rhic polarimetery in run9, sqrt (s)=200 gev

19
RHIC Polarimetery in Run9, sqrt(s)=200 GeV A.Bazilevsky for RHIC Polarimetry group RHIC Spin Collaboration Metting May 15 (Friday), 2009

Upload: salali

Post on 27-Jan-2016

26 views

Category:

Documents


1 download

DESCRIPTION

RHIC Polarimetery in Run9, sqrt (s)=200 GeV. A.Bazilevsky for RHIC Polarimetry group RHIC Spin Collaboration Metting May 15 (Friday), 2009. pC Rate history. s=200 GeV. Target changed in Yellow1 to thicker one (from Fill 10686)  Rate problems! - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

RHIC Polarimetery in Run9, sqrt(s)=200 GeV

A.Bazilevsky for RHIC Polarimetry group

RHIC Spin Collaboration MettingMay 15 (Friday), 2009

Page 2: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC Rate historys=200 GeV

Target changed in Yellow1 to thicker one (from Fill 10686) Rate problems!

Target changed in both Blues from fill 10732 Blue2 rates slightly increased

Page 3: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC measurements

Fills 10616 (Apr 18) – 10749 (May 15)

“Online” polarizations: 0.50-0.60

Pol-1 measure slightly lower than Pol-2: by ~6%

Blue1/Blue2: consistent within stat. uncertainties

Yell1/Yell2: shows variations above stat. uncertainties

Online Polarization (%), not normalized (!) vs fill

Page 4: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC-Blue vs HJet

Hjet/pC is stable over fills within (large) stat. errors (of HJet)

HJet: <P>=55%(fills 10616-10732)

HJet/Blue1 1.05

HJet/Blue2 0.99

Page 5: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC-Yellow vs HJet

Hjet/pC is stable over fills within (large) stat. errors (of HJet)

HJet: <P>=55%(fills 10616-10732)

HJet/Yell1 1.07

HJet/Yell2 1.00

Page 6: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

More precise Hjet-pC comparison

Hjet: fills combined in 9 periodsClear correlation between Hjet and pCConsistency vs period within 5%

pC-blueHJet

pC-yellowHJet

Page 7: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: Pol. Profile 2

2

P

IR

Polarimeters 2Polarimeters 1

Usual…R0.15 in previous years (100 GeV beams)R0.1 Experiments see 5% more polarization than Hjet

Horizontal profile

Vertical profile

Vertical profile

Vertical profile

Horizontal profile Horizontal profile

Page 8: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

Backups

Page 9: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

Summary

HJet: Running in stable conditions: <P>~55%

pC: Blue1 vs Blue2 consistent behavior

Yell1 vs Yell2 show systematic effects ~5-7% (may be due to larger rate effects in Yell1)

pC vs Hjet: consistent within stat. errorsHjet/Pol1 ~ 1.06; Hjet/Pol2 ~ 1

Might be ~5% drop in the pC measurements from the beginning to the end of the run due to detector degradation (“dead layer” increase)

Measurements are statistically consistent within a fill

Polarization decay Tdecay ~ 100-200 hours

Polarization profile no sharper than in previous yearsExperiments see ~5% more polarization than Hjet

Measurements at injection and flattop are consistent within ~2%

Page 10: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

Rate historys=500 GeV

Page 11: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

C Mass

Page 12: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: Polarization Profile

pC Scan C target over the beam cross:

Target Position

Inte

nsity

Pola

rizati

on

2

2

P

IR

I

P

2. Obtain R directly from the P(I) fit:

2

2

max 2exp)(

P

xPxP

2

2

max 2exp)(

I

xIxI

R

L

LPP

maxmax

P

I

Precise target positioning is NOT necessary

1. Directly measure I and P :

R=0.290.07

Page 13: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: Consistency within a Fill

Page 14: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: Consistency within a Fill

Clear polarization decay

Consistent between Pol1 and Pol2

10685-Blue10704-Blue

Page 15: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

Statistically Ok !

Page 16: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: Consistency within a Fill

Should be uniform if variations within a fill are only due to stat. errors

Should show higher density near 0 if there are “sizable” syst. effects

Prob(2, NDF) – from the fit to a constant in a fill

No systematic effects (comparable to stat. errors) are seen within a fill

Page 17: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

Pol. Decay

<Tdecay> = 80-180 hours

<Tdecay> = 250-800 hours

May need rate correction!

Run6:

<Tdecay> ~150 hours

Run8:

<Tdecay> ~400 hours

<Tdecay> ~100 hours

In a fill: fit to exp(-t/Tdecay)

Page 18: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: pol. at injection

Pol1 measures smaller values than Pol2 by ~5%(similar to flattop, or slightly smaller due to smaller rate effect at injection)

<P>~56%(similar to flattop)

Page 19: RHIC  Polarimetery  in Run9,  sqrt (s)=200  GeV

pC: injection vs flattop

On the average no difference within 2%:

Assuming that we don’t lose polarization on the ramp AN(inj)/ AN(ftp) is known within 2% Correction due to rate effect and polarization decay at store may be needed

Assuming that AN(inj)/ AN(ftp) is correct Polarization loss on the ramp <2%

(-2.11.2)%

(-0.91.1)%

(-1.21.0)%

(3.52.5)%