Page 1: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Vision :IMC aspires to be a leader in applied medical sciences, health care education and research.

Dr. Mohammed Emam

RAD 481

Radiotherapy physics & Equipments

Lecture’s Title: Beam Modification devices in Radiotherapy

Page 2: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Overview

Mission: IMC is committed to develop highly qualified and competent graduates able to provide leadership and excellence in

services to meet the health needs of the nation and the global community through wide range programs, nationally competitive

faculty, state-of-the art infrastructure, research applications, and a diverse environment with enriched engagements.

• Beam Modification defined as desirable modification in thespatial distribution of radiation, within the patient, byinsertion of any material in the beam path.

Page 3: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Types of beam modification

• There are four main types of beam modification:

– Shielding: To eliminate radiation dose to some special parts of thezone at which the beam is directed.

– Compensation: To allow normal dose distribution data to be appliedto the treated zone, when the beam enters a or obliquely throughthe body or where different types of tissues are present.

– Wedge filtration: Where a special tilt in isodose curves is obtained.

– Flattening: Where the spatial distribution of the natural beam isaltered by reducing the central exposure rate relative to theperipheral.

Page 4: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Problem in beam modification

• Radiation reaching any point, is made up ofprimary and scattered photons.

• Any introduction of the modification devicesresults in alteration of dose distribution, dueto these two phenomena.

• The phenomena scattering results in an“blurring” of the effect of the beammodification.

• Scattering is more in kilovoltage radiationthan in megavoltage radiation therapy.

Page 5: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Types of beam modification devices

• Field blocking and shaping devices:– Shielding blocks.

– Custom blocks.

– Asymmetrical jaws.

– Multileaf collimators.

• Compensators.

• Beam spoilers

• Wedge filters.

• Beam flattening filters.

• Bolus

• Breast cone.

• Penumbra trimmers.

• Electron beam modification

Page 6: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Shielding

• Since radiation attenuation is exponential and because of scattering, complete shielding can never be achieved.

• The aims of shielding are:– To protect critical organs.

– Avoid unnecessary irradiation to surrounding normal tissue.

– Matching adjacent fields.

Page 7: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Shielding

• An idea shielding material should have the following characteristics:

– High atomic number.

– High-density.

– Easily available.

– Inexpensive.

• The choice of the shielding material is also dictated by the type of beam being used!!

• The most commonly used shielding material for photons is lead.

Page 8: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Shielding

• The thickness used depends upon the energy of the radiation.

• For practical purposes, the shielding material which reduces beam transmission to 5% of its original is considered acceptable.

• The term half value-layer is a convenient expression for the attenuation produced by any material. Half-value layer is defined as the thickness of material, which will reduce the intensity of the primary beam by 50%.

Page 9: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Shielding

100%

50%

250 KV

4 MV

Lesser amount of scatteredradiation with megavoltage

radiation means that the attenuation produced by

shielding is also more.

The higher scatter contributionto the overall dose results in lower dosage adjacent to the shielded area in kilovoltage

radiation.

Beam energy Required lead thickness

4 MV 6.0 cm

6 MV 6.5 cm

10 MV 7.0 cm

Co60(1.25 MeV)

5.0 cm

Page 10: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Shielding

• In kilovoltage radiation shielding is readily achieved by placing sheets of lead on the surface directly.

• This is necessary, because of the lower penetrating power of the beam.

• In megavoltage radiation,

– Thicker blocks used.

– Placed higher up in shadow trays (15 -20 cm).

– Avoids increase in skin dose due to electron scatter.

– Also impossible to place the heavy block on the body !!

Page 11: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Custom blocks

• Material used for custom locking is known as the Lipowitz metal or Cerrobend.

• Melting point 70°C.

• Density 9.4 g /cm3 at 20°C (83%of lead).

• 1.21 times thicker blocks necessary to produce the same attenuation.

• Most commonly thickness of 7.5 cms used.

Lead,

26.70%

Bismuth,

50.00%

Cadmium,

10.00%Tin,

13.30%

Bismuth Lead

Tin Cadmium

Page 12: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Custom blocks

Outline of the treatment field being traced on

radiograph using a Styrofoam cutting

device.

Electrically heated wire pivoting around a point(simulating the source)

cutting the styrofoam block

Cavities in the styrofoam block

being used to cast the Cerrobend

blocks.

Page 13: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Custom Blocks

Page 14: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Custom blocks

• Shielding blocks can be of two types:

– Positive blocks, where the central area is blocked.

– Negative blocks, where the peripheral area is blocked.

• A Diverging block means that the block follows the geometric divergence of the beam.

• This minimises the block transmission penumbra.

Page 15: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Special Shielding

Page 16: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Independent Jaws

• Used when we want to block of the part of the field without changing the position of the isocenter.

• Independently movable jaws, allows us to shield a part of the field, and this can be used for “beam splitting”.

• Here beam is blocked off at the central axis to remove the divergence.

• Use of independent jaws and other beam blocking devices results in the shift of the isodose curves.

• This is due to the elimination of photon and electrons scatter from the blocked part of the field.

Page 17: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Multileaf Collimators

• Multileaf collimators are a bank of large number of collimating blocks or leaves

• Can be moved automatically independent of each other to generate a field of any shape.

• 40 pairs of leaves or more having a width of 1 cm on less (projected at the isocenter).

• Thickness = 6 – 7.5 cm• Made of a tungsten alloy. • Density of 17 - 18.5 g/cm3.• Primary x-ray transmission:

– Through the leaves < 2%.– Interleaf transmission < 3%.– For jaws 1%– Cerrobend blocks 3.5% .

Page 18: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Multileaf Collimators

• MLC systems may have double focus or single focus leaves.

• The latter are shaped to match the radiation beam along the Y axis only as the upper end is narrower than the lower.

• Single focus leaves are also rounded at the ends.

• This can lead to significant beam transmission (20%) when the leaves abut each other.

• Both designs are to ensure a sharp beam cut off at the edge.

• In order to allow fast interleaf movement, while reducing radiation transmission a tongue and groove design is often used.

• This design in turn leads to some under dosingin the region of the tongue (17 – 25%).

Beam

Tongue

Page 19: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Multileaf collimators

YX Y

X

Page 20: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Multileaf Collimators

• The degree of conformity between the planned field boundary and the boundary created by the MLC depends upon:

– Projected leaf width.

– Shape of target volume.

– Angle of collimator rotation.

• RCI = Treated Volume (inside 95% isodose curve) / PTV

• The direction of motion of the leaves should be parallel with the direction in which the target volume has the smallest cross-section.

Page 21: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Multileaf Collimators• The advantages are:

– Time for shaping and inserting of custom blocks is not required.

– The hardening of beam, scattered radiation, and increase in skin doses and doses outside the field, as seen with physical compensators is avoided.

– Automation of reshaping and modulation of beam intensity in IMRT.

– MLCs can also be used to as dynamic wedges and electronic compensators (2D).

• The disadvantages are:

– Island blocking is not possible.

– Because the physical penumbra is larger than that produced by Cerrobend blocks treatment of smaller fields is difficult, as is the shielding of critical structures, near the field.

– The jagged boundary of the field makes matching difficult.

– Practically smaller fields are used because MLC carriages and secondary jaws don’t move during IMRT delivery making matching of fields necessary.

Page 22: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Compensators

• A beam modifying device which evens out the skin surface contours, while retaining the skin-sparing advantage.

• It allows normal depth dose data to be used for such irregular surfaces.

• Compensators can also be used for – To compensate for tissue

heterogeneity. This was first used by Ellis, and is primarily used in total body irradiation.

– To compensate for dose irregularities arising due to reduced scatter near the field edges (example mantle fields), and horns in the beam profile.

Notice the reduction in the hot spot

Page 23: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Compensators

• The dimension and shape of a compensator must be adjusted to account for :– Beam divergence.– Linear attenuation coefficients of the

filter material and soft tissue.– Reduction in scatter at various depths

due to the compensating filters, when it is placed at the distance away from the skin.

• To compensate for these factors a tissue compensator is always has an attenuation less than that required for primary radiation.

• As the distance between the skin and compensator increases the thickness ratio(h’/h) decreases.

h'

h

d

d

h’/h

1

Page 24: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Compensators

• The thickness ratio depends on:– Compensator to surface distance.– Thickness of the missing tissue.– Field size.– Depth.– Beam quality.

• Of these, the distance is the most important factor when d is ≤ 20 cm.• Therefore, a fixed value of thickness ratio (τ) is used for most compensator work (~ 0.7).• The formula used for calculation of compensator thickness is given by: TD x (τ/ρc), where TD is the tissue deficit

and ρc is the density of the compensator.• The term τ/ρc can be directly measured by using phantoms.• The term compensator ratio is the inverse of the thickness ratio. (ρc /τ ).

Page 25: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Compensators

Two-dimensional compensators

• Used when proper mould room facilities are not available.

• Thickness varies, along a single dimension only.

• Can be constructed using thin sheets of lead, lucite or aluminum. This results in production of a laminated filter.

Page 26: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Compensators

Three-dimensional compensators• 3-D compensators are designed to

measure tissue deficits in both transverse and longitudinal cross sections.

• Various devices are used to drive a pantographic cutting unit.

• Cavity produced in the Styrofoam block is used to cast compensator filters.

• Medium density materials are preferred to reduce errors.

• Various systems in use for design of these compensators are:– Moiré Camera.– Magnetic Digitizers.– CT based compensator designing

systems.

Page 27: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

CompensatorsCompensating wedges:• Compensating wedges are useful

where the contour can be approximated with a straight line for an oblique beam.

• Three important differences between compensating wedges and wedge filters are:– Standard isodose curves, can be

used – No wedge transmission factors

are required.– Partial field compensation can

be done.

Set up– At the filter-surface distance

calculated ≥ 20 cm. – Nominal SSD measured from a

plane perpendicular to beam axis touching the highest point in the contour.

– In SAD technique the depth of the isocenter is measured from the same elevated point only.

Page 28: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Beam Spoilers

• Special beam modification device where shadow trays made from Luciteare kept at a certain distance from the skin.

• Based on the principle that relative surface dose increases when the surface to tray distance is reduced.

• First used by Doppke to increase dose to superficial neck nodes in head and neck cancers using 10 MV photon beams.

% d

ep

th d

ose

Depth (cm)

D = 5

D = 10

D = 20

D = 40

D = Tray to surface distance

.4 .8 1.2

100%

Page 29: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge Filters• A beam modifying device,

which causes a progressive decrease in intensity acrossthe beam, resulting in tilting the isodose curves from their normal positions.

• Degree of the tilt depends upon the slope of the wedge filter.

• Material: tungsten, brass. Lead or steel.

• Usually wedges are mounted at a distance of 15 centimeters from the skin surface.

• The sloping surface is made either straight or sigmoid in shade.

• A sigmoid shape produces a straighter isodose curve.

• Mounted on trays which are mounted on to the head of the gantry.

Page 30: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters

• Types of wedge systems:– Individualized wedge .– Universal wedge.– Dynamic wedges– Virtual wedges – Pseudo wedges

• The two dimensions of wedges are important – “X” or width and “Y” or length.

• All wedges are aligned so that the central axis of the beam is at the central axis of the wedge.

• If the X dimension of field is longer then we can’t use the wedge without risking a hot spot!!

X

This area will have a hot spot.

Page 31: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge Filters• Individual wedges are useful

in Cobalt beams

• Using bigger wedges than necessary will reduce output of the machine → increased treatment time.

• The width (W) of the wedge is fixed and important.

• The same wedge can however be used for fields with lesser lengths or breadths.

• The wedge systems available are:

– 6W ( x 15)

– 8W ( x 15)

– 10W ( x 15)

• All systems have the following four angles 15°, 30°, 45°, 60°.

Page 32: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge Filters

• Universal wedges are designed so that the same wedge can be used with all field sizes.

• This is useful as it saves time.

• However not suitable for cobalt beams because of excessive reduction of beam output with smaller fields.

• Come in one size of 20 x 30 cms (except 60°).

• Wedge angles used are: 60°, 45°, 30° & 15°.

Page 33: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters

• The wedge isodose angle (θ) is the complement of the angle through which the isodose curve is tilted with respect to the central ray of the beam at any specified depth.

• This depth is important because the angle will decrease with increasing depth.

• The choice of the reference depth varies:– 10 centimeters.– 1/2 - 2/3rd of the beam

width.– At the 50% isodose curve

(kV).

θ

Page 34: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters• The presence of the wedge decreases

output of the machine.• WTF = Dose with the wedge/ Dose

without the wedge (at a point in the phantom, along the central axis of the beam).

• Usually measured at a suitable depth below the Dmax usually 5 -10 cms! This minimises the error in calculation of PDD.

• The resultant reduction in output results in an increase in the treatment time.

Page 35: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters• In some isodose charts used in cobalt machines the wedge

transmission factor is already incorporated, and no further correction is necessary.

• Use of wedge will result in a preferential hardening - more pronounced in case of linear accelerators.

• This is because the Co 60 beam is monoenergetic .

• For small depths (<10 cms) most of the calculation parameters however remain unchanged.

• At larger depths however, the PDD can be altered specially in case of linear accelerator beams.

Page 36: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge Filters

Page 37: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters

Page 38: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters - Design

This angle is the wedge angle.

% DD at each point is calculated and tabulated

A B C D E

A B C D E

Non wedge isodose

40 55 67 68 62

Wedge isodose 35 39 43 76 105

Ratio (W: NW) .87 .71 .79 1.1 1.7

Transmission ratio

- .38 .42 .66 1

mm Pb - 15.2 13.6 6.5 0

Page 39: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters• Wedged fields are generally used for

relatively superficial tumors.• Beams are usually directed from the

same side of the patient.• The broad edges of the wedges

should be aligned together.• The wedge angle choosen depends

on the angle between the central rays of the two beams also called the “hinge angle”(φ).

• Wedges:– Reduce the hot spots at the

surface – Rapid dose falloff beyond the

region of overlap.• The overlap region is also called the

“plateau region”.

• Thus the 2 factors on which the wedge angle is choosen are:– The hinge angle.– The wedge separation.

• The wedge angle that will make the isodose curves parallel to each other and the hinge angle bisector is obtained using the equation.

φ

Θ = 90 – φ / 2

Page 40: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters

Problem Solution

A different value of wedge angle is required for different beam angles

A small range of hinge angles can be covered by a given wedge angle without producing significant variation(±5%).

The body contour can be more curved as a result, the isodose curves are not obtained in the manner desired.

Compensators may be used to overcome the deficit or a different wedge angle can be used, so that part acts as a compensator.

Page 41: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Wedge filters• Dynamic wedges or “motorized

wedges” as they were called once had a 60° wedge mounted in the treatment head itself.

• This wedge was moved into the field for part of the time to create the wedge beam profile desired.

• Virtual wedges or dynamic enhanced wedges are moving jaws that are moved by computer control to create wedge beam profiles.

• However use has not resulted in significant improvement over conventional wedges.

• Fixed jaws can be used to produce pseudo wedges where part of the treatment field requiring greater dose would be irradiated using smaller field sizes.

Page 42: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Flattening filters

• A beam flattening filter reduces the centralexposure rate relative to that near the edge of the beam.

• Used for Linear accelerators.• Due to the lower scatter the isodose curves

are exhibit “forward peaking”.• The filter is designed so that the thickest part

is in the centre.• Material: copper or brass.• Penetrating power should not increase as this

will alter the PDD as well as reduce the flattening.

• In cobalt beams:– The beam is almost monoenergetic.– Source emits uniform radiation all

around.

Page 43: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Flattening filters

• The beam flatness is specified at 10 centimeters.• The extent of flatness should be ± 3% along the central axis of

the beam at 10 centimeters.• Should cover 80% or more of the field, or reach closer than one

centimeter from the edge.• There is usually over flattening of isodoses, near the surface.

This results in production of “horns” or hot spots.• No point parallel to the surface should receive a dose > 107%

of the central axis dose.• Because of the thinner outer rim, the average beam energy is

lower at the periphery as compared to the center.

Page 44: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Bolus

• A tissue equivalent material used to reduce the depth of the maximum dose (Dmax).

• Better called a “build-up bolus”.

• A bolus can be used in place of a compensator for kilovoltage radiation to even out the skin surface contours.

• In megavoltage radiation bolus is primarily used to bring up the buildup zone near the skin in treating superficial lesions.

Page 45: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Bolus

• The thickness of the bolus used varies according to the energy of the radiation.

• In megavoltage radiation:– Co60 : 2 - 3 mm– 6 MV : 7- 8 mm– 10 MV : 12 - 14 mm– 25 MV: 18 - 20 mm

• Properties of an ideal bolus:– Same electron density and atomic number.– Pliable to conform to surface.– Usual specific gravity is 1.02 -1.03

Page 46: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Bolus• Commonly used materials are:

– Cotton soaked with water.

– Paraffin wax.

• Other materials that have been used:

– Mix- D (wax, polyethylene, mag oxide)

– Temex rubber (rubber)

– Lincolnshire bolus (sugar and mag carbonate in form of spheres)

– Spiers Bolus (rice flour and soda bicarb)

• Commercial materials:

– Superflab: Thick and doesn't undergo elastic deformation. Made of synthetic oil gel.

– Superstuff: Add water to powder to get a pliable gelatin like material.

– Bolx Sheets: Gel enclosed in plastic sheet.

Page 47: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Breast Cone

• A beam modifying and directing device used for a tangential fields therapy.

• Advantages:– Directs beam to the central axis of the area of

interest, where a tangential beam is applied to a curved surface.

– Helps position, the patient with an accurate SSD.– Endplate provides compensation, enhances

surface dose and presses down the tissue.– Effective shielding of lungs.

Page 48: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Penumbra trimmers• Refers to the region at the edge of the beam where the dose-

rate changes rapidly as a function of distance from the beam axis.• Types:

– Transmission penumbra: Transmission through the edge of the collimator block.

– Geometrical penumbra : Finite size of the source.• Physical penumbra: Lateral distance between to specified

isodose curves at a specific depth (90% & 20% at Dmax). • Takes scattered radiation into account.• Penumbra width depends upon:

– Source diameter.– SSD.– Depth below skin.– Source to diaphragm distance (inversely)

Page 49: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Penumbra trimmers

• Consists of extensible, heavy metal bars to attenuate the beam in the penumbra region.

• Increase the source to diaphragm distance, reducing the geometric penumbra.

• Another method is to use secondary blocks placed close to the patient ( 15 – 20 cms).

3. P = AB ( SSD + d – SDD)/ SDD

d

B

1. CD/ AB = MN/ OMSSD

SDD

C D

A

E

P

O

M

N

A

2. CD/ AB = SSD + d – SDD / SDD

Page 50: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Electron beams

• Electron field shaping is done using lead cutouts.

• For a low-energy electrons (<10 MeV), sheets of lead, less than 6 mm thickness are used.

• The lead sheet can be placed directly on the skin surface.

• Shields can also be supported at the end of the treatment cone if too heavy at the cost of greater inaccuracies.

• Design is easier, because the size is same as that of the field on the patients skin.

Page 51: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Electron Beam

• To avoid variation in output and electron scatter, jaws cannot be used to collimate electron beams.

• An electron beam cone is therefore used to provide the collimation.

• A primary collimator is provided close to source – defines the maximum field size.

• A secondary collimator, near the patient defines the treatment field.

Page 52: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Direct / Internal Shielding

• Used for electron beam shielding.

• A lead shield can be placed where shielding of structures against backscatter electrons is required.

• A tissue equivalent material is coated over the lead shield like wax/ dental acrylic/ aluminum.

• Example of areas requiring these techniques are the buccal mucosa and eye lids.

leadwax

Tissue to be shielded

Tissue to be treated

Page 53: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Scattering foil

• A device to widen the thin pencil beam (3 mm) of electrons.

• Metallic plates of tin, lead or aluminiumare used.

• Disadvantages:– Beam attenuation.– Generation of bremsstrahlung

radiation.• Advantages:

– Less prone to mechanical errors.– Less expensive.– Requires less instrumentation.

• Nowadays dual foil systems are used, which compare well with scanning beam systems.

Primary foil

Secondary foil

Electron cone

Page 54: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the

Conclusion• Beam modification increases conformity allowing a higher dose

delivery to the target, while sparing more of normal tissue simultaneously.

• Megavoltage radiotherapy is better suited for most forms of beam modification due to it’s favourable scatter profile.

• However any beam modification necessitates a close scrutiny of every phase of the planning and treatment process.

The price of safety in

radiotherapy is an eternal

vigilance.

Page 55: Radiotherapy physics & Equipments - WordPress.com · 2017-04-24 · zone at which the beam is directed. –Compensation: To allow normal dose distribution data to be applied to the