Semiconductor Detectors

Solid state Ionization Counters

Page 2: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 3: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

1/40 eV

Page 4: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 5: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

4 valence atoms

Page 6: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

No need to go to very low Temperature with Si:

Intrinsic carriers concentration

ISi ∼ 10-9 A, IGe ∼ 10-3 A,

Leakege current

Page 7: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

At room temperature: Measurable with Si detectors NOT Measurable with Ge detectors

ISi ∼ 10-9 A, IGe ∼ 10-3 A, Leakege current

Page 8: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

n-type p-type

To control the electrical conduction

p-type in contact with n-type give rise to a p-n junction

Page 9: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

n-type p-type

Page 10: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Increase of depletion region by applying REVERSE BIAS ⇒ E = Eint+ eVext Ge: Vext = 1000-3000 V, Si: Vext = 300 V

∇2ϕ = −ρ /ε

Ε(x) = − dϕdx

Page 11: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 12: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

∇2ϕ = −ρ /ε

Page 13: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

p-type n-type

Used for charged particles: 1 MeV e-, range = 1 mm 5 MeV α, range = 0.02 mm

Electric contact

Page 14: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 15: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 16: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 17: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

77 K (−196 °C) with LN2

Page 18: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 19: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

High Purity Ge Detectors impurity concentration N ~ 1010 atoms/cm3

600 µm

0.3µm

eNVd ε2

Characteristics size Ø~10cm, L~9cm shape coaxial n-type less sensitive to radiation damage operating temperature < 85 K rates ~ 10 kHz to prevent pile-up energy resolution 2 keV at 1.332 MeV (0.2 %) time resolution 4-5 ns (with CFD) 200-300 ns total rise time efficiency* up to 200% * relative to 7.5x7.5 cm NaI(Tl) for 1.33 MeV γ-rays emitted by 60Co source at 25 cm from detector (εa = 1.2 x 10-3)

active region d

V~2500-4500V

15%

150%

See also: Best choice HpGe detector, from ORTEC

Page 20: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Signal Pulse Shape depends on interaction point

V+

V-

V+

V-

Planar geometry

Coaxial geometry

Page 21: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

most severe in p-type detectors

Page 22: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Trapping Effect: - Reduction of pulse amplitude due to capture of carrier by trapping centers -  Deterioration of energy resolution due to variable amount of of charge lost per pulse

Page 23: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Energy resolution versus Temperature FW

HM

[ke

V]

FWHM

[ke

V]

Temperature [K] Temperature [K]

@122 keV @1332 keV

band structure effect

T(LN2) = 77 K (−196 °C)

Page 24: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to
Page 25: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Pulse shaping

true pulses

from preamp τ ~ 50µs

after shaping

FET

Preamplifier : FET (at 130 K, to minimize noise) Amplifier: CR-RC shaping circuit

pile-up

energy

time

τ

τ

t

out etEE −=if C1R1=C2R2=τ

τ ~ 15 µs τ ~ 15 µs is a good compromise

between reduced pile-up and good energy resolution

(depending on large charge collection)

mV

V

Page 26: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Preamplifier

R=input resistance C=input capacitance+ detector capacitance + cables …

τ = RC

operation mode for time information, high rates, …

operation mode for energy information

tc charge collection time ~100 ns

τ =RC decay time ~ 50 µs

Amplifier (RC-CR shaping)

RC-integrator (low-pass filter)

)1(...

/τtout

outin

eEE

EiRE

−−=

+=

CR-differentiator (high-pass filter)

τ/

...t

out

outin

EeE

ECQE

−=

+=

Page 27: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

γ-ray interaction

γσ EZpp ln2≈

ionization occurs in limited regions of the absorber

Ge

µ

ppCph σσσµ ++=

Linear attenuation coefficient (probability per unit path)

γ

γσEE

ZC

ln≈

54

5.3

−=

n

EZ n

phγ

σ

I/I0

t

e-µt

Page 28: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Detector response We detect recoil electrons

and NOT photons !

)(256.02

/21

22

2

cmEifMeVcm

cmEE

EEE

ee

eCEgap

>>=≈

+=−=

γ

γ

γγ

Egap

Important characteristics: §  energy resolution: δEγ/Eγ = FWHM/Eγ

§  peak-to-total: P/T = Areapeak/Areatotal

Egap

Page 29: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

Ge Response function (+ Anti-Compton Shield)

P/T~20%

P/T~60%

Anular detector

used material: BGO (Bi4Ge3O12) §  density ~ 7.3 g/cm3

§  Z = 83 §  3 times more efficient than NaI ⇒  ideal for very compact geometry (small spaces) N.B. in some cases NaI nose is used to improve the light output far away from PM tube

used with heavy metal collimators in front

Page 30: Semiconductor Detectorssleoni/TEACHING/Nuc-Phys-Det/PDF/... · 2012-12-21 · Semiconductor Detectors Solid state Ionization Counters . 1/40 eV . 4 valence atoms . No need to go to

incident γ

Compton scattering angular distribution

)cos1)(/(1 2'

θγ

γγ −+=

cmEE

Ee

high-energy γ-ray: forward scattering low-energy γ-ray: forward & backward

NaI nose: improvement of light output far away from PM tubes (low-energy γ-rays) BGO back-catcher: improvement of high-energy Compton scattering (high-energy γ-rays)