Central Annals of Sports Medicine and Research Cite this article: Murai A, Sakaue Y, Makikawa M (2014) Evaluation of Lower-Leg Swelling using Electromyograms Obtained with Voltage-Dividing Electrodes. Ann Sports Med Res 1(2): 1007. *Corresponding author Masaaki Makikawa, College of Science and Engineering, Ritsumeikan University, 1-1-1Noji-Higashi, Kusatsu, Shiga 525-8577, Japan, Tel: +81-77-561-2881; Fax: +81-77-561-2665; Email: Submitted: 18 November 2014 Accepted: 05 December 2014 Published: 10 December 2014 Copyright © 2014 Makikawa et al. OPEN ACCESS Keywords Lower leg swelling Internal impedance Voltage divider EMG Research Article Evaluation of Lower- Leg Swelling using Electromyograms Obtained with Voltage-Dividing Electrodes Akito Murai 1 , Yusuke Sakaue 1 and Masaaki Makikawa 2 * 1 Graduate School of Science and Engineering, Ritsumeikan University, Japan 2 College of Science and Engineering, Ritsumeikan University, Japan Abstract Lower leg swelling is caused by an increase in extracellularfluid in the lowerleg. Lower leg swelling is uncomfortable for the patient and increases the risk of blood flow abnormalities. Compression stockings, leg exercise, and a specific type of flooring material have been used to attempt to prevent lower leg swelling. To evaluate lower leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG) obtained with voltage dividing electrodes. Ten healthy men (23.3±2.1 years old) participated in this study. They were asked to work at a desk for 12 hours. The following measurements were obtained: EMG readings (obtained using voltage dividing electrodes), calf circumference, and lower leg impedance (obtained using an impedance meter). These measurements were obtained before, during and after the desk work. The results showed that the internal impedance decreased as the calf circumference increased. The reduction ratio of the internal impedance, as determined by EMG performed with voltage dividing electrodes, was significantly greater than the internal impedance measured using an impedance meter (P<0.05). Our new measurement method was thus proven to be effective in evaluation of lower leg swelling. ABBREVIATIONS EMG: Electromyogram; BIA: Bioelectrical Impedance Analysis INTRODUCTION In this paper, a new method for evaluation of lower leg swelling using electromyograms (EMG) obtained with voltage dividing electrodes is introduced [1]. Lower leg swelling occurs as a result of an increase in extracellular fluid in the lower leg. In daily life, extracellular fluid circulates around the body by blood and lymph flow, and both types of fluid circulation are driven by muscle and heart pumping activities. Are duction in these pumping activities causes poor circulation of extracellular fluid and an increase in extracellular fluid in the lower leg. Lower leg swelling is painful and uncomfortable and increases the risk of blood flow abnormalities. Compression stockings, leg exercises, and specific types of flooring materials have been used to attempt to prevent lower leg swelling [2-4]. Typically, lower-leg swelling is evaluated by measuring the calf circumference using a tape measure [5]. Lower leg swelling is easily measured by this method, but large measurement errors may result. Bioelectrical impedance analysis (BIA) is also used to evaluate lower leg swelling [6]. In BIA, the reduction in bioelectrical impedance associated with increased extracellular fluid is measured. To calculate the bioelectrical impedance, an alternating current (AC) of more than 100 kHz is applied to the human body [7]. There as on that low-frequency AC is not used in BIA is that low-frequency current poses risks of electrical shock and pacemaker malfunction. We have developed a new signal source estimation method that uses voltage divider technology [8]. Using this method, biological signals are measured with and without additional resistance between the signal electrode and the ground electrode, and the signal source potential is attenuated by a voltage divider consisting of internal impedance and additional resistance. The signal source position is estimated from the attenuation level of the signal source potential. In our study, we attempted to estimate the internal impedance using surface EMGs of lower extremities measured with and without use of a voltage divider. The theory and details of our measurement method are described in the next section. Our method employs surface EMG measurement, which poses no risk of electrical shock or pacemaker malfunction,

Upload: others

Post on 18-Jan-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central Annals of Sports Medicine and Research

Cite this article: Murai A, Sakaue Y, Makikawa M (2014) Evaluation of Lower-Leg Swelling using Electromyograms Obtained with Voltage-Dividing Electrodes. Ann Sports Med Res 1(2): 1007.

*Corresponding authorMasaaki Makikawa, College of Science and Engineering, Ritsumeikan University, 1-1-1Noji-Higashi, Kusatsu, Shiga 525-8577, Japan, Tel: +81-77-561-2881; Fax: +81-77-561-2665; Email:

Submitted: 18 November 2014

Accepted: 05 December 2014

Published: 10 December 2014

Copyright© 2014 Makikawa et al.

OPEN ACCESS

Keywords•Lower leg swelling•Internal impedance•Voltage divider•EMG

Research Article

Evaluation of Lower-Leg Swelling using Electromyograms Obtained with Voltage-Dividing ElectrodesAkito Murai1, Yusuke Sakaue1 and Masaaki Makikawa2*1Graduate School of Science and Engineering, Ritsumeikan University, Japan2College of Science and Engineering, Ritsumeikan University, Japan

Abstract

Lower leg swelling is caused by an increase in extracellularfluid in the lowerleg. Lower leg swelling is uncomfortable for the patient and increases the risk of blood flow abnormalities. Compression stockings, leg exercise, and a specific type of flooring material have been used to attempt to prevent lower leg swelling. To evaluate lower leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG) obtained with voltage dividing electrodes. Ten healthy men (23.3±2.1 years old) participated in this study. They were asked to work at a desk for 12 hours. The following measurements were obtained: EMG readings (obtained using voltage dividing electrodes), calf circumference, and lower leg impedance (obtained using an impedance meter). These measurements were obtained before, during and after the desk work. The results showed that the internal impedance decreased as the calf circumference increased. The reduction ratio of the internal impedance, as determined by EMG performed with voltage dividing electrodes, was significantly greater than the internal impedance measured using an impedance meter (P<0.05). Our new measurement method was thus proven to be effective in evaluation of lower leg swelling.

ABBREVIATIONSEMG: Electromyogram; BIA: Bioelectrical Impedance Analysis

INTRODUCTIONIn this paper, a new method for evaluation of lower leg

swelling using electromyograms (EMG) obtained with voltage dividing electrodes is introduced [1]. Lower leg swelling occurs as a result of an increase in extracellular fluid in the lower leg. In daily life, extracellular fluid circulates around the body by blood and lymph flow, and both types of fluid circulation are driven by muscle and heart pumping activities. Are duction in these pumping activities causes poor circulation of extracellular fluid and an increase in extracellular fluid in the lower leg. Lower leg swelling is painful and uncomfortable and increases the risk of blood flow abnormalities. Compression stockings, leg exercises, and specific types of flooring materials have been used to attempt to prevent lower leg swelling [2-4].

Typically, lower-leg swelling is evaluated by measuring the calf circumference using a tape measure [5]. Lower leg swelling is easily measured by this method, but large measurement errors may result. Bioelectrical impedance analysis (BIA) is also

used to evaluate lower leg swelling [6]. In BIA, the reduction in bioelectrical impedance associated with increased extracellular fluid is measured. To calculate the bioelectrical impedance, an alternating current (AC) of more than 100 kHz is applied to the human body [7]. There as on that low-frequency AC is not used in BIA is that low-frequency current poses risks of electrical shock and pacemaker malfunction.

We have developed a new signal source estimation method that uses voltage divider technology [8]. Using this method, biological signals are measured with and without additional resistance between the signal electrode and the ground electrode, and the signal source potential is attenuated by a voltage divider consisting of internal impedance and additional resistance. The signal source position is estimated from the attenuation level of the signal source potential. In our study, we attempted to estimate the internal impedance using surface EMGs of lower extremities measured with and without use of a voltage divider. The theory and details of our measurement method are described in the next section.

Our method employs surface EMG measurement, which poses no risk of electrical shock or pacemaker malfunction,

Page 2: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central

Makikawa et al. (2014)Email:

Ann Sports Med Res 1(2): 1007 (2014) 2/6

unlike BIA. The frequency band of the surface EMG signal passing through the human body is lower than the frequency band of the AC used in BIA. Low frequency AC passes only through only the extracellular fluid [9]. We therefore expected our method to be able to evaluate the change in internal impedance due to an increase in extracellular fluid with high sensitivity.

The purpose of this study was to evaluate lower leg swelling using EMGs measured using voltage dividing electrodes. The amplitude of an EMG measured without a voltage divider was compared to that measured with a voltage divider to calculate the bioelectrical impedance of the lower leg. Experiments were carried out to measure changes in the bioelectrical impedance of the lower leg before and after lower leg swelling.

MATERIALS AND METHODSTheory

A model of the lower leg is shown in Figure1A. To measure biosignals using surface electrodes, the human body can be replaced by an electrical circuit comprising a signal source, and the conductivity within the human body can be assumed to be uniform for simplicity [10]. Using the venin’s theorem, the human body can be replaced by a signal source and synthetic impedance between the signal source and electrodes [11,12]. The model for our measurement system is shown in Figure 1B, the measurement circuit is shown in Figure 1C. The muscle is replaced by a signal source with potential VS. In this study, skin surface EMGs were measured using the bipolar-lead method. Soft tissues inside the body are replaced by three synthetic impedances, Zb1, Zb2, and Zg, one to each electrode. The additional resistance RE is set to be positive or negative using a mechanical switch. The voltage divider consists of the additional resistance RE and the two internal impedances Zb1 and Zb2, and the signal source potential is attenuated at each electrode.

Because the input resistance Rin of the amplifier is very large (10TΩ), the signal source potential is not attenuated when the additional resistance is switched off. Eq.(1) expresses the output potential Vout when the additional resistance is switched off. Eq.(2) expresses the output potential V’out when the additional resistance is switched on, and the signal source potential is attenuated. The attenuation level of the signal source depends on the additional resistance and internal impedance, i.e., the soft body tissue. Eq.(3) is used to calculate the total value of the two internal impedances obtained from eqs.(1) and (2). In this study, the total value of the two the internal impedances, Zb, is used to evaluate lower-leg swelling.

.Sout VV ≅ (1)

.21

'S

Ebb

Eout V

RZZR

V++

= (2)

.'

'

21 Eout

outoutbbb R

VVV

ZZZ−

==+ (3)

Subjects

Ten healthy adult men (23.3±2.1years old) with no history of lower extremity problems participated in this study. The subjects were asked to work at a desk for 12 hours to induce extracellular retention in the lower leg, and the Zb of each subject was measured before, during, and after the desk work, using the procedure described in the next section. During the 12 hours of desk work, the subjects were not allowed to move from the desk except when necessary to eat lunch and use the washroom. The subjects were not allowed to wear such things as stockings, long socks, or tights that are considered to assist in blood flow and reduce lower leg swelling. They were asked to sleep for at least

(A) (B)

Zb1 +

-VS

Vout

Ground

RE

RinZb2

Lower leg

VS : Potential of signal sourceZb1 , Zb2 , Zg: Impedance between signal source and electrode

including skin contact impedanceRE : Additional resistanceRin : Input resistance of differential amplifier

Differentialamplifier

Signalelectrodes

Zg Groundelectrodes

Electrodes

V V

Z

Lower legMuscle

(C)

+

-

Vout

R E

R in +

TL071

-1 MΩ

1 kΩ

AD620

10 μF

1.6 kΩ

1,3,5,9,15 kΩ

Figure 1 (A) –Model of lower leg. (B) –Model of our measurement system. (C) –Measurement circuit of this study.

Page 3: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central

Makikawa et al. (2014)Email:

Ann Sports Med Res 1(2): 1007 (2014) 3/6

seven hours on the day before the experiment and not to drink alcohol. The experiments were conducted in accordance with the ethical principles of the Helsinki Declaration after obtaining informed consent from the subjects.

Measurement

EMG measurements performed using voltage- dividing electrodes were carried out before, during, and after the desk work. The target muscle of the surface EMG was the lateral gastrocnemius of the right foot. Figure 2 shows the positioning of the signal electrodes (e1 and e2) and the ground electrode (e3). A pair of signal electrodes was set on the muscle belly of the lateral gastrocnemius. The distance between the two signal electrodes was 25 mm. The ground electrode was set on the right tibia at the same height as the signal electrodes. Disposable electrodes (SP-00-S, Mets Co., Ltd.) were used in the experiments. Before the electrodes were attached, sebum and dirt were removed using alcohol swabs and wipes.

The surface EMG was measured for 60s as subjects at with his calf raised and his ankle joint held at a 30 degree angle. Figure 3 shows a photograph of the experiment. The subjects were asked to practice this experimental posture before the measurement for the purpose of EMG amplitude stabilization. Five additional resistances (1, 3, 5, 9, and 15 kΩ) were set between positive and negative inputs using a mechanical switch to confirm that the change in attenuation depended on the additional resistance. The

experimenter switched the five additional resistances on and off every10s. The total gain of the amplifier with a high pass filter of 10 Hz, shown in Figure 1C, was 84 dB. The sampling frequency was 1,000 Hz.

Before every EMG measurement, the calf circumference was measured using a tape measure and the bioelectrical impedance between electrode e1 and e2 (ZRef) were measured using impedance meter (35-3250, Hioki Co., Ltd.). The calf circumference was measured while a subject stood and relaxed his lower leg. The measurement point was marked to ensure that the same point on the calf of each subject was measured. ZRef was measured over a period of 10s using an impedance meter, with each subject assuming the same posture as in the calf circumference measurement. The sampling frequency was 1Hz, and the voltage applied ranged from 0.01V to a limit corresponding to a current less than 10 μA at which the subjects did not feel pain or numbness. The AC frequency between e1 and e3 was 50 kHz, as is typical in BIA to compare the impedance at low frequency with high frequency.

Data analysis

The EMG data for 60s intervals were divided into six sets of 10 s data. Next, data for between 2 and 8s were extracted, and artifacts contaminating the measurement data were eliminated using a digital band-pass filter of 15-500 Hz. For each set of data filtered from 15 to 500 Hz, the root mean square (RMS) was calculated. The RMS calculated from the EMG without additional resistance was denoted Vout, and that with additional resistance was denoted V’out. Five Zb values were obtained for each subject using eq. (3). In this study, the average of three Zb values for each subject, excluding the large stand smallest value, was defined as the Zb to take into account EMG amplitude variation. Zb values were normalized, and change ratios were calculated by comparison with the values obtained from measurements made before the desk work began.

The average ZRef over the measurement period of 10s was calculated. Both the calf circumference and bioelectrical impedance values were normalized, and change ratios were calculated by comparison with the values obtained from measurements made before desk work began.

RESULTS AND DISCUSSIONResults

Figure 4 shows an example of an EMG wave form measured before the desk work. The EMG amplitude was attenuated by the additional resistances applied in comparison to the EMG amplitude without additional resistance. The attenuation observed when the lowest additional resistance (1 kΩ) was applied was the largest of the attenuations observed for the five additional resistances. The attenuation decreased with each increase in the additional resistance.

Table 1 shows the all subjects of Zb, ZRef, and the calf circumference. For all of the subjects, the calf circumference was greater during and after the desk work than before the desk work. The calf circumference increased 0.2±0.1 cm on average from before to during the desk work and 0.3±0.2 cm on average from before to after the desk work.

Figure 2 Positioning of electrodes. (A)–Back view. (B)–Lateral view.

Figure 3 A photograph of the experiment.

Page 4: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central

Makikawa et al. (2014)Email:

Ann Sports Med Res 1(2): 1007 (2014) 4/6

Figure 4 Example of electromyogram (EMG) wave form (no= no resistance).

Figure 5 Change ratios of the total value of internal impedance (Zb) and the bioelectrical impedance measured using an impedance meter (ZRef).

subjectsZb[kΩ] ZRef[Ω] Calf circumference [cm]

Before During After Before During After Before During After

1 27.8 18.9 17.5 485.6 494.5 479.7 36.2 36.3 36.4

2 7.8 8.8 8.7 450.0 420.0 445.0 35.9 36.4 36.2

3 12.8 10.7 11.3 455.0 426.8 441.4 32.4 32.5 32.8

4 12.9 11.7 13.0 368.4 367.2 361.3 38.0 38.3 38.3

5 6.6 4.6 4.7 431.2 368.2 366.3 40.2 40.5 40.9

6 10.7 6.1 7.8 444.5 433.4 425.2 37.2 37.3 37.3

7 9.2 4.8 1.3 440.0 430.0 427.0 34.6 34.8 35.0

8 11.1 9.2 8.3 450.8 396.9 446.9 36.4 36.5 36.6

9 15.1 13.3 11.0 573.5 519.5 499.8 35.6 35.8 36.0

10 11.3 10.4 9.0 536.6 455.6 424.0 38.6 38.8 38.9

Average 12.4 9.3 9.1 440.7 417.1 424.1 36.4 36.6 36.7

S.D. 5.6 4.1 4.2 54.2 46.7 41.0 2.1 2.1 2.1

Table 1: All subjects of internal impedance (Zb), the bioelectrical impedance measured using an impedance meter (ZRef), and calf circumference (n=10).

Abbreviations: SD: Standard Deviation

Page 5: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central

Makikawa et al. (2014)Email:

Ann Sports Med Res 1(2): 1007 (2014) 5/6

Figure 5 shows the change ratios of Zb and ZRef. The Zb values were lower during and after the desk work than before the desk work for all but one subject. The average Zb decreased by 3.1±2.6 kΩ (-20.5%) from before to during the desk work and decreased by 3.3±3.2 (-25.1%) from before to after the desk work. For all of the subjects, ZRef was also lower during and after the desk work than before the desk work. The average ZRef decreased by 23.6±7.5 Ω (-6.7%) from before to during the desk work and decreased by 16.6±13.2 Ω (-6.4%) from before to after the desk work. The change ratio of Zb during versus after the desk work was significantly greater than that of ZRef (P<0.05).

Discussion

Long-term standing and sitting induce are duction in muscle pump activity and thus contribute to lower-leg swelling [6]. As shown in Table 1, calf circumference was found to increase with long-term deskwork. In this study, male subjects were asked to work at a desk to measure the change in internal impedance with mild lower-leg swelling. Calf circumference was found to increase by 0.3 cm on average.

The increases observed in calf circumference and volumes were brought about by pooling of extracellular fluid in the lower leg. As shown in Figure 5, Zb and ZRef measured using an impedance meter both decreased as the calf circumference increased. This finding suggests that long-term desk work induces pooling of extracellular fluid and decreased bioelectrical impedance in the lower leg. However, there were two subjects whose Zb did not decrease as a result of a lengthy period of desk work. There as on for this is that the subject had an athletic history: his heavy muscle mass inhibited pooling of extracellular fluid.

The impedance measured by conventional bioelectrical impedance method differs according to the frequency of AC current passing through the human body, and it is said that low frequency current should be used to measure the impedance of extracellular fluid [13]. The body tissue can be replaced by three electrical components that are the resistance of intracellular and extracellular fluid, and the capacitance of cell membrane [9]. The Zb measured at frequency-band of EMG signal mainly indicates the resistance of extracellular fluid, because low frequency current to intracellular fluid blocked by the capacitance of cell membrane. At high frequency, the impedance of cell membrane is reduced, and current can pass through both intracellular and extracellular fluid. Therefore, the change ratio of Zb may be significantly greater than that of ZRef at the current of 50 kHz, when using a conventional impedance meter. Previous study reported that the lower leg impedance decreased about 7% from morning to evening [14]. These results suggest that our measurement system using EMG signals can be used to measure changes in lower-leg impedance with higher sensitivity than a measurement system that uses high-frequency alternating current.

CONCLUSIONIn this study, we measured the internal lower-leg impedance

between a signal source and signal electrodes using EMG performed with voltage-dividing electrodes. Subjects were asked to work at a desk for a long period of time to induce changes in the internal impedance of the lower leg. The internal impedance

was found to decrease with increasing calf circumference, and our results are coincident with previous studies [1,5,6,13,14]. Low frequency AC current should be used to evaluate the lower leg swelling [14]. However, the low frequency AC current has risk of electrical shock e.g. pacemaker malfunction in BIA. In addition, EMG signal may be contaminated as noise in previous bioelectrical impedance methods. Therefore, this study indicates that our measurement system can be used to measure the change in internal impedance of the lower leg with high sensitivity and that EMG measurements performed with voltage-dividing electrodes can be used to evaluate lower-leg swelling and methods to reduce and prevent it.

ACKNOWLEDGEMENTSThis work was supported by JSPS Kakenhi Grant Number

25282137. We would like to thank Ms. Chika Soukawa and personnel with the R&D Division of Okamoto Co., Ltd., Nara, Japan, who provided technical advice.

REFERENCES1. Murai A, Sakaue Y, Makikawa M. Evaluation of lower-leg swelling

using EMG measured with voltage divider. Proceeding of 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC2014). 2014; 3751-3754.

2. Lin YH, Chen CY, Cho MH. Effectiveness of leg movement in reducing leg swelling and discomfort in lower extremities. Appl Ergon. 2012; 43: 1033-1037.

3. Mosti G, Partsch H. Occupational leg oedema is more reduced by antigraduated than by graduated stockings. Eur J Vasc Endovasc Surg. 2013; 45: 523-527.

4. Zander JE, King PM, Ezenwa BN. Influence of flooring conditions on lower-leg volume following prolonged standing. Int J Ind Ergon. 2004; 45: 279-288.

5. Kawano T, Nishida S, Hashimoto M. Development of measuring device for lower-leg swelling using strain gauge. JSME Int J Ser C. 2005; 48: 592-597.

6. Chester MR, Rys MJ, Konz SA. Leg swelling, comfort, and fatigue when sitting, standing, and sit/standing. Int J Ind Ergon. 2002; 29: 289-296.

7. Bedogni G, Malavolti M, Severi S, Poli M, Mussi C, Fantuzzi AL, et al. Accuracy of an eight-point tactile-electrode impedance method in the assessment of total body water. Eur J Clin Nutr. 2002; 56: 1143-1148.

8. Sakaue Y, Okada S, Makikawa M. Signal source estimation inside brain using switching voltage divider. Proceedings of XIII Mediterranean Conference on Medical and Biological Engineering and Computing 2013 IFMBE. 2013; 595-598.

9. Kanai H, Sakamoto K, Haeno M. Electrical measurement of fluid distribution in human legs: estimation of extra- and intra-cellular fluid volume. J Microw Power. 1983; 18: 233-243.

10. Musha T, Homma S. Do optimal dipoles obtained by the dipole tracing method always suggest true source locations? Brain Topogr. 1990; 3: 143-150.

11. Akamatsu N, Mori H. Body surface multi-electrode for inverse problem in electro cardiography and its Application. IEICE Trans Electron. 67: 104-111.

12. Sakaue Y, Okada S, Makikawa M. Development of brain electrical activity estimation device using resistance voltage divider. Proceedings of International Conference on Health care 2012. 2012; 172-174.

Page 6: Research Article Evaluation of Lower- Leg Swelling using...leg swelling, we attempted to estimate the change in internal impedance due to lower leg swelling using electromyograms (EMG)

Central

Makikawa et al. (2014)Email:

Ann Sports Med Res 1(2): 1007 (2014) 6/6

Murai A, Sakaue Y, Makikawa M (2014) Evaluation of Lower-Leg Swelling using Electromyograms Obtained with Voltage-Dividing Electrodes. Ann Sports Med Res 1(2): 1007.

Cite this article

13. Seo A, Rys M, Konz S. Measuring lower leg swelling: optimum frequency for impedance method. Med Biol Eng Comput. 2001; 39: 185-189.

14. Seo A, Yuji K, Fumitaka Y. A portable apparatus for monitoring leg swelling by bioelectrical impedance measurement. J Occup Health. 1997; 39: 150-151.