Teresa Jorge 1 , Manuel Graf 2 , Sergey Khaykin 3 , Yann Poltera 1 1 Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, Switzerland 2 Swiss Federal Laboratory for Materials Science and Technology 3 CNRS/INSU, LATMOS-IPSL, France Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH); and QCLAS and FLASH-B GRUAN ICM-11 Meeting Singapore, 23.05.2019

Upload: others

Post on 10-May-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Teresa Jorge1, Manuel Graf2, Sergey Khaykin3, Yann Poltera1

1Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, Switzerland 2Swiss Federal Laboratory for Materials Science and Technology 3CNRS/INSU, LATMOS-IPSL, France

Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH); and QCLAS and FLASH-B

GRUAN ICM-11 Meeting Singapore, 23.05.2019

Page 2: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Peltier Cooled Frostpoint Hygrometer (PCFH): Report on 2018 Activities & Status

slides adapted from

GAW-CH Landesausschuss Spring Meeting, 14 May 2019, Empa Dübendorf, Switzerland

T. Jorge1, F. G. Wienhold1, G. Cesbron1, S. Brossi2, T. Brossi2, P. Oelsner3, T. Naebert3, S. Meier3, and T. Peter1

1. Swiss Federal Institute of Tech., Zürich, Switzerland

2.mylab elektronik GmbH, Bubikon, Switzerland 3.GRUAN Lead Center, DWD Meteorological Observatory, Lindenberg, Germany

Page 3: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

PCFH Fundamentals

Principle of Operation

Peltier-cooled mirror hygrometer instead of cryogenically-cooled like FPH or CFH (phase down agreement of the Kigali Amendment to the Montreal Protocol, 2016).

Research and test instrument: includes two independently cooled mirrors and boards.

Page 4: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Field Campaigns at Lindenberg

1st Field Campaign in Lindenberg, 23 .. 27 July 2018: 2 flights

2nd Field Campaign in Lindenberg, 10 .. 14 December 2018: 3 flights

Aims of the Field Campaigns No frost point measurements!

Approval of basic concepts (optical, thermal, electrical, telemetry)

Characterization of Thermal Electric Cooler (TEC) : in lab in flight environment

Characterization of condensate formation and detection (mirror reflectivity monitoring)

Gather data for instrument numerical model implementation

Page 5: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

CFH

ECC O3 I-Met

COBALD

RS41 Twig®

RS41

OIF411

PCFH

1st Field Campaign: Flights & Payloads PCFH Flight 1 2018-0725-02UT_LI_LI236 (regular

science) 2018-0725-02UT_LI_PCFH001 Two separate RS41 telemetries

PCFH Flight 2 • 2018-0725-

20UT_LI_PCFH002: ECC O3, COBALD, CFH & PCFH

• One single RS41 CFH

ECC O3 box with OIF411

COBALD

RS41

Twig

®

spacer

PCFH

Power supply: sufficient for one flight duration

Robust: only battery replacement required after recovery.

Telemetry issues

PCFH

Page 6: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

TEC Performance

preflight: no ventilation

ITEC (a.u.)

∆T

= T h

ot –

Tm

irror

(K

)

1st Field Campaign: Selected Results Example Profile

2018

-072

5-02

UT_

LI_P

CFH

001

sub

-uni

t 1

Frost point temperature reached only up to tropopause level. Heat load was underestimated need lighter heat sink, protection

(radiation shield) and better thermal coupling.

Page 7: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

More space and better thermal coupling → better heat dissipation new electronics design

→ higher TEC Current achieved

1st Flights’ Version: not enough heat dissipation

PCFH Setup Revisions

Page 8: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

CFH

ECC O3 RS41 Twig®

PCFH COBALD

RS41 Twig®

PCFH

PCFH Flight 3 & 4 (“simultaneous” launch) • 2018-1212-16UT_LI_LI246-CFH

two single stage TECs • 2018-1212-16UT_LI_LI246-COB

single & double stage TECs

All Flights telemetry problem resolved

2nd Field Campaign: Flights & Payloads

CFH

ECC O3 RS41 Twig®

PCFH

COBALD

PCFH Flight 5 • 2018-1213-

22UT_LI_LI247 single & double stage TECs

Page 9: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

2nd Field Campaign Results: Cooling performance during ascent T h

ot ≅

Tai

r + 1

0 K

Black symbols: lab, without ventilation ∆

T =

T hot

– T

mirr

or (

K)

ITEC (A)

One assembly version achieved the wished performance.

Page 10: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Summary and Project Schedule Double stage TEC with increased spacing and with coupling with cupper block provides

adequate temperature range

Next steps: Improve Peltier hot side to heat sink coupling to minimize Thot – Tair (Nov-Dec 2019) → e.g. use larger cupper block

Numerical instrument model (2019- early 2020) Analyze ice formation dynamics and its effects on mirror reflectivity to introduce timing in

the numerical instrument model Add and test controller scheme in the numerical model

Field verification of the design changes Two campaigns in Lindenberg during November / December 2019 Four dual payload launches from Ny-Alesund in February 2020

Test Flights within Monitoring Networks is postponed by approximately one year (as of May 2019).

Page 11: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

[email protected] | May 2019 |

Quantum Cascade Laser based

Absorption Spectroscopy (QCLAS)

Balloon Instrument

alides from Manuel Graf, EMPA, Switzerland

Page 12: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Quantum Cascade Laser based Absorption Spectroscopy (QCLAS)

[email protected] | May 2019 | 1

- Direct absorption spectroscopy

- Excitation of a rovibrational transition at 1662 cm-1 (6.01 µm)

- Open path operation (L=6 meters)

- Novel circular multipass cell allows simple & robust setup1)

Principles Mid-IR: H-O-H bending vibrations

amount fraction, ppm [μmol/mol]

QC Laser MCT Detector

multipass cell

Temperature, pressure

Method

1) Graf et al. 2018, Optics Letters

𝑃𝑃𝑟𝑟 = 𝐶𝐶 ∗ 𝑃𝑃0 ∗ exp −(𝑁𝑁𝐻𝐻2𝑂𝑂 ∗ 𝜎𝜎𝑎𝑎𝑎𝑎𝑎𝑎,𝐻𝐻2𝑂𝑂 + 𝜖𝜖) ∗ 𝐿𝐿 + 𝑃𝑃𝑎𝑎

Page 13: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

The Instrument

gma | 13.11.2017 | 6

Specs

- On-board data storage

- Fully integrated electronics and batteries

- Thermal stabilization with phase change materials (PCM)

- 2 h autonomous operation

[email protected] | May 2019 | 2

size 30 x 20 x 10 cm3

mass 3 kg gas exchange (5 m/s ascent speed) 40 l/s power consumption 15 W measurement precision <1% measurement accuracy tba

bottom view Will be tested 2019-2020

Page 14: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Alexey Lykov, Central Aerological Observatory, Russia Sergey Khaykin, CNRS/INSU, LATMOS-IPSL, France

Central Aerological Observatory, ROSHYDROMET

Fluorescence Lyman-Alpha Stratospheric Hygrometer

for Balloon (FLASH-B)

www.flash-b.ru

slides from Sergey Khaykin

Page 15: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

The fluorescent method is based on the photodissociation of H2O molecules under UV lamp light at wavelengths below 137 nm and subsequent fluorescent relaxation of the exited OH* radical produced.

The instrument uses the open layout, where the optics is looking directly into the outside air.

Therefore, measuring only at night. For stratosphere, descent data only

Source of VUV radiation

λ=121,6 nm (Lyman-α) Н2О

Н(2S)

ОН (А2∑+)

ОН (Х2П )

PMT

300 < λ < 330 nm

Н2О + hν → ОН (А2∑+) + Н(2S)

fluorescence ОН (А2∑+) → ОН (Х2П) + hν’

Method of measurement and optical layout www.flash-b.ru

Page 16: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Range of measurement Resolution time Integration time Precision Total uncertainty Vertical resolution Temperature range Pressure range Required power Weight (w/o batteries) Size Insulation Box Flight weight Interface to Protocol XData

0.5...1000 ppmv 1 sec 4 sec 5.5% <10 % (1σ) at µ> 3 ppmv ~ 50 m (descent in UTLS) -950C … +400C 300… 5 hPa 9-30V, 1 W max 500 gr 265 mm x 155 mm x 105 mm 1000 gr Vaisala RS92, RS41 Meisei RS-06G, Meteolabor SRS34 20 byte/sec

Specifications www.flash-b.ru

Plug-and-play concept

Page 17: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Flights in Dolgoprudny www.flash-b.ru

3 consecutive flights of the same FLASH sonde in Dolgoprudny with a series of calibrations before and after each flight - Consistent water vapour vertical profiles in the stratosphere and good agreement with M10 and RS41 in the free troposphere - Small change of sensitivity factor (<10 %) in between the calibrations/flights

Tropospheric comparison with RS41 and M-10

Page 18: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Central Aerological Observatory, ROSHYDROMET

FLASH COBALD

Application on different platforms www.flash-b.ru

The Flash hygrometer was used on board stratospheric aircraft “Geophisia” M55 during StratoClim campaign and in test flight UAV.

Thank you for your attention! Questions?

Page 19: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

Backup slides

Thermal Electrical Cooler (TEC) Load Characteristic

double stage

single stage

Page 20: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Central Aerological Observatory, ROSHYDROMET

Flight preparation www.flash-b.ru

1.Recharge battery;

2.Fixing face down;

3.Clean optics;

4.Connect to radiosonde;

5.Switch on;

6.Check the data incoming;

7.Open cap;

8.Ready to fly!

Sealant recommended against contamination in stratosphere

Plug-and-play concept

Page 21: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Central Aerological Observatory, ROSHYDROMET

P: 3-1000 mBar T: 190 – 300 K µ: 1-1000 ppmv

To PC

Compressor

FLAS

H-B

Vacuum pump

dehu

mid

ifier

H2O

bu

bble

r

flow

co

ntro

ller

homogenizer

Reference dew point mirror hygrometer

MBW 373L ai

r out

flow

co

ntro

ller

Vacuum chamber

freezer

Air pressure

Calibration facility www.flash-b.ru

PCFH 2018 Activities & Status, GAW-CH Landesausschuss Spring Meeting, Dübendorf, 14 May 2019

Page 22: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Central Aerological Observatory, ROSHYDROMET

0

50

100

150

200

250

300

350

400

450

0

5000

10000

15000

20000

25000

30000

35000

132

464

797

012

9316

1619

3922

6225

8529

0832

3135

5438

7742

0045

2348

4651

6954

9258

1561

3864

6167

8471

0774

3077

5380

7683

9987

2290

4593

6896

9110

014

1033

710

660

1098

311

306

1162

911

952

1227

5

Signal MBW

Cal22 = 0.02x - 196.32

Cal25 = 0.0199x - 177.85 Cal26 = 0.0196x - 203.06

Cal29 = 0.019x - 175.82

Cal01 = 0.02x - 199.61

0

50

100

150

200

250

300

350

400

450

8000 13000 18000 23000 28000 33000

MB

W[p

pmv]

Signal [counts]

Calibration data www.flash-b.ru

The total relative error of the calibration amounts to 4%.

The calibration fit function is linear in the pressure range of 30 – 150 hPa and water vapor mixing range of 1 – 500 ppmv.

Page 23: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Central Aerological Observatory, ROSHYDROMET

Calculation formula www.flash-b.ru

WVMR = J*K(cal) * K(Quenching) * K(absorb.H20) * K(absorb.O2)

K(Quenching) = 1+(Q*N*(P/P0)*(T0/T)*106), where Q=2.3*10-11, P0,T0- surface pressure and temperature

K(absorb.H20) = EXP(-0.0001054*µ*P/T*L), where focus L=5 cm, P,T- pressure and temperature

K(absorb.02) = EXP(-0.0001536*P/T*L) Cal.point

If P < 36 μ = J*K(cal)*0.956*(1+0.00781*(t+273.16)/P) If 36 < P < 300 μ = J*K(cal) *(1+0.00031*P+0.00033*K(cal)2*P*J)

Page 24: Progress report on: the Peltier-cooled frostpoint hygrometer (PCFH… · 2019-11-28 · Principle of Operation Peltier-cooled mirror hygrometer instead of . cryogenically-cooled like

10th GRUAN Implementation and Coordination Meeting (ICM-10) Potsdam

Central Aerological Observatory, ROSHYDROMET

FLASH-B error budget www.flash-b.ru

Total uncertainty (calibration error + 1σ precision) is below 10% at stratospheric conditions.

Calibration error - MBW373L dew point uncertainty (0.1 K) - Pressure error (conversion to mixing ratio) - Non-linearity of calibration curve - Outgassing within calibration chamber - Random error (operator-related factors)

Total relative error of calibration is estimated at 4% (1 σ)

Measurement error - Instability of Lyman-alpha emission, including

temperature-related drifts (<3%) - Random error (5.5% precision for 4 s integration) - Detection limit 0.1 ppmv

Comparison with Pico-SDLA

Comparison with CFH Comparison with FPH In-flight repeatability

Comparison against other hygrometers and in-flight repeatability fully confirms the estimated uncertainty