service manual - rossmax · 6. recommended parts list: please refer the below parts to the exploded...

2
6. Recommended Parts List: Please refer the below parts to the exploded drawing Medication cup Mouthpiece Air tube Filter Adult and Child mask Cover and diverter Nozzle 5. Trouble Shooting: Symptom A Cannot be turned on Figure Root causes Defective power key Defective power wires Correction Replace the power key Replace the power wires Symptom B Motor operates but no spray is generated Figure Root Causes Run out of liquid medication Air tube is not connected properly Correction Refill the medication cup Re-connect the air tube and medication cup properly Symptom C Low spray volume Figure Root Cause Valve is open Nozzle is clogged Defective filter Correction Adjust the valve Rinse the nozzle with water Replace the filter Symptom D Machine shuts down suddenly during operation Figure Root Cause Motor automatically shuts down due to overheat Correction Wait for the temperature to drop, which takes about 20 minutes www.rossmax.com Rossmax Swiss GmbH, Tramstrasse 16, CH-9442 Berneck, Switzerland INXXXXXXXXXXXXXXX S&S_ServiceManual_NA100_EN_SW_ver1544 Service Manual Document No.: SM-11-012 Version: 2 ROSSMAX Model No. NA100 Nebulizer

Upload: phungthu

Post on 07-Nov-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

6. Recommended Parts List:Please refer the below parts to the exploded drawing

Medication cup Mouthpiece Air tube Filter Adult and Child mask Cover and diverter Nozzle

5. Trouble Shooting:Symptom A Cannot be turned onFigure

Root causes Defective power key Defective power wiresCorrection Replace the power key Replace the power wires

Symptom B Motor operates but no spray is generatedFigure

Root Causes Run out of liquid medication Air tube is not connected properlyCorrection Refill the medication cup Re-connect the air tube and

medication cup properly

Symptom C Low spray volumeFigure

Root Cause Valve is open Nozzle is clogged Defective filterCorrection Adjust the valve Rinse the nozzle with

waterReplace the filter

Symptom D Machine shuts down suddenly during operationFigure

Root Cause Motor automatically shuts down due to overheatCorrection Wait for the temperature to drop, which takes about 20 minutes

www.rossmax.com

Rossmax Swiss GmbH, Tramstrasse 16, CH-9442 Berneck, Switzerland IN

XXXX

XXXX

XXXX

XXX

S&S_Se

rviceMa

nual_N

A100_E

N_SW_

ver1544

Service ManualDocument No.: SM-11-012 Version: 2

ROSSMAX Model No.NA100 Nebulizer

CONTENT:1. Exploded Drawing ..........................................................2 2. Part List .................................................................................2-33. Cautionary Notes ............................................................3 4. Assembly Method.................................................................45. Trouble Shooting ........................................................5-6 6. Recommended Parts List ...................................................6

1. Exploded Drawing:

2. Part List:(No.) (Q’ty) Description Material (No.) (Q’ty) Description Material

1 1 Cover Plastic 8 1 Motor Plastic

2 1 Switch Plastic 9 1 Bottom ABS

3 1 Filter Cover ABS 10 6 Screw Metal

4 1 Filter Plastic 11 4 Rubber Pad NBR(65°C)

5 1 Upper Case ABS 12 1 out tube NBR(65°C)

6 1 Wire Plastic/Metal 13 1 In tube NBR(65°C)

7 4 Motor Sponge NBR (65°C)

3. Cautionary Notes:• This machine is not advised to be used in an environment with flammable and explosive materials.• Please read the instruction manual carefully before using the machine.• Please clean the medication cup, mouthpiece and masks before using the product for the first time.• The product is designed for single patient’s use. Please do not use the machine for more than one patient to avoid

cross transmission.• Do not use the machine if runs out of the filter.• Please unplug the machine after use.• Please clean then air-dry the nebulizer kit after every use to avoid clogging.• Please do not use the machine in an environment with possible exposure to water such a bathroom, kitchen and

etcetera to avoid electric shock. • Do not place the machine near a sink or bathtub. Please unplug the unit immediately if the above circumstances

occur.• Please do not use the machine during sleep. Children should not use the machine without the supervision and

instruction of adult.• Please make sure the machine remains off, unplugged and well-stored while not in use.• Please use the machine in an environment with good ventilation• Please stay from the machine plugged while filling the liquid medication to avoid electric shock.• The motor in the machine, equipped with a temperature control switch, will stop operating when it is overheated

and will be able to operate again after its temperature drops back to normal (approximately 20 minutes).• Do not attempt to disassemble the machine, which invalidate the warranty as well as might cause malfunction.• Please follow the physician’s instruction before using the machine. Also, please contact the local distributor if there

is any questions.

4. Assembly Method:1. Place the nozzle

into the Cover and diverter

2. Screw the cover tight to the medication cup

3. Slide the mouthpiece onto cover

4. Connect the air tube and the opening at the bottom of medication cup

5. Connect the other end of the air tube and the vent hole of the nebulizer

6. Take off the mouthpiece if you would like to use a mask

7. Replace the mouthpiece with a mask