reliability and validity of four instruments for …...reliability and validity of four instruments...

8
Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN, and MICHAEL P. FALL We studied the between-therapist reliability and the validity of four instruments in measuring lumbar spine curvature and pelvic tilt. The four instruments and their measurements were 1) a tape measure to measure the change in lumbar curvature during trunk flexion; 2) a gravity goniometer to measure pelvic angle and lumbar curvature during stance, trunk flexion, and trunk extension; 3) a parallelogram goniometer to measure lumbar curvature during stance, trunk flexion, and trunk extension; and 4) a standard goniometer to measure the angle between wooden pointers mounted perpendicularly to the spine to obtain pelvic angle and lumbar curvature during stance, trunk flexion, and trunk extension. We found no single instrument to be the most reliable or valid. Between-therapist reliability ranged from .64 to .93 (Pearson product-moment correlation) and from .60 to .92 (interclass correlation coefficient). The validities of the instruments compared with measurements from roentgenograms generally were low, ranging from -.13 to .76 (Pearson product-moment correlation) and -.73 to -.05 (inter- class correlation coefficient). Key Words: Lumbosacral region, Pelvis, Physical therapy. Objective methods of externally measuring the motion and curvature of the lumbar spine or the pelvic tilt are important in assessing spine and pelvic functions. Many methods of measuring the curvature or range of motion of the spine and the pelvic tilt have been in- vestigated. These methods can be clas- sified as standard goniometry, tape measurement, and contour measure- ment. Standard goniometry was shown by Fitzgerald et al to be reliable in meas- uring thoracolumbar extension and lat- eral flexion. 1 With this method, how- ever, restricting the measurement to the lumbar region is difficult. Pelvic tilt must be controlled closely or it will be included in the measurement. This method also is not useful in measuring pelvic tilt. Hart et al 2 and Reynolds 3 used a mod- ified goniometer called a spondylometer to measure total thoracolumbar flexion and extension. This method was relia- ble, and pelvic tilt did not affect the measurements because the sacrum sur- face was used as the reference for the measurement. This method, however, has many of the same limitations as standard goniometry: Lumbar motion is not measured separately from total spine motion, and neither pelvic tilt nor lumbar curvature during stance can be measured. Gravity goniometers have been used to measure spinal flexion, extension, and lateral flexion. 3-5 They measure the angle between the vertical plane and the tangent to the spine at the point of meas- urement. Regional motions can be measured by calculating the difference between angular motions at the two end points of the region. The gravity goni- ometer was shown by Reynolds to be reliable in measuring lumbar spine flex- ion, trunk extension, and lateral flex- ion. 3 Several methods of using a tape meas- ure to measure spine motion have been investigated. Macrae and Wright 6 and Fitzgerald et al 1 measured lumbar flex- ion using a skin distraction method based on that developed by Shöber, 7 and they found a high reliability with this method. Reynolds, however, found a low reliability when using the same method. 3 Moll et al developed a skin distraction method of measuring lateral flexion 8 and a method of measuring trunk extension using a plumb bob and a tape measure. 9 They found these methods to produce reliable measure- ments. Reynolds found this method of extension measurement to have good reliability, but not the lateral flexion measurement. 3 Frost et al measured total range of motion in trunk anterior flexion and lateral flexion as the distance from the fingers to the floor, and they measured trunk rotation and extension as the change in distance between bony land- marks. 10 The measurements of anterior flexion and lateral flexion had good re- liabilities, but the measurements of trunk rotation and extension did not. These methods cannot be used to meas- ure regional spinal or pelvic motion and, because of differences in arm length and leg length among individuals, they may not be appropriate methods for making comparisons among subjects. The meth- ods are reliable and easy to use, how- ever, for showing changes in total motion for an individual over time. Gajdosik et al calculated the angle of pelvic tilt by measuring the distance from the floor to the anterior-superior Dr. Burdett is Assistant Professor, Program in Physical Therapy, School of Health Related Profes- sions, University of Pittsburgh, Pittsburgh, PA 15261 (USA). Ms. Brown is Physical Therapy Supervisor, De- partment of Physical Therapy, Harmarville Reha- bilitation Center, Pittsburgh, PA 15238. Mr. Fall is a physical therapist with West Hart- ford Physical Therapy and Sports Medicine Asso- ciates, West Hartford, CT 06091. He was Staff Physical Therapist, Department of Physical Ther- apy, Harmarville Rehabilitation Center, when this study was conducted. This article was submitted January 24,1985; was with the authors for revision 14 weeks; and was accepted October 3, 1985. Volume 66 / Number 5, May 1986 677

Upload: others

Post on 03-Apr-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions

RAY G. BURDETT, KATHRYN E. BROWN, and MICHAEL P. FALL

We studied the between-therapist reliability and the validity of four instruments in measuring lumbar spine curvature and pelvic tilt. The four instruments and their measurements were 1) a tape measure to measure the change in lumbar curvature during trunk flexion; 2) a gravity goniometer to measure pelvic angle and lumbar curvature during stance, trunk flexion, and trunk extension; 3) a parallelogram goniometer to measure lumbar curvature during stance, trunk flexion, and trunk extension; and 4) a standard goniometer to measure the angle between wooden pointers mounted perpendicularly to the spine to obtain pelvic angle and lumbar curvature during stance, trunk flexion, and trunk extension. We found no single instrument to be the most reliable or valid. Between-therapist reliability ranged from .64 to .93 (Pearson product-moment correlation) and from .60 to .92 (interclass correlation coefficient). The validities of the instruments compared with measurements from roentgenograms generally were low, ranging from -.13 to .76 (Pearson product-moment correlation) and -.73 to -.05 (inter­class correlation coefficient).

Key Words: Lumbosacral region, Pelvis, Physical therapy.

Objective methods of externally measuring the motion and curvature of the lumbar spine or the pelvic tilt are important in assessing spine and pelvic functions. Many methods of measuring the curvature or range of motion of the spine and the pelvic tilt have been in­vestigated. These methods can be clas­sified as standard goniometry, tape measurement, and contour measure­ment.

Standard goniometry was shown by Fitzgerald et al to be reliable in meas­uring thoracolumbar extension and lat­eral flexion.1 With this method, how­ever, restricting the measurement to the lumbar region is difficult. Pelvic tilt must be controlled closely or it will be included in the measurement. This method also is not useful in measuring pelvic tilt.

Hart et al2 and Reynolds3 used a mod­ified goniometer called a spondylometer to measure total thoracolumbar flexion and extension. This method was relia­ble, and pelvic tilt did not affect the measurements because the sacrum sur­face was used as the reference for the measurement. This method, however, has many of the same limitations as standard goniometry: Lumbar motion is not measured separately from total spine motion, and neither pelvic tilt nor lumbar curvature during stance can be measured.

Gravity goniometers have been used to measure spinal flexion, extension, and lateral flexion.3-5 They measure the angle between the vertical plane and the tangent to the spine at the point of meas­urement. Regional motions can be measured by calculating the difference between angular motions at the two end points of the region. The gravity goni­ometer was shown by Reynolds to be reliable in measuring lumbar spine flex­ion, trunk extension, and lateral flex­ion.3

Several methods of using a tape meas­ure to measure spine motion have been investigated. Macrae and Wright6 and Fitzgerald et al1 measured lumbar flex­ion using a skin distraction method based on that developed by Shöber,7 and they found a high reliability with this

method. Reynolds, however, found a low reliability when using the same method.3 Moll et al developed a skin distraction method of measuring lateral flexion8 and a method of measuring trunk extension using a plumb bob and a tape measure.9 They found these methods to produce reliable measure­ments. Reynolds found this method of extension measurement to have good reliability, but not the lateral flexion measurement.3

Frost et al measured total range of motion in trunk anterior flexion and lateral flexion as the distance from the fingers to the floor, and they measured trunk rotation and extension as the change in distance between bony land­marks.10 The measurements of anterior flexion and lateral flexion had good re­liabilities, but the measurements of trunk rotation and extension did not. These methods cannot be used to meas­ure regional spinal or pelvic motion and, because of differences in arm length and leg length among individuals, they may not be appropriate methods for making comparisons among subjects. The meth­ods are reliable and easy to use, how­ever, for showing changes in total motion for an individual over time. Gajdosik et al calculated the angle of pelvic tilt by measuring the distance from the floor to the anterior-superior

Dr. Burdett is Assistant Professor, Program in Physical Therapy, School of Health Related Profes­sions, University of Pittsburgh, Pittsburgh, PA 15261 (USA).

Ms. Brown is Physical Therapy Supervisor, De­partment of Physical Therapy, Harmarville Reha­bilitation Center, Pittsburgh, PA 15238.

Mr. Fall is a physical therapist with West Hart­ford Physical Therapy and Sports Medicine Asso­ciates, West Hartford, CT 06091. He was Staff Physical Therapist, Department of Physical Ther­apy, Harmarville Rehabilitation Center, when this study was conducted.

This article was submitted January 24,1985; was with the authors for revision 14 weeks; and was accepted October 3, 1985.

Volume 66 / Number 5, May 1986 677

Page 2: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

Fig. 1. Modified gravity goniometer being used to measure the angle of the sacrum with respect to the horizontal plane.

Fig. 2. Parallelogram goniometer being used to measure the angle between the tangents to the spine at the sacrum and the thoracolumbar junction.

iliac spine (ASIS) and to the posterior-superior iliac spine (PSIS) and the dis­tance between the ASIS and PSIS.11

They found this method to be reliable for measuring pelvic position and ROM.

Various devices for measuring the contour of the spine have been devel­oped, including a simple, flexible rule that adapts to the contour of the back,4

a device for tracing the contour of the back,12 and an electronic device for re­cording the coordinates of points along the back.13 Other investigators have used pointers mounted perpendicularly to small bases that are taped to the back to indicate perpendicular planes to the spine.1415 Measurements of the angle

between the pointers are made from photographs. This method was com­pared to roentgenographic measure­ments and was reported to have a high validity for measurement of lumbar cur­vature during stance14 and trunk ante­rior flexion ROM.15

On the basis of this review of the literature, two basic techniques seem to have the potential of 1) isolating the position or motion of the lumbar spine or pelvis in the sagittal plane, 2) being valid and reliable, and 3) being clinically easy to use. One technique is the skin distraction method, which has been lim­ited mainly to measurement of spine flexion. The other technique is the

measurement of the curvature of a re­gion by measuring the angle between pointers mounted perpendicularly to the curve at two points.

The purpose of our study was to com­pare the reliability and validity of four instruments for measuring pelvic tilt during stance and lumbar curvature during stance, trunk flexion, and trunk extension. These four instruments were 1) a tape measure, 2) a gravity-operated goniometer modified for use on the back, 3) a parallelogram goniometer, and 4) a standard goniometer to meas­ure the angle between wooden pointers mounted perpendicularly to the back.

METHOD

Subjects Twenty-seven volunteers participated

as subjects for determining the validities of these measurement instruments, and 23 of these volunteers also participated as subjects for determining the reliabili­ties of the instruments. Informed con­sent was obtained from each of the sub­jects. All subjects were between the ages of 20 and 40 years, had no history of chronic back pain or disability, and were not grossly overweight. A skin-fold measurement taken over one PSIS of each subject had a mean value of 1.42 cm (±0.73 cm).

Instrumentation A tape measure marked in centime­

ters was used to measure the change in lumbar curvature between the positions of upright stance and trunk flexion. A modification of the method introduced by Shober7 was used. The skin was marked along the midline of the spine 10 cm above and 3 cm below the PSISs. The distance between these markers after trunk flexion was measured using the tape measure, and the change in distance between the marks was used to indicate the amount of lumbar flexion. This method cannot be used to measure pelvic tilt because the reference point for the measurement is on the pelvis. It also cannot be used to measure lumbar curvature for a particular position, but it can be used to measure the change in curvature between two positions. There­fore, we used a tape measure to measure only the change in lumbar curvature between the positions of upright relaxed stance and trunk flexion.

678 PHYSICAL THERAPY

Page 3: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

RESEARCH

Figure 1 shows the gravity goniom­eter* that was modified for use on the back by attaching a 2-cm x 3.5-cm rec­tangular platform on the base of the goniometer. This platform was placed on the back to form a tangent to the back at a particular point. The angle of the tangent with respect to the vertical plane then was read from the goniom­eter. Angles were measured at the tho­racolumbar (T-L) junction and on the sacrum. Pelvic tilt was defined as the angle between the horizontal plane and the perpendicular to the sacrum. This goniometer can be used to measure pel­vic tilt and lumbar curvature during any trunk position because the vertical grav­ity line is the reference for the measure­ment. This measure of pelvic tilt, how­ever, will not account for movement between the pelvis and the sacrum. In our study, this instrument was used to measure lumbar curvature and pelvic tilt during stance and lumbar curvature during trunk flexion and extension.

The parallelogram goniometer† that we used is shown in Figure 2. It was constructed so that the linkage between its two platforms resulted in a direct reading of the angular difference be­tween the surfaces of the two platforms. The angular difference between the cur­vature at the T-L junction and sacrum was measured with this goniometer. It cannot be used to measure pelvic tilt because the pelvis is the reference for the measurement.

Figure 3 shows the two 1.5-cm × 2.5-cm rectangular plastic platforms with 10-cm wooden pointers that we used. These platforms were attached with tape so that one was centered over the T-L junction and one was centered between the PSISs and the skin mark 3 cm below the PSISs. The angle between the point­ers was measured in two ways using a standard plastic goniometer. First, the angle was measured directly by aligning the pointers with the lines on the goni­ometer arms running perpendicularly to the axes of the goniometer arms. Pelvic tilt could not be measured directly this way because the pelvis was the reference for this measurement. Therefore, only lumbar curvature during relaxed upright stance, trunk flexion, and trunk exten­sion were measured with this method.

Fig. 3. Standard goniometer being used to measure the angle between wooden pointers mounted perpendicularly to the spine at the sacrum and thoracolumbar junction.

Fig. 4. Internal and external angles of pelvic tilt and lumbar curvature. IL = internal lumbar angle, EL = external lumbar angle, IP = internal pelvic angle, and EP = external pelvic angle.

Then, a photograph of the pointers on the body was made from the side, and the angle between the pointers was measured from the photograph by align­ing the long axes of the goniometer with the pointers. A vertical reference line placed in the field of view of the camera was used as the reference for measuring pelvic tilt. The camera was placed on a horizontal surface and was positioned visually by the therapist so that its film plane was parallel to the sagittal plane of the subject. Pelvic tilt was determined as the angle between the horizontal plane and a pointer mounted perpendic­ularly to the sacrum. Therefore, any mo­tion that occurred between the pelvis and the sacrum was included in this measurement. The photographs were used to measure pelvic tilt during stance and lumbar curvature during stance, trunk flexion, and trunk extension.

Reliability Determination

Two physical therapists who were un­familiar with these measurement meth­ods practiced until they became com­fortable using these instruments. Each therapist completed the following pro­cedures for each subject. With the sub­ject in the prone position, the low back area was exposed. The PSISs were pal­pated and marked. A line was drawn between them, and marks were placed along the midline of the spine 10 cm above and 3 cm below this line. The T-L junction also was palpated and marked. The subject then stood in a normal, relaxed posture with the heels about shoulder width apart. The angles of the sacrum and the T-L junction with respect to the vertical plane were meas­ured with the gravity goniometer. The parallelogram goniometer was used to measure the lumbar curvature between the T-L junction and the sacrum. The platforms with pointers were taped to the subject's back at the T-L junction and over the sacrum. The standard go­niometer was used to measure the angle between the pointers, and a photograph of the pointers was taken from the side.

The subject then sat on an adjustable-height seat and was positioned with his thighs parallel to the floor, knees in 90 degrees of flexion, and ankles in a neu­tral position. The subject flexed his trunk and hips from this position until his shoulders contacted his thighs or to his maximum amount of flexion. The subject then was instructed to place his forearms under his thighs and grasp his opposite elbows with his hands to be stabilized in this position. This was done to achieve a fixed position that could be reassumed for subsequent measure­ments. The angle between the pointers was measured with the standard goni­ometer, a photograph of the pointers was taken, and the platforms then were removed. The tape measure was used to measure the distance between the 10-cm and 3-cm marks, the gravity goni­ometer was used to measure the angles at the T-L junction and the sacrum, and the parallelogram goniometer was used to measure the angle between the T-L junction and sacrum.

The subject then was placed in the prone position on a table with a hinged top so that his ASISs were directly above the hinge. His thighs were strapped to the tabletop for stabilization, and the portion of the table under his upper

* Bowen and Co, Inc, Rockville, MD 20852. † Designed and constructed by Gary Debaucher,

Emory University Rehabilitation Center, Atlanta, GA 30322.

Volume 66 / Number 5, May 1986 679

Page 4: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

body was raised to cause hyperexten-sion. The position of the table was marked for each subject so that the same position could be achieved for subse­quent measurements. No attempt was made to keep the table position constant

among subjects or to take each subject to the end of his range. The angles of the T-L junction and sacrum were meas­ured using the gravity goniometer, and the angle between the T-L junction and sacrum was measured using the paral-

TABLE 1 Pearson Product-Moment Correlation Coefficients and Interclass Correlation Coefficients for Between-Therapist Reliabilitiesa

Measurement

Lumbar curvature during trunk flex­ion

Lumbar curvature during trunk ex­tension

Lumbar curvature during stance

Pelvic tilt during stance

Gravity Goni­

ometer

r

.93

.72

.93

.84

ICC

.91

.71

.92

.82

Parallel­ogram Goni­

ometer

r ICC

.93 .92

.64 .60

.90 .89

NM

Instrument

Platforms

Goni­ometer

r

.85

.77

.88

ICC

.85

.75

.88

NM

Photos

r

.87

.81

.88

.89

ICC

.87

.78

.88

.87

Tape Measure

r ICC

.71 .72

NMb

NM

NM

TABLE 2 Means and Standard Deviations of the Measurements Made by Each Therapist with Each Instrument

Measurement and Instrument

Lumbar curvature during trunk flex­iona

Gravity goniometer5

Parallelogram goniometer Goniometric measurement of plat­

forms Photographs of platforms Tape measure (cm)

Lumbar curvature during trunk ex­tensiona

Gravity goniometer Parallelogram goniometer Goniometric measurement of plat­

forms Photographs of platformsb

Lumbar curvature during stancea

Gravity goniometer Parallelogram goniometer Goniometric measurement of plat­

forms Photographs of platforms

Pelvic tilt during stancea

Gravity goniometer Goniometric measurement of plat­

forms6

Therapist 1

10.4 11.0

11.0 11.6 20.1

26.4 26.3

26.3 27.2

13.9 12.4

14.6 16.5

14.6

21.7

s

6.1 6.4

6.5 6.7 1.2

6.3 6.3

13.8 13.2

8.1 6.7

10.7 10.5

6.3

8.0

Therapist 2

11.3 11.0

11.8 12.1 20.1

25.6 24.4

24.2 23.7

14.5 12.9

14.3 16.3

13.7

20.2

s

5.3 5.4

6.1 6.9 1.1

9.2 8.0

10.8 12.2

9.4 7.8

9.5 9.8

7.8

7.5

lelogram goniometer. The platforms with the pointers then were placed on the T-L junction and sacrum, the angle between the pointers was measured with the standard goniometer, and a photo­graph was made of the pointers. The first therapist completed this procedure and erased all of the skin markings. Then, the second therapist immediately performed the same procedure on the same subject.

Validity Determination One of the therapists who assisted in

the reliability determination also partic­ipated in determining the validity of these instruments. Each subject was marked as previously described. The subject then was positioned in prepara­tion for a sagittal plane roentgenogram of the low back in one of the three measurement positions: upright relaxed stance, trunk flexion, or trunk exten­sion. Using all the instruments, the ther­apist measured the lumbar curve and sacral angle of the subject in this posi­tion. Then, while the subject was still in this position, a roentgenogram was taken of the low back region, Ten sub­jects were measured in upright relaxed stance, 6 in trunk flexion, and 11 in trunk extension. Only one position was measured for each subject to limit the exposure of each subject to radiation.

A line was drawn on the roentgeno­gram along the inferior surface of the 12th thoracic vertebral body (T12) and along the superior surface of the 1st sacral vertebral body (S1) (Fig. 4). The angle of pelvic tilt was defined as the angle between the horizontal plane and the line along the superior surface of S1. The base of the roentgenogram was used as the horizontal reference because the film was positioned so that its base was horizontal during the roentgenogram procedure. The lumbar curvature was defined as the angle between the lines along the inferior surface of T12 and the superior surface of S1. These two angles taken from the roentgenogram were used as the reference angles for deter­mining the validity of the external meas­urements.

The validity of the tape measure method of measuring change in lumbar flexion was not determined. Two roent­genograms of subjects—one in normal, relaxed stance and one in trunk flex­ion—would be needed to calculate the change in lumbar curvature internally

a All correlation coefficients are significantly different from zero at the p < .05 level. b NM = not measured.

a Except where indicated, all measurements are in degrees. b Significantly different at p < .05.

680 PHYSICAL THERAPY

Page 5: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

RESEARCH

to determine the validity of this method. We took only one roentgenogram for each subject, however, to minimize ex­posure to radiation.

Data Analysis

To determine the between-therapist reliabilities of the external measure­ments, Pearson product-moment corre­lation coefficients (r) and interclass correlation coefficients (ICCs) were calculated between the measurements made by the therapists. Because Pearson product-moment correlation shows only the parallelism of the two sets of meas­urements, an analysis of variance (AN-OVA) was performed to compare the means of the sets' measurements. Valid­ity of the external measurements also was determined by calculating Pearson product-moment correlation coeffi­cients and ICCs and by performing an ANOVA between external measure­ments and measurements from the roentgenograms. A probability level of .05 was selected for significance. A cor­relation coefficient of .80 or above was chosen as an indication of high between-therapist reliability and high validity.

RESULTS

The correlation coefficients for be­tween-therapist reliabilities for each of the instruments and measurements are shown in Table 1. These coefficients ranged from a low of .60 (ICC) and .64 (r) for measuring lumbar curvature dur­ing trunk extension with the parallelo­gram goniometer to a high of .92 (ICC) and .93 (r) for measuring lumbar cur­vature during trunk flexion using the parallelogram goniometer and for meas­uring lumbar curvature during stance using the gravity goniometer. Each of the measurements had a reliability cor­relation coefficient significantly differ­ent from zero, but not all measurements reached the value of .80 that was chosen as a clinically significant coefficient.

The least reliable of the measure­ments was lumbar curvature during trunk extension. Only the method using photographs of the pointers approached a .80 correlation coefficient. The tape measure method of measuring change in lumbar curvature between upright stance and trunk flexion also was unre­liable using this criterion. We saw no clear trend for the gravity goniometer, the parallelogram goniometer, or the platforms and standard goniometer to

TABLE 3 Results of Two-Way Analysis of Variance for the Measurement of Lumbar Curvature During Trunk Flexion

Instrument and Source Gravity goniometer

Therapists Subjects Therapist x subject

Parallelogram goniometer Therapists Subjects Therapist x subject

Platforms—goniometric meas­urement

Therapists Subjects Therapist x subject

Platforms—measurement from photographs

Therapists Subjects Therapist x subject

Tape measure Therapists Subjects Therapist x subject

df

1 22 22

1 22 22

1 22 22

1 22 22

1 22 22

SS

7.3 1375.4

60.7

0.0 1462.9

63.0

7.0 1636.0 132.0

3.1 1903.2 134.9

0.0 53.5 9.1

MS

7.8 62.5 2.8

0.0 66.5 2.9

7.0 74.4

6.0

3.1 86.5

6.1

0.0 2.4 0.4

F

2.85a

0.00

1.17

0.51

0.01

TABLE 4 Results of Two-Way Analysis of Variance for the Measurement of Lumbar Curvature During Trunk Extension

Instrument and Source Gravity goniometer

Therapists Subjects Therapist x subject

Parallelogram goniometer Therapists Subjects Therapist x subject

Platforms—goniometric measurement

Therapists Subjects Therapist x subject

Platforms—measurement from photographs

Therapists Subjects Therapist x subject

df

1 22 22

1 22 22

1 22 22

1 22 22

SS

7.8 2667.5 459.6

40.2 1851.2 437.3

52.2 5889.9

884.3

135.7 6441.9 679.8

MS

7.8 121.2 20.9

40.2 84.2 19.9

52.2 267.7

38.4

135.7 292.8 30.9

F

0.38

2.02

1.36

4.39a

exhibit a higher overall reliability. Each of these instruments was highly reliable for measuring lumbar curvature during trunk flexion and upright relaxed stance. In addition, the gravity goni­ometer and photographs of the plat­forms also were reliable for measuring pelvic tilt during stance.

Table 2 shows the means and stand­ard deviations of the measurements

made by the two therapists. Tables 3 through 6 show the results of two-way ANOVAs comparing the means of these measurements. Small, but significant, differences existed between the follow­ing measurements made by the two ther­apists: lumbar curvature during trunk flexion using the gravity goniometer, and pelvic tilt during stance and lumbar curvature during trunk extension using

a Significant at p < . 05.

a Significant at p < .05.

Volume 66 / Number 5, May 1986 681

Page 6: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

TABLE 5 Results of Two-Way Analysis of Variance for the Measurement of Lumbar Curvature During Relaxed, Upright Stance

Instrument and Source Gravity goniometer

Therapists Subjects Therapist x subject

Parallelogram goniometer Therapists Subjects Therapist x subject

Platforms—goniometric meas­urement

Therapists Subjects Therapist x subject

Platforms—measurement from photographs

Therapists Subjects Therapist x subject

df

1 22 22

1 22 22

1 22 22

1 22 22

SS

4.3 3263.6

142.7

2.6 2190.2

131.9

1.4 4238.3 277.6

0.5 4260.7

284.0

MS

4.3 148.3

6.5

2.6 99,6

6.0

1.4 192.7 12.6

0.5 193.7 12.9

F

0.66

0.44

0.11

0.04

TABLE 6 Results of Two-Way Analysis of Variance for the Measurement of Pelvic Tilt During Relaxed, Upright Stance

Instrument and Source Gravity goniometer

Therapists Subjects Therapist x subject

Platforms—measurement from photographs

Therapists Subjects Therapist x subject

df

1 22 22

1 22 22

SS

10.5 2036.2

198.5

25.1 2506.6 284.0

MS

10.5 92.5 9.0

25.1 113.9 12.9

F

1.17

3.59a

TABLE 7 Pearson Product-Moment Correlation Coefficients and Interclass Correlation Coefficients for Validities of External Measurements

Measurement

Lumbar curvature during trunk flex­ion

Lumbar curvature during trunk ex­tension

Lumbar curvature during stance

Pelvic tilt during stance

Gravity Goniometer

r

.73

.15

-.12

.10

ICC

-.11

-.73

-.62

-.45

Instrument

Parallelogram Goniometer

r

.46

.24

.03

ICC

-.19

-.71

-.59

NMb

Platforms

Goniometer

r

.70

.51

-.13

ICC

-.09

-.63

-.55

NM

Photos

r

.76a

.60

-.09

.08

ICC

-.05

-.62

-.49

-.20

the photographs of the platforms and pointers.

The correlation coefficients represent­ing the validities of the instruments and the measurements made by each instru­ment are shown in Table 7. The only correlation that was significantly differ­ent from zero was for the measurement of lumbar curvature during trunk flex­ion using the photographs of the plat­form pointers. This correlation coeffi­cient also approached the value of .80 chosen for clinical significance.

Table 8 shows the means and stand­ard deviations of the external measure­ments made with each instrument for each position and the internal measure­ments made from the roentgenograms. Results of the two-way ANOVA com­paring the means of these external and internal measurements are shown in Ta­ble 9. The measurements made exter­nally by the therapist with the gravity goniometer, parallelogram goniometer, and the regular goniometer directly or from photographs did not differ signifi­cantly for any of the measurement positions. The measurements made internally from the roentgenograms, however, differed significantly from all those made externally for all measure­ment positions.

DISCUSSION

The reliabilities of the methods of measurement found in our study are similar to those found in other studies of the measurement of trunk motion. Both our study and that of Reynolds3

found the gravity goniometer to be reli­able in measuring the curvature or po­sition of the spine. The reliabilities of the methods of measuring pelvic tilt in our study are similar to those found by Gajdosik et al11 in their study of pelvic angle measurement. Their method re­quires three measurements and a cal­culation; our method gives the angle of tilt directly. Their method, however, uses points of reference directly on the pelvis, and our method uses the sacrum as the reference surface, which may re­sult in invalid measurements because any motion between the sacrum and the pelvis would be included. We found the measurement of lumbar flexion using the tape measure to be unreliable using a criterion of 80 as a highly reliable correlation coefficient. This finding is similar to the findings of Reynolds3 for this method. Other investigators, how-

a Significant at p < .05.

a Significantly different from zero at the p < .05 level. b NM = not measured.

682 PHYSICAL THERAPY

Page 7: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

RESEARCH

ever, have found this method of meas­uring the change in lumbar curvature during trunk flexion to be highly relia­ble.1,6 Our finding of a relatively low reliability in this measurement may be because of the homogeneity of our sub­jects. The standard deviation was very small when this instrument was used to measure lumbar flexion compared with the standard deviations of the measure­ments obtained using the other instru­ments (Tab. 2). This difference would make a high correlation coefficient more difficult to obtain.

We did not find one of the methods studied to emerge as the most reliable when comparing the measurements each could make in common, but we did find that each method has its advan­tages and disadvantages. A common ad­vantage of these methods compared with many other methods, such as standard goniometry or reaching, is their ability to measure regional motion or position rather than combined mo­tions.

Measuring lumbar flexion using a tape measure is a simple method. A tape measure, however, cannot be used for measuring standing posture because standing is the reference position. Also, using a standard starting distance could result in different numbers of spinal seg­ments contributing to the motion for different individuals. Using a standard distance, however, may produce more reliable results than using the T-L junc­tion as a reference position because ther­apists may differ in their palpation skills. The other instruments are at least as reliable as the tape measure and can measure other motions or positions as well.

The gravity goniometer and the pho­tographs taken of the platform pointers are the most versatile of the methods we studied because they can measure both pelvic and lumbar spine positions. Re­liabilities of measurements made by the gravity goniometer depend on the even application of pressure so that the base of the goniometer is tangent to the spine. Uneven pressure, especially on a patient with a large skin fold over the area of measurement, could tilt the base of the goniometer with respect to the spine. An increase in the length of the base may minimize this problem.

The parallelogram goniometer cannot measure pelvic angle because a horizon­tal reference is needed, but it can meas­ure the lumbar curvature for any body

TABLE 8 Means and Standard Deviations of the Measurements Made Externally by One Therapist and Made Internally from Roentgenograms (in Degrees)

Measurement

Lumbar curvature during trunk flex­ion

Lumbar curvature during trunk ex­tension

Lumbar curvature during stance

Pelvic tilt during stance

Roentgen­ogram

-4 .1 a

65.6a

48.8a

29.8a

s

15.1

11.6

17.7

12.2

Gravity Goni­

ometer

16.0

34.5

16.9

10.2

s

6.5

4.9

5.7

6.3

Instrument

Parallel­ogram Goni­

ometer

15.0

33.1

16.2

s

6.3

5.9

5.8

NMb

Platforms

Goni­ometer

16.0

32.2

19.8

s

7.2

7.2

9.3

NM

Photos

15.5

32.4

21.3

15.1

s

7.1

6.9

11.3

11.2

TABLE 9 Results of Two-Way Analysis of Variance Comparing the Measurements Made Externallv and Internally Durina the Validity Determination

Measurement and Source Lumbar curvature during trunk

flexion Instrument Subjects Instrument x subject

Lumbar curvature during trunk extension

Instrument Subjects Instrument x subject

Lumbar curvature during stance

Instrument Subjects Instrument x subject

Pelvic tilt during stance Instrument Subjects Instrument x subject

df

4 5

20

4 10 40

4 9

36

2 9

18

SS

1888.3 1460.3 590.1

9395.3 1694.0 1230.3

7494.2 2040.0 3297.8

2080.9 1277.0 1549.1

MS

471.1 292.1

29.5

2348.8 169.4 30.8

1873.6 226.7

91.6

1040.5 141.9 86.1

F

16.0a

9.9

76.4a

5.5

20.5a

2.5

12.1a

1.7

position. Its reliability, similar to that of the gravity goniometer, is dependent on the even application of pressure to its bases. The parallelogram goniometer could be converted to an electrogoniom-eter by the addition of a potentiometer. As such, it could be used to measure continuously lumbar spine position or motion during an activity and also as a biofeedback device. The parallelogram goniometer used in our study can be constructed from readily available parts and tools.

The platforms and pointers can be applied easily to the back using double-sided tape. The main problem with the reliability of this method is skin motion, especially during trunk extension. Pa­tients with large skin folds under the platforms may cause the platform to tilt relative to the underlying spine during motions that tend to compress the skin. This method, however; is very versatile, especially when combined with photog­raphy. This method has been used with high-speed photography and videotap-

a Significantly different from the other means at p < .05 level. b NM = not measured.

a Significant at p < .05.

Volume 66 / Number 5, May 1986 683

Page 8: Reliability and Validity of Four Instruments for …...Reliability and Validity of Four Instruments for Measuring Lumbar Spine and Pelvic Positions RAY G. BURDETT, KATHRYN E. BROWN,

ing to measure spine and pelvic motion during activities.15,16

The validities of the lumbar curvature and pelvic tilt angles were very low, partly because of the high variability among the subjects between the tangent to the posterior surface of the sacrum and the angle of the surface of the S1. This high variability among the subjects resulted in a low validity for both the pelvic tilt measurements and the lumbar curvature measurements because the pelvic angle was used as a reference to measure this curvature. Use of another reference for pelvic angle from the roentgenograms, such as a line between the ASIS and PSIS, may increase the validity of these external measurements.

CONCLUSIONS

This study has compared the reliabil­ities and validities of four clinical meth­ods for measuring lumbar curvature and pelvic angle. Each of the methods has advantages and disadvantages that can affect the reliability and validity of the measurements or the ease of use of the instruments. No one instrument was found to be superior to the others. Lum­

bar curvature during stance or trunk flexion and pelvic tilt during stance were measured more reliably than was lum­bar curvature during trunk extension. Using the superior surface of S1 and the inferior surface of T12 as the references for the validity of the external measure­ments, we found that none of the instru­ments were able to measure any of the positions with a high validity. Some re­finement of the instruments and proce­dures used in this study may increase their reliabilities. Other methods may exist that are more reliable and valid for measuring regional spinal curvature or pelvic angle, and studies of these meth­ods are needed.

REFERENCES

1. Fitzgerald GK, Wynveen KJ, Rheault W, et al: Objective assessment with establishment of normal values for lumbar spinal range of mo­tion. Phys Ther 63:1776-1781,1983

2. Hart FD, Strickland D, Cliffe P: Measurement of spinal mobility. Ann Rheum Dis 33:136-139, 1974

3. Reynolds PMG: Measurement of spinal mobil­ity: A comparison of three methods. Rheumatol Rehabil 14:180-185,1975

4. Anderson JAD, Sweetman BJ: A combined flexi-rule/hydrogoniometer for measurement of lumbar spine and its sagittal movement. Rheu­matol Rehabil 14:173-179,1975

5. Loebl WY: Measurement of spinal posture and range of spinal movement. Annals of Physical Medicine 9:103-110,1967

6. Macrae IF, Wright V: Measurement of back movement. Ann Rheum Dis 28:584-587,1969

7. Shöber P: The lumbar vertebral column and backache. Meunchener Medizinische Woch-enschrift 84:336,1937

8. Moll JMH, Liyanage SP, Wright V: An objective clinical method of measuring lateral spinal flex­ion. Rheumatology and Physical Medicine 2:225-239,1972

9. Moll JMH, Liyanage SP, Wright V: An objective clinical method to measure spinal extension. Rheumatology and Physical Medicine 2:293-312,1972

10. Frost M, Stuckey S, Smalley LA, et al: Relia­bility of measuring trunk motions in centime­ters. Phys Ther 62:1431-1437,1982

11. Gajdosik R, Simpson R, Smith R, et al: Pelvic tilt: Intratester reliability of measuring the standing position and range of motion. Phys Ther 65:169-174,1985

12. Willner S: Spinal pantograph: A non-invasive technique for describing kyphosis and lordosis in the thoracolumbar spine. Acta Orthop Scand 52:525-529,1981

13. Gross C, Neuwirth M, Graham J, et al: Three-dimensional lordosimeter: A device for the-non-radkxjraphic assessment of spinal configura­tion. Bull Hosp Jt Dis Orthop Inst 42:151-171, 1982

14. Flint MM: Lumbar posture: A study of roent­genographic measurements and the influence of flexibility and strength. Research Quarterly 34:15-20,1963

15. Troup JDG, Hood CA, Chapman AE: Measure­ments of the sagittal mobility of the lumbar spine and hips. Annals of Physical Medicine 9:308-321,1968

16. Murray MP, Drought AB, Kory RC: Walking patterns of normal men. J Bone Joint Surg [Am] 46:335-360,1964

684 PHYSICAL THERAPY