sample conditioning - iberfluid · the sentry dccp reboiler is a complete panel that measures...

2
The Sentry® DCCP Reboiler is designed to optimize the analytical results for degassed cation conductivity. First, the panel uses an oversized heater and baffled vapor discharge chamber, which assures complete boiling and provides a steam blanket over the sample to eliminate any sample contamination that might occur due to discharge piping backflow. The vapor is then condensed and removed via a vent drain. After the sample is degassed, it is cooled with an additional sample cooler so that chemistry deviations in analyzer temperature compensation curves are reduced. MODEL DCCP BENEFITS The Sentry DCCP reboiler is a complete panel that measures specific cation and degassed conductivity or any individual measurement needed. The custom analysis panels include degassed conductivity, dissolved oxygen (DO) and corrosion products sampling. The panel is designed to work with a wide variety of conductivity analyzers and provides simple and compact cycle chemistry validation. FEATURES Capable of measuring specific conduc- tivity (SC), cation conductivity (CC) and degassed cation conductivity (DCC) Regenerative cooling option (no cooling water) Operates in accordance with ASTM D4519 Sample cooler ensures CC and DCC probes receive identical sample temperatures, eliminating errors Compact wall-mount design Surface mount, backplate option CE, SGS and CSA available DCC CC RC HTR-1 SC-1 SC 2 1 CIT 4 3 3 CIT FICV SC 1 CIT CC 1 6 3 CE 3 CE 3 CE 1 TIC 1 TE (1) The cooled sample flows through the first conductivity cell which measures specific conductivity (SC). (2) The cation resin bed exchanges positive ions in the sample for hydrogen ions. (3) The sample flows through the second conductivity cell which measures for cation conductivity (CC). (4) The digitally operated heater removes volatiles, such as CO 2 , in the sample. Turns off automatically on low sample flow. (5) The sample cooler returns the sample temperature to inlet conditions. (6) The sample flows through the third conductivity cell which measures the conductivity of the degassed sample (DCC). Operation DCCP Reboiler with analyzer panel SAMPLE CONDITIONING SENTRY DCCP REBOILER Single Line Conditioning Modules Gas Liquid & Slurry Solid & Powder Steam & Water

Upload: others

Post on 30-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: SAMPLE CONDITIONING - Iberfluid · The Sentry DCCP reboiler is a complete panel that measures specific cation and degassed conductivity or any individual measurement needed. The custom

The Sentry® DCCP Reboiler is designed to optimize the analytical results for degassed cation conductivity. First, the panel uses an oversized heater and baffled vapor discharge chamber, which assures complete boiling and provides a steam blanket over the sample to eliminate any sample contamination that might occur due to discharge piping backflow. The vapor is then condensed and removed via a vent drain. After the sample is degassed, it is cooled with an additional sample cooler so that chemistry deviations in analyzer temperature compensation curves are reduced.

MODEL

DCCP

BENEFITS

The Sentry DCCP reboiler is a complete panel that measures specific cation and degassed conductivity or any individual measurement needed. The custom analysis panels include degassed conductivity, dissolved oxygen (DO) and corrosion products sampling. The panel is designed to work with a wide variety of conductivity analyzers and provides simple and compact cycle chemistry validation.

FEATURES

�� Capable of measuring specific conduc-tivity (SC), cation conductivity (CC) and degassed cation conductivity (DCC)

�� Regenerative cooling option (no cooling water)

�� Operates in accordance with ASTM D4519

�� Sample cooler ensures CC and DCC probes receive identical sample temperatures, eliminating errors

�� Compact wall-mount design

�� Surface mount, backplate option

�� CE, SGS and CSA available

DCC

CC

RC

HTR-1

SC-1

SC

2

1CIT

4

3

3CIT

FICV

SC

1CIT CC

1

6

3CE

3CE

3CE

1TIC

1TE

(1)The cooled sample flows

through the first conductivity cell which measures specific

conductivity (SC).

(2)The cation resin bed exchanges positive ions in the sample for

hydrogen ions.

(3)The sample flows through

the second conductivity cell which measures for cation conductivity (CC).

(4)The digitally operated

heater removes volatiles, such as CO2, in the sample. Turns off automatically

on low sample flow.

(5)The sample cooler returns the sample temperature to inlet

conditions.

(6)The sample flows through

the third conductivity cell which measures the conductivity of the

degassed sample (DCC).Operation

DCCP Reboiler with analyzer panel

S A M P L E CO N D I T I O N I N G

SENTRY DCCP REBOILERSingle Line Conditioning Modules

Gas

Liquid & Slurry

Solid & Powder

Steam & Water

Page 2: SAMPLE CONDITIONING - Iberfluid · The Sentry DCCP reboiler is a complete panel that measures specific cation and degassed conductivity or any individual measurement needed. The custom

sentry-equip.com

966 Blue Ribbon Circle North, Oconomowoc, WI 53066 U.S.A. | +1-262-567-7256 | [email protected]

© 2017 Sentry Equipment Corp. All rights reserved. All product and company names are property of their respective owners. It is solely the responsibility of end users, through their own testing and analysis, to select products and materials suitable for their specific application requirements, ensure they are properly installed, safely applied and properly maintained, and to limit their use to their intended purpose. Improper selection, installation or use may result in personal injury or property damage.

S-SW-SPS-00149-14 4-17

1Sa

mpl

e In

let

Sam

ple

Out

let

Coo

ling

Wat

er In

let

Coo

ling

Wat

er O

utle

t

17 1

/2 [4

44]

with

out a

naly

zers

24 1

/2 [6

22]

with

ana

lyze

rs

0 [0]

0 [0

]

36 [914]

Heater

SampleCooler

Resin Column

Flow IndicatingControl Valve

Conductivity Cell

Conductivity Cell

Conductivity Cell

Analyzer (by customer)

Analyzer (by customer)

Temperature Indicating Controller

Con

dens

ate

Dra

in

SPECIFICATIONSvoltage cell holders flow indicator with valve analyzers/probes

1 = 110–120 VAC, 1 ph, 50/60 Hz 2 = 220–240 VAC, 1 ph, 50/60 Hz

0 = None1 = DCC2 = CC & DCC (with RC-100)3 = SC, CC & DCC (with RC-100)

0 = none1 = yes

0 = customer supplied1 = supplied by Sentry Equipment

dimensions without analyzer side plate: 36 x 17.5 x 15 in (91 x 44.5 x 38 cm) with analyzer side plate: 36 x 24.5 x 15 in (91 x 62 x 38 cm)

weight 61 lb (28 kg)

pressure rating 100 psig (7 barg)

temperature rating 130°F (54°C) with RC-100 before degas

sample flow 100–200 cc/min

sample inlet connection 1/4 in tube fitting

sample outlet connection 1/4 in tube fitting

sample cooler type DTC-DCCP

cooler cooling water connections 1/2 in FNPT

cooling water consumption 0.2 gpm (0.8 Lpm) at 22°C (71.6°F)

input power 110–120/220–240 VAC, 1 phase, 50/60 Hz

power consumption 1.6 KVA

cell holders (optional) 3/4 in FNPT, 316 SS

resin column (optional) Sentry RC-100

S E N T R Y D CC P R E B O I L E R > S I N G L E L I N E CO N D I T I O N I N G M O D U L E S > S A M P L E CO N D I T I O N I N G