Quick Guide Component R12 / R6 Cat. No. : CP-R12 / CP-R6
Setting Temperature Setting temperature -10 to 40℃ (R12) / -20 to 40℃ (R6) Setting temp. limit
Lid OpenStopStart
Max. RPM : 4,800 rpm
Max. RCF : 7,470 xg
Max. RPM : 5,000 rpm
Max. RCF : 7,261 xg
ComponentR12
ComponentR6
Serial No. Installation Date
Display ScreenStatus of RPM / RCF / Time / Temperature / Rotor ID / ACC / DEC / Program
Setting SpeedPress once : Setting RPMPress twice : Setting RCF
Setting TimeUp to < 100 hr Continuous run (00:00)
View SettingsVerifying the setting parameters during centrifugation
Program Save / Call Saving a program (up to 100)
Calling up a program
Setting the Rotor ID manually Enter the rotor ID. manually, when rotor id is not identified automatically
Setting ACC / DEC Rate ACC (0 ~ 9 steps, Max.9)
DEC (0 ~ 9 steps, Natural DEC.0)
Display & Control Panel
Saving / Calling Upa Program
Setting Run Parameters(RPM / RCF / Time / Temperature / ACC / DEC)
Temperature range that stops operation when the temperature is lower or higher than the settemperature.
Setting Temp. Limit
Press a parameterto be changed
Enter the desired values on the keypad
Press theEnter button
Press theEnter button
Enter the desired program number on the keypad
Press theSave or call button
Press theTemp. limit button
Press theEnter button
Enter the desired temperature range on the keypad
Connect the device to voltage
sources, which correspond to the
electrical requirements on the label
attached to the device.
Check the wall outlet is earth-
grounded, surge-protected and
can supply enough power.
Connect the power cable to the
distribution box, or install the plug
and socket that are packed with
the device.
Switch off the device after use.
Unplug the power plug before
cleaning or unused for a long
period of time.
Check the Line Voltage Grounded Outlet Power Connection Switch Off the Device
Quick Guide Component R12 / R6 Cat. No. : CP-R12 / CP-R6
Allow a safety clearance of 30 cm around the centrifuge during operation.
Rotors
Spread the grease lightly on the bucket where it contacts the rotor and the pivot.
Remove any dust regularly from the ventilation slits on the left side of the device.
There is a drain hose at the bottom of the device. The condensation water is automatically drained through the hose. Check and remove it regularly.
Component R6Component R12
30 cm
30 cm30 cm
Emergency Lid Open
If the centrifuge lid cannot be opened,insert the provided T-wrench into the emergency release hole and turn it clockwise. (Before activating emergency open, switch off the device and wait for the rotor to stop.)
Wipe up the moisture in the chamber with a dry cloth after use. Keep the lid open when not in use.
Clean the chamber and rotor by wipingwith a cloth soaked in mild detergent and dry completely.
Keep the Chamber Dry
Cleaning
Check the rotor, buckets and tubes are free of scratchand crack before operation.
Check the Accessories for Abnormalities
Load tubes symmetrically into the rotorwith identical amount.
Load Tubes Symmetrically
Place the lid on the wind shield rotor before centrifugation.
Check the Lid
After relocating the device or loading the rotor, balance them by using a leveler. If necessary, level the device by adjusting the ‘Leveling foot’ on the bottom of the device.
Balancing the Centrifuge and Rotor
3.0KVA (Component R12)
5.0KVA (Component R6)
220~240V 50/60Hz
Maintenance
Electrical Safety
Cautions for Use
Rev.1
Proudly made by Jiyoung Shim, Hyewon Jeong
Order & Inquiry l +82-2-3472-0727 l [email protected] l www.ihanil.com
Service l +82-2-3452-8966 l [email protected]
Before Operation After Operation
Maintenance of Swing-Out Rotor
Remove Dust Empty the Condensation Water
Allow a Safety Clearance of 30 cm
WS1000-6WS2400-6 6 x 1,000 mL / Max. 5k 6 x 2,400 mL / Max. 4.8k