Radiation Biodosimetry Test Development Continues Apace
Dr. Lynne K. WathenApril 18, 2018
![Page 2: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/2.jpg)
2Saving Lives. Protecting Americans.
21st Century: An Increasingly Complex & Dangerous World
![Page 3: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/3.jpg)
3Saving Lives. Protecting Americans.
ASPR’s Raison D’etre: “Unity of Command”Public Law 109-417 Pandemic & All-Hazards Preparedness Act
![Page 4: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/4.jpg)
4Saving Lives. Protecting Americans.
Save Lives,Protect
Americans from 21st
Century Health Security Threats
ASPR Mission
![Page 5: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/5.jpg)
5Saving Lives. Protecting Americans.
ASPR Priorities for Building Readiness for 21st Century Threats
Strong Leadership
Medical Countermeasures
Enterprise
Regional Disaster Health Response
System
Public Health Security Capacity
![Page 6: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/6.jpg)
6Saving Lives. Protecting Americans.
The BARDA Model
• BARDA develops and makes available medical countermeasures (MCMs) by forming unique public-private partnerships with industry partners
![Page 7: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/7.jpg)
7Saving Lives. Protecting Americans.
Radiation BiodosimetryProgram
Two types of biodosimetry tests are under development:- Point of Care Triage Screening Tests to discern individuals needing
medical evaluations from those who can evacuate- High Throughput Laboratory Tests to report the absorbed dose an
individual received and inform further care
Objective: Develop rapid, accurate FDA-cleared biodosimetry diagnostic assays/systems to inform patient management, improve health and psychosocial outcomes,
and save lives.
For Official Use Only
![Page 8: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/8.jpg)
8Saving Lives. Protecting Americans.
Biodosimetry Target Product Profiles
Point of Care Device (POC)
High Throughput Device (HT)
Type of result: Screening/Qualitative Quantitative (desired accuracy ± 0.5)
CONOPs: Initial Triage / Sorting Injury Assessment / Treatment Tool
Exposure level: 2 Gy - threshold Range: 0 – 10 Gy
Ease of operation: Easy to operate, minimal complexity, requires minimal training, CLIA waived
Laboratory instrument—more labor intensive, requires training
Device Characteristics: Integrated components—no separate sample preparation
May include separate components as needed. High automation desired.
Intended use: Tents, shelters, open settings Labs, hospitals, fixed facilities
# Patients / Event Up to 1,000,000 within 7 days Up to 400,000 within 7 days (may need multiple assessments)
Time to result: Rapid but individual sample result(15 to 30 minutes)
Up to 24 hours
![Page 9: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/9.jpg)
9Saving Lives. Protecting Americans.
Post-irradiation Symptoms And Treatment Strategies
Symptoms and Onset Mild(1-2 Gy)
Moderate (2-4 Gy)
Severe(4-6 Gy)
Very Severe (6-8
Gy)
Lethal(> 8 Gy)
Vomiting OnsetIncidence
≈ 2 hrs.10- 50%
≈ 1-2 hrs.70-90%
≈ 1 hr.100%
≈ 30 min.100%
≈ 10 min.100%
Diarrhea OnsetIncidence
None None 3-8 hrs.< 10%
1-3 hrs.> 10%
< 1 hr.≈ 100%
Headache OnsetIncidence
Slight Mild 4-24 hrs.50%
3-4hrs.80%
1-2 hrs.100%
Consciousness
OnsetIncidence
Alert Alert Alert Impairment Lost
Body Temperature
OnsetIncidence
Normal ↑ 1-3 hrs.10- 80%
↑ 1-2 hrs.80-100%
↑ < 1 hr.100%
↑ < 1 hr.100%
TreatmentStrategy
Outpatient Observation
General hospital
treatment
Specialized hospital
treatment
Specialized hospital
treatment
Palliativetreatment
Geographical location at the time of the blast and lymphocyte depletion kinetics will also help determine approximate absorbed dose.
![Page 10: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/10.jpg)
10Saving Lives. Protecting Americans.
Biodosimetry Pivotal Testing ProcessIntense pre-validation of the tests using samples from:• Normal Humans, ages 2 to 87, with percent ethnicities representative of the US population
• Individuals with potentially confounding conditions such as immunocompromised, pregnant, trauma, burn, diabetes, rheumatoid arthritis, sepsis, and radiotherapy
• Pre-Transplantation patients receiving fractionated irradiation
• Non-human primates (NHP) receiving total body irradiation either in single or fractionated doses
• Humans and NHPs after G-CSF or Leukine administration
• A host of common substances were also spiked into human blood and tested for potential assay interference (such as bilirubin, human serum albumin, human IgG, L-Ascorbic Acid, hemoglobin, acetylcysteine, captopril disulfide, and others depending on the assay chemistry)
![Page 11: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/11.jpg)
11Saving Lives. Protecting Americans.
Regulatory Strategy
• Demonstrating similarity of gene, protein, or cytological response to single and fractionated dose irradiation across species
• Seeking pre-Emergency Use Authorization status from the United States Food and Drug administration (FDA) One pre-EUA package filed in 2017 and under review
• Pursuing 1st biodosimetry in vitro diagnostics approval from FDA
NHP in vivoSingle Dose
NHP in vivo Fractionated
Human in vivo Fractionated
Human in vivoSingle Dose
![Page 12: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/12.jpg)
12Saving Lives. Protecting Americans.
Point of Care Biodosimetry
Completing Verification PhaseAnalytical and clinical validation will
follow
For Official Use Only
Developer Point Of CareTechnology Type
Sample Throughput per Instrument (per
hour)
SRI InternationalProtein Expression
ImmunoassayDual Lateral Flow w/ Reader & Cell
Extractor
≈ 24
![Page 13: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/13.jpg)
13Saving Lives. Protecting Americans.
Biomarker Panel is Predictive Beyond Day 7 in NHPs
0 1 14 16 18 21 24 0 14 16 18 21 24 26(Day)
Protein 1 elevated out to day 18 in 4 Gy samples
0 1 14 16 18 21 24 0 14 16 18 21 24 26(Day)
Protein 2 elevated out to day 18 in 4 Gy samples
For Official Use Only
![Page 14: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/14.jpg)
14Saving Lives. Protecting Americans.
High Throughput Laboratory-basedBiodosimetry Projects
For Official Use Only
Completing Verification PhaseAnalytical and clinical validation will
follow
Developer HTTechnology Type
MRI Global/AZState U/ Thermo
FisherGene expression Semi-automated
QuantStudio Dx
ASELL LLC Cytology -micronuclei
Semi-automated including Cytology
Microscopes
DxTerity Gene expressionSemi-automated
including ABI 3500 Dx
![Page 15: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/15.jpg)
15Saving Lives. Protecting Americans.
DxTerity Single Dose NHP Data
Confidence Interval Plot of Estimated Dose versus Actual Dose of Cross-validated NHP Single Dose (SD) data (893 samples) for all times post-irradiation combined (1 to 7 Days).
![Page 16: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/16.jpg)
16Saving Lives. Protecting Americans.
Human to NHP Fractionated Dose Comparison
For Official Use Only
ASU ARAD Biodosimeter (15 biomarkers)
![Page 17: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/17.jpg)
17Saving Lives. Protecting Americans.
ASELL Human Micronucleus Assay Analytical Performance
Irradiation of human blood samples ex vivo
• Apparently healthy adult human subjects used to establish current analytical performance
• Results demonstrate accuracy with 90% of results within 1.2 Gy of actual dose across 2-6 Gy
Error bars indicate +/-1 SD from the mean at each actual dose level.
For Official Use Only
![Page 18: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/18.jpg)
18Saving Lives. Protecting Americans.
Biodosimetry Test Next Steps
Developers are proceeding to: Product Validation / Clinical Testing Pre-Emergency Use Authorization filing FDA clearance Acquisition of Initial Test Stockpile Maintenance of Stockpile through 2026
For Official Use Only
![Page 19: Radiation Biodosimetry Test Development Continues Apace · 2018-05-22 · Radiation Biodosimetry Test Development Continues Apace Dr. Lynne K. Wathen April 18, 2018](https://reader031.vdocument.in/reader031/viewer/2022011903/5f153e4b603be963fe6458a8/html5/thumbnails/19.jpg)
19Saving Lives. Protecting Americans.
Conclusions
BARDA is advancing high-throughput and point of care biodosimetry tests toward:
• Attaining Pre-Emergency Use status to continue preparing the nation for a radiological incident
• Seeking FDA clearance for at least one high-throughput and one point of care biodosimetry test to enhance our preparedness position
• Developing a test implementation strategy with CDC, state and local stakeholders to help inform patient management, improve health and psychosocial outcomes, and save lives