Psychophysics in the Magnet

Sean MarrettFunctional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Page 2: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Imagine you are starting up your own FMRI lab

• Goal is to maximize the range of possible fMRI experiments and ease with which they can be set up.

• Exercise in this seminar is to consider range of essential equipment and pros and cons of different configurations

• Bare bones and build up from there

Page 3: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

What is needed• Experiment control and stimulus generation software (sometimes

the same, sometimes very different)• Interface with scanner to allow synchronization• Visual display • Device to allow subject feedback• Auditory stimulus delivery• Physiological monitoring setup with scanner synchronization• Method for recording eye-movement• Force measurement• Pain stimulation• Digression on head motion

– Prospective head movement correction

Page 4: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Hostile magnet• Strong static magnetic field can disrupt operation of many devices • Projectors Projectiles• RF emitted by coil(s) can interfere with operation and damage some

equipment and can also cause heating in devices / objects that can make them a hazard

• Rapidly changing magnetic field (gradient) can induce current in devices which can interfere or damage electronic devices

• Gradients also responsible for high levels of acoustic noise which make MR very loud (up to 120dB) which can damage hearing and also make it hard to communicate with subjects. Significant ear protection required

• Cables bringing electronic signals in and out of magnet room can bring in small signals that cause artifacts, especially because RF signals detected in FMRI are so tiny tat very low levels of spurious signals brought in from outside can produce prominent or subtle artifacts in reconstructed images.

Page 5: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

fORP FIU932

Interface box for button pressesConnects to user’s laptop via USBConnects to trigger box via BNCConnects to button box via proprietary 12-pin fiber connector.Compatible with legacy button boxes.

Current Designs FIU932 System

Synchronization Devices

Page 6: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Stimulus presentation software (Experimental control)

• What is needed • Commercial packages

– E-prime– Presentation

• Open source alternatives– PsychoPy– VisionEgg– Psychophysics Toolbox (+ Matlab($$))– Pygame

Page 7: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Visual stimulus

• Simplest setup includes external projector • Wave guide/ Long throw lens • Good mirror system & screen (DaLite)

Page 8: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Visual stimulus (cont)

• Projector sited inside magnet room• (Faraday cage)• Projector <> Projectile• Some video signals are right around NMR

resonance frequencies• Optical mux/demux

Page 9: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Hyperion Projector (3T-D / NMRF 3T)

Page 10: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Visual stimulus (cont)

• Integrated goggle systems (Resonance technologies, Nordic Neuro)

• Image guides (Avotec)• LCD Panels

Page 11: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Examples: integrated goggle systems

Page 12: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Video system practicalities

• Rear projection vs front projection (Easy set-up and more uniform field of view for diverse body types)

• Optics and long-throw lens (Buhl/Navitar is your friend)• Some modern DLP projectors very hard to modify• Inside magnet / Faraday cage• Outside magnet / wave guide / operation (turning it on

and off)• Timing and synchronization / video memory buffer• E.g. unreliable timing can make backward masking

problematic!

Page 13: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Delivering auditory stimulus in the magnet

• Very loud ambient noise• Headphones with transducers outside magnet• Headphones with electrostatics inside magnet

– Confon system• Opto-acoustic systems• Noise cancellation features• Gradients also responsible for high levels of acoustic noise

which make MR very loud (up to 120dB) which can damage hearing and also make it hard to communicate with subjects.

• Significant ear protection required• Close fitting head coils complicate ear protection and audio

delivery

Page 14: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Example audio delivery systems

Page 15: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Recording subject feedback/response (basic)

• Classic button box• Must be synchronized with scanner• Best if well-integrated with experimental

control software• Several vendors including CurrentDesign,

Psychology Software Tools, Cedrus• Range from classic button box to foot

response, VAS etc• Examples

Page 16: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Basic button boxes

Page 17: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Recording subject response• MRI compatible joystick / mouse• MRI compatible keyboard• Accelerometer/force measurement

Page 18: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Recording subject behavior

• Hand position/orientation • Video feed of subject face/limbs• Eye recording and tracking (next slide)

Page 19: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Eye movement system diagram

Page 20: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Eye movement recording systems

Page 21: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Recording subject behavior more advanced

• Musical instruments• Virtual reality

Page 22: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Subject/patient physiology

• Respiration• Pulse• Electrodermal activity• Integrated with scanner – easy to turn off/on• External system (Biopac or equivalent)

Page 23: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Biopac data acquisition system

Page 24: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Biopac Display

The 1st trace is from the respiration sensor / belt.The 2nd trace shows data from the pulse oximeter.The 3rd trace is the scanner’s TR trigger.The 4th trace is the filtered respiration data (similar to #1).The 5th trace is the filtered pulse oximeter data (similar to #2).

Page 25: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Touch, taste

• Ingeholm touch simulator• Taste – gustatory• Other specialized devices (shout out to

George Dold and Section on Instrumentation)

Page 26: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

John E. Ingeholm , George R. Dold , Lawerence E. Pfeffer , David Ide , Seth R. Goldstein , Kenneth O. Johnson , Ro...

The Helix: A multi-modal tactile stimulator for human functional neuroimaging

Page 27: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Gustatory stimulus deliveryJohn Ingeholm and George Dold and Tom Talbot and others

Page 28: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Other essential devices from Section on Instrumentation

Page 29: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Pain

• Medoc stimulator

Page 30: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Recording vocal responses (and communicating with subject)

• Noise cancelling microphones

Page 31: Psychophysics in the Magnet Sean Marrett Functional MRI Facility, NIMH/NIH/DHHS, Bethesda, MD, USA

Starting your own lab – FMRI device budget

• Scanner sync interface + basic button box ~ $2K and up

• Projector/visual display system - $4K and up (including mirror fabrication)

• Audio system - $5K and up• Eye tracking - $10K and up• External physiological monitor $5K and up