HAA-Ob-7/7-2/2 /Izdanje 6/Issue 6 stranica/page 1 od/of 7

PRILOG POTVRDI O AKREDITACIJI br.: 2091 Annex to the Accreditation Certificate No.:

Klasa/Ref. No.: 383-02/11-80/003

Urbroj/Id. No.: 569-02/4-14-7

Datum izdanja priloga /Annex Issued on: 2014-01-30

Zamjenjuje prilog/Replaces Annex:

Klasa/Ref. No.: 383-02/11-80/003

Urbroj/Id. No.: 569-02/4-13-3

Datum izdanja priloga /Annex Issued on: 2013-01-02

Norma: HRN EN ISO/IEC 17025:2007

Standard:(ISO/IEC 17025:2005+Cor.1:2006; EN ISO/IEC 17025:2005+AC:2006)

Akreditacija istječe: 2016-09-03

Accreditation expiry:

Prva akreditacija: 2006-09-05

Initial accreditation:

Akreditirani laboratorij

Accredited laboratory

LABORING d.o.o. za mjeriteljstvo i ispitivanje

Virjanska 22, HR-10000 Zagreb

Područje akreditacije:

Scope of accreditation:

Umjeravanje mjerila:duljine, temperature, tlaka, gustoće, obujma, momenta sile i

sile

Umjeravanje fokometara

Calibration of length, temperature , pressure, density, volume, torque and force

gauges Calibration of focimeters

Važeće izdanje Priloga dostupno je na web adresi: www.akreditacija.hr /

Valid issue of the Annex is available at the web address: www.akreditacija.hr

Ravnateljica:

Director General:

mr. sc. Biserka Bajzek Brezak, dipl. ing.

Page 3: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 2 od/of 7

PODRUČJE AKREDITACIJE / SCOPE OF ACCREDITATION

Mjerna

veličina/

Mjerilo

Measurand /

Calibration

item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration

and

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

1.

Duljina

(Pomična

mjerila)

Length

(Calibration of

Vernier-

callipers)

Od/from 0 do/to

300 mm

30 µm + 1 .10

-4 .

L

ISO 3599:1976

2.

Duljina

(Umjeravanje

mikrometara za

mjerenje

vanjskih

dimenzija u dvije

točke)Length

(Calibration of 2

point

micrometers for

outside

measurements)

Od/from 0 do/to

25 mm

5 µm + 1 .10

-5 . L

ISO 3611:1978

3.

Temperatura

(Stakleni

termometri

rastezanja)

Temperature

(Liquid-in-glass

thermometers)

Od/ from – 20 °C do/to 0

°C

0,03-0,0045 t°C

LAB-PO-0020, izdanje/Issue

08;2011-06-09 Umjeravanje

termometara punjenih tekućinom,

s cjelovitim uranjanjem

Liquid-in glass full immersion

thermometers calibration

LAB-PO-0021, izdanje/Issue

07;2011-06-09; Umjeravanje

termometara punjenih tekućinom,

s djelomičnim uranjanjem

Liquid-in-glass partial immersion

thermometers calibration

Vodena /glikolna

kupelj

dubina = 125

mm.

Water- glycol

bath Immersion

= 125 mm.

Od/ from 0 °C do/to +

100 °C

0,03+ 0,0004 t

°C

Kupka sa

silikonskim

uljem

Bath with silicon

oil

Od/from 100 °C do/to 275

°C

0,07 °C

> 275 °C do/to 400 °C

1,4 °C

Suhi kalibrator.

Dubina uranjanja

125 mm.

Dry block.

Immersion depth

125 mm.

Page 4: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 3 od/of 7

Mjerna

veličina/

Mjerilo

Measurand /

Calibration

item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration

and

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

4.

Temperatura

(Termometri s

direktnim

očitanjem, sa

platinskim

otporničkim

osjetnicima)

Temperature

(Direct reading

thermometers

whicha have

resistance

thermometers as

sensors )

Od/ from – 20 °C do/to

0 °C

0,03-0,0045 t°C

uronjenje/immer

sion > 20xD

LAB-PO-0023, izdanje/Issue

10;2013-10-21

Umjeravanje instrumenata sa

direktnim očitanjem

Direct reading instruments

calibration

Vodena /glikolna

kupelj

dubina = 125

mm.

Water- glycol

bath Immersion

= 125 mm.

Od/ from 0 °C do/to

+ 100 °C

0,03+ 0,0004 t

°C

uronjenje/immer

sion > 20xD

Od/from 100 °C do/to 275

°C

0,07 °C

uronjenje/immer

sion > 20xD

Kupka sa

silikonskim

uljem

Bath with silicon

oil

> 275 °C do/to 650 °C

1,4 °C

uronjenje/immer

sion > 20xD

Suhi kalibrator.

Dubina uranjanja

125 mm.

Dry block.

Immersion depth

125 mm.

5.

od/from 50°C do/to 650

°C

0,6 °C

uronjenje/immer

sion > 20xD

**LAB-PO-0023, izdanje/Issue

10;2013-10-21

Umjeravanje instrumenata sa

direktnim očitanjem

Direct reading instruments

calibration

Suhi kalibrator.

Dubina uranjanja

125 mm.

Dry block.

Immersion depth

125 mm.

Page 5: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 4 od/of 7

Mjerna

veličina/

Mjerilo

Measurand /

Calibration

item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration

and

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

6.

Temperatura

(Termometri s

direktnim

očitanjem s

termoparovima

kao osjetnicima)

Temperature

(Direct reading

thermometers

which have

thermocouples

as sensors )

Od/ from – 20 °C do/to

100 °C

0,6 °C

uronjenje/immer

sion > 20xD

LAB-PO-0023, izdanje/Issue

10;2013-10-21

Umjeravanje instrumenata sa

direktnim očitanjem

Direct reading instruments

calibration

Vodena /glikolna

kupelj

dubina = 125

mm.

Water- glycol

bath Immersion

= 125 mm.

Od/from 100 °C do/to 275

°C

0,7 °C

uronjenje/immer

sion > 20xD

Kupka sa

silikonskim

uljem

Bath with silicon

oil

> 275 °C do/to 650 °C

1,6 °C

uronjenje/immer

sion > 20xD

Suhi kalibrator.

Dubina uranjanja

125 mm.

Dry block.

Immersion depth

125 mm.

7.

Temperatura

(Termometri s

direktnim

očitanjem s

termoparovima

kao

osjetnicima)

od/from 50°C do/to 650

°C

0,8°C

uronjenje/imm

ersion > 20xD

**LAB-PO-0023, izdanje/Issue

10;2013-10-21

Suhi

kalibrator.

Dubina

uranjanja 125

mm.

Dry block.

Immersion

depth 125 mm.

8. Temperatura/

termostatirane

komore

Temperature/

Thermostatic

chambers

-20°C do/to 400 °C

1,0 °C

DKD-R 5-7 **

LAB-PO-0053,izdanje/Issue 01

2012-07-16

V< 2000 l

9. Temperatura/

Termostatirane

kupelji

Temperature/

Thermostatic

baths

Od/ from – 20 °C do/to

300 °C 0,05 °C

**LAB-PO-0050,izdanje/Issue

02, 2013-10-21

Određivanje

odstupanja,

horizontalnog i

vertikalnog

gradijenta, te

stabilnosti u

kupelji.

Determination

of deviation,

horizontal and

vertical

gradient and

stability in

baths

Page 6: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 5 od/of 7

Mjerna

veličina/

Mjerilo

Measurand /

Calibration

item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration

and

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

10.

Obujam

(Klipna mjerila

obujma)

Volume

(Piston

operated

volumetric

apparatus)

Od/ from 20 μl do/ to

40 μl

1 %

EN ISO 8655 -6:2002

(ISO 8655 -6:2002)

Od/ from 41 μl do/ to

100 μl

0,5 %

Od/ from 101 μl do/ to

100 ml

0,2%

11.

Obujam

(Staklene

pipete – in, ex)

Volume

(Glass

pipettes – in,

ex)

Od/ from 0,10 ml do/to

0,30 ml

0,6 %

HRN ISO 4787: 2000

(ISO 4787: 1984)

Mjerna

sposobnost

procjenjena je

za mjerenje

bez standardne

devijacije

Calibration

and

measurement

capability is

given for

measurement

without

standard

deviation.

Od/ from 0,31 ml do/ to

1,00 ml

0,3 %

Od/ from 1,01 ml

do/ to 500 ml

0,13 %

12.

Obujam

(Stakleno

odmjerno

posuđe,piknom

etri, graduirani

odmjerni

cilindri,

staklene birete)

Volume

(Laboratory

glass vessels,

pycnometers,

graduated

measuring

cylinders,

glass burettes)

Od/ from 1 ml do/ to

5000 ml

0,13 %

HRN ISO 4787: 2000

(ISO 4787: 1984)

Page 7: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 6 od/of 7

Mjerna

veličina/

Mjerilo

Measurand /

Calibration

item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration

and

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

13.

Moment sile

(Moment

ključevi)

Torque

(Torque

wrenches)

Od/ from 7 Nm do/ to

20 Nm

1,0 %

ISO 6789:2003

Od/ from 21 Nm do/

to 2000 Nm

0,5 %

14.

Moment sile

(Testeri za

moment

ključeve)

Torque

(Torque

wrench testers)

Od/ from 5 Nm do/ to

2000 Nm

0,2 %

Euramet cg-14; Guidelines on the

Calibration of Static Torque

Mesauring Devices , izd.02,

03/2011

za tester s

podjelom skale

0,1 Nm

for torque

wrench tester

with scale

division of 0,1

Nm

15.

Sila

(Tenziometri)

Force

(Cable

tensiometers)

Od/ from 5 N do/ to

2750 N

0,1 %

LAB-PO-0034:2006; izdanje/Issue

05;2011-02-17

Umjeravanje tenziometara

Cable tensiometers calibration

Upotreba

etalonskih

utega i

određene

čelične užadi

use of

standard

weights and

specific steel

cables

16.

Tlak

(Mjerila tlaka)

Pressure

(Pressure

gauges)

Od/ from –0,95 bar

do/ to 1 bar 0,5 mbar

**DKD-R 6-1:2003; Calibration of

Pressure Gauges

(EURAMET/cg-17/v.02:2011)

Tlačni medij:

plin (dušik,

zrak)/Pressure

medium: gas

(nitrogen, air)

Od/ from 1 bar do/ to

100 bar 0,07 bar

Tlačni medij:

ulje

Pressure

medium: oil

Od/ from 100 bar

do/ to 690 bar 0,5 bar

17.

Gustoća

(Areometri)

Density

(Hydrometers)

Od/ from 0,65 gcm-3

do/ to 1,15 gcm-3

0,0003 g cm-3

HRN ISO 649-2: 2001

(ISO 649-2: 1981)

Page 8: PRILOG POTVRDI O AKREDITACIJI br.: 2091 - Laboratorij … · 2014-11-17 · PRILOG POTVRDI O AKREDITACIJI br.: 2091 ... 0,2 % Euramet cg-14; Guidelines on the Calibration of Static

Prilog potvrdi o akreditaciji Annex to the Accreditation Certificate

2091

Datum izdanja priloga/ Annex Issued on 2014-01-30

Zamjenjuje prilog od /Replace Annex dated: 2013-01-02

HAA-Ob-7/7-2/2 / Izdanje 6/Issue 6 stranica/page 7 od/of 7

Mjerna

veličina/

Mjerilo

Measurand /

Calibration item

Mjerno područje

Measure range

Mjerna

sposobnost*

Calibration an

measurement

capability*

(CMC)

Metode umjeravanja

Calibration methods Napomene

Remarks

18.

Optika/

umjeravanje

fokometara

Optics/

Calibration of

focimeters

Mjerenje stražnje

tjemene jakosti leća

(od -20D do + 20D)

Mjerenje prizmatične

jakosti leća( od 2Δ

do 10 Δ ) Determination of

back vertex power

(from - 20 D to +

20 D) and

prismatic power (

from 2Δ to 10 Δ) of

lens

0,05 D

0,05 Δ

OIML R 93:1999

LAB-PO-0054, izdanje/Issue

01;2012-07-16

LAB-RU-0040, izdanje/Issue

01;2012-07-16 **

* Iskazano kao proširena mjerna nesigurnost (k = 2) izračunata u skladu s EA-4/02:1999 „Izražavanje mjerne

nesigurnosti pri umjeravanju“

Expressed as an expanded uncertainty (k = 2) according to EA-4/02:1999 „Expression of the uncertainty of

measurement in calibration“

** Metoda umjeravanja se provodi u laboratoriju i na terenu

Method of calibration performed in laboratory and on on-site.