let the sample clot for a minimum of 20 minutes. 2. Use the appropriate sample collection device. 5. Centrifuge the sample for at least 120 seconds at a minimum of 12,000 RCF. Refer to your centrifuge operator’s manual for centrifugation settings and times. 1. Use the appropriate lithium heparin tube. Do not use EDTA or sodium heparin for SNAP Total T 4 . 2. Use the appropriate sample collection device. 3. Draw the sample gently and transfer it to a lithium heparin tube. 4. Gently invert the sample for 30 seconds to mix. When using an evacuated tube, such as a BD Vacutainer* tube, allow the sample to draw naturally into the tube by vacuum. IDEXX Technical Support U.S./Canada 1-800-248-2483 Europe idexx.eu Australia 1300 44 33 99 New Zealand 0800 10 20 84 Distributors [email protected] Brazil 0800-777-7027 Latin America [email protected] China 400-678-6682 South Korea 080 7979 133 Taiwan 0800 291 018 Japan 0120-71-4921 5. Centrifuge the sample for at least 120 seconds at a minimum of 12,000 RCF. Refer to your centrifuge operator’s manual for centrifugation settings and times.

Upload: others

Post on 16-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Preparing Samples for Use on the IDEXX SNAPshot Dx Analyzer · sample to draw naturally into the tube by vacuum. 1. Use the appropriate serum separator tube. 3. Draw the sample gently

nStep 3: Prepare the SNAP* Device

© 2017 IDEXX Laboratories, Inc. All rights reserved. • 06-18022-08 *Vacutainer is a registered trademark of Becton, Dickinson and Company. SNAPshot Dx, SNAP, SNAP cPL, SNAP fPL, 4Dx, and Catalyst are trademarks or registered trademarks of IDEXX Laboratories, Inc. or its affiliates in the United States and/or other countries.

Preparing Samples for Use on the IDEXX SNAPshot Dx* Analyzer

2. Dispense the appropriate amount of conjugate into the same sample tube used in step 1.

1. Following the on-screen instructions on the SNAPshot Dx Analyzer, dispense sample into disposable sample tube. Ensure you use the correct pipette (as indicated on the screen).

5. Pour entire contents of sample tube into the sample well of a SNAP* device.

3. Gently invert the sample tube 4–5 times to mix.

4. Total T4 and Cortisol ONLY: Incubate the sample tube for 5 minutes.

Anticoagulated

Whole Blood (e.g., EDTA, heparin)

Test Type

Anticoagulated Whole Blood

Plasma

Serum

Pipette Used

Components Requiring Room Temperature

Warm-up

Warm-up Time

Required

SNAP* Total T4 † Purple Device 15 minutes

SNAP* Cortisol Gray or Blue (depending on protocol)

Device 10 minutes

SNAP* Bile Acids Gray Device and conjugate 30 minutes

SNAP* FIV/FeLV Combo Clear (provided) Device and conjugate 30 minutes

SNAP* Heartworm RT Clear (provided) Device and conjugate 30 minutes

SNAP* 4Dx* Plus Clear (provided) Device and conjugate 30 minutes

SNAP* cPL* Clear (provided) Device and conjugate 30 minutes

SNAP* fPL* Clear (provided) Device and conjugate 30 minutes

SNAP* Feline proBNP ‡ Clear (provided) Device and conjugate 30 minutes†Lithium heparin-based plasma‡EDTA-based plasma

1. Use the appropriate sample collection device.

2. Draw the sample gently and transfer it to a treated sample tube.

1. When the sample drawer is ready, remove the Catalyst* whole blood separator from the sample drawer.

2. Aspirate the plasma from the whole blood separator.

Serum Plasma

OR

nStep 2: Collect and Prepare the Sample

activation circle

6. When color first appears in the activation circle, press the activator.

7. Immediately load the SNAP device into an available port on the analyzer.

nStep 1: Verify Sample and SNAP* Device Requirements

‡ When using an evacuated tube, such as a BD Vacutainer* tube, allow the sample to draw naturally into the tube by vacuum.

1. Use the appropriate serum separator tube.

3. Draw the sample gently and transfer it to a serum tube.‡

4. Let the sample clot for a minimum of 20 minutes.

2. Use the appropriate sample collection device.

5. Centrifuge the sample for at least 120 seconds at a minimum of 12,000 RCF. Refer to your

centrifuge operator’s manual for centrifugation settings and times.

1. Use the appropriate lithium heparin tube. Do not use EDTA or sodium heparin for SNAP Total T4.

2. Use the appropriate sample collection device.

3. Draw the sample gently and transfer it to a lithium heparin tube.‡

4. Gently invert the sample for 30 seconds to mix.

‡ When using an evacuated tube, such as a BD Vacutainer* tube, allow the sample to draw naturally into the tube by vacuum.

IDEXX Technical Support U.S./Canada 1-800-248-2483

Europe idexx.eu

Australia 1300 44 33 99

New Zealand 0800 10 20 84

Distributors [email protected]

Brazil 0800-777-7027

Latin America [email protected]

China 400-678-6682

South Korea 080 7979 133

Taiwan 0800 291 018

Japan 0120-71-4921

5. Centrifuge the sample for at least 120 seconds at a minimum of 12,000 RCF. Refer to your

centrifuge operator’s manual for centrifugation settings and times.