q-gard® t pack - servapure.com · 4 remove the used q-gard pack. 5 the system indicates that the...

5
Q-GARD® T PACK Cat. # QGARDT1X1 /QGARDT2X1 / QGARDT3X1 Replacement Where used • Milli-Q* Advantage A10* • Milli-Q Reference and Milli-Q Reference A+ system Q-Gard T Pack Instructions sheet The information in this document is meant to be used by a'Qualified Service Representative PF09758 - Rev. 5,01/2013 Q-GARD®T PACK 1/5

Upload: vanthu

Post on 13-Dec-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Q-GARD® T PACK - servapure.com · 4 Remove the used Q-Gard Pack. 5 The system indicates that the ... See User Manual. Note AID Cleaning is not needed for Milli-Q Reference and Milli-Q

Q-GARD® T PACKCat. # QGARDT1X1 /QGARDT2X1 / QGARDT3X1

Replacement

Where used • Milli-Q* Advantage A10*• Milli-Q Reference and Milli-Q Reference A+ system

Q-Gard T Pack

Instructions sheet

T h e i n f o r m a t i o n i n t h i s d o c u m e n t i s m e a n t t o b e u s e d b y a ' Q u a l i fi e d S e r v i c e R e p r e s e n t a t i v eP F 0 9 7 5 8 - R e v . 5 , 0 1 / 2 0 1 3 Q - G A R D ® T P A C K 1 / 5

Page 2: Q-GARD® T PACK - servapure.com · 4 Remove the used Q-Gard Pack. 5 The system indicates that the ... See User Manual. Note AID Cleaning is not needed for Milli-Q Reference and Milli-Q

Replacement procedure

When? The Q-Gard Pack should be replaced when one of the following Alert or Alarmmessages is displayed.• REPLACE Q-GARD PACK, or• MILLI-Q RES < SP. REPLACE Q-GARD AND QUANTUM.

Removal Follow the steps below to remove the Q-6ard Pack.

Step A c t i o n Resul t

1 Make sure the Milli-Q SystemIs in STANDBY Mode.

S T R N D B Y

11 ftpr 2006 07:35

Ready ^

2 • Push the Q-PQD Plunger (orPOD Plunger) down once todepressurise the Milli-QSystem.

• After water stops being

dispensed, push down theQ-POD Plunger (or PODPlunger) again. i

3 • Open the Milli-Q System leftdoor.

• Lift up the Pack LockingH a n d l e .

1Continued on next page

THE INFORMATION IN THIS DOCUMENT IS MEANT TO BE USED BY A QUALIFIED SERVICE REPRESENTATIVEP F 0 9 7 5 8 - R e v . 5 . 0 1 / 2 0 1 3 q - g a r d ® t p a c k 2 / 5

Page 3: Q-GARD® T PACK - servapure.com · 4 Remove the used Q-Gard Pack. 5 The system indicates that the ... See User Manual. Note AID Cleaning is not needed for Milli-Q Reference and Milli-Q

Replacement procedure, Continued

R e m o v a l

(continued) Step A c t i o n Resul t

4 Remove the used Q-Gard Pack.

5 The system indicates that theQ-Gard Pack is removed.

S T fl l

Q - G R R O P A C K O U T

='RESS ■>

1 U

■»

Installation Follow the steps below to install the Q-Gard Pack.

Step A c t i o n Resul t

1 • Remove the covers on the

2 ports of the Q-Gard Pack.• Look inside the ports.• Make sure the rubber 0-rings j i W t * n

are firmly in placeI l l "

• Wet the 0-rings with water.Push the top of the Q-6ard Packinto the ports on the Milli-QSystem.

2 Push on the bottom of the Q-GardPack.

' 'i"

Continued on next page

THE INFORMATION IN THIS DOCUMENT IS MEANT TO BE USED BY A QUALIFIED SERVICE REPRESENTATIVEP F 0 9 7 5 8 - R e v . 5 , 0 1 / 2 0 1 3 q - g a r d ® t p a c k 3 / 5

Page 4: Q-GARD® T PACK - servapure.com · 4 Remove the used Q-Gard Pack. 5 The system indicates that the ... See User Manual. Note AID Cleaning is not needed for Milli-Q Reference and Milli-Q

Replacement procedure, Continued

Ins ta l l a t i on

(continued) Step A c t i o n

The system indicates that a newQ-Gard Pack is installed.

• Push the Pack Locking Handled o w n .

• Close the left door.

Press 0

R e s u l t

INSTRLL Q-eRRO

fl o e u O - S a r d T 1 h a s b c - e n

i i \ s t a i l e d .

C»tal09u4 N* » QGFlRDTlXlLot H* = F6DN27329. *■

STANDBY

11 Apr 2006 07:35

Ready 4

THE INFORMATION IN THIS DOCUMENT IS MEANT TO BE USED BY A QUALIFIED SERVICE REPRESENTATIVEP F 0 9 7 5 8 - R e v . 5 , 0 1 / 2 0 1 3 Q - G A R D ® T P A C K 4 / 5

Page 5: Q-GARD® T PACK - servapure.com · 4 Remove the used Q-Gard Pack. 5 The system indicates that the ... See User Manual. Note AID Cleaning is not needed for Milli-Q Reference and Milli-Q

Rinsing

Procedure The Quantum Cartridge and the Q-Gard Pack need to be rinsed out after theyare installed.

Step A c t i o n

1 Locate the clear tubing and barbed fitting from the Accessories Bag.

2 Screw the barbed fitting onto the Q-POD Dispenser.

3 Push the clear tubing onto the barbed fitting. Place the other end ina sink.

4 Dispense water for about 10 minutes.5 Perform an AID Cleaning at this time. See User Manual.

N o t e

AID Cleaning is not needed for Milli-Q Reference and Milli-QReference A+ system.

6 Leave the Milli-Q System in READY Mode for at least 6 hours orovernight.

7 Dispense water for about 10 minutes.8 Remove the barbed fitting and clear tubing.

9 Install a POD Pak.

1 0 Leave the Milli-Q System in READY Mode.

Manual TOC Only for a Milli-Q Reference A+ system, perform a manual TOG Curve Check.Curve Check See user manual.

THE INFORMATION IN THIS DOCUMENT IS MEANT TO BE USED BY A QUALIFIED SERVICE REPRESENTATIVEP F 0 9 7 5 8 - R e v . 5 , 0 1 / 2 0 1 3 q - g a r d ® t p a c k 5 / 5