quantitative analysis of bpa and bpf - christina drescher
TRANSCRIPT
Quantitative Analysis of Bisphenol A and Bisphenol F Leaching Out of Baby Feeding Bottles Using Capillary Electrophoresis
Christina DrescherSupervisor: Dr. Kingsley Donkor
Introduction Bisphenol A (BPA) and Bisphenol F (BPF) are endocrine disrupters
that can easily leach out of plastic materials baby feeding bottles
An endocrine disrupter can easily lead to harmful health effects due to its ability to mimic bodily hormones
Area of high interest in the plastic industry, especially when regardingthe safety of infants
The goal of this research is to develop a sensitive and rapid capillaryelectrophoresis (CE) method to analyze and quantify Bisphenol A andBisphenol F within baby feeding bottles from Indonesia and China.
CE experimental conditions were established to successfully separateBPA and BPF with good resolution and timely migration.
Although there has been some research conducted with regards toBisphenol A and F in baby bottles, to the best of our knowledge, this is the first analysis done using capillary electrophoresis (CE).
Method Development Four months of method development was involved for determination of
a CE separation method for BPA and BPF Gallet-Ayala et al. method1 was used as a starting point Optimization of the CE buffer was neededThe buffer comprised of SDS (sodium dodecyl sulfate) and phosphate
adjusted to pH 2.5After multiple trials and increasing concentrations of SDS, an optimum
buffer concentration was obtained for timely separation of BPA and BPF (Figure 2 & Table 1)
ExperimentalNeat Calibration StandardsFive BPA standards were prepared ranging from 200-900 ppb BPAEach standard was prepared from a 1000 ppb BPA stock solution and
diluted with 18 MΩ H2OEach standard contained 50 µL of dilution buffer (Table 1) Sample PreparationBPA was leached from baby bottles according to the flow chart2
The boiling water bath was placed in a shaker to ensure full BPA exposure to the water sample within the bottles
Key steps for this procedure are shown in Figures 3-5Total leaching/extraction procedure took 6 hStandard Addition Standards36 standard addition standards were prepared (3 real replicates for both
China and Indonesia bottles)Known BPA concentration ranged from 0-500 ppbEach standard also included 50 µL of dilution buffer, 100 µL sample,
and the remaining diluted with 18 MΩ H2OTable 2: CE Beckman P/ACE MDQ Capillary Electrophoresis System Instrument Parameters
DiscussionThe qualitative separation of BPA and BPF was
achieved (Figure 2) The quantification of BPA within the baby bottle
samples was also determinedThe leaching process followed was effective in
extracting the BPA from the samplesStandard addition was found to be the most
advantageous technique to determine quantitative results for such low concentrationsat ppb level
The BPA levels found were below the BPA threshold, but still raise concern with regards to infants
Data and Results
Thank you to Dr. Kingsley Donkor and to Laiel Soliman for all of their continuous support, knowledge, and wisdom.
Acknowledgements
Conclusions The method developed for BPA detection and quantification using
capillary electrophoresis was successful Both types of feeding bottles were determined to contain BPA The bottles from Indonesia appear to leach a significantly higher
amount of BPA compared with the bottles from China Quantification of BPF within the bottles has not yet been completedThe method is not yet flawless for the detection of BPA and BPF, and
future work is needed
Future Work Continued replication of this method is needed to confirm any
results found to date. Collect more data to give full certainty of concentrations determined.
Continued work adjusting the method for the separation of BPA and BPF within a sample
Analysis of other samples, aside from baby bottles, such as foodcontainers (canned tomato soup and paste)
References 1. Gallart-Ayala, H.; Nunez, O.; Moyano, E.; Galceran, M. T.
Electrophoresis, 2010, 1550-1559
2. Lateef, S. S. Agilent Applications, 2011, 1-11
Table 3: Calibration and standard addition curve data
Curve Type Equation R2
Neat BPA Calibration y = 3.3344x + 166.95 0.9948
Indonesia Sample Standard Addition y = 0.9301x + 268.56 0.9661
China Sample Standard Addition y = 1.3373x + 225.17 0.9560
Table 4: BPA concentrations determined for each baby feeding bottle
Sample[BPA]found µg/kg bw
Country ppb avg 12 month infantIndonesia
(150 mL bottle) 9.62 0.145China
(250 mL bottle) 3.37 0.084Table 1: Optimum Buffer Optimum Concentration (mM)
Buffer Component mM, diluted with 18 MΩ H2OSeparation/Rinse Buffer Sample Dilution Buffer
SDS 75 150Phosphate Buffer
(pH 2.5) 25 25
Figure 3: Solid Phase Extraction Apparatus
Figure 1: Structures of BPA and BPF
Figure 2: BPA and BPF optimized CE separation electropherogram
Figure 4: Ice bath to cool samples post-
leaching
Figure 5: N2 Gas Evaporation Apparatus
Figure 6: Beckman Coulter P/ACE MDQ CE with samples loaded
Figure 7: CE Standard Addition BPA electropherogram
Capillary Fused silica, 50 µm O.D. x 50 cm total length (40 cm to detector)Operating Temperature 25°C
Detection UV, 214 nm (indirect)Rinse pressure 20 psi: 3.0 min (0.1 M NaOH), 2.0 min (Water), 5.0 min (Buffer)
Injection of Sample Pressure, 0.5 psi for 15.0 sSeparation voltage + 25 kV
Separation time 10 min, 0.17 min ramp time
Minutes
4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4 7.6 7.8 8.0
AU
0.0006
0.0008
0.0010
0.0012
0.0014
0.0016
0.0018
0.0020
Minutes
16.4 16.6 16.8 17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6 18.8 19.0
AU
-0.03
-0.02
-0.01
0.00
0.01
0.02
0.03
0.04
BPA
BPF