Quality Control Testing of Screen Speed. Quality Control Testing of Screen Speed. QC testing of screen speed should occur on acceptance and then yearly. Evaluate first whether similar cassettes marked with the same relative speed are the same using the following procedure. Procedure . - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Quality Control Testing of Screen Speed

Quality Control Testing of Screen Speed

Page 2: Quality Control Testing of Screen Speed

Quality Control Testing of Screen Speed QC testing of screen speed should

occur on acceptance and then yearly.

Evaluate first whether similar cassettes marked with the same relative speed are the same using the following procedure

Page 3: Quality Control Testing of Screen Speed

Procedure Make an exposure of a step-wedge or

homogenous phantom onto an image receptor so that the center of the image has an optical density of about 1.5

Expose each image receptor to the same technical factors.

Process, and take optical density readings of the same center area in each. If all have the same relative speed, the optical density should not vary by more than +-0.05

Page 4: Quality Control Testing of Screen Speed
Page 5: Quality Control Testing of Screen Speed

Screen Uniformity Testing During manufacturing processes,

inconsistencies may occur in which the phosphor layer is applied more thickly at one portion of the screen than at another.

Sometimes during cleaning excessive rubbing may remove more of the phosphor layer.

Screen uniformity testing should be performed on acceptance and then yearly.

Page 6: Quality Control Testing of Screen Speed

Procedure Make an exposure of a step-wedge or

homogenous phantom onto an image receptor so that the center of the image has an optical density of about 1.5

Process and take optical density readings in the center and in each of the four quadrants of the image. It should not be more than +-0.05.

Any film/screen image receptors that exceed this limit should be removed from service.

Page 7: Quality Control Testing of Screen Speed

Spectral Matching

The film and intensifying screen should match each other because the film are sensitive to a specific color.

A blue-violet emitting screen phosphor should be used with a monochromatic films.

A green-emitting screen phosphor should use with an orthochromatic film.

Page 8: Quality Control Testing of Screen Speed

Panchromatic

Is a type of black-and-white photographic film that is sensitive to all wavelengths of visible light.

A panchromatic film therefore produces a realistic image of a scene. Almost all modern photographic film is panchromatic,

Page 9: Quality Control Testing of Screen Speed
Page 10: Quality Control Testing of Screen Speed

Orthochromatic

Orthochromatic photography refers to a photographic emulsion that is sensitive to only blue and green light, and thus can be processed with a red safelight.

It is used on all radiographic films.

Page 11: Quality Control Testing of Screen Speed

Screen Resolution

Intensifying screens should be able to demonstrate clear images of patient anatomy so that the proper diagnosis can be obtained.

The ability of an imaging system to accurately display images.

There are two types of resolution Contrast resolution Spatial resolution

Page 12: Quality Control Testing of Screen Speed

Contrast Resolution

Is the ability of an imaging system to distinguish structures with similar x-ray transmission as separate entities (in short separate shades of grays)

It is affected by the sensitivity of the image receptor speed and the amount of radiographic mottle (noise).

If the radiographic mottle is increased, contrast resolution is decreased.

Page 13: Quality Control Testing of Screen Speed
Page 14: Quality Control Testing of Screen Speed

The Relationship Between The Quantum Noise, Contrast Resolution And Screen Speed Is that the faster the screen speed

screens, the lower the mAs values are used, which in turns increases the quantum mottle and lowers the contrast resolution.

Page 15: Quality Control Testing of Screen Speed

Spatial Resolution

Is the ability of an imaging system to create separate images of closely spaced objects.

In other words, do the two object appear sharp and clear, or do they blur together?

This is determined by the amount of light diffusion that occurs between the screen and film.

Page 16: Quality Control Testing of Screen Speed

Spatial Resolution

It is affected by: Screen thickness Phosphor crystal size Film/screen contact.

The most common method of measuring spatial resolution is the spatial frequency.

The unit of spatial frequency is the line pairs per millimeter (lp/mm).

Page 17: Quality Control Testing of Screen Speed

Line Pairs Per Millimeter Resolution Chart. a line pair is a space and line each

being 0.1mm wide. The greater the line pairs per

millimeter value the smaller the object that can be imaged and the better the spatial resolution.

The human eye can read up to 5lp/mm, but most screen system cannot provide this level of spatial resolution.

Page 18: Quality Control Testing of Screen Speed
Page 19: Quality Control Testing of Screen Speed
Page 20: Quality Control Testing of Screen Speed
Page 21: Quality Control Testing of Screen Speed

Other Methods Of Measuring Spatial Resolution Point spread function (PSF) Line spread function (LSF) Edge spread function (ESF) Modulation transfer function (MTF)

Page 22: Quality Control Testing of Screen Speed

Point Spread Function Is a graph that is obtained with a pinhole

camera and a microdensitometer. The pinhole camera creates a black dot in

the center of a film and microdensitometer is used to take readings.

The values are plotted on a graph versus the distance from the center of the point.

The narrower the peak on the graph, the better spatial resolution and image quality.

Page 23: Quality Control Testing of Screen Speed

Line Spread Function Is a graph that is more accurate and

easier to obtain than the PSF graph. It requires an aperture with a slit that

is 10um wide instead of the pinhole camera.

The density readings are taken of the centerline and plotted.

Page 24: Quality Control Testing of Screen Speed
Page 25: Quality Control Testing of Screen Speed

Edge Spread Function

It requires a sheet of lead to be placed on a cassette and exposed.

Density readings are taken at the border between the black and white areas and plotted on a graph.

Page 26: Quality Control Testing of Screen Speed

Modulation Transfer Function Is a numeric value that is used to

measure the spatial resolution and is obtained from the LSF graph with a mathematical process known as fourier transformation.

Page 27: Quality Control Testing of Screen Speed
Page 28: Quality Control Testing of Screen Speed
Page 29: Quality Control Testing of Screen Speed

Screen Condition

Intensifying screens must be free of dirt, stains and defects to properly function.

A regular schedule at least every 6 months of screen cleaning with an antistatic solution should be a standard department policy.

A UV light may be used to examine the surface of the screen if there are any stains.

Page 30: Quality Control Testing of Screen Speed

Why do we need to have a QA/QC on Just an intensifying screen?

Page 31: Quality Control Testing of Screen Speed

Laboratoryfilm/screen contact test

Page 32: Quality Control Testing of Screen Speed

film/screen contact test

Frequency of test Yearly As necessary

Equipment required Cassette to be tested Test tools:▪ Box of paper clips or sheet of perforated zinc or

fine wire mesh, large enough to cover a 35 x 43cm (14 x 17” film), with a square hole, about 10cm from one edge, approximately 2 to 2cm square)

Page 33: Quality Control Testing of Screen Speed

Method Load the cassette to be tested and

place it face up on the tabletop Cover the whole of the cassette with

the test tool (if cassette distribute evenly).

Set a FFD (SID) of 150cm[60”] ( the longer FFD (SID) reduces geometric unsharpness)

Collimate to cover whole of cassette Make exposure using a 50kVp and

6mAs Process film

Page 34: Quality Control Testing of Screen Speed

Evaluation If a densitometer is available measure the image. Inspect the image, looking for areas that look

blurred. A noticeable area of unsharpness could be

caused by: Damaged cassette Screen packing, deterioration An air pocket

When using a close mesh wire test the poor film/screen contact areas may also have a higher density.

Page 35: Quality Control Testing of Screen Speed