reduces the risk of air embolism - · pdf fileopen-ended tubing is inadequate due to...

23
Reduces the Risk of Air Embolism 1,* In-Service Training Distributed by: Manufactured by: 1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003:17;329-35. *Compared to open-ended tubing CO 2 delivery in an in vitro model.

Upload: vonguyet

Post on 12-Mar-2018

221 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

Reduces the Risk of Air Embolism1,*

In-Service Training

Distributed by: Manufactured by:

1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003:17;329-35. *Compared to open-ended tubing CO2 delivery in an in vitro model.

Page 2: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

Intended Audience

• CV Surgeons • Perfusionists • CV Anesthesiologists • CV OR Team

Page 3: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

3

Clinical Challenge: Preventing Air Emboli

Preventing Air Emboli During Open Heart Surgery

• Air enters the heart during open-heart surgery and/or when the lung veins and the great vessels are opened1

• Air is trapped at the highest level in each cavity of the left side of the heart (Fig. 1)

• Air bubbles are ejected into the aorta during weaning from Cardio Pulmonary Bypass (CPB)1

• Arterial air emboli can occur (Fig. 2)1

Fig. 1: Air trapped at the highest level in each cavity of the left side

of the heart

Fig. 2: Air Embolism

1. Svernarud et al. Circulation 2004;109:1127-32.

Page 4: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

4

Open Heart Surgery De-Airing: Common Methods

Common methods used to prevent air embolism at the end of open heart surgical procedures:

• CO2 delivered through open-ended tubing1,2 CO2 has been used for de-airing since 1953 because of two important properties: CO2 is heavier than air and it dissolves very quickly.

• Venting of air from the ascending aorta by insertion of a needle at the highest portion1,2

• TEE (Trans Esophagus Echo) guided surgical maneuvers, such as shaking of the heart, suctioning through a needle inserted in the left ventricle1,2

• Trendelenburg maneuver (head down; Fig. 1)2

• Hyperinflation of the lungs2

Fig. 1: Trendelenburg maneuver

1. Martens et al., Ann Thorac Surg 2008;85:543–7. 2. Survey Results from CryoLife market research of US Perfusionists (n=6), Boston, MA, April 30, 2017.

Page 5: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

5

Challenge: Open-ended Tubing and Air Bubbles

Open-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet generates” which causes CO2 to exit the cavity.1

Open-end Tubing Challenges:

• Small open-ended tubing = high velocity and turbulence which causes CO2 to exit cavity1

• Risk for air embolism1

• Risk for infections3,4

The open-end tubing method is like turbulent water flowing into a bucket and most of it

splashing back out.1

1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003;17:329-35. 2. Svernarud et al. Circulation 2004;109:1127-32. Median number of microemboli after discontinuation of cardiopulmonary bypass was zero at 19 minutes with Open-ended Tubing in 3 areas of interest (left ventricle, left atrium, and proximal part of ascending aorta taken together shown by TEE). 3. Persson and van der Linden. Journal of Hospital Infection 2004;56:131-6. 4. Persson et al. Surgical Endoscopy 2005;19:91-4.

Page 6: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

6

Solution: CarbonAid and CarbonMini CO2 Diffuser

CarbonAid and CarbonMini CO2 Diffuser devices have been designed to:

• Reduce the Risk of Air Embolism1,*

• By replacing air with >99% CO2 (p<0.001)1,*

CarbonAid and CarbonMini CO2 Diffuser devices replace air with CO2 which:

• Significantly Reduces Time for Microemboli to Disappear • ~12 minutes (p<0.01)2,3

• May Prevent Infections

• By creating a >99% CO2 bacteriostatic atmosphere4,5

*Compared to open-ended tubing CO2 delivery in an in vitro model. 1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003;17:329-35. 2. Svernarud et al. Circulation 2004;109:1127-32. 3. Figure 2 in Svernarud et al., median number of microemboli after discontinuation of cardiopulmonary bypass was 0 with CarbonAid at 8 minutes vs. 20 minutes with Open-ended Tubing in 3 areas of interest (left ventricle, left atrium, and proximal part of ascending aorta taken together shown by TEE). 4. Persson and van der Linden. Journal of Hospital Infection 2004;56:131-6. 5. Persson et al. Surgical Endoscopy 2005;19:91-4.

Open-ended tubing

<1% Air

CarbonAid and CarbonMini

18% - 96% Air

18% - 96% Air

<1% Air

Page 7: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

7

CarbonAid CO2 Diffuser vs. Open-ended Tubing

“Conventional open-ended tubes provided a poor and varying de-airing of the wound cavity model (18%-96% remaining air) because of CO2 jets with calculated velocities between 1.3 and 34 meter/second […]

The [CarbonAid] gas diffuser provided an almost complete de-airing of the model (0.2% remaining air) at flows of 5 to 10 L/min. This was a result of a uniform distribution of CO2 with calculated velocities of about 0.1 meter/second.”1

“Distribution of mean air contents in % (n=10) in (A) the vertical plane in the direction of the CO2 insufflation and (B) in the horizontal plane at half the depth of the model, during CO2 insufflation with a 2.5-mm tube, a 1/4-in (6.35 mm) tube and a gas diffuser at a CO2

flow of 7.5 L/min. The dashed line marks the direction of the CO2 insufflation. The air content in positions No. 6, 7, and 8 in the horizontal plane was not actually measured but was because of symmetry assumed to be equal to positions No. 2, 3, and 4, respectively.”

1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003;17:329-35.

Page 8: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

8

CarbonAid CO2 Diffuser vs. Open-ended Tubing (continued)

Conclusion: “These data imply that de-airing of a cardiothoracic wound by CO2 insufflation depends on flow and outflow velocity. To compensate for diffusion with ambient air, the CO2 flow should be > 5 L/min, and the outflow velocity should be about 0.1 m/s or less to avoid turbulence in the wound. This is only attainable with a [CarbonAid CO2] gas diffuser.”1

The mean air content in % (n=80) at half the depth of the model plotted versus the calculated outflow velocity (m/s) of the 2.5 mm tube, the 1/4-in tube, and the gas diffuser at CO2 flows of 2.5, 5, 7.5, and 10 L/min. The 4 values of the gas diffuser are located close to the origin of coordinates because of low air contents caused by low outflow velocities. The curve represents an exponential regression of the data.1

1. Persson and van der Linden. Journal of Cardiothoracic and Vascular Anesthesia 2003;17:329-35.

Page 9: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

9

Indication for Use: CarbonAid and CarbonMini CO2 Diffusers

To be used during open heart surgery procedures for insufflation of CO2 into the thoracic cavity to reduce the risk of air embolism.1,2 Including the following:

• Valve surgery • Atrial/ ventricular septum defect closure • Left ventricular reconstruction • Aortic surgery • Single clamp coronary surgery • Insertion of ECMO through sternotomy • Insertion of LVAD • Heart transplantation

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 10: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

10

CarbonAid and CarbonMini CO2 Diffuser Sizes

CarbonAid CO2 Diffuser For larger wound openings >8 cm in diameter1

CarbonMini CO2 Diffuser For smaller wound openings <3–8 cm in diameter2

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 11: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

11

In-Service Training

This in-service training is intended for general instruction of CarbonAid and CarbonMini CO2 Diffusers. Variations in use may occur in specific procedures due to individual techniques and patient conditions.

Page 12: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

12

CO2 Connection1,2

Step 1:

• CO2 Source: Determine in advance how CO2 gas is delivered in the OR. It may be from a central source or a tank.

• Secure a CO2 gas source with the capacity to deliver CO2 flow via a CO2 flow meter during the estimated time for procedure. Approximately 1 kg CO2 /hour is consumed when the flow is set for 10 liter/minute.

• If a tank is being used, confirm that a backup tank is available for lengthy or concomitant cases.

• Connect tubing to CO2 source. The gas source must include a pressure regulator and a CO2 flow meter.

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 13: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

13

Flow Meter and Pressure Regulator1,2

Step 2:

• The CO2 flow meter shall contain or be connected to a pressure regulator. • The service pressure for the CO2 gas is set between 3.5-4.5 bar

(50-65 psi or 350-450 kpa). • The set gas flow should continue until the closure of the heart, the lung veins, and the

great vessels is completed.

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 14: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

14

Device Selection and Flow Rate

Step 3: • Select the appropriate product and recommended flow rate

Diameter of surgical wound opening

CO2 flow Recommended

Recommended Product

<3 cm 3 L/Min CarbonMini1

3-4 cm 5 L/Min CarbonMini1

5-8 cm 8 L/Min CarbonMini1

>8 cm 10 L/Min CarbonAid2

1. Instructions for Use CarbonMini CO2 Diffuser. 2. Instructions for Use CarbonAid CO2 Diffuser.

Page 15: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

15

Positioning1,2

Step 4:

Open Sternotomy CarbonAid Positioning: Place the foam tip at the caudal end of the wound. Bend the thin tubing so the foam tip is placed approximately 2 inches (5 cm) below the skin surface.

Minimally Invasive CarbonMini Positioning: Place the foam tip at the caudal end of the wound. Bend the thin tubing so the foam tip is placed approximately 1-2 inches (2-4 cm) below the skin surface.

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 16: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

16

Positioning (continued)

Step 4 (continued): CarbonMini positioning during minimally invasive cases is surgeon preference. Field experience has reported placement options may be: • Placement in MI cavity or • Placement in a specially created port just

superior of the incision site.

Page 17: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

17

Immobilize the Tubing1,2

Step 5: Immobilize the tubing per surgeon preferred method near the caudal end of the wound. Attachment methods may include: • Sutures or • Adhesive sterile tape

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 18: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

18

Foam Tip Positioning1,2

Step 6: • The foam tip should be positioned a few

centimeters below skin surface. • Avoid submerging the foam tip in fluid.* • If the foam tip gets soaked with fluid its

full function can quickly be restored by squeezing the tip with two fingers to remove the trapped fluid.

*If the foam tip is completely covered in fluid, CarbonAid and CarbonMini may not be able to deliver CO2 gas properly. 1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 19: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

19

Device Selection and Flow Rate1,2

Step 7: • Connect tubing to CO2 Source. The gas source must include a pressure regulator and

a CO2 flow meter. • Start the recommended CO2 gas flow before opening of the heart, the lung veins

and/or the great vessels to ensure saturated CO2 atmosphere as follows: – CarbonAid: CO2 gas flow should be started at least one (1) minute before the heart, the

lung veins and/ or the great vessels are opened.1 – CarbonMini: CO2 gas flow should be started at least two (2) minutes before the heart, the

lung veins and/ or the great vessels are opened.2

1. Instructions for Use CarbonAid CO2 Diffuser. 2. Instructions for Use CarbonMini CO2 Diffuser.

Page 20: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

20

Intermittent Suctioning1

Step 8: Use intermittent suction within the cavity by having the suctioning device outside of the cavity when not in use. If a suction device is used in the cavity, avoid suctioning when no fluid is present and try to minimize the suction rate to avoid depletion of CO2. This is to be coordinated with the perfusionist.

1. Svenarud, Persson and van der Linden. J Thorac Cardiovasc Surg 2003;25:1043-9.

Page 21: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

21

Bubbles on Echo1

Step 9: Inform the CV anesthesiologist that at the end of surgery, CO2 bubbles often show up on echo and is often confused with air bubbles. The difference is: CO2 bubbles will dissipate almost immediately, unlike oxygen bubbles which take more time to be dissolved, sometimes 15 to 20 minutes.1

1. Cardia Innovation

Page 22: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

22

Device Removal

Step 10: At completion of a procedure, gently remove the CarbonAid or CarbonMini CO2 Diffuser.

Page 23: Reduces the Risk of Air Embolism - · PDF fileOpen-ended tubing is inadequate due to “high outflow velocities” and “the air turbulence the jet ... outflow velocities . The curve

On-X Life Technologies, Inc. 1300 East Anderson Lane, Bldg. B Austin, Texas 78752 USA Phone: (512) 339-8000 Fax: (512) 339-3636 Email: [email protected]

CryoLife, Inc. 1655 Roberts Boulevard, NW, Kennesaw, Georgia 30144 USA Phone: 888-427-9654 Fax: 770-590-3753 Email: [email protected]

Distributed by: Manufactured by:

CarbonAid and CarbonMini are distributed by CryoLife, Inc. The CarbonAid and CarbonMini products are manufactured by Cardia Innovation AB. CryoLife, the snowflake design, Life Restoring Technologies are Registered trademarks owned by CryoLife, Inc. CarbonAid and CarbonMini are registered trademarks of Cardia Innovation AB. All other trademarks are owned by their respective owners. © 2017 CryoLife Inc. All rights reserved. MLENG1119.000 (08/2017)