SAVe II
Easy Set Up Guide
2
Vent Overview
PEEP
Tidal Volume
Confirmation button
Alarm Silence
Respiratory Rate
Single Manual Breath
Power
Peak Pressure Limit
Height Selection
![Page 2: SAVeII easy guide - Compatibility Mode...Microsoft PowerPoint - SAVeII easy guide - Compatibility Mode Author medgu Created Date 6/26/2019 8:34:31 AM](https://reader033.vdocument.in/reader033/viewer/2022051812/602a70d85bde382f1d14ade2/html5/thumbnails/2.jpg)
6/26/2019
2
Circuit Setup
3
4
Start-up
1. Press power
2. Select pt height
3. Press confirm
![Page 3: SAVeII easy guide - Compatibility Mode...Microsoft PowerPoint - SAVeII easy guide - Compatibility Mode Author medgu Created Date 6/26/2019 8:34:31 AM](https://reader033.vdocument.in/reader033/viewer/2022051812/602a70d85bde382f1d14ade2/html5/thumbnails/3.jpg)
6/26/2019
3
5
Add PEEP/ Start Ventilator
1. Increase PEEP to 5
2. Press Confirm
*If different Tidal Volume is desired, refer to chart on next slide.
6
![Page 4: SAVeII easy guide - Compatibility Mode...Microsoft PowerPoint - SAVeII easy guide - Compatibility Mode Author medgu Created Date 6/26/2019 8:34:31 AM](https://reader033.vdocument.in/reader033/viewer/2022051812/602a70d85bde382f1d14ade2/html5/thumbnails/4.jpg)
6/26/2019
4
7
Add Supplemental FiO2
1. Remove black cap(do not discard cap or white filter)2. Attach green side of large bore tubing to vent3. Fully extend large bore tubing4. Attach small bore tubing to O2 source and turn flow to 6LPM
Monitor Patient
• Maintain end tidal CO2 between 35-45
• Increase RR or VT to lower end tidal CO2
• Decrease RR or VT to raise end tidal CO2
• Maintain SpO2 above 94%
• Increase FiO2 or PEEP (do not raise above 10) to raise SpO2
8