s8110croft
TRANSCRIPT
John Michael Croft
Part 3:
4) Figure 4.1 below shows an NP Chart for number of defects each day. Day 1-20, sample sizes were 30. Subsequently, new equipment was installed and the sample sizes were increased to 40 per day.
a. No out of control signals or run rule violations appear below. The process appears to be in statistical control to monitor performance going forward.
37332925211713951
10
8
6
4
2
0
Sample
Sam
ple
Cou
nt
__NP=4.4
UCL=10.34
LCL=0
Figure 4.1: NP Chart for Equipment Final Inspections
Tests performed with unequal sample sizes
b. Figure 4.2 below shows a standardized Z-chart for the above for comparison purposes. Again, no OOC signals or run rule violations occur.
37332925211713951
3.0
1.5
0.0
-1.5
-3.0
Observation
Stan
dard
ized
Dat
a
0
UCL=3
LCL=-3
30 40
37332925211713951
4
3
2
1
0
Observation
Mov
ing
Ran
ge
__MR=1.128
UCL=3.686
LCL=0
30 40
Figure 4.2: Standardized Z-MR Chart
File:document.docx Page 1 of 6 John Michael Croft
John Michael Croft
c. Figure 4.3 below shows an NP Chart by stages (old and new equipment). Again no OOC signals or run rule violations occur. Notice the newer equipment has a wider variance due to lack of use suggesting it may need to be monitored and calibrated to reduce the variance of defects per day.
37332925211713951
12
10
8
6
4
2
0
Sample
Sam
ple
Cou
nt
__NP=5.45
UCL=11.96
LCL=0
Old New
Figure 4.3: NP Chart by Stage
5) Below are a series of evaluations on the X5 variable assuming the first 30 subgroups as a baseline with very good control of the process.
a. Figure 5.1 below shows an IMR chart of the first 30 subgroups to estimate the mean and standard deviation.
28252219161310741
70
65
60
55
50
Observation
Indi
vidu
al V
alue
_X=60.21
UCL=72.62
LCL=47.81
28252219161310741
16
12
8
4
0
Observation
Mov
ing
Ran
ge
__MR=4.66
UCL=15.24
LCL=0
Figure 5.1: IM-R of Subgroups 1-30
File:document.docx Page 2 of 6 John Michael Croft
John Michael Croft
Process Mean=60.21
ProcessStandard Deviation= 4.661.128
=4.13
b. Figure 5.2 below attempts to monitor the remaining subgroups based on the parameter estimates from the first 30 subgroups using an IMR chart. Notice several OOC and run rule violations in the MR chart that need to be addressed prior to making meaningful decisions based on the I chart. However, the I chart displays one OOC and 2 run rule violations.
554943373125191371
80
70
60
50
Observation
Indi
vidu
al V
alue
_X=60.21
UCL=72.6
LCL=47.82
554943373125191371
20
15
10
5
0
Observation
Mov
ing
Ran
ge
__MR=4.66
UCL=15.22
LCL=0
5
5
1
11
2
2
1
1
Figure 5.2: IM-R all Subgroups
c. Figure 5.3 simulates the above again, but with a EWMA chart. Again, notice several OOC points beyond the UCL suggesting the process is unstable and needs to be calibrated to remove the unwanted variability.
File:document.docx Page 3 of 6 John Michael Croft
John Michael Croft
554943373125191371
64
63
62
61
60
59
58
57
Sample
EWM
A
__X=60.21
UCL=63.052
LCL=57.368
Figure 5.3: EWMA Chart
d. Figure 5.4 below simulates the above again but with a CUSUM Chart. Again we notice several OOC signals toward the end of the process suggesting the need to refine and reduce variability.
554943373125191371
40
30
20
10
0
-10
-20
Sample
Cum
ulat
ive
Sum
0
UCL=20.65
LCL=-20.65
Figure 5.4 CUSUM Chart
e. To summarize the above results:
File:document.docx Page 4 of 6 John Michael Croft
John Michael Croft
i. Subgroups 1-30 represent a stab process with a mean, 60.21 and standard deviation, 4.13.
ii. Figures 5.2 – 5.4 all suggest assignable cause variation or possible mean shifts while monitoring subgroups 31-60.
iii. Figure 5.2 suggest an OOC signal at subgroup 46 and run rule violations at 48 and 50.
iv. Figure 5.3 show several OOC signals within the cluster of subgroups 45 – 54 suggesting the process lost control between 31 and 43.
v. Figure 5.4 show several OOC signals from 46 – 57. vi. Recommend eliminating assignable cause variation and closely
monitoring to reduce variability within the process. 6) Below are several charts evaluating two parameters, separately and together, monitoring
process performance in a multivariate setting. a. Figures 6.1 and 6.2 below show X6A and X6B to be independently in statistical
control with no OOC signals or run rule violations.
28252219161310741
40
35
30
25
20
Observation
Indi
vidu
al V
alue
_X=29.65
UCL=42.38
LCL=16.93
28252219161310741
16
12
8
4
0
Observation
Mov
ing
Ran
ge
__MR=4.78
UCL=15.63
LCL=0
Figure 6.1: IM-R X6A
File:document.docx Page 5 of 6 John Michael Croft
John Michael Croft
28252219161310741
25.0
22.5
20.0
17.5
15.0
Observation
Indi
vidu
al V
alue
_X=20.16
UCL=26.50
LCL=13.82
28252219161310741
8
6
4
2
0
Observation
Mov
ing
Ran
ge
__MR=2.384
UCL=7.790
LCL=0
Figure 6.2: IM-R X6B
b. Figure 6.3 performs a multivariate T^2 Chart considering both variables together. Subgroup 30 appears to be OOC as it exceeds the UCL on the T^2n chart. This appears consistent with the spike in the Generalized Variance.
28252219161310741
20
15
10
5
0
Sample
Tsqu
ared
Median=2.25
UCL=14.31
28252219161310741
1.6
1.2
0.8
0.4
0.0
Sample
Gen
eral
ized
Var
ianc
e
|S|=0.442
UCL=1.443
LCL=0
Figure 6.3: Multivariate Chart
c. Considering the interaction of the two effect together allows for OOC signals to be detected where univariate charts might not.
File:document.docx Page 6 of 6 John Michael Croft