2017-06-07 | Dr. Andreas Scheel | University Hospital Cologne
Quality control of PD-L1 in NSCLC
- Results of two German ring trials
![Page 2: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/2.jpg)
Step-wise approach
Material Assays Scoring
Technicalaspects
Predictive value
Biologicalsignificance
![Page 3: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/3.jpg)
1.) Internal ring trial (DGP1) -
Gain experience; identify challenges
1a, Scoring: Observer-concordance
1b, Staining: Laboratory-concordance
Assays: Comparisons
Assays Scoring
Step-wise approach
2.) Open ring trial (QuIP GmbH2) -
Quality assessment
Workshop
Open ring trial
Assays
Scoring
1 'DGP' = German Association for Pathologyhttps://www.pathologie-dgp.de/
2 'QuIP' = 'Quality-Initiative Pathology'= German EQA Institution
http://www.quip-ringversuche.de
![Page 4: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/4.jpg)
1a) Internal ring trial: Scoring
1a: Internal ring trial,PD-L1 Scoring
![Page 5: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/5.jpg)
PD-L1 Scoring
Scheel AH et al. Der Pathologe 2016. 37(6):557-567
Various cut-off criteria (assays; line of treatment)
Criteria can be integrated into 6-step score
Currently most important:
≥50% and ≥1% stained tumor cells (TC)
![Page 6: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/6.jpg)
HE 28-8 22C3 SP142 SP263
1st round, 'Training Set':
15 cases NSCLC
Central staining
2 LDTs
9 Observer
2nd round, 'Validation Set':
15 cases NSCLC
Central staining
4 Clinical trial assays
9 Observer
1a) Internal ring trial: Scoring
![Page 7: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/7.jpg)
1a) Interobserver-Concordance
Landis JT, Koch GG. Biometrics 1977. 33(1);159-174
Scheel AH et al. Modern Pathology 2016. 29(10):1165-72
![Page 8: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/8.jpg)
1b) Internal ring trial: Staining
1b: Internal ring trial,PD-L1 Staining
![Page 9: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/9.jpg)
Trial TMA, NSCLC specimens
NSCLC-TMA, 2mm cores, n=21
Sections cut at central lab
Staining and scoring at local labs
- 10 institutes
- 4 clinical trial assay (CTAs)
(22C3, 28-8, SP263, SP142)
- 11 lab-developed tests (LDTs)
- Total: 27 stained sections
Central review of stainings
Controls: Cell-lines; Tonsil
1b) Internal ring trial: Staining
![Page 10: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/10.jpg)
#High SD because most samples were ≥1%
Landis JT, Koch GG. Biometrics 1977. 33(1);159-174
Scheel AH et al. submitted
1b) Interlaboratory-Concordance
![Page 11: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/11.jpg)
1b, Interassay-Comparison
1b) Internal ring trial: Interassay
![Page 12: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/12.jpg)
NSCLC-TMA, CTAs
![Page 13: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/13.jpg)
LDTs: QC+ and QC-
n=11 TMA-sections stained with LDTs
Employed Abs:22C3 n=428-8 n=3SP263 n=2E1L3N n=1QR1 n=1
Manual review
Staining similar to22C3 / 28-8 pharmDx= QC+ , n=6
Employed Abs:22C3 n=128-8 n=1SP263 n=2E1L3N n=1QR1 n=1
Staining divergent= QC- , n=5
Employed Abs:22C3 n=328-8 n=2
Image-analysis / quantitation
![Page 14: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/14.jpg)
NSCLC-TMA, LDTs
![Page 15: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/15.jpg)
Quantitation by image-analysis
IHC: PD-L1
Hematoxylin
DAB
- Per core: analysis of
0.46mm² @ 200x
- Invasive carcinoma
manually selected
- Color-deconvolution
- Read-outs: Hem. intensity,
DAB-area relative to area
of invasive carcinoma
Color-deconvolution
Manual selection
Positivemask
![Page 16: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/16.jpg)
- Bars: Mean of sites; antennae: SD
- QC+: Similar to 28-8 / 22C3; QC-: Less DAB Scheel et al, Submitted
Image-analysis
![Page 17: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/17.jpg)
Summary internal ring trial
1. Reproducible PD-L1 scoring is achievable
2. Reproducible PD-L1 staining is achievable
- Assays ('Kits') quite reliable
- LDTs possible, calibration critical
- Different primary antibody possible:
28-8, 22C3, SP263, E1L3N, others
Results match other studies, in particular
Adams J., / InCA-Network; WCLC 2016, Vienna
![Page 18: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/18.jpg)
2: Open ring trial / QuIP#-EQA
2.) Open ring trial / QuIP#
# 'QuIP' = 'Quality-Initiative Pathology'= German EQA Institution
http://www.quip-ringversuche.de
![Page 19: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/19.jpg)
Open ring trial; workshop
Workshop PD-L1 in NSCLC,
Charité, Berlin 09/2016
Lectures & live microscopy
PD-L1 scoring criteria
Challenges in scoring:
Cell type
Inclusion of low-intensity
Heterogeneity
Assay-calibration (LDTs)
![Page 20: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/20.jpg)
Open ring trial; work-flow
Case selection 3 lead institutes 14 cases Resection specimens
Case verification 5 panel institutes IHC: CTAs and LDTs Concordance ≥ 7/8
Ring trial 10 NSCLC cases 2 unstained slides / case Participants: n=87
CologneGöttingen
Berlin
Hannover
Jena
Heidelberg
Munich (LMU / TU)
QuIP GmBH, used with permission
![Page 21: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/21.jpg)
Open ring trial; evaluation
Per case Correct score 2 points Not evaluable 1 point
Total Optimal 20 points 'Successful' ≥ 18 points
Optional Return of stained slides for review
QuIP GmBH, used with permission
![Page 22: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/22.jpg)
Results
Returned scores: 83
participated successful: 60 (72%)
participated: 23 (28%)
Composition
NSCLC resection specimens, 9 cases
(One case excluded; 'equivocal')
Category 0, TC < 1%: 2 cases
Category 1, TC 1%-49%: 2 cases
Category 2, TC 50-100 %: 5 cases
Results-form: Category, not percentage
QuIP GmBH, used with permission
Open ring trial; cases, results
![Page 23: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/23.jpg)
Employed clones
QuIP GmBH, used with permission
![Page 24: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/24.jpg)
Deviating scores, stainings
Slide-sets returned n=18Correct staining / score n= 1Scoring incorrect n= 4Staining incorrect n= 13
A B
QuIP GmBH, used with permission
![Page 25: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/25.jpg)
QuIP - summary
87 participants, 83 results (95%), 60 successful (72%)
Success possible with CTAs and LDTs
LDTs: Different clones may be used
(E1L3N particullary successful)
Deviating results: calibration > interpretation
QuIP: 2nd open ring trial PD-L1 in NSCLC ongoing,
n=97 participants
QuIP GmBH, used with permission
![Page 26: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/26.jpg)
- TMAs of PD-L1+ cell-lines:
useful for LDT-calibration
- Here: Reference (22C3),
CTA-analogue (QC+),
divergent (QC-)
LDT-calibration with cell-lines
Scheel et al. submitted
![Page 27: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/27.jpg)
Summary
Internal ring trials
Reproducible PD-L1 scoring and staining possible
Assays: Reliable;
22C3 ≈ 28-8 ≈ SP263; ≠ SP142
LDTs: LDTs ≈ 22C3/28-8 possible;
6/11 protocols successful;
calibration essential
Open ring trial
60/83 successful (72%)
Success possible with CTA or LDT
LDTs: Different antibodies possible
![Page 28: Quality control of PD-L1 in NSCLC - Results of two German ...nordiqc2017.dk › wp-content › uploads › 2_NordiQC_Scheel_2017-06-04.pdfDr. Andreas Scheel University Hospital Cologne](https://reader036.vdocument.in/reader036/viewer/2022070803/5f033bf87e708231d4083354/html5/thumbnails/28.jpg)
Dr. Andreas ScheelUniversity Hospital Cologne
Institute of PathologyKerpener Str. 62, 50937 Cologne
Thank you for
your attention!
University Hospital ColognePathology IHC-Lab: Julia BongardAlexandra FlorinStefanie MarxMarion MüllerUrsula RommerscheidCorinna SeidelSusan Zupp
DGP PD-L1 IHC Harmonisation trialPathologists: Industry partners:Reinhard Büttner BMSGustavo Baretton MSDManfred Dietel RocheLukas Heukamp AstraZenecaKorinna Jöhrens VentanaThomas Kirchner Dako / AgilentIver Petersen Targos Molecular PathologySimone ReuJosef RüschoffAndreas ScheelHans-Ulrich SchildhausPeter SchirmacherMarkus TiemannArne WarthWilko Weichert
Lung Cancer Group Cologne (LCGC)http://lungcancergroup.de/Jürgen WolfRieke FischerMeike ThuratAnna KostenkoLucia NogovaSebastian MichelsMatthias Scheffler...
QuIP GmbHhttp://www.quip-ringversuche.deManfred DietelKorinna JöhrensJörg MaasThomas Pilz...