redesign of ecmo circuit pressure alarm system
DESCRIPTION
University of Pittsburgh Senior Design - BioE1161. Redesign of ECMO Circuit Pressure Alarm System. Desiree Bonadonna Apryle Craig Laura Gilmour. Summary. Market review Problems Regulations Solutions Implementing the design Testing Future work. Need for Pressure Monitor Redesign. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/1.jpg)
Redesign of ECMO Circuit Pressure Alarm System
Desiree Bonadonna
Apryle Craig
Laura Gilmour
University of PittsburghSenior Design - BioE1161
![Page 2: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/2.jpg)
Summary
• Market review
• Problems
• Regulations
• Solutions
• Implementing the design
• Testing
• Future work
![Page 3: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/3.jpg)
Need for Pressure Monitor Redesign
• >120 Neonatal/Pediatric ECMO Centers in United States, increasing at a rate of ~2-5/year
• Growing number of private perfusion groups
• Sold through distributors and consultants
• Clots in the circuit are the most common mechanical complication (19%)
![Page 4: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/4.jpg)
• Expenditure may be between $80,000 and $100,000 per life saved
• Insurance reimbursement rate 60-70%
Expense Breakdown Cost/day Cost/patient
ECMO $112
Disposables - $4,000
PICU bed $3,400
Physician $520
Respiratory +
Blood Bank +
Need for Pressure Monitor Redesign
![Page 5: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/5.jpg)
• Current system with roller pump is analog• Current system with centrifugal is digital
Redesign of the ECMO pressure system includes:• Digital system for roller pump• Additional pressure monitor • Audible and visual alarm
Users: Healthcare Workers•Perfusionists•ECMO Technicians•Nurses•Physicians
![Page 6: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/6.jpg)
User Requirements
• Accurate to within +/- 15mmHg
• Refresh rate should not exceed 15 seconds
• Cost effective
• Safe and reliable
• Meets medical device regulations
![Page 7: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/7.jpg)
Imprecise/inaccurate readings
• Convert to digital display
Unknown post-heat exchanger pressure• Add a third pressure gauge
Requires constant monitoring
• Add audible and visual alarms
Cluttered
• Design the device to hang on an IV pole
Current ECMO Circuit Pressure Problems and Proposed Solutions:
![Page 8: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/8.jpg)
Regulations
Classification: • Class II, CFR 870.2100
• Cardiovascular blood flowmeter
Predicate device:• Digibio Digital Blood Pressure Monitor
Electronic Regulations:• IEC 61000-4-2
• IEC 60601-1-2
![Page 9: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/9.jpg)
Goal #6: Improve the effectiveness of clinical alarm systems.a) Implement regular preventive
maintenance and testing of alarm systems.
b) Assure that alarms are activated with appropriate settings and are sufficiently audible with respect to distances and competing noise within the unit.
JCAHO 2004 National Patient Safety Goals
![Page 10: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/10.jpg)
Project Objectives and Features: Programmable Alarm Range
“Assure that alarms are activated with appropriate settings…”
Adjustable range for anthropometric differences and
varying pathologies
![Page 11: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/11.jpg)
Project Objectives and Features: Audible Alarm
“…are sufficiently audible with respect to distances and competing noise
within the unit.”
• Immediate notification of deviation
• Alarm=120dB• Mean unit noise level = 80-89dB
• Startle response occurs at 30dB > mean noise
• Max Impulses = 1016-P/10 = 8,913
![Page 12: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/12.jpg)
Project Objectives and Features:Visual Alarm
Indicates which pressure is deviant from set range
![Page 13: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/13.jpg)
Project Objectives and Features: Digital Display
Human FactorsDecrease Human Error
Ease of Use
![Page 14: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/14.jpg)
Project Objectives and Features: Additional Pressure Indicator
Pressure drop across heat exchanger can be determined and differentiated
from patient
![Page 15: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/15.jpg)
Design Alternatives
Based on human factors
![Page 16: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/16.jpg)
Design Alternatives
![Page 17: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/17.jpg)
Engineering Technologies/Methodologies
Technologies
• SolidWorks
• PSpice
• Rapid Prototyping
• Excel
Methodologies
• Circuitry analysis
• Digital logic
![Page 18: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/18.jpg)
Schematic of One Display Unit
ADC & Display
Visual Alarm
Audible Alarm
Op-Amp
![Page 19: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/19.jpg)
2 Stages of A/D Converter Choice:•1st type produced larger error and required more circuitry•2nd type (ICL7107) included built-in display driver
![Page 20: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/20.jpg)
3 Stages of Op-Amp Design:•Scaled down 0-10V to 0-.2V and included another inverting op-amp•Scaled down 0-1V to 0-.2V •Eliminated 2nd inverting op-amp since input was determined to be negative
![Page 21: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/21.jpg)
Visual Alarm Highlights:•Programmable range•Comparators for high and low•Switches to view range while adjusting it•LED lights when out of range
![Page 22: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/22.jpg)
Audible Alarm Highlights:•Inverters needed between comparator and OR-gate •OR-gate will be 6:1 in final design
![Page 23: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/23.jpg)
Experimental DesignA/DTP-001
• Tests linearity of input voltage to output reading
PCTP-001• Tests deviation of first-generation prototype
from known pressure
VACTP-001• Tests accuracy and independence of visual
alarms
AACTP-001• Tests accuracy and independence of audible
alarm
![Page 24: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/24.jpg)
Display vs. Voltage Input
y = 4712.9x - 21.333
R2 = 0.9982
y = 4707.9x - 20.935
R2 = 0.9982
y = 4711.9x - 20.195
R2 = 0.9983
-200
0
200
400
600
800
1000
0.00 0.05 0.10 0.15 0.20 0.25
Voltage Input (V)
Dis
play
Trial 1
Trial 2
Trial 3
Linear (Trial 3)
Linear (Trial 2)
Linear (Trial 1)
A/DTP-001 Results
![Page 25: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/25.jpg)
PCTP-001 Results
PCTP-001 Results Trial 1
40
60
80
100
120
140
160
180
1 6 11
Sample
Pre
ssu
re (
mm
Hg
)
Manometer
Existing
Proposed
![Page 26: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/26.jpg)
PCTP-001 Results
PCTP-001 Results Trial 2
0
50
100
150
200
250
300
0 5 10 15
Sample
Pre
ssu
re (
mm
Hg
)
Manometer
Existing
Proposed
![Page 27: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/27.jpg)
Statistical Analysis
% error vs. known pressure
0.00
2.00
4.00
6.00
8.00
10.00
12.00
0 50 100 150 200 250 300
Pressure (mmHg)
% e
rror
Existing 1
Proposed 1
Existing 2
Proposed 2
Within the physiological range, the percent error of our device is lower than that of the currently used technology at CHP.
![Page 28: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/28.jpg)
Statistical Analysis
Statistical Analysis of Deviation from Manometer
0
2
4
6
8
10
12
Mean Maximum
Pre
ssu
re (
mm
Hg
)
Exisitng 1
Proposed 1
Existing 2
Proposed 2
![Page 29: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/29.jpg)
Competitive Analysis
• Analog Pressure Gauges
• Custom Made Digital Monitors
• COBE Cardiovascular
• SIII Pump Modules
• Console mounted control unit
• Strengths• Smaller• Not part of a kit• Less expensive• Hangs on IV pole
• Weaknesses• Does not monitor all circuit information such as temperature, etc.
![Page 30: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/30.jpg)
Team Roles
![Page 31: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/31.jpg)
Project ManagementBioE 1160 Goals BioE 1161 Accomplishments
Model circuitry in PSpice X
Build working circuit X
Test A/D Converter X
Test First Generation Prototype X
Add Visual Alarm X
Test Addition of Visual Alarm X
Implement Results of Testing
Add Audible Alarm X
Test Addition of Audible Alarm
Implement Results of Testing
Design Casing X
![Page 32: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/32.jpg)
Future Work
• Further Testing
• Visual Alarm VACTP-001
• Audible Alarm AACTP-001
• Electrical Prototyping
• Second Generation Prototype
![Page 33: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/33.jpg)
Acknowledgements
University of Pittsburgh
Department of Bioengineering
Department of Electrical Engineering
Michael Shaver, CCP
Steven Jacobs, PhD
Vikram Sundararaman
![Page 34: Redesign of ECMO Circuit Pressure Alarm System](https://reader036.vdocument.in/reader036/viewer/2022062423/56814561550346895db235b9/html5/thumbnails/34.jpg)
Questions?