sfc332p - us operators manual - dynamic flow computers · 2020. 3. 2. · dynamic flow computers...

122
Date: 3/10/2021 SFC332P - US OPERATORS MANUAL Flow Computer Prover Version 11104 W.Airport Blvd, Suite 108 Stafford, Texas 77477 USA (281) 565-1118 Fax (281) 565-1119

Upload: others

Post on 30-Jan-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

  • Date: 3/10/2021

    SFC332P - US

    OPERATORS MANUAL Flow Computer

    Prover Version

    11104 W.Airport Blvd, Suite 108

    Stafford, Texas 77477 USA

    (281) 565-1118

    Fax (281) 565-1119

  • Date: 3/10/2021

    WARRANTY

    Dynamic Flow Computers warrants to the owner of the Flow Computer that the product

    delivered will be free from defects in material and workmanship for one (1) year

    following the date of purchase.

    This warranty does not cover the product if it is damaged in the process of being installed

    or damaged by abuse, accident, misuse, neglect, alteration, repair, disaster, or improper

    testing.

    If the product is found otherwise defective, Dynamic Flow Computers will replace or

    repair the product at no charge, provided that you deliver the product along with a return

    material authorization (RMA) number from Dynamic Flow Computers.

    Dynamic Flow Computers will not assume any shipping charge or be responsible for

    product damage due to improper shipping.

    THE ABOVE WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESS

    IMPLIED OR STATUTORY. BUT NOT LIMITED TO ANY WARRANTY OF

    MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY

    WARRANTY ARISING OUT OF ANY PROPOSAL, SPECIFICATION, OR SAMPLE.

    LIMITATION OF LIABILITY:

    DYNAMIC FLOW COMPUTERS SHALL HAVE NO LIABILITY FOR ANY

    INDIRECT OR SPECULATIVE DAMAGES (INCLUDING, WITHOUT LIMITING

    THE FOREGOING, CONSEQUENTIAL, INCIDENTAL AND SPECIAL DAMAGES)

    ARISING FROM THE USE OF, OR INABILITY TO USE THIS PRODUCT.

    WHETHER ARISING OUT OF CONTRACT, OR UNDER ANY WARRANTY,

    IRRESPECTIVE OF WHETHER DFM HAS ADVANCED NOTICE OF THE

    POSSIBILITY OF ANY SUCH DAMAGE INCLUDING, BUT NOT LIMITED TO

    LOSS OF USE, BUSINESS INTERRUPTION, AND LOSS OF PROFITS.

    NOTWITHSTANDING THE FOREGOING, DFM’S TOTAL LIABILITY FOR ALL

    CLAIMS UNDER THIS AGREEMENT SHALL NOT EXCEED THE PRICE PAID

    FOR THE PRODUCT. THESE LIMITATIONS ON POTENTIAL LIABILITY WERE

    AN ESSENTIAL ELEMENT IN SETTING THE PRODUCT PRICE. DFM NEITHER

    ASSUMES NOR AUTHORIZES ANYONE TO ASSUME FOR IT ANY OTHER

    LIABILITIES

  • Date: 3/10/2021

    CHAPTER 1: QUICK START.................................................................................................................... 1-1 Introduction: ............................................................................................................................................ 1-1 Technical Data: ........................................................................................................................................ 1-2 Parts List .................................................................................................................................................. 1-3 SFC332P Prover Flow Computer: Dimensions ...................................................................................... 1-4 Starting and Installing theWindow Software: .......................................................................................... 1-5 Website - DFC Configuration Software .................................................................................................. 1-6 Getting acquainted with the flow computer wiring: ................................................................................ 1-8

    Back terminal wiring: .......................................................................................................................... 1-8 Back Panel Jumper .............................................................................................................................. 1-9 Memory Jumper................................................................................................................................. 1-10

    INPUT/OUTPUT: Assigning and Ranging Inputs ............................................................................... 1-14 Input/Output Assignment .................................................................................................................. 1-14 How to assign a transmitter to an I/O point ....................................................................................... 1-14 Ranging the Transmitter Inputs: ........................................................................................................ 1-15

    WIRING: ............................................................................................................................................... 1-16 Wiring of Analog Inputs: ................................................................................................................... 1-16 Wiring the analog inputs 1-4 : ........................................................................................................... 1-17 Wiring of Analog Inputs 5,6 : ............................................................................................................ 1-18 Wiring of RTD Input : ....................................................................................................................... 1-19 Wiring of Analog Output:.................................................................................................................. 1-20 Turbine Input Wiring ......................................................................................................................... 1-21 Turbine Wiring for Passive (dry contact) Pulse Generators .............................................................. 1-24 Density Input Wiring ......................................................................................................................... 1-25 RS-232 Connection ........................................................................................................................... 1-26 RS-485 Connection ........................................................................................................................... 1-27 Wiring of Status Inputs ...................................................................................................................... 1-28 Wiring of Switch/Pulse Outputs ........................................................................................................ 1-29 I/O Expansion .................................................................................................................................... 1-30 Wiring of Brooks Prover: .................................................................................................................. 1-33 Wiring of Calibron Prover: ................................................................................................................ 1-34

    Calibration Through Windows™ Software ........................................................................................... 1-35 Analog Input of 4-20mA or 1-5 Volt Signal ...................................................................................... 1-35 RTD Calibration: ............................................................................................................................... 1-36 Calibration of Analog Output: ........................................................................................................... 1-37 Multi-Variable Transmitters (Model 205) – DP and Pressure ........................................................... 1-38 Multi-Variable Transmitters (Model 205) –Temperature .................................................................. 1-39 Verifying Digital Inputs and Outputs ................................................................................................ 1-40

    Calibration Through DOS Software ...................................................................................................... 1-41 Analog Input 4-20mA or 1-5 volt signal: .......................................................................................... 1-41 Verifying Digital Inputs and Outputs ................................................................................................ 1-42

    Pulse Interpolation: ................................................................................................................................ 1-43 Double Chronometry: ............................................................................................................................ 1-43

    CHAPTER 2: Data Entry ............................................................................................................................ 2-1 Configuration File through Window Program ......................................................................................... 2-1

    New ..................................................................................................................................................... 2-1 Open .................................................................................................................................................... 2-1 Close .................................................................................................................................................... 2-1 Save ..................................................................................................................................................... 2-1 Save As ................................................................................................................................................ 2-1

    VIEW ...................................................................................................................................................... 2-2 View Drawings .................................................................................................................................... 2-2

    TOOLS .................................................................................................................................................... 2-3 Com Settings ....................................................................................................................................... 2-3 Meter Configuration ............................................................................................................................ 2-4

  • Date: 3/10/2021

    Security .............................................................................................................................................. 2-17 Connect to Device ............................................................................................................................. 2-18 Go Offline .......................................................................................................................................... 2-18 Settings .............................................................................................................................................. 2-18

    Calibration ............................................................................................................................................. 2-18 Parameter Overrides: ............................................................................................................................. 2-19

    Prover Request .................................................................................................................................. 2-19 Prover/Master Temperature Override ................................................................................................ 2-19 Prover/Master Pressure Override ....................................................................................................... 2-19 Prover Shaft Temperature Override ................................................................................................... 2-19 Prover Gravity Override .................................................................................................................... 2-19 Meter Temperature Override ............................................................................................................. 2-19 Meter Pressure Override .................................................................................................................... 2-19 Equilibrium Pressure Override .......................................................................................................... 2-19 Alpha T E-6 Override ........................................................................................................................ 2-19 CTL or CPL Override ........................................................................................................................ 2-19 Meter Cumulative Total Override ..................................................................................................... 2-19 Reset Prove Data Area ....................................................................................................................... 2-20 Date and Time ................................................................................................................................... 2-20 Clear System ...................................................................................................................................... 2-20

    HISTORICAL DATA ........................................................................................................................... 2-21 CAPTURE AND STORE .................................................................................................................. 2-21 View Report in Modubs Address Format .......................................................................................... 2-23 Viewing previously captured reports ................................................................................................. 2-24 Printing Reports ................................................................................................................................. 2-24

    CHAPTER 3: Flow Equations ..................................................................................................................... 3-1 Porve Equation ........................................................................................................................................ 3-1

    Prove IV Flow Rate(Pipe Prover) ........................................................................................................ 3-1 Prove IV Flow Rate(Compact Prover) ................................................................................................ 3-1 Prove IV Flow Rate(Master Meter Proving) ....................................................................................... 3-1 Corrected Prover Volume(Pipe or Compact Prover) ........................................................................... 3-1 Prove Volume (IVP - Mater Meter Proving) ....................................................................................... 3-1 Corrected Prover Volume(Mater Meter Proving) ................................................................................ 3-1 Meter Volume (IVM) .......................................................................................................................... 3-1 Corrected Meter Volume ..................................................................................................................... 3-1 Prove Meter Factor .............................................................................................................................. 3-1

    DENSITY EQUATIONS ........................................................................................................................ 3-4 Sarasota Density GM/CC – US Unit ................................................................................................... 3-4 UGC Density GM/CC (US UNIT) ...................................................................................................... 3-5 Solartron Density GM/CC – US UNIT ............................................................................................... 3-6

    CHAPTER 4: MODBUS DATA ................................................................................................................. 4-1 MODBUS PROTOCOL .......................................................................................................................... 4-1

    TRANSMISSION MODE ................................................................................................................... 4-1 ASCII FRAMING ............................................................................................................................... 4-1 RTU FRAMING .................................................................................................................................. 4-1 FUNCTION CODE ............................................................................................................................. 4-2 ERROR CHECK ................................................................................................................................. 4-2 EXCEPTION RESPONSE .................................................................................................................. 4-2 BROADCAST COMMAND ............................................................................................................... 4-2 MODBUS EXAMPLES ...................................................................................................................... 4-3 MODBUS TABLE .............................................................................................................................. 4-5 SFC332P - Alarm Status ................................................................................................................... 4-41

  • Date: 3/10/2021

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-1

    Date: 3/10/2021

    CHAPTER 1: QUICK START

    Introduction: The SFC332 Prover Flow Computer was designed after careful consideration to our customers in all sectors

    of the oil and gas industry. It was built to address the different needs for refineries, chemical plants, gas

    processing plants, offshore platforms, pipeline and transmission, remote gas wells, and storage caverns.

    The focus has been to bring the different needs and requirements of these specialized industries into one

    hardware platform and therefore reducing the spare parts requirements, the training process, calibration,

    and overall cost of ownership. We believe the SFC332 Prover Flow Computer has delivered and met the

    design intentions.

    The SFC332 Prover Flow Computer combines the following features:

    User Friendly

    Flexible

    Easy to understand and configure

    Rugged

    Economical to install and maintain

    Accurate

    We hope that your experience with the SFC332P Prover Flow Computer will be a simple pleasant

    experience, not intimidating in any way.

    General Description:.

    The SFC332P Prover Flow Computer has a host of inputs and outputs beyond its turbine inputs: 4

    additional analog inputs, two 4 wire RTD inputs, 2 analog outputs, RS-232 and RS-485 with Modbus

    protocol, 4 status inputs and 5 switch and pulse outputs.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-2

    Date: 3/10/2021

    Technical Data:

    POWER

    VOLTAGE RANGE 12-30 VDC

    WATTAGE 4 WATT

    OPERATING CONDITIONS

    TEMPERATURE - 40 TO 185 °F

    HUMIDITY 100%

    HOUSING NEMA 4X CLASS 1 DIV. 1

    FEATURES

    DISPLAY PLASMA 2 LINES 16 CHARACTER

    PROCESSOR 32-BIT MOTOROLA 168332 @ 16.7 MHz

    FLASH ROM 4 MB @ 70 NANO SECONDS

    ROM 2 MB @ 30 NANO SECONDS

    FREQUENCY INPUT 3 CHANNELS 0 - 5000 Hz >70 mV FOR SIN WAVE > 6 VOLTS FOR SQUARE WAVE

    ANALOG INPUT FOUR 24-BIT CHANNELS EXPANDABLE TO 6

    RTD INPUTS 2 CHANNELS 4 WIRES

    ANALOG OUTPUT 2 CHANNELS 12 BIT SINGLE ENDED

    DIGITAL OUTPUT OUTPUTS 1 & 2 PULSE/SWITCH 0.5 AMPS RATING OUTPUTS 3 TO 5 ARE SWITCH OUTPUTS 0.25 AMPS RATING

    STATUS INPUTS 4 ON/OFF TYPE SIGNAL

    ALL INPUTS AND OUTPUTS ARE OPTICALLY ISOLATED

    SERIAL 1 RS485 @ 9600 BAUDS VARIABLE

    1 RS232 @ 9600 BAUDS VARIABLE

    COMMUNICATION PROTOCOL MODBUS

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-3

    Date: 3/10/2021

    Parts List

    Part Description

    332-01P Controller (CPU) Board for SFC332/1000 w/prover option.

    332-02 Terminal (BP) Board for SFC332/1000.

    332-03 Analog Board for SFC332/1000.

    332-04 Display (LCD) for SFC332/1000.

    332-05 Rosemount Interface Board for SFC1000.

    332-06 Prover Option for SFC332/1000.

    332-07 Enclosure for SFC332/1000.

    332-08 Mounting Bracket w/captive screws for SFC332/1000 Boards.

    332-09 Adapter between SFC1000 and Rosemount 205.

    332-10 Center portion of housing for SFC332/1000 enclosure.

    332-11 Glass Dome Cover for SFC332/1000 Enclosure.

    332-12 Blank Dome Cover for SFC332/1000 Enclosure.

    332-13 O'ring for SFC332 Enclosure.

    332-14 External I/O Expansion.

    332-15 Battery Replacement for SFC332/1000.

    332-16 1/2 Amp 250V Fuse for SFC332/1000.

    332-17 EPROM for SFC332/1000 (set of two).

    RS232 External RS232 Connection for all models.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-4

    Date: 3/10/2021

    SFC332P Prover Flow Computer: Dimensions

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-5

    Date: 3/10/2021

    Starting and Installing theWindow Software: First make sure your computer has the minimum requirements to install Dynamic’s Dynacom software.

    System Minimum Requirements

    In order to install this software product the following requirements must be met:

    Windows Operating System (Win95, Win98, Win98SE, win2000, WinNT, WinXP, Vista, Windows 7, Windows 8, Windows 10)

    For Windows NT, 2000, XP or Vista: Administrator level access to create an ODBC system DNS.

    Minimum disk space available: 16 MB.

    1 Serial Communication Port

    If your computer meets these requirements, you can run the setup file downloaded from our website

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-6

    Date: 3/10/2021

    Website - DFC Configuration Software

    Step 1. Go to our website WWW.DYNAMICFLOWCOMPUTERS.COM

    Step 2. Click on the “Downloads”

    Step 3. Select either Windows or

    DOS software based on Step 2.

    Step 4. On the new screen

    presented to you click on the

    application that you are trying to

    download. Once you hit the link it

    will ask you if you want to run or

    save the file in your computer.

    Select SAVE. (See illustration 1)

    http://www.dynamicflowcomputers.com/

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-7

    Date: 3/10/2021

    Step 5. The file will start to

    transfer to your computer. The

    download time depends on your

    Internet connection speed and the

    type of application that being

    downloaded.

    Step 6. When the download if

    finish. Press the OPEN button to

    start the setup process. (See

    Illustration)

    Step 7. Follow the steps in the

    application setup.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-8

    Date: 3/10/2021

    Getting acquainted with the flow computer wiring:

    To wire the flow computer, get familiar with the drawings provided in the software’s Wiring menu. Use

    the arrow left and right keys to move to and select Wiring, then press .

    Back terminal wiring: The back terminal wiring indicates the overall positions of the terminal plugs and their functions. Though

    the back panel’s jumpers are also shown, refer to the next drawing, “Back Panel Jumpers”, for information

    on their settings and functions.

    The Flow Computer receives its power via the two topmost pins on Terminal P1, on the left of the terminal

    board. Also on Terminal P1 are, from top to bottom, inputs from the two turbines and the RS-485 serial

    connection.

    To the right (P4), from top to bottom, is status input 1, density frequency input, and switch output 1 and 2.

    Terminal P3, at the lower bottom, handles analog inputs and outputs. These are, in order from right to left,

    analog inputs 1-4 and analog outputs 1 and 2.

    Terminal P5, top middle, is the RTD terminal block, "100 platinum RTD input".

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-9

    Date: 3/10/2021

    Back Panel Jumper In this illustration, a jumper is “ON” when the jumper block is used to connect the jumper’s to wire prongs.

    “OFF” means the jumper block is completely removed or attached to only one of the two wire prongs.

    Note: R11 and R3 could have a vertical orientation instead of a horizontal orientation on certain Flow Computer models.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-10

    Date: 3/10/2021

    Memory Jumper

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-11

    Date: 3/10/2021

    Steps to clear memory through removing the memory jumper

    (1) Turn off the power, move the jumper to the next two pins, wait for 5 seconds

    (2) Put the jumper back

    Memory cleared and Flow Computer ID is set to 1, 9600 baud rate, RTU mode

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-12

    Date: 3/10/2021

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-13

    Date: 3/10/2021

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-14

    Date: 3/10/2021

    INPUT/OUTPUT: Assigning and Ranging Inputs

    Input/Output Assignment We will now configure your SFC332P Computer’s inputs and outputs. The flow computer allows the user

    to configure the inputs and outputs. The flow computer will not use the unassigned inputs.

    How to assign a transmitter to an I/O point

    1 Click “Configure Device”, configuration menu is prompted

    2 On configuration menu, click “Input Assignment”

    3 Enter assignments for DP, temperature, pressure, density and spare inputs.

    4 Assignment (1-n). Assignments 1-4 are analog inputs attached to terminal of the

    back panel. These inputs accept 4-20mA or 1-5 volts input and are suitable for temperature,

    pressure, density, or spare inputs. An assignment 5 is strictly RTD (temperature) input only

    for the meter, densitometer. Assignment 7 indicates a density frequency input; it is assigned

    automatically once you choose live density frequency input in the setup menu at density type

    Assignment 10 (module 1) is for Rosemount multi-variable module only. DP, pressure, and

    temperature for the meter can be assigned. When a frequency type primary element is hooked

    to the flow computer, the Multi .Variable pressure and temperature can be used and the DP

    becomes a spare input that could be assigned for strainer differential.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-15

    Date: 3/10/2021

    Ranging the Transmitter Inputs:

    1. Enter the range values: after assigning the inputs scroll down the transducer inputs

    assignment menu to scale the 4-20mA. Enter the value at …@4mA and …@20mA. Enter both

    values similar to the way the transmitter is ranged. 1-5 volts is equivalent to 4-20mA. Enter

    the 1 volt value at the 4mA, and 5 volt value at 20mA.

    2. Enter the high and low limits: high limits and low limits are simply the alarm points in

    which you would like the flow computer to flag as an alarm condition. Enter these values

    with respect to the upper and lower range conditions. Try to avoid creating alarm log when

    conditions are normal. For example: If the line condition for the pressure is between 0 to 500

    PSIG, then you should program less than zero for low-pressure alarm, and 500 or more for

    high-pressure alarm. High and low limits are also used in the scale for the Modbus variables.

    3. Set up the fail code: Maintenance and Failure Code values tell the flow computer to

    use a default value in the event the transmitter fails. The default value is stored in

    Maintenance. There are three outcomes: the transmitter value is always used, no matter

    what (Failure Code = 0); the Maintenance value is always used, no matter what

    (Failure Code = 1); and the Maintenance value is used only when the transmitter’s

    value indicates that the transmitter has temporarily failed (Failure Code = 2).

    RTD inputs will skip 4-20 mA assignment because RTD is a raw signal of 50 (ohms) to 156. Readings

    beyond that range require a 4-20 mA signal to the flow computer.

    Density coefficients for raw frequency inputs are programmed in this menu. The menu will only show

    parameters relevant to the live density selected (i.e., Solartron or UGC, etc.).

    NOTE: Solartron density input requires values in US units, i.e. PSI, and F. Request a calibration sheet in US units from the densitometer

    manufacturer.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-16

    Date: 3/10/2021

    WIRING: Wiring to the flow computer is very straightforward and simple. But still it is very important to get familiar

    with the wiring diagram.

    Wiring of Analog Inputs: Use your arrow keys to move the cursor to the menu item Wiring and then use up/down arrow keys to

    select Analog Wiring. Press ENTER. Typical wiring for analog inputs 1 and 2 are shown in the

    drawing. Analog inputs 3 and 4 are to the left of analog 1 and 2. Note that the analog input has only one

    common return, which is the -ve signal of power supply powering the transmitters.

    When wiring 1-5 volts, make sure to calibrate the flow computer for the 1-5 volt signal because the flow

    computer calibration defaults for the 4-20 mA which is different from the 1-5 volts. JP5 must be cut for 1-

    5 volt inputs. The jumpers for analog 1-4 are in order from right to left. It is possible to cut the first two

    jumpers for analog 1 & 2 in for 1-5 volts signal and have analog in 3 & 4 as 4-20 mA signal. Signal line

    impedance provided by our flow computer is less than 250. Therefore, when using a transmitter that

    requires a minimum of 250 resistance in the loop, an additional resistor at the flow computer end needs to

    be installed in series with the 4-20 mA loop in order to allow the hand held communicator to talk to the

    transmitter.

    NOTE: The 4-20mA or 1-5 volt DOES NOT source power to the transmitters. You can use the DC power feeding the flow computer to power the 4-20mA loop IF that power supply is FILTERED.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-17

    Date: 3/10/2021

    Wiring the analog inputs 1-4 :

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-18

    Date: 3/10/2021

    Wiring of Analog Inputs 5,6 : The ‘Flow Computer’ can be configured as ‘6 analog inputs’ or ‘4 analog inputs and 2 RTD inputs’

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-19

    Date: 3/10/2021

    Wiring of RTD Input : Go to the wiring menu where it says RTD and press . The flow computer shows wiring to RTD 1

    and RTD 2. 100 platinum can be used; a temperature range of -43F to +300F can be measured. RTD 1

    is to the right where P5 designation is. In the figure below notice that each side of the RTD requires two

    wire connections. When using less than 4 wires a jumper must be used to make up for the missing lead.

    Internal excitation current source generated is approximately 7 mA.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-20

    Date: 3/10/2021

    Wiring of Analog Output: The analog outputs are located on the left side of P3 connector. Go to the wiring diagram and scroll down

    to Analog Output and press . Wiring diagram shows typical Analog output wiring. Notice

    that analog output will regulate 4-20 mA current loop but DOES NOT source the power for it. External

    power is required.

    ASSIGNING /RANGING THE 4-20MA ANALOG OUTPUTS :

    Go to the I/O assignment main menu and scroll to Analog Output Assignment. Press

    . A selection menu is prompted. Select the analog output number and then enter what the 4 mA output will indicate and the 20 mA. Make sure that the 20 mA assignment value exceeds the upper range

    limit of what you assigned the Analog output for, otherwise the analog output will not update beyond 20

    mA.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-21

    Date: 3/10/2021

    Turbine Input Wiring Scroll to Turbine under Wiring and press . Two drawings above each other will show

    typical wiring for turbine meter 1 and turbine meter 2. When dual pick ups from the same turbine are

    connected, use the inputs for turbine 1 for pickup 1 and turbine 2 for the second pickup coil. When

    connecting sine wave directly from the pickup coil make sure the distance from the pickup coil to the flow

    computer is very short–less than 50 feet with shielded cable. In the event there is presence of noise, the

    distance must be shortened. When connecting sine wave signal, the R11 jumper for meter 1 must be

    installed and R3 jumper for meter 2 must be installed. (JP3 and JP2 must be off when using sine wave).

    On the other hand, when using square wave, the square wave signal can be sinusoidal but has to be above 5

    volts peak to peak with less than 0.4 volts offset in order for the flow computer to read it. R11 and R3 must

    be off and JP3 on for meter 1; JP2 must be on for meter 2.

    Note: When connecting square wave input, the JP3 and JP2 connect the turbine return to the flow computer power return. Therefore, signal polarity is very important. Reverse polarity could result in some damage or power loss. When sine wave is used the signal polarity is usually of no significance.

    The turbine input is immediately under the power input on terminal P1. The third pin down from the top is

    Turbine/PD "minus", and below it is Turbine plus. The second pulse input for Turbine/PD meter 2 or the

    second pickup coil is below turbine one input on P1. The fifth pin down from the top is turbine 2 "minus"

    signal and below it is Turbine/PD 2 plus signal.

    Note: R11 and R3 are oriented vertically in some flow computers.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-22

    Date: 3/10/2021

    TURBINE - S INE WAVE

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-23

    Date: 3/10/2021

    TURBINE -SQUARE WAVE

    Note: R11 and R3 are oriented vertically in some flow computers.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-24

    Date: 3/10/2021

    Turbine Wiring for Passive (dry contact) Pulse Generators

    Some mass flow meters have pulse outputs that do not provide power but instead require external power, they are referred to as passive outputs, dry outputs, open collector, etc. (For example the Krohne UFM 3030 Mass meter). In these cases the wiring should be as shown on the below diagram. The pull up resistor can be adjusted to limit the current sink by the Mass meter. For Turbine Input 1 JP3 must be ON and R11 OFF and if using Turbine Input 2 then JP2 must be ON and R3 OFF.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-25

    Date: 3/10/2021

    Density Input Wiring When using a live densitometer input with frequency signal, the signal can be brought into the Flow

    Computer in its raw form. The Flow Computer accepts a sine wave or square with or without DC offset.

    Example for density wiring can be seen in the wiring diagram. Use the arrow keys to get to Wiring |

    Density and press . Find two drawings, one with barrier and the other without. Barriers are

    used for area classification. Notice that the RTD wiring is also drawn to show how to hook the density

    RTD signal.

    Note: When wiring the density input polarity is of significance and reverse polarity could result in some damage or power loss. The density signal is on connector P4, the third and fourth pin down from the top. The third pin down is density plus, the fourth down is density minus. When Density input is 4-20mA it should be connected as a regular 4-20mA

    signal to the analog input and not the density frequency input.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-26

    Date: 3/10/2021

    RS-232 Connection The RS-232 is not located on the terminal board. The RS-232 is a green 5 pin terminal block with screw

    type connector located on the display side of the enclosure. Scroll to Wiring | New RS-232 and press

    . Termination jumpers for the RS-232 are located at the top corner of the board on the same side of the RS-232 connector. The two jumpers at the top are for terminating the transmit line and below it is

    the receive line.

    The RS-232 port can be used for printing reports, Modbus communication, or interfacing to the

    configuration program. If the port is configured as printer port in the flow computer communication setup,

    then reports get printed (i.e. interval and daily reports

    Note: Twisted shielded cable is required.

    WARNING: When the RS-232 terminal is used with a modem, external protection on the phone line is required.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-27

    Date: 3/10/2021

    RS-485 Connection RS-485 wiring is shown in the wiring diagram under RS-485. The RS-485 termination jumper is JP4

    located on the back terminal. The maximum distance when 18 gauge wire is used is 4000 feet.

    Note: Twisted shielded cable is required.

    WARNING: When the RS-485 terminal is used, external transient protection and optical isolation is required, especially for long distance wiring.

    RS485/232 Adapter

    Dynamic recommends B&B Electronics. We generally use Model 485D9TB, which is a port power converter requiring only a 2-Wire connection. The 485D9TB has a terminal block which makes the wiring more convenient and provides the option of external 12V power for low power serial ports. Model 485SD9R can also be used, but it has a DB9 terminal which requires additional cables. With Model 485SD9R the pins that connect to the flow computer are pin 3 on the DB9 to TX on the flow computer and pin 8 on the DB9 goes to RX on the flow computer. For a USB to RS485 converter, we recommend Model USTL4 which is also port powered and supports half and full duplex networks.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-28

    Date: 3/10/2021

    Wiring of Status Inputs There is one status input standard and an optional three more on the back of the CPU board. The standard

    status input is shown in the wiring diagram under Status Input. It has 4 volts of noise hysteresis,

    with a trigger point of 5 volts and an off point of 1 Volt. Status inputs 2, 3, and 4 require the I/O expansion

    connector and its wires be installed; refer to wiring drawing IO-Exp. Connection numbers 6, 7, and 8 are

    the status in (positive) for inputs 2, 3, and 4, respectively, and 11 is the return for all three inputs.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-29

    Date: 3/10/2021

    Wiring of Switch/Pulse Outputs Scroll to Switch output under Wiring and press . The wiring diagram shows switch 1

    and 2 and the return. Please note that switches 3, 4, and 5 cannot be used for pulse output; switches 1 and 2

    can be used for pulse or switch output. See also I/O Expansion. Notice that the switch outputs are

    transistor type outputs (open collector type with maximum DC rating of 350 mA continuous at 24 VDC)

    and require external power.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-30

    Date: 3/10/2021

    I/O Expansion The I/O expansion is 16-pin connector next to the RS-232 terminal. Eleven pins of the 16-pin connector

    are utilized. When the flow computer is ordered with the I/O expansion feature, the wires and the plug are

    provided with the flow computer. There will be 11 wires with the wire number tag at the outer end of the

    wire. The tag will indicate the wire number. The following is the sequence for the wires. On the top right

    edge of the connector towards the top outer side of the CPU board is pin 1, across from it is pin 9.

    Connection Purpose Comments 1 detector switch 1 Requires prover option CPU to operate.

    Rating: 5-36 VDC 2 detector switch 2

    3 switch output 3 Maximum rating: 75 mA @24 volts Range: 5-36 VDC

    4 switch output 4

    5 switch output 5

    6 status input 2

    Rating: 6-36 VDC 7 status input 3

    8 status input 4

    9 Return: detector switches 10 Return: switches 3, 4, 5

    11 Return: status 2, 3, and 4

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-31

    Date: 3/10/2021

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-32

    Date: 3/10/2021

    Prover/Expansion

    Connection No. Purpose Comments

    1 detector switch 1 Requires prover option CPU to operate. Rating: 5-36 Vdc 2 detector switch 2

    3 switch output 3 Maximum rating: 75mA @24 volts Range: 5-36 Vdc

    4 switch output 4

    5 switch output 5

    6 status input 2

    Rating: 6-36 Vdc 7 status input 3

    8 status input 4

    9 Return: detector switches 10 Return: switches 3, 4, 5

    11 Return: status 2, 3, and 4

    12 RS232 TX

    13 RS232 RX

    14 RS232 RTS

    15 RS232 ret

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-33

    Date: 3/10/2021

    Wiring of Brooks Prover:

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-34

    Date: 3/10/2021

    Wiring of Calibron Prover:

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-35

    Date: 3/10/2021

    Calibration Through Windows™ Software

    Calibrations are performed under Calibration. Select inputs to be calibrated, and then select full,

    single, offset calibration method.

    Analog Input of 4-20mA or 1-5 Volt Signal

    OFFSET CALIBRATION :

    For simple offset type calibration simply induce the signal into the analog input and make sure the

    SFC332P is reading it. After you verify that the SFC332P recognized the analog input, enter the correct

    mA reading, and then click OK. The offset type calibration is mainly used when a small offset adjustment

    needs to be changed in the full-scale reading. The offset will apply to the zero and span. Offset is the

    recommended method for calibrating the temperature input.

    FULL CALIBRATION METHOD:

    To perform full calibration, be prepared to induce zero and span type signal.

    1. Induce the low end signal i.e. 4mA in the analog input.

    2. Click inputs to be calibrated under calibration menu, click full calibration, enter the first point

    - the analog input value i.e. 4mA, and then click OK button.

    3. Now be ready to enter the full-scale value. Simply induce the analog signal and then enter

    the second value i.e. 20mA, and then click OK button

    4. Induce live values to verify the calibration.

    TO USE DEFAULT CALIBRATION

    1. Select Analog Input 2. Select Reset calibration method 3. Now verify the live reading against the flow computer reading

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-36

    Date: 3/10/2021

    RTD Calibration: RTD Calibration is a 2-step process. The first step is a one time procedure to verify transducer linearity and is done at the time the meter is being setup. The second step is the routine calibration sequence. Step 1 – Linearity Verification

    1- Use a Decade box with 0-150 °F settings. 2- Connect RTD cable to this resistive element for verification of linearity. Verify low and high

    points. It must be within ½ degree. 3- Connect the actual RTD element and compare with a certified thermometer. 4- If not within ½ degree do a Full Calibration (See Full Calibration below). If problem persists

    verify other elements such as RTD Probe, connections, shield, conductivity of connectors, etc.

    The purpose of the above procedure is to verify zero and span and make sure that the two points fall within the expected tolerance. Step 2 – Routine Calibration Once Linearity has been verified through Step 1, the routine calibration procedure is reduced to simply connecting the actual RTD and doing an offset point calibration (see offset calibration below). Calibration after that will be simple verification for the stability of the transmitter. If it drifts abnormally then you need to verify the other parts involved. Calibration Procedures through Windows™ Software At the top menu, go to Calibration and Select RTD Input. RESET TO DEFAULT CALIBRATION 1. Select Reset calibration method 2. Now verify the live reading against the flow computer reading OFFSET CALIBRATION: 1. Select offset calibration method. 2. Induce a live value and wait for 10 seconds for the reading to stabilize. Then enter the live value. The value entered must be in Ohm only. 3. Now verify the live reading against the flow computer reading FULL SCALE CALIBRATION: 1. Prepare low range resistive input (i.e., 80 Ohm.) and High range resistive input (i.e., 120. Ohm). 2. Go to the calibration menu and select RTD full calibration method. Induce the low end (80

    Ohm.) resistive signal and then wait 10 seconds, enter live value in Ohm, and click OK button. 3. Induce the High range signal (120 Ohm.) and wait 10 seconds, then enter 120 Ohm and click OK button. 4. Now verify the live reading against the flow computer reading.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-37

    Date: 3/10/2021

    Calibration of Analog Output: Follow the following steps to calibrate the analog output against the end device

    1. Go to the calibration menu, select analog output, and then select method. Full calibration will

    cause the flow computer to output the minimum possible signal 4 mA. Enter the live output

    value reading in the end device i.e. 4 mA and click OK button. Now the flow computer will

    output full scale 20 mA. Enter the live output i.e. 20 then click OK button.

    2. Now verify the output against the calibration device.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-38

    Date: 3/10/2021

    Multi-Variable Transmitters (Model 205) – DP and Pressure Calibrations are performed under Calibration. . Select inputs to be calibrated, and then select full,

    single, offset calibration method.

    OFFSET CALIBRATION

    1. Induce live value for pressure or DP.

    2. Select Multivariable DP or pressure.

    3. Select offset calibration method, enter offset, and click OK button.

    4. Now read induce live values to verify the calibration.

    FULL SCALE CALIBRATIO N

    1. Induce live value for pressure or DP.

    2. Select Multivariable DP or pressure

    3. Select full calibration method

    4. Induce the low range signal, enter the first point, and then click OK button.

    5. Induce the high range signal, enter the second point, and then click OK button.

    6. Now verify the live reading against the flow computer reading.

    TO USE DEFAULT CALIBRATION

    1. Select Multivariable DP or pressure

    2. Select Reset calibration method

    3. Now verify the live reading against the flow computer reading

    While doing calibration before downloading any of the calibrated values, it is a good practice to verify that

    the Flow Computer close reading to the induced value.

    The DP reading must be re-calibrated for the zero offset after applying line pressure.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-39

    Date: 3/10/2021

    Multi-Variable Transmitters (Model 205) –Temperature

    RTD Calibration is a 2-step process. The first step is a one time procedure to verify transducer linearity and is done at the time the meter is being setup. The second step is the routine calibration sequence. Step 1 – Linearity Verification

    1. Use a Decade box with 0-150 °F settings. 2. Connect RTD cable to this resistive element for verification of linearity. Verify low and high points. It must be within ½ degree. 3. Connect the actual RTD element and compare with a certified thermometer. 4. If not within ½ degree do a Full Calibration (See Full Calibration below). If problem persists verify other elements such as RTD Probe, connections, shield, conductivity of connectors, etc.

    The purpose of the above procedure is to verify zero and span and make sure that the two points fall within the expected tolerance. Step 2 – Routine Calibration Once Linearity has been verified through Step 1, the routine calibration procedure is reduced to simply connecting the actual RTD and doing an offset point calibration (see offset calibration below). Calibration after that will be simple verification for the stability of the transmitter. If it drifts abnormally then you need to verify the other parts involved. Calibration Procedures through Windows™ Software At the top menu, go to Calibration and Select RTD Input. RESET TO DEFAULT CALIBRATION 1. Select Reset calibration method 2. Now verify the live reading against the flow computer reading OFFSET CALIBRATION: 1. Select offset calibration method. 2. Induce a live value and wait for 10 seconds for the reading to stabilize. Then enter the live value. The value entered must be in Degrees only. 3. Now verify the live reading against the flow computer reading FULL SCALE CALIBRATION: 1. Prepare low range resistive input (i.e., 80 Ohm.) and High range resistive input (i.e., 120. Ohm). 2. Go to the calibration menu and select RTD full calibration method. Induce the low end (80 Ohm.) resistive signal and then wait 10 seconds, enter the equivalent temperature in degrees, and click OK button. 3. Induce the High range signal (120 Ohm.) and wait 10 seconds, then enter the temperature degrees equivalent to 120 Ohm and click OK button. 4. Now verify the live reading against the flow computer reading.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-40

    Date: 3/10/2021

    Verifying Digital Inputs and Outputs Use the diagnostic menu. A live input and output is displayed. On the top of the screen pulse inputs and

    density frequency input are shown. Compare the live value against the displayed value on the screen.

    Failure to read turbine input could be a result of a bad preamplifier or the jumper selection for sine and

    square wave input are not in the correct position. Refer to wiring diagram View | Wiring

    Drawing| Turbine for proper turbine input wiring. Density input can be sine or square wave with

    or without DC offset. Minimum accepted signal has to be greater than 1.2 volt peak to peak. Status input

    is shown below the frequency input to the left of the screen. When the status input is on, the live diagnostic

    data will show ON. Minimum voltage to activate the status is 6 volts with negative threshold of 2 volts.

    The switch outputs are open collector and require external voltage.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-41

    Date: 3/10/2021

    Calibration Through DOS Software

    Analog Input 4-20mA or 1-5 volt signal:

    Calibrations are performed under I/O | Calibration.

    Use the arrow keys to scroll to 0 followed by pressing the function key.

    Now verify the live reading against the flow computer reading.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-42

    Date: 3/10/2021

    Verifying Digital Inputs and Outputs Use the diagnostic menu. Scroll down by using the arrow keys to Diag | Diagnostic Data and

    press . A live input and output is displayed. On the top of the screen pulse inputs and density frequency input are shown. Compare the live value against the displayed value on the screen. Failure to

    read turbine input could be a result of a bad preamplifier or the jumper selection for sine and square wave

    input are not in the right position. Refer to wiring diagram Wiring | Turbine for proper turbine

    input wiring. Density input can be sine or square wave with or without DC offset. Minimum accepted

    signal has to be greater than 1.2 volt peak to peak. Status input is shown below the frequency input to the

    left of the screen. When the status input is on, the live diagnostic data will show ON. Minimum voltage to

    activate the status is 7 volts with negative threshold of 2 volts. Switch outputs are to the right of the status

    inputs. Press in the diagnostic menu to manually activate the switch outputs to the on and off

    position. After the screen freeze the cursor will point to switch output one. Use the space bar to toggle the

    switch on/off and the key to advance to the next switch. Press to exit diagnostic screen. The switch outputs are open collector and require external voltage.

  • Dynamic Flow Computers SFC332P Prover Manual Quick Start — 1-43

    Date: 3/10/2021

    Pulse Interpolation:

    When interfacing with Small Volume Provers (SVP), Dynamic uses Dual chronometry as the pulse

    interpolation method which is required for provers providing less then 10,000 unaltered pulses.

    Pulse interpolation is a pulse counting technique used to calculate the total number of meter pulses between

    two detectors, including fractional pulses. A conventional pipe prover requires a minimum of 10,000

    unaltered pulses to ensure an accurate test. This minimum is based upon the assumption there is a potential

    error of 1 pulse each time a detector is passed, which equates to a 2-pulse error per run. The value of 2 pulses

    per 10,000 equates to a potential 0.02 % error. SVPs require less than 10,000 pulses, which would equate to

    a potential error greater than 0.02 %.

    To enhance the meter's pulse output, Small Volume Provers use pulse interpolation for fractional meter pulse

    counting. To interpolate fractional meter pulses or to mathematically interpolate partial pulses, Dynamic uses

    the double chronometry method for this purpose since it is the method most widely used.

    Double Chronometry:

    Double chronometry pulse interpolation increments time precisely every millisecond. A high frequency

    master oscillator operates two time counters referred to as Time A and Time B. Time A starts when the first

    detector switch is tripped, and Time B starts with the leading edge of the first flow meter pulse after Time A

    has started. Time A is stopped when the final detector is tripped and Time B is stopped with the leading edge

    of the first flow meter pulse after Time A has stopped (see figure below). Using the ratio of Time A and Time

    B allows for counting of a fraction of the flow meter pulse.

    D

    C

    BTime

    ATimeK

    _

    _

    K = K Factor, or counts per unit Volume, from the flow meter.

    A = Time for displaced volume.

    B = Time for whole meter pulses.

    C = Total number of whole meter pulses.

    D = Displaced Volume

    Double Chronometry Figure

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-1

    Date: 3/10/2021

    CHAPTER 2: Data Entry

    and Configuration Menus

    The SFC332P software is constructed around a menu-driven organization

    Configuration File through Window Program

    New Create a new file to store all the programmed information for one Flow Computer. After a file is opened it

    becomes the currently active file, its contents can be viewed and its parameters can be edited.

    Open Use this function to open an existing configuration file. After a file is opened it becomes the currently

    active file, its contents can be viewed and its parameters can be edited. When this function is chosen a list

    of existing configuration files is displayed. Select the file to be opened.

    Close Close or exit configuration file.

    Save When permanent modifications are performed on a file, user must save the new changes before exiting the

    program, or proceeding to open a different file.

    Save As Use Save As to save the parameters in the currently active file (that is, the parameter values currently being

    edited) to a new file. The original file will remain in memory.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-2

    Date: 3/10/2021

    VIEW

    View Drawings Select the wiring diagram to be displayed. (See details in chapter 1)

    Back Panel

    Analog Input

    RTD

    Analog Output

    Status Input

    Switch Output

    Turbine

    Densitometer

    RS 232

    RS 485

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-3

    Date: 3/10/2021

    TOOLS

    Com Settings

    PORT - COMMUNICATION PORT NUMBER (1,2,3,4)

    Enter the PC port used to communicate with the Flow Computer.

    Baud Rate

    Note: this parameter must be set the same for both the PC and the Flow Computer for communication to occur.

    Baud rate is defined as number of bits per second. The available selections are 1200, 2400, 4800, 9600, or

    19200.

    Parity

    Note: this parameter must be set the same for both the PC and the Flow Computer for communication to occur.

    RTU - NONE

    ASCII - EVEN or ODD

    Set the parity to match the Modbus Type.

    Data Bits

    Options available: 5, 6, 7, or 8. Generally used: 8 for RTU mod, 7 for ASCII mode.

    Stop Bits

    Options available: 1, 1.5,or28. Generally used: 1.

    Modbus Type

    Note: this parameter must be set the same for both the PC and the Flow Computer for communication to occur.

    The Modbus Communication Specification is either Binary RTU or ASCII.

    Unit ID Number

    The Unit ID Number is used strictly for communication purposes; it can take any value from 1 to 247. Only

    one master can exist in each loop.

    Note: Do not duplicate the Unit ID number in a single communication loop! This situation will lead to response collisions and inhibit communications to units with duplicate ID numbers.

    Time Out

    The amount of time in seconds the program will wait for an answer from the flow computer.

    Retry Times

    Retry times for the program to communicate with the flow computer in case of timeout.

    Auto Detect Sett ings

    Click this button and the configuration program will attempt to communicate with a single Flow Computer

    at different baud rates and formats.

    Failure to communicate can occur because of a wiring problem, wrong PC port selection, communication

    parameter mismatch between PC and Flow Computer (Modbus type, parity, baud rate, etc.) or lack of

    power to the Flow Computer. To use this feature, the user must insure that only one Flow Computer is

    connected to the PC. More than one Flow Computer in the loop will cause data collisions and

    unintelligible responses.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-4

    Date: 3/10/2021

    Meter Configuration

    METER DATA SETTINGS

    Company Name

    Up to 20 characters. The company name appears in the reports.

    Select Flow Rate (0=Hour,1=Day,2=Min.)

    The flow rate will be based on daily basis, hourly, or minute.

    Flow Rate Average Second

    Flow rate averaged 1-5 is for fluctuating flow rate conditions. This number is used to perform averaging on the current flow rate by adding correct to the previous seconds flow rate, and then display an averaged smoothed number. Only low-resolution pulse meter requires this option.

    Atmospheric Pressure

    Local pressure or contracted atmospheric pressure. (i.e. 14.73 PSIA in US unit or 1.01325 Bar or 101.325 KPA in metric unit)

    Old Table Gravity Unit – SG or API (Available for version 1.17.12 or newer)

    Select SG to use old table 23/24 or select API to use old table 5/6.

    METER DATA F ILE

    Meter ID

    Up to 8 characters. This function will serve as meter tag.

    Lease Name

    Up to 16 characters.

    Serial Number

    Up to 8 characters.

    Model Name

    Up to 8 characters.

    Meter Location

    Up to 16 characters.

    Meter Size

    The size of meter

    K Factor

    The number of pulses per unit volume (i.e. 100 pulses/barrel in Volume Unit, 100 pulses/lb in mass unit)

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-5

    Date: 3/10/2021

    Table

    US Unit Selection

    Table

    0

    1 6A

    2 5B/6B

    3 6B

    4 23A/24A

    5 24A

    6 23B/24B

    7 24B

    8 OLD23/24

    9 OLD24

    10 24C

    11 6C

    12 ASTM1550-Butadiene

    13 NEW23/24

    14 NEW24

    US Units Selection

    Table Range Limits

    0 5A/6A

    1 6A

    2 5B/5B

    3 6B

    4 23A/24A

    5 24A

    6 23B/24B

    7 24B

    8 OLD 23/24, 5/6 0.37 < SG < 1.0, -50 < Deg.F < 200

    9 OLD24, 5/6 0..35 < SG < 1.0, -50 < Deg.F < 200

    10 24C

    11 6C

    12 ASTM 1550-Butadiene

    13 NEW23/24 0.35 < SG < 0.688, -50 < Deg.F < 200

    14 NEW24 0.35 < SG < 0.688, -50 < Deg.F < 200

    Table A is for Crude, the Table B is for refined products, the Table C is for special products - butadiene, toluene. OLD Tables are used for LPG and NGLs. Crude oil, natural gasoline, drip gasoline JP4 use 6A,24A if density is known, and 5A/6A,23A/24A if live densitometer is used. Gasoline, naphthalene, jet fuel, aviation fuel, kerosene, diesel, heating oil, furnace oil use 5B/6B,23B/24B when live density is used. 6B,24B when no live density is used. Benzene, toluene, styrene, orthoxylene, metaxylene, acetone use 6C/24C.

    Gravity

    Enter zero to use live gravity.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-6

    Date: 3/10/2021

    PROVE DATA

    Prove Report Number

    The prover report number will appear in the prover report.

    Prover Serial Number

    The prover serial number will appear in the prover report.

    Prover Size

    Enter the size of prover.

    Prover Type

    Selection Description

    0 Ballistic type prover with piston action

    1 Reduced type prover uni-directional with total pulse count less than 10,000

    2 Reduced volume bi-directional with less than 10,000 pulses in each direction

    All the above will use the double chronometry method (pulse interpolation)

    Selection Description

    3 Uni-directional big volume prover with pulse count over 10,000 counts

    4 Bi-directional large volume prover with pulse count in each direction exceeding 10,000 in each direction

    5 Master Meter Proving

    6 Smith ballistic type prover with piston action

    7 Calibron with Single Detector

    Prove Method

    Selection US Unit

    0 Volume - Barrel

    1 Mass - LB

    Number of Prove Runs to Average

    Numbers 1-10 are accepted to achieve good meter factor and pulse deviation limit. After the number of consecutive prove runs to average is achieved within the pulse deviation limit, the prove runs is complete

    Number of Total Runs

    Numbers 1-20 are the total number of allowable runs to achieve repeatability. The Flow Computer will attempt to achieve consecutive runs within the repeatability limit.

    Prove Abort Time Out in Seconds

    If single prove run was not complete in the set time, then the sequence of prove will be aborted. The reason for this function is to accommodate the possibility the second detector switch was not triggered for one reason or another, or the proving was launched from the wrong direction.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-7

    Date: 3/10/2021

    Repeatability - Pulse Deviation (Repeatability)

    The pulse deviation limit is for setting the repeatability of runs. The Flow Computer will try to achieve consecutive runs within the programmed limits. If master meter proving is selected Ratio = Pay Meter Counts / Master Meter Counts Pulse Deviation % = Highest of Ratio - Lowest of Ratio x 100 Otherwise Pulse Deviation % = Highest of Counts - Lowest of Counts x 100 Lowest of Counts

    Meter/Prove Volume Resolution

    Enter 0 to set 5 decimal places or enter 1 to set 6 decimal places for the meter and prove volume resolution.

    Master Meter K Factor

    The number of pulses per unit volume (i.e. 100 pulses/barrel in volume proving or /lb in mass proving for US unit, pulses/M3 in volume proving or /KG in mass proving)

    Prover Run Counts

    Enter prover run counts if master meter proving is configured. The flow computer will use the master meter pulse counts to constitute one run.

    Detector Switch

    Provers are always fitted with dual detector switches. Because of wiring considerations, many times both detectors are paralleled together. Select single detector, when both detectors are paralleled together. Select dual detector, when two isolated detector signals are used. Prover operation will not with the wrong prover controller chip. 333 chip is our most common prover controller chip. It uses a single detector switch. It is normal operation is for Active Low switches commonly found in Piston SVPs like Calibron and Brooks. It also supports Active High switches (sometimes called Reverse Polarity detector switch). 331 chip is generally used for Ball provers with dual detector switches like most traditional prover controllers, one switch for start and one to stop. It also supports Active High single detector provers.

    Single Detector Delay

    When single detector is used, hysteresis time is required to insure proper de-bounce time. 100 milliseconds are sufficient in most cases.

    Prover Volume

    The prover volume in barrels(us unit) at reference conditions per water draw.

    Percentage of pre-travel in Volume

    After the second detector switch is tripped. The Flow Computer allows for time delay period proportional to time between prover run first and second detector switch. After the time expires, the launch sequence will re-start.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-8

    Date: 3/10/2021

    Prover Diameter

    Enter the diameter of the prover in inches

    Prover Wall Thickness

    The thickness of prover wall in inches assuming single wall prover.

    Modulus of Elasticity E+7

    The elasticity coefficient of prover wall material.

    Prover Temperature Sample Period

    The prover temperature should be stable before proving. This would indicate the meter and the fluid through the prover has similar temperatures. The sample period configured is a command to the Flow Computer to sample the prover temperature every so many minutes and compares the temperature to the previous sample.

    Prover Temperature Allowable Change in Deg.F(US) or in Deg.C(Metric)

    Each time the prover temperature is sampled and compared to the previous sample. The deviation should not exceed the set limit. Once the prover temperature is stable the Flow Computer will give the command to initiate a prove.

    Flow Rate Change per Sample Period

    The maximum flow rate fluctuation acceptable before initiating a prove. The temperature sample period is used to monitor flow rate.

    Prover and Meter Temperature Deviation

    This will set the limit between the meter and prover temperature. If the difference between the meter and prover temperature is above the set limit the prove will be aborted

    Prover Base Temperature

    Enter the temperature (Deg.C in metric unit or Deg.F in US unit) in which the prover volume was established. All steel correction factors will be corrected to base temperature.

    Prover Base Pressure

    The prover pressure correction factors will be based on the reference calibration pressure. The modulus of elasticity combined with the live pressure is used to calculate wall expansion due to pressure.

    Coefficient of Expansion on Displacer Shaft E-7 (Piston Prover)

    This is the coefficient of thermal expansion of the prover piston shaft

    mild carbon 62.0 e-07/Deg.F 1.12 e-05/Deg.C

    316 stainless 88.3e-7/Deg.F 1.59e-5/Deg.C

    304 stainless 96.0e-7/Deg.F 1.73e-5/Deg.C

    17-4ph stainless 60.0e-7/Deg.F 1.08e-5/Deg.C

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-9

    Date: 3/10/2021

    Area Thermal Coefficient E-7 (Piston Prover)

    This is the coefficient of expansion for prover wall

    mild carbon 62.0 e-07/Deg.F 1.12 e-05/Deg.C

    316 stainless 88.3e-7/Deg.F 1.59e-5/Deg.C

    304 stainless 96.0e-7/Deg.F 1.73e-5/Deg.C

    17-4ph stainless 60.0e-7/Deg.F 1.08e-5/Deg.C

    Upstream Signal Polarity

    The upstream signal polarity signals high or low is the ready to launch signal. The logic polarity of the piston upstream can be programmed as low or high to indicate ready to launch.

    Run Output Signal Polarity

    When using ballistic provers, a signal is required to start prover, or bring the piston to the launch position.

    Detector Mounting Section

    When piston type prover is selected, the detector switches distance will change with temperature. For ballistic piston type prover, the user must select "on calibration section mounting". Pipe provers are "off calibration section mounting".

    Coefficient of Cubical Expansion E-7

    mild carbon 186.0 e-7/Deg.F 3.35 e-5/Deg.C

    316 stainless 265.0 e-7/Deg.F 4.77 e-5//Deg.C

    304 stainless 288.0 e-7/Deg.F 5.18 e-5/Deg.C

    17-4ph stainless 180.0 e-7/Deg.F 3.24 e-5/Deg.C

    Prove Seal

    When prover seal signal is not available., the option for requiring prove ready signal is waived.

    Master Meter Flow Rate/K Factor

    Enter master meter k factor. K Factor is the number of pulses per unit volume. Enter the different correction factors for the meter at different flow rate. The master meter will use interpolated K factor for volume calculation.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-10

    Date: 3/10/2021

    INPUT ASSIGNMENTS

    The Flow Computer provides 4 analog inputs, 4 status input, 5 switch outputs, one density frequency input, two turbine inputs, and two 4 wire RTD inputs. In order for the Flow Computer to read the live input, the input must be properly assigned and properly wired.

    Use RTD 5/6 for Analog Inputs 5/6

    The ‘Flow Computer’ can be configured as either ‘4 Analog and 2 RTD Inputs’ or ‘6 Analog Inputs’. (For

    version SFC332P-US 1.17.0 and higher).

    Enter ‘1’ to use ‘6 Analog Inputs’.

    Meter Temperature Assignment

    Selection Description

    0 Not Used

    1 Analog Input #1

    2 Analog Input #2

    3 Analog Input #3

    4 Analog Input #4

    5 RTD#1/Analog#5

    6 RTD#2/Analog#6

    10 Multi.Variable

    Meter Pressure Assignment

    Selection Description

    0 Not Used

    1 Analog Input #1

    2 Analog Input #2

    3 Analog Input #3

    4 Analog Input #4

    5 Analog#5

    6 Analog#6

    10 Multi.Variable

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-11

    Date: 3/10/2021

    Prover Temperature Assignment

    Selection Description

    0 Not Used

    1 Analog Input #1

    2 Analog Input #2

    3 Analog Input #3

    4 Analog Input #4

    5 RTD#1/Analog#5

    6 RTD#2/Analog#6

    Use Meter Temperature 1=Yes

    If there is no prover temperature mounted on the prover, assign the meter temperature to be used for prover corrections.

    Prover Temperature Transducer 0=Single,1=Dual

    Inlet or outlet temperature transmitters are accepted. Inlet is used in forward direction and outlet in the reverse direction.

    Shaft Temperature Assignment

    Selection Description

    0 Not Used

    1 Analog Input #1

    2 Analog Input #2

    3 Analog Input #3

    4 Analog Input #4

    5 RTD#1/Analog#5

    6 RTD#2/Analog#6

    Prover Pressure Assignment

    Selection Description

    0 Not Used

    1 Analog Input #1

    2 Analog Input #2

    3 Analog Input #3

    4 Analog Input #4

    5 Analog#5

    6 Analog#6

    10 Multi.Variable

    Use Meter Pressure 1=Yes

    If there is no prover pressure mounted on the prover, assign the meter pressure to be used for prover corrections.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-12

    Date: 3/10/2021

    Spare Assignment

    Spare input is used for display and alarms purpose only. It is not used in the calculation process. To read spare input value, use the diagnostic screen.

    Spare Resolution

    The resolution is used for calibrations and display only.

    ANALOG INPUT

    4mA

    Enter the 4mA value for the transmitter.

    20mA

    Enter the 20mA value for the transmitter.

    Low/High Limit

    Enter the low and high limits. When live value exceeds high limit or less than low limit, an alarm log will

    be generated.

    Maintenance Value

    The value to be used when the transmitter fails, or while calibrating. For calibration, set fail code to 1 while

    calibrating.

    Fail Code

    Fail Code 0: always use the live value even if the transmitter failed.

    Fail Code 1: always use the maintenance value

    Fail Code 2: use maintenance value if transmitter failed. I.e. 4-20mA is above 21.75 or below

    3.25)

    Note: Multi-variable default value is used when transmitter connection fails.

    RTD INPUT

    Low/High Limit

    Enter the low and high limits. When live value exceeds high limit or less than low limit, an alarm log will

    be generated.

    Maintenance Value

    The value to be used when the transmitter fails, or while calibrating. For calibration, set fail code to 1 while

    calibrating.

    Fail Code

    Fail Code 0: always use the live value even if the transmitter failed.

    Fail Code 1: always use the maintenance value

    Fail Code 2: use maintenance value if transmitter failed

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-13

    Date: 3/10/2021

    DENSITY TYPE

    If live density is connected to the meter, user must enter the density type. Raw density frequency or 4-20mA can be selected.

    Type Density

    0 None

    1 4-20mA

    2 UGC

    3 Sarasota

    4 Solartron

    Density 4-20mA Type

    Type Description

    0 Specific Gravity 4-20mA.

    1 API Gravity 4-20mA

    2 Density Signal 4-20mA in GM/CC

    Use Meter Temperature as Density Temperature

    To allow the meter temperature to calculate the effect of temperature on the densitometer. Make sure the meter and density temperature are similar to avoid measurement errors.

    Use Meter Pressure as Density Pressure

    This feature allows the user to assign the meter pressure to calculate the effect of pressure on the densitometer. Take caution and make sure the pressure is very similar on the meter and densitometer to avoid measurement errors.

    Sarasota, UGC, or Solartron Constants

    Enter the densitometer constants accordingly with the type selection.

    Density Correction Factor

    Enter the correction factor for the densitometer.

    Density Period Low/High Limits

    Density Period is the time period in microsecond. The densitometer fails if the density period exceeds the density period low or high limits. If the densitometer fails and density fail code is set to 2, the maintenance value will be used.

    TRANSDUCER TAG ID

    Up to 8 alphanumeric ID number. The transmitters are referred to according to the TAG ID. All alarms are labeled according to TAG ID.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-14

    Date: 3/10/2021

    ANALOG OUTPUT ASSIGNMENT

    4-20mA selection must be proportional and within the range of the selected parameter. The 4-20mA signal is 12 bits.

    Assignment Description

    1 IV Flow Rate

    2 Meter Temperature

    3 Meter Pressure

    4 Prover Temperature

    5 Prover Pressure

    6 Shaft Temperature

    7 Left Temperature

    8 Right Temperature

    9 Density

    10 Densitometer Temperature

    11 Densitometer Pressure

    12 API

    13 API@60

    14 SG

    15 SG@60

    16 Spare#1

    17 Spare#2

    18 Remote Control

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-15

    Date: 3/10/2021

    STATUS INPUT ASSIGNMENT

    User can select any one of status input and assign it to input point.

    Assign Description

    1 Dialog Scroll

    2 Dialog Select - to select and change some parameters without lap-top or PC

    3 N/A

    4 Prove Request (Single Prove Run)

    5 Prove Request (a complete prove sequence)

    6 When status is "ON", the display will stop scrolling to allow for continuous monitoring.

    7 the display will scroll as the user toggle the status

    8 N/A

    9 Prover Ready (High) / Upstream Polarity

    SWITCH OUTPUT ASSIGNMENT

    The Flow Computer switch outputs are open collector type and require external D.C power.

    Assignment - Contact Type Outputs

    Assignment Description

    1 Prove meter

    2 Prove Launch Forward

    3 Prove Launch Reverse

    4 Prove Launch Forward/Reverse

    5 Prove in Progress

    6 Compact Prove Run

    7 Prove Complete

    8 Prove Abort

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-16

    Date: 3/10/2021

    FLOW COMPUTER D ISPLAY ASSIGNMENT

    Display assignment selections are up to 16 assignments. The Flow Computer will scroll through them at the assigned delay time.

    Assignment Description

    1 IV Flow Rate

    2 Meter Temperature

    3 Meter Pressure

    4 Prover Temperature

    5 Prover Pressure

    6 Prover Shaft Temperature

    7 Prover Left Temperature

    8 Prover Right Temperature

    9 Density

    10 Densitometer Temperature

    11 Densitometer Pressure

    12 API

    13 API@60

    14 SG

    15 SG@60

    16 Densitometer Period

    17 Un-Corrected Density

    18 Density Frequency

    19 Prove Mode

    20 Next Prove Meter ID

    21 Next Prove Request

    22 Spare Data

    23 Date/Time

    MODBUS SHIFT

    To assign variable 4 bytes, 2 bytes and scaled variables for easy polling and convenience. The Flow Computer will repeat the assigned variables into the selected locations. Example : 3801=4201. Register size compatibility could cause rejection for certain address assignments. Valid two bytes modbus address are from 2534 to 2732. Valid four bytes registers are from 3131 to 4821.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-17

    Date: 3/10/2021

    Security

    SECURITY CODES

    The desktop application provides 4 security areas to prevent users from entering data into certain areas. The

    four areas are:

    Configuration: Allow user to modify device configuration settings.

    Override: Allow user to change values directly on the device.

    Calibration: Let the user to calibrate the device inputs.

    Image File Download: Let user download an image file to the device. This procedure will erase all the

    information and configuration stored in the device.

    Master Access: Once the master access is granted, the user can access to all four areas.

    Use the Tools|Security Codes option to modify the access code; a form will appear showing the five

    different security areas and the actual access status (at bottom of the form). To put a new access code log in

    to the desired security area and press Change security Code, type in the code and retype it on the confirm

    space to avoid mistyped codes. Then click [OK].

    The system will update the security access every time the application connects to the device and every time

    data is written to the device it will check for security access before writing.

    NOTE: In case the access code is forgotten contact our offices for a reset code.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-18

    Date: 3/10/2021

    Connect to Device

    Click to establish the communication. If the communication is failed, check information in the “Communication Port Settings”.

    Go Offline

    Click to disconnect the communication.

    Settings

    REPORT TEMPLATES

    Templates are HTML files that are used to generate the reports. They can be modified with any HTML

    editor. Specify the new location if you want to use the formatted report.

    Calibration Calibrations are performed under Calibration. . Select inputs to be calibrated, and then select full,

    single, offset calibration method. (See details in chapter 1)

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-19

    Date: 3/10/2021

    Parameter Overrides:

    Prover Request Enter the meter number to prove and then enter prove sequence type. Single run sequence is one run one way in uni-directional prover and round trip in bi-directional. Sequence will be a sequence of run to achieve repeatability and meter factor.

    Prover/Master Temperature Override When there is no live prover/master temperature connected to the flow computer, user can enter override value to be used in the proving calculations, or use the meter temperature.

    Prover/Master Pressure Override When there is no live prover/master pressure connected to the flow computer, user can enter override value to be used in the proving calculations, or use the meter pressure.

    Prover Shaft Temperature Override The shaft temperature override is used in piston type provers. The temperature for the shaft is normally the ambient temperature.

    Prover Gravity Override Prover gravity override will be used for liquid corrections for prover.

    Meter Temperature Override This value is entered when no live temperature is available, or different value from live value should be used.

    Meter Pressure Override Pressure override can be used when no live pressure transmitter is connected to the Flow Computer.

    Equilibrium Pressure Override Enter the override value in PSIG. Zero in the data entry indicates using calculated value.

    Alpha T E-6 Override Enter Alpha T Override. Alpha T is the thermal expansion coefficient for the Table 6c or Table 24C. Enter number assuming that is divided by 1000000.(Example : 0.000355 = 355 / 1000000)

    CTL or CPL Override Enter the override value, the proving meter temperature and pressure correction will default to the override value. Zero in the data entry indicates using calculated value.

    Meter Cumulative Total Override Enter meter cumulative total before starting prove. The Meter Cumulative Total appears in the report.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-20

    Date: 3/10/2021

    Reset Prove Data Area Enter meter number to reset previous prove data area .

    Date and Time Change the date and time of the flow computer.

    Clear System

    Enter reset system code to reset all data.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-21

    Date: 3/10/2021

    HISTORICAL DATA

    CAPTURE AND STORE

    To retrieve historical data, go to Historical Data menu. Check the report to retrieve and then click “Get

    Report” option. The Get Report option retrieves the information from the flow computer, shows it on the

    screen and stores it on the database.

    The valid data entries are shown at the bottom of the dialog. The available types of reports are:

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-22

    Date: 3/10/2021

    PREVIOUS PROVE REPORT

    A previous proves for each meter can be stored and retrieve at any time.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-23

    Date: 3/10/2021

    S INGLE PROVE REPORT

    The previous single prove run report.

    PROVE ABORT REPORT

    The previous prove abort report.

    Check HTML FILE box to generate additional historical reports in HTML format.

    Every time a report is generated it is saved in our proprietary DFM format. This file can be only viewed using our Dynacom software. In order to exchange information with other systems different that Dynacom, we have provided additional report formats. As follows: HTML Report: HTML is a widely used document format that can be viewed with endless number of software applications, among them are Internet Explorer, Microsoft Excel, Microsoft Word, etc. This format allows almost any user to view and print our reports without having to use our Dynacom Application.

    View Report in Modubs Address Format Toggle “Show Modbus Address” button to display report in modbus address or value format.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-24

    Date: 3/10/2021

    Viewing previously captured reports

    Once a report is stored in the database using the Historical Data|Open Saved Report option to view the

    Previously Captured Reports.

    When the option is selected, a dialog will appear asking for the name of the report you want to see. The

    browse button can be used to locate the list of reports stored in the disk.

    Printing Reports

    The Print Button (shown on the picture) lets you print the report to any printer installed in your computer. The printed version will look just like it is shown on the screen.

    The Save Report Button allows the user to save the report into a file.

  • Dynamic Flow Computers SFC332P Prover Manual Data Entry — 2-25

    Date: 3/10/2021

    MASTER PROVING REPORT (MASS PROVER )

  • Dynamic Flow Com