rapidly monitoring released n-glycan profiles during process … · 2015. 8. 27. · 1 waters...

5
1 WATERS SOLUTIONS Rapi Fluor-MS™ Glycan Performance Test Standard (p/n 186007983) ACQUITY ® QDa ® Detector ACQUITY UPLC ® H-Class Bio System (FTN) ACQUITY UPLC Fluorescence Detector (FLR) ACQUITY UPLC Glycan BEH Amide Columns Empower ® 3 Chromatography Data Software Waters ® Fraction Manager – Analytical KEY WORDS Glycans, mass detection, H-Class, ACQUITY, QDa, Rapi Fluor-MS, IgG APPLICATION BENEFITS Rapid feedback on glycoprofiles during production to ensure product quality Reduced sample preparation times for released N-glycans Increased throughput for N-glycan analysis Specificity for N-glycan species by incorporating mass detection INTRODUCTION As glycosylated biotherapeutics move through the development pipeline, the glycoprofile and N-glycan species present are characterized. In addition, as new protein therapeutics progress through development, manufacturing conditions are carefully studied and evaluated during scale-up to ensure consistent safety and efficacy in preparation for clinical studies, and eventual commercialization. As part of this process, the critical quality attributes are often monitored closely to ensure production batches remain within defined acceptance criteria, and to identify those parameters that are critical, often as part of a quality-by-design (QbD) approach. In particular, the N-glycan profile is often monitored closely due to the importance of glycans on the safety and efficacy of protein biotherapeutics. Monitoring of released N-glycan profiles has historically been burdened with labor intensive sample preparation, which often takes several hours to days. This makes monitoring of the impact of manufacturing conditions on N-glycan profiles challenging. In addition, the analysis of released and labeled N-glycans frequently requires long analysis times. When monitoring of specific structures is desired, users often rely on optical detection for identification and quantification. In this application note we present the use of Rapi Fluor-MS, a novel reagent for rapidly labeling released N-glycans. Rapi Fluor-MS dramatically reduces overall released N-glycan sample preparation times to 30 minutes, while improving fluorescence signal by up to 14x and MS signal by up to 1000x, compared to traditional labeling techniques. In conjunction with reduced sample preparation times, we geometrically scaled a highly resolving chromatographic method to one having a total cycle time of 10 minutes. Finally, we incorporated the ACQUITY QDa Detector to monitor specific glycan species using selected ion recording (SIR), which provides a selective means of monitoring species, even if they co-elute. We will discuss how the combination of Rapi Fluor-MS and the ACQUITY QDa Detector provides a powerful solution for obtaining meaningful data rapidly and efficiently. Rapidly Monitoring Released N-Glycan Profiles During Process Development Using Rapi Fluor-MS and the ACQUITY QDa Detector Eoin F.J. Cosgrave, Robert Birdsall, and Sean M. McCarthy Waters Corporation, Milford, MA, USA

Upload: others

Post on 29-Jan-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • 1

    WAT E R S SO LU T IO NS

    RapiFluor-MS™ Glycan Performance Test

    Standard (p/n 186007983)

    ACQUITY® QDa® Detector

    ACQUITY UPLC® H-Class Bio System (FTN)

    ACQUITY UPLC Fluorescence Detector (FLR)

    ACQUITY UPLC Glycan BEH Amide Columns

    Empower®3 Chromatography Data Software

    Waters® Fraction Manager – Analytical

    K E Y W O R D S

    Glycans, mass detection, H-Class,

    ACQUITY, QDa, RapiFluor-MS, IgG

    A P P L I C AT IO N B E N E F I T S ■■ Rapid feedback on glycoprofiles during

    production to ensure product quality

    ■■ Reduced sample preparation times for

    released N-glycans

    ■■ Increased throughput for N-glycan analysis

    ■■ Specificity for N-glycan species by

    incorporating mass detection

    IN T RO DU C T IO N

    As glycosylated biotherapeutics move through the development pipeline, the

    glycoprofile and N-glycan species present are characterized. In addition, as new

    protein therapeutics progress through development, manufacturing conditions

    are carefully studied and evaluated during scale-up to ensure consistent safety

    and efficacy in preparation for clinical studies, and eventual commercialization.

    As part of this process, the critical quality attributes are often monitored closely

    to ensure production batches remain within defined acceptance criteria, and to

    identify those parameters that are critical, often as part of a quality-by-design

    (QbD) approach. In particular, the N-glycan profile is often monitored closely due

    to the importance of glycans on the safety and efficacy of protein biotherapeutics.

    Monitoring of released N-glycan profiles has historically been burdened with

    labor intensive sample preparation, which often takes several hours to days.

    This makes monitoring of the impact of manufacturing conditions on N-glycan

    profiles challenging. In addition, the analysis of released and labeled N-glycans

    frequently requires long analysis times. When monitoring of specific structures is

    desired, users often rely on optical detection for identification and quantification.

    In this application note we present the use of RapiFluor-MS, a novel reagent for

    rapidly labeling released N-glycans. RapiFluor-MS dramatically reduces overall

    released N-glycan sample preparation times to 30 minutes, while improving

    fluorescence signal by up to 14x and MS signal by up to 1000x, compared to

    traditional labeling techniques. In conjunction with reduced sample preparation

    times, we geometrically scaled a highly resolving chromatographic method to

    one having a total cycle time of 10 minutes. Finally, we incorporated the

    ACQUITY QDa Detector to monitor specific glycan species using selected ion

    recording (SIR), which provides a selective means of monitoring species, even

    if they co-elute. We will discuss how the combination of RapiFluor-MS and the

    ACQUITY QDa Detector provides a powerful solution for obtaining meaningful

    data rapidly and efficiently.

    Rapidly Monitoring Released N-Glycan Profiles During Process Development Using RapiFluor-MS and the ACQUITY QDa DetectorEoin F.J. Cosgrave, Robert Birdsall, and Sean M. McCarthyWaters Corporation, Milford, MA, USA

    http://www.waters.com/waters/partDetail.htm?partNumber=186007983http://www.waters.com/waters/partDetail.htm?partNumber=186007983www.waters.com/waters/nav.htm?cid=134761404www.waters.com/waters/nav.htm?cid=10166246www.waters.com/waters/nav.htm?cid=514222http://www.waters.com/waters/nav.htm?cid=134815665www.waters.com/waters/nav.htm?cid=513188www.waters.com/waters/nav.htm?cid=134782878

  • 2

    E X P E R IM E N TA L

    Released N-glycans were prepared from commercially

    available trastuzumab following the protocol provided within

    the RapiFluor-MS sample preparation kit. High mannose species

    used in spiking studies were isolated from RNase-B following

    release and labeling with RapiFluor-MS. Mannose species were

    chromatographically separated and collected using the Waters

    Fraction Manager – Analytical. Collected samples were dried down

    using a CentriVap and reconstituted in water. For each analysis the

    mass load was approximately 32 pmol of released and labeled

    N-glycan on column. LC-MS grade acetonitrile and water were

    purchased from Pierce. Ammonium formate was prepared using

    Ammonium Formate Solution-Glycan Analysis (p/n 186007081)

    by pouring the entire contents of the solution into 1 L of water

    and mixed. The UPLC® system used was dedicated for applications

    which do not require non-volatile salts to reduce the likelihood of

    adduct formation in the mass detector.

    LC conditionsLC system: ACQUITY UPLC H-Class Bio

    Detectors: ACQUITY UPLC FLR ACQUITY QDa

    Columns: High resolving method: ACQUITY UPLC Glycan BEH Amide, 130Å, 1.7 µm, 2.1 mm x 150 mm (p/n 186004742)

    High throughput method: ACQUITY UPLC Glycan BEH Amide, 130Å, 1.7 µm, 2.1 mm x 50 mm (p/n 186004740)

    Column temp.: 60 °C

    Sample temp.: 10 °C

    FLR settingsData rate: 5 points/sec

    Excitation wavelength: 265 nm

    Emission wavelength: 425 nm

    QDa settingsSample rate: 5 points/sec

    Mass range: 500 – 1250 Da

    Cone voltage: 15 V

    Capillary voltage: 1.5 kV

    Probe temp.: 400 °C

    Mode: Positive ion

    Mobile phase A: Acetonitrile (Pierce, LC-MS Grade)

    Mobile phase B: 50 mM ammonium formate, pH 4.4, (LC-MS Grade, ammonium formate concentrate)

    Mobile phase C: Acetonitrile (LC-MS Grade)

    Mobile phase D: Acetonitrile (LC-MS Grade)

    Gradient table high resolution method:

    Flow Time (mL/min) %A %B %C %D

    Initial 0.400 75 25 0 0

    35.0 0.400 54 46 0 0

    36.5 0.200 0 100 0 0

    39.5 0.200 0 100 0 0

    42.5 0.200 75 25 0 0

    47.4 0.400 75 25 0 0

    55.0 0.400 75 25 0 0

    Gradient table high throughput method:

    Flow Time (mL/min) %A %B %C %D

    Initial 0.800 75 25 0 0

    5.8 0.800 54 46 0 0

    6.1 0.400 0 100 0 0

    6.6 0.400 0 100 0 0

    7.1 0.400 75 25 0 0

    8.0 0.800 75 25 0 0

    10.0 0.800 75 25 0 0

    Data management

    Empower 3 Chromatography Data Software (CDS)

    Rapidly Monitoring Released N-Glycan Profiles During Process Development Using RapiFluor-MS and the ACQUITY QDa

    http://www.waters.com/waters/partDetail.htm?partNumber=186007081http://www.waters.com/waters/partDetail.htm?partNumber=186004742http://www.waters.com/waters/partDetail.htm?partNumber=186004740

  • 3

    R E SU LT S A N D D IS C U S S IO N

    EU

    0

    10

    20

    30

    40

    Minutes 2 3 4 5

    10 minute method

    EU

    0

    10

    20

    Minutes 5 10 15 20

    55 minute method

    *

    *

    Figure 1. A high resolution separation (top chromatogram) was scaled to a high throughput method by scaling the gradient geometrically while reducing column length and flow rate. While resolution is reduced, selectivity remains constant.

    Figure 2. Selected ion chromatograms for N-Glycan species separated using high throughput method. Co-eluting M5 and A2G(4)1are easily discriminated by mass detection.

    During characterization of released N-glycans, a

    highly resolving method is often used to provide

    accurate identification and quantification of the

    species present in samples. While these methods

    can be effectively scaled, there is a corresponding

    loss in resolution as overall run time decreases

    when using the same chromatograph and particle

    size column. Often, some loss in resolution will be

    tolerated if the benefit of speed is achieved, however

    critical structures must remain clearly identifiable.

    As shown in Figure 1, moving from a higher resolving

    55 min method to a 10 min high throughput method

    preserves much of the resolution between N-glycan

    species, however there is loss of resolution between

    the indicated peaks when moving to the shorter

    method. This loss of resolution complicates accurate

    monitoring by optical detection as there is no ability

    to discriminate between two species.

    Due to the improved MS response, we introduced

    the use of the ACQUITY QDa Detector as part of

    the detector stream to selectively monitor each of

    the species present in the sample. By using the SIR

    function of the ACQUITY QDa Detector we were able

    to collect independent chromatographic traces for

    each of the components to overcome the challenge

    of using optical only detection. As shown in

    Figure 2, we can clearly discriminate between

    different glycoforms by using selected ion recording.

    For each species the corresponding peak, or peaks

    for species with resolved positional isomers, can be

    clearly identified and integrated for quantification.

    Inte

    nsity

    0

    50000

    Inte

    nsity

    0

    200000

    Inte

    nsity

    0

    1x10 6

    2x10 6

    Inte

    nsity

    0

    50000

    100000

    Inte

    nsity

    0

    1x10 6

    Inte

    nsity

    0

    200000

    400000

    Inte

    nsity

    0

    20000

    Minutes

    1 2 3 4 5 6

    A2G(4)1 896.4 m/z

    F(6)A2 888.4 m/z

    F(6)A2G(4)1 969.4 m/z

    F(6)A2G(4)2 1050.5 m/z

    A2 815.3 m/z

    F(6)A2G(4)2S1 865.5 m/z

    M5 774.1 m/z

    Rapidly Monitoring Released N-Glycan Profiles During Process Development Using RapiFluor-MS and the ACQUITY QDa

  • 4

    With a high throughput method developed and the ability to

    selectively monitor particular species with mass detection, we

    investigated the linearity of response for the target N-glycan

    species Mannose 5 (M5). As described in the experimental section,

    the RapiFluor-MS labeled M5 species was isolated from the labeled

    N-glycan pool of RNAse-B. After collection, the collected material

    was dried and reconstituted in water. The reconstituted sample

    was added to a RapiFluor-MS labeled released N-Glycan sample

    from trastuzumab at various levels. We investigated the linearity

    of the response by selectively monitoring the peak area of M5 in

    relation to the volume added to the sample. As shown in Figure 3,

    the chromatographic reproducibility was quite good. In addition,

    the peak area for each volume added was highly linear (Figure 4),

    Inte

    nsity

    0

    50000

    100000

    150000

    200000

    250000

    Minutes 1.00 2.00 3.00 4.00 5.00

    SIR: M5

    Figure 3. Overlay of chromatograms over a range of M5 spike levels. Spiked amounts ranged from 1-6 µL of reconstituted M5. Absolute concentrations were not determined.

    Figure 4. Plot of amount of M5 added vs. peak area for spike M5 samples (data shown in Figure 3).

    Figure 5. SIR for co-eluting glycan structures. Left: fluorescence profiles of trastuzumab N-glycans with increasing M5 (A to F). The indicators denote the retention times for co-eluting glycans M5 and A2G1. Middle: SIR of M5 for each of the glycan samples A to F. Right: SIR for the co-eluting structure, A2G1. Use of ACQUITY QDa SIR enables the quick determination of glycan structure responsible for changing peak area in fluorescence profiles.

    R = 0.9929

    0.0E+00

    1.0E+05

    2.0E+05

    3.0E+05

    4.0E+05

    5.0E+05

    6.0E+05

    7.0E+05

    8.0E+05

    0 2 4 6 8

    Peak

    are

    a (u

    V s

    ec)

    Amount M5 added (µL)

    EU

    0

    20

    40

    60

    EU

    0

    20

    40

    EU

    0

    20

    40

    EU

    0

    20

    40

    EU

    0

    20

    40

    EU

    0

    20

    40

    Retention time (min)

    1 2 3 4 5

    Inte

    nsity

    0

    100000

    200000

    Inte

    nsity

    0

    100000

    200000

    Inte

    nsity

    0

    100000

    200000

    Inte

    nsity

    0

    100000

    200000

    Inte

    nsity

    0

    100000

    200000

    Inte

    nsity

    0

    100000

    200000

    Retention time (min)

    1 2 3 4 5

    Inte

    nsity

    0

    50000

    100000

    150000

    Inte

    nsity

    0

    50000

    100000

    150000

    Inte

    nsity

    0

    50000

    100000

    150000

    Inte

    nsity

    0

    50000

    100000

    150000

    Inte

    nsity

    0

    50000

    100000

    150000

    Inte

    nsity

    0

    50000

    100000

    150000

    Retention time (min)

    1 2 3 4 5

    FLR

    A

    B

    D

    E

    F

    C

    ACQUITY QDa M5 SIR 774.3 m/z

    A2G1 SIR 896.4 m/z

    AM5

    BM5

    DM5

    EM5

    FM5

    CM5

    AA2G1

    BA2G1

    DA2G1

    EA2G1

    FA2G1

    CA2G1

    *

    *

    *

    *

    *

    *

    strongly indicating that the mass detector provides a response

    suitable for quantification.

    After determining the linearity of response for spiked M5 species,

    we simulated a bioreactor process in which the relative amount of

    M5 was increasing. For this study we selected the A2G1 species as

    the reference for relative quantification and spiked in increasing

    amounts of M5. As shown in Figure 5, the abundance of A2G1

    (right column) remains largely constant over the course of the study

    while the M5 species (middle column) increases as expected with

    increased spiking levels. In addition, the FLR trace (left column)

    demonstrates an increase in peak area for these two species

    (labeled peak), however in the absence of mass information the

    precise cause of this increase cannot be determined.

    Rapidly Monitoring Released N-Glycan Profiles During Process Development Using RapiFluor-MS and the ACQUITY QDa

  • Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com

    Waters, The Science of What’s Possible, ACQUITY, QDa, ACQUITY UPLC, UPLC, and Empower are registered trademarks of Waters Corporation. RapiFluor-MS is a trademark of Waters Corporation. All other trademarks are the property of their respective owners.

    ©2015 Waters Corporation. Produced in the U.S.A. August 2015 720005438EN AG-PDF

    CO N C LU S IO NS

    For routine high throughput assays, RapiFluor-MS with the ACQUITY UPLC

    H-Class Bio System and the ACQUITY QDa Detector provides a novel approach for

    accurately monitoring released N-glycan species. Reduced sample preparation

    times and greatly improved MS response when coupled with the ACQUITY UPLC

    and ACQUITY QDa enable the ability to more closely monitor released N-glycan

    profiles, something which has previously not been possible. While FLR detection

    was used in this example, for high throughput methods requiring only relative

    quantification this may not be needed as each species can be monitored with the

    ACQUITY QDa Detector. As discussed here, complete sample preparation and

    analysis can be completed in 40 minutes. In addition to reproducible sample

    preparation, separation and quantification are reproducible and quantitative,

    allowing scientists to make meaningful decisions rapidly.

    Reference

    1. Lauber MA, Brousmiche DW, Hua Z, Koza SM, Guthrie E, Magnelli P, Taron CH, and Fountain KJ. Rapid Preparation of Released N-Glycans for HILIC Analysis Using a Novel Fluorescence and MS-Active Labeling Reagent. 2015, Waters Application Note P/N 720005275EN.

    http://www.waters.com/waters/library.htm?lid=134829002