2
CTA: 99929 69389 Ø10 40 60 30 40 40 80 20 Needle puncture test fixture with variable test angle Applications The needle puncture test fixture is used to determine the puncture force of needles to ISO 11040-4. Typical medical applications are, for example, puncture tests on skin replacement material or penetration tests on standard films. Function description The needle puncture test fixture consists of a base body with a swivel able specimen holder. The material to be punctured is gripped in the specimen holder. The test angle can be set to four different angles by means of a lock-pin. Advantages and features • Variable test angle to test axis; can be locked at 30°, 45°, 60° and 90° • Needles with a diameter up to 8 mm or injection needles and other connectors with Luer cones can be mounted in the optionally available holders • ZwickRoell's testing software combined with the Traceability option fulfills the requirements necessary for compliance with all FDA 21 CFR Part 11 criteria Product Information Needle puncture test fixture 1318214 0119

Upload: others

Post on 12-May-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Product Information Needle puncture test fixture€¦ · Needle puncture test fixture with variable test angle 1318214 0119 Applications The needle puncture test fixture is used to

CTA

: 999

29 6

9389

Ø10

40

6030

4040

80

20

Needle puncture test fixture with variable test angle

ApplicationsThe needle puncture test fixture is used to determinethe puncture force of needles to ISO 11040-4. Typicalmedical applications are, for example, puncture tests onskin replacement material or penetration tests onstandard films.

Function descriptionThe needle puncture test fixture consists of a base bodywith a swivel able specimen holder.

The material to be punctured is gripped in the specimenholder. The test angle can be set to four different anglesby means of a lock-pin.

Advantages and features• Variable test angle to test axis; can be locked at 30°,

45°, 60° and 90°• Needles with a diameter up to 8 mm or injection

needles and other connectors with Luer cones can bemounted in the optionally available holders

• ZwickRoell's testing software combined with theTraceability option fulfills the requirements necessaryfor compliance with all FDA 21 CFR Part 11 criteria

Product InformationNeedle puncture test fixture

1318

214

0119

Page 2: Product Information Needle puncture test fixture€¦ · Needle puncture test fixture with variable test angle 1318214 0119 Applications The needle puncture test fixture is used to

Technical data

Type Needle puncture test fixture

Item no. 022358

Maximum test load (Fmax) 50 N

Specimen dimensions

Specimen width 20 mm

Specimen thickness up to 1 mm

Specimen gripping and puncture opening Ø 10 mm

Test angle, can be locked in with a pin 30°, 45°, 60° and 90°

Connecting diameter Ø 20 mm

Ambient temperature +10 ... +35 °C

Accessories required

1 x requiredDescription Item number

Needle holder (non-corrosive), to mount needles with a diameter of 0 to 8 mm 022357

Mounting to fix injection needles and other connectors with Luer cones 024560