shimadzu atomic absorption aa-7000
TRANSCRIPT
AA-7000 Series
ShimadzuAtomic Absorpt ion Spectrophotometers
C122-E057
Reaching even greater heights Unprecedented sensitivity, thanks to a newly developed 3D optical system
Advanced safety technology
System configuration evolving to suit the needs of the user
High Performance
The newly developed 3D optical system is capable of flame: Pb 0.015 ppm detection, and furnace: Pb 0.00005 ppm detection.
Equipped with the world’s first*1 vibration sensor, a multimode automatic gas leak check function, and other advanced safety technology.
Evolving from the basic flame model to suit the needs of the user. The dual atomizer system*2 is the world’s smallest*1, and a furnace model is also available.
Easy-to-understand, user-friendly WizAArd software supports FDA 21 CFR Part 11 compliance. QA/QC functionality and hardware validation software are also provided.
*1 Survey by Shimadzu in November 2008.*2 The optional auto atomizer changer is required.
Reaching even greater heights Unprecedented sensitivity, thanks to a newly developed 3D optical system
Advanced safety technology
System configuration evolving to suit the needs of the user
High Performance
The newly developed 3D optical system is capable of flame: Pb 0.015 ppm detection, and furnace: Pb 0.00005 ppm detection.
Equipped with the world’s first*1 vibration sensor, a multimode automatic gas leak check function, and other advanced safety technology.
Evolving from the basic flame model to suit the needs of the user.The dual atomizer system*2 is the world’s smallest*1, and a furnacemodel is also available.
Easy-to-understand, user-friendly WizAArd software supports FDA 21 CFR Part 11 compliance. QA/QC functionality and hardware validation software are also provided.
*1 Survey by Shimadzu in November 2008.*2 The optional auto atomizer changer is required.
FLAME MODELHighly stable, enhanced flame model
FLAME MODELHighly stable, enhanced flame model
And it doesn’t stop there...
FURNACE MODELCutting-edge furnace model, achieving unprecedented sensitivity
And beyond…
DUAL ATOMIZER SYSTEMDual atomizer system, compatible with a variety of analyses
Basics
Optics
Wavelength range
Bandwidth
Lamp mode
Background correction
Number of HC lamps
Atomizer position adjustment
Concentration computation
Repeat analysis
Analog output
Data processing
Re-analysis
Electronic Records
Ambient temperature / humidity
Flame (continuous mode / micro sampling method*1)Burner head Titanium 10 cm slot (for Air-C2H2 flame)
Nebulizer Pt-lr capillary with PTFE orifice, ceramic impact bead (capable of handling hydrofluoric acid)
Chamber Engineering plastics
Flow rate control Automatic fuel gas flow rate setting, automatic search of optimum gas flow rate
Safety measures Automatic flame extinction via flame vibration sensor, automatic gas leak check, etc.
FurnaceHeating temperature range Ambient to 3,000 °C
Heating control system
Setting heating conditions
Power requirements Choose from 200 VAC, 220 VAC, 230 VAC or 240 VAC. ±5%, 6 kVA, 50/60 Hz combined
Dimensions and weight (power supply unit)
260 W × 555 D × 527 H mm, 46 kg
Positioning Lateral/vertical manual adjustment,with option mounted,automatic atomizer switching / automatic search of optimum burner height
Flame continuous method, flame micro sampling method *1, furnace method
Optical double-beam, automatic switching to single beam during furnace measurementAberration-corrected Czerny-Turner spectrometer, photomultiplier tube
185.0 to 900.0 nm, automatic wavelength setting via line search
0.2, 0.7, 1.3, 2.0 nm (4-step automatic switching)
BGC-SR (high-speed self-reversal method) and BGC- D2 (D2 lamp method)
6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement)
Manual (standard), automatic (with option mounted)
Calibration curve method, standard addition method, simple standard addition method
Up to 20 repetitions, automatic exclusion of deviant values by setting SD and %RSD
2 channels (atomic absorption/energy signal, background signal)Output range: Both 4-level switching *2
Calculation of final concentrations, QA/QC function
Automatic dilution and re-analysis of unknown samples (flame micro sampling method, furnace method)
Compatible with FDA 21 CFR Part 11 Compliance
Choose from 100 VAC, 120 VAC, 220 VAC, or 230 VAC. ±10%, 230 VA, 50/60 Hz combined
AA-7000F: 700 W × 588 D × 714 H mm, 72 kgAA-7000G: 700 W × 580 D × 538 H mm, 65 kg (excluding protrusion and options)
10 to 35 °C, 20 to 80% (less than 70% when temperature is higher than 30 °C)
Power requirements *3
Dimensions and weight *4
Printed in Japan 3295-01910-30ANS
SHIMADZU CORPORATION. International Marketing Division3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax. 81(3)3219-5710URL http://www.shimadzu.com
870
588
280
714
390
960
340
Dimensions (Dual Atomizer System)
Maximum 20 stages Inner gas type: Dual automatic switching typeHigh-sensitivity mode setting Enrichment in furnace: Maximum 20 timesOptimum temperature program search support function
Drying: Digital current control with automatic temperature calibration functionAshing, Atomization: Digital temperature control via optical sensor
EMISSION, NON-BGC, BGC-D2, BGC-SR
Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. We maintain a global network of sales, service, technical support and applications centers on six continents, and have established long-term relationships with a host of highly trained distributors located in over 100 countries. For information about Shimadzu, and to contact your local office, please visit our Web site at www.shimadzu.com
Personal Computer / MonitorOperating system Microsoft Windows Vista Business or XP Professional
CPU
RAM 1 GB or higher (Vista) or 512 MB or higher (XP)
Monitor XGA (1024 × 768 dots) or higher
Storage device One CD-ROM drive (for installing software)Requires 60 MB min. hard disk space for installation
I/O port One serial port (for AA control)
Intel® Celeron 420 (1.60 GHz) or higher
*1 Optional autosampler and micro sampling kit are required for the flame micro sampling method.*2 1V F.S. fixed in EMISSION mode.*3 Power is required separately for the personal computer.*4 Protruding parts and optional equipment are not included.*Windows is a registered trademark of Microsoft Corp., U.S.A.