reproductive laboratory - university of … · web viewreference: keel ba, webster bw. crc handbook...

79
Revised: 10/06/2005 REPRODUCTIVE LABORATORY This College of American Pathologists (CAP) Laboratory Accreditation Program (LAP) Checklist is provided as a Microsoft® Word 2000 electronic file for convenience and for educational purposes. It represents the fully-approved version for use in the LAP as of the date given in the header. Newer approved versions of this Checklist may be found via the Internet at the CAP Web site (http://www.cap.org/apps/docs/laboratory_accreditation/checklists/ checklistftp.html ) for both viewing and download to your computer. If you are currently enrolled in the CAP LAP and are preparing for an inspection, please note: The Checklists undergo frequent revision, and the contents may have changed after you receive your inspection packet. If a Checklist has been updated since receiving your packet, you will be inspected based upon the Checklists that were mailed to you in your application or reapplication packet. For questions about the use of Checklists in the inspection process, please e-mail the CAP at [email protected], or call (800) 323-4040, ext. 6065. Suggestions for content improvement should be sent by e-mail to LAP at [email protected]. All checklists are © 2005 College of American Pathologists. All rights reserved.

Upload: trinhkhuong

Post on 04-Apr-2018

216 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

Revised: 10/06/2005

REPRODUCTIVE LABORATORY

This College of American Pathologists (CAP) Laboratory Accreditation Program (LAP) Checklist is provided as a Microsoft® Word 2000 electronic file for convenience and for educational purposes. It represents the fully-approved version for use in the LAP as of the date given in the header.

Newer approved versions of this Checklist may be found via the Internet at the CAP Web site (http://www.cap.org/apps/docs/laboratory_accreditation/checklists/checklistftp.html) for both viewing and download to your computer.

If you are currently enrolled in the CAP LAP and are preparing for an inspection, please note:

The Checklists undergo frequent revision, and the contents may have changed after you receive your inspection packet. If a Checklist has been updated since receiving your packet, you will be inspected based upon the Checklists that were mailed to you in your application or reapplication packet.

For questions about the use of Checklists in the inspection process, please e-mail the CAP at [email protected], or call (800) 323-4040, ext. 6065. Suggestions for content improvement should be sent by e-mail to LAP at [email protected].

All checklists are © 2005 College of American Pathologists. All rights reserved.

Page 2: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

OUTLINE

SUMMARY OF CHANGESINSPECTION TECHNIQUES – KEY POINTS

ANDROLOGY AND EMBRYOLOGYLABORATORY SAFETYPROFICIENCY TESTINGQUALITY CONTROL AND QUALITY MANAGEMENT

SUPERVISIONPROCEDURE MANUALREQUISITIONS, SPECIMEN RECEIPT, AND RESULTS REPORTINGGENERAL QUALITY CONTROLREAGENTS AND CULTURE MEDIA

ANDROLOGY PROCEDURES AND TESTSCOMPUTER-ASSISTED SEMEN ANALYSIS (CASA) AUTOMATED INSTRUMENTS

Calibration and Quality ControlMANUAL SEMEN ANALYSIS

Sperm ConcentrationSperm MotilitySemen Stained Smear - Sperm DifferentialBiochemical Tests

ANTI-SPERM ANTIBODY (ASA) TESTSSPERM PROCESSING FOR THERAPEUTIC INSEMINATION

ANDROLOGY PERSONNELEMBRYOLOGY

OOCYTE AND EMBRYO HANDLINGCULTURE OF SPERM, OOCYTES, AND EMBRYOSMICROMANIPULATION OF EMBRYOSEMBRYO TRANSFER PROCEDURES

RECORDSINSTRUMENTS AND EQUIPMENTEMBRYOLOGY PERSONNEL

EMBRYOLOGY LABORATORY DIRECTORCRYOPRESERVATION OF SPERM, OOCYTES, AND EMBRYOS

PHYSICAL FACILITIES

REPRODUCTIVE LABORATORY Page 2 of 59

Page 3: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

SUMMARY OF CHANGESREPRODUCTIVE LABORATORY Checklist

10/6/2005 Edition

The following questions have been added, revised, or deleted in this edition of the checklist, or in the two editions immediately previous to this one.

If this checklist was created for a reapplication, on-site inspection or self-evaluation it has been customized based on the laboratory's activity menu. The listing below is comprehensive; therefore some of the questions included may not appear in the customized checklist. Such questions are not applicable to the testing performed by the laboratory.

Note: For revised checklist questions, a comparison of the previous and current text may be found on the CAP website. Click on Laboratory Accreditation, Checklists, and then click the column marked Changes for the particular checklist of interest.

NEW Checklist Questions

Question Effective Date None.

REVISED Checklist Questions

Question Effective Date RLM.00100 12/29/2004RLM.00200 12/29/2004RLM.11400 12/29/2004

DELETED Checklist Questions

Question Effective Date RLM.13500 10/06/2005RLM.14100 12/29/2004

REPRODUCTIVE LABORATORY Page 3 of 59

Page 4: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The checklists used in connection with the inspection of laboratories by the Commission on Laboratory Accreditation (“CLA”) of the College of American Pathologists have been created by the College and are copyrighted works of the College. The College has authorized copying and use of the checklists by College inspectors in conducting laboratory inspections for the CLA and by laboratories that are preparing for such inspections. Except as permitted by section 107 of the Copyright Act, 17 U.S.C. sec. 107, any other use of the checklists constitutes infringement of the College’s copyrights in the checklists. The College will take appropriate legal action to protect these copyrights.

****************************************************************

INSPECTION TECHNIQUES – KEY POINTS

****************************************************************

I. READ – OBSERVE – ASK – the three methods of eliciting information during the inspection process. These three methods may be used throughout the day in no particular order. Plan the inspection in a way that allows adequate time for all three components.

READ = Review of Records and DocumentsDocument review verifies that procedures and manuals are complete, current, available to staff, accurate and reviewed, and describe good laboratory practice. Make notes of any questions you may have, or processes you would like to observe as you read the documentation.

OBSERVE – ASK = Direct Observation and Asking QuestionsObserving and asking questions accomplish the following:

1. Verifies that the actual practice matches the written policy or procedure2. Ensures that the laboratory processes are appropriate for the testing performed3. Ensures that outcomes for any problem areas, such as PT failures and issues/problems

identified through the quality management process, have been adequately investigated and resolved

4. Ensures that previously cited deficiencies have been corrected

Use the following techniques: Observe laboratory practices – look at what the laboratory is actually doing. Compare the

written policy/procedure to what you actually observe in the laboratory to ensure the written policy/procedure accurately reflects laboratory practice. Note if practice deviates from the documented policies/procedures.

Ask open ended, probing questions – these are starting points that will allow you to obtain large amounts of information, and help you clarify your understanding of the documentation you’ve seen and observations you’ve made. This eliminates the need to ask every single checklist question, as the dialogue between you and the laboratory may address multiple checklist questions.

REPRODUCTIVE LABORATORY Page 4 of 59

Page 5: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

Ask open-ended questions that start with phrases such as “show me how…” or “tell me about …” or “what would you do if…”. By asking questions that are open-ended, or by posing a hypothetical problem, you will avoid “cookbook” answers. For example, ask “Could you show me the specimen transport policy and show me how you ensure optimum specimen quality?” This will help you to determine how well the technical staff is trained, whether or not they are adhering to the lab’s procedures and policies, and give you a feel for the general level of performance of the laboratory.

Ask follow-up questions for clarification. Generally, it is best not to ask the checklist questions verbatim. For example, instead of asking the checklist question “Is there documentation of corrective action when control results exceed defined tolerance limits?” ask, “What would you do if the SD or CV doubles one month?” A follow-up probing question could be, “What would you do if you were unable to find a cause for the change in SD or CV?”

II. Evaluate Selected Specimens and Tests in Detail

For the Laboratory General Checklist: Follow a specimen through the laboratory. By following a specimen from collection to test result, you can cover multiple checklist questions in the Laboratory General checklist: questions on the specimen collection manual; phlebotomy; verbal orders; identification of patients and specimens; accessioning; and result reporting, including appropriate reference ranges, retention of test records, maintaining confidentiality of patient data, and proper handling of critical values and revisions to reports.

For the individual laboratory sections: Consult the laboratory’s activity menu and focus on tests that potentially have the greatest impact on patient care. Examples of such tests include HIV antibodies, hepatitis B surface antigen, urine drugs of abuse, quantitative beta-hCG, cultures of blood or CSF, acid-fast cultures, prothrombin time and INR reporting, and compatibility testing and unexpected antibody detection. Other potentially high-impact tests may be identified by looking at very high or low volume tests in the particular laboratory, or problems identified by reviewing the Variant Proficiency Testing Performance Report.

To evaluate preanalytic and postanalytic issues: Choose a representative specimen and “follow" the specimen through the laboratory or section of the laboratory, reviewing appropriate records in the preanalytic and postanalytic categories.

To evaluate analytic processes: Choose 2 or 3 analytes and perform a comprehensive review of records, including procedure manuals, quality control and proficiency testing records, instrument maintenance records and method performance validations for the last 2 years, selecting timeframes at the beginning, mid-point, and end of this timeframe. Compare instrument print-outs to patient reports and proficiency testing results to ensure accurate data entry. If problems are identified, choose additional tests or months to review.

III. Verify that proficiency testing problem have been resolved: From the inspector’s packet, review the Variant PT Performance Report that identifies, by analyte, all of the PT scores below 100%. Correlate any PT problems to QC or maintenance records from the same time period. Be thorough

REPRODUCTIVE LABORATORY Page 5 of 59

Page 6: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

when reviewing these representative records, selecting data from the beginning, middle and end of the period since the last on-site inspection.

IV. Review correction of previous deficiencies: Review the list of deficiencies from the previous on-site inspection provided in the inspector’s packet. Ensure that they have been appropriately addressed.

#################################################################

ANDROLOGY AND EMBRYOLOGY

#################################################################

*****************************************************************

LABORATORY SAFETY

*****************************************************************

The inspector should review relevant questions from the Safety section of the Laboratory General checklist, to assure that the reproductive laboratory is in compliance. Please elaborate upon the location and the details of each deficiency in the Inspector's Summation Report.

*****************************************************************

PROFICIENCY TESTING

*****************************************************************

**REVISED** 12/29/2004

RLM.00100 Phase II N/A YES NO

Is the laboratory enrolled in the appropriate available required CAP Surveys or a CAP-approved alternative proficiency testing (PT) program for the patient testing performed?

NOTE: The list of analytes for which CAP requires proficiency testing is available on the CAP website [http://www.cap.org/apps/docs/laboratory_accreditation/ptgraded.html] or by phoning 800-323-4040 (or 847-832-7000), option 1. The laboratory’s participation in proficiency testing must include all analytes on this list for which it performs patient testing. Participation in proficiency testing may be through CAP Surveys or a CAP-approved proficiency testing provider. Laboratories will not be penalized if they are unable to enroll in an oversubscribed program. If unable to enroll, however, the laboratory must implement an alternative assessment procedure for the affected analytes. For regulated analytes, if the CAP and CAP-approved alternative PT programs are oversubscribed, CMS requires the laboratory to attempt to enroll in another CMS-approved PT program.

REPRODUCTIVE LABORATORY Page 6 of 59

Page 7: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

COMMENTARY:

N/A

REFERENCES: 1) Department of Health and Human Services, Health Care Financing Administration. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7146 [42CFR493.801]; 2) Tholen DW. Reference values and participant means as targets in proficiency testing. Arch Pathol Lab Med. 1993;117:885-889; 3) Westgard JO, et al. Laboratory precision performance. State of the art versus operating specifications that assure the analytical quality required by clinical laboratory improvement amendments proficiency testing. Arch Pathol Lab Med. 1996;120:621-625; 4) NCCLS. Continuous quality improvement: essential management approaches and their use in proficiency testing; proposed guideline GP22-P. Wayne, PA: NCCLS, 1997; 5) College of American Pathologists, Commission on Laboratory Accreditation. Standards for laboratory accreditation; standard III. Northfield, IL: CAP, 1998.

**REVISED** 12/29/2004

RLM.00200 Phase II N/A YES NO

For tests for which CAP does not require PT, does the laboratory at least semiannually 1) participate in external PT, or 2) exercise an alternative performance assessment system for determining the reliability of analytic testing?

NOTE: Appropriate alternative performance assessment procedures may include: split sample analysis with reference or other laboratories, split samples with an established in-house method, assayed material, regional pools, clinical validation by chart review, or other suitable and documented means. It is the responsibility of the laboratory director to define such alternative performance assessment procedures, as applicable, in accordance with good clinical and scientific laboratory practice. Participation in ungraded/educational proficiency testing programs also satisfies this checklist question.

COMMENTARY:

N/A

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7184 [42CFR493.236(c)(1)]; 2) Jorgensen n, et al. Semen analysis performed by different laboratory teams: an intervariation study. Int J Androl. 1997;20:201-218; 3) Shahangian S, et al. A system to monitor a portion of the total testing process in medical clinics and laboratories. Feasibility of a split-specimen design. Arch Pathol Lab Med. 1998;122:503-511.

REPRODUCTIVE LABORATORY Page 7 of 59

Page 8: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.00300 Phase II N/A YES NO

Does the laboratory integrate all proficiency testing samples within the routine workload, and are those samples analyzed by personnel who routinely test patient samples, using the same primary method systems as for patient samples?

NOTE: Replicate analysis of proficiency testing samples is acceptable only if patient specimens are routinely analyzed in the same manner. If the laboratory uses multiple methods for an analyte, proficiency samples should be analyzed by the primary method. There must not be any interlaboratory communication on proficiency testing data before results are reported. The educational purposes of proficiency testing are best served by a rotation that allows all technologists to be involved in the proficiency testing program. Records of these studies must be kept and can be an important part of the competency and continuing education documentation in the personnel files of the individuals. When external proficiency testing materials are not available, the semi-annual alternative performance assessment process should also be integrated within the routine workload.

COMMENTARY:

N/A

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992 (Feb 28): 7146 [42 CFR 493.801(b)].

RLM.00400 Phase II N/A YES NO

Is there evidence of evaluation and, if indicated, prompt corrective action in response to "unacceptable" results on the proficiency testing report and results of the alternative performance assessment system?

NOTE: The evaluation must document the specific reason(s) for the "unacceptable" result(s) and actions taken to reduce the likelihood of recurrence. This must be done within one month after the laboratory receives its proficiency testing evaluation. In addition, each ungraded challenge, each educational challenge, and each episode of nonparticipation must be reviewed and corrective action instituted as appropriate.

COMMENTARY:

N/A

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7173 [42CFR493.1407(e)(4)(iv)]; 2) NCCLS. Using proficiency testing (PT) to improve the clinical laboratory; approved guideline GP27-A. Wayne, PA: NCCLS, 1998.

REPRODUCTIVE LABORATORY Page 8 of 59

Page 9: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.00500 Phase II N/A YES NO

Is there documented evidence of ongoing evaluation by the laboratory director or designee of the proficiency testing and alternative performance assessment results?

COMMENTARY:

N/A

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7173 [42CFR493.1407(e)(4)(iii)]; 2) NCCLS. Using proficiency testing (PT) to improve the clinical laboratory; approved guideline GP27-A. Wayne, PA: NCCLS, 1998.

*****************************************************************

QUALITY CONTROL AND QUALITY MANAGEMENT

*****************************************************************

A detailed quality control program specific for the techniques in use is essential for semen analysis, special sperm tests, the identification of oocytes, and the maintenance of the developmental capacity of these oocytes during the procedures of fertilization, embryo culture, and embryo transfer.

The quality control program should be clearly defined and documented, including general policies and delegation of responsibilities.

-----------------------------------------------------------------

SUPERVISION

-----------------------------------------------------------------

Before patient results are reported, QC data must be judged acceptable. The laboratory director or designee must review QC data at least monthly. Beyond these specific requirements, a laboratory may (optionally) perform more frequent review at intervals that it determines appropriate. Because of the many variables across laboratories, the CAP makes no specific recommendations on the frequency of any additional review of QC data.

All quality management (QM) questions in the Laboratory General Checklist pertain to the reproductive laboratory.

RLM.00600 Phase II N/A YES NO

Does the reproductive laboratory have a written QC/QM program?

REPRODUCTIVE LABORATORY Page 9 of 59

Page 10: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

NOTE: The QM/QC program must be clearly defined and documented. The program must ensure quality throughout the preanalytic, analytic, and post-analytic (reporting) phases of testing, including patient identification and preparation; specimen collection, identification, preservation, transportation, and processing; and accurate, timely result reporting. The program must be capable of detecting problems in the laboratory’s systems, and identifying opportunities for system improvement. The laboratory must be able to develop plans of corrective/preventive action based on data from its QM system.

Appropriate laboratory personnel must judge QC data acceptable before patient results are reported. The laboratory director or designee must review QC data at least monthly. Beyond these specific requirements, a laboratory may (optionally) perform more frequent review at intervals that it determines appropriate. Because of the many variables across laboratories, the CAP makes no specific recommendations on the frequency of any additional review of QC data.

COMMENTARY:

N/A

REFERENCES: 1) Berwick DM. Continuous improvement as an ideal in health care. New Engl J Med. 1989;320:53-56; 2) Byrd W. Quality assurance in the reproductive biology laboratory. Arch Pathol Lab Med. 1992;116:418-422; 3) Michelmann HW. Quality management in the andrology laboratory. Int J Androl. 1997;20(suppl 3):50-54; 4) Souter VL, et al. Laboratory techniques for semen analysis: a Scottish survey. Health Bull (Edinb).1997;55:140-149.

RLM.00700 Phase II N/A YES NO

Is there a documented procedure describing methods for patient identification, patient preparation, specimen collection and labeling, specimen preservation, and conditions for transportation, and storage before testing, consistent with good laboratory practice?

COMMENTARY:

The laboratory must have a completely documented procedure describing methods for patient identification, patient preparation, specimen collection and labeling, specimen preservation, conditions for transportation, and storage before testing. Such protocols must be consistent with good laboratory practice.

REFERENCE: Byrd W. Quality assurance in the reproductive biology laboratory. Arch Pathol Lab Med. 1992;116:418-422.

REPRODUCTIVE LABORATORY Page 10 of 59

Page 11: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.00800 Phase II N/A YES NO

Is there evidence of ongoing evaluation of instrument maintenance and function, temperature, etc.?

COMMENTARY:

N/A

REFERENCE: Byrd W. Quality assurance in the reproductive biology laboratory. Arch Pathol Lab Med. 1992;116:418-422.

RLM.00900 Phase II N/A YES NO

Is there a documented system in operation to detect and correct significant clerical and analytical errors, and unusual laboratory results?

COMMENTARY:

The laboratory must have a documented system in operation to detect and correct significant clerical and analytical errors, and unusual laboratory results. One common method is review of results by a qualified person (technologist, supervisor, pathologist) before release from the laboratory, but there is no requirement for supervisory review of all reported data. The selective use of delta checks also may be useful in detecting clerical errors in consecutive samples from the same patient. In computerized laboratories, there should be automatic "traps" for improbable results.

REFERENCES: 1) Laessig RH, et al. Quality control and quality assurance. Clinics Lab Med. 1986;6:317-327; 2) Bachner P. Quality assurance: an accreditation perspective. Lab Med. 1989;20:159-162; 3) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988. Fed Register. 2003 (Jan 24):905 [42CFR493.1230,42CFR493.1239(a)(b), 42CFR493.1240, 42CFR493.1290, 42CFR493.1299(a)(b)] ; 4) Dufour D, et al. The clinical significance of delta checks. Am J Clin Pathol. 1998;110:531.

RLM.01000 Phase II N/A YES NO

Does the system provide for the timely correction of errors?

COMMENTARY:

The system for detecting clerical errors, significant analytical errors, and unusual laboratory results must provide for timely correction of errors, i.e., before results become available for clinical decision making. For suspected errors detected by the end user after reporting, corrections must be promptly made if such errors are confirmed by the laboratory.

REPRODUCTIVE LABORATORY Page 11 of 59

Page 12: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

REFERENCES: 1) Kilshaw D. Quality assurance. I. Philosophy and basic principles. Med Lab Sci. 1986;43:377-381; 2) Byrd W. Quality assurance in the reproductive biology laboratory. Arch Pathol Lab Med. 1992;116:418-422.

RLM.01100 Phase II N/A YES NO

In the absence of on-site supervisors, are the results of tests performed by personnel reviewed by the laboratory director, general supervisor or technical person in charge of the reproductive laboratory on the next routine working shift?

NOTE: The CAP does NOT require supervisory review of all test results before or after reporting to patient records. Rather, this question is intended to address only that situation defined under CLIA-88 for "high complexity testing" performed by trained high school graduates qualified under 42CFR493.1489(b)(5) when a qualified general supervisor is not present.

COMMENTARY:

N/A

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7182 [42CFR493.1463(a)(3) and 42CFR493.1463(c)]: 7183 [42CFR493.1489(b)(1) and 42CFR493.1489(b)(5)].

RLM.01200 Phase II N/A YES NO

Is there evidence of active review by the director or designee of the following?

1. Quality control of routine procedures2. Documentation of the type of reagents that are used and their source3. Instrument function checks, including temperature checks, adequate gas flow, and

concentration in the incubators

NOTE: Before patient results are reported, QC data must be judged acceptable. The laboratory director or designee must review QC data at least monthly. Beyond these specific requirements, a laboratory may (optionally) perform more frequent review at intervals that it determines appropriate.

COMMENTARY:

N/A

REFERENCES: 1) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, A, 2, 10; 2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule.

REPRODUCTIVE LABORATORY Page 12 of 59

Page 13: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

Fed Register. 2003(Jan 24):7166 [42CFR493.1254]; 3) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press. 1992. ----------------------------------------------------------------

PROCEDURE MANUAL

-----------------------------------------------------------------

The procedure manual should be used by personnel at the workbench and should include: test principle, clinical significance, specimen type, required reagents, test calibration, quality control, procedural steps, calculations, reference intervals, and interpretation of results. The manual should address relevant pre-analytic and post-analytic considerations, as well as the analytic activities of the laboratory. The specific style and format of procedure manuals are at the discretion of the laboratory director.

The inspection team should review the procedure manual in detail to understand the laboratory's standard operating procedures, ensure that all significant information and instructions are included, and that actual practice matches the contents of the procedure manuals.

RLM.01300 Phase II N/A YES NO

Is a complete procedure manual available at the workbench or in the work area?

NOTE 1: The use of inserts provided by manufacturers is not acceptable in place of a procedure manual. However, such inserts may be used as part of a procedure description, if the insert accurately and precisely describes the procedure as performed in the laboratory. Any variation from this printed or electronic procedure must be detailed in the procedure manual. In all cases, appropriate reviews must occur.

NOTE 2: A manufacturer's procedure manual for an instrument/ reagent system may be acceptable as a component of the overall departmental procedures. Any modification to or deviation from the procedure manual must be clearly documented.

NOTE 3: Card files or similar systems that summarize key information are acceptable for use as quick reference at the workbench provided that:

a. A complete manual is available for reference.b. The card file or similar system corresponds to the complete manual and is subject to

document control.

NOTE 4: Electronic (computerized) manuals are fully acceptable. There is no requirement for paper copies to be available for the routine operation of the laboratory, so long as the electronic versions are readily available to all personnel. Such electronic versions must be subjected to proper document control (i.e., only authorized persons may make changes, changes are dated/signed (manual or electronic), and there is documentation of periodic

REPRODUCTIVE LABORATORY Page 13 of 59

Page 14: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

review). Current paper copies of electronically stored procedures should be available at the time of the CAP inspection, or rapidly generated at the request of the inspector.

COMMENTARY:

A documented procedure manual must be developed for the reproductive laboratory and be available at the workbench. Its elements should include: test principle, clinical significance, specimen type(s), required reagents, calibration, quality control, procedural steps, calculations, reference intervals, and interpretation, as applicable.

NOTE 1: The use of inserts provided by a manufacturer is not acceptable in place of a procedure manual. However, such inserts may be used as part of a procedure description, if the insert accurately and precisely describes the procedure as performed in the laboratory. Any variation from this procedure must be detailed in the procedure manual. In all cases, appropriate reviews must occur.

NOTE 2: A manufacturer's procedure manual for an instrument/reagent system may be acceptable as a component of the overall departmental procedures. Any modification to or deviation from the procedure manual must be clearly documented.

NOTE 3: Card files or similar systems that summarize key information are acceptable for use as quick reference at the workbench provided that:

a. A complete manual is available for reference,b. The card file or similar system corresponds to the complete manual and is subject to

document control.

NOTE 4: Electronic (computerized) manuals are fully acceptable. There is no requirement for paper copies to be available for the routine operation of the laboratory, so long as the electronic versions are readily available to all personnel. Such electronic versions must be subjected to proper document control (i.e., only authorized persons may make changes, changes are dated/signed (manual or electronic), and there is documentation of periodic review). Current paper copies of electronically stored procedures should be available at the time of the CAP inspection, or rapidly generated at the request of the Inspector.

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1251]; 2) NCCLS. Clinical laboratory technical procedure manuals - fourth edition; approved guideline GP2-A4. Wayne, PA: NCCLS, 2002.

RLM.01400 Phase II N/A YES NO

Is there documentation of at least annual review of all policies and procedures in the laboratory by the current laboratory director or designee?

REPRODUCTIVE LABORATORY Page 14 of 59

Page 15: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

NOTE: The director must ensure that the collection of policies and technical protocols is complete, current, and has been thoroughly reviewed by a knowledgeable person. Technical approaches must be scientifically valid and clinically relevant. To minimize the burden on the laboratory and reviewer(s), it is suggested that a schedule be developed whereby roughly 1/12 of all procedures are reviewed monthly. While each procedure should have evidence of annual review, signature or initials on each page of a procedure is not required. A single Title Page or Index of all procedures with only one signature is not sufficient documentation that each procedure has been carefully reviewed.

COMMENTARY:

There must be documentation of at least annual review of all policies and procedures in the laboratory by the current laboratory director or designee. The director is responsible for ensuring that the collection of technical protocols is complete, current, and has been thoroughly reviewed by a knowledgeable person. Technical approaches must be scientifically valid and clinically relevant. To minimize the burden on the laboratory and reviewer(s), it is suggested that a schedule be developed whereby roughly 1/12 of all procedures are reviewed monthly. While each procedure should have evidence of annual review, signature or initials on each page of a procedure is not required. A single title page or index of all procedures with only one signature is not sufficient documentation that each procedure has been carefully reviewed.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 1992(Feb 28):7173 [42CFR493.1407(e)(13)].

RLM.01500 Phase II N/A YES NO

Does the laboratory have a system documenting that all personnel are knowledgeable about the contents of procedure manuals (including changes) relevant to the scope of their testing activities?

COMMENTARY:

The laboratory must have a system documenting that all personnel are knowledgeable about the contents of procedure manuals (including changes) relevant to the scope of their testing activities. This does not specifically require annual procedure sign-off by testing personnel. The form of this system is at the discretion of the Laboratory Director.

RLM.01600 Phase II N/A YES NO

If there is a change in directorship, does the new director ensure (over a reasonable period of time) that laboratory procedures are well-documented and undergo at least annual review?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 15 of 59

Page 16: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

If there is a change in directorship of the laboratory, the new director must ensure (over a reasonable period of time) that all chemistry laboratory procedures are well-documented and undergo at least annual review.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493.1251(d)].

RLM.01700 Phase II N/A YES NO

When a procedure is discontinued, is a paper or electronic copy maintained for at least 2 years, recording initial date of use, and retirement date?

COMMENTARY:

A paper or electronic copy of a discontinued procedure must be maintained for at least 2 years, recording initial date of use, and retirement date.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493. 1251(e), 42CFR493.1105(a)(2)].

-----------------------------------------------------------------

REQUISITIONS, SPECIMEN RECEIPT, AND RESULTS REPORTING

-----------------------------------------------------------------

RLM.01800 Phase II N/A YES NO

Are there specific patient instructions for collection and prompt delivery of a semen sample to the laboratory?

NOTE: This should be written in simple terms in a language readily understood by the patient. Elements should include the need to abstain from ejaculation for 2-5 days before masturbation, avoidance of lubricants and other contamination, completeness of collection, use of the supplied container, maintenance of sample temperature, and prompt delivery.

COMMENTARY:

Patients must be provided with specific instructions for collection and prompt delivery of a semen sample to the laboratory. This should be written in simple terms in a language readily understood by the patient. Elements should include the need to abstain from ejaculation for 2-5 days before masturbation, avoidance of lubricants and other contamination, completeness of collection, use of the supplied container, maintenance of sample temperature, and prompt delivery.

REPRODUCTIVE LABORATORY Page 16 of 59

Page 17: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.01900 Phase II N/A YES NO

Are there documented criteria for the rejection of unacceptable specimens or the special handling of sub-optimal specimens?

NOTE: This question is not intended to imply that all "unacceptable" specimens be discarded or not analyzed. For example, if an unacceptable specimen is received, there must be a mechanism to notify the requesting physician, and to note the condition of the sample on the report. Many semen samples are sub-optimal; all samples should be evaluated and unusual properties noted. The laboratory may wish to record that a dialogue was held with the requesting physician.

COMMENTARY:

Documented criteria must be developed for acceptability or rejection of specimens. If compromised specimens are accepted, the final report must have a note indicating the nature of the problem and, if applicable, caution in interpreting the result. The laboratory may wish to record that a dialogue was held with the requesting physician.

RLM.02000 Phase II N/A YES NO

Are semen specimens accompanied by the following collection information, and are records maintained on the following?

1. Method of collection2. Type of specimen container3. Days of abstinence4. Collection or transport problems (e.g., incomplete specimen, exposure to

temperature extremes)5. Abnormalities of liquefaction6. Time of specimen receipt and analysis

COMMENTARY:

Semen specimens must be accompanied by the following collection information, with records maintained.

1. Method of collection2. Type of specimen container3. Days of abstinence4. Collection or transport problems (e.g., incomplete specimen, exposure to temperature

extremes)5. Abnormalities of liquefaction6. Time of specimen receipt and analysis

REPRODUCTIVE LABORATORY Page 17 of 59

Page 18: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.02100 Phase II N/A YES NO

Are all semen specimens given sufficient time for liquefaction before testing?

COMMENTARY:

All semen specimens must be given sufficient time for liquefaction before testing.

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.02200 Phase II N/A YES NO

Are semen specimens mixed thoroughly before testing?

COMMENTARY:

Semen specimens must be mixed thoroughly before testing.

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.02300 Phase II N/A YES NO

Are all characteristics of the semen specimens noted and reported (e.g., gelatinous clumps, viscosity, contaminants, erythrocytes, etc.)?

COMMENTARY:

All characteristics of the semen specimens must be noted and reported.

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

REPRODUCTIVE LABORATORY Page 18 of 59

Page 19: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.02400 Phase II N/A YES NO

Are patient results reported in a legible, easy-to-interpret format that clearly delineates the clinical significance of the results?

COMMENTARY:

Patient results must be reported in a legible, easy to interpret format that clearly indicates the clinical implication of the result. Data must be legible, accurate, reported in clearly designated units of measurement, and reported to persons authorized to receive and use information.

RLM.02500 Phase II N/A YES NO

Where possible, are all patient results reported with accompanying reference (normal) intervals or interpretations?

COMMENTARY:

The laboratory must report reference (normal) intervals or interpretations with patient results, where such exist. This is important to allow proper interpretation of patient data. In addition, the use of high and low flags (generally available with a computerized laboratory information system) is recommended.

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7162 [42 CFR 1291(d)]; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.02600 Phase I N/A YES NO

Are results routinely available within a time compatible with clinical needs?

COMMENTARY:

All results, whether "routine" or "STAT," must be available within a time compatible with clinical needs. There should be sufficient personnel to promptly provide all necessary services as required. Mechanisms should be in place to provide back-up for laboratory personnel as required.

-----------------------------------------------------------------

GENERAL QUALITY CONTROL

-----------------------------------------------------------------

REPRODUCTIVE LABORATORY Page 19 of 59

Page 20: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.02700 Phase II N/A YES NO

Is there a schedule or system AVAILABLE AT THE INSTRUMENT for the regular checking of the critical operating characteristics for all instruments in use?

NOTE: This must include, but is not limited to electronic, mechanical, and operational checks. The procedure and schedule must be as thorough and as frequent as specified by the manufacturer.

COMMENTARY:

There must be a routine plan or schedule available at the instrument for the regular checking of the critical operating characteristics of all the instruments in use. The laboratory should have an organized system for monitoring and maintaining all instruments. This must include, but is not limited to electronic, mechanical, and operational checks. The procedure and schedule must be as thorough and as frequent as specified by the manufacturer. Function checks should be designed to check the critical operating characteristics to detect drift, instability, or malfunction, before the problem is allowed to affect test results. All servicing and repairs should be documented. The inspector should have identified the specific instruments involved at the summation conference.

RLM.02800 Phase II N/A YES NO

For QUALITATIVE tests, is a positive and negative control included with each run of patient specimens?

COMMENTARY:

For qualitative tests, a positive and negative control must be included with each run of patient specimens.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7166 [42CFR493. 1256(d)(3)(ii)].

RLM.02900 Phase II N/A YES NO

Are control specimens tested in the same manner and by the same personnel as patient samples?

COMMENTARY:

It is implicit in quality control (QC) that control specimens are tested in the same manner as patient specimens. Moreover, QC specimens must be analyzed by personnel who routinely perform patient testing - this does not imply that each operator must perform QC daily, so long as each instrument and/or test system has QC performed at required frequencies, and all analysts participate in QC on a

REPRODUCTIVE LABORATORY Page 20 of 59

Page 21: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

regular basis. To the extent possible, all steps of the testing process must be controlled, recognizing that pre-analytic and post-analytic variables may differ from those encountered with patients.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7166 [42CFR493. 1256(d)(8)].

RLM.03000 Phase II N/A YES NO

Are the results of controls verified for acceptability before reporting results?

COMMENTARY:

Controls must be reviewed before reporting patient results. It is implicit in quality control that patient test results will not be reported when controls do not yield acceptable results.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7166 [42CFR493. 1256(f)].

RLM.03100 Phase II N/A YES NO

Are quality control data organized and presented so they can be evaluated daily by the technical staff to detect problems, trends, etc?

COMMENTARY:

Results of controls must be recorded or plotted to readily detect a malfunction in the instrument or in the analytic system. These control records must be readily available to the person performing the test. Quality control data must be organized and presented so they can be evaluated daily by the technical staff to detect problems, trends, etc.

REFERENCE: NCCLS. Statistical quality control for quantitative measurements: principles and definitions - second edition; approved guideline C24-A2. Wayne, PA: NCCLS, 1998.

------------------------------------------------------------

REAGENTS AND CULTURE MEDIA

-------------------------------------------------------------

REPRODUCTIVE LABORATORY Page 21 of 59

Page 22: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.03200 Phase II N/A YES NO

Are reagents and solutions properly labeled, as applicable and appropriate, with the following elements?

1. Content and quantity, concentration or titer2. Storage requirements3. Date prepared or reconstituted by laboratory4. Expiration date

NOTE: The above elements may be recorded in a log (paper or electronic), rather than on the containers themselves, providing that all containers are identified so as to be traceable to the appropriate data in the log. While useful for inventory management, labeling with "date received" is not routinely required. There is no requirement to routinely label individual containers with "date opened"; however, a new expiration date must be recorded if opening the container changes the expiration date, storage requirement, etc. The inspector will describe specific issues of non-compliance in the Inspector's Summation Report.

COMMENTARY:

N/A

REFERENCES: 1) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493. 1252(c)]; 2) Jones HW, et al. In vitro fertilization. Baltimore, MD: Williams and Wilkins, 1986:178.

RLM.03300 Phase II N/A YES NO

Are all reagents used within their indicated expiration date?

COMMENTARY:

Reagents must not be used beyond their stated or assigned expiration date.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7164 [42CFR493. 1252(d)].

RLM.03400 Phase II N/A YES NO

Is the water used in preparing culture media type I water or special purpose water which has been defined and documented as being suitable for use with human in vitro fertilization culture media?

REPRODUCTIVE LABORATORY Page 22 of 59

Page 23: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

COMMENTARY:

Type I water or special purpose water which has been defined and documented as suitable for use with human in vitro fertilization must be used in preparing culture media. Water is one of the most important agents used in the media preparation since contaminants in water are quite frequently encountered that can have a substantially detrimental effect on oocyte and embryo growth.

REFERENCE: NCCLS. Preparation and testing of reagent water in the clinical laboratory, third edition; approved guideline C3-A3. Wayne, PA: NCCLS, 1997.

RLM.03500 Phase II N/A YES NO

Are explicit procedures for media preparation documented?

NOTE: Mark this question "N/A" if no media are prepared by the laboratory.

COMMENTARY:

Explicit procedures for media preparation must be documented. Procedures for media preparation should be clearly documented and readily available to laboratory personnel. If there is no de novo preparation of media in the laboratory, procedures for preparation of purchased media for use need not be written. All media preparation must be performed with sterile technique, including location and environment appropriate for media preparation. The laboratory has a responsibility for ensuring that any media purchased or prepared is sterile and capable of supporting culture of gametes and embryos.

REFERENCE: Jones HW, et al. In vitro fertilization. Baltimore, MD: Williams and Wilkins, 1986;178.

RLM.03600 Phase II N/A YES NO

Are media storage and expiration requirements documented?

COMMENTARY:

Media storage and expiration requirements must be documented at all times in order to ensure active reagents in all media.

REFERENCE: Jones HW, et al. In vitro fertilization. Baltimore, MD: Williams and Wilkins, 1986;178.

REPRODUCTIVE LABORATORY Page 23 of 59

Page 24: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.03700 Phase II N/A YES NO

Does the laboratory have a method for quality control of media and is it documented?

COMMENTARY:

The laboratory must have a documented method for quality control of the media and all media supplements. Culture media must be able to support the viability of gametes and/or and the growth of embryos. Media must be evaluated using a bioassay system such as the one or two cell mouse embryo culture assay or a sperm motility assay. If culture media or protein supplement is purchased, there must be documentation that it has been tested by the supplier. It is highly recommended that new media be tested on site. Quality control testing is highly recommended when commercial media is purchased and used within its labeled expiration period, since pretesting by the manufacturer may not reflect media suitability when in actual use in the laboratory. Documentation of quality control testing using an appropriate bioassay system must always be supplied by the manufacturer

REFERENCE: American Fertility Society guidelines for human embryology laboratories. August 1990, section IV, C, 1c.

RLM.03800 Phase I N/A YES NO

Does the laboratory test and document the quality of the contact material using a bioassay?

COMMENTARY:

Contact materials should be evaluated routinely by a bioassay.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58(suppl 1):IV,C, 1a and VI, A, 5.

RLM.03900 Phase II N/A YES NO

Do the records indicate that, when components are prepared that do not meet the quality control requirements, immediate corrective action was taken and documented?

COMMENTARY:

The records must indicate that when components are prepared that do not meet the quality control requirements, immediate corrective action is taken and documented.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58(suppl 1):IV, A, 2.

REPRODUCTIVE LABORATORY Page 24 of 59

Page 25: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

****************************************************************

ANDROLOGY PROCEDURES AND TESTS

*****************************************************************

-----------------------------------------------------------------

COMPUTER-ASSISTED SEMEN ANALYSIS (CASA) AUTOMATED INSTRUMENTS

-----------------------------------------------------------------

Varieties of systems are in use and some questions may not apply to every system. The questions are intended to check factors common to all automated systems. Inspectors should use individual judgment in applying the questions to the particular type of system being used.

RLM.04000 Phase I N/A YES NO

Where applicable, are the optical and imaging components used those recommended by the manufacturer or a validated substitution?

COMMENTARY:

Where applicable, the optical and imaging components used should be those recommended by the manufacturer or the substitution should be validated by the laboratory.

.................................................................

Calibration and Quality Control

.................................................................

Several different methods may be used for calibration and quality control in the automated analysis of semen characteristics. "Calibration" techniques include use of:

1. Multiple analyzed sperm specimens,2. Stabilized preparations of sperm cells (e.g., fixed or preserved),3. Sperm surrogates (e.g., latex particles),4. Videotaped sperm specimens.

NOTE: If stabilized control materials are used, they should represent different analytic levels (e.g., normal and high). Similarly, retained patient specimens should be of differing counts and/or motility, as applicable.

REPRODUCTIVE LABORATORY Page 25 of 59

Page 26: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.04100 Phase II N/A YES NO

Is calibration verified with materials appropriate to the reportable range of the instrument, and is verification documented?

COMMENTARY:

The quality control procedure for the automated instrument must include calibration and evaluation using defined limits of agreement with manually counted semen smears or stored digital images, as appropriate for the particular system. Computer-assisted semen analysis laboratories must periodically establish that their computer assisted semen analysis equipment is functioning correctly and there is a protocol to determine if the analysis is in control.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58(suppl 1):I, B, 6.

RLM.04200 Phase II N/A YES NO

Does the laboratory perform and document calibration and quality control methods for the analyzer during each day of use, using an appropriate number of controls with varying ranges of values?

COMMENTARY:

The laboratory must perform and document calibration and quality control methods for the analyzer during each day of use, using an appropriate number of controls with varying ranges of relevant values.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58(suppl 1):I, B, 6.

RLM.04300 Phase II N/A YES NO

Does the laboratory have a procedure for recalibration of CASA instrument parameter(s) when problems are encountered?

COMMENTARY:

The laboratory must have a procedure for recalibration of CASA instrument parameter(s) when problems are encountered.

REFERENCE: World Health Organization. Laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992.

REPRODUCTIVE LABORATORY Page 26 of 59

Page 27: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.04400 Phase II N/A YES NO

Has the material used for calibration been validated using primary reference procedures (e.g., manual counts)?

COMMENTARY:

The material used for calibration must have been validated using primary reference procedures.

REFERENCES: 1) World Health Organization. Laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Krause W. [Value of computer-assisted sperm analysis (CASA). reproducibility--online documentation--prognostic value]. [Article in German]. Fortschr Med. 1996;114:470-473; 3) Tsuji T, et al. Automated sperm concentration analysis with a new flow cytometry-based device, S_FCM. Am J Clin Pathol. 2002;117:401-408.

RLM.04500 Phase II N/A YES NO

If a manual method is used as the system control for automated or semi-automated sperm counts, is its accuracy verified and documented at intervals appropriate for laboratory volume?

COMMENTARY:

When a manual method is used as the system control for automated or semi-automated sperm counts, its accuracy must be verified and documented periodically.

REFERENCES: 1) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press. 1994; 2) Lenzi A. Computer-aided semen analysis (CASA) 10 years later: a test-bed for the European scientific andrological community. Int J Androl. 1997;20:1-2; 3) Mahmoud AM, et al. Performance of the sperm quality analyser in predicting the outcome of assisted reproduction. Int J Androl. 1998;21:41-46; 4) Tsuji T, et al. Automated sperm concentration analysis with a new flow cytometry-based device, S_FCM. Am J Clin Pathol. 2002;117:401-408.

RLM.04600 Phase II N/A YES NO

Are tolerance limits established for the value of each quality control sample?

COMMENTARY:

Tolerance limits must be established for the value of each sample used in quality control.

REPRODUCTIVE LABORATORY Page 27 of 59

Page 28: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.04700 Phase II N/A YES NO

If sperm motility and/or count are assessed by CASA, is a procedure established to determine that the concentration of the specimen is within the range appropriate for automated counting?

COMMENTARY:

If sperm motility and/or count are assessed by CASA, a procedure must be established to determine that the concentration of the specimen is within the range appropriate for automated counting.

REFERENCES: 1) Vantman DD, et al. Computer-assisted semen analysis: evaluation of method and assessment of the influence of sperm concentration on linear velocity determination. Fertil Steril. 1988;49:510-515; 2) Yeung CH, et al. A technique for standardization and quality control of subjective sperm motility assessments in semen analysis. Fertil Steril. 1997;67:1156-1158; 3) Sidhu RS, et al. accuracy of computer-assisted semen analysis in prefreeze and post-thaw specimens with high and low sperm counts and motility. Urology. 1998;51:306-312; 4) Tsuji T, et al. Automated sperm concentration analysis with a new flow cytometry-based device, S_FCM. Am J Clin Pathol. 2002;117:401-408.

RLM.04800 Phase I N/A YES NO

Has the laboratory established reference intervals (normal values) for the motion variables measured?

COMMENTARY:

The laboratory should establish reference ranges for the motion variables measured during sperm motility assessment.

REFERENCE: Schieferstein G, et al. Sperm motility index. Arch Androl. 1998;40:43-48.

RLM.04900 Phase II N/A YES NO

Are upper and lower limits of all reportable parameters on CASA instruments defined so that results that fall outside these limits are verified before reporting?

COMMENTARY:

The laboratory must verify the linearity range for each reportable parameter of its instruments. Results that fall outside of these limits may be verified by repeating the test, using an alternative method or diluting/concentrating the specimen, as appropriate.

REFERENCE: Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press. 1994.

REPRODUCTIVE LABORATORY Page 28 of 59

Page 29: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.05000 Phase II N/A YES NO

Are criteria established for method calibration verification?

NOTE: Criteria must be established for method calibration verification. Criteria for determining the need for calibration verification typically include:

1. At complete changes of reagents, unless the laboratory can demonstrate that changing reagent lots does not affect either the range used to report patient test results or the control values,

2. When indicated by quality control data,3. After major maintenance or service,4. When recommended by the manufacturer,5. At least every six months.

COMMENTARY:

Criteria must be established for method calibration verification. Criteria for determining the need for calibration verification typically include:

1. At complete changes of reagents,2. When indicated by quality control data,3. After major maintenance or service,4. When recommended by the manufacturer,5. At least every six months.

When calibration verification criteria are exceeded, the laboratory must recalibrate.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7165 [42 CFR 493.1255].

----------------------------------------------------------------

MANUAL SEMEN ANALYSIS

-----------------------------------------------------------------

.................................................................

Sperm Concentration

.................................................................

REPRODUCTIVE LABORATORY Page 29 of 59

Page 30: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.05100 Phase I N/A YES NO

Are the lines in the counting or motility chambers bright, and are the chambers clean and free of scratches?

NOTE: The inspector should examine several chambers.

COMMENTARY:

All lines on the counting and motility chambers should be bright, and the chambers should be clean and free of scratches.

RLM.05200 Phase II N/A YES NO

Is each sperm sample counted in duplicate, usually through counting both chambers of a standard hemocytometer?

NOTE: Defined limits of agreement between replicate counts must be established.

COMMENTARY:

If sperm counts are performed manually by pipette dilution and chamber count, each sample must be counted in duplicate, plating both chambers of a standard hemocytometer. Defined limits of variation must be established.

RLM.05300 Phase II N/A YES NO

Does the laboratory indicate on the report that results may be inaccurate due to the presence of cell clumps or debris in the counting chamber?

COMMENTARY:

The laboratory must indicate on the report that results may be inaccurate due to the presence of cell clumps or debris in the counting chamber.

RLM.05400 Phase I N/A YES NO

Is there an additional procedure beyond unstained bright-field microscopy to ensure the accurate distinction of leukocytes from other round cells (e.g., Wright's or PAP stain, leukocyte alkaline phosphatase, myeloperoxidase)?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 30 of 59

Page 31: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

There must be an additional procedure beyond unstained bright-field microscopy to ensure the accurate distinction of leukocytes from other round cells (e.g., Wright's or PAP stain, leukocyte alkaline phosphatase, myeloperoxidase).

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York: Cambridge University Press, 1992; 2) Fishel TJ, et al. Increased polymorphonuclear granulocytes in seminal plasma in relation to sperm morphology. Hum Reprod. 1997;12:2418-2421; 3) Zimmermann BS, et al. Relationship of bacteriological characteristics to semen indices and its influence on fertilization and pregnancy rates after IVF. Acta Obstet Gynecol Scand. 1997;76:964-968; 4) Trum JW, et al. Value of detecting leukocytospermia in the diagnosis of genital tract infection in subfertile men. Fertil Steril. 1998;70:315-319.

RLM.05500 Phase II N/A YES NO

For azoospermic specimens, as well as post-vasectomy checks for reproductive sterility, is a concentrating technique employed?

COMMENTARY:

For azoospermic specimens, as well as post-vasectomy checks for reproductive sterility, a concentrating technique must be employed. Without such an approach, the presence of both motile and non-motile sperm may not be detected.

REFERENCES: 1) Jones CD, Cornbleet PJ. Wright-Giemsa cytology of body fluids. techniques for optimal cytocentrifuge slide preparation. Lab Med. 1997;28:713-716; 2) Jaffe TM, et al. Sperm pellet analysis: a technique to detect the presence of sperm in men considered to have azoospermia by routine semen analysis. J Urol. 1998;159:1548-1550.

.................................................................

Sperm Motility

.................................................................

RLM.05600 Phase II N/A YES NO

Is sperm motility percent and progression routinely evaluated within 1 hour of receipt?

COMMENTARY:

Sperm motility percent and progression must be evaluated within 1 hour of receipt.

REPRODUCTIVE LABORATORY Page 31 of 59

Page 32: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.05700 Phase II N/A YES NO

Has the laboratory established a standard specimen temperature range for semen analysis assessment, and are deviations from this temperature noted on the report?

COMMENTARY:

The laboratory must establish a standard specimen temperature range for semen evaluation assessment, and deviations from this temperature must be noted on the report, since specimen motility is temperature-dependent.

REFERENCE: World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992.

RLM.05800 Phase I N/A YES NO

Has the laboratory defined when viability measurements should be performed on specimens with abnormally low percent motility (e.g., less than 30%)?

COMMENTARY:

Non-motile sperm may represent forms that were originally non-viable in the ejaculate, or previously motile forms that have subsequently lost motility. Thus, viability assessment is useful in making the distinction, and is commonly performed with a dye-exclusion method such as eosin-nigrosin. The laboratory should define when viability measurements should be performed on specimens with abnormally low percent motility (e.g., less than 30%).

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction: fourth edition. New York, NY: Cambridge University Press, 1999; 2) Gunalp S, et al. A study of semen parameters with emphasis on sperm morphology in a fertile population: an attempt to develop clinical thresholds. Hum Repro. 2001;16:110-114.

RLM.05900 Phase II N/A YES NO

Does the laboratory have an appropriate procedure for evaluating a sufficient number of separate and randomly chosen microscopic fields and sperm cells?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 32 of 59

Page 33: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The laboratory must determine an appropriate procedure for evaluating a sufficient number of separate and randomly chosen microscopic fields and number of motile sperm cells.

REFERENCE: World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992.

RLM.06000 Phase II N/A YES NO

Are manual measures of percent sperm motility quantified objectively?

COMMENTARY:

Manual measures of sperm motility must be quantified objectively.

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Yeung CH, et al. A technique for standardization and quality control of subjective sperm motility assessments in semen analysis. Fertil Steril. 1997;67:1156-1158.

RLM.06100 Phase II N/A YES NO

Is forward progression of sperm evaluated?

COMMENTARY:

The forward progression of sperm must be evaluated.

REFERENCES: 1) World Health Organization laboratory manual for the examination of human semen and sperm-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 2) Vulcano GJ, et al. A lineal equation for the classification of progressive and hyperactive spermatozoa. Math Biosci. 1998;149:77-93.

RLM.06200 Phase II N/A YES NO

Does a procedure exist to verify the sperm motility method used (e.g., video tapes of specimens with known percent motility and/or specific motion quality)?

COMMENTARY:

A procedure must exist to measure and verify the sperm motility method used (e.g., video tapes of specimens with known motility).

REPRODUCTIVE LABORATORY Page 33 of 59

Page 34: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

REFERENCES: 1) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994; 2) Yeung CH, et al. A technique for standardization and quality control of subjective sperm motility assessments in semen analysis. Fertil Steril. 1997;67:1156-1158.

.................................................................

Semen Stained Smear - Sperm Differential

.................................................................

RLM.06300 Phase I N/A YES NO

Are stains used to facilitate classification of cell types (as opposed to performing differentials of unstained preparations)?

COMMENTARY:

Stains should always be used to facilitate classification of cell types, as opposed to performing differentials on unstained preparations.

REFERENCE: Coetzee K, et al. Predictive value of normal sperm morphology: a structured literature review. Hum Reprod Update. 1998;4:73-82.

RLM.06400 Phase II N/A YES NO

Are all stains checked for intended reactivity each day of use?

COMMENTARY:

Stains must be checked each day of use for intended reactivity to ensure predictable staining characteristics.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7166 [42CFR493.1256(e)(2)].

RLM.06500 Phase I N/A YES NO

Examine a smear. Is the quality satisfactory (uniform cell distribution, properly stained, ready recognition of cell types that are reported)?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 34 of 59

Page 35: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The quality of the smear should be satisfactory (uniform cell distribution, properly stained, ready recognition of cell types that are reported).

REFERENCE: Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.06600 Phase II N/A YES NO

Are slides adequately identified?

COMMENTARY:

Slides must be adequately identified. Slide identification must include a unique specimen or accession number, patient name and/or number, date.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): IV, D.

RLM.06700 Phase II N/A YES NO

Is the classification method used indicated on the report?

COMMENTARY:

The classification method must be indicated on the report, since many different classification systems are in use.

RLM.06800 Phase I N/A YES NO

Are the slides retained for at least 30 days for future reference?

COMMENTARY:

Slides should be retained for at least 30 days for future reference.

RLM.06900 Phase I N/A YES NO

Does the laboratory have a defined, documented system to ensure consistency of morphologic observations among all personnel performing microscopic morphologic classification of sperm and other cells?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 35 of 59

Page 36: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The laboratory must have a documented system that ensures that all personnel report microscopic morphologic data on patient samples in a similar fashion. For initial accuracy as well as consistency in serial samples from the same patient, the laboratory must be able to document that all of its staff are consistent with respect to morphologic classification. Suggested methods to accomplish this include:

1. Circulation of stained semen smears with defined specific qualitative abnormalities of sperm, and/or

2. Multi-headed microscopy, and/or3. Use of reliable published atlases.

REFERENCES: 1) Freund M. Standards for the rating of human sperm morphology: a cooperative study. Int J Fertil. 1966;11(suppl):1-9; 2) Calamera JC, Vilar O. Comparative study of sperm morphology with three different staining procedures. Andrologia. 1979;11:255-258; 3) Urry R. seminal fluid. In: Kjeldsberg CR, Knight JA. Body fluids. laboratory examination of amniotic, cerebrospinal, seminal, serous, & synovial fluids: a textbook atlas, 2nd edition. Chicago, IL: American Society of Clinical Pathology, 1986:117-127; 4) Adelman MM, Cahill EM. Atlas of sperm morphology. Chicago, IL: American Society of Clinical Pathology, 1989; 5) World Health Organization laboratory manual for the examination of human semen and semen-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992; 6) Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.07000 Phase I N/A YES NO

Is there a file of unusual slides?

COMMENTARY:

There should be a file of unusual slides that are retained for training and demonstration.

RLM.07100 Phase II N/A YES NO

Are the stains used (e.g., Wright’s, Papanicolaou, eosin/nigrosin, etc.) of sufficient quality to demonstrate the cellular characteristics for which they are designed?

COMMENTARY:

The selection of semen stains is at the discretion of the director. Of those used, they must be of sufficient quality to demonstrate the cellular characteristics for which they are designed.

RLM.07200 Phase II N/A YES NO

Are controls (where available and when appropriate) routinely run on all special stains?

REPRODUCTIVE LABORATORY Page 36 of 59

Page 37: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

COMMENTARY:

Controls (where available and when appropriate) must be routinely run on all special stains. The expected staining of normal cells on the semen smear may be absent for technical reasons. Failure to evaluate the expected reaction of normal cells may cause diagnostic errors.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): VII, A, 1.

RLM.07300 Phase II N/A YES NO

Are stains examined for contamination and replaced, as appropriate?

COMMENTARY:

Stains must be examined periodically for contamination and replaced, as appropriate.

.................................................................

Biochemical Tests

.................................................................

RLM.07400 Phase II N/A YES NO

For biochemical tests such as fructose and zinc, are positive and negative controls run with each assay?

COMMENTARY:

Positive and negative controls must be run with each biochemical test such as fructose and zinc.

REFERENCE: Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

-----------------------------------------------------------------

ANTI-SPERM ANTIBODY (ASA) TESTS

-----------------------------------------------------------------

REPRODUCTIVE LABORATORY Page 37 of 59

Page 38: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.07500 Phase II N/A YES NO

Are serum and follicular fluid specimens used for indirect ASA testing heat-inactivated before use?

COMMENTARY:

All serum and follicular fluid specimens used for indirect ASA testing must be treated to inactivate complement.

REFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press, 19RLM.185.

RLM.07600 Phase I N/A YES NO

If the testing for ASA requires motile sperm, are specimens assayed with minimal delay and is the motility assessed and recorded?

COMMENTARY:

When ASA testing requires motile sperm, specimens should be assayed with minimal delay, and the motility should be assessed and recorded.

REFERENCE: Mortimer D. Practical laboratory andrology. New York, NY: Oxford University Press, 1994.

RLM.07700 Phase II N/A YES NO

For indirect antibody testing, are positive and negative controls run with each assay?

COMMENTARY:

Positive and negative controls must be run with each assay.

REFERENCES: 1) Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press, 19RLM.185; 2) Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7166 [42CFR493. 1256(d)(iii)]; 3) Evans ML, et al. A convenient mixed immunobeads screen for antisperm antibodies during routine semen analysis. Fertil Steril. 1998;70:344-349.

REPRODUCTIVE LABORATORY Page 38 of 59

Page 39: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

-----------------------------------------------------------------

SPERM PROCESSING FOR THERAPEUTIC INSEMINATION

-----------------------------------------------------------------

RLM.07800 Phase II N/A YES NO

Are special handling requirements for insemination specimens defined and followed (e.g., aseptic technique, processing with minimum delay), as necessary?

COMMENTARY:

Special handling requirements for insemination specimens must be defined and followed (i.e., aseptic technique, processing with minimum delay) as necessary.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section I, B, 5.

RLM.07900 Phase II N/A YES NO

Are there documented procedures for preparing sperm for insemination (e.g., swim-up technique)?

COMMENTARY:

There must be documented procedures for preparing sperm for insemination (e.g., swim-up technique).

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section IV, B, 2.

RLM.08000 Phase II N/A YES NO

Is there a system to verify and maintain the identity of the specimen throughout receipt, storage, processing, and disposition?

COMMENTARY:

Procedures must be in place to ensure that insemination specimens are correctly identified for the recipient.

REFERENCE: Byrd W. Quality assurance in the reproductive biology laboratory. Arch Pathol Lab Med. 1992;116:418-422.

REPRODUCTIVE LABORATORY Page 39 of 59

Page 40: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.08100 Phase II N/A YES NO

Is there documentation of all individuals handling the specimen from receipt to final disposition?

COMMENTARY:

Documentation must be in place to identify the individual handling the specimen for all phases of sperm processing.

REFERENCE: World Health Organization laboratory manual for the examination of human semen and semen-cervical mucus interaction, 3rd edition. New York, NY: Cambridge University Press, 1992.

*****************************************************************

ANDROLOGY PERSONNEL

*****************************************************************

The laboratory should have an organizational plan, personnel policies, and job descriptions that define qualifications and duties for all positions. Personnel files should contain qualifications, references, performance evaluations, health records, and continuing education records for each employee. Ideally, these files should be located in the laboratory. However, they may be kept in the personnel office or health clinic if the laboratory has ready access to them (i.e., easily available to the inspector).

The director of the andrology laboratory should have the appropriate training and background to assume responsibility for the overall operation and administration of the laboratory, including hiring competent personnel, formulating laboratory policies and protocols, and communicating regularly regarding patient progress and patient protocols as they affect laboratory aspects of treatment. The director must be accessible to the laboratory for on-site, telephone or electronic consultation as needed.

RLM.08200 Phase II N/A YES NO

Does the director and all other personnel in the andrology laboratory meet the requirements described in the Laboratory General Checklist?

NOTE: The director of the andrology laboratory and other andrology laboratory personnel must meet the personnel qualifications for "high complexity" testing laboratories (CLIA-88 terminology) as delineated in the Laboratory General Checklist Personnel Section.

The director should have two years documented experience in a laboratory performing andrology procedures. This experience should include quality management, quality control, inspection, accreditation and licensing procedures, as well as tissue culture, ART and andrology procedures.

REPRODUCTIVE LABORATORY Page 40 of 59

Page 41: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

COMMENTARY:

N/A

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Oct 1):1056[42CFR493.1443], 1058[42CFR493.1449], 1066[42CFR493.1459], 1067[42CFR493.1461], 1070[42CFR493.1489].

#################################################################

EMBRYOLOGY

#################################################################

Embryology laboratories may have separate andrology facilities that are not accredited by the College of American Pathologists. However, if the embryology laboratory either processes sperm for therapeutic insemination, oocyte insemination, or performs any form of a semen analysis, it must complete questions in the preceding SPERM PROCESSING FOR THERAPEUTIC INSEMINATION section and/or the ANDROLOGY section.

If no embryology procedures are performed in the laboratory, mark all questions in this section "N/A" and continue with the PERSONNEL section.

*****************************************************************

OOCYTE AND EMBRYO HANDLING

*****************************************************************

-----------------------------------------------------------------

CULTURE OF SPERM, OOCYTES, AND EMBRYOS

-----------------------------------------------------------------

RLM.08300 Phase II N/A YES NO

Are sterile techniques employed in the handling, assessment and culturing of human sperm, oocytes, and embryos?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 41 of 59

Page 42: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

Sterile techniques must be employed in the handling, assessment and culturing of human sperm, oocytes, and embryos. All procedures relating to recovery of eggs from follicular aspirates must be performed using sterile technique. All procedures relating to embryo transfer must be performed using sterile technique, as far as possible.

RLM.08400 Phase II N/A YES NO

Are there documented criteria for evaluation/assessment of oocyte maturity and embryo quality?

COMMENTARY:

There must be documented criteria for evaluation and assessment of oocyte maturity and embryo quality. Documented protocols should include description of oocyte and embryo quality and maturity. The stage of embryo development at transfer must be documented.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VIII, B, 2.

RLM.08500 Phase II N/A YES NO

Are there documented criteria for insemination relative to oocyte maturity?

COMMENTARY:

There must be documented criteria for insemination relative to oocyte maturity. Documented protocols must include remedial steps to be used for immature and/or atretic oocytes.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section B, 1.

RLM.08600 Phase II N/A YES NO

Are there defined criteria for volume and number of sperm used for insemination of each egg?

COMMENTARY:

There must be defined criteria for volume and number of sperm used for insemination of each egg. Techniques must be documented for estimation of sample parameters for concentration, motility, and morphology along with techniques for insemination with respect to count and motility for both normal and male factor patients.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section V, 1-4.

REPRODUCTIVE LABORATORY Page 42 of 59

Page 43: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.08700 Phase II N/A YES NO

Is there a documented policy for the disposition of oocytes with an abnormal number of pronuclei?

COMMENTARY:

There must be a documented policy for the disposition of poly-pronuclear oocytes. Embryos with three or more pronuclei should not be transferred.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 8.

RLM.08800 Phase II N/A YES NO

Is there a defined period for examination of oocytes for fertilization?

COMMENTARY:

There must be a defined period for examination of oocytes for fertilization. The interval from oocyte insemination to examination should be 14-20 hours.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 2.

RLM.08900 Phase II N/A YES NO

Does the laboratory have documented criteria for re-insemination, using either in vitro fertilization or intracytoplasmic sperm injection?

COMMENTARY:

The laboratory must have documented criteria for re-insemination. Documented procedures for re-insemination of oocyte and/or micromanipulation should include time frame for re-insemination, criteria for use of initial sample, time frame for re-examination of these oocytes and the hierarchy for their use at embryo transfer.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 7.

REPRODUCTIVE LABORATORY Page 43 of 59

Page 44: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

-----------------------------------------------------------------

MICROMANIPULATION OF EMBRYOS

-----------------------------------------------------------------

RLM.09000 Phase II N/A YES NO

Is there a documented program to train and evaluate personnel in their competency to perform micromanipulation?

COMMENTARY:

For laboratories performing micromanipulation, there must be a training program for new personnel, using animal model systems or nonviable human oocytes.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 2.

RLM.09100 Phase II N/A YES NO

Does the laboratory have a program to ensure that micromanipulation procedures are performed at an acceptable level?

COMMENTARY:

There must be a documented program to ensure that micromanipulation procedures are providing acceptable levels of performance. This would include fertilization of oocytes, survival following zona hatching, and pregnancy rates using micromanipulated embryos.

-----------------------------------------------------------------

EMBRYO TRANSFER PROCEDURES

-----------------------------------------------------------------

RLM.09200 Phase II N/A YES NO

Are there documented procedures for the length of time that embryos are cultured before transfer?

COMMENTARY:

There must be documented procedures for the length of time that embryos are cultured before transfer.

REPRODUCTIVE LABORATORY Page 44 of 59

Page 45: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 7.

RLM.09300 Phase II N/A YES NO

Does the laboratory document the status and quality of embryos before transfer?

COMMENTARY:

The laboratory must document the status and quality of embryos before transfer.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 2.

RLM.09400 Phase II N/A YES NO

Is the identity of the patient specimen (sperm or embryos) checked against the identity of the patient prior to transfer or insemination and is this identification documented?

COMMENTARY:

There must be an established chain of custody for all reproductive gametes or embryos that are transferred back to a patient. This includes documentation of the patient specimen identification (ID), as well as the patient's ID. When it is not possible for the laboratory staff to check the patient's ID, then this should be done and documented by a nurse, physician, or other health care provider before transfer.

RLM.09500 Phase II N/A YES NO

Does the laboratory check the catheter for any embryos left after transfer?

COMMENTARY:

The laboratory must check the catheter for any embryos left after transfer. Embryos can be retained in the catheter. These should be identified and the transfer may be repeated or the embryos frozen.

REFERENCE: Poindexter AN, et al. Residual embryos in failed embryo transfer. Fertil Steril. 1986;46:262-267.

REPRODUCTIVE LABORATORY Page 45 of 59

Page 46: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

*****************************************************************

RECORDS

*****************************************************************

RLM.09600 Phase II N/A YES NO

Are records generated and retained for each individual patient's treatment cycle?

COMMENTARY:

Documented records must be generated and retained for each individual patient's treatment cycle.

REFERENCE: Quigley MM. Data management in an in vitro fertilization and embryo transfer program. In: Human in vitro fertilization and embryo transfer. Wolf DP, Quigley, MM, eds. New York, NY: Plenum Press, 1984;383-402.

RLM.09700 Phase II N/A YES NO

Do the records contain information concerning the following?

1. Results of oocyte retrieval procedure2. Semen analysis before and after processing3. Disposition of all oocytes collected4. Outcome of insemination (e.g., fertilization)5. Outcome of any culture (e.g., cleavage)6. Disposition of all embryos7. Relative timing of protocol events (incubation hours, etc.)

NOTE: The inspector will detail any specific deficiencies in the Inspector’s Summation Report.

COMMENTARY:

The records must contain the results of oocyte retrieval, semen analysis before/after processing, disposition of all oocytes collected, information concerning the outcome of insemination (fertilization), and outcome of any culture (cleavage). Additionally, documented protocols must be established for the disposal of any poly-pronuclear or otherwise abnormal embryos (manner of discarding or nature of experimentation: micromanipulation, culturing, microscopy, etc.). The record must contain information concerning the relative timing of protocol events (incubation hours, etc.).

REFERENCES: 1) Sharma V, et al. An analysis of factors influencing the establishment of a clinical program in an ultrasound-based ambulatory in vitro fertilization program. Fertil Steril. 1988;49:468-478; 2) Byrd W, Wolf DP. Oogenesis, fertilization and early development. In: Human in

REPRODUCTIVE LABORATORY Page 46 of 59

Page 47: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

vitro fertilization and embryo transfer. Wolf DP, Quigley MM, eds. New York, NY: Plenum Press, 1984;213-273; 3) Veeck LL, et al. Maturation and fertilization of morphologically immature human oocytes in a program of in vitro fertilization. Fertil Steril. 1983;39:594-602; 4) Gerrity M. Quality control and laboratory monitoring. In: Human in vitro fertilization and embryo transfer. DP Wolf (ed). New York, NY: Plenum Press, 1988;25-45; 5) Tarkowski AK. Recent studies in parthenogenesis in the mouse. Reprod Fert suppl. 1971;14:31-39.

RLM.09800 Phase II N/A YES NO

Does the record identify the individual(s) performing each procedure?

COMMENTARY:

For quality assurance of IVF programs, the minimal standards suggested are that all certified embryology laboratory personnel perform at least 20 complete IVF procedures per year.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 2.

RLM.09900 Phase II N/A YES NO

Does the record identify specific media and protein solutions used in each phase of the procedure?

COMMENTARY:

To maintain proper quality control and quality assurance of IVF procedures, documentation of the source and lot number of each batch of culture medium and the source/lot of every protein supplement must be available for review.

REFERENCE: Gerrity M. Quality control and laboratory monitoring. In: Human in vitro fertilization and embryo transfer. DP Wolf (ed). New York, NY: Plenum Press, 1988;25-45.

RLM.10000 Phase II N/A YES NO

Is there a mechanism for reporting unusual or abnormal events to the supervisor, director, or physician?

COMMENTARY:

There must be a mechanism for reporting unusual or abnormal events to the supervisor, director, or physician.

REPRODUCTIVE LABORATORY Page 47 of 59

Page 48: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.10100 Phase II N/A YES NO

Is a copy of the record retained for laboratory files?

COMMENTARY:

A copy of the record must be retained in the laboratory.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, B, 2.

RLM.10200 Phase II N/A YES NO

Does the laboratory periodically review clinical outcome in relation to all data collected?

COMMENTARY:

The laboratory must periodically review clinical outcome in relation to all data collected. The laboratory should keep statistical records and periodically review clinical outcome in relation to this data.

REFERENCES: 1) Gerrity M. Quality control and laboratory monitoring. In: Human in vitro fertilization and embryo transfer. DP Wolf (ed). New York, NY: Plenum Press, 1988;25-45; 2) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): 1S-16S.

*****************************************************************

INSTRUMENTS AND EQUIPMENT

*****************************************************************

The laboratory should have an organized system for monitoring and maintaining all instruments.

RLM.10300 Phase II N/A YES NO

Does the laboratory have a documented procedure for the preparation of glassware used in therapeutic procedures?

COMMENTARY:

The laboratory must have a documented procedure for the preparation of contact glassware used in therapeutic procedures. Glassware must be prepared in a manner such that all toxins are removed before use. All contact glassware used for therapeutic procedures must be carefully processed.

REPRODUCTIVE LABORATORY Page 48 of 59

Page 49: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

Glassware washing protocols including detergent type and source, type of water used, number of rinses and exact procedure to be followed must be strictly defined. Heat sterilization should be used whenever possible.

REFERENCES: 1) Veeck LL, Maloney M. Insemination and fertilization. In: In vitro fertilization. Norfolk HW, et al (eds). Baltimore, MD: Williams and Wilkins, 1986;178-200; 2) Gerrity M. Quality control and laboratory monitoring. In: Human in vitro fertilization and embryo transfer. DP Wolf (ed). New York, NY: Plenum Press, 1988;25-45; 3) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): 1S-16S.

RLM.10400 Phase II N/A YES NO

Are there documented criteria for use of gas mixtures?

COMMENTARY:

There must be documented criteria for gas mixtures used. The gaseous environment that the oocytes and embryos are maintained in throughout the entire procedure must be outlined in the procedure manual.

REFERENCES: 1) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): 1S-16S; 2) Veeck LL, Maloney M. Insemination and fertilization. In: In vitro fertilization. Norfolk HW, et al (eds). Baltimore, MD: Williams and Wilkins, 1986;178-200.

RLM.10500 Phase II N/A YES NO

Is there documentation of checks of instrument function each day of use using an independent measuring device for the following?

1. Gas concentrations in incubators2. Temperature of incubators

COMMENTARY:

Daily recordings of gas concentrations and temperature must be made to enable corrective measures to be taken if values are not within acceptable documented limits.

REFERENCES: 1) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): 1S-16S; 2) Veeck LL, Maloney M. Insemination and fertilization. In: In vitro fertilization. Norfolk HW, et al (eds). Baltimore, MD: Williams and Wilkins, 1986;178-200; 3) Gerrity M. Quality control and laboratory monitoring. In: Human in vitro fertilization and embryo transfer. DP Wolf (ed). New York, NY: Plenum Press, 1988;25-45.

REPRODUCTIVE LABORATORY Page 49 of 59

Page 50: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.10600 Phase II N/A YES NO

Are acceptable limits of temperature, gas content, and humidity defined?

COMMENTARY:

Acceptable limits of temperature, CO2, content, and humidity must be defined. Only if limits are defined can variances be reported and action taken to correct the problem.

REFERENCES: 1) Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): 1S-16S; 2) Abramczyk JW, Lopata A. Incubator performance in the clinical in vitro fertilization program: importance of temperature conditions for the fertilization and cleavage of human embryos. Fertil Steril. 1986;46:132-134.

RLM.10700 Phase II N/A YES NO

Are digital displays of incubator temperature calibrated with a certified thermometer?

COMMENTARY:

Digital displays of incubator temperature must be calibrated with a certified thermometer. Displays need to be checked and calibrated to a NIST-traceable thermometer.

RLM.10800 Phase II N/A YES NO

Does the laboratory have a method to detect and prevent incubator gas control failure?

COMMENTARY:

The laboratory must have a method to detect and prevent incubator gas control failure. The CO2 needs to be checked and recorded daily from the display and a fyrite test kit if the incubator is in use.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section V, A, 1.

RLM.10900 Phase II N/A YES NO

Does the laboratory's incubator for embryos have emergency backup power, and is it tested periodically?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 50 of 59

Page 51: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The laboratory's incubator for embryos must have a backup power supply. The backup power supply must be checked periodically.

REFERENCE: Revised minimum standards for in vitro fertilization, gamete intrafallopian transfer, and related procedures. Fertil Steril. 1998;70(suppl 2):1S-5S.

RLM.11000 Phase II N/A YES NO

Does the laboratory have a method to monitor and maintain adequate liquid nitrogen levels?

COMMENTARY:

The laboratory must have a method to monitor and maintain adequate nitrogen levels.

REFERENCE: Standards and technical manual reproductive cells and tissues. American Association of Tissue Banks, 1992: D.3000.

RLM.11100 Phase II N/A YES NO

Are all critical incubator, storage, refrigeration, and freezing units monitored and checked periodically?

COMMENTARY:

There must be evidence of monitoring of all critical incubator, storage, refrigeration, and freezing units. Alarm systems, if used, must be checked periodically.

REFERENCE: Revised minimum standards for in vitro fertilization, gamete intrafallopian transfer, and related procedures. Fertil Steril. 1998;70(suppl 2):1S-5S.

RLM.11200 Phase II N/A YES NO

Are the alarms monitored 24 hours/day (either remote or in the laboratory)?

COMMENTARY:

The alarms must be monitored 24 hours/day (either remote or in laboratory). Audible alarms are only effective if someone is able to respond to the difficulty and is trained to follow the appropriate methodology to correct the problem or take alternative measures.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section IV, A, 2, 10.

REPRODUCTIVE LABORATORY Page 51 of 59

Page 52: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.11300 Phase II N/A YES NO

Does the laboratory have a policy for implementing backup capability (refrigerators, freezers, incubators)?

COMMENTARY:

The laboratory must have a procedure for use of backup equipment available (refrigerators, freezers, incubators).

NOTE 1: If any unit begins to fail, a repair or replacement would probably not be able to be purchased and delivered soon enough to avoid loss of contents. It is therefore necessary to have an emergency procedure to provide backup units with adequate storage capacity to allow complete transfer of contents. The backup units must be periodically tested to ensure their functionality if needed.

NOTE 2: Documented procedures for use of backup equipment, location, and contact personnel must be part of the procedure manual.

*****************************************************************

EMBRYOLOGY PERSONNEL

*****************************************************************

-----------------------------------------------------------------

EMBRYOLOGY LABORATORY DIRECTOR

-----------------------------------------------------------------

**REVISED** 12/29/2004

RLM.11400 Phase II N/A YES NO

Does the director of the embryology laboratory have proper qualifications through education and experience to provide direction and administration of the laboratory?

NOTE: The director of the embryology laboratory should have 2 years of documented experience in a laboratory performing in vitro fertilization or assisted reproductive technologies-related procedures. Directors of embryology laboratories who are not physicians or qualified doctoral scientists, but who were functioning as embryology directors on or before July 20, 1999 are considered in compliance with the personnel requirements in the Laboratory General Checklist, as long as they meet all other requirements. Effective January 1, 2006, all new laboratory directors should hold HCLD (High Complexity Laboratory Director), ABB-ELD (American Board of Bioanalysis Embryology Laboratory

REPRODUCTIVE LABORATORY Page 52 of 59

Page 53: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

Director), or equivalent certification. Directors in laboratories performing testing covered by CLIA ’88 must meet CLIA laboratory director requirements for high complexity testing.

COMMENTARY:

N/A

REFERENCES: 1) United States general accounting office. December 1989, p 17, 7d; 2) Guidelines for human embryology and andrology laboratories Fertil Steril. 1992;58 (suppl1): section II, A, 1; 3) College of American Pathologists. Standards for accreditation, reproductive laboratory accreditation program. Northfield, IL: CAP, current edition.

#################################################################

CRYOPRESERVATION OF SPERM, OOCYTES, AND EMBRYOS

#################################################################

RLM.11500 Phase II N/A YES NO

Does the laboratory have a documented procedure(s) for cryopreservation of sperm, oocytes, and/or embryos?

COMMENTARY:

The laboratory must have a documented procedure for cryopreservation according to stage of development of the embryo or the type of gamete. The survival of human embryos requires the use of techniques specific to the stage of embryo development.

REFERENCES: 1) Fehilly CB, et al. cryopreservation of cleaving embryos and expanded blastocysts in the human. a comparative study. Fertil Steril. 1985;44:638; 2) Lasalle B, Testart J. Human embryo features that influence the success of cryopreservation with the use of 1,2 propanediol. Fertil Steril. 1985;44:645-651.

RLM.11600 Phase II N/A YES NO

Does the laboratory have a reliable method for labeling and tracking of cryopreserved specimens?

COMMENTARY:

The laboratory must have a reliable method for labeling and tracking cryopreserved specimens.

REFERENCE: American Association of Tissue Banks. Standards for tissue banking, 1997.

REPRODUCTIVE LABORATORY Page 53 of 59

Page 54: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.11700 Phase II N/A YES NO

Are records of all patient specimens, donor specimens, and patient/donor matches retained and easily accessible?

COMMENTARY:

The laboratory must maintain permanent records of gametes and embryos that are used for insemination or donated. There must be permanent records for each person inseminated with donor sperm or who receives embryos that resulted from donated oocytes or embryos, and be able to recover the identity of the donor(s).

REFERENCE: American Association of Tissue Banks. Standards for tissue banking, 1997.

RLM.11800 Phase II N/A YES NO

Are duplicate records maintained in a separate area from the originals?

COMMENTARY:

Duplicate records of cryopreserved specimens must be maintained in a separate area from the originals.

REFERENCE: Guidelines for human embryology and andrology laboratories. Fertil Steril. 1992;58 (suppl 1): section VI, A, 7.

RLM.11900 Phase II N/A YES NO

Are procedures adequate to ensure that cryopreserved patient specimens can be easily retrieved?

COMMENTARY:

There must be a mechanism by which patient specimens can easily be retrieved.

RLM.12000 Phase II N/A YES NO

Is the laboratory able to document the disposition and current inventory of all specimens that have been stored in its cryobanks?

COMMENTARY:

REPRODUCTIVE LABORATORY Page 54 of 59

Page 55: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

The laboratory must be able to report the disposition of all samples that have been stored in the bank, and the current remaining inventory of samples in the bank.

REFERENCE: American Association of Tissue Banks. Standards for tissue banking, 1997.

RLM.12100 Phase II N/A YES NO

Is there a policy to cover inventoried samples that cannot be located in the bank?

COMMENTARY:

There must be a policy that addresses the issue of samples that cannot be located. This may require a periodic physical inventory of the bank.

RLM.12200 Phase II N/A YES NO

Are procedures adequate to verify specimen identity and integrity throughout the entire process of cryopreservation?

COMMENTARY:

The identity of the specimen must be ensured through all steps of the procedure. There must be unique identifying numbers on all specimens.

REFERENCE: Department of Health and Human Services, Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988; final rule. Fed Register. 2003(Jan 24):7173 [42CFR493. 1232, 42CFR493.1242(a)(3)].

RLM.12300 Phase II N/A YES NO

Does the laboratory have a program to ensure that cryopreservation is capable of providing viable recovery rates?

COMMENTARY:

The laboratory must have a program to ensure that cryopreservation is capable of providing acceptable recovery rates of viable sperm, oocytes, and/or embryos.

REFERENCE: American Association of Tissue Banks. Standards for tissue banking, 1997.

REPRODUCTIVE LABORATORY Page 55 of 59

Page 56: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.12400 Phase II N/A YES NO

Is there a documented procedure to cover length of storage, informed consent and long-term disposition of cryopreserved gametes or embryos?

COMMENTARY:

There must be consent forms that cover the length of storage of gametes, embryos, or reproductive tissue, and their long-term disposition. Good practice also dictates that the consent form for all procedures is on file and readily available to the laboratory staff.

REFERENCES: 1) Trounson A, et al. Ultrarapid freezing: a new low-cost and effective method of embryo cryopreservation. Fertil Steril. 1987;48:843-50; 2) Harrison KL, et al. The optimal concentration of albumin as embryo cryoprotectant. J In Vitro Embryo Transfer. 1987;4:288-291; 3) American Association of Tissue Banks. Standards for tissue banking, 1997.

RLM.12500 Phase II N/A YES NO

If samples are transferred to another facility, is detailed information provided regarding sample identity, specimen quality, and thawing techniques for that specimen?

COMMENTARY:

There must be documented procedures for the release of any specimen and its transport to another facility. Detailed information must be provided to another facility when specimens are transferred to that facility. This information must include, but not be limited to, the number of specimens, the identity of the specimens, the quality of the specimen, any quality control data on freeze/thawing of the specimen and the appropriate thawing technique to use with each specimen.

*****************************************************************

PHYSICAL FACILITIES

*****************************************************************

RLM.12600 Phase I N/A YES NO

Is there adequate space for administrative functions?

COMMENTARY:

Additional space should be provided for administrative functions.

REPRODUCTIVE LABORATORY Page 56 of 59

Page 57: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.12700 Phase I N/A YES NO

Is there adequate space for clerical work?

COMMENTARY:

Additional space should be provided for clerical work.

RLM.12800 Phase I N/A YES NO

Is there adequate space for technical work (bench space)?

COMMENTARY:

Additional space should be provided for technical work (bench space).

RLM.12900 Phase I N/A YES NO

Is there adequate space for shelf storage?

COMMENTARY:

Additional space should be provided for shelf storage.

RLM.13000 Phase I N/A YES NO

Is there adequate refrigerator/freezer storage space?

COMMENTARY:

Additional space should be provided for refrigerated/frozen storage.

RLM.13100 Phase I N/A YES NO

Is the available space efficiently utilized?

COMMENTARY:

The existing space should be used more efficiently.

REPRODUCTIVE LABORATORY Page 57 of 59

Page 58: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

RLM.13200 Phase II N/A YES NO

Is the space available so there is no compromise of the quality of work, safety of personnel, or limitation of quality control activities?

COMMENTARY:

Existing space limitations were so severe as to interfere with the quality of work, the safety of personnel, and/or the ability of personnel to carry out adequate quality control procedures with appropriate documentation.

RLM.13300 Phase I N/A YES NO

Are floors and benches clean, free of clutter and well-maintained?

COMMENTARY:

Floors and benches should not be cluttered or otherwise deteriorated. Better cleaning and/or maintenance procedures should be implemented.

RLM.13400 Phase I N/A YES NO

Are water taps, sinks, and drains adequate?

COMMENTARY:

Water taps, sinks, and drains should be improved to support the types of procedures and workload of the laboratory.

RLM.13600 Phase I N/A YES NO

Are electrical outlets adequate?

COMMENTARY:

Electrical outlets should be improved to support the types of procedures and workload of the laboratory.

RLM.13700 Phase I N/A YES NO

Is ventilation adequate?

REPRODUCTIVE LABORATORY Page 58 of 59

Page 59: REPRODUCTIVE LABORATORY - University of … · Web viewREFERENCE: Keel BA, Webster BW. CRC handbook of the laboratory diagnosis and treatment of infertility. Boca Raton, FL: CRC Press,

College of American Pathologists Revised: 10/06/2005

COMMENTARY:

Ventilation should be improved to support the types of procedures and workload of the laboratory.

RLM.13800 Phase I N/A YES NO

Is lighting adequate?

NOTE: Direct sunlight should be avoided because of its extreme variability and the need for low light levels necessary to observe various computer consoles, etc. Lighting control should be sectionalized so general levels of illumination can be controlled in areas of the room if desired.

COMMENTARY:

Lighting should be adequate.

REFERENCE: College of American Pathologists. Medical laboratory planning and design. Northfield, IL: CAP, 1986.

RLM.13900 Phase I N/A YES NO

Is temperature/humidity control adequate?

COMMENTARY:

Temperature/humidity control should be improved to support the types of procedures and workload of the laboratory.

RLM.14000 Phase I N/A YES NO

Are telephones conveniently located, and are calls easily transferred?

COMMENTARY:

Telephones (number and/or location) should be improved to support the types of procedures and workload of the laboratory.

REPRODUCTIVE LABORATORY Page 59 of 59