Problem Report & ResolutionSummary
• VEP, v1.1.7• Device = ADIS16228• PROBLEM
– The Alarm Status Form does not appear to detect Alarm conditions properly
– See Following slides for summary of observation• RESOLUTION
– The process of observation omitted one critical step: enable alarms
– See pages 7 – 9 for step‐by‐step instructions on enabling the alarms.
– See page 13 for demonstration of expected results
![Page 2: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/2.jpg)
Step #1 – Start the software
![Page 3: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/3.jpg)
Step #2 – Press Start to run a Manual FFT capture, analysis and waveform update
![Page 4: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/4.jpg)
Step #3 – Press Alarms > Alarm Settings to access the window to set the alarm levels
![Page 5: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/5.jpg)
Step #4 – Press Read Default Values
![Page 6: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/6.jpg)
Step #5 – Press Write to DUT, then exit Alarm Settings
![Page 7: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/7.jpg)
Step #6 – Press Register Access from the Main Screen
![Page 8: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/8.jpg)
Step #7 – Scroll down and find ALM_CTRL
![Page 9: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/9.jpg)
Step #8 – Set ALM_CTRL[2:0] = 111Set New Hex Value to 87, then press Write
![Page 10: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/10.jpg)
Step #9 – Exit Register AccessWindow, then Press Alarms > Alarm Status Form
![Page 11: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/11.jpg)
Step #10 – Press > Start to run a Manual FFT
![Page 12: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/12.jpg)
Step #11 – Change vertical scale in Waveform displays
![Page 13: Problem Report Resolution Summary · 2013. 12. 16. · Problem Report & Resolution Summary • VEP, v1.1.7 • Device = ADIS16228 • PROBLEM – The Alarm Status Form does not appear](https://reader035.vdocument.in/reader035/viewer/2022081620/6117b650d953fb730179cca3/html5/thumbnails/13.jpg)
Step #12 – While shaking device, press > Start to run another Manual FFT