Page 1: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 1

Service Guide

Page 2: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 2

Table of Contents

1. Incoming Check ................................................................................... 4

1.1 Optical check ................................................................................................... 4

1.2 Initialisation ..................................................................................................... 4

1.3 Functions check .............................................................................................. 4

1.4 Sensors voltages and matrix check (control menu ) .................................... 5

1.5 Flowing sensors with span gas ..................................................................... 5

2. Single Steps for Disassembly and Re-building ............................... 6

2.1 Dismount instrument ...................................................................................... 6

2.2 Clean the pump ............................................................................................... 7

2.3 Change the printer unit ................................................................................... 7

2.4 Control the condensation trap or gas cooler ................................................ 7

2.5 Check of accessories ...................................................................................... 8

3. Details About some Specific Components ....................................... 9

4. Frequent Errors and Their Identification ........................................ 10 4.1 Error list ........................................................................................................ 11

5. Upgrades / Changes .......................................................................... 12

5.1 EN2–R (with Soot Measurement) ................................................................. 13

5.2 EN2 with TRGI ................................................................................................ 13 5.3 EN2 - F ............................................................................................................ 13

6. Outgoing Check ................................................................................. 14

6.1 Initialisation ................................................................................................... 14

6.2 Function check .............................................................................................. 14

6.3 Temperature, soot measurement and draught measu rement alignment . 14

6.4 Flowing sensors with span gas and adjustment ........................................ 16

6.5 Completion ..................................................................................................... 16

6.6 Check Setup Settings ................................................................................... 17

7. Programming ..................................................................................... 19

Page 3: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 3

7.1 Software ......................................................................................................... 19

7.2 Printout ........................................................................................................... 19

7.3 Fueltypes ........................................................................................................ 19

7.4 Activation of sensors .................................................................................... 20

8. Changes at the software version ..................................................... 21

9. Service Information ........................................................................... 24

9.1 Analogue values with reference values ....................................................... 24

9.2 Matrix .............................................................................................................. 24 9.3 Charging information .................................................................................... 25 9.4 Change to new battery type .......................................................................... 25

10. Care Tips .......................................................................................... 26

11 Details EN2 - F ................................ ................................................ 27

12 Details Ecom R (Remote Display) ................ ................................. 28

13 Upgrade Bluetooth ............................. ........................................... 32

14. Pictures from boards .......................... .......................................... 35

15. Climate cabinet ............................... ............................................... 37

Page 4: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 4

1. Incoming Check

1.1 Optical check

• Possible damages (at transport case or else)

• Check soiling grade of filters

1.2 Initialisation

• Switch on instrument

• Check mains power / battery operation (does the unit operate on battery?)

• Check error messages (error message must appear and then disappear)

• Release T-Gas sensor and then re-plug it

• Release T-Air sensor and then re-plug it

• Release water trap plug and then re-plug it

• Check all connectors (loose contact / damages)

• Check clock

• Note operation hours

1.3 Functions check

• Tightness check of probe � Low pressure must be constant

• Tightness check gas pump

• Check printing intensity

• Check draught sensor (function)

• Check draught measurement

• If needed be control heating check

• If needed be check 4Pa Test

Page 5: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 5

1.4 Sensors voltages and matrix check (control menu )

• O2 sensor must show approx 12000 mV ( +/- 2000mV). Value must be stable.

• CO sensor must show approx 0 (+/- 40 mV )

• Check matrix values.

To be able to check the matrix- and analogue values you have to enter the service menu

which can be done as follows:

1. confirm control in the main menu

2. slide the magnet over the read contact (app. 2 cm left of the interface cover)

1.5 Flowing sensors with span gas

• Flow sensors with CO span gas (at least 2-3 minutes for all sensors) -all toxic sensors must have a tolerance of +/- 5 %

-O2 sensors have a tolerance of +/- 0,1 vol.%

• Flow with CO/H2 span gas

� value may not go reverse, otherwise the sensor is defect

• Check CO switch off (manual switch-off)

� switch off CO prior to flowing with span gas

� by new switch on the value should not exceed 50 ppm

� after having reached the end value, the value should not run back

otherwise the sensor is defect

• By instruments for chimney-sweepers the CO sensor must additionally be checked with

a approx 1700 ppm concentration

• If needed be calibrate with different span gases

� Caution: Always apply span gas pressure free!

Page 6: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 6

1.6 Check of accessories

• Check probe (type SCD)

• Check sampling probe (type SCD)

• Check possible damages at connectors

• Dismount probe pipe and check tight position of thermocouple

• Check shifting ability of cone along pipe

• Check tubing (damages/cracks?)

• Check possible soiling and corrosion state of condensation trap

• Check possible cracks on water trap bowl

2. Single Steps for Disassembly and Re-building

2.1 Dismount instrument

• Loosen the screws at the bottom side of the

case and take out the instrument

• Loosen the screws down at the left and right

side of the instrument and open the

instrument

• by loosen the black riveting the instrument

can be dismantled in two parts

• Check O2 sensor on humidity and possible

leak

• Release sensors (if necessary)

• Release tubing (if necessary)

• Blow through gas channel plate with compressed air

• Renew the tubing of internal filters

Page 7: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 7

2.2 Clean the pump

• If needed be dismantle and clean housing of the peristaltic pump

• loosen pump body of the gas pump

• Check wearing state of seal and membrane

• Check soiling state of pump body and clean

• Re-assemble the measurement gas pump

2.3 Change the printer unit

• loosen the two screws

• take out carefully the printer

(confection at the unit)

• re-assemble the printer unit and carefully

• fix the screws (finger-tightened)

2.4 Control the condensation trap or gas cooler • Dismount condensation trap

• � after loosen the condense body the circlip

• has to be removed and the POM part has

• to be beaten off the cooling body

POM Part Condensation trap Cooling body

Seals clean with acid-free grease

Page 8: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 8

• Attention: special care must be taken for the cable.

• The gas cooler is constructed the

same way. The only difference is the

build in peltier element and the fan

which is situated above the cooling

body.

2.5 Check of accessories

• Dismount probe

• Check and grease O-rings O-ring

• Re-assembly probe • Check strain relief and soldering joints at all plugs

• Accessory check completed!

Outer pipe with grip Inner pipe Thermocouple

Page 9: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 9

3. Details About some Specific Components

Gas probe (type SCD) with 3-chamber tubing: • Performs the gas temperature and provides with a frictionless passing of the gas � Therefore always keep clean probe and tubing! • Draught measurement

Condensation trap:

• Also responsible for the frictionless passing of the gas

• serve for the admission of the humidity of the waste gas

• Furthermore it secures the condensation evacuation (error message) � Therefore always keep clean the condensation trap!

Gas pump:

• Low pressure must achieve 0,6 bar at a minimum

• Low pressure must remain constant � if not ascertained, measurement drifts can be the consequence • Sucking in the gas

Analogue values:

• The values in the service menu in sub point “Analogues values” are itemised labelled

Matrix:

• The values in the service menu in sub point “Matrix” are itemised labelled

� The CO% value must be in the matrix on 20000 (Only after upgrad a CO% sensor, and on each of the 3 pages)

Sensors:

• As a precaution the O2 sensor should be changed every 3,5 years

• Every 4 years should be recommended to change the CO sensor (Depends on the

working time of the CO sensor in ppm)

• Further special features:

• Screenshot by long simultaneous pressing of “store” and “Print”

• Error messages for T-room and T-gas can be ignored by pressing Key ESC

Page 10: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 10

4. Frequent Errors and Their Identification

Error description Troubleshooting / Err or source

Error message „Empty water trap“ - Dry contacts

- Clean water trap

- The pins which will be fitted with a

piece of shrinking sleeve

(see News 40/2009)

- Remove filter screen (metal).

Error message „Flash ...- ?“ Software update to version 1.4 /

02.09.2010

Unit does not charge LED (IC Max 4073) on charger board V3 lights although no AC adapter is plugged in. Exchange to charging board V4. (Built from # 2101)

Display shut down after 5 sec. from turn on. Contact/Wiring from display to CPU board has bad connection (cleaning)

Soot measurement does not work and not

show on the Display.

- Switch On with Power Supply. Power supply board transfer data only then

to the main board. (Only for the first

turn on)

- Change Fuel type´s

Tubing fold (peristaltic pump) cable sleeve support

(tubing order number: 55199)

Page 11: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 11

The model EN2 offers the possibility to print out error codes which enable to trace which error has been displayed last while the customer operated the instrument. The error history of the EN2 can be printed by selecting in the extended service menu the point PRINT (in the display of the operating hours, error counter etc). The error code is issued 4 Byte hexadecimal: 1* $8000 $0000 = O2 sensor (Error came below 2000 mV) 2* $4000 $0000 = CO sensor (over 200 mV) 3* $2000 $0000 = NO sensor (over 200 mV) 4* $1000 $0000 = 200 mv NO sensor 5* $0800 $0000 = NO2 sensor (over 200 mV) 6* $0400 $0000 = SO2 sensor (over 200 mV) 7 $0200 $0000 = T.Air sensor?

8 $0100 $0000 = T.Gas sensor 9 $0080 $0000 = Empty water trap 10 $0040 $0000 = condensation trap connected 11* $0020 $0000 = Operation temperature 12* $0010 $0000 = Put probe to fresh air 13 $0008 $0000 = Service required 14* $0004 $0000 = Reset clock 15 $0002 $0000 = Check sum CO 16 $0001 $0000 = Pump performance Switzerland

17* $0000 $8000 = Flash EEE error 18* $0000 $4000 = Flash Atmel 19* $0000 $2000 = ext. EEProm (I2C) 20* $0000 $1000 = EEProm Check sum 21* $0000 $0800 = MMC card SPI error 22* $0000 $0400 = MMC card not compatible

23* $0000 $0200 = MMC card full 24* $0000 $0100 = MMC card error 25* $0000 $0080 = setup intern 26* $0000 $0040 = Switzerland remove seal stoppers 27* $0000 $0020 = reserved 28* $0000 $0010 = reserved 29* $0000 $0008 = reserved 30* $0000 $0004 = reserved 31* $0000 $0002 = reserved 32* $0000 $0001 = reserved

* - Faults are stored

green - Faults aren’t stored Caution: The specifically developed software also can for the transformation of the code do used "Errorcode.exe" become !!! Please contact rbr Messtechnik for this software.

Page 12: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 12

5. Upgrades / Changes

• Changes the additional board Old additional board V1 (for sensor)

NO2

SO2 / CO%

Old additional board V2 (for sensor) Please connect the board on the right side. On the left side must be a hole free!

6. sensor CO %

SO2

free NO2

Additional board V4 for EN2 and J2KN (At the same place as the V2)

Page 13: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 13

• Important changes listed by serial numbers:

From serial number Reason

1198 WH EN2 with ODU plug

1799 O2 sensor Envitec, half high

2101 Supply Board V4 built in

2734 filter screen (metal) removed

4000 equipped with new PT2000 temperature sensor

02.04.2014: all EN2-R are equipped with the LifePo4 battery (103418)

5.1 Activation of ANA Board version Swiss:

- Resistor R12 changes from 1K Ohm to 10 k Ohm.

5.2 EN2–R (with Soot Measurement)

• Works also in accumulator mode

• Probe heating starts automatically as soon as following conditions are complied with:

1. You are in the mode „Soot“

2. A fuel type for soot measurement has been selected (for example oil)

• No software change necessary

• Difference to standard EN2:

1. Circuit board is new (will be detected automatically)

2. Transport case is deeper

3. Heated probe

• Attention: The retrofitting of a standard EN2 is not possible!!!

5.2 EN2 with TRGI

• Retrofitting just for TRGI is possible for all EN2 instruments.

• No software change necessary.

• You need a under case with all the material necessary (e.eg. UNO).

5.3 EN2 - F (FFFFlight case)light case)light case)light case)

• condensate trap must be set in upper position, otherwise you get the error message

"Connecting the condensate trap"

• device with lithium-ion battery (extended battery life)

• Unit can be retrofitted with Bluetooth for remote control "Ecom R"

• Hard case is suitable for travel in a plane.

• For details see chapter 11 and 12

• Hardware is the same as EN2 with a few exceptions:

- AC/DC adapter in the device

- Micro switch on the condensate trap (to control the position)

- Li-ion accu (for longer operating hours). (Battery with a different connector!)

- Hard case

- Cooling fan in case

Page 14: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 14

6. Outgoing Check

6.1 Initialisation

- Switch on the instrument on battery (without accessories)

- Connect mains charger

- Check error messages

- Does the error message „T-Gas sensor“ appear?

- Connect T-Gas sensor

- Does the error message „T-Gas sensor“ disappear?

- Does the error message „T-Air sensor“ appear?

- Connect T-Air sensor

- Does the error message „T-Air sensor“ disappear?

- Does the error message „Connect condensation trap“ appear?

- Connect plug of condensation trap

- Does the error message „Connect condensation trap“ disappear?

- Check possible loose contacts at T-Gas and T-Air sensors

- Check matrix values (are all 3 temperature values available?)

6.2 Function check

• Check all instrument functions • Check and adjust date/time

• Perform tightness check with probe Probe with pump tester

• Check condensation evacuation function

• Error message „Empty water trap“ must appear

• Check T-Gas and T-Air with heat source (temperature increase positive?)

• Check battery operation (lines by connection of mains charger)

6.3 Temperature, soot measurement and draught measu rement alignment

• Calibrate gas temperature with simulator � Wait for termination of stabilisation time of 3 minutes

- Adjust value on 22,2 °C

- Check value with 84 °C (+/- 1 °C)

- Check value with 222 °C (+/- 1 °C)

• Compare T-Air with T-Gas value (place the two sensors side by side)

� Wait for termination of stabilisation time of 3 minutes

Page 15: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 15

• Compare values and, if need be, adjust T-Air on T-Gas value

• Calibrate and check draught measurement

• Connect draught tubing and check (does the draught sensor react when air is blown

thru the probe?)

• if necessary calibrate soot measurement

• if necessary check soot measurement with

• volume control equipment (1,63 l < 1min)

Mark by 1,63 l

• Caution: You can check the soot measurement in the calibration mode:

„F2“ – calibration

„F3“ – test

Page 16: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 16

6.4 Flowing sensors with span gas and adjustment

• These setting will be done in the service menu

• Perform new calibration phase

• Flow CO sensor with span gas and adjust if need be

� Wait for termination of 3 mn stabilisation time

• Issue a printout as record for service centre (piece of evidence in case of later claim)

• Issue a printout as record for customer

• Check CO switch-off (switch on the fresh air pump manually prior to flow with COH2

gas)

• Flow with COH2 span gas and check

� By 2,5 % O2 switch off the fresh air pump � The CO value should not exceed 50 ppm � Wait for termination of stabilisation time of 3 minutes

• Issue a printout as record for the customer

• If installed, perform this procedure also for other sensors

• Issue a printout as record for the customer

• If needed be calibrate with different span gas

6.5 Completion

• Zero the operation counter by a maintenance or instrument cleaning � go to the service menu and when the operating hours are displayed put a 100 Ohm resistor in the t-room socket and press „F2“ By this the operation hours, the CO-fresh air purging and the measuring cycles are set to zero.

Put 150 Ohm resistor into the t-room socket and press “F2” to reset the overall operating hours to zero!

� Attention: service date will be set to the new date!

• Install instrument in its transport case

• Check instrument completeness (incl. accessories)

Page 17: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 17

6.6 Check Setup Settings

� for all settings a 100 Ohm resistor has to be put in the T – room socket and by pressing „F2“ the corresponding point will be confirmed

Important note: You may only modify those settings and configurations if you are fully aware of the characteristics of the according parameters. In case of questions contact the rbr service!

EN2setup Parameters

Flags: 1 with or without peristaltic pump (only EN2)

2 with Pa low pressure sensor (lower pressure range for Pa resolution with one decimal place, Italy)

1

3

5

7

9

11

13

15

17

19

21

23

25

27

29

31

2

4

6

8

10

12

14

16

18

20

22

24

26

28

30

32

33

34

35

36

37

38

39

40

41

42

43

44

45

46

Page 18: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 18

3 with 4Pa PCB (for 4Pa test) (hardware has to be installed)

4 with H2 display

5 low NOx measurement with one decimal place for NO, NO2 and NOx (e.g. USA)

6 O2 reading with two decimal places (depends on version)

7 with LiFePo4 battery (only if Li-Io battery is installed, from V1.8_07/2012, EN2-F + D standard)

8 O2 sensor with non-linear characteristic curve (e.g. 5FO, 5Ox)

9 with procedures, which guide the user through the measuring steps (e.g. France)

10 with UNI10389 (for Italy)

11 with mean value calculation

12 temperature unit Fahrenheit °F (only USA)

13 time display AM/PM (only USA)

14 with display of the sensors‘ temperature (USA)

15 without CO limit (Netherlands)

16 with toxicity index, CO/CO2 ratio (Netherlands)

17 with service reminder (when the set operating hours respectively the set date is reached)

18 with dew point display

19 with NOx calculation from NO (without NO2 sensor) – refer to 34

20 with NO2 display instead of NO at mg/m3 (Austria/Switzerland)

21 with CO and air duct measurement

22 with 4Pa + heating check (hardware has to be installed)

23 with internal flow measurement (hardware has to be installed)

24 with dongle for calibration

25 version engine

26 with data logger function on SD memory card

27 t-room and t-gas display with one decimal place

28 without 21% O2 limit upwards

29 Chinese font GB2312-SIMSUN

30 version France (with CO ambient measurement)

31 version BImSchV (with specific mean value measurement)

32 version Weishaupt

Values and factors: 33 Step size for CO display without CO ppm

34 NO factor for NO2 calculation without separate NO2 sensor (factor 50)

35 cross sensitivity between CO2 and O2 (X.XXX%/%CO2)

36-44 currently not used

45 pump speed 0…70 (default 35), for ecom-D insignificant

46 1=NO2 2=SO2 3=CO% or template: 4=NO2 08=SO2 16=CO% (e.g. 28=NO2+SO2+CO%)

Page 19: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 19

7. Programming

If the device has a total crash then you must make at first the basic software with the cyclone pro. The next step is the normal software update 7.1with the PC. � Caution: Please hold ON/OFF button during the update with cyclone pro !

7.1 Software • Confirm the point „Software - Update“ in the instrument

• Connect instrument to PC via USB cable

• Confirm „Yes“ with F1

• Start PC software EN2Setup

• At transfer confirm the selection „Software update“ or „Software update with default

initialisation“

� Software update: the firmware will be updated

� Software update with default initialisation: Here it can be selected which components

shall be transferred apart from the pure software (fuel types, Print, Parameters or

Procedures).

• Select file and open the file

• Start with OK

� The desired procedures of the customer can and the desired parameters

of the customer can be stored beforehand in a new .com file . The parameters can

also be stored in the service menu afterwards. (see point 5.6 „Check set-up

settings“).

7.2 Printout • Connect instrument to PC via USB cable

• Switch on instrument (Don’t press any button unless you got an error message)

• Start PC software EN2Setup and click the point „Customer Data“ in the sub-menu Edit

• Select file

• Fill up text fields as desired

• Hereby the serial number can be changed „Overwrite serial number“ (Need expert

modus)

• start with „Transfer to EN2“

7.3 Fueltypes • Connect instrument to PC via USB cable

• Switch on instrument (Don’t press any button unless you got an error message)

• Start PC software EN2Setup and click the point „Fueltypes“ in the sub-menu Edit

• Select file

• Click „Write to Transfer L1 / L2 / L3“ to start the programming procedure

Page 20: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 20

7.4 Activation of sensors

• The instrument automatically identifiers and activates NO sensors

• All other sensors must be activated

• Enter in the service menu and select „Setup”

• confirm the point „with 4.th sensor“

• select the sensor „1=NO2 ; 2=SO2 ; 3=CO%“

• After Software Version 1.7 :

• There is no longer a selection in the service menu for “4th sensor”

• You activate the sensor in the internal service menu with numbers, for example:

NO2 = 4, SO2 = 8, CO% = 16, when you have NO2 + SO2 it´s 12 to put in the box.

Page 21: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 21

8. Changes at the software version V1.1 : 1) SUI version (only SUI) 2) CO - automatic 3) M.Flow 4) DataLogger(MMC) 5) information HC (only German) 6) Count Down 7) T.G/T.R with comma 8) SUI message " Removing seal stoppers!" (only SUI) V1.2 : 1) Staudruckplatine -> internal flow possible 2) Circuit board is recognized automatically 3) Change SUI Hotkeys (only SUI) V1.3 : 1) Changes BImSchV 2) UNI10389 (only ITA) 3) Changes FRA 4) BImSchV in setup (only German) 5) Weishaupt without calibration 6) Operation temperature to -5°C 7) Enhanced protocol with loses 8) Header CSV file V1.4 : China Fonts, correction Öl pressure printer V1.5 1) soot measurement 2) Flash error no displayed 3) CO limit to 20000 (13.12.10) V1.6 1) Pressure test (04.04.11) 38400 Baudrate 2) @ for keyboard (GI in enhanced Protokoll) 3) O2 also in X.XX% 4) Message "Finish without storage?" V1.7 1) France "Attestation" + + NO2/SO2 stored values -> CL protocol

2) CO limit to 65500 if only CO% (Copf <> 0 and <1000), 3) Fixed bug Italy store averages 4) TRGI light stays on always 5) back pressure improves detection Switzerland:

Message ID Russ + warning changed with deactivated soot. 6) Switzerland Predefining dynamic pressure values changed 7) changes to France (H.Wolf list) 8) Sensor activation changed. NO2/SO2/CO% with values enable 08/04/16 9) Sensor SO2 and NO2 simultaneously (sensor value 12 in 4th), Korea! V1.8 1) Bolzano integrated / error Fixed sending dial to UNI / average values as J2KN-3.2

2) Average measurement changed (starting and stopping with F2 and F4) 3) Image MMC card inserted 4) Italy ppm CO / 0% O2 is always relative to 0% O2.

V 1.8 dated 03.07.2012 1) New software switch for LiFePo4 battery 2) Display of battery slower

Software Changes EN2 Version 1.9 dated 29.08.2012 1) Service history added in “Control” picture 3. + Service (100R+picture working hrs.) 2) Procedures 18/19/20 added 3) CO/NO/NO2 ppm-hrs. added (reset with J2K2probase2.exe) 4) Service date set with added service history

Page 22: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 22

4) Display CL0/100 dependent of T.room Resistor 5) Bluetooth module automatic setup with name and serial number of EN2 6) New features for enhanced protocol + auto select at 0R T.room in startup menu 7) few changes sensor configuration + CO purge at H2 overflow

Software Changes EN2 Version 1.9 dated 10.01.2013 1) New Dongle function with SD-Card 2) Bug data acquisition search function solved 3) WLAN possible with additional pcb 4) H2S Sensor

Software Changes EN2 Version 1.9 dated 06.02.2013 Upper limit for “Tightness test” (Pressure test ) stabilization time was changed to 120 minutes. Software Changes EN2 Version 1.9 dated 14.02.2013 1) Added parameter for read of printout in enhanced protocol. 2) Added Software Dongle compatible to ECOM-CL and ECOM-B 3) Fixed bug for Italy and ETA(K)

Software Changes EN2 Version 1.9 dated 27.02.2013 Bug France Attestation + empty SD card solved

Software Changes EN2 Version 1.9 dated 07.03.2013 Added new Flash AT45DB161E

Software Changes EN2 Version 1.9 dated 11.03.2013 In USA Version (°F+AM/PM active ) pump is not swit ched off in menu Main + menu Adjustments + menu Control

Software Changes EN2 Version 1.9 dated 18.03.2013 WLAN without authentication (Open)

Software Changes EN2 Version 1.9 dated 10.05.2013 PT2000 up to 500°C, Problem with H2 display solved

Software Changes EN2 Version 1.9 dated 27.05.2013 1) Improved stability of PLL 2) Pressure test „Usage Property“, measurement time in seconds 3) UNI11137 („Usage Property“) Italy changed

Software Changes EN2+D Version 2.0 dated 04.11.201 3 1) ECOM-D implementation with lot of changes for printer, battery charging… 2) Not used menus are not displayed (ex. Bluetooth) 3) Menu 4Pa is activated together with Pa option (Italy)

Page 23: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 23

4) New Option „O2 Eco Sensor“ and „without 21%O2 limit“ 5) “USB“ Menu ( 0R t.room) in service mode is activated at menu “draft” and higher 6) Some adjustments like Ref.O2 or Unit are restored from Memory Card 7) China: CO2 Display is valid for O2 0..21%

Software Canges EN2+D Version 2.0 dated 05.02.2014 1) Printout recalculation same as J2KN 2) Error level low NOx changed 3) Internal flow calibration limit changed 4) Switzerland Input additional values only with number keys

Software Changes EN2+D Version 2.0 dated 24.02.201 4 - ecom-D and ecom-EN2 compatibility, software automatically detects the hardware (e.g. A/D converter) and PCB - menus that are not required are disabled (e.g. Bluetooth) - 4Pa modifications for Italy - new option “O2 eco sensor” and “without 21%O2 limit” - adaptions of the “USB” submenu - settings such as unit or reference O2 are read from the MMC card - China: CO2 display is valid for O2 0…21% - SDHC memory cards and FAT32 are supported - Switzerland: modifications input “further values” Software Changes EN2+D Version 2.0 dated 20.03.201 4 - special version Jenbacher added Software Changes EN2+D Version 2.0 dated 27.03.201 4 - special version Engine added Software Changes EN2+D Version 2.1 dated 26.05.2014 - America Flow function improved - Enhanced protocol „next service“ added - IR-CO2 added Software Changes EN2+D Version 2.0 dated 29.07.201 4 - Data logging time up to 999 sec Software Changes EN2+D Version 2.0 dated 11.08.2014 - Pressure measurements (TRGI) measurement time corrected - special changes for France

Page 24: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 24

9. Service Information

9.1 Analogue values with reference values

Temp. Gas (0) T-Gas-Comp (9xxx) Temp. Room (9xxx) Temp. CO (9xxx)

Temp. Cooler (32xxx) Ext. Analog (0) Accu (7xxx) GND (-32xxx)

O2 (1xxxx) CO (50) H2 (50) NO (0)

NO 200mV (0) Russ/Soot–D (0) Russ/Soot–S (0) GND1 (0)

add. PCB NO2 (0) add. PCB SO2 (0) add. PCB CO% (0) add. PCB CxHy (0)

add. PCB CH5 (0) add. PCB CH6 (0) add. PCB CH7 (0) add. PCB CH8 (0)

Draft normal (1xx) Draft low (2xxx) Draft 4Pa (0) Draft 4Pa low (-2xxx)

Water trap (5xx) Light sensor (2xxx) Temp. Printer (2xxx) ---

Caution: The reference to the values is displayed down left in the display!

Caution: The values in () are only standard values and depend on the instrument features.

9.2 Matrix

• Temp hi / Temp mid / Temp low / other calibration values

� Caution: The reference to the values is displayed down left in the display!

CO H2 NO SO2

CO x x 0 0

H2 x x 0 0

NO 0 0 2225 0

SO2 0 0 0 5000

Page 25: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 25

9.3 Charging Information:

Information for charging board V3 and V4: External LED: LED green - battery is fully charged LED red - battery is charging or power is on Flashing red and green - Error loading / battery fault Internal LED on board: Every 4 seconds is displayed flashing on the frequency of the flashing between two intervals, the status of the software are recognized: 1 time = start charging with current and voltage regulation only when the battery is flat. 2 times = constant current charging 1 A 3 times a constant = 7.2 V current charging. 4 times = battery full. From below the charging cycle to 6.35 V, or when the Device is performed maintenance charge (= 6.5 volts) Constant blinking = battery failure, no current flow.

9.4 Change to new battery type: - Software version V1.8 from 07/2012 or later.

- Tick the box “LiFePo4 accu” in the setup

- Change the accu and connector.

- Reduce the bolts in the accu case.

- Use cellular rubber to fix the accu in the case.

Page 26: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 26

10. Care Tips

• Never stock the instrument in humid or cool rooms overnight!!! � Condensation formation

• Clean condensation trap � Exchange filter and clean gas paths with compressed air

• Regular filter change (soot filter, fine dust filter)

� Protection of the pump against soiling, condensation trap

• Clean probe (type SCD) and 3-chamber tubing � Hang up tubing overnight

� Clean probe & tubing with compressed air

Page 27: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 27

11. Details for EN2-F

Connection air

temperature

Slot for multi-media card

Connections Pa pressure

sensor

Connection gas

temperature

Connection USB

Connection sampled

gas

Connection draught

Connection pressure

Connection AUX

Integral printer

Accu compartment

Graphic display

Optical flow control

Condensation trap

(with peltier cooler)

with fine dust filter

Bluetooth Antenna

Mains connection

Charge control

Sensors compartment

Page 28: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 28

12. Remote monitor ecom-R

Connection with ecom-EN2-F (the ecom-R is only connectable with one ecom-EN2-F) Program connection: -switch on ecom-EN2-F (only one instrument with Bluetooth near the ecom-R ) -choose at the ecom-EN2-F in menu „Adjusments/Internal/ Bluetooth“ the „ enhanced Protocol“ -switch on ecom-R (programming will be done automaticly) Delete connection: -hold the red key (approx. 5 s) until „Unpair:“ appears -when „Unpair: OK“ shown, the connection is deleted Operating

ecom -EN2-F must have Bluetooth option!

(possible data transfer distance approx. 10 m)

Change display:

Choose line

Adjust line

Store adjustment

Break off adjustment

Change display

Start printout

at ecom-EN2-F

Choose dislay

(4 displays

Switch on/off ecom-R

(hold approx. 1 s)

READY = Connection exists

CALIB = EN2-F in calibration phase

WAIT = Searching for connection

Batterie voltage

Page 29: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 29

Battery Change -open battery box (operate unlocking) -take out used batteries -put in new batteries (respect polarity) -close battery box Important: Always dispose of used batteries and accumulators in official recycling containers only (e.g. in battery shops). Model with integrated air temperature sensor If there is no air temperature sensor connected at the ecom-EN2, then the air temperature sensor of the ecom-R supplies the measured values.

General information: - Battery life approximately 16 hours. - Push the red button one second to turn on. - Backlight turns off automatically after 15 min. - Batteries are charged via a cable connection (EN2 must be turned on). - It is recommended to use the extended data transfer mode to operate, because only in this protocol it shows errors and additional information. - Fault at the flue gas analyzer is shown with an exclamation point.

Battery box

Batteries 1,5 V AA or Accus 1,2 V AA

Unlocking

Air temperature sensor

Page 30: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 30

Connect with a ecom unit:

Unlink:

In operation, press the red button and hold it. The device turns off but hold the button. 5 seconds later you get the message "Unpair".

Now release the button. If the pairing is canceled you get the message "Unpair: OK".

A communication error between the radio module and microchip with "Unpair: Error" is acknowledged.

link:

The next time you restart the device connects automatically (Note: Only turn on a device).

Internel setup:

Push the button (1) and hold them and swipe with a magnet on the right side (2).

Each of the four keys activates a menu item.

• Setup parameters (low NOx, TG comma, Int. temp., O2 x.xx%). • With Unpair BT cancels the pairing between Ecom R and ecom-unit. • BT <-> Serial is to connect with a J2KN with a cable (terminal application) • Exit leaves the ecom-R to return to measurement mode

Page 31: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 31

Software update

To make software updates you need a cyclone Pro and an image file. Remove the board of the ecom-R from the case. For this unscrew the two case sides. Then pull out the the board and the front panel together. Cyclone Pro: Check that the jumpers on the Cyclone Pro are plugged as shown.

Then plug the connector as shown.

The Cyclone Pro connecting with the MicroStac plug on the board, select the SAP file and start the program. The power supply is provided by CyclonePro.

The update is finished. Assemble the equipment. Unlink it. And reboot.

Updates The supplier of the Bluetooth module changed the internal Bluetooth protocol in a way that a software update to version 1.9 was necessary.

When a new Bluetooth module was used in an ecom-R, a connection between the ecom-R and the measuring instrument could not established. Version 1.9 solves this problem.

Due to the new Bluetooth module a new approach is necessary when connecting the ecom-R with the measuring instrument:

1. Disengage the ecom-R from the measuring instrument.

Page 32: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 32

2. Switch on the ecom-R. If the ecom-R connects to the measuring instrument, go to point 6.

3. Switch off ecom-R.

4. Switch on ecom-R again.

5. The ecom-R connects to the measuring instrument.

6. The ecom-R is ready to use.

13. Upgrade EN2 Bluetooth module 1) Parts required (SET= Part No.100761 in service center, Part No. 102249 by rbr

service)

2) Solder the antenna (ground / signal)

Page 33: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 33

3) Plug in the Confection. Fit board and fasten with 2 screws.

4) Put away the protection of the glue part. Install the antenna with the nut.

Page 34: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 34

5) the finished upgrade:

6) Settings in the device in the internal menu.

7) Maybe you have to enter a password at the receiver. It is in the delivery state either 0000 or 1234

Page 35: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 35

Page 36: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 36

Page 37: Service Guide - Novotech · 2018-08-14 · Contact/Wiring from display to CPU board has bad connection (cleaning) Soot measurement does not work and not show on the Display. - Switch

• Erstellungsdatum: 09.08.2012 • KR – updated 12.08.2014 VH 37

15. Prepare climate chamber

Use the Aux Connector on the front of the unit to a RS 232 connection.

Switch on the unit and start the software with the choice “EN”.

You don’t necessarily need T-room or T-gas. If it does not work you should be work with a crossing wiring.