rad 354 chapt. 28 multi-slice ct in short – ct is a thin band (fan) of radiation directed toward...

16
RAD 354 Chapt. 28 Multi-slice CT • In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt. is measured by a receptor and the response transmitted to a computer for reconstruction of the image(s)

Upload: gloria-vanessa-thornton

Post on 17-Jan-2016

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

RAD 354 Chapt. 28 Multi-slice CT

• In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt. is measured by a receptor and the response transmitted to a computer for reconstruction of the image(s)

Page 2: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

“Slip rings”

Page 3: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Dates/Activities to Remember

• 1970 – Sir Godfrey Hounsfield,Ph.D.., demonstrated the technique of CT scanning – “EMI” head unit

Page 4: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Dates/Activities Con’t.

• 1982 Dr. Hounsfield and Dr. Alan Cormack share the Nobel Peace Prize in physics

• CT Generations and important items:– 1st – translate-rotate; “pencil beam” - one full

rotation for 1 image; head only; 5 min. per image!– 2nd – translate-rotate, but with a “fan” beam;

produced INCREASED INTENSITY at the edge of the beam – “bow-tie” filter used to remedy this; 30 sec. image time

Page 5: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Generations con’t.

– 3rd – used ARRAY of DETECTORS – produced “ring artifacts”

– 4th rotate – stationary, fan beam, detector array; short image time and NO re-con time!

– 5th is NOW here! “Spiral/slip ring” (It is slip ring technology that has enabled this latest generation!)

Page 6: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Major Components of CT

• Gantry – tube-receptor array, collimators (pre-pt. and pre-detector), generator, pt. couch

• Computer – microprocessor – recon time was a BIG issue until recently

• Operating console – usually two monitors – one for RIS/HIS and the second for the imaging portion (often times a “remote” monitor is set up at the radiologist’s desk – can really be a hassle)

Page 7: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Radiation Dose Comparisons

Page 8: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Current – 5th Generation Terms

• Spiral• Helical• 16/32/64 slice• Slip Rings CT

Page 9: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Interpolation/Extrapolation

• Interpolation – information ALONG the axis WITHIN established values

• Extrapolation – information OUTSIDE axis values established, BUT an ESTIMATED value based upon previously measured parameters IF they continue

Page 10: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

“Spiral” Terms

• Pitch – relationship between couch movement and x-ray beam collimation (pitch ratio)– Increasing pitch ratio enables much MORE of the pt. to be

imaged in ONE BREATH HOLD! Breath holds = misrepresentation of the images.

• Slip ring technology made “spiral CT” possible• Three slip rings in the gantry (usually)

– One for hi voltage power to the tube and hi voltage generator– Low voltage power to the rotating gantry portion– Transfer of digital data from gantry receptors to the computer

Page 11: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Multi Slice CT

• Multi detector arrays– Two at once with one beam = two slices per 360

degrees– With “spiral” enables two detectors to produce 16

slices• Envision a person on the table/couch in a

screw/nut fashion going through the gantry

Page 12: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Spiral/slip ring

Page 13: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Tube (fan beam)/receptors

Page 14: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Z-axis

Page 15: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Maximum Intensity Projection (MIP)

• Pre selecting pixel levels to be viewed AFTER obtaining a “rubic’s cube of information”– Then only exhibit those at the PRE-DETERMINED

value – like an MRA or CTA

Page 16: RAD 354 Chapt. 28 Multi-slice CT In short – CT is a thin band (fan) of radiation directed toward the pt. and the remnant radiation emitted from the pt

Spiral Advantages of Conventional CT

• Motion blur is reduced – fewer artifacts• Lowered imaging time (fewer breath holds

required = less motion and less mis-representation)

• Larger volumes can be imaged with LESS part overlap artifacts